Teseo NSG 3040 User manual

1
NSG 3040 EMC TEST SySTEM
USER MANUAL
601-279B

NSG 3040 EMC Test System
NSG 3040 EMC TEST SySTEM
USER MANUAL

1 Explanation of symbols 7
2 Introduction 8
2.1 General description 8
3 Standards and applications 10
3.1 Burst test 10
3.2 Combination wave test 11
3.3 Mains quality test 12
3.4 Magneticeldswithmainsfrequency(option) 12
3.5 Pulsedmagneticelds(option) 12
4 Safety instructions 13
4.1 General 13
4.2 Installation 14
4.3 InstallationofanEUTpowerswitch 15
4.4 Applicablesafetystandards 16
4.5 Testexecution 17
4.6 Userwarnings-Generator 18
4.7 Userwarnings-EUT 19
5 Installation 20
5.1 NSG 3040 installation 21
5.2 Groundingthetestsystemtothegroundreferenceplane 21
5.2.1 Rear ground brackets 22
5.3 Mountingina19”rack 22
6 Mainframe description 23
6.1 Frontpanel 23
6.1.1 EUToutputconnection 23
6.1.2 Highfrequencygroundterminal 24
6.1.3 Surgeoutputsockets 25
6.1.4 Burstoutputsocket 25
CoNTENTS

6.1.5 IndicatorLEDs 25
6.1.6 Touchpanelanduserinterface 26
6.2 Rearpanel 26
6.2.1 Mainspowerinput 26
6.2.2 EUTpowerinput 27
6.2.3 ACEUTmainsinput 28
6.2.4 DCEUTinput 29
6.2.5 Groundconnectionpoint 29
6.2.6 Systeminterfaceconnector25pinDsub 30
7 Touch screen interface 32
7.1 Mainmenu 33
7.2 Systemwindow 34
7.3 Systemwindow–General 34
7.4 Systemwindow–Equipment 37
7.5 Systemwindow–Communications 38
7.6 Systemwindow–Monitoring 40
7.6.1 EUTsupplyvoltage&EUTsupplyfrequency 40
7.6.2 TestactionbyEUTfailinput 40
7.7 Systemwindow-SD-cardproperties 41
7.7.1 UsingtheSD-cardtoupdateTSIsoftware 41
7.7.2 RemovingandreplacingtheSD-card 42
7.8 Systemwindow-Language 45
8 Setting test parameters 46
8.1 Theredmenubar 46
8.2 Thebottombar 49
8.3 ElectricalFastTranisient(EFT)/bursttesting 53
8.3.1 Testcongurationforpowerlinecoupling 53
8.3.2 Testcongurationswithexternalcoupling 53
8.3.3 SettingEFT/bursttestparameters 54
8.3.4 Voltage parameter 55
8.3.5 Frequency 55
8.3.6 Phase 55
8.3.7 Coupling 56
8.3.8 Bursttime 57
8.3.9 Repetitiontime 57
8.3.10 Testduration 57
8.3.11 Burstgeneratortechnicaldata 58

5
8.3.12 DTAselectionwindowforburst 58
8.4 Combinationwave(CW)/surgepulsetesting 58
8.4.1 Testcongurationforpowersupplylinecoupling 58
8.4.2 Testcongurationswithexternalcoupling 59
8.4.3 Settingsurgetestparameters 60
8.4.4 Voltage 60
8.4.5 Impedance 61
8.4.6 Phase 61
8.4.7 Coupling 61
8.4.8 Repetitiontime 62
8.4.9 Testduration 62
8.4.10 Combinationwavegeneratortechnicaldata 63
8.5 MainsPowerQualityTeting(PQT) 63
8.5.1 Samplegraphsofvoltagedips/drops 64
8.5.2 SettingmainsPQTtestparameters 65
8.5.3 VoltageUVar 65
8.5.4 Phase 65
8.5.5 Repetitiontime 66
8.5.6 T-Event 66
8.5.7 Testduration 66
8.5.8 Dips&dropsvalue 67
8.6 Variationtest 67
8.7 Pulsedmagneticeldoption 70
8.8 PowermagneticeldinconjunctionwithMFO 70
9 System interface functions 71
9.1 Interlock 72
9.2 Triggertoscopeoutputsignal 73
9.2.1 Triggersignalwithasurgepulse 73
9.3 Synchronizationoutputsignal(sync) 73
9.4 Pulseenable/nextstepinput 74
9.5 EUTfailinput 74
10 Internal coupling network 75
11 Maintenance and function check 77
11.1 General 77
11.2 Cleaning 77
11.3 Functioncheck 78
11.4 Calibration 79

