iDangers, Warnings, Cautions, and Battery
SoundPro®Sound Level Meter Outdoor Measuring System
SOUNDPRO® SOUND LEVEL METER
OUTDOOR MEASURING SYSTEM
(SP-OMS)
Dangers, Warnings, Cautions & Battery
DANGER!
Failure to observe the following procedures may result in serious personal injury:
Risk of Electrical Shock. Avoid nearby overhead wires and electrical fixtures. The conductive
electrical mast will tend to attract stray voltages and currents, including the possibility of Lightning.
Risk of Fire or Explosion. DO NOT connect to AC power while the lid is sealed or closed. DO NOT
connect to AC power while the lid is sealed or closed. DO NOT modify the enclosure to provide AC
power from external sources while the lid is closed. Hydrogen gas may be released from the lead-
acid batteries during charging leading to a risk of explosion. The cover must be left open during the
re-charge process.
Risk of Chemical Burn. Although the batteries provided are Absorbent Glass Mat (AGM) Sealed
Lead-Acid (SLA) batteries that can be stored, trans- ported, operated, and re-charged in any
orientation, there is still a slight risk of exposure to sulfuric acid if the case is cracked or punctured.
These batteries can be shipped as “dry cargo” and do not require special hazardous material
shipping.
Not for use in explosive or hazardous locations. This product is not intrinsicallysafe.
WARNING!
Failure to observe the following procedures could damage the instrument:
Read the manual before operation.
DO NOT operate in temperatures exceeding 50°C (122°F). Excessively hot and extreme sun
exposure may lead to over-heating. Use additional sun shields to reduce running temperature, if
necessary. Repeated or extended time in extreme environments will lead to premature battery failure.
Substitution of components may impair the accuracy of the instrument. Repair should only be
performed by authorized service personnel.
Caution!
General
Before first use, insure that the voltage select switch on the charger is set to the line voltage in your
location. It may be set to either nominal 115V or 230VAC. To preserve battery cycle life, store the
batteries fully charged. Turn OFF the power at the power distribution panel when not in use. Charge
batteries as soon as possible after any use and before storing them for long periods.
Intended Use:
The SoundPro Outdoor measurement system provides protection to re-chargeable batteries for extended outdoor
operation of the SoundPro Sound Level Meter. Consult your company’s safety professional for local standards, or call
TSI at 1-800-680-1220 (USA) or (651) 490-2860.

ii Table of Contents
SoundPro®Sound Level Meter Outdoor Measuring System
TABLE OF CONTENTS
Dangers, Warnings, Cautions & Battery..........................i
Intended Use: .....................................................................i
Introduction.......................................................................1
SoundPro Outdoor Measuring System (SP-OMS) .........1
Assemble the Outdoor Kit: Field Instructions ...............2
Mounting the outdoor microphone and mast to the
weatherproof case. .....................................................2
Assemble the Outdoor Kit: Internal Connections..........3
Making the internal connections .....................................3
Make the measurements ................................................4
Turn-off and disassemble ...............................................4
Battery: First Assembly and Replacement .................... 4
Maintenance and Care Instructions................................ 5
Recharging the internal batteries ................................... 5
Reactivating the desiccant ............................................. 6
Storage Recommendations............................................ 6
SoundPro Outdoor Kit Specifications............................ 7
Customer Service............................................................. 8
Technical Support Contacts ........................................... 8
Service Contact Information........................................... 8
Returning for Service...................................................... 8
Warranty............................................................................ 9
FIGURES
Figure 1: Assembly for outdoor kit .........................................2
Figure 2: Connecting the battery power ................................3
Figure 3: On/Off switch for the SoundPro Outdoor Kit...........3
Figure 4: Connecting two SoundPro outdoor
battery assemblies.................................................. 5
Figure 5: Sample SoundPro outdoor kit assembly parts........ 7
TABLES
Table 1: SoundPro Outdoor kit parts .....................................3

