manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. Tektronix
  6. •
  7. Portable Generator
  8. •
  9. Tektronix FG501 User manual

Tektronix FG501 User manual

Tektronix, Inc.
p.10.
Box 500
Beaverton, Oregon 97077
COMMITTEDTO EXCELLENCE
INSTRUCTION
MANUAL
Serial Number
First printing August
1972
TABLE OF CONTENTS
SECTION
1
OPERATING INSTRUCTIONS
SECTION
2
SPECIFICATION AND PERFORMANCECHECK
Page
1-1
2-
1
I
WARNING
)
THE FOLLOWING SERVICING INSTRUCTIONS ARE FOR USE BY QUALIFIED PER-
SONNEL ONLY. TO AVOID PERSONAL INJURY, DO NOT PERFORM ANY SERVICING
OTHER THAN THAT CONTAINED IN OPERATING INSTRUCTIONS UNLESS YOU ARE
QUALIFIED TO DO SO.
SEICTION
3
ADJUSTMENT
3-1
SE:CTION
4
MAINTENANCEAND INTERFACING INFORMATION
4-
1
SE:CTION
5
CIRCUIT DESCRIPTION
5-
1
SE:CTION 6 OPTIONS 6-
1
SECTION
7
REPLACEABLEELECTRICAL PARTS
7-
1
SEiCTlON
8
DIAGRAM AND CIRCUIT BOARD ILLUSTRATION
8-1
SEICTION
9
REPLACEABLEMECHANICAL PARTS AND EXPLODED VIEW
9-1
CHANGEINFORMATION
Fig.
1-1.
FG
501
Function Generator
REV.
B
JUL
1978
Section
1-FG
501
INSTRUCTIONS
INTRODUCTION
The FG !501 FunctionGenerator isdesignedtooperate
in a
TM
500-Series power module. Low distortion sine,
square,tria~ngle,pulse,andrampwaveformsfrom0.001
Hz
to 1
MHz
as well as a $2.5 volt square-wave trigger are
available atthefront panel.Variable DCoffset ofk7.5volts
is also provided. A "hold1' feature allows the generator
output to be abruptly halted at its instantaneous voltage
level and held there until manually switched on again.
A voltage-controlled frequency (VCF)inputisprovided
to control the output frequency from anexternal voltage
source. The output frequency can be swept above or
below the selected frequency to a maximum of 1000:l
depending onthe polarityandamplitudeof the VCF input
and the selected output frequency.
Also included is an external gate input that allows the
generator to beturned onforthe durationof an externally
applied gating signal. This mode provideseither a single
cycle output or a train (burst) of preselected waveforms
depending on the gating signal width and the generator
frequency setting. The phase (startlevel) of thewaveform
burst can be varied
k90°
by a front-panel control.
The variety of swept and modulated signals available
from the IFG 501 make it especially useful for such
applications as testing servo-system or amplifier
response, distortion, and stability; FM generation and
frequency multiplication; or simply used as a variable
beat-frequency oscillator, repetition-rate, or tone-burst
generator. 'The square-wavetrigger output can beusedas
a source for transistor-transistor logic (TTL) or to syn-
chronize an external device such as an oscilloscope
or
counter.
The FG 501 is calibrated and ready for use when
received. Itisdesignedtooperateinanycompartment of
a
TM 500-Series power module only. Refer to the power
module Instruction Manual for line voltage requirements
-
and power module operation.
Installation and Removal
CAUTION
D
Turnthepower module off beforeinsertingtheplug-
in; otherwise, damage may occur to the plug-in
circuitry. Because of the high current drawn by the
FG
507,
it is also recommended that the power
module be turned off before removing the FG
507.
Refer to Fig.
7-2.
Check
to
see that the plastic
barriers on the interconnecting jack of theselected
power module compartment match the cut-outs in
the
FG
501
circuit board edge connector.
Align the FG 501 chassis with the upper and lower
guides of the selected compartment. Pushthe module in
and press firmly to seat the circuit board in the inter-
connecting jack.
Pull the Power switch on the front panel of the power
module to apply power to the FG 501. Observe that the
POWER indicator light on the FG 501 comes on.
Removethe FG 501 from the power module by pulling
the release latch at the bottom of the front panel and
sliding the unit straight out of the power module.
InterconnectingJack
Bottom Groove
Fig.
1-2.
Plug-in installationand removal.
REV.
C
JUL
1978
Operating Instructions-FG
501
OPERATING CONSIDERATIONS
NOTE
Before using the
FG
501
for the first time, read the
Operating Considerations in this section and the
description of the front-panel controls, connectors,
and indicators in Fig.