6
NSG 3040 EMC Test System
11.5 Warranty 79
12 CE conformity 80
13 Technical data 81
13.1 Dimensions/weight 82
13.2 Options 82
13.3 AccessoriesforIEC/EN61000-4-11 83
13.4 AccessoriesforIEC/EN61000-4-8/-4-9 83
14 System description 84
15 Addresses 86
WARNING - Lethal danger from high voltages and the
risk of radiating illegal electromagnetic interference.
The NSG 3040 may only be installed and used by autho-
rized and trained EMC specialists.
The NSG 3040 must only be used for EMC tests as
specied in these operating instructions.
Personnel tted with a heart pacemaker must not
operate the instrument and must not be in the vicinity
of the test setup while it is in operation.
When the system is used in conjunction with options,
accessories or other equipment the safety instructions
concerning those devices must also be observed.

7
Thearrowsymbolshownbelowindicatesanote.Notesareusedtoprovide
additionalinformationthatwillenabletheusertoachieveoptimaltestperfor-
manceandefciency.
Example:
1 ExpLANATioN of SyMboLS
NOTE - This connection must not be confused with the
EUT power input.
Theexclamationpointsymbolshownbelowindicatesawarning.Awarning
meansthatinjuryordeath,aswellasdamagetotheinstrumentsandEUT,are
possibleiftheinstructionsarenotobeyed.
Example:
WARNING - Never connect or disconnect the EUT while
the system is performing a test.

8
NSG 3040 EMC Test System
2.1 General description
TheNSG3040testsystemisamultifunctiongeneratorthatsimulatescable-
borneelectromagneticinterferenceeffectsforimmunitytestingtointernational,
national,andmanufacturers’standards.
Thesystemisdesignedtofulllconductedelectromagneticcompatibility(EMC)
testrequirementsforcompliancetestingofhousehold,ofce,lightindustrial
orcommercialequipment,includingcombinationwavesurge,ElectricalFast
Transient(EFT)pulses,andPowerQualityTesting(PQT).
TheNSG3040’smodulararchitectureandindustrystandardinterfacesallowit
tobeeasilyexpandedandcustomizedtomeetindividualtestingneeds.
Thesystemisdesignedasaseriesofinteroperablefunctionunitswithamaster
controllerthathandlestherealtimefunctionsandcommunicateswiththe
functionmodules.Eachfunctionunitcontainsaslavecontroller;allfunction
unitsareconnectedtogetherthroughtheirslavecontrollersandnetworked
withthecentralmastercontrollerviaaeldbus(Interbus).Informationconcern-
ingspecialfeaturesandtheiradjustableparametersarestoreddirectlyinthe
functionmodules.
2 iNTRodUCTioN

9
Thismodularityenablesthefunctionunitstobecombinedintocustomized
testsystems,andlaterreconguredtoaddresschangingtestingrequirements.
Thefunctionunitscanbereadilymodiedtoaddresstherequirementsofnew
standards,andnewfunctionunitsfornewparametersmaybeincorporated
inexistingsystems.
TheNSG3040iscontrolledthroughitsstandarduserinterfaceviaatouch
paneldisplay.ThesystemcanalsobecontrolledbyaremotePCviaitsEthernet
interface.
Toensureoptimaluserandequipmentsafety,onlyindustry-standardandcor-
rectlyspeciedplugsandsocketsareusedthroughout.Highvoltageoutputs
areswitch-protected.