1Introduction
SoundPro Outdoor Measuring System (SP-OMS)
SoundPro®Sound Level Meter Outdoor Measuring System
Introduction
The SoundPro®Outdoor Measuring System helps protect the SoundPro
Sound Level Meter from environmental exposure, such as wind, rain, snow,
chemicals, particulates, animals, vandalism and theft. It is also used for
extended battery life with up to one week of continuous monitoring (two
weeks with optional second battery).
The weatherproof case holds the SoundPro instrument and battery pack with
room for accessories and storage of the system components while not in
use. The exposed components are made of stainless steel or plastic to resist
corrosion.
SoundPro Outdoor Measuring System (SP-OMS)
How it works? The weatherproof case provides a stable base for mounting
the microphone mast. There is an internal battery monitor circuit to reduce
the risk of damaging the batteries by over-discharge. The circuit also
protects against over-voltage, over-temperature (70°C or 158°F), and over-
current (2 Amps, each output). The on-board universal input battery charger
fully re-charges empty batteries in about 16 hours or less.
Provisions in the case design allow customer-supplied padlocks and cables
to be used to lock the case and secure it to a stationary object.

2Assemble the Outdoor Kit: Field Instructions
Mounting the outdoor microphone and mast to the weatherproof case
SoundPro®Sound Level Meter Outdoor Measuring System
Assemble the Outdoor Kit: Field Instructions
To assemble the outdoor kit refer to Figure 1 and Table 1.
Mounting the outdoor microphone
and mast to the weatherproof case
a. Assemble the mast pieces by screwing the threaded
coupler of the shaft segment with coupler of item 3
(see Figure 1/Table 1), to the threaded base shaft
(see 2). You may use multiple extensions (see 3) to
increase the height of the microphone.
b. From inside the case (see 1), feed the smaller female
end of the cable assembly (9) through the hole in the
case cover.
c. For easier insertion, you may lubricate the connector
boot with a bit of petroleum jelly. Feed the female
end of the cable through the mast, and then
completely through the preamp adapter (see 4). The
insertion tool (see 11) will be necessary to push the
connector through the preamp adapter (see 4).
d. Assemble the preamp adapter (see 4) to the mast.
Then assemble the mast to the case by screwing the
threaded portions together.
Caution: | DO NOT over tighten.
e. Screw the microphone (see 6) onto the preamp (see
5). Connect the preamp to the female end of the
cable by properly orienting and inserting the
connector into the preamp. Hold the cable connector
while turning the knurled knob on the preamp to
tighten the connectors.
f. Grasp the preamp and push the connector’s rubber
boot into the preamp adapter until the preamp is
completely seated.
g. Insert the windscreen (see 7) inside of the bird
spikes (see 8). Place this assembly over the
microphone preamp. Gently spread the bird spike
wires and insert them into the holes on the preamp
adapter.
NOTE: Be cautious when bending the wires to
prevent them from deforming.
Figure 1: Assembly for outdoor kit

3Assemble the Outdoor Kit: Internal Connections
Making the internal connections
SoundPro®Sound Level Meter Outdoor Measuring System
Table 1: SoundPro Outdoor kit parts
Description
Part Number
Description
Part Number
1. Weatherproof case assembly.
057-832
10. Case plug w/ 059-330 O-ring (both
parts required)
056-467
2. Base Shaft with sealed nut.
(1.0 m; 39”)
056-472
11. Cable insertion tool.
056-475
3. Shaft segment with coupler.
(1.4 m, 56”). Two provided for optional
extra height. (1.9 m; 73”)
056-473
12. SoundPro outdoor 12V power
cable assembly.
057-241
4. Preamp adapter 2 in./5 cm tall.
056-462
13. SoundPro outdoor battery
assembly (1 or 2) provided.
057-242
5. SoundPro Preamp assembly (included
with SoundPro Sound Level Meter).
053-809
14. SoundPro outdoor power
distribution panel.
057-826
6. Microphone (various models included
with SoundPro Sound Level Meter and
identified in Appendix A of user manual)
(See Part # on
delivery note.)
15. Universal Input (Switch-able)
12V/5A Battery Charger.
053-562
7. Windscreen/weather shield.
056-463
16. USA 115V IEC320 Line cord
#18AWG.
068-028
8. Bird Spikes.
056-464
17. European 230V IEC320 Line cord
#18AWG (Optional).
054-462
9. SoundPro outdoor extension cable
assembly.
057-240
18. Desiccant SG-40 Rechargeable
canister.
057-235
Assemble the Outdoor Kit: Internal Connections
Making the internal connections
1Connect the other end of the cable (see 9in Figure 1/Table 1 ) to the meter inside the case.
2For brief runtimes (4 hours or less), you may use the internal battery power of the SoundPro Sound Level Meter. For
extended runtime, connect the internal 12-Volt power cable (see Figure 2) to the meter at its bottom inlet. Then turn
On the power switch at the power distribution panel (see Figure 3).
Power cord connected
On/Off switch Power distribution panel (see
16 in Table 1 for part details)
Figure 2: Connecting the battery power
Figure 3: On/Off switch for the SoundPro Outdoor Kit
3When you have turned On the power switch and the SoundPro Sound Level Meter, verify that the power cord is
connected for external power. (NOTE: The battery icon , will appear on the display of the SoundPro Sound
Level Meter as a fully charged icon when connected with AC power cord.)

