TECNA PROVASET 2P User manual

PROVASET 2P
PRESSURE DECAY LEAK TESTER
INSTRUCTION MANUAL FOR:
code 2P2B00 - full sc le 2 b r
code 2P6B00 - full sc le 6 b r

WHITE
WHITE
PAGE
PAGE

PROVASET 2P code 2P2B00 - 2P6B00 - Instruction m nu l
rev. 14/02/2013 p ge 3
Used conventions
! CAUTION: Safety rules for the operator
% WARNING: Instructions to be followed carefully for the correct equipment operation
and to avoid any damage
NOTE: Contains important prescriptions for the equipment maintenance
NOTE: Contains important supplementary information or useful suggestions
! CAUTION:
THIS TESTER USES C MPRESSED AIR.
THE PERAT R MUST BE PR TECTED AGAINST P SSIBLE DAMAGE CAUSED BY
THE ACCIDENTAL DETACHMENT F PARTS UNDER PRESSURE F THE PR DUCT
T BE TESTED.
THE TESTED PR DUCT MUST C NF RM WITH THE BINDING REQUIREMENTS
(ex. directive "Simple pressure tanks").
General recommendations
•The power socket needs compulsorily ground connection.
•Alw ys keep the power supply c ble in good conditions.
•Cle n the equipment outer housing fter disconnecting the power supply c ble. Use
only w ter with neutr l detergents nd soft cloths. Never use lcohol or thinners.
•Do not pour fluids on the equipment: this could d m ge some of its p rts perm nently,
especi lly those under volt ge.
•Use the equipment only for the intended purposes, described in this m nu l.
Different uses c n compromise the equipment s fety.
NOTE:
THIS MANUAL MUST BE READ CAREFULLY BEF RE USING THE EQUIPMENT F WHICH IT IS
INTEGRAL PART.
PRESERVE WITH CARE.
Reserved rights
This m nu l is offici lly only h nded out to the leg l owner of the corresponding equipment
for conforming uses.
The copy nd the distribution of ny p rt of this m nu l, under ny form, re forbidden
without the express consent of tecna srl.

WHITE
WHITE
PAGE
PAGE

PROVASET 2P code 2P2B00 - 2P6B00 - Instruction m nu l
rev. 14/02/2013 p ge 5
INDEX OF CONTENTS
1 INTRODUCTION
2 DESCRIPTION AND INSTALLATION
2.1 Front p nel
2.2 Re r p nel
3 PNEUMATIC DIAGRAM
4 TESTING CYCLE
4.1 Le k test cycle with three-ph ses
4.1.1 Oper ting principle
4.1.2 Testing cycle procedure
4.1.3 Three-ph ses testing cycle (T1, T2, T3)
4.2 Le k test cycle with two-ph ses
4.2.1 Oper ting principle
4.2.2 Two-ph ses testing cycle (T1, T2)
4.3 Errors rel ting to the testing cycle
4.4 Test results
4.5 Al rm indic tions: C uses nd solutions
4.6 Obstruction test
5 MESSAGE DISPLAY
5.1 B se menu structure
5.2 Mess ge displ yed t power on
5.3 LCD contr st
5.4 Mess ge menu
5.5 "TEST COUNTERS" menu
5.6 Inst nt neous re ding of the pressure sensor
6 PROGRAMMING
6.1 Using the keybo rd
6.2 "TEST TABLE selection" select menu
6.3 "TEST TABLES" modify menu
6.4 "CONFIGURATION AND OPTIONS" menu
7 CONNECTION TO PROGRAMMABLE LOGIC CONTROLLER /PLC
7.1 Interf ce with the PLC: Fe tures, sign ls nd connection
8 CALIBRATING THE PRESSURE SENSOR
8.1 C libr tion method
8.2 "PRESSURE CALIBRATION" menu
9 TECHNICAL FEATURES
DECLARATION OF CONFORMITY
GUARANTEE
TECHNICAL SERVICE REQUEST
USER INFORMATION EUROPEAN DIRECTIVE 2002/96/EC

