6. Position and engage threads of end cap into the housing by hand until snug.
7. Tighten end cap with the combination end cap wrench, employing the wrench
opening marked AC as shown. Position wrench with marked side facing end cap
for best wrench control.
Housing
Autochuck Turbine
Wrench
Opening
Note: Replace the turbine when the handpiece fails to achieve full speed and/or
increases sharply in noise level. The replacement cartridge is available from your
dealer and is easily replaced in the ofce. To order a replacement, see 'Replacement
Parts' section for StarDental part numbers.
Turbine Installation
1. Using wrench supplied, remove old end cap and turbine.
2. Inspect housing before cleaning. Remove any obvious debris to prevent damage
and to insure performance of the new turbine. Also, remove o-ring(s) from housing
at bur opening, if present.
3. Connect handpiece and run air through turbine housing without turbine installed
for 5 seconds.
4. Clean housing thoroughly with a cotton swab and isopropyl alcohol. Dry with syringe
air.
CAUTION : Use burs that have an overall length of less than 24 mm and shank
length of at least 7.3 mm.
Autochuck Operation
Autochuck Bur Changing System
1. Insert bur into autochuck until resistance is felt.
2. Depress autochuck end cap button and continue to feed-in bur until it clicks and
stops.
3. Release end cap button. Now, apply light inward force to the dental bur. This will
help set the chucking mechanism for initial start up. (Always tug on bur after it has
been installed to conrm that it is secured.)
1. Always tug on bur after chuck is loaded to conrm that the bur is secured.
2. Before use, verify concentric rotation by operating the handpiece outside the oral
cavity in a safe place. Do not use burs that do not run concentrically or produce
excessive noise.
3. Never operate handpiece without a bur installed.
4. Always use burs with an overall length of less than 24 mm. Shank length must be
at least 7.3 mm.
5. Use only recommended burs to prevent damage to autochuck and in some cases to
prevent burs from improper or incorrect placement. An improperly clamped bur may
dislodge during use, causing injury to you or your patient. A sudden, high-pitched
sound or increased vibration while handpiece is operating alerts the operator to the
possibility that the chuck is not clamped sufciently to retain the bur or a bent bur
is inserted.
6. When using the autochuck, never depress end cap while bur is rotating.
7. For maximum life, never operate handpiece above 45 psi.
CAUTION : Never depress the autochuck end cap button or rest the end cap
against any part of the oral anatomy during handpiece operation. Depressing the end
cap while a bur is rotating can activate the chuck mechanism and result in end cap
heat-up and/or loss of bur retention.
4. To remove bur, grasp handpiece rmly so your thumb touches the end cap and push
until a slight ‘click’ is noticed.
5. Pull bur straight out.
6. Release endcap.
Replace the turbine when the handpiece fails to achieve full speed and/or increases
sharply in noise level. This is an indication of a failing turbine. A failing turbine will reduce
the efency of the handpiece and may result in end cap heat up. See "Turbine Installation".
Always insert a bur the full length of the chuck. If you attempt to extend it from the
chuck, the bur may dislodge during use and cause injury to you or your patient.
DO NOT USE FOR ORAL SURGERY
5. Place new turbine assembly into housing and gently insert until fully seated.
Clean debris from
inside of housing
Cleaning
Instructions
Note After each patient, operate handpiece for 20 seconds with air
and water spray activated.
Cleaning
External
Surfaces
To remove debris from external surfaces, clean using a small
toothbrush with isopropyl alcohol. Dry thoroughly.
SAFETY
INSTRUCTIONS
WARNING
CAUTION
StarDental® Instrument Solutions
065720 D Date of Issue 2016-05-02
StarDental®
1816 Colonial Village Lane
Lancaster, PA 17601 U.S.A.
Printed in the U.S.A.
©2016 DentalEZ, Inc.
DentalEZ Dental Products (GB) Ltd.,
Cleveland Way,
Hemel Hempstead, Hertfordshire
HP2 7DY, ENGLAND
EC REP
Caution: Federal law restricts this device to
sale by or on the order of a dentist.
™
6
-05-
0
s
(
GB
)
e
land
r
t
f
ord
s
E
N
G
L
0
2
D
entalEZ Dental Product
s
C
lev
e
Hemel Hempstead, He
r
HP2 7DY,
E
866.DTE.INFO (383.4636)
www.DentalEZ.com
Autochuck Turbine
Replacement Kit
Instruction Manual
BEFORE INSTALLING THE TURBINE, CAREFULLY READ AND FOLLOW ALL OF
THESE INSTRUCTIONS. SAVE THESE INSTRUCTIONS FOR FUTURE REFERENCE
AND OBSERVE ALL WARNINGS AND CAUTIONS.

