tempmate S2 User manual


empower your supply chain.
V1.0-11/2022-EN · Technical Changes and Errors excepted 3
1. Introduction ............................................................................................................................................................................................ 3
2. Intended use ...........................................................................................................................................................................................3
3. tempmate.®-S2 models ................................................................................................................................................................... 4
4. Device description .................................................................................................................................................................................6
5. Display indicators ................................................................................................................................................................................. 7
6. Quick Start Guide...................................................................................................................................................................................8
7. Operation and use ..............................................................................................................................................................................10
Conguration..............................................................................................................................................................................10
Startlogger................................................................................................................................................................................10
Set mark ......................................................................................................................................................................................11
Stoplogger.................................................................................................................................................................................11
Manual readout .........................................................................................................................................................................11
Readout with tempbase software.......................................................................................................................................12
8. Important notes ..................................................................................................................................................................................13
9.MainTechnicalSpecicationstempmate.®-S2 T...................................................................................................................14
10.MainTechnicalSpecicationstempmate.®-S2TH............................................................................................................15
13. Contact information ........................................................................................................................................................................16
Table of contents
Thetempmate.®-S2isdesignedtobemountedona
broadcast or stationary and measure relevant para-
meters such as temperature and optionally relative
humidity. The device records data and stores it on an
internal memory.
1. Introduction 2. Intended use
The tempmate.®-S2 is designed to be mounted on
shipments or stationary and to record relevant parame-
ters as mentioned in the datasheet. Any use or opera-
tionrequiringspecicrequirementsandstandardsnot
explicitly mentioned in the data sheet must be validated
and tested at the customer‘s responsibility.

empower your supply chain.
V1.0-11/2022-EN · Technical Changes and Errors excepted 5
3. tempmate.®-S2 models
Single-use
Temperature
Rel. humidity
LCD

empower your supply chain.
V1.0-11/2022-EN · Technical Changes and Errors excepted 7
4. Device description 5. Display indicators
LCD
LED Indicator
Start/Stop
Button
Sensor rel.
Humidity
Not started
Started*
Stopped
Number
measured values
Highest
measured value
Lowest
measured value
Averagevalue
MKT
Time & Date
*Ifthestartdelayisset,thissymbolashesaftera
successful start until the time limit has elapsed.
LCD
LED Indicator
Start/Stop
Button
PresstheStartbuttonrepeatedlytocalluptheindividualpagesonthedevicedisplay.

empower your supply chain.
V1.0-11/2022-EN · Technical Changes and Errors excepted 9
RECORD STATUS
Rec.
Stop
Mark
Mark
3 sec.
Rec.
Rec.
Stop
Stop
RECORD STATUS
Rec.
Stop
Mark
Mark
3 sec.
Rec.
Rec.
Stop
Stop
6. Quick Start Guide

empower your supply chain.
V1.0-11/2022-EN · Technical Changes and Errors excepted 11
7. Operation and use
STEP 1 Conguration*optional
Thisstepisonlynecessaryifyouwanttoadaptthealreadypre-installedcongurationtoyourapplication.
- Download the free tempbase 2 software. – https://www.tempmate.com/de/download/
- Install the tempbase 2 software on your PC.
- RemovethecapandconnectthenotstartedloggertoyourPC.
- Openthetempbase2softwareandselectthe„LoggerSetup“button(1).
- Makethedesiredsettingsandsavethemonyourdeviceusingthe„SaveParameter“button(2)
- RemovetheloggerfromyourPCandreplacethecapsecurely.
STEP 2 Startlogger
- Press and hold the blue button for 5 seconds.
- AsuccessfulstartisindicatedbythegreenLEDonyourdeviceashing10times.
- Important:Ifanotherornoashingsignalappears,donotusetheloggerandcontactsupport.
STEP 3 Set mark
- Press the blue button twice in succession to set a mark.
- Asuccessfullysetmarkerisindicatedbytheword„MARK“andthenumberofmarkerssetsofar
in your display.
- Important: Up to 10 marks can be set per operation.
STEP 4 Stoplogger
- Press and hold the blue button for 5 seconds.
- AsuccessfulstopisindicatedbytheredLEDonyourdeviceashing10times.
Alternative stop modes
Automaticstop(defaultsetting)
o Thedevicewillstopautomaticallywhenthemaximumnumberofreadingsinthe
data memory is reached and no manual stop is performed beforehand.
o This stop mode works in addition to the manual stop..
Softwarestop(optional)
o Thissettingcanbemadeinthetempbase2software.(seeSTEP1)
o ThestopistriggeredautomaticallybyconnectingtheloggertothePCand
openingthesoftware.
o Amanualstopisnotpossibleinthisconguration.
STEP 5 Manual readout
- RemovethecapandconnecttheloggertoyourPC.
- AsuccessfulconnectionisindicatedbybothLEDsashing.TheabbreviationsCSVandPDFappear
one after the other in the display.