WHITE
WHITE
PAGE
PAGE

PROVASET 2P code 2P2B00 - 2P6B00 - Instruction m nu l
rev. 14/02/2013 p ge 7
1 INTRODUCTION
PR VASET 2P is n electro-pneum tic le k tester dedic ted to perform pressure dec y le k
tests: It m kes it possible to bench test sm ll nd medium-size products, both stiff nd
flexible.
PR VASET 2P fe tures very comp ct dimensions, simple use nd high perform nces.
The pressure re ding offers 0.1 mb r resolution on the model 2P2B00 up to 2 b r full sc le
r nge nd model the model 2P6B00 up to 6 b r full sc le r nge.
The simple nd effective pneum tic circuit gu r ntees the reli bility shown on site by the
previous PR VASET models.
The c se is m de of unp inted st inless steel th t is suit ble for ny environment, especi lly
the white rooms of the biomedic l sector.
Beyond the le k test, PR VASET 2P llows to lso c rry out multiple obstruction test fter
the relev nt en bling.
PR VASET 2P h s 16 progr mm ble test p r meter t bles: e ch t ble includes the
p r meters djusting test cycle nd it is therefore possible to pre-set different test
conditions for different products.
PR VASET 2P h s six select ble units of me surement rel ting to the pressure.
PR VASET 2P h s five select ble l ngu ges for displ y mess ges.
PR VASET 2P h s n ided procedure m king it possible to c libr te the pressure sensor.
This vers tility, together with high ccur cy nd reli bility level nd n user friendly control,
m kes it possible to use the le k tester both in m nuf cturing dep rtments nd Qu lity
M n gement l bor tories.

PROVASET 2P code 2P2B00 - 2P6B00 - Instruction m nu l
rev. 14/02/2013 p ge 8
2 DESCRIPTION AND INSTALLATION
2.1 Front pane
PR VASET 2P - Front p nel
PICTURE 1
KEYS:
1. START push-button: Button st rting the test cycle (m nu l st rt).
If pressed continuously for more th n one second, it c ncels the test being c rried out
(RESET) t ny moment.
2. GREEN light:
It blinks for second fter the tester st rting up.
It blinks if the l st test result is PASSED.
3. RED light:
It blinks for second fter the tester st rting up.
It blinks when the device is in ALARM st te or if the l st test result is ERROR or FAILED.
4. B ck-lit gr phic 128x64 LCD displ y th t visu lises ny inform tion
rel ting to the testing ph ses, the progr mm ble p r meters nd the l rms.
5. Air outlet connection:
Tube holder (tube 6x4mm) for the connection of the product to be tested.
NOTE: The connection tube to the tested product must be as short as possible.
Any volume added to the tested product reduces the measure
sensitivity.
The tests should be performed always with the same type of connection
to have repetitive test results.
Check often the integrity of the connection between the instrument and
the tested products.
Do not tighten the tube stopping nut more than necessary (if you use a
tool, pay special attention).
NOTE: For the use of keys, see paragraph 6.1.

PROVASET 2P code 2P2B00 - 2P6B00 - Instruction m nu l
rev. 14/02/2013 p ge 9
2.2 Rear pane
PR VASET 2P - Re r p nel
PICTURE 2
KEYS:
6. Pressure reducer:
It djusts the ir pressure th t supplies the pneum tic circuit.
Adjust the pressure reducer to the progr mmed test r ted pressure in c se the three-
time le k test is selected.
Adjust the pressure reducer t pressure th t is slightly higher th n the progr mmed
test r ted pressure, in c se the two-time le k test is selected.
To djust the pressure, unlock the knob by unscrewing the locking nut, rot te it
clockwise to incre se nd counter-clockwise to decre se:
At the end, rot te the nut g in up to the locking position.
Adjusting field:
2P2B00 100 mb r : 2 b r ;
2P6B00 200 mb r : 8,5 b r;
Inlet m x. pressure : 10 b r.
% WARNING: Do not adjust the pressure reducer at a pressure that is higher than the
device full scale not to damage the inner pressure sensor.
7. PLC connector nd START FOOT SWITCH:
15-pin socket "D" connector for the PR VASET 2P - PLC connection nd for the FOOT
SWITCH START CONTROL connection s expl ined in ch pter 7.
8. M in two-pole 0/I (OFF/ON) switch for power supply
Socket:
IEC320 socket for the det ch ble supply c ble.
The socket h s two del yed fuses of the T 630mA, 5x20mm type.

