Stryker EOLE DC 32" User manual

2017/12 836002-5210 V3.4 www.stryker.com
EOLE DC Powered Support Surface
2871
Operations/Maintenance Manual


www.stryker.com 836002-5210 V3.4 3
Table of Contents
Symbols and Denitions................................................................ 4
Symbols ......................................................................... 4
Warning/Caution/Note Denition ...................................................... 5
Technical Specication................................................................. 6
Introduction.......................................................................... 7
Contraindications.................................................................. 7
Intended Use of Product ............................................................ 7
Expected Service Life .............................................................. 7
Product Description ................................................................ 7
Contact Information ................................................................ 7
Product Serial Number Location/Identication ........................................... 8
Summary of Safety Precautions.......................................................... 9
Product Description .................................................................. 10
Control Unit Front................................................................. 10
Control Unit Rear ................................................................. 10
Control Panel .................................................................... 10
Instructions..........................................................................11
Installing the Control Unit............................................................11
Product Functions ................................................................ 12
Transport mode .................................................................. 13
Storage ........................................................................ 13
Cleaning and Disinfection.............................................................. 14
Troubleshooting ..................................................................... 15
Service Information .................................................................. 16
Top Cover Replacement ........................................................... 16
Air Cell Replacement .............................................................. 16
Control Unit Replacement .......................................................... 16
Hose Replacement ............................................................... 16
CPC Tube Replacement ........................................................... 16
Filter Replacement................................................................ 16
Preventive Maintenance............................................................... 17
Checklist ....................................................................... 17
Appendix A: EMC Information .......................................................... 19
Guidance and Manufacturer’s Declaration- Electromagnetic Emissions: ...................... 19
Guidance and Manufacturer’s Declaration- Electromagnetic Immunity: ....................... 20
Warranty........................................................................... 22
Limited Warranty ................................................................. 22
To Obtain Parts and Service ........................................................ 22
Return Authorization .............................................................. 22
Damaged Merchandise ............................................................ 22
International warranty clause........................................................ 22

4836002-5210 V3.4 www.stryker.com
Return To Table of Contents
Symbols and Denitions
SYMBOLS
TUV marking
CE marking
Warning / Caution, consult accompanying documentation
Type BF equipment
Double Insulation
Fuse
Temperature Limitation, Operating: 10°C to 40°C, Storage: -15°C to 50°C
Humidity Limitation, 10% - 90%
Refer to instruction manual/ booklet
Disposal: Contact local distributor who will take the necessary steps according to your
national market.
Do Not Iron
Maximum washing temperature 60°C, normal process, only for top cover of mattress.
Chlorinated Bleach
Do Not Tumble Dry
Do Not Dry Clean
Allow to Completely Air Dry
Manufacturer
IP24
First Digit (Solids) Protected against touch by ngers (>12.5mm); Second Digit (Liquids)
Water splashing against the enclosure from any direction shall have no harmful effect.
Authorized representative in the European community
Catalogue Number (model)
SN
Serial Number
CPR
Do Not Open with Cutter

www.stryker.com 836002-5210 V3.4 5
Return To Table of Contents
WARNING/CAUTION/NOTE DEFINITION
The words WARNING, CAUTION and NOTE carry special meanings and should be carefully reviewed.
WARNING
Alert the reader about a situation which, if not avoided, could result in death or serious injury. It may also
describe potential serious adverse reactions and safety hazards.
CAUTION
Alert the reader of a potentially hazardous situation which, if not avoided, may result in minor or moderate
injury to the user or patient or damage to the equipment or other property. This includes special care necessary
for the safe and effective use of the device and the care necessary to avoid damage to a device that may
occur as a result of use or misuse.
NOTE
Provide special information to make maintenance easier or important instructions clearer.
Symbols and Denitions

