TECHCON SYSTEMS TS566R User manual

TS566R SMART CONTROLLER
FOR
SPRAY VALVES
User Guide
Copyright © OK International

2
CONTENTS
Page Number
1. Safety ………………………………………………………………… 3
2. Symbol Definitions ………………………………………………. 4
3. Specifications ……………………………………………….………. 5
4. Features ……………………………………………………………. 5
5. To Control Spray Valve………………………………….................…. 5
5.1 Connecting the unit…….……………………………………… 6
5.1.1 Login …………………………………………………… 6
5.1.2 Pressure Calibration.……………………………………. 7
5.1.3 Controlling mode ………………………………………. 8
5.2 Spraying ……………….……………………………............ 10
5.2.1 Pressure Adjustment …………………………………. 10
5.2.2 To Change Pressure Unit Display ………….………... 10
5.2.3 Spray Time Setting ……………………………………... 11
5.2.4 Manual/Purge Cycle ……………………………………. 12
5.2.5 Automatic Cycle ……………………………………… 12
5.2.6 Teach Mode Setting ……………………………………. 13
5.3 Stored Program in Memory Cell ……………………………. 14
5.3.1 To store dispense parameters …………………………… 14
5.3.2 To run in Continuous mode ……………………………. 14
5.3.3 To run in Single Sequence mode ………………………. 15
5.3.4 To run in Continuous Sequence mode ……………………17
5.3 Cycle Counter ………………………………………………... 17
5.4 To Change Password …………………………………………. 18
5.5 To Connect to Wi-Fi network ………………………………… 19
6. Internet of Thing (IoT) ………………………………………………... 21
7. Software Upgrade ……………………………………………………. 27
8. Troubleshooting ……………………………………………………. 29
9. Maintenance …………………………………………………………. 30
10. Warranty ……………………………………………………………... 30
11. I/O Configuration and End of Cycle Switch ………………………. 31

3
1. SAFETY
1.1 Intended Use:
WARNING: Use of this equipment in ways other than those described in this User
Guide may result in injury to persons or damage to property. Use this equipment
only as described in this User Guide.
OK International cannot be responsible for injuries or damages resulting from
unintended applications of its equipment. Unintended uses may result form taking
the following actions:
•Making changes to equipment that has not been recommended in the User
Guide
•Using incompatible or damaged replacement parts
•Using unapproved accessories or auxiliary equipment
1.2 Safety Precautions:
•Do not operate this unit in excess of maximum ratings/settings
•Always wear appropriate personal protective clothing or apparel
•The fluid may be toxic and/or hazardous. Refer to Material Safety Data Sheet
for proper handling and safety precautions
•Do not smoke or use open flame when flammable materials are being sprayed
•This equipment is for indoor use only

4
2. SYMBOL DEFINITIONS
Symbol
Description
Symbol
Description
Run (Activate)
Setup
Valve Pressure
Counter reset
Atomized Air
Pressure
Mac address
Timed mode
Accept change
Interrupt mode
Pressure calibration
Teach mode
Save
Purge mode
Reset time in Teach
mode
Spray
IP address
Pre-Spray
Post-Spray
Login/Logout
Cancel
Run Method
Sequence Mode
Continuous Mode
Sequence Continuous
Mode
E-Stop
Change Password
Remote Server
On line
Off Line
Stand Alone Control
Mode
Robotic Mode
Max Input Pressure
Setting

5
3. SPECIFICATIONS
4. FEATURES
Figure 1.0
Size
290mm x 212mm x 98mm
(11.4” X 8.3” X 3.9”)
Weight
3.02 kg (6.65l bs)
Input Voltage
24VDC
Output Voltage Range
0-24 VDC
Rated Power
15W
Air Input
100 psi (6.9 bars) Max.
Air Output
0-99.9 Psi (6.9 bar)
Pollution Degree
II
Installation Category
I
Indoor Use
Altitude up to 2,000m (6,562ft)
Operating Temperature
0˚C to 50˚C (32˚F to 122˚F)
Storage Temperature
-10˚C to 60˚C (14˚F to 140˚F)
Max. Relative Humidity
80% for temperature up to 31˚C (87.8˚F) Decreasing
linearly to 50% relative humidity at 40˚C (104˚F)
Timer
0.008-99.99 seconds
Cycle Mode
Timed, Interrupt, Teach, Purge
Timing Repeat Tolerance
+/- 0.001%
Cycle Rate
900 cycles/min
Display
Touch Screen, Resistive
Meets or exceed
CE, TUV and NRTL

