TabTimer Careousel MK3/11 User manual

TabTimer® Careousel MK3/11
Automac Pill Dispenser
Instrucons
Note: When opening for the rst me, the lid is unlocked.
The key and baery case securing screw can be found inside the dispenser.

2
TABTIMER® CAREOUSEL MK3/11
AUTOMATIC PILL DISPENSER
The pack also includes:
Pharmacy label
carrier and cover
Three reversible Day /
Time discs Key
Baeries
4 x AA Alkaline
Baery case
screw
Lock
Dispenser Tray
Lid
Display
Buons 1,2,3
Apply Instrucon label
Figure 1.
+
-
+
+
-
-
-
+

3
INTRODUCTION
The TabTimer® Careousel MK3/11 ® Automac Pill Dispenser reminds the user by means of an
alarm and ashing light when to take their medicine. It will make the correct dose available at
the correct me of day or night, whilst keeping other pills locked out of sight. Medicine is
dispensed from an internal tray containing 29 secons. Maximum number of daily alarm
sengs - 28.
When the alarm sounds the internal tray will rotate one posion. The correct dosage will
come into view in the lid opening. Pills should be removed from the dispenser by lng it and
pouring the pills into the hand or suitable container. This will also cancel the alarm signal and
the ashing light.
Technical data:
Dimensions - length 190mm, width 180mm, height 56mm
Weight - 380g excl. baeries, 480g incl. baeries
Compartment volume - approx. 10ml per compartment
Power supply - 6V (4 x 1.5V Alkaline baeries)
Baery life - up to 1 year
Open the lid as illustrated. Place an index
nger in the lid opening and press in the
catch with a thumb. Li the lid.
The key and baery case securing screw
can be found in one of the pill
compartments.
Open the lid to the baery box
(underneath the dispenser) and
acvate the dispenser by removing the
baery isolang strip. (Note the correct
baery polarity + / - )
USING THE DISPENSER FOR THE FIRST TIME Figure 2.
Figure 3.

4
FILLING THE DISPENSER
1. Time Discs — three day/me discs are supplied which comprise:
Two printed discs for between 1 - 4 medicine prompts per day (use either side) and one
blank disc for more than 4 medicine prompts per day. The discs should be completed by hand
indicang the medicine mes*. Select the appropriate disc for the number of doses per day.
*NB: Before lling out the day/me discs, calculate which will be the date and me of the
user’s rst dose with the newly lled medicine tray (parcularly if a pharmacy will rell the
tray) and start to the right of the red secon (see gure 4. below). Place the disc on top of
the tray. Secure the disc in place by bending down the four tabs (a spot or two of glue will
secure it further if preferred).
2. Medicine Remove the tray from the dispenser. Starng from the right of the solid red
secon (do not ll this secon), place the correct dosage in each secon of the tray.
Note: if a pill is inadvertently dropped in an incorrect compartment,
a pair of tweezers can be helpful to easily remove the pill.
Figure 5.
3. Replace the tray in the dispenser
Place the tray in the dispenser with the red secon showing
in the lid opening, so that when the lid is closed the rst
dose to be taken is to the right of the lid opening. At the
next alarm me the tray will rotate one secon clockwise
and make available the correct medicine. (See gure 4.).
Note: To avoid damaging the mechanism, the medicine
tray must always be reposioned by liing it out and then
lowering directly into the case in the correct posion.
Do not force the tray to rotate.
4. Complete the pharmacy label carrier
Remove the re-posable adhesive pad cover on the
back and sck the pharmacy label carrier to the
baery case cover. Slide the plasc cover into
posion to protect it. (See gure 5.).
Note: Re-usable spare trays with lids and me discs
can be purchased to allow pre-lling.
First Dose
Figure 4.

5
Menu A (accessed through Buon 1)
Set CLOCK:
1. Press Buon 1 repeatedly unl ‘CLOCK’ is visible on the display.
2. Set the correct me by using Buons 2 and 3. You can fast forward or reverse the me by
pressing and holding down these buons.
Set DOSES per day:
1. Press Buon 1 repeatedly unl ‘DOSES’ is visible on the display.
2. Wait a few seconds. Adjust to the correct number of doses per day using Buons 2 and 3.
Set ALARM mes:
1. Press Buon 1 repeatedly unl ‘ALARM’ is visible on the display.
2. Set the 1st (1) alarm me using buons 2 and 3.
3. Press Buon 1 again for the 2nd (2) alarm me and repeat the procedure for all
subsequent alarm mes required.
Press and hold down buon 1 for a few seconds to exit the menu and revert to normal
operaon.
DISPENSER SETTINGS
The dispenser is set using 3 buons:
Buon 1 is used for moving to the next funcon.
Buons 2 and 3 are used to change the sengs in each funcon.
There are two seng menus:
MENU A:- Buon 1: DOSES ALARM CLOCK
MENU B:- Buon 3: BEFORE AFTER SOUND LIGHT HOURS DLS
DATE START RECORD TEST
NB: 1. It important that the DATE is set to allow DLS and START funcons to operate.
2. Aer compleon of sengs in either Menu A or B press buon 1 for three seconds
to restore to normal operaon. If nothing is done the display will automacally
return to normal aer 30 seconds.