4Battery: First Assembly and Replacement
Make the measurements
SoundPro®Sound Level Meter Outdoor Measuring System
Make the measurements
1Follow the SoundPro Sound Level Meter user manual to prepare and run your measurements. (Note: This may
require a pre-calibration, setup changes, and pressing the run key or programming an auto-run.)
2When you are ready, place the SoundPro Sound Level Meter in the pocket of the case, verify that the SoundPro
Sound Level Meter is running on external power, and close up the case.
Turn-off and disassemble
1When you are finished, turn Off the SoundPro Sound Level Meter and then turn Off the power switch at the power
distribution panel.
2Remove and disassemble the mast, microphone, and windscreen and place the parts back into the case.
3Remove and store the case plug with O-ring (see 10 in Figure 1/Table 1).
Battery: First Assembly and Replacement
1. Open and unpack the batteries.
•NOTE: The battery is packaged separately from the outdoor kit to
ensure the shipping weight of each carton is under the 70-pound
limit that is imposed by common freight carriers such as Federal
Express (FEDEX) and United Parcel Service (UPS).
2. (Please skip step 2 for first assembly.)
a. If replacing the batteries, disconnect and remove the old batteries.
b. Disconnect the battery’s connectors from the power distribution
panel (see Figure 3).
c. Remove the battery cover as discussed in Step 3.
d. While holding the case open, stand the case on its right side (away from the wheel end).
e. Gently slide the battery out of the compartment (one at a time if more than one battery).
3. Open the enclosure and remove the battery cover.
•Unlatch and open the outdoor system enclosure.
NOTE: The cover is held in place with six Phillips-Head screws. The power distribution panel and the charger
are also mounted on the battery cover. It is optional to remove the shaft segments, the accessory
parts, and some of the foam layers.
•Carefully remove the six larger screws around the perimeter and remove the cover.
4. Install batteries into enclosure.
•While holding the outer enclosure lid open, tip the entire case on its side, so that the wheel-side is up, the
battery enclosure is near the top, and the right handle is on the floor. This will make it much easier to slide the
batteries down into their compartments. Position each battery so that the wires and posts can still be seen as
you slide it into position. Lay the enclosure back down in its normal orientation.
•As displayed in Figure 4 below, it is optional to connect two SoundPro outdoor battery assemblies (see 13 in
Table 1) inside the case for additional battery power.

5Maintenance and Care Instructions
Recharging the internal batteries
SoundPro®Sound Level Meter Outdoor Measuring System
Battery cables routed
Installed SoundPro outdoor
battery assembly (see 13 in
Table 1 for part details)
Figure 4: Connecting two SoundPro outdoor battery assemblies
5. Route the battery cables.
•While positioning the battery cover over the batteries in the correct orientation, take one of the battery wires and
its green connector, and feed it through one of the rubber grommets. Take the other connector and feed it
through the other grommet. Carefully route the wires under the cover so that they will not be pinched (see
Figure 4)
6. Replace cover and screws.
•Position and secure the cover in place with the six (6) screws you removed in Step (3).
7. Connect the Batteries.
•Connect the two battery cables into the Power Distribution Panel (see Figure 4) at the two connectors near the
RED fuses. These two connectors are the only connectors that may be used for the batteries.
Maintenance and Care Instructions
Recharging the internal batteries
1. Open the case and leave it open for the duration of the charging. See the WARNING: “Risk of FIRE or EXPLOSION”
at beginning of this manual. Keeping the case open also allows the charger to stay cool.
2. You may leave the master On/Off switch in either position. If you leave the master On, you may also leave the
meter connected, but doing that will increase the charging time by about 10%.
3. Before the first use of the charger, verify that the input voltage selector on the charger module is set to the
appropriate line voltage. The selector is a small red slide switch near the power inlet that is moved into the correct
position using a small screwdriver. In the USA, it is 115 Volts (the UP position). With the European cord you will use
230 Volts (the Down position). When it is in its proper location, you will see your line voltage displayed.
4. Connect your line cord to the charger power inlet.
5. Connect the cord to a power outlet.
6. There is a status lamp on the charger that is located on the opposite end from the power inlet. The lamp has three
states.
•Yellow: Indicates Charging.
•Green: Indicates Finished Charging.
•Blinking: Indicates Charging Problem.