1-3.
Output Connections
The output of the FG 501 is designed to operate as a
voltage source inseries with 50
f2
andworking intoa 50
f2
load. At the higher frequencies, an unterminated or
improperly terminated output will cause excessive
aberrations on the output waveform (see lmpedance
Matching discussion).Loadslessthan50Rwill reducethe
waveform amplitude.
Excessive distortion or aberrations due to improper
termination is lesslikely tooccur at thelowerfrequencies
(especiallywith sineandtrianglewaveforms). However,to
ensure that waveform purity is preserved, observe the
following precautions:
1. Use quality 50
f2
coaxial cables and connectors.
2.
Makeall connections tight andas short as possible.
3.
Use quality attenuators, if necessary, to reduce
waveform amplitude to sensitive circuits.
4. Useterminators or impedance-matching devicesto
avoid reflections when using long cables, i.e., six feet or
more.
5. Ensure that attenuators, terminations, etc. have
adequate power-handling capabilities for the output
waveform (approximately 0.5 W into a 50
f2
load).
Power output isdetermined bytheselected waveform,
itsamplitude, and the amount of offset voltage selected.
Thephysical and electricalcharacteristicsof thepulse-
transmitting cable determine the characteristic im-
pedance, velocity of propagation, and amount of signal
loss. Signal loss, due to energy dissipation in the cable
dielectric, is proportional to the frequency; therefore, a
few feet of cable can attenuate high-frequency informa-
tioninafast-rise pulse.Itisimportanttokeepthesecables
as short as possible.
When signal comparison measurements or time
difference determinations are made, the twosiignalsfrom
the test device should travel through coaxial cables with
identical loss and time-delay characteristics.
Ifthereisadcvoltage acrosstheoutput load,theoutput
pulse amplitude will be compressed; or insorne cases, if
the voltage exceeds
f
10
V,
it may short the output. To
prevent this from occurring, the output must be coupled
through a dc blocking capacitor to the load. The time
constant of the coupling capacitor and load must belong
enough to maintain pulse flatness.
Risetime and Falltime
If the output pulse from the FG 501 is used for
measuring the rise or falltime of a device, the risetime
characteristics of associated equipment may have to be
considered. If the risetime of the device under test is at
least 10times greater than the combined risetimesof the
FG 501 plus the monitoring oscilloscope and associated
cables, the error introduced will not exceed 1% and
generally can be ignored. If the rise or falltime of the test
device, however, is less than 10 times as long as the
combined risetimes of the testing system, the actual
/-
risetimeof the device will havetobedeterminedfromthe
risetime of each component making up the system. This
equals the square root of the sum of the squares of the
individual risetimes. Conversely,therisetimeof thedevice
under test can be found from the same relationship if all
theactual risetimes inthesystemareknownexceptthatof
the device under test.
lmpedance Matching
Reflections.
As a pulse travels down a transmission
line, eachtime itencountersamismatch,or animpedance
different than the transmission line, a reflection is
generated and sent back alongthe linetothe source. The
amplitude and polarity of the reflections are determined
by the amount of the encountered impedance inrelation
to the characteristic impedance of the cable. If the
mismatch impedance is higherthantheline,thereflection
will be of the same polarity as the applied signal; if it is
lower, the reflection will be of opposite polarity. If the
reflectedsignal returns beforethe pulseisended, itadds
to or subtracts from the amplitude of the pulse. This
distorts the pulse shape and amplitude.
MatchingNetworks.
Thefollowingdescribes methods
for matching impedance networks into relatively low
impedances. If the FG 501 is driving a high impedance,
--
such as the
1
MR inputimpedanceof the vertical inputfor
an oscilloscope,thetransmission linemust beterminated
REV.
B
JUL
1978
Operating Instructions-FG
501
Fig.
1-3.
Operating controls and connectors.
REV
C,
MAR
1979
Operating Instructions-FG
501
