SUPMEA U-SUP-EC8.0-EN2 User manual

User's Manual
U-SUP-EC8.0-EN2
Conductivity controller
Supmea Automation Co.,Ltd.
www.supmea.com
Headquarters
5th floor,Building 4,Singapore Hangzhou Science Technology Park,No. 6 street,
Hangzhou Economic Development Area,Hangzhou 310018,China
Singapore
2 Venture Drive #11-30 Vision Exchange Singapore

I
Preface
Thank you for purchasing conductivity controller. Please read this manual carefully
before operating and using it correctly to avoid unnecessary losses caused by
false operation.
Note
Modification of this manual’s contents will not be notified as a result of
some factors, such as function upgrading.
We try our best to guarantee that the manual content is accurate, if you find
something wrong or incorrect, please contact us.
This product is forbidden to use in explosion-proof occasions.
Version
U-SUP-EC8.0-EN2

II
Safety Precautions
In order to use this product safely, be sure to follow the safety precautions
described.
About this manual
Please submit this manual to the operator for reading.
Please read the operation manual carefully before applying the instrument.
On the precondition of full understanding.
This manual only describes the functions of the product. The company
does not guarantee that the product will be suitable for a particular use by
the user.
Precautions for protection, safety and modification of this product
To ensure safe use of this product and the systems it controls, Please read
carefully the operation manual and understand the correct application
methods before putting into operation, to avoid unnecessary losses due to
operation mistakes. If the instrument is operated in other ways not
described in the manual, the protections that the instrument give may be
destroyed, and the failures and accidents incurred due to violation of
precautions shall not be borne by our company.
When installing lightning protection devices for this product and its control
system, or designing and installing separate safety protection circuits for
this product and its control system, it needs to be implemented by other
devices.
If you need to replace parts of the product, please use the model
specifications specified by the company.
This product is not intended for use in systems that are directly related to
personal safety.Such as nuclear power equipment, equipment using
radioactivity, railway systems, aviation equipment, marine equipment,
aviation equipment and medical equipment.If applied, it is the responsibility

III
of the user to use additional equipment or systems to ensure personal
safety.
Do not modify this product.
The following safety signs are used in this manual:
Hazard, if not taken with appropriate precautions, will result in serious
personal injury, product damage or major property damage.
Warning:Pay special attention to the important information linked to
product or particular part in the operation manual.
Confirm if the supply voltage is in consistent with the rated voltage
before operation.
Don’t use the instrument in a flammable and combustible or steam area.
To prevent from electric shock, operation mistake, a good grounding
protection must be made.
Thunder prevention engineering facilities must be well managed: the
shared grounding network shall be grounded at is-electric level,
shielded, wires shall be located rationally, SPD surge protector shall be
applied properly.
Some inner parts may carry high voltage. Do not open the square panel
in the front except our company personnel or maintenance personnel
acknowledged by our company, to avoid electric shock.
Cut off electric powers before making any checks, to avoid electric
shock.
Check the condition of the terminal screws regularly. If it is loose, please
tighten it before use.
It is not allowed to disassemble, process, modify or repair the product
without authorization, otherwise it may cause abnormal operation,
electric shock or fire accident.
Wipe the product with a dry cotton cloth. Do not use alcohol, benzine or

IV
other organic solvents. Prevent all kinds of liquid from splashing on the
product. If the product falls into the water, please cut off the power
immediately, otherwise there will be leakage, electric shock or even a
fire accident.
Please check the grounding protection status regularly. Do not operate if
you think that the protection measures such as grounding protection and
fuses are not perfect.
Ventilation holes on the product housing must be kept clear to avoid
malfunctions due to high temperatures, abnormal operation, shortened
life and fire.
Please strictly follow the instructions in this manual, otherwise the
product's protective device may be damaged.
Don’t use the instrument if it is found damaged or deformed at opening
of package.
Prevent dust, wire end, iron fines or other objects from entering the
instrument during installation, otherwise, it will cause abnormal
movement or failure.
During operation, to modify configuration, signal output, startup, stop,
operation safety shall be fully considered. Operation mistakes may lead
to failure and even destruction of the instrument and controlled
equipment.
Each part of the instrument has a certain lifetime, which must be
maintained and repaired on a regular basis for long-time use.
The product shall be scrapped as industrial wastes, to prevent
environment pollution.
When not using this product, be sure to turn off the power switch.
If you find smoke from the product, smell odor, abnormal noise, etc.,
please turn off the power switch immediately and contact us in time.

V
Disclaimer
The company does not make any guarantees for the terms outside the
scope of this product warranty.
This company is not responsible for damage to the instrument or loss of
parts or unpredictable damage caused directly or indirectly by improper
operation of the user.
No.
Name
Quantity
Note
1
Conductivity controller
1
2
Manual
1
3
Certificate
1
After opening the box, please confirm the package contents before starting the
operation.If you find that the model and quantity are incorrect or there is physical
damage in appearance, please contact us.

