TDT PZ2 User manual

6-3
PZ2 PreAmp
PZ2PreAmp
PZ2Overview
The PZ2 is a high channel count preamplifier
suitable for extracellular recordings. The PZ2
preamplifier features a custom 18-bit hybrid A/
D architecture that offers the advantages of
Sigma-Delta converters at significantly lower
power and a fast fiber optic connection capable
of simultaneously transferring up to 256
channels. The extended bandwidth offered by
this connection supports sampling rates up to
~50 kHz and improves signal fidelity, spike
discrimination, sorting, and analysis. Used
exclusively with Z-Series base stations, PZ2
preamplifiers are available in 32, 64, 96, 128,
or 256-channel models.
Note: When sampling at a rate of ~50 kHz only the first 128 amplifier channels will be
available.
SystemHardware
All PZ2 channels are organized into groups of 16 channel banks with each bank
corresponding to a rear panel headstage connector and front panel LED display.
Recorded signals are digitized, amplified, and transmitted to the RZ2 base station via
a single fiber optic connection for further processing. In addition, configuration
information is sent from the RZ2 to the PZ2 preamplifier across the fiber optic
connection.
A standard configuration for neurophysiology recordings includes electrodes (chronic or
acute), one or more Z-Series high impedance headstages, a PZ2 preamplifier, and
an RZ2 base station.
HardwareSet‐up
The diagram below illustrates the connections necessary for PZ2 preamplifier
operation.

6-4 System 3
PZ2 PreAmp
One or more Z-Series headstages can be connected to the input connectors on the
PZ2 back panel. A 5-meter paired fiber optic cable is included to connect the
preamplifier to the base station. The connectors are color coded and keyed to ensure
proper connections. The PZ2 battery charger connects to the round female connector
located on the back panel of the PZ2 preamplifier.
Important! To avoid introducing EMF noise, DO NOT connect the charger to the PZ2 while
collecting data.
PoweringON
To turn the preamplifier on, move the three position battery switch located on the
front panel of the PZ2, to either the Bat-A or Bat-B position.
PoweringOFF
To turn the preamplifier off, move the three position battery switch located on the
front panel of the PZ2, to the OFF position.
Important! Channels are grouped by 16-channel banks and each bank will only power up when
a headstage is connected. This design helps to increase battery life.
PZ2SoftwareControl
The preamplifier’s hardware operation (power options and indicator LEDs) can be
configured using the PZ2_Control macro within the RPvdsEx control circuits running
on the RZ2 base station.

System 3 6-5
PZ2 PreAmp
Double-clicking the macro in RPvdsEx displays the macro properties and allows users
to easily configure the macro. Additional information on using the macro is available
in the macro properties dialog box.
This macro is not required for preamplifier operation but is recommended if the user
requires more control over the amplifier power/up or power/down status or front
panel LEDs. See the relevant sections below for more information about these
features.
PZ2Features
ClipWarningsandActivityDisplay
256 front panel LEDs can be used to indicate spike activity and/or clip warning
depending on display mode and configuration. See “Display Button” and “Status
LED” below for more information.
RecordingChannelLEDs
When enabled, LEDs for each channel may be lit green to indicate activity or red to
indicate a clip warning.
Green:Activity|Red:ClipWarning
Clip Warning When the input to a channel is greater than -3dB from the
preamplifier's maximum voltage input the LED for the
corresponding channel is lit red indicating clipping may occur.
Activity Whenever a unit (spike) occurs (the sensitivity threshold can
be configured with the PZ2_Control macro) the LED for the
corresponding channel is lit green.
Note: The LED Indicators are also mirrored on the RZ2 LCD display.
DisplayButton
The Display button located on the front panel of the PZ2 toggles the clip warning
and activity display LEDs between software control and standard operation.
To toggle between display modes:
•Press the Display button.
StatusLED
When recording, the status LED located below the Display button indicates the
current display mode of the LED Indicators.
Green Software Control of LEDs
Use the PZ2_Control macro to configure LED Indicators. LEDs
are turned off until enabled through software control.
Orange LEDs enabled for standard operation

6-6 System 3
PZ2 PreAmp
In this mode, LEDs are automatically enabled for default activity
and clip warning display as described above.
ExternalGround
The external ground is optional and should only be used in cases where the subject
must occasionally make contact with a metal surface that isn’t tied to the animal
ground, such as a lever press. When contact is made, a ground loop is formed that
temporarily adds extra noise to the system. Grounding this metal surface directly to
the TDT hardware removes this ground loop at the cost of raising the overall noise
floor a small amount.
A banana jack located on the back of the PZ2 (directly to the right of the charger
input) provides connections to common ground for the first bank of channels (1-
16). A cable kit is also provided to ensure cables used with the external ground are
suitable for this use. Each kit includes: one male banana plug to male banana plug
pass through and one male banana plug to alligator clip pass through. These cables
also include ferrite beads to remove any potential RF noise that might travel through
the cable. For best results position the ferrite bead close to the source of the RF
noise.
BatteryOverview
The PZ2 preamplifier features two Lithium ion batteries to allow for longer record
times. A three-position switch selects the active battery between Bank-A, Bank-B, or
both banks off.
MaximizingBatteryLife
To increase battery life, individual banks of channels will only power up when a
headstage is connected to the corresponding input.
The PZ2_Control macro can also be added to the circuit running on the RZ2 to
further specify how PZ2 channel banks are powered. When a headstage is
connected, banks may be powered on or off statically through the Power Control
options within the macro or dynamically by using the PZ2_Control macro inputs.
See the internal macro help for more information.
BatteryStatusLEDs
Battery Level: Eight LEDs indicate the voltage level of the selected battery. These
LEDs can be found on the front of the PZ2 preamplifier by the heading Level. When
the battery is fully charged, all eight LEDs will light green. When the battery voltage
is low, only one green LED will be lit. If the voltage is allowed to drop further, the