into a 50
f2
attenuator and a 50
f2
termination at the
oscilloscope input. The attenuator isolates the input
capacity of the device. Distortion can be caused bythis
input capacity.
A simple resistive impedance-matching network that
provides minimumattenuation isillustrated inFig. 1-4. To
match impedancewith the illustrated network,thefollow-
ing conditions must exist:
(R1
+
Z2)R2must equal
zI
Ri
+
Z2+ R2
and
RI
+
-
'IR2
must equal
~2
ZI
+
R2
Therefore:
R1R2
=
2122;and RIZl
=
R2(Z2
-
Z1)
For example;tomatcha 50
f2
system toa125
f2
system,
Z1equals 50
Cl
and Z2 equals 125
0.
Therefore:
and R2
1
50dT125
-
50
=
64.6
f2
When constructing such a device, the environment
surrounding the components should also be designed to
provide a transition between the impedances. Keep in
mindthat the characteristic impedanceof acoaxialdevice
isdetermined bythe ratio betweentheoutsidediameter of
the inner conductor to the inside diameter of the outer
conductor.
z,
=
1381
E
loglo Dld, where D isthe inside
diameter of the outer conductor, and d is the outside
diameter of the inner conductor.
E
isthe dielectric cons-
tant (1 in air).
Attenuation Ratios.
Though the network in Fig. 1-4
provides minimum attenuation for a purely resistive
impedance-matching device,theattenuation asseenfrom
one end does not equal that seen from the other end.
A
signal (El) applied from the lower impedance source (Z1)
encounters a voltage attenuation (A1)which is greater
than 1 and less than
2,
as follows:
Zl
-
(Low)
2
-
(High)
Fig.
1-4.
Impedance-matching network that provides minimum
attenuation.
A signal (E2) appliedfromthehigher impedancesource
(22) encounters a greater voltageattenuation
(A2)
which is
greater than 1 and less than 2 (Z~IZI):
Inthe example of matching 50
f2
to 125
f2.
The illustrated network can be modified to provide
different attenuation ratios by adding anot.her resistor
(less than R1)between Z1and the junction of R1and R2.
Duration of Ramps and Pulses
The duration of ramp and pulse waveforrns is always
equal tothe half-cycle time of thesine, square, ortriangle
waveform frequency. For MULTIPLIER setti~ngsof 1 or
greater,the retraceloff time issuchthat thewaveformhas
a
duty cycle of approximately 80°/o, i.e., frequency equals
approximately 1.6X FREQUENCY
Hz
dial setting. For
MULTIPLIER settings less than 1, the retrace/off time is
from 10 ms to 100 ms, which results in duty cycles
---
approaching 10O0/0;i.e., frequency equals approximately
2X FREQUENCY
Hz
dial setting.
REV.
B
JUL
1978
Operating Instructions-FG
501
OPERATION
-A
Free-Running Output
The following procedure provides a free-running
waveform output with variable frequency and amplitude.
1. Set the AMPL control tothe fully counterclockwise
position and the OFFSET control to the 0 (centered)
position. Check that the PHASEcontrol ispushedin(off).
2.
Set the FUNCTION selector to the desired
waveform (see Fig. 1-5).
Fig.
1-5.
Output waveforms availablefrom the
FG
501.
3.
Select the desired frequency withthe MULTIPLIER
selector and FREQUENCY Hz dial. For example, if the
MULTIPLIER selector is set to the
lo5
position and the
FREQUENCYHz dial is at 5, output frequency is500 kHz;
i.e., MULTIPLIER setting
X
FREQUENCYHz setting. The
output frequency is calibrated when the FREQUENCY
VERNIER control is in the fully clockwise position. The
durationof rampand pulsewaveforms isdependent onthe
MULTIPLIER setting. See Duration of Rampsand Pulses
under Operating Considerations for further information.
4.
Connect the load to the OUTPUT connector and
adjust theAMPL control for the desired output amplituae.
Variable
DC
Offset
Pull outwardon theOFFSETcontrol (pullswitchadded
at SN B020000) to position the dc level (baseline) of the
output waveform. Forexample,$5
V
of offsetwill increase
the dc
+
peak ac voltage of a 7.5
V
p-p output to+5 and
+
12.5
V
dc
.t
peak ac while -5
V
of offset will reducethe
dc
+
peak ac output to +2.5
V
and -5
V.
Gated (Burst) Output and Variable Phase
A gating signal of
2
to 15
V
amplitude applied to the
GATE INPUT connector with the PHASE control pulled
out will provide a burst of cycles at the OUTPUT
connector. Thedurationof theburstand number of cycles
inthe burst depend on the gating signal durationandthe
