Status Instruments DM650HM User manual

D2614-01-03 CN5819
DM650HM USER GUIDE
BATTERY POWERED %RH, °C/°F METER
WITH RELAY TRIP AND DATA LOGGING FUNCTION
Important - Please read this document before installing.
Every effort has been taken to ensure the accuracy of this document; however, we do
not accept responsibility for damage, injury, loss or expense resulting from errors and
omissions, and we reserve the right of amendment without notice.
IMPORTANT – CE, UKCA & SAFETY REQUIREMENTS
The instrument is designed to be directly attached to the sensor probe assembly.
Remote Probes may be used but the user must ensure all sensor and cable entries
maintain environmental protection to at least IP65 rating.
To maintain CE EMC requirements, input wires must be less than 3 metres.
The product contains no serviceable parts, or internal adjustments. No attempt must
be made to repair this product. Faulty units must be returned to supplier for repair.
This product must be installed by a qualified person. All electrical wiring must be
carried out in accordance with the appropriate regulations for the place of installation.
Before attempting any electrical connection work, please ensure the battery is
removed
ABSOLUTE MAXIMUM CONDITIONS (To exceed may cause damage to the unit).
Battery Voltage
± 3.7 VDC (Protected for over-voltage and reverse connection)
Ambient
Temperature (-30 to 70) °C, Humidity (10 to 95) % RH (Non-
condensing)
Relays
50 VDC 40 VAC RMS
Important – Read this
document before installing.
---------------------------------------------------------------------------------
1~DESCRIPTION.
The DM650HM is a battery powered indicator with built-in sensors designed to
display % relative humidity, temperature, dew point and delta temperature. The
instrument offers an advanced display-mode, allowing the user to display not only
process variables but one or two (1–32) character messages. Additional to the
messages, the user may select an alert event to occur when the process is within a
band. The change-over relay can be used together with the messaging and alert to
provide switching at user-set points. In addition to this, on board datalogging is also
available to record process details at up to 5000 points.
2~RECEIVING AND UNPACKING.
Please inspect the packaging and instrument thoroughly for any signs of transit
damage. If the instrument has been damaged, please notify your supplier immediately.
3~SPECIFICATION.
Refer to data sheet for full specification.
Factory default
Displays %RH, °C, relay not set
4~INSTALLATION AND WIRING
4.1~MECHANICAL.
The instrument is a high accuracy digital thermometer. In order to ensure correct
operation, the following must be observed:
The product must be stored in a dry clean environment and remain in original
packaging prior to installation.
The instrument must not be installed adjacent to electro-mechanical starters,
controllers, thyristor power units or electrical switch gear.
Any cleaning of the instrument must be done using a mild detergent and soft cloth. No
solvents or abrasive cleaners should be used.
Stated ambient operating conditions must not be exceeded. Battery life will reduce
with higher ambient temperature operating conditions.
4.2~ELECTRICAL.
For a wiring diagram please refer to the rear panel of the DM650HM inside the
case housing.
Connections: NO (Normally Open), NC (Normally Closed), and C (Common).
Two-part connectors are used for input and relay connections, allowing the unit to be
easily removed from the housing for reprograming or data download if this is not
possible in situ. On insertion of battery the unit will run through a power up check;
during this time the relay may change state.
IMPORTANT: Always remove battery before any wiring takes place. Gain access to
the connectors and battery holder by unscrewing the cap ring to release front panel
assembly from case.
RELAY CONNECTION: A relay with changeover contacts is available. Screw
terminals are provided for connection for wire size 16 to 20 AWG. The relay contacts
are rated at 48 VDC 28 VAC RMS @ 1 A (5 mA minimum current) see DM650HM
data sheet.
BATTERY: To remove battery, use screwdriver to ease the positive end of the battery
out of holder. Insert new battery negative end first then press into place. (Observe
polarity). Battery type 3.6 V Lithium (2.4 A/Hr) CR14505 (IEC) AA case style. Please
dispose of the battery in a responsible way.
5~USER CONFIGURATION. IMPORTANT
READ COMPLETE SECTION BEFORE ATTEMPTING CONFIGURATION.
The instrument is provided with a USB interface for direct connection to a PC. Free
software USBSpeedLink is available. Please refer to the USBSpeedLink software for
further information on configuration. The software can be downloaded from
www.status.co.uk.
DISPLAY: The display provides six 14 segment characters for display of temperature
and alpha-numeric messages, together with an 8 segment bar graph and six icons.
The display can operate in an ambient temperature range of (-30 to 70) °C, but at
temperatures lower than -5 °C (due to the slow LCD speed) scrolled messaging is not
practical.
9
1 2 34 6
8
7
8
1. NFC - The Symbol is on when a NFC field is detected. When a detected field is lost the symbol
will turn off after a few seconds.
2. TRANSMIT/RECEIVE: Symbol on when either NFC or USB communication is active.
3. USB: Symbol on when USB port is connected to a PC. Please note battery is not required during
configuration.
4. LOG: Not used
5. Not used
6. BAR GRAPH: Indicates the state of the logger. The condition is dependent on the selected
logger mode, either Single or Rolling mode.
Single Mode (Log to the maximum number of logs then stop)
Start of Log Mid Log End of Log (alternating)
Rolling Mode (Log to the maximum number of logs then as each new log is taken the oldest log is
discarded)
Start of Log Mid Log Rolled over Log (max bar toggles)
7. DEG: Degree symbol “°” used to indicate either °C or °F on the last digit.
8. DIGITS: Six-digit 14 segment display with – sign, range 9999.9 to -9999.9. Advanced mode offers
two temperature-dependent 32-character message options.
9. WARNING ICON: This symbol will toggle on and off to indicate a warning. The warning symbol will
be active either when the sensor signal is out of range, not connected or when the battery is low