6836002-5210 V3.4 www.stryker.com
Return To Table of Contents
Item Specication
Power Supply AC 230V, 50Hz, 0.07A
Fuse Rating T1AL, 250V
Dimension (L x W x H) 29.5 x 14.5 x 19.2 cm / 11.5” x 5.7” x 7.6”
Weight 2.4 kg / 5.3 lb
Cycle Time 12 minutes
Environment
Atmospheric
Pressure Operation: 70-106 hPa
Temperature
• Operation: 10°C to 40°C (50°F to 104°F)
• Storage: -15°C to 50°C (5°F to 122°F)
• Shipping: -15°C to 70°C (5°F to 158°F)
Humidity
• Operation: 10% to 90% non-condensing
• Storage: 10% to 90% non-condensing
• Shipping: 10% to 90% non-condensing
Classication
• Class II, Type BF, IP24
• Applied Part: Air Mattress
• Not suitable for use in the presence of a ammable anesthetic
mixture (No AP or APG protection)
Air Mattress Specication
Model EOLE DC 32” (80cm) EOLE DC 35” (90cm)
Model Number 2871
Flame Retardant Standards EN 597-1 and EN 597-2
Safe Working Load 200 kg / 441 lb
Dimension (L x W x H) 200 X 80 X 20 cm
78.74 X 32 X 7.84 inches
200 X 90 X 20 cm
78.74 X 35.43 X 7.84 inches
Weight 4.75 kg / 10.47 lb 5.45 kg / 12 lb
Technical Specication

www.stryker.com 836002-5210 V3.4 7
Return To Table of Contents
This manual is designed to assist with the operation and maintenance of the EOLE DC Powered Support
Surface. Carefully read this manual thoroughly before using or beginning maintenance on the support
surface. To ensure safe operation of this equipment, it is recommended that methods and procedures are
established for educating and training staff.
CONTRAINDICATIONS
None known.
INTENDED USE OF PRODUCT
EOLE DC is a constant low pressure powered support surface intended to provide pressure redistribution to
aid in the prevention and treatment of pressure ulcers. The system consists of a control unit combined with
an alternating air cell mattress. The air cells redistribute the weight of the patient over the surface and aid in
the reduction of tissue interface pressure. It is recommended that the product be operated by personnel who
are qualied to perform general nursing procedures and have received adequate training in the prevention
and treatment of pressure ulcers.
This support surface is intended to be used with human patients in a general hospital, nursing home or
homecare environment and for patients at risk of developing pressure ulcers, as well as those who require
therapy for pre- existing pressure ulcers. The safe working load for EOLE DC is 200 kg/ 441 lb; the patient
must not exceed safe working load specied by the support surface, frame, and accessories. Patients shall
meet the minimum age requirement of 2 years old.
EOLE DC shall be used with a mattress cover at all times.
The support surface is not intended to be a sterile product nor is it intended to include a measuring function.
EXPECTED SERVICE LIFE
The products are intended to offer safe and reliable operation when in use or installed according to the
instructions provided by Stryker Medical. Stryker Medical recommends that the system be inspected and
serviced by authorized technicians if there are any signs of wear or concerns with device function and
indication on products. Otherwise, service and inspection of the devices generally should not be required.
The control unit thereof has an expected service life of 3 years and the mattress thereof has an expected
service life of 2 years.
PRODUCT DESCRIPTION
EOLE DC is powered support surface focusing on equalizing pressure and enhancing comfort.
Introduction

8836002-5210 V3.4 www.stryker.com
Return To Table of Contents
CONTACT INFORMATION
Contact Stryker Customer Service or Technical Support at: (800) 327-0770 or (269) 324-6500.
Stryker Medical
3800 E. Centre Avenue
Portage, MI 49002
USA
Please have the serial number (A) of your Stryker product available when calling Stryker Customer Service
or Technical Support. Include the serial number in all written communication.
PRODUCT SERIAL NUMBER LOCATION/IDENTIFICATION
The serial number (A) is located at the mattress
cover near foot right corner of the mattress as shown
in Figure 1. To access the serial number, unzip the
cover about one foot.
▼Figure 1
EOLE DC
Mattress Air CellCPR A
Format:
2871
MY Y M M - SSSSS
• M = Mattress
• YY = Year
• MM = Month
• SSSSS = Sequence (Numeric)
Model Number Legend (X)
2871 EOLE DC
Month Legend (MM)
January 01
February 02
March 03
April 04
May 05
June 06
July 07
August 08
September 09
October 10
November 11
December 12
Year Legend (YY)
2014 14
2015 15
2016 16
2017 17
2018 18
Introduction