output frequency selected.Whenthegatingsignal goesto
the zero level, the generator completes its last cycle and
remains quiescent until the next gating signal.
Single cycles can be obtained by applying a gating
signal with a periodapproximately equal tothe period of
the FG 501 output waveform. The number of cycles per
burst can be approximated by dividing the gating signal
duration by the period of the FG 501 output frequency
GATE INPUT
PHASE
control at
90"
LAG
GATE INPUT
PHASE
control at
0"
GATE INPUT
PHASE
control at
90"
LEAD
1
43
1
-04
Fig.
1-6.
Single cycle output with variable phase.
REV.
C
JUL
'1978
Operating Instructions-FG
501
The phase (start level) of the waveform burst can be
varied +90° by pullingout andturningthe PHASEcontrol
either counterclockwise or clockwise from the 0
(centered)position (see Fig. 1-6). The phaseof theoutput
burst isreferencedtothesineortrianglewaveform 0° start
point.
Output frequency can be varied during the burst
duration by applying a voltage-controlled frequency
(VCF) signal tothe VCF INPUT connector.
Voltage-Controlled Frequency (VCF) Output
Theoutputfrequency of any selected waveformcanbe
swept within a range of 1000:l by applying a 0 to 10 V
signal to the VCF INPUT connector. The polarity of the
VCF input signal determines which direction the output
frequency sweeps from the frequency slet by the
MULTIPLIER selector and FREQUENCY Hz dial; i.e., a
+
,-,
signal sweeps the frequency upward as shown in Fig.
1-7(A), a
-
signal sweeps the frequency doiwnward as
shown in Fig. 1-7(B).
Themaximum sweptfrequencyrangeof 1000:1encom-
passes the sensitive uncalibrated range of the FRE-
QUENCY Hz dial, i.e., <.l to 1. Therefore, to ensure that
the frequency does sweep at least a range of 1000:1, it is
recommended that the FREQUENCY Hzdial be set at 10
and a Oto-1OVsignal beappliedtotheVCF
IN1
connector.
Theoutput will thussweep downwardat least 1000:lfrom
a FREQUENCY Hz dial setting of 10 as shown in Fig.
1-7(B). It may be necessary to vary the CAL control to
obtain the full 1000:l swept range or the lowest swept
frequency desired.
SWEPT FREQUENCY
RANGE
(A) Uncalibrated
VCF
INPUT INPUT INPUT
-5
OUTPUT OUTPUT OUTPUT
Fig.
1-7.
Swept Frequency rangewith
10
V signals applied to VCF
IN
connector.
REV.
B
JUL
1978
Operating Instructions-FG
501
An input signal that varies symmetrically about a 0 V
level will also sweep the generator symmetrically about
the center frequency set bythe MULTIPLIER selectorand
-
FREQUENCY Hz dial as shown in Fig. 1-7(C).
Sincethe VCF input amplitude vs frequency isalinear
relationship, the frequency output range can be deter-
mined from the VCF input amplitude.
Hold
Mode
Three detented HOLD positionsare provided between
the lowest three MULTlPLlER selector positions. By
switchingtoany oneoftheHOLDpositions,thegenerator
can bestop~pedat itsinstantaneousvoltagelevel and held
there until the MULTIPLIER selector setting is changed.
Trigger
Olutput
A
TTL-corn pati
ble
+
2.5
Vsquarewaveisavailablefrom
the TRlG (OUTPUT connector. The frequency of the
trigger outlput is determined by the output frequency
selected by the MULTIPLIER selector and FREQUENCY
Hzdial (seeFig. 1-8).Whenthe FUNCTIONselector isset
for rampor pulse,thetrigger outputfrequency
is
about 1.6
.
times the dial indications. Output impedance is 600
n.
Triangle
Waveform
I
Sine Wave
I
I
I
I
I
I
I
Square Wave
TR
lG
OUTPUT
Waveform
I
I
I
I
Ramp Waveform
Duty cycle of the TRlG OUTPUT waveform follows
the duty cycle of the OUTPUT waveform.
1431-18
Fig. 1-8. Phase relationships between various waveforms from
OUTPUT
and
TRlG OUTPUT
connectors.
REV.
B
JUL
1978
Operating Instructions-FG 501
APPLICATIONS
Response
Analysis
The FG 501 is particularly suited for determining
response characteristics of circuits or systems. This
application utilizesthe VCF input of the FG 501 tosweep
the generator over arangeof frequencies.Byapplyingthe
desired waveformfromanother FG501(orequivalent) toa
device under test and sweepingthewaveform frequency
over a selected range, various response characteristics
can be observed on a monitoring oscilloscope.
The following procedure describes a technique for
determining response characteristics of any frequency-
