Tera TOM 584 TF User manual

TERA Ohmmeter
TOM 584 TF
The TERA Ohmmeter TOM 584 works with the Current - Voltage- Measurement.
This is the inexpensive version of the TOM 374.
High resistances up to 2x1012Ohm (2 Tera-Ohm) can be measured with 100 volts
and a precision of ±1 x10x

2
List of contents
More Instruments from „Kleinwächter GmbH“ 3
1.Description 4
2. Operation Instruction 4
2.1 Start up 4
2.2 Charging the battery 4
2.3 Ranges and Voltagees 5
3. Set – Up Functions 5
3.1 Timer 6
4. Other Display Information’s 6
4.1 Display Contrast 6
5. PC – Operating 6
5.1 System requirements 6
5.2 Softwareinstallation 6
5.3 Starting the Software 7
5.4 Software Manual 7
6. Warranty 8
7. Safety Notice 8
8.Specifications 8
9. Parts of Delivery 9
10. Optional accessories 9
11. Maintenance / Calibration 9
12. Legend 10
Conformity explanation 11

3
More Instruments from „Kleinwächter GmbH“
TERA Ohmmeter TOM-374
The TERA Ohmmeter TOM-374 works with the voltage measuring process. An Electro Field Meter is
integrated as input amplifier with its very high impedance. Through it, it is possible, to measure resistances
directly against earth. High resistances up to 1014 Ohm (100 Tera Ohm) can be measured with a maesure
voltage of only 100 volts and an accuracy of 5%.
Through the built-in microprocessor, a very high comfort is guaranteed.
Charge Plate Monitor CPM-374
The CHARGE PLATE Monitor 374 measures the voltage potential on its integrated Plate-Electrode
according the Fieldmill-Influence-Principal and is suitable to determine the Static Decay Time from a
selected higher voltage threshold to a lower voltage threshold level.
The unit meets the requirements of the European Norm EN 100015 and the American Standard
EOS/ESD S.3.1.
At the upper side of the appliance, there is a plate condenser 150 x 150mm with a capacity of 20pF
±2pF.
This plate will be charged by an internal High Voltage supply with >± 1200 V and then measured the
discharge time in an ionization air. The start and stop voltage can be selected. The pos. and neg.
discharge time as well as the offset voltage will be shown in the display after the end of measurement.
The quality of the area ionization can be decided from it.
Electro field meter
EFM 022 Small hand-held Electro - Fieldmeter with digital display designed to measure electrostatic
voltage potentials (with pre-selected distance) according the fieldmill influence principal.
EFM 120 small very solid Electro - Field meter with high sensitivity to measure the
Electro – DC – Field and electrostatic Charge. The display is a 21 point LED
Bar graph with green Zero – LED in the middle.
EFM 2XX supersensitive Electro - Field Meters for laboratory use, to measure Electrostatic charges
down to < 1V. Appliances in pistol casings with analog - or digital display.
All the Electro - Fieldmeters works with the „Feldmühlen-Influenz-Prinzip“. By the use of a microcomputer,
it will increase the service comfort as well as the operational safety further.
No radioactive substances are used.
More informations !
info@kleinwaechtergmbh.de
Forschungs-, Entwicklungs-, Produktions- und Vertriebsgesellschaft mbH
D-79688 Hausen · Krummattstr. 9 · Tel. + 49 (0) 7622 / 6863 -- 0 · FAX – 40