6
MENU B (Accessed through Buon 3)
To access the funcons in this Menu press and hold down buon 3 unl ‘BEFORE’ is
displayed on the screen. The individual funcons in this menu can then be accessed by
pressing buon 1 repeatedly. At any me Hold Buon 1 down to return to clock.
‘BEFORE’ - Taking medicaon before the pre-set alarm me (EARLY DOSE)
It is possible to allow medicine to be taken up to 5 hours before the normal pre-set alarm
me. To set this funcon go to ‘BEFORE’.
Press Buons 2 or 3 to set the me required (up to 5 hours before in 5 minute increments).
If this funcon has been set the user may access their medicine in advance by lng the
dispenser upside down twice. When upright again the dispenser will then rotate one
secon, the alarm will sound, and the medicine will become available. Tilt the dispenser
again to dispense the medicine and stop the alarm.
NB: If medicine is taken before the normal me, the dispenser will not acvate again unl the
following medicine is due.
If this funcon is not required set at 00.00
We do not recommend that the BEFORE funcon is used for those with confusion.
AFTER (DURATION)
The length of me the alarm will sound / light will ash can be adjusted by the ‘AFTER’
funcon:
Press Buon 1 repeatedly unl ‘AFTER’ is displayed. Adjust the required alarm duraon by
pressing Buons 2 or 3 (minimum 5 minutes / maximum 1 hour).
SOUND
There are four alarm sounds available. To choose the alarm sound:
Press Buon 1 repeatedly unl ’SOUND’ is displayed. Choose your preferred sound
01, 02, 03, or 04 by pressing Buons 2 or 3. Should no sound be required set at 00.
The alarm will sound every 4 seconds for the rst minute, and every 30 seconds thereaer
unl medicine is taken or the alarm duraon limit is reached (see ‘AFTER’ funcon above).
LIGHT
Press Buon 1 repeatedly unl ‘LIGHT’ is displayed.
If the light is not required press buon 2 once – ’00’ will be displayed.
If the light is required press buon 3 once – ’01’ will be displayed.
HOURS - 24hr / AM-PM clock sengs.
The dispenser is pre-set to display the me as a 24hr display.
Should you wish to change this to AM/PM clock seng:
Press Buon 1 repeatedly unl ‘HOURS’ is displayed. Choose AM/PM or 24hrs by pressing
buons 2 or 3. NB. Using the 24hr clock is recommended.

7
DLS (DAYLIGHT SAVING)
To enable the dispenser to automacally adjust for daylight saving.
Press Buon 1 repeatedly unl ‘DLS’ is displayed.
Using Buons 2 or 3 set at ‘01’ to enable Daylight Saving. Set at ‘00 ‘ to disable.
DATE
Press Buon 1 repeatedly unl ‘DATE’ is displayed. Wait a few seconds. YY will be displayed.
Using buons 2 and 3: Set the Year (YY) then press 1
Using buons 2 and 3: Set the Month (MM) then press 1
Using buons 2 and 3: Set the Date (DD) then press 1
START (START DELAY FUNCTION)
Aer programming the doses and alarm mes it is possible to delay the me at which the
dispenser will start to operate. This funcon is useful if the dispenser is being lled osite, for
example, by a pharmacy and not being delivered to the user unl a later date.
Press Buon 1 repeatedly unl ‘START’ is displayed.
Wait a few seconds. YY will displayed.
Using buons 2 and 3: Set the Year (YY) then press 1
Using buons 2 and 3: Set the Month (MM) then press 1
Using buons 2 and 3: Set the Date (DD) then press 1
Using buons 2 and 3: Set the Hour (HH) then press 1.
The me (HH) should be in advance of the rst dose to be dispensed for the user.
If this funcon is not required do not set.
Note: If at any me you go back to check the START seng at a later me the display will show
the pre-set DATE sengs. The START seng will not be displayed and it should be re-set.
RECORD
The dispenser will record the actual mes the previous 28 medicines were dispensed.
Press Buon 1 repeatedly unl ‘RECORD’ is displayed. The me that the last medicine was
taken is displayed. Press Buons 2 or 3 for all of the mes medicine was dispensed over the
previous medicine cycle to be shown.
TEST FUNCTION
The dispenser has an inbuilt test funcon. To test the rotaon and alarm sounds:
Press Buon 1 repeatedly unl ‘TEST’ is displayed.
Press Buon 3 – the tray will rotate, ‘MOVING’ and then ‘TAKE’ is displayed.
The alarm will sound. Tilt the dispenser to stop the alarm. ‘TEST’ is then displayed again.
To test the ‘BEFORE’ funcon, Press Buon 1 unl ‘TEST’ is displayed. Press Buon 2.
Close the lid and then turn the dispenser upside down twice. Turn back to the upright
posion. The tray will rotate and the alarm will sound. Tilt the dispenser again to cancel.
Hold Buon 1 down to return to clock.

