manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. Status Instruments
  6. •
  7. Measuring Instrument
  8. •
  9. Status Instruments SEM1600B User manual

Status Instruments SEM1600B User manual

D2534-01-04
SEM1600B
USER INSTRUCTIONS
Important - Please read this document before installing.
Every effort has been taken to ensure the accuracy of this document; however, we do
not accept responsibility for damage, injury, loss or expense resulting from errors and
omissions, and we reserve the right of amendment without notice.
IMPORTANT – CE, UKCA & SAFETY REQUIREMENTS
Product must be DIN rail mounted, inside a suitable enclosure providing environmental
protection to IP65 or greater.
To maintain CE UKCA requirements, input and supply wires must be less than 30
metres.
The product contains no serviceable parts, or internal adjustments. No attempt must be
made to repair this product. Faulty units must be returned to supplier for repair.
Before attempting any electrical connection work, please ensure all supplies are
switched off.
ABSOLUTE MAXIMUM CONDITIONS (To exceed may cause damage
to the unit).
Supply voltage (SELV)
± 50 Vdc Protected for over-voltage
and reverse connection
Current with over-voltage
± 200 mA
Input voltage
± 10 VDC between any terminals
Input current
± 100 mA between any terminals
Environmental protection
IP65 or greater required
Ambient
Temperature (-30 to 75) °C
RH (10 to 95)% non-condensing
Important – Read this
document before installing.
--------------------------------------------------------------------------------------------------------------
1~DESCRIPTION.
The SEM1600B is a powered bridge amplifier for use with strain
gauges or load cell signals. The product has a built-in capability to
scale the input signal to a process value while the output stage offers
either voltage, bipolar voltage or active/passive current retransmission
signals. The device uses ratio metric measurement to obtain high
stability.
The product comes with an AC/DC power supply that will operate in
the range (10 to 48) VDC and (10 to 32) VAC making the device ideal
for battery operation. An additional volt-free contact input is available
for tare setting using a remote switch. The high precision input stage
of the device allows for a bridge excitation voltage of 5 VDC to be
used, as opposed to the traditional 10 VDC. This reduces the power
requirement for the bridge supply and up to four bridges (cells) may
be connected to the input.
To configure: connect a standard USB cable between the SEM1600B
and a PC. The free configuration software will guide you through any
changes you wish to make. The SEM1600B does not need to be
wired to a power supply for configuration.
2~RECEIVING AND UNPACKING.
Please inspect the packaging and instrument thoroughly for any signs of transit
damage. If the instrument has been damaged, please notify your supplier immediately.
3~SPECIFICATION.
Refer to the datasheet for full specification. Download at www.status.co.uk
Factory
defaults
Calibration factor 2 mV/V, Balance 0.0
Sample rate 10 per s, Filter Off
Scale 2 points
Tare Setpoint 0, Tare Offset 0
Process Output 0 to 100 = (4 to 20) mA
Button locks Off
4~INSTALLATION AND WIRING.
Important safety requirements
This equipment is suitable for Environment Installation BS EN61010-1
Pollution Degree 2; Installation CAT II; CLASS I and is classed as
"PERMANENTLY CONNECTED EQUIPMENT". The equipment is
intended for industrial and commercial application only and is not
suitable for domestic or medical use.
The equipment must be mounted inside an enclosure that provides
protection >= IP65. In NORMAL USE, the equipment will only be
accessed for maintenance by qualified personnel.
Please ensure the equipment is mounted vertically with terminals (10,
11 and 12) at the bottom. This will provide maximum ventilation.
This equipment may generate heat. Ensure the enclosure size is
adequate to dissipate heat. Be sure to consider any other equipment
inside the enclosure.
The equipment surfaces may be cleaned with a damp cloth. Use a
mild detergent/water. Ensure the supply is off before cleaning and, on
completion of cleaning, the equipment is completely dry before the
supply is turned back ON.
This equipment must be installed by a qualified person. All electrical
wiring must be carried out in accordance with the appropriate
regulations for the place of installation.
4.1~MECHANICAL.
Dimensions in mm
56.4 mm17.5 mm
1
2
11 1210
987
4 5 6
1 2 3
90mm
ADJUST
ST AT E
The equipment must be mounted on a DIN rail style DIN EN50022
inside a plastic or metal enclosure with a protection level >= IP65. All
wiring must be secured. Maximum cable sizes 2.5 mm². Connection is
via screw clamp terminals.
4.2~ELECTRICAL/CONNECTIONS
For wiring connections refer to the side label on the SEM1600B and
this document.
Supply (10 to 48) Vdc, (10 to 32) Vac
To maintain CE EMC requirements, input and supply wires must be
less than 30 metres.
Input cable lengths > 3 m it is recommended to use screened or
twisted pair.
Maximum mA output cable run = 1000 metres. The output loop
should be grounded at a single point.
Before installation, care must be taken to ensure enough voltage is
available in any loop to drive the total loop load.
D2534-01-04
4.2~ELECTRICAL (continued)
SEM1600B Connections
1
+
SUPPLY
2
Vdc Vac
INPUT
4
Voltage
5
3
6
mA
6
5
mA Externally Powered
OUTPUT
Vs
Tare
Vexc
Bridge Input
Vexc
8
7
10
11
9
12
V
out
+
+
++
+
+
5~FEATURES.
5.1~STATE LED
The State LED is GREEN under normal run conditions indicating an
in-range input signal. If the input signal is out of range or is lost, the
State LED will light RED.
5.2~TARE
If PINs 9 and 12 are shorted (momentarily) the input value will be
overwritten by the TARE setpoint value (normally zero).
5.3~ACTIVE TRIM
Down Adjust button press: Configures the low scale output against the
live input signal.
Up Adjust button press: Configures the high scale output against the
live input signal.
Press and keep button pressed - the state light will go off for a few
seconds then flash at a fast rate before returning to a steady state.
Release button. Adjust operation complete. Output adjusts to levels
pre-set in the software configuration.
6~USER CONFIGURATION.
The SEM1600B can be configured using a Windows PC.
Live input and output values can also be viewed on a PC or a suitable
Android device.
6.1~PC CONFIGURATION USBSpeedLink Software
During configuration the equipment takes its power from the
USB port, therefore no power connection is required. The equipment
can be configured whilst powered but the computer used must be
portable battery-powered or a USB isolator should be used to isolate
the SEM1600B from the supply earth to avoid grounded earth loop
effects.
Observe any warning information given in the software.
PC Configuration steps
1
Download and install the USBSpeedLink software from
www.status.co.uk
2
Run the software and open to the correct screen for the SEM1600B
3
Connect to the PC using an A to Mini B USB lead.*1
4
Read the SEM1600B configuration into the software.
5
Configure the device to the required settings for operation.
5.1
SEM1600B configuration options:
Model:Free type field
Calibration factor: From sensors/Load cell datasheet.
Balance:From Load cell calibration certificate if available.
Tag:Free type field
Sample rate:Samples per second
Filter: Rate in seconds
Scale points:2 up to 6 points
mV to Process value table:mV input to process table
Units:Free type field for Engineering unit
Tare Clear: Removes Tare
Tare Setpoint:Value the input is set to on Tare action
Tare Offset:
Difference between input and Tare setpoint on
tare action
Output Signal:mA, V, ±V
Output Range:Any within limits
Button locks:On, Off
6
Read Data:Live data can be displayed showing input and output
values. This can only be done if the device is powered as well as
connected to the software via the USB lead.
*3
7
Write/Save:the configuration to the device. *2
*1 Once only, on the first time connecting to the SEM1600B, drivers will
install to the PC, allow time for this before proceeding.
*2 The configuration is not saved onto the device unless the configuration
screen is sent.
*3 The SEM1600B can be configured whilst connected and powered, but a
portable battery powered computer or USB isolator must be used to avoid
the effects of ground loops.
.
Configuration screen
6.2~ANDROID MONITORING USBView Software
Using a suitable OTG USB lead to connect the SEM1600B to an
Android device, live data reading can be taken.
The USBView app. can display input temperature/value, output mA/V
and the Tag information.
USB Software can be downloaded free of charge from
www.status.co.uk
www.status.co.uk
This guide is also available online at www.status.co.uk
Status Instruments Ltd, Status Business Park, Gannaway Lane, Tewkesbury, Gloucestershire, UK, GL20 8FD,
Web Page: www.status.co.uk,
Email: [email protected]
Technical Support: support@status.co.uk
Tel: +44 (0) 1684 296818, Fax: +44 (0) 1684 293746

