430 SWL®and 430 SW™High-Speed Handpieces are designed to t a StarDental®360° quick disconnect swivel. The 430 K®handpiece is a conventional
non-swivel high speed handpiece. The 430 SWL connects to both StarDental and standard 4/6-Line ISO ber optic (F.O.) tubing. The 430 SW and 430 K
connect to standard 4 hole tubing.
• Specications
• Handpiece Installation
• Handpiece Operation
• Safety Instructions
• Lubrication - For Lubricated
Handpieces ONLY
• Cleaning & Sterilization
• Troubleshooting
• Turbine Replacement
Specications
430 SWL: Type: High speed ber optic with a swivel and replace-
able autochuck turbine (Swivel sold separately.)
Operating Air Pressure: 34 to 42 psi
Air Consumption: 1.9 SCFM @ 34 psi
No Load Speed: 430,000 rpm
Connection: 360° quick-disconnect swivel with 4-Line and
6-Pin ISO conguration and ber optics
Power: 14 watts
Weight: 49g
430 SW: Type: High speed with a swivel and replaceable
autochuck turbine (Swivel sold separately.)
Operating Air Pressure: 30 to 32 psi
Air Consumption: 1.9 SCFM @ 30 psi
No Load Speed: 430,000 rpm
Connection: 360° quick-disconnect swivel with 4-Line ISO
conguration or 2/3-Line ISO conguration
Power: 14 watts
Weight: 47g
Note: Filter air and water supplies, recommended ltration: 25 microns. Soften and deionize water. Be sure lters operate efciently.
Handpiece Installation
Note: This joint loosens over time, resulting in a loss of power. Check for tightness periodically.
Air and Water Adjustment
1. Air Pressure –
NOTICE To prevent premature damage to the turbine, do not exceed 34 psi when using a HiFlo swivel.
Use an air pressure gauge, StarDental Part #262591 for 4/6-Line: set pressure to:
34-42 psi - 430 SWL
30-32 psi - 430 SW
30-32 psi - 430 K.
If a gauge is not available, set pressure at delivery head allowing 1 psi pressure loss for every foot of tubing.
(See Specications for recommended operational pressure.)
Handpiece Operation
For example, if you are using the 430 K that you want to operate at 32 psi and you are using either 5 feet or 12 feet of tubing, then the pressure should
be set at the delivery head to compensate for the air pressure losses.
5 feet tubing 32 psi + 5 psi = 37 psi
or 12 feet tubing 32 psi + 12 psi = 44 psi
2. Water Pressure – Adjust water volume at the delivery head to obtain a satisfactory spray mist.
3. Non-Star 6-pin Fiber Optic System:
If the handpiece is connected to a non-StarDental 6-pin ber optic (F.O.) system, consult your operator’s manual of F.O. system for correct voltage.
Non-StarDental F.O. systems must have a rated amperage output between 1.5 Amps dc and greater than .73 Amps dc at 3.1 Vdc. Use StarDental Part #
262962 bulb for non-StarDental F.O. systems that meet these requirements. Adjust the voltage per instructions with the 3.1 Vdc bulb.
If the non-StarDental 6-pin F.O. system has a rated amperage output of at least 1.5 Amps dc at 3.3 VDC, then the StarDental Part #262522 bulb may be
used. Adjust the voltage per instructions with the 3.3 VDC bulb.
Use standard length concentric diamond and carbide burs with .0628" nominal diameter friction grip
shanks designed for high speed handpieces. Every StarDental Diamond Bur meets specications below
for recommended lengths. Contact your dealer for a complete list of StarDental Diamond Burs. StarDental
autochuck turbines have the following bur requirements:
Autochuck Bur Changing System
Maximum
24.0 mm
OVERALL LENGTH
Minimum
7.3 mm
Shank Length
Use only burs with hardened, tempered steel shanks or carbide shanks. Every StarDental Bur maintains
required hardness. Before use, verify concentric rotation by operating handpiece outside the oral cavity
in a safe place.
Note: Do not use burs that produce excessive noise or do not run concentrically. Do not use burs that are
modied, eccentrically designed, or have bent or ‘soft’ (non-heat treated) shanks.
Autochuck Bur Changing System
1. Insert bur into autochuck until resistance is felt.
2. Depress autochuck end cap button and continue to feed-in bur until it clicks and stops.
3. Release end cap button. Now, apply light inward force to the dental bur. This will help set the chucking mechanism for initial start up. (Always tug on bur
after it has been installed to conrm that it is secured).