www.stryker.com 836002-5210 V3.4 9
Return To Table of Contents
WARNING
• Check patient’s skin regularly. Consult physician if any redness or skin break occurs. Serious injury could
result if the patient’s skin condition is left untreated.
• Do not place the control unit in the patient’s bed, in contact with the patient, or under sheets or other
coverings. Doing so could cause serious injury or could affect control unit performance.
• Do not use in the presence of a ammable anesthetic mixture or with oxygen (O2) or nitrous oxide (N2O).
• Verify bed side rails are compatible with bed frame and existing mattress. A risk assessment must be
performed by a suitably qualied person, especially when side rails are prescribed, to ensure that the
bed meets the IEC 60601-2-52 bed standard.
• Use with appropriate top sheet and minimize layers of bedding between patient and mattress.
• Assess patient’s risk of entrapment according to protocols and monitor accordingly.
• Close supervision is necessary when this product is used on or near children. Electrical burns or choking
may result from a child swallowing a small part detached from the device.
• Use this product only for its intended use as described in this manual.
• Do not operate product if the power cord or plug has been damaged.
• Keep the power cord away from heated surfaces.
• Never block any air openings of this product or place it on soft surfaces, where openings may be blocked.
Keep the air opening free of lint, hair, and other similar particles.
• Do not modify this equipment without the authorization of the manufacturer.
• Mattress covers have passed skin sensitization and skin irritation tests. However, if you suspect that
the patient or caregiver you may have had or is having an allergic reaction, please consult a physician
immediately.
• The power cord to the Control Unit should be positioned to avoid a strangulation hazard and/or damage
to the cord. Careful consideration is required when routing the power cable. It is recommended that plac-
ing the cord under the bed frame and attaching it to an electrical outlet at the head of bed.
• Serious injury or death can result from the use (potential entrapment) or non-use (potential patient falls)
of side rails or other restraints. The safe use of the support surface is maximized when used in conjunc-
tion with side rails; there may be an increased risk of falls when side rails are not present. Local poli-
cies regarding the use of side rails should be taken into account. Whether and how to use side rails is a
decision that should be based on each patient’s individual needs and should be made by the physician,
operators, and responsible parties.
• When cleaning the support surface, ensure that no liquid is allowed to seep into the zipper area and
watershed cover barrier (underside); uids allowed to come in contact with the zipper may leak into the
support surface.
• Do not expose the mattress to excessive moisture. Personal injury or equipment damage could occur.
• The use of quaternaries containing glycol ethers and/or accelerated hydrogen peroxides may compro-
mise the cover integrity and legibility.
• Be aware of devices or equipment placed on the top of the support surface. Damage to the surface
may occur due to the weight of the equipment, heat generated by the equipment, or sharp edges on the
equipment.
• Do not put overlays or accessories inside the cover. Doing so may reduce pressure redistribution
performance.
• It is the responsibility of the caregiver team to evaluate the appropriate CPR protocol to be used with the
surface.
• If there is a possibility of electro-magnetic interference with mobile phones, please increase the distance
(3.3m/10.8 feet) between devices or turn off the mobile phone.
• Ensure the waterproof cap to the power switch is present and unbroken before use. Failure to do so could
increase risk of electric shock.
• Mattress contains metal snap buttons and delrin zippers and should not be exposed under X-rays entirely.
Always use X-ray cassette during portable X-ray procedure.
NOTE
The EOLE DC support surface must be used with a mattress cover at all times. The support surface cover
may interact with all external skin.
Summary of Safety Precautions

