13
6. Commissioning
At the first commissioning of the M-Sens 2 it is necessary to calibrate the sensor.
Please consider:
•Correct connection between sensor, C1-box and transmitter.
•Correct installation of the sensor in respect to the wall thickness.
In case of negative results in spite of consideration of the points as stated above, please contact SWR.
Commissioning of the M-Sens 2
After the delivery the sensor is not calibrated to the product(s), so each calibration and parameterization
has to be executed during the commissioning. Therefore it is necessary to assign the measured moisture to
the desired display and to the initial value. The menu functions are very self-explanatory. In the following a
short introduction to the menu overview:
Put the installation CD in the computer and follow the installation menu. The program is compatible with
Windows operating systems 98, NT, XP, Vista, Win 7, Win 8 and Win 10. The connection can take place via
the USB interface on the front plate, using the supplied cable, or via the integrated RS 485 interface (bus-
compatible) on the screw terminals 11 and 12. By allocating different addresses to the analysis unit, the
devices can be addressed individually in the bus using the ModBus protocol. A brief introduction follows to
serve as an overview.
All changed values are adopted by leaving the menu level and confirming the save function.
Entering the After starting the M-Sens 2 configuration program, the COM1 to COM8
menu interface must be selected on the separate PC.
Set baud rate to 9600 Bd as a fixed value. Set address of the analysis unit
(standard = 1).
Basic function For the measurement using the sensor, at least 2 measuring points should be
calibrated as a reference point and the corresponding humidity value should
be entered. To do this, the reference point measurement is started in the
calibration. After the measurement, the analysis device expects the input of the
humidity content determined in the lab. While the material is conveyed (in the
loading state), the 2nd point is set to a known humidity value and the calibration
is also carried out. If only one material humidity value is available, then minor
humidity fluctuations and corresponding fluctuations in the humidity values
determined in the lab can be used to save at least 2 calibration points.
Adjustment For the adjustment of user requirements with regard to material,
measuring units etc., sub-menu items 1 to 3 are available in menu 1.
Alarms The user can select operation modes “Fault” or “Alarm” in the menu Alarms (2).
A shortfall or exceeding of an adjustable measuring value can be output via the
alarm relay as “open” or “close”.