Always insist on StarDental parts for best performance.
Description Item Number
LubeFree Ceramic Autochuck Turbine Replacement Kit
Single ............................................................................................................................................... 263739
Four Pack ........................................................................................................................................ 264184
E6 LubeFree Ceramic Autochuck Turbine Replacement Kit
Single ...............................................................................................................................................265807
Lubricated Ceramic Autochuck Turbine Replacement Kit
Single ...............................................................................................................................................263968
Four Pack ........................................................................................................................................263453
Steel Bearing Lubricated Autochuck Turbine Replacement Kit
Single ...............................................................................................................................................260296
Four Pack ........................................................................................................................................260295
LubeFree Ceramic Autochuck 430 Torque Turbine Replacement Kit
Single ............................................................................................................................................... 265877
Lubricated Ceramic Autochuck 430 Torque Turbine Replacement Kit
Single ...............................................................................................................................................265885
End Cap Wrench .........................................................................................................................203048
DentaLube II .................................................................................................................................262539
Lube Tip for use with DentaLube II .....................................................................................202954
Junk-Out High-Speed Chuck Cleaner ................................................................................. 263790
Note: Always use ISO burs to prevent locking and slipping of burs in autochuck. If
problems of locking and slipping continue, replace worn autochuck turbine
immediately - see 'Replacement Parts' section for StarDental part numbers.
If any problem persists, contact dealer for further assistance; or, contact StarDental
Technical Service at 1-866-DTE-INFO.
For maximum turbine life do not exceed recommended pressure.
LubeFree Ceramic Bearing Autochuck Turbine
- This turbine uses a sophisticated ceramic bearing design and, therefore, requires
no lubrication. To ensure the life of this turbine, do not lubricate it. If it gets lubricated
even once, it MUST then continue to be lubricated for the rest of its life.
Lubricated Ceramic and Steel Bearing Autochuck Turbine
Note: Lubricate handpiece with DentaLube II, StarDental® Part #262539,prior to each
sterilization procedure.
1. Conrm that the bur is in the chuck. (Never operate handpiece without a bur in place.)
2. After each patient, operate handpiece for 20 seconds with air and water spray
activated.
3. Disconnect handpiece from swivel (or directly from tubing if swivel is not being used).
4. Assemble conical red tip applicator, StarDental® Part #202954, into DentaLube® II
bottle cap.
5. Place three (3) drops of DentaLube II, StarDental® Part #262539, into drive air at rear
of handpiece drive air slot or drive air tube as shown.
Lubricating
6. Reconnect handpiece to tubing or to swivel and operate for 20 seconds to expel
excess lube.
7. Remove bur from chuck prior to sterilization.
Troubleshooting
Problem: LACK OF POWER
Cause: Worn out turbine
Correction: Refer to Turbine Installation.
Problem: BUR LOCKS IN AUTOCHUCK
Cause: Debris buildup in chucking mechanism
Correction: Clean with Junk-Out Chuck Cleaner P/N 263790
Cause: Use of non-ISO burs