PROVASET 2P code 2P2B00 - 2P6B00 - Instruction m nu l
rev. 14/02/2013 p ge 10
! CAUTION: The protective earth terminal is connected to the tester chassis.
Check the good state of the earth connection while installing the
equipment and later periodically according to standards.
! CAUTION: Disconnect the power supply wire before replacing the fuses.
9. Compressed ir inlet connector:
Quick connector for tube 6x4 mm.
Inlet m x. pressure: 10 b r.
% WARNING: The supply pressure must be kept constant to allow a correct operation
of the pressure reducer; pressure changes can modify the test pressure.
! CAUTION: Use only dry air, without condensate, filtered, not lubricated and free
from explosive gases.
10. L bel:
It indic tes the producer d t , the tester ex ct code, its seri l number nd the d t
rel ting to the necess ry electric l volt ge supply.

PROVASET 2P code 2P2B00 - 2P6B00 - Instruction m nu l
rev. 14/02/2013 p ge 11
3 PNEUMATIC DIAGRAM
PR VASET 2P - Pneum tic di gr m
PICTURE 3
The pneum tic di gr m of the device is simple, but effective nd s fe.
It is m de up of:
•A pressure reducer (see point 6, p r. 2.2)
•Two 3/2-w y N.C. solenoid v lves: Both v lves gu r ntee tight se l nd c n be e sily
inspected for m inten nce nd cle ning purposes.
•A pressure tr nsducer: It me sures the testing pressure.
Both v lves 1 nd 2 re opened ( ctiv ted) only during the filling nd re closed
(de ctiv ted) during the settling nd the pressure dec y me surement.
In this w y, the piece being tested is connected to the pressure reducer (inlet ir) only during
the filling, while - during the settling nd the me surement - the piece being tested is
completely isol ted from the inlet ir, bec use the closed v lve 2 is ctu lly connected to
the disch rge outlet.
A possible blow-by or f ilure of v lve 2 during the test c n therefore only c use pressure
loss in the piece being tested nd not pressure incre se: The possibility th t n error of the
pneum tic circuit c n hide pressure le k of piece to be rejected is therefore excluded.

PROVASET 2P code 2P2B00 - 2P6B00 - Instruction m nu l
rev. 14/02/2013 p ge 12
4 TESTING CYCLE
! CAUTION: This tester uses compressed air.
The oper tor must be protected g inst possible d m ge c used by
the ccident l det chment of p rts under pressure of the product to be
tested.
The tested product must conform with the binding requirements (ex.
directive "Simple pressure t nks").
4.1 Leak test cyc e with three phases
4.1.1 Operating princip e
% WARNING: The operator must carry out the test after ensuring that the supply
pressure adjustment is optimal (see point 6, par. 2.2).
The inlet pressure in particular must be maintained constant to allow a
good adjustment of the test rated pressure.
Before starting the testing cycle, it is necessary to set the operative
parameters of the "TEST TABLES" menu (see par. 6.3) and of the
"C NFIGURATI N AND PTI NS" menu (see par. 6.4).
The test cycle st rts fter receiving START comm nd.
PR VASET 2P - Three-time pressure dec y le k test
PICTURE 4

PROVASET 2P code 2P2B00 - 2P6B00 - Instruction m nu l
rev. 14/02/2013 p ge 13
The three-ph ses test cycle is b sed on 5 progr mm ble p r meters.
T1 Fi ing time
Time T1 is st rted fter the START comm nd: It is the filling ph se. Air comes out from
the outlet connection on the front p nel (see p r. 2.1 point 5) to fill the piece being
tested t the test pressure Pr.
If the filling of the piece to be tested is provided outside the equipment, the time T1
c n be set t 0 (zero): t the START comm nd the equipment skips the filling ph se
nd, without opening the v lves, the test directly st rts from time T2.
By selecting the option Pr/T1 in the configur tion p r meters for the type of test with
three ph ses, the filling ph se is utom tic lly interrupted when the test pressure Pr is
re ched, without w iting for the end of time T1.
T2 Sett ing time
It is necess ry to w it until the el stic p rts in the tested piece re settled.
At the end of T2, if the pressure inside the tested piece is within the set percent ge
limits %Pr m x nd %Pr min, the pressure v lue is stored s reference point for
checking the Dp le k limit in time T3.
T3 Measuring time
The pressure dec y inside the tested piece is me sured.
At the end of time T3, the tester stores the test result nd displ ys the pressure dec y
th t w s me sured t the test end.
Pr Test rated pressure
It is the pressure v lue t which (with progr mmed percent ge toler nce %Pr m x
nd %Pr min) the le k test is c rried out.
The percent ge toler nce limits %Pr m x nd %Pr min of the test r ted pressure Pr re
progr mmed in the “configur tion nd options” menu.
Dp Pressure De ta
It is the m ximum limit of pressure dec y or pressure incre se which is ccepted
during the T3 time.
A pressure dec y is indic ted s neg tive v lue, with minus sign, pressure
incre se is indic ted s positive v lue, without sign.

