PRECAUTIONS
This instrument is an electronic precision unit, so be sure to use and store it in a place controlled under
normal living temperature, humidity and atmospheric pressure, and avoid direct exposure to sunlight.
To ensure right operation, install the instrument on the wall and check for a precise horizontal and vertical
alignment. Also do not put anything on top of the instrument.
To ensure an optimal coverage of polarized tests, align the center of LCD with the line of sight of the patient.
Wear the polarized or Red/Green lenses and select in CC-100 Series a dissociated test, check the horizontal
and vertical alignment until the best coverage for each eye is reached.
Connect all cables properly before using.
Use the power at a rated voltage.
When not in use, switch off the power source and protect the unit from sunlight and dust.
For accurate functioning, keep the instrument clean and free of spots and dust.
Technical Specifications :
93-42 EC Directive
Classification
Electrical Specifications:
AC 110-120 V or 220-240V -- 50-60 Hz
220-240 V: 20x5mm 2.5A Antisurge
Electromagnetic Compatibility
This product conforms to the EMC standard (IEC 60601-1-2:2007).
MEDICAL ELECTRICAL EQUIPMENT needs special precautions regarding EMC and need to be
installed and put into service according to the EMC information provided in the accompanying
documents.
Portable and mobile RF communications equipment can effect medical electrical equipment
The use of accessories and cables other than those supplied with the instrument, exception for
cables sold by the manufacturer of the equipment as replacement parts, may result in increase
emission or decrease immunity of the equipment or system.
The equipment or system should not be used adjacent to or stacked with other equipment. If
adjacent or stacked use is necessary, the equipment or system should be observed to verify normal
operation in the configuration in which it will be used.