the meter by yourself, the method is follow:
Power on the meter, smear a little of coupling paste
on the surface of the metal block buit-in the shell of the
meter. Put the probe on and press it, when coupling
symbol appears, press button “Adjust” for two seconds,
four bars appear on the screen. Keep to press the probe,
button “Adjust” can release. You can see the bars
disappear gradually.Finally “4.0 mm” displays, the meter
will be calibrated.
Ⅷ.Notices
1. The ultrasonic wave velocity listed in Table(1) is for
reference. Because the component of your material may
not be identical with that listed in Table(1), had better
make sample block to test the wave velocity in it.
2. Poriferous matter, or uncompacted material isn’t a good
ultrasonic wave conductor, can not be measured by the
meter. Casting has large crystals, if its surface is rough,
it is difficult for us to test thickness by the meter.
3. When measuring thickness under 2.5mm, sometimes an
error named as “double refraction” may occur. Another
error is named as “ pulse envelope or cycle jump”. They