PROVASET 2P code 2P2B00 - 2P6B00 - Instruction m nu l
rev. 14/02/2013 p ge 14
4.1.2 Testing cyc e procedure
The tester ccepts test st rting comm nds t ny moment nd during the visu lis tion of
ny menu p ge, p rt from those rel ting to the pressure sensor c libr tion.
The test t ble used when testing is directly displ yed when st rting or fter "reset" of the
test in progress.
At the end of test or during the tests, the test t ble c n be displ yed in the mess ge menu
(see p r. 5.1) nd c n be selected (figure from 1÷16) in the "Test t ble selection" menu (see
p r. 6.2).
Any ch nge of the test t bles c rried out during test in progress will be ctive only for the
next test.
4.1.3 Three-phases testing cyc e (T1, T2, T3)
Re dy nd w iting for the next test, th t will be c rried out
with the displ yed t ble no. XX.
The green or the red light is blinking, depending on the result
of the l st test performed.
Only for 2P2B00 model
In c se the m nu l pressure disch rge option w s selected,
the green nd the red lights re both off.
Test running:
Filling time T1
xx.x: Rem ining T1 time (decre sing v lue).
yyyy : Line pressure re d by the pressure sensor.
The v lves, opened when st rting, re closed t the end of
time T1.
Test running:
Settling time T2
xx.x: Rem ining T2 time (decre sing v lue).
yyyy : Line pressure re d by the pressure sensor.
Test running:
Time T3, me surement of the pressure dec y Dp
xx.x: Rem ining T3 time (decre sing v lue).
yyyy : Pressure dec y detected (with sign).
TEST
TABLE n XX
TEST: FILLING
T1 s XX.X
mmH2O YYYY
TEST
TABLE n XX
TEST: Dp MEASURE
T3 s XX.X
mmH2O - YYYY
READY
TABLE n XX
T1 xx.x mmH20
T2 xx.x r YYYY
T3 xx.x Dp XX
TEST
TABLE n XX
TEST: SETTLING
T2 s XX.X
mmH2O YYYY

PROVASET 2P code 2P2B00 - 2P6B00 - Instruction m nu l
rev. 14/02/2013 p ge 15
4.2 Leak test cyc e with two phases
4.2.1 Operating princip e
The two-ph ses le k test is simil r to the three-ph ses test: The only difference is th t in the
two-ph ses test, time T1 includes both the filling nd the settling ph se nd it is not possible
to set it t 0 (zero)
Time T2 is the me suring ph se of the pressure dec y.
The v lves, opened when st rting, re closed when the pressure re ches the set test r ted
v lue.
4.2.2 Two-phase testing cyc e (T1, T2)
Test running:
Filling/settling time T1
xx.x: Rem ining T1 time (decre sing v lue).
yyyy : Line pressure re d by the pressure sensor.
The v lves, opened when st rting, re closed when the
pressure re ches the Pr r ted v lue.
Test running:
Time T2, me surement of the pressure dec y Dp
xx.x: Rem ining T2 time (decre sing v lue).
yyyy : Pressure dec y detected (with sign).
4.3 Errors re ating to the testing cyc e
This p ge is displ yed if, t the end of the "Filling" or "Settling"
time, the line pressure is not within the set %Pr m x nd %Pr
min limits or if the test h s been interrupted by n l rm.
The word "F iled" is displ yed if the pressure is lower th n the
set %Pr min limit t the end of the settling time (L rgely F iled).
The red led blinks up to the beginning of new test.
In c se of error with l rm, the type of error th t interrupted
the test is indic ted.
The red led blinks up to the beginning of new test.
TEST
TABLE n XX
TEST: FILLING
SETTLING
T1 s XX.X
mmH2O YYYY
TEST
TABLE n XX
TEST: Dp MEASURE
T2 s XX.X
mmH2O -YYYY
END OF TEST
TABLE n XX
result:
ERROR/FAILED
r % MIN
mmH2O Dp
XXXXX xxx
END OF TEST
TABLE n XX
result:
ERROR
mmH2O Dp
XXXXX xxx