6Maintenance and Care Instructions
Reactivating the desiccant
SoundPro®Sound Level Meter Outdoor Measuring System
7. Leave the power cord connected until you see the GREEN (FINISHED) indicator. The charging rate is 5 Amps. In
the unlikely event that the batteries (84 Amp-Hours) are completely emptied (more than eight days of discharge), it
will take at least 84AH /5 A = 16.8 Hours (overnight), to completely re-charge them. For each hour of charging, you
should be able to get about 10 hours of runtime.
8. At 80% depth of discharge, the batteries should be good for at least 200 cycles. The batteries will age and lose
capacity over time, and will likely need to be replaced about three to five years from purchase.
9. When charging has completed you may unplug the charger and store the cable away. The enclosure lid can safely
be closed as soon as the charger is unplugged.
Reactivating the desiccant
When you first use the silica gel canister in the area to be protected, it may become saturated rather quickly as it
"drinks up" residual moisture. Once the residual dampness is removed, you can maintain a dry condition with less
frequent need to reactivate the silica gel. Reactivation times below are minimums. Sometimes, when the silica gel
becomes overly saturated, it is good practice to extend the reactivation period. However, do not vary from the
recommended temperatures.
When the blue silica gel beneath the inspection window turns pink, reactivate as follows:
•Place the unit in a vented 150°C (300°F) oven for at least 3 hours. (Or until the silica gel turns blue again).
Note:The desiccant can be reactivated as many times as you want to.
Storage Recommendations
The batteries have an extremely low self-discharge rate, thus providing extended storage capability while maintaining
high state-of- charge levels for dependable operation. These batteries have a shelf life more than two times that of
conventional lead batteries. To assure maximum reliability, TSI recommends that all stored batteries be recharged (boost
charged) at least every 12 months or when the open circuit voltage drops to 12.00 volts per battery, whichever occurs
earlier. Check the batteries more frequently if storage temperature regularly exceeds 25°C (77°F).
The outdoor measurement system should be stored in a cool dry location with the lid closed whenever possible. High
battery storage temperatures will lead to excessive self-discharge and premature battery failure.

7SoundPro Outdoor Kit Specifications
Storage Recommendations
SoundPro®Sound Level Meter Outdoor Measuring System
SoundPro Outdoor Kit Specifications
Windscreen Attenuation
0.5 dB @ 2 kHz, 1 dB @ 6 kHz, 1.4 dB @ 10 kHz, 2 dB @ 20 kHz
Battery type
(1) 12 Volt, 42 Amp-Hour battery
Battery runtime
Total capacity : 34 Amp-Hours of 12 Volt power
Approximately 5.5 days, with nominal 3400 mAH load and de-rated to 80% depth of
discharge (3400 mAH / 250 mA / 24 (H/day) = approximately 5.5 days
When other additional accessories are used, reduced runtimes are calculated from
the 80% de-rated 34 Amp-Hour Capacity.
With the optional second battery installed, runtime will increase up to 7 days
(doubling the runtime).
Operating Temperature
0°C to 50°C (32°F to 122°F)
Excessively hot and extreme sun exposure may lead to over-heating. Use additional
sun shields to reduce running temperature, if necessary.
Repeated or extended time in extreme environments will lead to premature battery
failure.
Overall Dimensions
In Transit
(W x D x maximum H)
64 cm x 51 cm x 38 cm (25 in. x 20 in. x 15 in.)
Ready to Use
(W x D x maximum H)
64 cm x 51 cm x 200 cm; (25 in. x 20 in. x 78 in.)
Overall Weight
30 kg; 67 lbs (with one battery)
45 kg; 100 lbs (with two batteries)
SoundPro Sound Level Meter is sold separately.
Figure 5: Sample SoundPro outdoor kit assembly parts