CAUTION Never depresstheautochuck endcapbutton or rest the end cap against any part of the
oral anatomy during handpiece operation. Depressing the end cap while a bur is rotating can activate
the chuck mechanism and result in end cap heat-up and/or loss of bur retention.
1. To remove, grasp handpiece rmly so your thumb touches the end cap and push until a slight ‘click’
is noticed.
2. Pull bur straight out.
3. Release end cap.
430 K:
1. Slide tubing coupler until the end of the ISO connector is exposed; hold in hand as shown
with smaller ngers securing the tubing.
2. Align handpiece tubes into the ISO connector holes.
3. While applying slight pressure to hold handpiece tubes in the connector holes, slide coupler
forward and engage threads. Continue threading coupler clockwise until hand tight against
the handpiece.
1. Grasp the swivel in front of the bearing assembly between the thumb and index nger as shown.
2. Slide tubing coupler until the end of the tubing connector is exposed; hold in hand as shown with ngers securing the tubing.
3. Align and insert swivel tubes into the tubing connector holes.
4. While applying slight pressure to hold ttings in the connector holes, slide coupler forward and engage threads. Continue threading coupler clockwise
until hand tight against the swivel.
5. Using the wrench as shown, tighten the swivel to the tubing coupler until snug. Be sure to turn the tubing coupler, not the swivel. DO NOT OVER TIGHTEN.
6. Connect handpiece to swivel until it clicks.
430 SWL and 430 SW:
• Replacement Part Numbers
• Warranty
Sterilizable up to 135°C
135°C
5
Turn tubing coupler while
holding swivel in
place
Wrench
Tubing Coupler
Swivel ISO Connector
Tubing
Coupler
Tubing
Bearing
Assembly
1,2,3,&4
BEFORE OPERATING THE HANDPIECE, CAREFULLY READ AND FOLLOW ALL OF THESE INSTRUCTIONS. SAVE THESE
INSTRUCTIONS FOR FUTURE REFERENCE AND OBSERVE ALL WARNINGS AND CAUTIONS.
Indications For Use
The 430 High-Speed Dental Handpieces are pneumatically driven, hand-held devices intended for use by trained dental professionals for drilling
in the oral cavity. Procedures include, but are not limited to, caries removal, restorative work and crown preparations.
430 K: Type: High speed with a replaceable autochuck
turbine
Operating Air Pressure: 30 to 32 psi
Air Consumption: 1.9 SCFM @ 30 psi
No Load Speed: 430,000 rpm
Connection: 4-line xed ISO conguration
Power: 14 watts
Weight: 47g
ISO Connector
Tubing
Coupler
Tubing
CAUTION
SAFETY
INSTRUCTIONS
WARNING
To minimize risk of serious injury:
• Always insert bur the full length of the chuck. If you attempt to extend it from the chuck, the bur may dislodge during use and possibly injure you or
your patient.
To minimize risk of injury or damage to handpiece or auto chuck:
• Replace turbine when handpiece does not run at full speed or runs noisily.
• Use only recommended burs. An improperly clamped or bent bur, usually indicated by sudden high-pitched sound or increased vibration during
use, may dislodge injuring you or your patient.
• Follow these safety instructions:
- Conrm bur is secure by tugging on it after it is loaded
- Verify concentric rotation by operating handpiece outside oral cavity. Do not use if it does not run concentrically or runs with excessive noise.
- Do not operate handpiece without an installed bur.
- Always use burs with overall length of less than 24mm. Shank length must be at least 7.3 mm. This includes most short shank burs. (see Handpiece
Operation)
- When using autochuck, never depress end cap while bur is rotating.
LUBRICATED Ceramic Bearing Turbine:
For maximum turbine life do not exceed recommended pressure.
Note: Lubricate handpiece daily with DentaLube
II
, StarDental Part #262539, and prior to each sterilization
procedure.
1. Disconnect handpiece.
2. Assemble the conical red tip applicator, StarDental Part #202954, into the DentaLube II bottle cap.
3. Find drive air tube at rear of handpiece and place three drops of DentaLube II.
(See illustrations for more detail).
Note: Never operate handpiece without bur in place.