sensitive device that operates within the frequency range
of the FG 501. Refer totheVoltage-Controlled Frequency
(VCF) Output discussion under Operation for additional
information.
1. Connect the equipment as shown in Fig. 1-9.
2.
Set theMULTlPLlER selector and FREQUENCY Hz
dial for the desired upper or lower frequency limit
(depending on the direction you wish to sweep).
3.
Apply the desired waveform to the VCF INPUT
connector. (A positive-going waveform will sweep the
frequency upwardsfrom the FREQUENCYHzdial setting
while a negative-going waveform will sweep clownwards.
4.
Adjust the amplitudeof theVCF input waveform for
the desired output frequency range.
5. Observe the response characteristics on the
monitoring oscilloscope.
The frequency at whi splayed respon
characteristic occurscan bedetermined byfirst removing
the VCF input waveform, then manually adljusting the
FREQUENCY Hz dial to again obtain the particular
characteristic observed inthe swept display and reading
that frequency on the FREQUENCY Hz dial.
WAVEFORM
I
UNDER
,
TEST
1
Fig.
1-9.
Analyzing circuit or system response.
REV.
B
JUL
1978
Operating Instructions-FG
501
Tone-Burst Generation or Stepped Frequency
Multiplication
The FG 501 can be used as a tone-burst generator or
\
._
frequency multiplier for checking tone-controlled
devices. This application utilizes a ramp generator, such
as the TEKTRONIX RG 501,as a VCFsignal source anda
pulse generator, such as the TEKTRONIX PG 501, as a
gating signal source.
The following procedure describes a technique for
obtaining a tone-burst or frequency multiplied output
from the FG 501. Refer to the Gated (Burst) Output and
Variable Phase and the Voltage-Controlled-Frequency
(VCF) Output discussions under Operation for additional
information.
1. Connect the equipment as shown in Fig. 1-10.
2.
Pull out the FG 501 PHASE control. Set the ramp
generator for the desired ramp duration and polarity.
3.
Adjust the pulse generator period for the desired
number of bursts within the selected ramp duration.
Adjust the pulse generator duration for the desired burst
width.
4. Select the sweep frequency range by adjusting the
FREQUENCY
Hz
dial for one end of the swept range
(upper or lower limit depending on the polarity of the
ramp). Then, adjust the ramp generator amplitudefor the
other swept frequency limit.
Various other tone-burst or frequency multiplied
characteristics can be obtained by usingdifferent gating
input waveforms, i.e., triangle, sine, square, etc.
VCF INPUT*
GATE INPUT*
UUUL
Tone-Burst or
Stepped Frequency
Output
*Gating pulsesynchronized with VCF RAMP.
PULSE
GEN
DEVICE
UNDER
Fig.
1-10.
Tone-burst generationor stepped frequency multiplication.
REV
A,
JUN
1978
Section
2-FG
501
SPECIFICATION
AND PERFORMANCE CHECK
SPECIFICATION
Performar~ceConditions
The electrical characteristics are valid only if the
FG 501 has been calibrated at an ambient temperature
between+20° Cand+30° C andisoperating at an ambient
temperature between 0" C and +50° C unless otherwise
noted. Forced air circulation is required for ambient
tem
perature above 4-40' C.
Only those items listed in the Performance Re-
quirements column of the Electrical Characteristicstable
are normallyverified when doingthe PerformanceCheck
procedure of thismanual. ItemslistedintheSupplemental
lnformation column are either explanatory notes or
minimum performance characteristics for which no
tolerance ranges are specified, and which normally re-
quireverification only after repairs or parts replacement.
Table
2-1
ELECTRICAL CHARACTERISTICS
Characteristic
Frequency
Range
Sine VVave, Square Wave,
and Tiriangle
Accuracy
Resolution
Stability
Temperature
Time
Pulse and Ramp range
Time Symmetry
Sine Wave, Square Wave,
and Triangle
--
-
--
-
Performance Requirement
0.01
Hz
to 1
MHz
in
9
decade steps.
Within 3% of full scale 1 to 10;
.1 to 1 uncalibrated.
Within 1% from 0.001
Hz
to
1
MHz
on calibrated portion (1to 10)
of FREQUENCY
Hz
dial, +20°C to
+50° C.
Supplemental lnformation
1 part in
lo4
of full scale with
FREQUENCY VERNlER control.
Within 2% from 0.1
Hz
to 1 MHz,
and within 10% from 0.001
Hz
to
0.1
Hz,
0°C to +SO0 C.
Within 0.1% for 10 minutes.
Within 0.25% for 24 hours.