PROVASET 2P code 2P2B00 - 2P6B00 - Instruction m nu l
rev. 14/02/2013 p ge 16
4.4 Test resu ts
The test is ended with fai ed result if, t the end of the "Dp
me sure" time, the me sured pressure dec y - s bsolute
v lue without sign - is equ l or higher th n the set Dp
m ximum limit.
The red led blinks up to the beginning of new test.
The test is ended with good result if, t the end of the "Dp
me sure" time, the me sured pressure dec y - s bsolute
v lue without sign - is lower th n the set Dp m ximum limit.
The green led blinks up to the beginning of new test.
If the m nu l pressure disch rge option w s selected, the
pressure drops down when the oper tor disconnects the
tested p rt, nd then the green light turns off nd the displ y
ch nges to the mess ge indic ting re dy for new test with
the selected t ble.
4.5 A arm indications: Causes and so utions
Under l rm conditions, the test in progress is c ncelled.
- TABLE No. XX
Error in the p r meters stored in the eeprom.
Reprogr m the test t ble.
- CONF GURAT ON
Error in the p r meters stored in the eeprom.
Reprogr m a the p r meters of the "CONFIGURATION AND OPTIONS" menu.
- CAL BRAT ON
Error in the c libr tion const nts of the pressure tr nsducer.
C rry out the c libr tion procedure.
- PRESSURE
Error in the re ding of the pressure tr nsducer.
It is necess ry to rep ir the device.
NOTE: Until these operations are not carried out in the correct way, the device won’t
be ready to start the test.
END OF TEST
TABLE n XX
result:
FAILED
mmH2O Dp
XXXXX xxx
END OF TEST
TABLE n XX
result:
ASSED
mmH2O Dp
XXXXX xxx
ALARMS
- TABLE n XX
- CONFIGURATION
- CALIBRATION
- RESSURE

PROVASET 2P code 2P2B00 - 2P6B00 - Instruction m nu l
rev. 14/02/2013 p ge 17
4.6 Obstruction test
En ble the obstruction test inside the "Configur tion nd options" menu (see p r. 6.4).
The obstruction test function m kes it possible for the oper tor to check, during the settling
ph se, whether second ry p rts of the piece being tested re obstructed.
After START comm nd, t the end of the filling ph se, (th t is to s y during the settling
ph se), the oper tor c n open the closing plug of second ry p rt of the tested line nd
check the sudden pressure dec y th t t kes pl ce if the line is not obstructed: After re-
closing the plug, the le k test rest rts utom tic lly from the filling ph se.
The norm l procedure to c rry out the obstruction test is the following one:
1. Prep re the piece to be tested;
2. After START comm nd, the test st rts with the filling ph se, followed by the settling one;
3. Before the end of the settling ph se, open the plug th t closes the p rt th t you wish to
test;
4. If the p rt is not obstructed, sudden pressure dec y t kes pl ce nd is sign lled by the
sound of the buzzer for 1 second. The test rest rts from the filling ph se (st rt of time T1);
5. Re-close the plug on the tested p rt;
6. At the end of the filling ph se, the obstruction test c n be repe ted on nother
devi tion or the test c n go on like norm l le k test.
If during the obstruction test with opened plug the pressure dec y is not sudden, this dec y
c n be considered like settling nd does not c use ny sign lling: the test is c rried out
like norm l le k test.
NOTE: The obstruction test can be only applied with higher test pressure
of approx. 200 mbar.