10
NSG 3040 EMC Test System
TheNSG3040testsystemisdesignedprimarilyforcable-bornetransientinter-
ferencetestsasspeciedintheEuropeangenericstandardsIEC/EN61000-6-1
coveringequipmentforhousehold,ofceandlightindustrialuse,andIEC/EN
61000-6-2forapplicationsinindustrialenvironments.TheNSG3040generates
thesetestsinaccordancewithIEC/EN61000-4-2,-4,-5,and-11.Accessoriesare
alsoavailableforgeneratingoptionalteststoIEC/EN61000-4-8and-9.
TheEUdirectiveno.89/336/EEC(fortheassignmentoftheCEsymbol)tothese
standardsandtothistypeofequipment.
ESDtests(inaccordancewithIEC/EN61000-4-2)mustbeperformedwitha
separatesimulator,suchas theTeseqNSG435,NSG437 orNSG438. The
standardcallsforbothairandcontactdischarges,andthesimulatorissupplied
withspecialtipsforeachtypeoftest.Inthecaseofairdischargesthesimulator
isdischargedbyholdingthetipclosetotheEUT.Then,whiledepressingthe
trigger,movingitclosertothetargetareauntiladischargeoccurs.Contact
dischargesoccurwiththetipofthesimulatorindirectcontactwithEUT.
Bursttests(incompliancewithIEC/EN61000-4-4)simulatethehighvoltage/high
frequencyinterferencepulsestypicallyproducedwhenaninductivelyloaded
switchisoperated.Withoutcounter-measures,suchinterferencemayoccur
whenacurrentthroughanelectromagneticdevice,e.g.motor,circuitbreaker,
relay,uorescentlamp,etc.isswitchedoff.
Thistypofinterferencecanaffectotherequipmentineitheroftwoways.Firstly,
theinterferencecanbecoupleddirectlyintothetargetequipmentviathemains
powercable.Theinterferencecanbetransmittedfromthesourcealongthe
mainspowercableconnectedtothetarget.Interferencefromthemainscan
3 STANdARdS ANd AppLiCATioNS
3.1 Burst left

11
reachanyotherpieceofequipmentconnectedtothesamepowersourcein
asimilarway,wherebythisdoesnotallhavetooccurinthesamesectionof
a building.
Alternatively,theinterferencecanbecapacitivecoupledintoanytargetdevice
inthevicinity.
Thetestsystemenablesatesttobeerformedusingbothcouplingmethods.
TheEUTisconnectedtothemainspowersocketonthefrontpanelofthetest
systemforthedirectmainsinjectiontest.Capacitivecoupledtestsrequirethe
interferencetobesuperimposedontheEUTcableviaanexternalcouplingunit
thatisconnectedtotheBurstoutputonthefrontpanelofthesystem.
3.2 Combination wave test
Thesurgetest,incompliancewithIEC/EN61000-4-5,duplicateshighenergy
interferenceasexperiencedwithalightningstrike.Generallyspeaking the
interferencendsitswayintohouseholdequipmentviathelowvoltagemains
power supply.
Thiskindofinterferencecanaffectequipmentineitheroftwoways.Firstly,the
interferencecanbecoupleddirectlyintotheequipmentviathemainssupply.
Theinterferenceisconveyeddirectlyfromthesource(e.g.lightningstriketo
external power cables). Everyitemof equipmentconnected to thispower
sourcewillbeaffectedbytheinterferencepulses.
Alternatively,thepulsesfromthesourceoftheinterferenceoritsassociated
mainscablescanbecoupledintootherequipmentpositionednearby.
Surgepulseinterferencecanalsooccuronsignalanddatalinesthroughcou-
plingeffectsandelectricaldischarges.
Thetestsystemenablesteststobecarriedoutusingbothcouplingmethods.
TheEUTisconnectedtothemainspowersocketonthefrontpanelofthetest
systemforthedirectmainsinjectiontest.Externallycoupledtestsrequirethe
interferencetobesuperimposedontotheEUTpowerfeedcableorsignal/data
linesviaanexternalcouplingunitthatisconnectedtothesurgeoutputonthe
frontpanelofthesystem.