8
GENERAL INFORMATION
Lock
The dispenser can be locked with the key supplied (see
gure 6.). Turn the key a quarter of a turn clockwise to
lock, an-clockwise to unlock.
Please note that the lock is a simple device intended as
an addional barrier only.
Baery Cover Screw
A screw is provided to prevent inappropriate access to the baeries (should this be
necessary). Place the screw in the hole in the baery cover securing clip and ghten.
The screw must be removed when the baeries need changing. It should then be replaced.
Medicines:
There are some pills that should not be exposed to the air and therefore should not be used
in the dispenser. If in doubt please check with your pharmacist.
Pharmacy lling: Many pharmacies will ll the TabTimer® Careousel MK3/11 Dispenser.
The device meets the requirements of the Quality Care Pharmacy Program (QCPP).
For the suggested Standard Operang Procedure for pharmacy lling of the TabTimer®
Careousel MK3/11, please ask your pharmacist to contact TabTimer Pty Ltd.
Fault handling:
The alarm will sound connuously in the event of the dispenser failing to rotate properly.
This might be caused by an obstrucon between the body of the dispenser and the internal
tray. In this event the display will show ‘ERROR’. A baery should be removed and then
immediately replaced. Any obstrucon should be removed.
If the dispenser is upside down at the alarm me, the internal tray will not move unl the
dispenser is turned upright. If the dispenser is upside down for all the me between the
‘BEFORE’ and ‘AFTER’ sengs a warning signal will sound and ‘TILTED’ will be displayed.
Press Buon 3 once to cancel. Check the posion of the internal tray to ensure that the
correct medicine due will be available at the next alarm me.
If the dispenser was upside down for the duraon of 1 or more alarms, the medicine due to
be presented at the next scheduled alarm may not be the correct one for that me of day.
(See General Observaons).
Figure 6.

9
Baeries:
Baeries (4 x AA Alkaline) are supplied with the dispenser when new.
Baeries will last in normal use for up to 12 months.
Baery life can be checked using the following procedure:
1. Insert baeries in the dispenser.
2. Immediately watch the LCD display for the indicaon BATLEV.
3. At full charge a reading of between 1550 - 1600 (or more) will be shown.
4. At a reading of 1325 the baery warning indicator ‘’LOWBATT’’ will iniate.
5. At a reading of 1275 the unit will not operate, will blink ERROR and sound alarm
sound 04 for 2 seconds, silence for 28 seconds, repeated. This is dierent to the
standard alarm which sounds for 2 seconds repeang every 2 seconds.
NB: 1. Rechargeable baeries should not be used as they have an insucient charge to
reliably operate the unit.
2. When the baery charge is reduced to a reading of 1325 the display will indicate
LOWBAT. In normal operaon the LOWBAT warning will be acve for 1-2 months
depending on how many daily doses are set. This gives early warning that a baery
change is required.
Remember when changing baeries to insert them the correct way around (see gure 3. – p.3).
Removing Baeries: If the baeries are removed and replaced immediately the alarm
sengs are retained; only the clock would need to be reset if the baeries are not replaced
immediately.
Cleaning:
Dispenser - clean external surfaces of the dispenser body and lid with a disinfectant
spray and / or wipe with a damp cloth.
Internal tray - the tray may be washed in warm water or in a dishwasher (max. 35C).
Transport: If for any reason the dispenser is despatched by post, courier or other carrier, it is
important that all baeries are removed from the dispenser.
Important - Reposioning the Tray. To avoid damaging the mechanism, the medicine tray
must always be reposioned by liing it out and then lowering directly into the case in the
correct posion. Do not force the tray to rotate.
TabTimer cannot accept responsibility for faults or any other consequences arising from
misuse e.g. dropping, water damage, inseron of foreign objects, forced turning of the pill
tray or forced opening of the dispenser.