System 3 6-7
PZ2 PreAmp
last LED will flash red. TDT recommends charging the battery before this flashing
low-voltage indicator comes on. While charging, the Level LEDs will flash green.
ChargingtheBatteries
Operate the preamplifier with the charging cable disconnected. Connecting the PZ2
charger will simultaneously charge both batteries. TDT recommends putting the three-
position switch in the OFF (middle) position while charging the PZ2.
ChargingIndicators
When powered on, the PZ2 battery status LEDs are also used for each battery to
indicate which battery, if any, is charging. These LEDs are found next to the Level
LEDs by the headings -A- and -B-. A green indicator denotes the battery bank is
fully charged while a red indicator designates the battery is currently charging. When
the device is in operation (charger is not connected) the -A- and -B- LEDs are
not lit.
An external battery pack is also available to provide longer battery life for extended
recording sessions. See “PZ-BAT External Battery Pack for the PZ Amplifiers” on
page 6-23.
PZ2TechnicalSpecifications
Status Description
8 Green Fully Charged
1 Green, 7 Unlit Low Voltage
1 Flashing Red Low Voltage - Charge
Immediately!
8 Green Flashing Charging in Progress
Status Description
Red Charging
Green Fully Charged
Unlit Operation Mode (charger not connected)
A/D Up to 256 channels, 18-bit hybrid
Maximum Voltage In +/- 10 mV
Frequency Response 3 dB: 0.35 Hz – 7.5 kHz
6 dB: 0.2 Hz – 8.5 kHz
Anti-Aliasing Filter 4th order Lowpass (24 dB per octave)
S/N (typical) 73 dB

6-8 System 3
PZ2 PreAmp
*Note: When sampling at a rate of 48.828 kHz the PZ2 preamplifier is limited to a
maximum of 128 channels.
**Note: If longer cable lengths are required, contact TDT.
InputConnectors
PZ2 Preamplifiers have up to 16, 26-pin headstage connectors on the back of the
unit. A1 – A16 represent the 16 channels coming from each connected headstage.
The PZ2 channels are marked next to the respective connector on the preamplifier.
So, for the connector for channel 1 – 16, A1 is channel 1 while on the connector
for channels 17 – 32, A1 is channel 17.
Important! Each input connector uses its own unique ground and reference. When using multiple
headstages, ground pins on all headstages should be connected together to form a
single common ground. See “Headstage Connection Guide” on page 6-99.
Distortion (typical) < 1%
A/D Sample Rate Up to 48828.125 Hz*
Sample Delay
Dependent on Sample Rate and RPvdsEx input method
Rate Pipe Input MC Input
6 kHz 16 samples 15 samples
12 kHz 17 samples 16 samples
25 kHz 20 samples 19 samples
50 kHz 26 samples 25 samples
Input Impedance 105Ohms
Power Requirements 2 Lithium Ion cells at 10 AmpHours each
Battery
Eight hours to charge both cells
Battery life between charges, per cell:
32 ch ~ 13 hrs
64 ch ~ 11 hrs
96 ch ~ 9.5 hrs
128 ch ~ 8 hrs
256 ch ~ 5 hrs
Charger External 6 V, 3 A power supply
Indicator LEDs Up to 256 status or clip warning, battery life, active battery
bank
Input referred noise 2 μV rms typical 300- 7000 Hz, 8 μV peak typical
Fiber Optic Cable 5 meters standard, cable lengths up to 20 meters**

System 3 6-9
PZ2 PreAmp
PinoutDiagram
Note: TDT technical support (386-462-9622 or [email protected]) before attempting to
make any custom connections to pins 6, 18, or 19.
Pin Name Description Pin Name Description
1 A1 Analog Input Channels 14 V+ Positive Voltage (+1.5V)
2 A2 15 GND Ground
3A3 16GND
4 A4 17 V- Negative Voltage (-1.5V)
5 Ref Reference 18 HSD Headstage Detect
6 HSD Headstage Detect 19 HSD
7A5
Analog Input Channels
20 A6
Analog Input Channels
8A7 21A8
9A9 22A10
10 A11 23 A12
11 A13 24 A14
12 A15 25 A16
13 GND Ground 26 NA Not Used

6-10 System 3
PZ2 PreAmp
Table of contents