6
5. TO CONTROL SPRAY VALVE
CAUTION: A 5-micron filter (TSD800-6) must be installed with the unit to ensure proper
air filtration.
Figure 2.0
5.1 Connecting the Unit: (Refer to Figure 1.0 and 2.0)
1. Connect the power cord, foot switch to the back of the unit.
2. Connect Valve air hose to Port A (Not Regulated)
3. Connect Atomized air hose to Port B (Regulated)
4. Press the Power switch to turn on the unit.
Items
Description
1
Air Filter
2
Valve (not included)
3
Pressure Pot (not included)
4
Display
5
Power Adapter
6
Foot Switch

7
5.1.1 Login
1. Touch the Login Icon to enter login screen
2. Enter “0000” in Password window
3. Touch the Check Mark icon to save and exit
5.1.2 Pressure Calibration
Note: Pressure calibration must be performed when the unit is activated for
the first time.
1. Using a pressure gauge, verify that the input pressure line is delivering
approximately 100 psi
Note: If input pressure is not 100 psi, calibration will result in a
mismatch between the display pressure and the actual pressure on output
port B
2. Once you have verified input pressure, touch the Calibration icon
to enter calibration screen.

8
Please keep in mind that the exact pressure being fed must be indicated
correctly in this menu prior to calibrating the pressure. Not doing so may
cause the pressure display to show the incorrect pressure.
3. Turn pressure adjustment knob counterclockwise until the wheel can no
longer be turned.
4. Touch the 0 icon to set the pressure to 0
5. Turn pressure adjustment knob clockwise until display output pressure is
100 psi
6. Touch the 100 icon to set the pressure to 100
Warning: Do not turn knob all the way clockwise. 100 psi upper bound
should be set as soon as display output pressure goes from 99 to 100 psi.
Continuing to turn knob clockwise despite pressure on display already
being set at 100psi will result in incorrect output pressure readings if the
100 icon were to be pressed at that time
Note: The digital values shown at “0” and “100” icons are for reference
only. The actual calibrated values will be different
7. Touch the Reboot icon to save the settings and reboot the system
8. Wait until the system completes the rebooting sequence and the home
screen is displayed
The unit is now calibrated and ready to operate

9
5.1.3 Controlling Mode
The TS560R can be used as stand-alone controller or integrated to any
robot/automation system.
A. To use as stand-alone controller:
Go to setup screen to check for controlling mode. If the icon “C” is
showing, then it is already set as stand-alone controlling mode. If the
icon “R” is showing, press this icon to change to “C”
B. To use as integrated controller to robot/automation
Press icon “C” to change to “R”
Note: Users will experience inconsistent dispensing and intermittent
behaviors with their dispensing valves if attempting to run the TS550R
in controller mode when interfacing with the robot.
Stand alone controlling
mode
Robot mode (integrate
to robotic system)

10
5.2 SPRAYING
5.2.1 Pressure Adjustment
Note: Pressure in Port A (I) is not regulated. It delivers the same pressure
as the supplied inlet pressure. This pressure is to activate the valve. Make
sure the supplied pressure is at least 70 psi (4.8 bar)
Pressure in Port B is regulated. This pressure is to control the atomized air
spray. Adjust the pressure accordingly to achieve desired spray results.
1. Touch the “Pressure B” icon to enter the pressure setup screen
2. Touch the Up and Down arrows to set the desired atomized pressure
3. Touch the Check Mark icon to save and exit
5.2.2 To Change Pressure Unit Display
Note: The default pressure unit is PSI. To change pressure unit to BAR,
follow below instructions.
1. Touch the Setup icon to enter setup screen
2. Press the “BAR” icon to change pressure unit to BAR
3. Touch the X icon to save and exit

11
5.2.3 Spray Time setting
Note: A complete cycle consists of Pre-Spray, Spray and Post-Spray.
Pre-spray: Atomized air turned on before the valve is opened.
Post-spray: Atomized air stayed on after the valve is closed.
1. Touch the “Pre”icon to setup pre-spray time
2. Touch Up and Down arrows to set the desired pre-spray time
3. Touch the Check Mark icon to save and exit
4. Touch the “Post”icon to setup post-spray time
5. Touch Up and Down arrows to set the desired post-spray time
6. Touch the Check Mark icon to save and exit
5.2.4 Manual/Purge Cycle Setting
1. Touch the Purge icon to select the purge cycle.
The Purge icon will turn to green color.
2. Press and hold down the foot switch to activate the purge cycle.