10
GENERAL OBSERVATIONS
The dispenser is not suitable for everyone and assessment is very important.
Most problems occur due to misuse/interference.
Users who wilfully try to get at their medicine by force will eventually damage the
unit. Look out for knife marks and evidence of someone trying to force it open.
Look out for pencil/biro marks in the pill compartments. This may indicate that the
user might need a lile more training regarding pping the pills out. The dispenser
must be turned over to allow the pills to fall into the hand or suitable container. It
may be that the person lling it has put the wrong pill in and tried to get it out with
a pen or pencil. Tweezers are very useful to li out a pill incorrectly placed.
Where it is thought necessary, keep the baery cover grub screw in place to
prevent access to the baery case.
Try to keep the dispenser reasonably horizontal. If it is le upside down it will not
rotate at the next alarm me and will give out the ERROR sound (slow bleep).
On turning over it will then rotate and dispense medicine as usual (for duraon of
AFTER me). Check the posion of the internal tray to ensure that the correct
medicine will be available at the next alarm me (see also FAULT HANDLING p.8).
If DOSES are set to 00 then even if the alarms are set, it will not alarm.
This is useful if you do not want it to alarm for a period of me. Remember to
reinstate the DOSES when the alarms are required again (or use the Start Delay
Funcon. See p7.)
Spares: Spare keys, baery cover screws, pharmacy label carriers, spare trays and
me discs are all available. Please contact TabTimer® for assistance.
Figure 7.
Note: If the user repeatedly leaves the dispenser
upside down, they may nd a Careousel Dispenser
Tipper very helpful .
(see gure 7.).

11
Pharmacy lling:
Many pharmacies will ll TabTimer® dispensers.
The device meets the requirements of the Quality Care Pharmacy Program (QCPP)*.
For the suggested Standard Operang Procedure for pharmacy lling of the
MK3/11 dispenser, ask your pharmacist to contact
TabTimer® .
The new Pharmacy Filler tray is now available for
easier Pharmacy lling of dispenser trays.
Ask your Pharmacist to contact TabTimer® on 1300 822 846.
COMMONLY REPORTED ISSUES
‘I have just received my new dispenser but it is locked and there is no key.’
The unit is not locked but there is a knack to open (see Figure 2.) the key is inside.
‘The alarm sound/light has stopped working.’
Check the sound / light sengs – they may have been inadvertently set to ‘00’
‘The dispenser alarmed during the night when it should not have’.
Check the sengs, usually it has been incorrectly programmed to p.m. instead of a.m.
Set to 24 hour clock to avoid this happening.
‘The dispenser didn’t allow for the clocks going back/forward.’
The date needs to be programmed into the unit. Check DLS is set to ‘01’.
‘The unit is not rotang.’
Check that there is no obstrucon; check the baery level, check it is programmed correctly.
Use the TEST funcon in Menu B.
‘The compartments are too small I can’t get my ngers in to get the pills out, also the
alarm does not stop ringing.’
The pills should be pped out into the hand, which also stops the alarm ringing.
‘There is an intermient alarm sound and ‘ERROR’ is indicated on the screen.’
If the dispenser has been unable to rotate due to an obstrucon it will show ‘ERROR’ on the display
and emit the error alarm. This is a intermient sound with a longer interval ( 30 seconds) compared to
the normal ‘dispensing’ alarm interval. To stop alarm, take a baery out and then replace it.
Clear the obstrucon. If the ERROR message connues there may be another fault in which case
contact TabTimer® for inspecon.
‘The tray is rotang slowly.’
Check the baery level and replace if necessary. (Note- Do not use rechargeable baeries, they do not
have sucient power for extended use).
‘I have dropped the dispenser, should I send it back for checking?’
Use the TEST funcon in menu B. If all is operang sasfactorily then it is probably operang
normally. For extra reassurance reset the alarms and actually watch/listen to the rotaon.
‘If I take the baeries out will I have to reset all the alarms again?’
No, the alarm sengs are retained, only the clock would need to be reset if the baeries are not
replaced immediately.
‘The dispenser is not working and ‘TILTED’ is displayed?’
Dispenser has been le upside down at the dosage me and all of the AFTER me.
See FAULT HANDLING p.8.

12
TabTimer® Pty Ltd, PO Box 7395, ALEXANDRIA NSW 2015
T: 1300 822 846 / E: info@tabmer.com.au
www.TabTimer.com.au
Copyright: TabTimer Pty Ltd Made in Sweden.
Important Noce:
TabTimer® Automac Pill Dispensers are not a medical device, they are aids only, designed to
assist the user in remembering to take their medicine. They are not suitable for everyone, and
assessment is very important.
It is not a medicine safe or vault and is not 100% tamperproof. With determinaon a paent
may be able to break into the device and access more medicine than intended. If there is any
doubt that a paent will have an uncontrollable urge to access more medicine than prescribed,
then please do not use this device. TabTimer® has alternave devices available.
TabTimer® does not accept responsibility for faults or other consequences arising from any
misuse of the dispenser, for example, damage from dropping, moisture, liquids or heat, inser-
on of foreign objects, forced or unauthorised opening of the dispenser. As with all electronic
devices, do not leave in direct sunlight.
It is the user’s responsibility to ensure that medicine is taken as prescribed.
Full Terms and Condions are available on our website : www.TabTimer.com.au