4. With bur in place connect handpiece to tubing, operate for 20 seconds to expel excess lubricant.
StarDental®
Instrument Solutions
Red Tip
Applicator
Drive Air Tube
Lubrication - For Lubricated Handpieces ONLY
NOTICE LubeFree Handpieces Contain A LubeFree Ceramic Bearing Turbine -
This turbine uses a sophisticated ceramic bearing design and, therefore, requires no lubrication. To ensure the life of, and to prevent
damage to, this turbine - do not lubricate it. If it gets lubricated even once, it MUST then continue to be lubricated for the rest of its life.
It is the responsibility of the user to use only FDA compliant sterilizing equipment and accessories that are
capable of meeting the manufacturers recommendations below.
CAUTION Do not exceed 135°C
To prevent damage to the bur, use an autoclavable bur block.
Do not use:
• dry heat sterilizer • cold sterilizers • ultrasonic cleaners
• chemical disinfectors • immersion techniques • surgical milk
• surface wipes • hand soaps • detergents
• sterilizer bags •surface sprays
containing disinfectants
Limits On Reprocessing Repeated processing has minimal effect on this handpiece. End of life is normally determined by wear and
damage due to use.
INSTRUCTIONS
WARNING Clean and sterilize handpieces and burs as soon as practical before rst use and after each patient.
Containment and transportation: It is recommended that handpieces are reprocessed as soon as is reasonably practical following use.
Preparation for cleaning: Remove handpiece from swivel. No further disassembly is required.
Cleaning - Automated: Washer-disinfectors are not recommended.
Cleaning - Manual: Cleaning Handpieces:
Internal Surface
Optical Surface
(1 of 2)
Remove debris from external surfaces using a small brush and
isopropyl alcohol. Clean internal surfaces of the handpiece at
the swivel connection location using a small brush and
isopropyl alcohol. Optical surfaces can be cleaned with a
cotton swab soaked in isopropyl alcohol. Dry thoroughly.
Repeat after each patient.
Cleaning Diamond Burs:
• After each use, clean the diamond bur by running it against a wet diamond cleaning stone.
• To mechanically clean, the diamond bur should be held in a solution of either enzymatic
presoak, Sparkl® Dental Instrument Cleaner StarDental Part# 205472, or water.
• Rinse the bur thoroughly and allow it to dry, then Sterilize – See Sterilization.
Lubrication: Lubricate - For Lubricated Handpieces ONLY - See 'Lubrication' instruction section
Disinfection: Use steam sterilization exclusively for complete sterility as outlined below.
Maintenance: Use Junk-Out High Speed Chuck Cleaner Kit (263790) once a week prior to sterilization to ensure
optimum working condition of the chuck. Instructions for cleaning the chuck are included in the Junk-Out kit.
See Turbine Replacement section of this manual for additional information.
Handpiece should be routinely maintained, checked for corrosion, worn parts, etc. and should be replaced
on an as needed basis.
Inspection & function testing: See Handpiece Operation or Troubleshooting sections of this manual.
Packaging: Single: A standard sterilization packaging material must be used. Ensure that the pack is large enough to
contain the handpiece without stressing the seals.
In sets: Handpieces may be loaded into dedicated instrument trays or general purpose sterilization trays.
Ensure that cutting edges are protected and do not exceed the tray manufacturer’s limit per tray. Wrap the
trays using appropriate packaging materials and methods.
Wrapped Instrument
Sterilization:
Gravity Type Autoclave: Pre-Vacuum Autoclave:
• Minimum of 15 minutes at 132° C • Minimum of 4 minutes at 132° C
• Minimum of 30 minutes at 121° C
Do not exceed 135° C
Drying: Dry the handpiece for 30 minutes minimum using the sterilization dry cycle and allow the handpiece
to cool to room temperature in bag prior to use. If moisture is present, increase your sterilizer dry cycle
appropriately.
Storage: Store in bag after sterilization until ready to use.
Additional information: Do not exceed the sterilizer manufacturer’s guidelines for load capacity.
Manufacturer contact: See front or back panels of this manual.
Cleaning & Sterilization
Water
Drive Air
Coolant
Air
Drive Air
430 K 430 SWL and 430 SW
Caution: Federal law restricts this device to
sale by or on the order of a dentist.
866.DTE.INFO (383.4636)
www.DentalEZ.com
0344
064561 H Date of Issue 2019-03-31
StarDental®
1816 Colonial Village Lane
Lancaster, PA 17601 U.S.A.
Printed in the U.S.A.