10 836002-5210 V3.4 www.stryker.com
Return To Table of Contents
CONTROL UNIT FRONT
2
1
3
◄Figure 2
1. Power Switch On/Off
2. Front Panel
3. Power Socket
CONTROL UNIT REAR
4
5
◄Figure 3
4. Hanger
5. CPC Connector
◄Figure 4
6. EOLE DC Mattress
7. Air Cell
8. Air Hose
9. CPR Strap
CONTROL PANEL
10 11 12 13 14 15
◄Figure 5
10. Lock/Unlock
11. Mode Selection
12. Comfort Level
13. Maxrm/Seat mode
14. Alarm
15. Mute
Product Description

www.stryker.com 836002-5210 V3.4 11
Return To Table of Contents
INSTALLING THE CONTROL UNIT
1. Place control unit on at surface or suspend control unit on end of bed using attached hooks. See Fig-
ure 2 and Figure 3. Remove the plug to disconnect the device. Do not position the equipment such that
it is difcult to operate the disconnecting device.
2. Position the mattress on bed frame.
3. Connect the hose assembly between the mattress air cell and the control unit. Connect the adaptor
from control unit onto the air valve.
4. Plug the power cord and Maxrm/Seat mode will be inated automatically. Note: The unit will take
approximately 40 minutes to inate the mattress. Nurse can adjust the comfort level or mode with the
patient on the initial stage.
5. After installation, make sure the ap is not folding upwards to avoid uid seeping through mattress
cover.
NOTE
Make sure the control unit is suitable for the local power voltage and frequency.
6. Position patient on the mattress.
WARNING
Deate before CPR or CPR could be ineffective.
To deate mattress for CPR:
When there is an emergency to perform CPR on the patient, quickly pull the CPR strap from the mattress
to release air. The quick connector on the pump unit can be disconnected for even faster deation. The
air cell will deate in approximately 15 seconds. Proceed with CPR procedures.
Resetting CPR:
After CPR, re-plug CPR and make sure the CPR plug is xed on the mattress.
Instructions

12 836002-5210 V3.4 www.stryker.com
Return To Table of Contents
PRODUCT FUNCTIONS
THERAPY
1. Maxrm / Seat mode
When connected to the power for the rst time, the control unit automatically inates to maximum ination
and the indicator light of Maxrm/Seat comes on. This insures the control unit is able to reach its maximum
operating pressure. Once the maximum pressure level is reached, the pump will automatically switch
into alternating mode. User can also use this function as full mattress ination while ingress/egress the
patient for better support. Nurse or professional operator can adjust the comfort level manually at the
maxrm stage.
On the alternating or static mode, nurse can operate the maxrm button to implement the maxrm or
return to the former stage.
a. Alternate mode
In Alternate therapy mode, the mattress system will alternate every 12 minutes. User can select for
best comfort.
b. Comfort level:
Press and to adjust the pressure level for the patient’s comfort.
c. Static mode
Press THERAPY button to suspend alternating function, if needed. The pressure inside of air cells will
be adjusted to the same softness. Press the THERAPY button again; it will switch back to alternating
mode. Under the static mode, cell pressure level will be lowered compare to the same pressure level
from alternating mode.
2. Alarm Mute
Press alarm mute to deactivate the alarm sound. If the problem continues, the alarm sounds again after
3 minutes.
a. Power Failure Alarm
During power failure situation, the Power Failure indicator will light on with sound. Upon power
restoration, press the power switch to disable the audible and visual alarm and LED.
b. Low Pressure Alarm
The audible low pressure alarm is not active during initial mattress ination. The audible alarm will be
active after approximately 50 minutes has elapsed from the time the unit has been turned on.
If there is a loss of mattress pressure with the unit ON and the alarm switch activated, an alarm will
sound and ash intermittently. In addition, the low pressure light will be illuminated.
c. Service Alarm
This feature will light during mechanical failure situation. User can notify the technician for repair.
3. Lock
Patient or caregiver can hold the lock button 3 seconds to activate or deactivate lock mode. In lock mode,
Patient or caregiver can press Maxrm/Seat button for maximum ination.
The panel of control unit will be lock automatically without any operation after 3 minutes.
Instructions