Correction: Always use ISO burs less than 24 mm overall length and a
shank length of at least 7.3 mm.
Problem: BUR SLIPS IN AUTOCHUCK
Cause: Use of non-ISO burs
Correction: Always use ISO burs less than 24 mm overall length and a
shank length of at least 7.3 mm.
Cause: Bur not properly installed
Correction: Remove bur from autochuck per 'Autochuck Operation' section's
steps 4, 5 and 6. Then, re-install bur per 'Autochuck Operation'
section's steps 1, 2 and 3.
Cause: Debris build-up in chucking mechanism
Correction: Clean with Junk-Out Chuck Cleaner P/N 263790
DO NOT USE OIL.
Replacement Part Numbers
StarDental®
Instrument Solutions
Red Tip
Applicator
Drive
Air
Drive Air
Swivel
Back End
Fixed
Back End
For new products, manuals and technical information, call
1-866-DTE-INFO or visit our web site at www.dentalez.com.
2
3
1
Limited Warranty
O-Rings and Seals are specifically excluded from this warranty.
DentalEZ® warrants the Autochuck Turbine to be free of defects in material and workmanship, under
normal usage, under the following terms:
StarDental® Products: Warranty Period**:
Replacement Turbines:
LubeFree Ceramic Bearing Autochuck Turbine 1 Year from date of insertion*
E6 LubeFree Ceramic Bearing Autochuck Turbine 6 Months from date of insertion*
Lubricated Ceramic Bearing Autochuck Turbine 2 Years from date of insertion*
Lubricated Steel Bearing Autochuck Turbine 1 Year from date of insertion*
Please note the following additional terms of our warranty and return policy:
• Warranties cover manufacturing defects only and do not cover defects resulting from abuse,
improper handling, cleaning, care or maintenance, normal wear and tear or non-observance of
operating, maintenance or installation instructions. Failure to use authorized parts or an authorized
repair facility voids this warranty.
• Liability is limited to repair or replacement of the defective product at our sole discretion. All
other liabilities, in particular liability for damages, including, without limitation, consequential
or incidental damages are excluded.
• THIS WARRANTY IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESSED OR IMPLIED,
INCLUDING ANY IMPLIED WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A
PARTICULAR PURPOSE. NO EMPLOYEE, REPRESENTATIVE OR DEALER IS AUTHORIZED TO
CHANGE THIS WARRANTY IN ANY WAY OR TO GRANT ANY OTHER WARRANTY.
* The warranty on a replacement turbine begins on the date of insertion into your handpiece. The
turbine must be registered on that date. . Visit www.dentalez.com/warrantyreg. If turbine is not
registered, warranty will begin on date of purchase. Registration must take place within 12 months
of purchase.
** Provided conditions in warranty are met.
Note to ensure that your High-Speed chuck continues to be in
optimum working condition - clean once a week with Junk-Out
High-Speed Chuck Cleaner Kit P/N 263790:
Shake well before use
- Install brush to handle. See diagram at right.
- Use one new brush per cleaning
application.
- Do not reuse brushes.
- Remove bur from chuck.
- Dip brush into bottle.
- Remove excess cleaner from the brush.
- Insert brush into bur end of turbine.
- Clean gently by rotating and inserting the brush in and out of the
turbine.
- Repeat process if necessary.
Reseal cap on bottle immediately after use to reduce
evaporation.
WARRANTY REGISTRATION
PLEASE VISIT ONE OF THE WEBSITES BELOW TO REGISTER YOUR STARDENTAL PRODUCT. THIS WILL ENSURE
PROPER REGISTRATION AND PROMPT AND ACCURATE SERVICE SHOULD ANY OF YOUR PURCHASES NEED
EVALUATION OR REPAIR DURING THE WARRANTY PERIOD.
For Identa® Oral Cancer Screening System products: www.identa.net/warranty
For all Other StarDental® Products: www.dentalez.com/warranty
This manual suits for next models
8
Other StarDental Dental Equipment manuals
Popular Dental Equipment manuals by other brands

Wieland
Wieland AGC Micro Plus operating instructions

DentalEZ
DentalEZ Heritage CORE Chair user manual

KaVo
KaVo SMARTmatic PROPHY S31 Instructions for use

KaVo
KaVo ESTETICA E30 Short instructions for use

Dabi Atlante
Dabi Atlante PROFI NEO owner's manual

Durr Dental
Durr Dental VISTACAM CL 2106 Installation and operating instructions