12
NSG 3040 EMC Test System
3.3 Mains quality test
Themainsqualitytestincludesthesimulationofdipsanddrop-outsinthemains
powersupplyinaccordancewithIEC/EN61000-4-11andforDCpowersupplies
inaccordancewithIEC/EN61000-4-29.
Avoltagedipissaidtooccurwhenthesupplyvoltagefallsconsiderablybelow
thenominallevelforarelativelyshorttime,e.g.forafewcycles,whereasa
drop-outmeansthatthevoltagefallstozeroforasimilarperiod.
3.4 Magnetic fields with mains frequency (option)
Mainsfrequencymagneticeldtests,or,POWERMtests,involvethesimulation
ofthemagneticeldstypicallygeneratedbythecurrentowinpowersupply
cablesasspeciedintheIEC/EN61000-4-8standard.Suchmagneticelds
mayaffecttheoperationofitemsofequipmentthataresensitivetothem.The
NSG3040performsthistestbycausingacurrenttoowinamagneticeld
coilsuchthatthecurrentandfrequencyproduceaproportionaleldwithin
thecoilparameters.
Themagneticeldcoils,availableasaccessories,areconnectedtothemag-
neticeldoption(MFO)which,inturn,isconnectedtothesystem.
3.5 Pulsed magnetic fields (option)
Testswithpulsedmagneticelds,orPULSEMtests,simulatethekindofinter-
ferenceproducedbysurgepulsesasaresultoflightningstrikestobuildings
andothermetallicstructuressuchasfree-standingmasts,groundconductors,
groundingnetworks,etc.asspeciedinIEC/EN61000-4-9.Magneticeldsof
thistypemayupsettheoperationofinstallationsthatndthemselveswithin
suchelds.NSG3040performsthistestbycausingaheavycurrenttoowin
amagneticeldcoilsuchthattheamplitudeofthepulsecurrentproducea
proportionaleldwithinthecoilparameters.
Themagneticeldcoils,availableasaccessories,areconnectedtothesurge
pulseoutputsocketviatheINA752pulseshapingnetwork.

13
TheNSG3040systemanditswithaccessoriesoperateathighvoltages.
4 SAfETy iNSTRUCTioNS
WARNING - Improper or careless operation can be
fatal.
Theseoperatinginstructionsformanessentialpartoftheequipmentandmust
beavailabletotheoperatoratalltimes.Theusermustobeyallsafetyinstruc-
tions and warnings.
NeitherTeseqAG,Luterbach,Switzerland,noranyofitssubsidiarysalesorga-
nizationscanacceptanyliabilityforpersonal,materialorconsequentialinjury,
lossordamagethatmayresultfromimproperuseofequipmentandacces-
sories.
4.1 General
TheNSG3040mustbeoperatedonlybyauthorizedandtrainedspecialists.
Thegeneratoristobeusedonlyforthepurposespeciedbythemanufacturer.
Theuserisdirectlyresponsibleforensuringthatthetestsetupdoesnotcause
excessiveradiatedinterferencewhichcouldaffectotherinstrumentation.The
testsystemitselfdoesnotproduceanyexcessiveEMradiation.However,the
injectionofinterferencepulsesintoaEUTcanresultinitand/oritsassociated
cables radiating electromagnetic energy. Toavoid unwantedradiation, the
standardsorganizationsrecommendthatthetestsetupbeoperatedinsidea
Faradaycage.