Other Status Instruments Measuring Instrument manuals

Status Instruments DM4500U User manual

Status Instruments

Status Instruments DM4500U User manual

Status Instruments DM650HM User manual

Status Instruments

Status Instruments DM650HM User manual

Status Instruments DM3600A User manual

Status Instruments

Status Instruments DM3600A User manual

Status Instruments DM3600U User manual

Status Instruments

Status Instruments DM3600U User manual

Status Instruments DM4000U User manual

Status Instruments

Status Instruments DM4000U User manual

Status Instruments SEM1620 User manual

Status Instruments

Status Instruments SEM1620 User manual

Popular Measuring Instrument manuals by other brands

Oval EL4101 Key operations guide

Oval

Oval EL4101 Key operations guide

Omron F3EM datasheet

Omron

Omron F3EM datasheet

PCB Piezotronics Triaxial ICP 356A12/010CA Installation and operating manual

PCB Piezotronics

PCB Piezotronics Triaxial ICP 356A12/010CA Installation and operating manual

Magtrol 6530 user manual

Magtrol

Magtrol 6530 user manual

MULTISPAN MFM 14DC manual

MULTISPAN

MULTISPAN MFM 14DC manual

Cirrus Research Trojan2 User's manual guide

Cirrus Research

Cirrus Research Trojan2 User's manual guide

PCB Piezotronics ICP J352C34 Installation and operating manual

PCB Piezotronics

PCB Piezotronics ICP J352C34 Installation and operating manual

Major tech MT145 instruction manual

Major tech

Major tech MT145 instruction manual

Hilti PR 25 operating instructions

Hilti

Hilti PR 25 operating instructions

Endress+Hauser Micropilot M FMR231 operating instructions

Endress+Hauser

Endress+Hauser Micropilot M FMR231 operating instructions

Bosch BEA 055 Original instructions

Bosch

Bosch BEA 055 Original instructions

Human Elisys Quattro user manual

Human

Human Elisys Quattro user manual

ATD Tools 12551 owner's manual

ATD Tools

ATD Tools 12551 owner's manual

YOKOGAWA 96001 user manual

YOKOGAWA

YOKOGAWA 96001 user manual

Xylem WTW Multi 9430 operating manual

Xylem

Xylem WTW Multi 9430 operating manual

Marine Magnetics SeaSPY2 Operation manual

Marine Magnetics

Marine Magnetics SeaSPY2 Operation manual

BRONKHORST FLEXI-FLOW Compact quick start guide

BRONKHORST

BRONKHORST FLEXI-FLOW Compact quick start guide

RS PRO RS-136 instruction manual

RS PRO

RS PRO RS-136 instruction manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.