©2019 DentalEZ, Inc.
Emergo Europe
Prinsessegracht 20
2514 AP The Hague
The Netherlands
EC REP
order of a dentist.
ricts this device to
r of a denti
430 High-Speed Handpieces
Includes Lubricated and LubeFree
StarDental® Instruments
Standard and Satin Finish
Non-Sterile
Medical Device
DO NOT USE FOR ORAL SURGERY
CAUTION
NOTICE Any serious incident occurring from the use of this device should be reported to StarDental immediately. If incident occurred in a EU
Member State the Competent Authority must be notied.

Troubleshooting Turbine Replacement
1. Remove old end cap and turbine assembly.
2. Inspect housing before cleaning. Remove any obvious debris to prevent damage and to insure performance of turbine.
3. Connect handpiece and run air through turbine housing without turbine installed for 5 seconds.
4. Clean housing thoroughly with cotton swab and isopropyl alcohol. Dry with syringe air.
5. Place turbine assembly into housing and gently insert as shown.
6. Position and engage threads of end cap into the housing by hand until snug.
7. Tighten end cap with the combination end cap wrench employing the wrench opening marked AC as shown above. Position wrench with marked side
facing end cap for best wrench control.
Autochuck Turbine Assembly
Housing
Wrench Opening
Replacement Part Numbers
Always insist on StarDental Parts for best performance.
Note: Swivel assemblies sold separately.
Description Item Number
Problem: NO LIGHT OR INTERMITTENT LIGHT
Cause: Electrical connections
Correction: Verify connections: wall outlet-power pack, power pack-lamp control, and lamp control-tubing connector.
Cause: Crimped signal air line to lamp control
Correction: Straighten signal air line to allow air ow.
Cause: Corrosion on lamp contact pins
Correction: If 6-pin swivel, remove lamp from swivel. Clean lamp contact pins with isopropyl alcohol.
Cause: Swivel lamp burned out
Correction: Replace Xenon-Halogen lamp with StarDental®Part #262522, or, when operating with a non-StarDental
6-pin F.O. system, use StarDental Part #262962. (Consult your Operator’s Manual for correct voltage.)
Note: See Handpiece Installation, Non-Star 6-pin F.O. System.
Problem: YELLOWISH LIGHT
Cause: Debris on optical surfaces
Correction: Clean all exposed optical surfaces. (Refer to Cleaning.)
Problem: REDUCED LIGHT
Cause: Low voltage
Correction: Refer to StarBright®Fiber Optic Electrical System & Handpiece Tubing Instruction Manual or
contact StarDental Customer Service at 1-866-DTE-INFO.
Cause: Swivel lamp burned out
Correction: Replace Xenon-Halogen lamp with StarDental Part #262522, or, when operating with a non-StarDental
6-pin F.O. system, use StarDental Part #262962. (Consult your Operator’s Manual for correct voltage.)
Note: See Handpiece Installation, Non-Star 6-pin F.O. System.
Problem: LACK OF POWER
Cause: Air pressure too low
Correction: Verify proper handpiece air pressure. (Refer to Handpiece Installation.)
Cause: Crimped tubing
Correction: Straighten tubing inside and outside delivery head.
Cause: Debris buildup
Correction: Clean inside of housing and end cap with swab and alcohol.
Cause: Worn out turbine
Correction: Refer to Turbine Replacement.
Cause: Swivel/Handpiece not threaded fully onto tubing
Correction: Refer to Handpiece Installation.
Problem: WATER LEAKAGE
Cause: Brittle handpiece and/or swivel rear seal
Correction: Replace rear seal for 4-line, “Green” StarDental Part #202897, and for 6-line,
“Green” StarDental Part #262520. For 4-line HiFlo swivel, replace “Black” StarDental
Part #263734 and for 6-pin HiFlo swivel, “Black” StarDental Part #263770.
Cause: Swivel O-rings worn
Correction: Replace both O-rings at the front of swivel using StarDental Swivel O-ring Kit.
See Replacement Part Numbers.
Problem: SWIVEL WILL NOT DISENGAGE FROM HANDPIECE
Cause: Buildup of debris
Correction: Place two drops of DentaLube II into swivel/tubing connector area. Rotate swivel to circulate lube.
Disengage handpiece. Contact dealer if unable to disengage swivel.