4
1. Description
The TERA Ohmmeter TOM 584 is suitable very well for the mobile use in industrial areas through its
compact housing and the battery power supply. It can also be used stationary with a plug power supply.
The TOM 584 is menu-controlled, has only two keys and is very operator-friendly.
All parameters are shown in the LC - Display.
The included PC-Software is for remote controlled usage, and additionally measuring results can be
administered and processed.
This TOM 584 works with the current - voltage - principle.
Besides the electrode connection sockets, also an additional ground socket exists. The appliance should be
grounded to eliminate parasitic effects (influence effect).
To the transactions of the measurements anuses of tied-up Norm DIN EN 61340 / EOS-ESD 4.1/6.1 can
the measuring time be selected with the internal timer.
With Timer = ON the measurement values inclusive the air parameters can be stored in the internal
EEPROM. They don’t are deleted by switching off the unit. The values can be read out trough the PC-
Interface at all times.
2. Operating Instruction
2.1. Start operation
The TERA Ohmmeter TOM 584 is delivered with a 9V NiMH – Battery, and is directly ready to start.
Only the enclosed plug power supply should be used for charging the battery, and stationary work.
• The electrodes are connected to the corresponding sockets (s. legend) and placed on the probe.
• The measuring instrument should be grounded.
• During the measurement of very high-ohm-resistors is to be kept that no influence effect is happen on
the measure input.
• To Switch On the unit press key „B“.
After switching on, first the software version is shown. Provided the timer is „ON“, in the display appear
press start. You can start the measure by pressing key „A”
The current resistance value and the timer attitude are shown afterwards.
If the timer should be turned off, immediately appears in the display:
>2.0xE12Ω
ΩΩ
ΩT=OFF
• To switch off the appliance press again key „B”.
• In the battery mode the unit switches off, if no button was pressed, automatically after approximately
5 minutes.
2.2 Charging the 9V Battery
The TOM 584-TF is allowed to operate only direct with the enclosed power supply. Also only with this the
installed NiMH – battery should be loaded. Connect the power supply to the integrated socket, and but the
plug power supply to a power socket. The NiMH - battery is completely charged after maximum 14 hours
and supplies the TOM 584 approximately 2 hours in the continuous operation.

5
2.3. Measure ranges and measure voltages
The TOM 584 has an automatic ranging approve the hole resistor range. The appliance, accordingly the
adjacent resistance, selects the Measure voltage automatically.
• Measured resistor less then 200 kΩ—- Measure Voltage 10 VDC
• Measured resistor higher then 200 kΩ—- Measure Voltage 100 VDC
3. „set-up“ Function
Through simultaneous pressing the two keys „A” and „B” the „set-up" function is called.
In the display appears shortly:
SET TIMER !
After that it appears: ↓
↓↓
↓ON TIMER OFF↓
↓↓
↓
By pressing key „B“, the timer is switched off and the unit makes a reset.
By pressing key „A“, the timer is switched on and the countdown - time can be selected between1 and 240
seconds. In the display appears:
↓
↓↓
↓OK T=001s STEP↓
↓↓
↓
By pressing key „B“, it appears T= 002s and than in 1 second Steps up to10 second.
After that the timer starts with T=010s - T=015s - T=020s the changing steps are now 10 Seconds up to 60
Seconds, after this you get 60 second steps up to 240 seconds.
By pressing the „A“ key you choice the timer time and in the display appears:
↓
↓↓
↓AVERAGE LAPSE↓
↓↓
↓
By pressing key „A“, the middle value over the measure time will be displayed (AVERAGE).
By pressing key „B“, the last value after the end of the measure time will be displayed (LAPSE).
Now, it is asked whether they should be deleted in the EEPROM stored data (file). In the display, it appears:
↓
↓↓
↓Y DEL FILES NO↓
↓↓
↓
By pressing key „B", no files will be deleted, and the next files are saved with File No x+1, than the unit
makes a reset.
By pressing key „A“, all saved files will be deleted, and the next file to save in the EEPROM starts with File
No. 001. In the display appears for approximately 2 seconds:
WAIT . . .
After that appears FILES DELETED !
Now the Unit makes an Reset and after that in the display appears:
↓
↓↓
↓START T=xxxs
By pressing key A “Start” the next measurement will be started.
The timer decrements in 1 second steps up to Zero, then the measure value will appears in the display as
Ravg by AVERAGE - Mode and Rlap by Lapse – Mode
After 5 second it appears:
↓
↓↓
↓FN:xxx T=xxxs
By pressing key A „Start“ the measure value together with the air temperature and the Humidity will be
saved in the EEPROM using the displayed File - No. These values then later can be read out over the
PC - Interface.