www.stryker.com 836002-5210 V3.4 13
Return To Table of Contents
TRANSPORT MODE
In case of power failure or transport: Disconnect the CPC connector and
interconnect Male to Female part of the air hose connector to slow deation.
For transport purpose, interconnect Male to Female part of the air hose
connector. When a “click” is felt or heard, the connection is completed and
secured; then air from mattress is sealed off.
STORAGE
1. To quickly deate the mattress for storage, take off the CPR
strap and the CPC connectors. It will make the air release
quickly.
2. Lay the mattress out at and upsides down.
3. Fold in half and place the control unit inside.
4. Roll from the head end towards the foot end.
5. The power cord could be wrapped around the pump bumper on
the back of pump.
6. Place the whole system into the carrying bag.
Instructions

14 836002-5210 V3.4 www.stryker.com
Return To Table of Contents
Cleaning and Disinfection
The control unit housing, tubing, and mattress should be cleaned between patients.
• To clean, use water and a clean cloth to wipe down the Control Unit, power cord, hoses, mattress top
cover, and bottom cover. Do not use abrasive cleaners on the mattress. Note: Blood and other body
uids must be thoroughly cleaned from all surfaces before applying disinfectants.
• Apply disinfectants to the external surfaces of the control unit, hoses and mattress top cover, and bot-
tom cover by wiping. Stryker recommends a chlorine-based solution with a concentration less than or
equal to 1000 ppm or 70% alcohol twice a week.
• To wash the top cover of mattress by washing machine with normal process under the temperature
60°C in 45 minutes.
• It is not recommended to disinfect the internal parts of the mattress on a regular basis, but only as
needed for particular instance, the air cell could be wiped with a cloth and disinfectants as recommend-
ed above.
• Wipe down the mattress with a clean, dry cloth to remove any excess of disinfectant.
• If other detergent or other cleaning agent is used, choose one that will not have adverse chemical ef-
fects on the surface of the plastic case of the control unit, mattress cover and any other component of
the device.
• When cleaning or disinfecting the support surface, ensure that no liquid is allowed to seep into the zip-
per area and watershed cover barrier (underside); uids allowed to come in contact with the zipper may
leak into the support surface.
• Avoid dust and proximity to dusty areas.
• All components should be air dried thoroughly before use.
The waterproof cap of power switch should be on the power switch.
• Avoid using sharp tools on the waterproof cap over the power switch.
• Please reply to your distributor if the cap is broken or missed off.
WARNING
• Do not use phenolic based products for cleaning.
• Do no dry the mattress in direct sunlight.

www.stryker.com 836002-5210 V3.4 15
Return To Table of Contents
Troubleshooting
Problem Solution
Loss of power Check if the plug is connected to mains.
Low pressure alarm noises
1. Check if the CPR is sealed.
2. Check if the air cell is broken.
3. Check if the connection tube is tightly secured.
4. Check if there is any leakage on air cells.
Patient is bottoming out Pressure setting might be inadequate for the patient. Adjust comfort range
1 to 2 levels higher and wait for a few minutes for best comfort.
Air Mattress is not secure
1. Check if all the snap buttons or straps of mattress are all securely
fastened.
2. Check if the mattress is xed to the bed frame by elastic straps.
Air cells fail to inate
Make sure the air hose is not kinked, cracked, or split. Verify that the power
switch is illuminated, signifying the control unit has power. Verify that the
air hoses are fully inserted with a positive connection.