To prevent recurrence, refer to Cleaning
Problem: IMPROPER SPRAY
Cause: Clogged lters, tubes or water ports
Correction: Inspect and clean water lter in delivery head. Disconnect handpiece.
Depress foot control; verify water exits the swivel.
(If water does not come from swivel, refer to the swivel's instruction manual.)
Cause: Brittle handpiece/swivel rear seal
Correction: Replace rear seal
Cause: Swivel O-rings worn
Correction: Replace both O-rings at the front of swivel using StarDental Swivel O-ring Kit.
See Replacement Part Numbers.
Cause: Clogged spray ports
Correction: Use clean-out wire to unclog ports.
Problem: BUR LOCKS IN AUTOCHUCK
Cause: Debris build-up in chucking mechanism
Correction: Clean with Junk-Out Chuck Cleaner P/N 263790
Cause: Use of non-ISO burs
Correction: Always use ISO burs.
Problem: BUR SLIPS IN AUTOCHUCK
Cause: Debris build-up in chucking mechanism
Correction: Clean with Junk-Out Chuck Cleaner P/N 263790
Cause: Use of non-ISO burs
Correction: Always use ISO burs.
Clean debris
from inside of
housing
Bearings
Swivel O-rings at
the front of the
swivel
Spray Port
Note: Always use ISO burs to prevent locking and slipping of burs in autochuck. If problems of locking and slipping continue, replace
worn autochuck turbine immediately with the appropriate StarDental Autochuck Turbine Replacement Kit for the handpiece
type (LubeFree or Lubricated) - See Replacement Part Numbers.
If any problem persists, contact dealer for further assistance; or, contact StarDental Technical Service at 1-866-DTE-INFO.
Note: Replace the turbine when the handpiece fails to achieve full speed and/or increases sharply in noise level. The replacement cartridge is available
from your dealer and is easily replaced in the ofce. When ordering a replacement, be sure to order the appropriate StarDental Autochuck Turbine
Replacement Kit for the handpiece type (LubeFree or Lubricated) - See Replacement Part Numbers.
Limited Warranty
O-Rings and Seals are specically excluded from this warranty.
DentalEZ®warrants the 430 SWL®, 430 SW™and 430 K®Handpieces for 1 year from the date of
purchase* with the exception of the ber optic glass which carries a warranty for 5 years from date
of handpiece purchase*.
Please note the following additional terms of our warranty policy:
• Warranties cover manufacturing defects only and do not cover defects resulting from abuse,
improper handling, cleaning, care or maintenance, normal wear and tear or non-observance
of operating, maintenance or installation instructions. Failure to use authorized parts or an
authorized repair facility voids this warranty.
• Liability is limited to repair or replacement of the defective product at our sole discretion.
All other liabilities, in particular liability for damages, including, without limitation,
consequential or incidental damages are excluded.
• THIS WARRANTY IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESSED OR IMPLIED, INCLUDING
ANY IMPLIED WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE.
NO EMPLOYEE, REPRESENTATIVE OR DEALER IS AUTHORIZED TO CHANGE THIS WARRANTY IN
ANY WAY OR TO GRANT ANY OTHER WARRANTY.
*Provided conditions in warranty are met.
For new products, manuals and technical information, call
1-866-DTE-INFO or visit our web site @ www.dentalez.com
Important Notice About Returning Product To The DentalEZ Inc.
Anyone wishing to return any DentalEZ Inc. product to a DentalEZ facility MUST obtain the proper
RETURN AUTHORIZATION NUMBER from the Customer Service Department. The RA Number must appear
clearly marked on both the outside and inside of the returned product carton. When calling for a return
authorization number you will be asked to provide the following information:
To Obtain A
Return Authorization
Number Call
1-866-DTE-INFO
• Product Name, Model Number or Part Number
• Serial Number of Product
• Dealer Return Purchase Order Number
• Reason For Return
• Copy of Invoice Showing Purchase
Items will be evaluated. Return Authorization Number is not a guarantee for credit. Your cooperation with
this procedure will help us expedite processing of any returns.
WARRANTY REGISTRATION
PLEASE VISIT THE WEBSITE BELOW TO REGISTER YOUR STARDENTAL PRODUCT. THIS WILL
ENSURE PROPER REGISTRATION AND PROMPT AND ACCURATE SERVICE SHOULD ANY OF YOUR
PURCHASES NEED EVALUATION OR REPAIR DURING THE WARRANTY PERIOD.