14
NSG 3040 EMC Test System
WARNING - Because of its construction, the NSG 3040
is not suitable for use in an explosive atmosphere.
WARNING - Persons tted with a heart pacemaker must
neither operate the instrument nor approach the test
setup while a test is being executed.
Onlyapprovedaccessores,connectors,adapters,etc.aretobeusedtoensure
safeoperation.
WARNING - The NSG 3040 will start a self test when it is
switched on. This includes the function of the EUT cou-
pling relays and the detection of the EUT-input power
for synchronisation reason. Therefore the EUT-output is
to be considered as carying EUT power all the time!
4.2 Installation
TheNSG3040testsystemconformstoprotectionclass1.Localinstallation
regulationsmustberespectedtoensurethesafeowofleakagecurrents.
WARNING - Operation without a ground connection is
forbidden!
Twoindependentgroundconnectionsarenecessary-oneforthetestsystem
andonefortheEUT.Thesemustbeconnectedbacktothelocalpermanent
installationortoaxed,permanentgroundconductor.
Operatetheequipmentonlyindrysurroundings.Anycondensationthatoccurs
mustbeallowedtoevaporatebeforeputtingtheequipmentintooperation.Do

15
notexceedthepermissibleambienttemperatureorhumiditylevels.Useonly
ofciallyapprovedconnectorsandaccessoryitems.
Ensure that a reliable return path for the interference current is provided
betweentheEUTandthegenerator.Thereferencegroundplaneandtheground
connectionstotheinstruments,asdescribedintherelevantteststandards,
servethispurposewell.
Thetestsystemmayonlybeopeneduponspecicinstructiongivenbythe
manufacturer.Sincetheinstrumentworks,onprinciple,withtwoindependent
powersupplies(oneforthegeneratorandonefortheEUT),theNSG3040must
bedisconnectedfrombothsourcesbeforeanymodicationstothetestsetup
areundertaken.Besidesthemainsconnectionsthemselves,certaincompo-
nentsalsooperateathighvoltages,andarenotprovidedwithanyformofextra
protection against accidental contact.
4.3 Installation of an EUT power switch
TheEUTinputshouldbeconnectedaproperlyratedpowerswitchdevice,which
shouldbelocatedclosetothetestsetup.Inordertoensureaneasyandquick
accesstotheEUTpower,theswitchshouldbeclearlyandvisiblylabelledasa
devicefor“EUTpoweron/off”.
Thein-housepowerdistributionmustbeequippedwithapropercircuitbreaker
andanemergencyoffbuttonasperIEC61010-1:2001.
The test setup should only be accessible to trained per-
sonnel.
DimensioningofmainssupplyandratingoffuseprotectionofACorDCpower
supplymustconformwithnationalprescriptionsandEUTrequirements.
Inapropriatearrangement,mounting,cablingorhandlingoftheEUTorthepro-
tectiveelementscanhamperornegatetheeffectivenessoftheNSG3040’s
safetyfeatures.

16
NSG 3040 EMC Test System
4.4 Applicable safety standards
TheNSG3040conformstothesafetyrequirementsspeciedinIEC348and
offersallthefeaturesnecessaryforsaveandefcientoperation.
TheNSG3040wasdevelopedandismanufacturedincompliancewithISO
9001.
ThesystemcomplieswiththesafetyrequirementsofIEC/EN61010-1(Safety
requirementsforelectricalequipmentformeasurement,controlandlaboratory
use).
Likeallmainspower-drivengenerators,thesystemisdesignedforhighvoltage
workingsafetyinaccordancewithVDE0104.
InterferenceimmunityhasbeentestedinaccordancewithEN60326-1.
Itistheuser’sresponsibilitytoensurethatthetestsetupdoesnotemitexces-
siveelectromagneticinterference(EMI)whichmightaffectotherequipment.
Thetestsystemitselfdoesnotproduceanyexcessiveradiation;however,the
injectionofinterferencepulsesintotheEUTcanresultinthedeviceand/or
itsassociatedcablesradiatingEMI.Toavoidradiatingunwantedinterference
thestandardsorganizationsrecommendthatthetestsetupbelocatedina
Faradaycage.
Sincethepurposeofthetestsystemistoproduceinterferencesignalsfor
interferenceimmunitytesting,therequirementsinIEC/EN61000-6-1,-2,-3,-4
concerninglimitingtheradiatedEMIcanonlybecompliedwithbyoperating
thetestsysteminsideaFaradaycage.