D2614-01-03 CN5819
5.1~USB CONFIGURATION
Using the USBSpeedLink software to program a new configuration into the device
work from left to right through the option tabs.
The USB port is located inside the housing on the rear panel.
Configuration steps
Install and run the USBSpeedLink software
Select the correct programming page in the software (follow any on screen
safety precautions)
Connect the USB lead to the PC, connect the USB lead to the device
Read the unit configuration into the software
Re-configure or adjust configuration options as required
Send the new configuration to the unit
5.1.1~Sensor option tab
Use this tab to pre-set RH and temperature values; this can be used for system testing
and diagnostics.
5.1.2~Relay tab (showing live readings in the “Data” section)
Use this tab to set the relay controls.
Relay options for %RH, temperature, dew point and delta t. source
Set point
Dead band
Low alarm
Low control (inverted low alarm)
High alarm
High control (inverted high alarm)
Alarm inside band
Alarm outside band
Low battery alarm
Alarm off
A diagram will display to show alarm type function
5.1.3~Basic display options tab
Use this tab to select the display value options, °C or °F and LED function
!
5.1.4~Advanced display options tab
Use this tab to set up the advance messaging options.
5.1.4~Advanced display options tab (continued)
6 user set ranges/bands can be set for either %RH, temperature, dew point or delta t.
Each range can have 1 or 2 messages assigned that will display on the screen when
the input source signal is within the set band. Also, the LED can be set to flash, within
the selected range.
Up to 7 user messages can be entered or pre-set messages can be used.
Relay in or out of alarm trip can also be used to prompt specific messages to be
displayed.
5.1.5~Other display options tab
Use this tab to set the “Tag” and “Contact” free type fields stored to the device. A
passcode to protect the device log from being overwritten by an Android device can be
added.
Longitude and latitude data can be entered, this will show in the log files and can help
identifying data logs recorded from multiple devices.
6~DATA LOGGING
USB and Android NFC software allows the user to read existing logs, change the log
manifest, start a new log, synchronise the instrument clock and reset the
maximum/minimum readings.
Logs can run to a fixed number and stop or continually roll over, up to 5000 log points
can be recorded. The start of the log can be delayed up to one month.
Data from can be viewed on screen as a graph or table and can also be saved to a
csv file.
6~1 USB LOGGING
USB LogLink software
6~2 NFC LOGGING
Status Instruments Ltd, Status Business Park, Gannaway Lane, Tewkesbury,
Gloucestershire, UK, GL20 8FD,
Web Page: www.status.co.uk, Email: sales@status.co.uk
Technical Support: support@status.co.uk
Tel: +44 (0) 1684 296818, Fax: +44 (0) 1684 293746
USB LOGGER INTERFACE (connector inside housing)
The USB interface allows the instrument to communicate with a PC running
the USBLogLink software
The prime function of the interface is to read logged data from the device
using free software available to download.
The software allows the user to read existing logs, change the log manifest,
start a new log, synchronise the instrument clock and reset the
maximum/minimum readings.
USBLogLink is available from the manufacturer or supplier.
NFC LOGGER INTERFACE
The NFC interface allows the instrument to communicate with an Android
device using NFC connectivity.
The prime function of the interface is to read logged data from the device
using a free app, which is available for downloading to Android devices.
The app allows the user to read existing logs, change the log manifest,
start a new log, synchronise the instrument clock and reset the
maximum/minimum/average readings. Logs can run to a fixed number and
stop or continually roll over, up to 5000 log points can be recorded. The
start of the log can be delayed up to one month.
Note: For larger logs the data may take over a minute to fully download via
the NFC interface.
Note: Not all Android devices fully implement the near field communication
(NFC) standard and may not hold the full 5000 log points.
Download Status Instruments apps from the Google Play Store
MULTIFUNCTION ALERT LED
The alert LED normal state is off, on alert the LED will emit an
intense white light pulse every 5 seconds. The LED can be
programmed to pulse on any of the following combined events:
Mode Description
No events The LED never operates, extending battery life.
(Factory default setting)
Battery Alert on low battery detect.
Trip Alert when relay trip is on.
Temperature In advanced mode only the alert LED can be made to
alert in any one of eight user-set temperature bands. Example, to
alert operator when temperature is outside a safe operating range.
The function of the alert LED can be further enhanced with the option
of displaying an alert message in advanced display mode.
Other Status Instruments Measuring Instrument manuals

Status Instruments
Status Instruments DM4000U User manual

Status Instruments
Status Instruments DM3600A User manual

Status Instruments
Status Instruments SEM1600B User manual

Status Instruments
Status Instruments DM4500U User manual

Status Instruments
Status Instruments DM3600U User manual

Status Instruments
Status Instruments SEM1620 User manual
Popular Measuring Instrument manuals by other brands

PCB Piezotronics
PCB Piezotronics ICP J352C34 Installation and operating manual

Major tech
Major tech MT145 instruction manual

Hilti
Hilti PR 25 operating instructions

Endress+Hauser
Endress+Hauser Micropilot M FMR231 operating instructions

Bosch
Bosch BEA 055 Original instructions

Human
Human Elisys Quattro user manual