[↓][↑]
[↓][↑]
MESSAGE
MENU
►►Test phases
►►Results
►►Errors/Alarms
ROGRAMMING
TEST TABLE
SELECTION
ROGRAMMING
TEST TABLES
ROGRAMMING
CONFIGURATION
AND O TIONS
READY
TABLE n XX
T1 xx.x mmH20
T2 xx.x r YYYY
T3 xx.x Dp XX
SELECT
TABLE n XX
T1 xx.x mmH20
T2 xx.x r YYYY
T3 xx.x Dp XX
MODIFY
TABLE n XX
T1 xx.x mmH20
T2 xx.x r YYYY
T3 xx.x Dp XX
CONFIGURATION ->
Language:
Test table:
Test pressure:
[MAX + %] XXX
[MIN - %] YYY
ROVASET 2
Model: Rev:
2 XB00 X.X
ressure:
mmH2O YYYY
RESSURE
CALIBRATION
FOLLOW
CAREFULLY THE
INSTRUCTION !
CONFIGURATION ->
Start Command:
Enable test for
Obstruction:
Final pressure
Discharge:
CONFIGURATION
Buzzer:
rogramming:
[table n. ] XX
[parameters] XX
TEST COUNTERS
TESTS XXXXXX
ASSED XXXXXX
FAILED XXXXXX
TEST COUNTERS
TESTS XXXXXX
ASSED XXXXXX
FAILED XXXXXX
RESET +YES -NO
CONFIGURATION ->
ressure scale:
Test phases:
Time base:
[↓][↑] [↓][↑] [↓][↑] [↓][↑]
[↓][↑]
[↓][↑]
[↓][↑]
P PP
P
P
P
PROVASET 2P code 2P2B00 - 2P6B00 - Instruction m nu l
rev. 14/02/2013 p ge 18
5 MESSAGE DISPLAY
5.1 Base menu structure
The BASE MENU is subdivided into MENUS (mess ge groups or progr mm ble p r meters).
E ch menu c n be recognised th nks to its initi l p ge.
PR VASET 2P - Structure of the b se menu
PICTURE 5
Press button [P↵] to scroll the b se menu nd select the menu you w nt to ccess; it c n be
recognised by the initi l p ge which is visu lised by the tester.
After selecting the menu, press the rrow down button [â] to enter. The instructions to move
inside the b se menu by using the keys re described in p r gr ph 6.1.
NOTE: PRESS BUTT N [P↵] R THE ARR W UP BUTT N T RETURN T THE BASE MENU.

PROVASET 2P code 2P2B00 - 2P6B00 - Instruction m nu l
rev. 14/02/2013 p ge 19
5.2 Message disp ayed at power on
When switching on, the device c rries out l mp test of the front p nel nd of the inner
buzzer nd displ ys the following mess ge:
x.x : Softw re revision
WARNING: While starting up, the operator must make sure that all the lights are on and
that the buzzer sounds for one second.
For the next 3 seconds, the displ y shows the following mess ge:
This p ge, shown t power up, informs the oper tor th t
the LCD displ y contr st c n be djusted by keeping pressed
the key [+] or the key [-] nd then pressing, together, with
pulses, the key [P].
5.3 LCD contrast
When switching on the device, the ch r cters displ yed on the LCD must be cle rly
re d ble with good contr st between light nd d rk.
The position of the oper tor from the device displ y, higher or lower, ch nges the sight
ngle nd the contr st.
NOTE: The contr st c n be ch nged only inside the p ge of the b se menu th t
shows the mess ge "READY TABLE n. XX", with the test t ble which is re dy for
the next test.
To modify the LCD contr st, press first nd keep pressed the key [+], to incre se,
or the key [-], to decre se, then press simult neously nd repe tedly, with
pulses, the key [P].
5.4 Message menu
The "Mess ge Menu" cont ins ll the inform tion sent by the tester, ccording to the current
st te, to the oper tor nd rel ting to:
- Indic tions on the test st te (see ch p. 4);
- Displ y of the oper tive test t ble;
- Piece counter;
- Inst nt neous re ding of the pressure sensor.
To ccess this inform tion, you need to use the [↓] or [↑] keys.
tecna srl
ROVASET 2
Model: Rev:
2 XB00 X.X
* tecna srl
ROVASET 2
SET LCD CONTRAST
KEYS [+][-]&[ ]

PROVASET 2P code 2P2B00 - 2P6B00 - Instruction m nu l
rev. 14/02/2013 p ge 20
5.5 "Test COUNTER" menu
TESTS
xxxxxx: Number of tests to be c rried out.
R nge (0 ÷ 9999).
PASSED
xxxxxx: Number of tests ended with p ssed result.
R nge (0 ÷ 9999).
FAILED
xxxxxx: Number of tests ended with f iled result.
R nge (0 ÷ 9999).
Press the [P↵] key to en ble the zero setting request, then
press the [+] key to zero.
5.6 Instantaneous reading of the pressure sensor
From the Mess ge Menu, t ny time, it is possible to displ y the
re ding of the sensors by using the keys [↓] or [á].
TEST COUNTERS
TESTS XXXXXX
ASSED XXXXXX
FAILED XXXXXX
P
TEST COUNTERS
TESTS XXXXXX
ASSED XXXXXX
FAILED XXXXXX
RESET +YES -NO
ROVASET 2
Model: Rev:
2 XB00 X.X
ressure:
mmH2O YYYY