17
WARNING - The test area must be organized that no
unauthorized persons have access during the execu-
tion of a test. If a safety contact (Interlock) is used as a
means of access control to the test zone (e.g. a Faraday
cage), then an additional contact connected in series
is necessary to provide protection for parts of the EUT
that are likely to be touched EUT, its accidentally.
During a test, the EUT, its accessories and cables are
to be considered live at all times. The test system must
be stopped and the EUT supply disconnected before
any work can be carried out on the EUT. This can be
achieved simply by opening the Interlock circuit.
The EUT is to be tested only in a protective cage or
under a hood which provides protection against electric
shock and all manner of other dangers pertaining to the
particular EUT (see: User warnings - Generator).
The safety instructions for all the instruments and asso -
ciated equipment involved in the test setup.
Test setup conguration is to be strictly in compliance
with the methods described in the relevant standard
to ensure that the test is executed in a compliant
manner.
4.5 Test execution

18
NSG 3040 EMC Test System
WARNING - Users must be aware of the following
dangers that can occur during testing:
Local burning, arcing, ignition of explosive gases.
EUT supply current surge caused by a ashover
or breakdown resulting from the superimposed high
voltage.
Disturbance of other, unrelated electronics, telecom-
munications, navigational systems and heart pace-
makers through unnoticed high frequency radia-
tion.
Ground contacts, including the EUT ground, can be at
elevated voltage levels that would make them
dangerous to touch.
4.6 User warnings - Generator

19
WARNING - Users must be aware of the following
dangers that can occur during testing:
EUTs are often functional samples that have not yet
been subjected to safety tests. It is therefore pos-
sible that the EUT could be damaged by internal
overloads or may even start to burn.
As soon as the EUT shows signs of being disrupted
the test should be stopped and the power to the EUT
switched off.
Internal disruption of the electronics can result in
the interference voltage or the EUT supply voltage
being present on the EUT’s outer casing.
Electrical breakdown or arcing from and within
plugged connections that are overstressed by
voltage during the test.
Explosion of components with re or fragmentation
as a result of energy dissipated, e.g. from the resul-
tant supply current or ignition of vaporized plastics
materials.
Faulty behavior by the EUT, e.g. a robot arm strikes
out or a temperature controller fails, etc.
4.7 User warnings - EUT

20
NSG 3040 EMC Test System
Thischaptercontainsashortcheck-listwithstepsthatshouldbetakenbefore
theinstrumentisswitchedonandputintooperation.
Upondelivery,checkthepackagingforsignsofdamageintransit.Iftheindicator
panelontheshock-watchiscoloreditmeansthatthepackagehassuffered
ashockduringtransit.Anydamageshouldbereportedimmediatelytothe
transportation company.
LifttheNSG3040testsystemoutofitspackagingbygraspingthemounted
grips.
Usingthefollowinglist,checkthatalltheitemsorderedhavebeendelivered:
1. NSG 3040 generator
2. Usermanual
3. 1Mainspowercableforthetestsystem
4. 1Dummyplug(interlockblindconnector)
5. 1Groundcable(togroundreferenceplane)
6. 1EUTpowerinputconnectorwithcable
7. 1EUTpoweroutputconnector
8. Optionalitems,asordered
Checktheinstrumentforsignsoftransportdamage.Anydamageshouldbe
reportedtothetransportationcompanyimmediately.
5 iNSTALLATioN
NOTE: Do not dispose of packaging materials. All packag-
ing should be retained in the event that the instrument
or any of its accessories should need to be returned to
a Teseq service center for repair or calibration.
Table of contents