16 836002-5210 V3.4 www.stryker.com
Return To Table of Contents
TOP COVER REPLACEMENT
Tools Required: None
Procedure:
1. Disconnect the hose assembly between the mattress air cell and control unit.
2. Unzip the top cover.
3. Discard the old cover.
4. Place the new cover.
5. Carefully zip the cover.
6. Verify proper operation of the unit before returning it to service.
AIR CELL REPLACEMENT
Tools Required: None
Procedure:
1. Disconnect the hose assembly from the air valve of mattress.
2. Unzip 2-way zipper from either way to remove the top cover and remove the CPC tubes.
3. Remove and discard the old air cell.
4. Place the new air cell, connect the tubes and zip the cover to close.
5. Verify proper operation of the unit before returning it to service.
CONTROL UNIT REPLACEMENT
Tools Required: None
Procedure:
1. Disconnect the plug from mains power and hose.
2. Discard the old control unit.
3. Place the new control unit and connect the plug to mains power and hose.
4. Verify proper operation of the unit before returning it to service.
HOSE REPLACEMENT
Tools Required: None
Procedure:
1. Disconnect the hose from control unit and mattress.
2. Discard the old hose.
3. Connect the new hose to control unit and mattress.
4. Verify proper operation of the unit before returning it to service.
CPC TUBE REPLACEMENT
Tools Required: None
Procedure:
1. Disconnect the tube from control unit and mattress.
2. Discard the old tube.
3. Connect the new tube to control unit and mattress.
4. Verify proper operation of the unit before returning it to service.
FILTER REPLACEMENT
Tools Required: None
Procedure:
1. Discard the old lter.
2. Verify proper operation of the unit before returning it to service.
WARNING
Any replacement of non-authorized or wrong parts may cause the unpredictable risk rise. Please check the
replaced part is suitable for Stryker Medical’s EOLE DC Powered Support Surface, Model 2871.
Service Information

www.stryker.com 836002-5210 V3.4 17
Return To Table of Contents
Preventative maintenance should be performed annually, at a minimum. A preventative maintenance
program should be established for all Stryker Medical equipment. Preventative maintenance may need to be
performed more frequently based on the usage level of the product.
CHECKLIST
_______ Cover zipper opens and closes properly and has no visible damage.
_______ No tears, rips, holes, cracks, or other openings in the mattress cover.
_______ Check labels for legibility, proper adherence, and integrity.
_______ Support surface cover straps and snaps are intact and are not damaged.
_______ Straps properly secure the support surface assembly to the crib.
_______ Components have not degraded or come apart.
_______ Check main power cord and do not plug if there is an abrasion or excessive wear.
_______ Check airow from the air hose.
_______ Check the air hose if there is kink or breaks
_______ Verify proper operation of the unit before returning it to service.
_______ Check the waterproof cap of power switch.
Preventive Maintenance
Product Serial Number:
Completed by: ___________________________________________ Date: _____________________

18 836002-5210 V3.4 www.stryker.com
Return To Table of Contents
The parts and accessories listed on this page are currently available for purchase. Some of the parts identied
on the assembly drawing parts in this manual may not be individually available for purchase. Please call
Stryker Customer service USA at 1-800-327-0770 for availability and pricing.
Product Part Number
EOLE DC POWERED SUPPORT SURFACE 32” (80cm) 2871-000-002
EOLE DC POWERED SUPPORT SURFACE 35” (90cm) 2871-000-001
EOLE DC Control Unit 2871-001-000
Service Part Name Part Number
Manual, EOLE DC 2871-009-001
Mattress, Top Cover 32” (80cm) 2871-019-006
Single Air Cell, Orange PU 32” (80cm) 2871-019-007
Single Air Cell, Clear 32” (80cm) 2871-019-008
Mattress, Top Cover 35” (90cm) 2871-002-000
Single Air Cell, Orange PU 35” (90cm) 2871-004-001
Single Air Cell, Clear 35” (90cm) 2871-004-002
Air hose, PVC, EOLE DC 2871-004-003
Plug, Replacement, QTY 1 2871-004-004
Manifold 2871-004-005
Pump, Button Overlay, EOLE DC 2871-001-001
Pump, Fuse 2870-001-002
Tube, CPC 2870-001-003
Pump, Compressor 2870-001-004
Air Filter 2870-001-005
Accessory Part Number
Pump, UK Plug 2870-019-001
Transport Bag 2870-019-002
Pump, FR Plug 2870-019-003
Quick Reference Replacement Parts

