The SPL Guard Arizona Thawing Station is a solution for the controlled thawing of samples in a
single rack. The thawing station can only be used as a standalone instrument.
COMPONENTS
SPL Guard Arizona Thawing Station 1
12 AC/DC adaptor 1
GETTING STARTED
With the SPL Guard Arizona
INSTALLATION SETUP
• Step 1: Remove the SPL Guard Arizona
Thawing Station from the packaging.
• Step 2: Plug the adaptor into the DC-plug on
the back of the unit.
• Step 3: Connect the adaptor plug to an
electrical outlet.
• Step 4: Position the unit with the display side
facing out.
WARNING: Do not put your hand into the machine
when the ventilators are rotating!
OPERATION INSTRUCTION
Step 1: Place a rack in the rack-bay. Ensure that
at least one tube is located at the IR-sensor po-
sition, read “sensor location per rack format“ for
more information.
• Step 2: Press the blue button on the
front-panel. The machine will now start
thawing.
• Step 3: When the tubes reach their desired
temperature, the machine will turn off and a
beeping sound will be made. Press the blue
button again to conrm that the rack can be
removed.
• Step 4. Remove the rack from the rack-bay.
SENSOR LOCATION PER RACK
FORMAT
The SPL Guard Arizona uses an Infra-Red Sensor
to determine the temperature of the tubes.
It is important to at least have one tube at the
pre-scripted tube position when turning on the
machine.
The pre-scripted positions are as following (arrow
is indicating the front side of the machine)*:
1. For 96-well format
2. For 48-well format
3. For 24-well format
*NOTE: For accuracy purposes, it can be wise to
place one tube next to the marked tube location,
in each direction.