=2X dial setting with MULTI-
PLIER at to+.6X dial set-
ting with MULTIPLIER AT
lo5
setting.
Within 10% on uncalibrated por-
tion (0.1 to 1) of FREQUENCY
Hz
dial.
REV.
6
JUL
'1978
Specification and Performance Check-FG
501
Table
2-1
(cont)
Characteristic
Amplitude (excluding offset)
Stability
Temperature
Time
Performance Requirement
SN B130000-up: 20
V
p-p open
circuit. 10
V
p-p into 50
fl
load.
Sine, triangle, and square wave
amplitudes matched within 5% for
singlesetting of AMPLlTUDEcon-
trol. Below SN B130000,OUTPUT
control provides 15
V
p-p open
circuit and 7.5
V
p-p into
50
fl
load.
Within 0.1% for 10 minutes.
Within 0.25% for 24 hours.
--
Supplemental
I
nformiation
Power-supply limiting causes com-
pression of output waveform
when maximum amplitude and max-
imum offset are used simu~lta-
neously.
Within 2% from 0.1 Hzto 1 MHz.
Wtihin 10% from 0.001 Hz to
0.1 Hz, 0°C to
+SO0
C.
Hold Mode Stability Within 5% of full output voltage in 1
hour at +25OC on 0.001
Hz:
range.
Offset
Amplitude
Range
I
I
Into Open Circuit
Into50 ohm Load
SN B130000-up:
+
or
-
7.5
V
SN below 8130000:
+
or
-
5
V
SN B130000-up:
+
or
-
5
V
SN below B130000:
+
or
-
2.5
V
-6
V
peak signal plus offset.
SN below B130000: At least
+
Into Open Circuit
Into 50 ohm Load
and
-
5
V
peak signal plus
offset.
At least
+
and
-
15
V
peak signal
plus offset.
SN B130000-up:At least
+
and
Amplitude
Output Impedance
Trigger Output
3
+2.5
V
square wave into a
600
n
load.
50
n.
Triangle and Ramp
Linearity (between
10%and 90% points)
Frequency
Within 1% from 0.081 Hz to
100 kHz excluding first 200 ns
after switch points.
Same as frequency at output con-
nector.
Within 2% from 100 kHz to 1 MHz,
excluding first 200 ns after
switch points.
-
REV
C,
MAR
1979
Specificationand PerformanceCheck-FG 501
Table
2-1
(cont)
characteristics
-
Ramp Duration
Sine Wave Distortion
Square Wave and Pulse
Outputs
Aberrations
Pulse Dluration
External Gate lnput
lnput Signal
Burst Length
Phasing
lnput lrnpedance
External V'oltage-
Controlled Frequency
(VCF) lnput
Output Frequency
Range
Slew Rate
Performance Requirement
1% or less from 0.001 Hz to 1 Hz.
0.5% or less from 1 Hz to 20 kHz.
1% or less from 20 kHz to 100 kHz.
2.5% or less from 100 kHz to 1 MHz
at 10' MULTIPLIER setting.
100 ns or less.
5% or less measured p-p with out-
put amplitude at 10 V into exter-
nal 50
0
load.
At least 1000:l with 10 V VCF
input. Negative-going voltage
decreases frequency; positive-
going increases frequency.
f,,
=
10X MULTIPLIER setting,
f.
=
mln
MULTlPLlER setting
100
Supplemental Information
--
(see Operating Consider-
2f ations).
Applies to calibrated portion of
dial only (1 to 10). Valid from
+lO°C to+50°C.
=
1
(see Operating Consider-
2f ations).
Square wave at least +2 V, but
not to exceed +15 V. Output bursts
are synchronized with gate input.
Determined by selected output
frequency and gating pulse width.
Continuously variable from -90@
to +90° referred to 0° sine or
triangle start points.
REV
C,
MAR
1979
Specificationand Performance Check-FG
501
Characteristics
Ternperature
Operating
Table
2-2
ENVIRONMENTAL CHARACTERISTICS
Storage
Altitude
Operating
Storage
Vibration
Operating and non-operating
Shock
Operating and non-operating
Transportation
Information
To 15,000feet. Maximum operating temperature decreased by
1
O
C/100 feet from 5000 to 15,000feet.
-
To 50,000 feet.
With the instrument complete, vibration frequency swept from
10to 55 to 10 Hz at 1 minute per sweep. Vibrate 15 minutes in
each of the three major axes at 0.015" total displacement. Holld
10 minutes at any major resonance; or, if none, at 55 Hz. Totid
time, 75 minutes.
30 g, 1/2 sine, 11 ms duration, 3 shocks in each direction
along 3 major axes, for a total of 18 shocks.
-
-
Qualified under National Safe Transit Committee Test Procedure
1A, Category
11.
Table 2-3
PHYSICAL CHARACTERISTICS
Characteristic
I
Dimension
Overall Size
(measured at maximum
points)
Height
Width 5.0 in (12.7 cm)
2.6 in (6.6 cm)
Length
Net Weight
(Instrument only)
12.2 in (31 em)
2 Ibs (0.906 kg)
Specification and Performance Check-FG
501
PERFORMANCE CHECK
Introduction
This procedurechecks the electrical characteristics of
the FG 501 that appear intheSpecification section of this