www.stryker.com 836002-5210 V3.4 19
Return To Table of Contents
Appendix A: EMC Information
GUIDANCE AND MANUFACTURER’S DECLARATION- ELECTROMAGNETIC EMISSIONS:
This device is intended for use in the electromagnetic environment specied below. The user of this device
should make sure it is used in such an environment.
Emissions Test Compliance Electromagnetic Environment-Guidance
RF emissions
CISPR 11 Group1 The device uses RF energy only for its internal
function. Therefore, its RF emissions are very
low and are not likely to cause any interference in
nearby electronic equipment
RF emissions
CISPR 11 Class B
Harmonic emissions
IEC61000-3-2 Class A The device is suitable for use in all establishments,
including domestic establishments and those
directly connected to the public low-voltage power
supply network
Voltage uctuations / Flicker
emissions
IEC61000-3-3
Complies
WARNING
1. The device should not be used adjacent to or stacked with other equipment. If adjacent or stacked use
is necessary, the device should be observed to verify normal operation in the conguration in which it will
be used.
2.Use of accessories, transducers and cables other than those specied or provided by the
manufacturer of this equipment could result in increased electromagnetic emissions or decreased
electromagnetic immunity of this equipment and result in improper operation.
3. Portable RF communications equipment (including peripherals such as antenna cables and external
antennas) should be used no closer than 30 cm (12 inches) to any part of the Pump, including cables
specied by the manufacturer. Otherwise, degradation of the performance of this equipment could result.

20 836002-5210 V3.4 www.stryker.com
Return To Table of Contents
Appendix A: EMC Information
GUIDANCE AND MANUFACTURER’S DECLARATION- ELECTROMAGNETIC IMMUNITY:
This device is intended for use in the electromagnetic environment specied below. The user of this device
should make sure it is used in such an environment.
Basic EMC
standard
Immunity Test Levels Compliance
Levels
Electromagnetic
Environment-Guidance
HOME HEALTHCARE
ENVIRONMENT
Electrostatic
Discharge (ESD)
IEC61000-4-2
±8kV contact
±15kV air
±8kV contact
±15kV air
Floors should be wood, concrete or
ceramic tile. If oors are covered
with synthetic material, the relative
humidity should be at least 30 %.
Electrical fast
transient/ burst
IEC61000-4-4
±2kV for power supply line
±1kV for input/output line
±2kV for power
supply line
±1kV for input/
output line
Mains power quality should be that
of a typical commercial or hospital
environment
Surge
IEC61000-4-5
± 1 kV line(s) to
line(s)
± 2 kV line(s) to earth
± 1 kV line(s) to
line(s)
Mains power quality should be that
of a typical commercial or hospital
environment.
Voltage dips,
short interruptions
and voltage
variations on
power supply
input lines
IEC61000-4-11
Voltage Dips:
i) 100% reduction for 0.5
period,
ii) 100% reduction for 1
period,
iii) 30% reduction for 25/30
period,
Voltage Interruptions:
100% reduction for 250/300
period
230V (UT) (1)
Voltage Dips:
i) 100%
reduction for 0.5
period,
ii) 100%
reduction for 1
period,
iii) 30%
reduction for
25/30 period,
Voltage
Interruptions:
100% reduction
for 250/300
period
Mains power quality should be that
of a typical commercial or hospital
environment. If the user of this
device requires continued operation
during power mains interruptions,
it is recommended that the device
be powered from an uninterruptible
power supply or a battery.
Power frequency
(50/60Hz)
magnetic eld
IEC61000-4-8
30 A/m 30 A/m Power frequency magnetic elds
should be at levels characteristic
of a typical location in a typical
commercial or hospital environment.
Conducted RF
IEC 61000-4-6
3 Vrms
0,15 MHz – 80 MHz
6 Vrms in ISM and amateur
radio bands between
0,15 MHz and 80 MHz
80 % AM at 1 kHz (4)
6Vrms Portable and mobile RF
communications equipment should
be used no closer to any part of this
device, including cables, than the
recommended separation distance
calculated from the equation
applicable to the frequency of the
transmitter.