StarDental® Products: www.dentalez.com/warranty
Note: Swivel assemblies sold separately.
430 SWL Satin LubeFree Handpiece............................................................................................................................................................................................................................268840
430 SWL® Fiber Optic LubeFree™Handpiece with Autochuck Turbine.........................................................................................................................................................264450
430 SWL®Fiber Optic Lubricated Handpiece with Autochuck Turbine........................................................................................................................................................ 264451
HiFlo Swivel Assembly: 4-line Fiber Optic .......................................................................................................................................................................................................... 263758
HiFlo Swivel Assembly: ISO/C (6-pin 3.3 Vdc) StarBright®..............................................................................................................................................................................263772
HiFlo Swivel Assembly: ISO/C (6-pin 3.1 Vdc) (for non-StarDental F.O. System)....................................................................................................................................263773
StarBright® Plug n' Play™ LED Swivel Assembly.....................................................................................................................................................................................................265646
430 SW Satin LubeFree Hanpiece ................................................................................................................................................................................................................................. 268841
430 SW Satin Lubricated Handpiece ...........................................................................................................................................................................................................................268861
430 SW™LubeFree Handpiece with Autochuck Turbine ..................................................................................................................................................................................... 262057
430 SW™Lubricated Handpiece with Autochuck Turbine .................................................................................................................................................................................. 264951
Swivel Assembly 430 SW, ISO 2/3-line....................................................................................................................................................................................................................261550
Swivel Assembly 430 SW, ISO 4-line.........................................................................................................................................................................................................................261547
430 K®LubeFree Handpiece with ISO 4-line Fixed Connection and Autochuck Turbine ......................................................................................................................262052
430 K®Lubricated Handpiece with ISO 4-line Fixed Connection and Autochuck Turbine ...................................................................................................................264952
Ceramic Bearing Autochuck Turbine Replacement Kit for LubeFree High-Speed Handpieces.............................................................................................................263739
Ceramic Bearing Autochuck Turbine Replacement Kit for Lubricated High-Speed Handpieces.........................................................................................................263968
End Cap Replacement for Autochuck Turbine ......................................................................................................................................................................................................... 261294
Replacement Lamp, 6-pin 3.3 VDC for swivel connected to StarBright F.O. System
(Package of 2)......................................................................................................................................................................................................................................................................... 262522
Replacement Lamp, 3.1 VDC 6-pin for swivel connected to non-StarDental F.O. System
(Package of 2).........................................................................................................................................................................................................................................................................262962
Rear Seal, 2/3-line (Package of 3) ..................................................................................................................................................................................................................................203103
Rear Seal, 4-line (Package of 3) ...................................................................................................................................................................................................................................... 202897
Rear Seal, 6-line (Package of 3) ...................................................................................................................................................................................................................................... 262520
Rear Seal, HiFlo 6-Pin, Fiber Optic Swivel (Package of 3)......................................................................................................................................................................................263770
Rear Seal, HiFlo 4-Line, Fiber Optic Swivel (Package of 3) .................................................................................................................................................................................. 263734
End Cap Wrench ...................................................................................................................................................................................................................................................................203048
Swivel O-Ring Kit for All StarDental Swivels (2 Large, 2 Small) .........................................................................................................................................................................259884
Air Pressure Gauge (ISO 4/6-line)...................................................................................................................................................................................................................................262591
Cleaning Wire (Gauge: 0.008, Package of 2)................................................................................................................................................................................................................256719
DentaLube II........................................................................................................................................................................................................................................................................... 262539
ISO Tubing Set Screw (Package of 2)..............................................................................................................................................................................................................................261355
Allen Wrench .035 ................................................................................................................................................................................................................................................................280830
Junk-Out High-Speed Chuck Cleaner........................................................................................................................................................................................................................... 263790
Sparkl® Dental Instrument Cleaner................................................................................................................................................................................................................................ 205472
This manual suits for next models
12
Other StarDental Dental Equipment manuals
Popular Dental Equipment manuals by other brands

Aseptico
Aseptico ADU-08SX Sealant Express Operation and maintenance instruction manual

Satelec
Satelec Cocoon Hygienist user manual

Whip Mix
Whip Mix VeriPulse IOS user manual

Sirona
Sirona Intego Brief operating instructions

Cavitron
Cavitron Plus G136 Directions for use

KaVo
KaVo MASTERmatic LUX M10 L Instructions for use