VI
Contents
Chapter 1 Product introduction......................................................................................... 1
Chapter 2 Parameter.......................................................................................................... 2
Chapter 3 Instrument installation......................................................................................3
Chapter 4 Wiring................................................................................................................. 5
Chapter 5 Electrode connection figure............................................................................ 5
Chapter 6 Relay contact protection..................................................................................6
Chapter 7 Display................................................................................................................8
Chapter 8 Key function.......................................................................................................9
Chapter 9 HOLD Type..................................................................................................... 10
Chapter 10 Setting............................................................................................................11
Chapter 11 Default............................................................................................................22
Chapter 12 Password.......................................................................................................25
Chapter 13 Error code......................................................................................................25
Chapter 14 RS485 communication................................................................................ 26
Chapter 15 Warranty & After-sales Service..................................................................30

Chapter 1 Product introduction
www.supmea.com
1
Chapter 1 Product introduction
The product package contains 1 controller, 1 manual.
Controllers are used in industrial measuring of the temperature, conductivity,
Resistivity, salinity and total dissolved solids, such as waste water treatment,
environmental monitoring, pure water, sea farming, food production process, etc.
The instrument can be panel, wall or pipe mounted.
The instrument provides two current outputs. The maximum load is 500Ω.
The instrument provides 3 relays. It can pass though a maximum of 5A/250VAC or
5A/30VDC.

Chapter 2 Parameter
www.supmea.com
2
Chapter 2 Parameter
Functions
EC
Resistivity
Salinity
TDS
Measuring range
0.00uS-2000
mS
0.00-20.00
MΩ-CM
0.00-78.00
g/Kg
0-133000
ppm
Resolution
0.01/0.1/1
0.01
0.01
1
Accuracy
±1%F.S.
±1%F.S.
±1%F.S.
±1%F.S.
Temp. compensation
Pt1000/NTC30K
Temp. range
-10.0 to +130.0℃
Temp. compensation range
-10.0 to +130.0℃
Temp. resolution
0.1℃
Temp. accuracy
±0.2℃
Cell constant
0.001 to 20.000
Ambient temp. range
0 to +70℃
Storage temp.
-20 to +70℃
Display
Back light, dot matrix
EC current output1
Isolated, 4 to 20mA output , max. load 500Ω
Temp. current output 2
Isolated, 4 to 20mA output , max. load 500Ω
Current output accuracy
±0.05 mA
RS485
Modbus RTU protocol
Baud rate
9600/19200/38400
Maximum relay contacts
capacity
5A/250VAC,5A/30VDC
Cleaning setting
ON: 1 to 1000 seconds, OFF: 0.1 to 1000.0 hours
Multifunction relay
clean/period alarm/error alarm
Relay delay
0-120 seconds
Data logging capacity
500,000
Language selection
English/traditional Chinese/simplified Chinese
Waterproof grade
IP65
Power supply
90~260 VAC, power consumption < 7 W
Installation
panel/wall/pipe installation
Weight
0.85Kg

Chapter 3 Instrument installation
www.supmea.com
3
Chapter 3 Instrument installation
The instrument can be panel, wall or pipe mounted installation.
Panel Installation: Make a 138x138 mm square cutout and insert the instrument.
Screw in the fixed block with the screws and fixed bar.
Cutout dimension
Overall dimension

Chapter 3 Instrument installation
www.supmea.com
4
Installation figure

Chapter 4 Wiring
www.supmea.com
5
Chapter 4 Wiring
Wiring label
Chapter 5 Electrode connection figure
4-electrode
2-electrode

Chapter 6 Relay contact protection
www.supmea.com
6
Chapter 6 Relay contact protection
Electrical spark at the relay contact may affect the life of the relay, especially in an
inductive and capacitive load. In order to inhibit the spark and arc, user should use
an RC circuit to extend the life of the relay.
AC protection, use for inductive load
1. load
2. RC eliminate spark, using in 220VAC, R=100 ohm1W,C=0.1uF630V
3. Relay contact

Chapter 6 Relay contact protection
www.supmea.com
7
DC protection ,
A1: inductive load A2: 1N4007
A3: relay contact
AC/DC protection ,
B1: capacitive load
B2: 0.8 Ohm/1W(DC24V)
B3: relay contact
Resistive load ,
C1:lamp bulb
C3:relay contact

Chapter 7 Display
www.supmea.com
8
Chapter 7 Display
1. Main display
2. Unit
3. Temperature and unit
4. Date and time
5. Temp. compensation: auto(ATC) or manual(MTC)
6. First current output
7. Second current output
8. Relay 1
9. Relay 2
Note:
If the EC readings are over the range, it will display 99.99/999.9/9999
If the temperature readings are under or over the range, it will display -99.9/999.9.