12
Alternately, touch and hold the Run icon on the display to activate the
purge cycle.
5.2.5 Automatic Spray Cycle Setting
1. Touch the “Timed” mode icon to set the spray cycle time.
The icon will turn to green color.
2. Touch the “Spray”icon to enter the setup screen
3. Touch the Up and Down arrows to set the desired spray time
Note: The minimum activation is 0.008 second
4. Press the Check Mark icon to save and exit
5. Press the foot switch to activate the “Timed” spray cycle.
Alternately, touch the Run icon on the display to activate the “Timed”
spray cycle
Note: The unit has an “Interrupt” mode feature. In this mode, the
“Timed” spray cycle can be disrupted if the foot switch is released and
resumed when the foot switch is depressed again.
6. Touch the “Interrupt” icon to activate “Interrupt” mode. The icon will
turn to green color
5.2.6 Teach Mode Setting
In the teach mode, the spray time will be accumulated as long as the foot
switch is depressed. This is helpful in determining the required spray
time when the dispense output is known.
1. Touch the “Teach” icon to enter the teach mode

13
2. Touch the “Time Reset” icon to set timer to zero
3. Press and hold down the foot switch, the spray time will be accumulated
4. Release the foot switch when the desired amount of fluid has sprayed
5. Touch the “Timed” icon to transfer the accumulated spray time to “Timed”
mode
6. The unit is now set to repeat this “Timed” cycle
5.3 Stored Program in Memory Cell
The unit has 50 memory cells to store all parameters. The controller can activate
all memory cells in sequence mode.
5.3.1 To store dispense parameters
1. Touch the forward or backward arrow to select the desired memory cell.
2. Enter all the desired parameters then touch the “Save”
icon to save the data
5.3.2 To run in Continuous Mode
The controller can be setup to repeat the run continuously.
1. Touch the Setup icon to enter the setup screen

14
2. Touch the “Run Method” icon
3. Enter the delay time (ex. 5000 ms)
NOTE: The minimum delay time is 15ms
4. Touch the “Continuous Run” icon
5. Touch the Check Mark icon to save
6. Touch the X icon to exit
The screen will look similarly to the screen below:
Note: If the controller is set to activate memory cell 1 and the delay time is
set at 5000 ms, the controller will activate memory 1 continuously with
5000 ms delay between each activation.

15
5.3.3 To run in Single Sequence Mode
1. Touch the “Run Method” icon to enter sequence mode setup
2. Enter number of memories to be run in sequence (ex. 1 to 3)
3. Touch the “Sequence Mode” icon
4. Touch the Check Mark icon to save
5. Touch the X icon to exit
The screen will look similarly to the screen below:

16
Notes:
A. If there is no delay time entered in the setting, the operator has
to press the foot switch or touch the Run icon after each
memory cell is completed to activate the next memory cell.
B. If delay time is entered in the setting, the controller will activate
the next memory cell in sequence automatically. Delay time
must be at least 15 ms.
5.3.4 To run in Continuous Sequence Mode
1. Follow steps 1 to 2 above and enter the waiting time “delay time” between
each activation.
2. Touch the “Sequence Mode” icon
3. Touch the “Continuous Mode”icon
4. Touch the Check Mark icon to save
5. Touch the X icon to exit
The screen will look similarly to the screen below:
Note: If the sequence mode is set to activate memory cells 1 –3, and the
delay time is set at 5000 ms, the controller will activate memory 1 to 3
continuously with 5000 ms delay between each activation. The delay time
must be at least 15 ms.

17
5.4 Cycle Counter
The cycle counter records the numbers of spray cycle being activated. Up to
999,999 cycles can be recorded. To reset the counter, follow steps below:
1. Touch the Setup icon to enter setup screen
2. Touch the Counter Reset icon to reset the counter
3. Press the check mark icon to confirm
4. Touch the X icon to exit
5.5 To Change Password
Note: The default password is “0000”. To change password, follow the
instructions below.
1. Touch the Login icon to enter the login screen

18
2. Touch the change “Password” icon
3. Enter the old password, then enter the new password
4. Touch the Check Mark icon to save
5. Touch the X icon to exit

19
6. INTERNET of Thing (IoT)
6.1 Setup webserver on the computer
1. Go to Techcon Website (www.techconsystems.com), under resources tab,
download “Software download for IoT”, and focus mainly on copying the
following files to your computer:
a. The “portal.exe” file
b. The “timezone.zip” file
2. Create a new folder and name “IoT” or any desired name
3. Copy the “portal.exe” file and the “timezone.zip” file in the “IoT” folder
4. Click on the “Windows”start icon
5. Type: “cmd” in the box (as show below)
6. Press the “Enter” button
The screen should appear as shown below

20
7. Type: “cd :\iot”then press the “Enter”button
8. Type: “set OKI_IOT_PREFIX=MyPrefix” then press the “Enter” button
Table of contents
Other TECHCON SYSTEMS Controllers manuals