manual. Thils procedure can also beused byan incoming
inspection facility to determine acceptability of perfor-
mance. If the instrument fails to meet the requirements
given in this performance check, the adjustment
procedure should be performed.
Theelectrical characteristics inTable 2-1arevalidonly
if the FG 5011 is calibrated at an ambient temperature of
+20°C to+:30° C and operated at an ambient temperature
of 0°C to 4-50°C. Forced air circulation is required for
ambient temperature above +40° C.
Tolerances that are specified in this performance
checkprocedureapplytotheinstrument undertestanddo
not include test equipment error.
Test Equipment Required
Thetest equipment listed inTable 2-4, or equivalent, is
required to perform the performance check. Test equip-
ment characteristics listed are the minimum required to
verify the performance of the equipment under test.
Substitute equipment must meet or exceed the stated
requirements. All test equipment is assumed to be
operating within tolerance.
Table
2-4
LIST
OF
TEST EQUIPMENT REQUIREMENTS
Performance
Description
Oscilloscope
Power Module
Digital Voltmeter
Frequency Counter
Pulse Generator
Variable dc
Power Supply
Requirement
Bandwidth dc to 15 MHz;
deflection factor 10 mV/
div to 5 V/div; sweep
rate 20 ns/div to 1 ms/
div.
Three compartments or
more.
Range 0 to +20 V dc; dis-
played error less than
0.5O/o.
Frequency range 0.1 Hz
to above 1 MHz; accuracy
within one part of
lo5
+1 count.
0 to +2 V square-wave
output into 50 load.
Period 0.2 ms; duration
0.1 ms.
Output 0 to 20 V at 0.4
A
or greater.
Application
Steps 1, 2, 3,
5,
7,
8,
and 9.
All tests.
VCF INPUT and Offset
range checks.
Basic timing
&
VCF
INPUT.
Phase range check.
Check VCF INPUT
Example
TEKTRONIX T921 or equiv-
alent.
TEKTRONIX TM 503,
TM 504, or equivalent.
TEKTRONIX DM 501a
TEKTRONIX DC 504"
or equivalent.
TEKTRONIX PG 501"
or equivalent.
TEKTRONIX PS 501"
or equivalent.
"~equires
Thll
500-Series power module.
Specificationand PerformanceCheck-FG
501
Table
2-4
(cont)
Performance
Distortion
Analyzer to at least 600 kHz. Dis-
tortion resolution<0.5O/o.
Application Example
Hewlett-Packard 334~Dis-
tortion Analyzer or equiv-
alent.
Check sine wave dis-
tortion.
50
R
Feedthrough
Termination (2) bnc connectors. Steps 1, 2, 3, 5, 6,
8,
and 9. Tektronix Part No.
011-0049-01.
600
R
Feedthrough
Termination bnc connectors. TRIG OUTPUT Amplitude
check Tektronix Part No.
011-0092-00.
bnc connectors.
50
R
Coaxial Cables
(2 ea) Tektronix Part No.
012-0057-01.
All.
Adapter bsm-to-bnc. TRIG OUTPUT Amplitude
check. Tektronix Part No.
103-0036-00.
Dual banana plug-to-bnc
female.
Adapter VCF INPUT check. Tektronix Part No.
103-0090-00.
bnc connectors.
Tee Connector Basic timing check. Tektronix Part No.
103-0030-00.
IOX
Attenuator bnc connectors
50
R
impedance. Square wave checks. Tektronix Part No.
011-0059-02.
PRELIMINARY PROCEDURE PERFORMANCE CHECK PROCEDURE
1. Ensurethat the correct nominal line selector block
has been installed on the lineselector pinson the power
module interface board, andtheregulatingrangeselected
includes the applied line voltage. Refertothe installation
section of the power module manual.
1.
Dial Alignment
a. Set the FG 501 controls as follows:
FUNCTION Triangle
AMPL'
FUIIY
cw
OFFSET^
Midrange and In
PHASE In
MULTIPLIER 1
o3
FREQ VERNIER Fully cw
FREQUENCY Hz Near 10
2. Ensurethat all test equipment issuitably adaptedto
the applied line voltage.
3. Install the FG 501 into the power module and, if
applicable, install the TM 500-Series test equipment into
the test equipment power module. b. Adjust the oscilloscope vertical for dc coupling at
2 V/div sensitivity. Set the time base sweep speed to
.1 ms/div. Set the triggering controls to internadsource+
slope operation.
4. Connect the equipment under test and the test
equipment to a suitable line voltage source. Turn on all
equipment and allow at least20minutesfortheequipment
to stabilize.
mi el ow
SN
B130000
AMPL control is labeled OUTPUT.
2~orSN below
B130000
set OFFSET to zero.
c. Connect theOUTPUT of the FG 501through
a
50
R
coaxial cable and a 50
R
termination, with the 50
R
termination at the vertical input of the oscilloscope.
-
REV
A,
MAR
1979