6
3.1 Timer
If the timer was not activated, the TOM 584 directly switches on the resistance display in the on-line
operation. In the display appears:
=5.5xE10Ω
ΩΩ
ΩT=OFF
4. Other Display Information’s
If the maximum measurement is exceeded appears in the display:
>2.0xE12 (T=OFF)
If the minimal measurement becomes under-stride appears in the display:
<2.0xE04Ω
ΩΩ
ΩT=OFF
If the Battery Voltage is lower than 7,6 Volt, report the unit in the change with the current line:
LOW BATTERY !
The accumulator must be charged again, however the current measurement can quietly be completed. If the
Battery Voltage is lower as 7,0 volts, following display appear one after the other. Then the unit
automatically switches off, to avoid a destroying of low Voltage of the battery.
LOW BATTERY
AUTO OFF
SWITCHING OFF
4.1 Display contrast
On the right side of height of the display there is a hole trough it you can screw a potentiometer for adjust
Contrast of the LCD - Display.
5. PC - Operating
5.1 System requirement
• PC with 486er CPU or higher
• 3,5” Floppy disk drive 1,44MB (HD)
• free serial Interface COM1 or COM2
• Windows98, 2000, XP with EXCEL
5.2 Installations
Put the disk with the TOM software into the floppy drive. Go to „Start“ then Execute. Type Install and press
<Enter>. Now the software will be installed.

7
5.3 Start the Software
Connect the TERA – Ohmmeter with the PC – Interface cable to a free serial PC Interface
(COM 1 or COM 2) from your PC, and press key „A“ to switch on the TERA – Ohmmeter.
Start under Windows your EXCEL Program.
Open in Excel the file „UT584.xls“ in the directory c:\tom584
For more information about this program use the HELP bottom
5.4 Software Manual
To select the Menüpunkte use Coursor - Buttons ↑↓. With the <Enter> Button Menüpunkt will be started.
→Change start condition
Timer ? (0-240s) give in direct the time for the count-down (0 – 240 sec).
Average ? [y/n] with “y” the mid-value (Average) of all the measurements, during
the count-down timer is running, are calculated.
with „n” the last measured value is shown in the display (Lapse).
Important ! The changed items are not stored in the EEPROM !
→Start Measurement
With <Enter> the count-down timer for the measurement cycle will be started. After the count-down
timer returned to zero it is possible to store the complete measurement-curve.
(100ms – sample-time).
Save Data ? [y/n] with “y” the measured values will be stored in the file MV.DAT .
These File (MV.DAT) can be imported and modified with a
spread-sheet analysis (ex. EXCEL).
For separation of the data-cells a comma is used.
Important ! With each saving the file MV.DAT will be overwritten !
If you want to store different files, you need to rename the actual file MV.DAT.
Then you can store the next MV.DAT.
→Save Data - Files from TOM
with <Enter> the complete data-memory (in the EEPROM of the TOM) are transmitted and stored
as the file “FILE.DAT”. The modification- procedure is like for the “MV.DAT”.
→Reset Tera
with <Enter> a reset the processor in the TOM will be started and the selected values in the memory
of the TOM are transmitted to the PC.
→Exit
with <Enter> the programm will be closed.