Chapter 8 Key function
www.supmea.com
9
Chapter 8 Key function
Key name
Meas. status
Setting status
Cal. status
Record status
MODE
Enter password
Exit
Exit
Exit
SHIFT
None
Move digit
Move digit
Move digit
UP
Enter record
Inc
Inc
Inc
DOWN
None
Dec
Dec
Dec
ENTER
ON/OFF back
light
Enter
Enter
Enter

Chapter 9 HOLD Type
www.supmea.com
10
Chapter 9 HOLD Type
HOLD Type is a safe mode. It is for Calibration, Setting, Record and Clean. In this
mode all the relays are open( inactive), current output follows the setting by
user(last current or fixed current).
The instrument will enter HOLD Type when user presses into Calibration, Setting,
Record or the instrument works in clean mode.
It will in HOLD Type around 10 seconds when it goes back to measurement mode
form the above mentioned 4 modes then left HOLD Type.
The instrument will go into the HOLD Type when turn on the power. Current output
in HOLD Type:
User has two choices: fixed current output or last current output.
Fixed current:
User can set the output current from 4.00 to 20.00mA when instrument goes
into HOLD Type.
Last current:
User can set the output current keep at the last current when instrument
goes into HOLD Type.
Relays in HOLD Type: All relays are opened.(inactive)

Chapter 10 Setting
www.supmea.com
11
Chapter 10 Setting
Press MODE key to enter the password menu and then press UP/DOWN/SHIFT
key to input password 2008 then press ENTER will enter to setting mode or press
MODE key to exit. If no key is pressed and over 10 minutes then it will go back to
measurement mode.
10.1 Main display
Press UP/DOWN key to choose functions, press ENTER key enter the function.
Page1
Page2
Note:
1. When the input data is not in correct range then it will display ERROR on
the top of LCD
2. After input data user needs to press ENTER to save the data.
3. Press MODE to exit.
4. No key is pressed in 10 minutes then it will go back to measurement mode.

Chapter 10 Setting
www.supmea.com
12
10.2 Current 1 settings
1. Set the corresponding 4.00mA to EC/ Resistivity/Salinity/TDS.
2. Set the corresponding 20.00mA to EC/ Resistivity/Salinity/TDS, the
difference between 4.00mA and 20.00 mA at least for EC is
1.00/10.0/100(uS/mS), for Resistivity is 1.00MΩ,for Salinity is 1.00g/kg,for
TDS is 100 ppm.
3. Set the offset current ,the range is± 1.00mA.
4. The filter time range is 0-120 seconds, the low pass filter of software will
active when the current from one point to another point if user sets the filter
time.
5. Set the current 1 output mode(fixed / last) when instrument enter into HOLD
Type.
10.3 Current 2 settings
1. Set the corresponding 4.00mA to temperature.
2. Set the corresponding 20.00mA to temperature, the difference at least
between 4.00mA and 20.00mA is 10.0℃.

Chapter 10 Setting
www.supmea.com
13
3. Set the offset current of temperature, the range is±1.00mA.
4. The filter time range is 0-120s, the low pass filter of software will active
when the current from one point to another point if user sets the filter time.
5. Set the current 2 output mode(fixed / last) when instrument enter into HOLD
Type.
10.4 Relay 1 settings
1. Press UP/DOWN key to ON/OFF (enable/ disable) relay1.
2. Close set point: active point for EC/ Resistivity/Salinity/TDS.
3. Open set point: inactive point for EC/ Resistivity/Salinity/TDS.
4. Delay time: the range is 0-120s. Relay needs to delay first then active if the
measuring data is reach to close set point.
5. For example: If user wants turn on the pump at EC 1000uS and turn off it at
EC 400uS, then the close S.P. needs to set to 1000uS, Open S.P. sets to
400uS.
10.5 Relay 2 settings
1. Press UP/DOWN key to ON/OFF (enable/ disable) relay2.
Table of contents
Other SUPMEA Controllers manuals
Popular Controllers manuals by other brands

WAGO
WAGO 8003 Series Operating and assembly instructions

Delta Electronics
Delta Electronics AC Motor Drive VFD-EL user manual

Mitsubishi Electric
Mitsubishi Electric Q172DCPU user manual

Popp
Popp POPE009204 manual

Sonnet
Sonnet Tempo edge SATA Pro 6Gb ExpressCard/34 user guide

Smartgen
Smartgen MGC120 user manual

HomeMatic
HomeMatic HM-LC-Sw4-SM-2 Installation and operating manual

Ecus
Ecus POWER_AVR SVR TRI 3/3 user manual

Murphy
Murphy PowerCore MGC1000 Operation manual

AMX
AMX Control Card RDC-PDC Installation sheet

Sturtevant Richmont
Sturtevant Richmont TCV operating instructions

DT Research
DT Research DT1363 Basic operation guide