Installation
The unit may be mounted in a panel of any thickness up to a
maximum of 6mm (0·25in). Leave enough space behind the
instrument to allow for bends in the connection cables. As the
front of panel enclosure conforms to IP52, it is protected from
dripping water. The unit is intended for use in a reasonably
stable ambient temperature within the range -10 to +55°C. Do
not mount the unit where there is excessive vibration or in
excessive direct sunlight.
Safety
The unit is designed in accordance with BS EN 61010-1:2001
(IEC 61010-1:2001) – Permanently connected use, Normal
condition, pollution degree 2, Measurement Category III
Reinforced Insulation
EMC Installation Requirements
Whilst this unit complies with all relevant US EMC (electro-
magnetic compatibility) regulations, any additional precautions
necessary to provide proper operation of this and adjacent
equipment will be installation dependent and so the following can
only be general guidance:
•Avoid routing wiring to this unit alongside cables and
products that are, or could be, a source of interference.
•The signal to the unit should not be subject to excessive
interference. In some cases, a supply line filter may be
required.
•To protect the product against incorrect operation or
permanent damage, surge transients must be controlled. It
is good EMC practice to suppress transients and surges at
the source. The unit has been designed to automatically
recover from typical transients; however in extreme
circumstances it may be necessary to temporarily reduce the
magnitude of the measured signal to the point where the
instrument turns off for a period of greater than 10 seconds
to restore correct operation.
•It is good practice to install sensitive electronic instruments
that are performing critical functions in EMC enclosures that
protect against electrical interference causing a disturbance
in function.
Wiring
Input connections are made to terminals posts. Use cable lugs
suitable for 6-32 UNC screw. Wiring should conform to national
Electrical Code Class 1 and sized appropriate to CT secondary
current ratings.
The current inputs of this product are designed for
connection into systems via current transformers only.
Instrument transformers used for connection to the meter must
be of approved type and compliant with ANSI/IEEE C57.13 or
IEC 60044-1, selected and sized appropriate to the supply
network being monitored. CT secondary’s must be grounded in
accordance with local regulations. It is desirable to make
provision for shorting links to be made across CT secondary’s to
permit easy replacement of a unit should this ever be necessary.
This instrument is intended for panel mounting. Terminals must
be enclosed within the panel. Terminal nuts should be tightened
to 8.0 lbs in (1.0Nm) only.
Earth/Ground Connections
For safety reasons, current transformer secondary connections
should be grounded in accordance with local regulations. Under
no circumstances should the product be operated without this
Earth connection.
Maintenance
In normal use, little maintenance is needed. As appropriate for
service conditions, isolate from electrical power, inspect the unit,
and remove any dust or other foreign material present.
Periodically check all connections for freedom from corrosion and
screw tightness, particularly if vibration is present.
The front of the case should be wiped with a dry cloth only. Use
minimal pressure, especially over the viewing window area. If
necessary wipe the rear case with a dry cloth. If a cleaning agent
is necessary, isopropyl alcohol is the only recommended agent
and should be used sparingly. Water should not be used. If the
rear case exterior or terminals should be contaminated
accidentally with water, the unit must be thoroughly dried before
further service. Should it be suspected that water might have
entered the unit, factory inspection and refurbishment is
recommended.
Contains no serviceable parts:
In the unlikely event of a repair being necessary, it is
recommended that the unit be returned to the factory or nearest
Crompton Instruments / TE Connectivity service centre.
Specification
Measurement Inputs
Nominal Input Current 5A AC RMS
Max. Continuous input Overload 6A AC RMS
Max. Short duration input Current 50A for 1 second.
Maximum VA Burden 3 VA
Frequency 60Hz +/- 2%
Maximum Total Harmonic Distortion 30%
Maximum Crest Factor 1.68
Measurement Category CAT III / 600V
Range of Use
Values of measured quantities, components of measured
quantities, and quantities which affect measurement errors to
some degree, for which the product gives meaningful readings:
Warnings
•During normal operation, voltages hazardous to life
may be present at some of the terminals of this unit.
•Installation and servicing should be performed only
by qualified, properly trained personnel abiding by
local regulations.
•Ensure all supplies are de-energised before
attempting connection or other procedures.
•Terminals should not be user accessible after
installation and external installation provisions must
be sufficient to prevent hazards under fault
conditions.
•Never open-circuit the secondary winding of an
energized current transformer.
•This product should only be operated with CT
secondary connections Earthed.
•If this equipment is used in a manner not specified
by the manufacturer, protection provided by the
equipment may be impaired.
•If sufficient torque setting of 8.0Ibs in (1.0Nm) cannot
be applied to the rear fixing, the product should not
be installed.
•This product contains a Lithium battery. Do not
dispose of in a fire, as the battery may explode.
Where ever possible the product should be recycled,
or disposed of in accordance with local regulations.
Caution: Risk of
Electric Shock and
Danger