8
6. Warranty
We provide 24 month warranty from date of receiving the unit, if the unit was handled properly.
The damage of the NiMH-battery caused by improper handling as well as mechanical damage on the unit
are excluded from warranty. Warranty is not provided, if the housing was opened.
7. Safety Advice
The measurement unit is not approved for the use in explosive areas!
The unit is not approved for the use in power utilities!
Only measurements can be executed at voltage-free objects.
8. Specifications
Housing : Plastic Housing 200 mm x 100 mm x 40 mm (L x B x H)
Weight : app. 350 g
Display : alphanumeric Display, 1 line with 16 Letters,
Display field 60 mm x 15 mm
Measure Ranges : Resistor 20 kΩ- 2.0 TΩTolerance ± 1xEXX Ω
PC - Interface : serial Interface, 9-pin SUB-D - Connector
PC - Software : Excel – Software for Data transmitting and remote control
Battery : NiMH 9V / 160 mAh to IEC 6 F 22
Operating time of completely loaded battery: approximately 1 hours of
continuous operation
Charge time with enclosed power adapter: maximum 14 hours
Power Supply : 9V-DC / 300 mA
9. Parts of Delivery
• Tera – Ohmmeter TOM 584
• 9V NiMH - Battery
• Plug Power supply 9V-DC / 300mA
• 2 pce. Teflon insulated measurement lead, 1m
• 1 groundig cable with aligator clip
• Serial PC - Cable 9 pol.
• 3,5“ Disk with PC - Software
• Foam Packing
• Operating Instructions
• Certificate of factory calibration

9
10. Optional accessories
To the TOM 584 are offered optional different Measure Electrodes with the appropriate cables:
Accessories set with:
ESD - suitcase with sufficiently place for
the measuring instrument plus all accessories
2,50 kg Electrode pair trough EOS/ESD S 4.1/S 7.1
DIN EN 61340 –5-1, -4-1, -2-3
For Measuring Grounding Resistors and Point to Point
Resistors, specially for installed ESD Table- or floor
coverings
2 pce. Teflon isolated measure cords 5m
• Ring Electrode to EOS/ESD S 11.11 + DIN EN 61340 –5-1, -2-3
comprising to the measurement of transit and surface resistances, 1 shielded and 1 no shielded
Measure cord.
• Other electrodes are available on inquiry
Info: Make sure, that there is no voltage on your measuring. External Voltage gives you a wrong
measure value and can destroy your instrument !
11. Maintenance / Calibration
The appliance is maintenance-free. If the appliance should get dirty, it can be cleaned with a lint-free cotton
cloth with solvent-free detergent. Don’t open the appliance. If the appliance should not be used for a longer
time, remove the 9V Battery from the appliance.
The recommended interval for factory calibration is 1 year.

10
12. Legend
1 Power Supply connector 9-12V DC
2 Serial PC – Interface
3 Measure Voltage Output
4 Measure Input
5 Contrast – Potentiometer for the LC - Display
6 Ground connector

11
Konformitätserklärung
DieFirma
Kleinwächter GmbH
Krummattstr. 9
79688 Hausen i.W.
erklärt in alleiniger Verantwortung, dass die Produkte
TERA Ohmmeter TOM 584 / TOM 584 TF
Messgeräte zur Messung hochohmiger Widerstände bis 2 Teraohm
auf die sich diese Erklärung bezieht, die Forderung
der EWG-Richtlinie über die elektromagnetische Verträglichkeit
Richtlinie des Rates vom 3.5.1989 (89/336) Stand: Juli 1993
erfüllt, insbesondere der Normen
EN 61010 „Safety“
EN 55011 Gruppe 1, Klasse B
EN 50082-2/92
für ISM-Geräte.
Die Firma Kleinwächter GmbH hält folgende technischen Dokumentationen zur Einsicht bereit :
• vorschriftsmässige Bedienungsanleitung
• Baupläne
• technische Dokumentationen
Hausen, den 09.01.2002
Jürgen Brunner
Geschäftsführer
This manual suits for next models
2
Table of contents
Other Tera Measuring Instrument manuals
Popular Measuring Instrument manuals by other brands

socomec
socomec DIRIS A40 operating instructions

Wabtec
Wabtec R-450 Installation

Kane
Kane KM9106 Operator's manual

Ballantine
Ballantine 340 instruction manual

Agilent Technologies
Agilent Technologies 16700 SERIES LOGIC ANALYSIS SYSTEM 16700 Product overview

Digital-Ally
Digital-Ally Laser Ally Speed LIDAR Operator's guide