Other manuals for FG501

2

Other Tektronix Portable Generator manuals

Tektronix HDLG7 User manual

Tektronix

Tektronix HDLG7 User manual

Tektronix RTX130B User manual

Tektronix

Tektronix RTX130B User manual

Tektronix AWG520 User manual

Tektronix

Tektronix AWG520 User manual

Tektronix CFG253 User manual

Tektronix

Tektronix CFG253 User manual

Tektronix FG501 User manual

Tektronix

Tektronix FG501 User manual

Tektronix ECO422D Manual

Tektronix

Tektronix ECO422D Manual

Tektronix AFG2000 User manual

Tektronix

Tektronix AFG2000 User manual

Tektronix VITS 201 User manual

Tektronix

Tektronix VITS 201 User manual

Tektronix AWG5012 User manual

Tektronix

Tektronix AWG5012 User manual

Tektronix SW503 User manual

Tektronix

Tektronix SW503 User manual

Tektronix AWG7000 Series User manual

Tektronix

Tektronix AWG7000 Series User manual

Tektronix DTG5078 User manual

Tektronix

Tektronix DTG5078 User manual

Tektronix TSG-271 User manual

Tektronix

Tektronix TSG-271 User manual

Tektronix DG2030 User manual

Tektronix

Tektronix DG2030 User manual

Tektronix DTG5078 User manual

Tektronix

Tektronix DTG5078 User manual

Tektronix AWG5204 User manual

Tektronix

Tektronix AWG5204 User manual

Tektronix PPG3000 Series User manual

Tektronix

Tektronix PPG3000 Series User manual

Tektronix AWG410 User manual

Tektronix

Tektronix AWG410 User manual

Tektronix AWG5000B Series User manual

Tektronix

Tektronix AWG5000B Series User manual

Tektronix SG 504 User manual

Tektronix

Tektronix SG 504 User manual

Tektronix AWG70001A Use and care manual

Tektronix

Tektronix AWG70001A Use and care manual

Tektronix TG700 User manual

Tektronix

Tektronix TG700 User manual

Tektronix SPG8000A User manual

Tektronix

Tektronix SPG8000A User manual

Tektronix SPG70UP User manual

Tektronix

Tektronix SPG70UP User manual

Popular Portable Generator manuals by other brands

Fischer Panda 8000 NE Operation manual

Fischer Panda

Fischer Panda 8000 NE Operation manual

Onan CCKB Operator's manual

Onan

Onan CCKB Operator's manual

Bohmer-AG W Series instruction manual

Bohmer-AG

Bohmer-AG W Series instruction manual

Amico AH7500E owner's manual

Amico

Amico AH7500E owner's manual

Hydra-Stop Hydra-Power Operation instructions

Hydra-Stop

Hydra-Stop Hydra-Power Operation instructions

Festo OVEM Series operating instructions

Festo

Festo OVEM Series operating instructions

MQ Power DCA220SSJU4F2 Operation manual

MQ Power

MQ Power DCA220SSJU4F2 Operation manual

WhisperPower W-GV/2 user manual

WhisperPower

WhisperPower W-GV/2 user manual

BOSSCO ODOROX owner's manual

BOSSCO

BOSSCO ODOROX owner's manual

Watt & Sea CRUISING 300 Installation and instruction manual

Watt & Sea

Watt & Sea CRUISING 300 Installation and instruction manual

BOSSCO DY8500 Series owner's manual

BOSSCO

BOSSCO DY8500 Series owner's manual

Concept2 SENTINEL S35 manual

Concept2

Concept2 SENTINEL S35 manual

home diesel power 7500W owner's manual

home diesel power

home diesel power 7500W owner's manual

Emax EGES14020T operating instructions

Emax

Emax EGES14020T operating instructions

Sel Genesis 260AC-DC Use and maintenance manual

Sel

Sel Genesis 260AC-DC Use and maintenance manual

Delta OHM HD2040 manual

Delta OHM

Delta OHM HD2040 manual

Novatech Instruments 408A instruction manual

Novatech Instruments

Novatech Instruments 408A instruction manual

Wesco WES-250 manual

Wesco

Wesco WES-250 manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.