empower your supply chain.
V1.0-11/2022-EN · Technical Changes and Errors excepted 13
- TheloggerwillautomaticallyopenasanexternaldriveonyourPC.Thisprocessmaytakeafew
minutesdependingontheamountofdata.
- OpenthedriveandcopythePDFandCSVreportstoredtoalocalfolder.
- Important:AreportisautomaticallygeneratedasPDFand/orCSVwhenthedeviceisstopped.
Thedevicecanstillbereadoutduringarunningmeasurementandanintermediatereportcan
be downloaded.
Readoutwithtempbase2software(optional)
- RemovethecapandconnecttheloggertoyourPC.
- Openthetempbase2softwareandselectthe„Export/Import“button(3).
- Selectthedesiredleformat(PDF/XLS/IME)forexportandthelelocationandconrmthedownload.
8. Important notes
- Thecongurationofyourdevicecannotbechangedduringrecording.
- IfyouconnectthedevicetothePCwithoutamanualstop,atemporarystopoftherecordingis
triggered.Youcannowdownloadaninterimreport.AfterdisconnectingthedevicefromyourPC,the
recordingwillcontinueautomatically.
- Alwaysdisposeofdevicesaccordingtoyourcountry‘sregulations.
- Do not place the device in corrosive liquids and do not expose it to direct heat.

empower your supply chain.
V1.0-11/2022-EN · Technical Changes and Errors excepted 15
MainTechnicalSpecicationstempmate.®-S2 T
RecordingOptions Single-Use
Dimension [mm] 93 x 47 x 8 mm
Weight[g] 30g
Casing EcofriendlyHardcase
Battery CR2450 lithium metal button cell battery
Connection Interface USB2.0,A-Type(integrated)
Display Multifunction LCD
Protection Class IP65
Shelf Life 24 months
TemperatureRange –30 °C to 70 °C
Temperature Accuracy ±0.3°C(-20to40°C)±0.5(others)
Temperature Resolution 0.1 °C
HumidityRange n/a
HumidityAccuracy n/a
HumidityResolution n/a
AmbientLightSensor Yes
MemoryCapacity(measurements) 32,000points(T)
RunTime(LoggingInterval) Upto180days(10min.)=StandardModel
Other models only on request
Data Export PDF&CSV
AlarmConguration upto5pointscustomizable,singleorcumulative
Startup Mode Button
Stop Mode Button or by Software
OtherSpecications
TemporaryPDF ThePDFReportisavailableatanytime
withoutstoppingtheLogger
Software PDFreader+optionaltempbase2Software
Reprogrammable Yes,withoptionaltempbase2Software
Backup-Log-Feature Yes
Custom ID and Travel Description Yes
CalibrationCerticate Yes,with6points(hardcopy)&PDFinmemory
Conformity EN12830,CE,RoHS
SerialNumberIdentication Yes
MainTechnicalSpecicationstempmate.®-S2 TH
RecordingOptions Single-Use
Dimension [mm] 93 x 47 x 8 mm
Weight[g] 30g
Casing EcofriendlyHardcase
Battery CR2450 lithium metal button cell battery
Connection Interface USB2.0,A-Type(integrated)
Display Multifunction LCD
Protection Class IP52
Shelf Life 24 months
TemperatureRange –30 °C to 70 °C
Temperature Accuracy ±0.3°C(-20to40°C)±0.5(others)
Temperature Resolution 0.1 °C
HumidityRange 0-100%rH
HumidityAccuracy ±3%rH(20to80%rH)±5%(others,at25°C)
HumidityResolution 0.1%rH
AmbientLightSensor Yes
MemoryCapacity(measurements) 32,000points(T/rH)
RunTime(LoggingInterval) Upto180days(10min.)=StandardModel
Other models only on request
Data Export PDF&CSV
AlarmConguration up to 5 points temperature and 2 points humidity customizable
Startup Mode Button
Stop Mode Button or by Software
OtherSpecications
TemporaryPDF ThePDFReportisavailableatanytime
withoutstoppingtheLogger
Software PDFreader+optionaltempbase2Software
Reprogrammable Yes,withoptionaltempbase2Software
Backup-Log-Feature Yes
Custom ID and Travel Description Yes
CalibrationCerticate Yes,with6points(hardcopy)&PDFinmemory
Conformity EN12830,CE,RoHS
SerialNumberIdentication Yes
Sinceenvironmentalinuencescanaffectthelifespanofbatteries,werecommendthatallourproductsbestoredintheconditions
thatareoptimalforthem.Inthisway,theshelflifegivenatthetimeofpurchasecanbeguaranteed.
Optimal storage conditions: +15°C to +23°C / 45% to 75% rH.

empower your supply chain.
V1.0-11/2022-EN · Technical Changes and Errors excepted 17
Do you have any questions? Please contact us - our
experienced team will be happy to support you.
+49713163540
Contact information

tempmateGmbH
Wannenäckerstr. 41
74078Heilbronn,Germany
Tel.+49-7131-6354-0
www.tempmate.com