- 4 -
Sponsoring members of NACE International, NACE Foundation QMF-570 104-154
V. Operating Instructions
A. Pearson Survey(Also see condensed Operating Instructions)
The following components are required for a Pearson Survey of a buried pipeline:
1. Oscillator with connecting cables
2. Twelve-volt storage battery (not furnished)
3. Receiver
4. Two sets of shoe cleats
5. Cleat connecting cables
6. Terminal Boards
The audio Oscillator is placed along the pipeline at a point where an electrical connection can be made to
the pipe. Test point lead wires are convenient connections. The pipe lead wire is connected to the terminal of the
Oscillator marked PIPE. Connect a wire from the terminal marked GROUND to some metal structure, which is in
good electrical contact with the earth. If no structure is available, a rod can be driven into the earth at a point from
20 to 30 feet from the pipe at right angles to the pipeline.
Connect a 12-volt storage battery to the battery terminals, observing correct polarity (red is positive or +,
black is negative or -). Set the output voltage switch to 2.5, the interrupter switch to “OFF”, and the battery switch
to “ON”. The Fine Adjust knob should be advanced all the way to the Right (MAX). Next, advance the output
voltage control until the LED lights on the Oscillator show Red. Then move the Fine Adjust knob to the Left (MIN)
until the LED lights show Green.
NOTE: Model PD-C (Rev. B) uses a new crystal that produces an interrupt frequency at a faster rate than
previous models. Users familiar with this instrument may notice the difference.
If overloaded, circuit breaker will trip. To reactivate, reduce driving voltage and reset circuit breaker.
Turn the interrupter switch “ON”, if desired, and the Oscillator is ready for use.
For the most recent Oscillator, PD-C (Rev. B), the instructions are:
1. Connect 12 volt battery, observing polarity.
2. Connect to pipe and ground. Output Voltage to 2.5, Fine Adjust all the way to the Right (MAX).
3. Interrupter to OFF. Battery to ON.
4. Advance Output Voltage until Red LEDs light. Move Fine Adjust towards MIN until Green LED lights.
5. Interrupter to ON. Instrument is now ready for use.
NOTE: Model PD-C (Rev. B) uses a new crystal that produces an interrupt frequency at a faster rate than
previous models. Users familiar with this instrument may notice the difference.
The terminal boards should be attached to the belts of the operators. (The board with the cord and plug is
used by the operator who carries the Receiver). Both operators fasten cleats on their shoes, then attach cleat cable
to each cleat, running the cables inside their pant legs. The key operator connects with wires from both of his cleats
to the lower terminal on his terminal board. The secondary operator connects the wires from both of his cleats to
the lower terminal on his terminal board. The secondary operator connects the wires from both of his cleats and
one end of the long connecting cable to his terminal board. The other end of the connecting cable goes to the
upper terminal of the key man’s terminal board. The plug coming from the key man’s terminal board is plugged into
the jack marked CLEATS, the Receiver is turned “ON”, and the apparatus is ready for use.