8Customer Service
Technical Support Contacts
SoundPro®Sound Level Meter Outdoor Measuring System
Customer Service
This section gives directions for contacting TSI Incorporated for technical information and directions for returning the
SoundPro Outdoor Measuring System.
Technical Support Contacts
If you have any difficulty setting or operating the instrument, or if you have technical or application questions about this
system, contact TSI’s Technical Support.
North America and Asia Pacific
Telephone: 1-800-680-1220 (USA);
+1 651-490-2860 (Outside USA)
Fax: +1 651-490-3824
E-mail: [email protected]om
Europe, Middle East, and Africa
Service Contact Information
If your instrument does not operate properly, or if you are returning the instrument for service, visit our website at
http://rma.tsi.com for a Return Material Authorization, or contact Customer Service.
North America and Asia Pacific
TSI Incorporated
1060 Corporate Center Drive
Oconomowoc, WI 53006-4828
Europe, Middle East, and Africa
TSI Instruments Ltd.
Stirling Road
Cressex Business Park
High Wycombe, Bucks
HP12 3ST
United Kingdom
Telephone: 1-800-680-1220 (USA);
+1 651-490-2860 (Outside USA)
E-mail: [email protected]om
Telephone: +44 (0) 149 4 459200
E-mail: [email protected]
Returning for Service
Visit our website at http://rma.tsi.com and complete the on-line “Return Merchandise Authorization” form or call TSI at
1-800-680-1220 (USA) or (651) 490-2860, or 001 651 490 2860 (International) for specific return instructions.
Customer Service will need the following information:
The instrument model number
The instrument serial number
A purchase order number (unless under warranty)
A billing address
A shipping address
Use the original packing material to return the instrument to TSI. If you no longer have the original packing material,
seal off any ports to prevent debris from entering the instrument and ensure that the display and the connectors on the
instrument front and back panels are protected. This instrument is very fragile and must be packed in a manner
appropriate for a precision instrument.

9Warranty
SoundPro®Sound Level Meter Outdoor Measuring System
Warranty
(For country-specific terms and conditions outside of the USA, please visit www.tsi.com.)
Seller warrants the goods, excluding software, sold hereunder, under normal use and service as described in the operator's manual, to
be free from defects in workmanship and material for 12 months, or if less, the length of time specified in the operator's manual, from
the date of shipment to the customer. This warranty period is inclusive of any statutory warranty. This limited warranty is subject to the
following exclusions and exceptions:
a. Hot-wire or hot-film sensors used with research anemometers, and certain other components when indicated in specifications,
are warranted for 90 days from the date of shipment;
b. Pumps are warranted for hours of operation as set forth in product or operator’s manuals;
c. Parts repaired or replaced as a result of repair services are warranted to be free from defects in workmanship and material, under
normal use, for 90 days from the date of shipment;
d. Seller does not provide any warranty on finished goods manufactured by others or on any fuses, batteries or other consumable
materials. Only the original manufacturer's warranty applies;
e. This warranty does not cover calibration requirements, and seller warrants only that the instrument or product is properly
calibrated at the time of its manufacture. Instruments returned for calibration are not covered by this warranty;
f. This warranty is VOID if the instrument is opened by anyone other than a factory authorized service center with the one exception
where requirements set forth in the manual allow an operator to replace consumables or perform recommended cleaning;
g. This warranty is VOID if the product has been misused, neglected, subjected to accidental or intentional damage, or is not
properly installed, maintained, or cleaned according to the requirements of the manual. Unless specifically authorized in a
separate writing by Seller, Seller makes no warranty with respect to, and shall have no liability in connection with, goods which
are incorporated into other products or equipment, or which are modified by any person other than Seller.
The foregoing is IN LIEU OF all other warranties and is subject to the LIMITATIONS stated herein. NO OTHER EXPRESS OR
IMPLIED WARRANTY OF FITNESS FOR PARTICULAR PURPOSE OR MERCHANTABILITY IS MADE. WITH RESPECT TO
SELLER’S BREACH OF THE IMPLIED WARRANTY AGAINST INFRINGEMENT, SAID WARRANTY IS LIMITED TO CLAIMS OF
DIRECT INFRINGEMENT AND EXCLUDES CLAIMS OF CONTRIBUTORY OR INDUCED INFRINGEMENTS. BUYER’S
EXCLUSIVE REMEDY SHALL BE THE RETURN OF THE PURCHASE PRICE DISCOUNTED FOR REASONABLE WEAR AND
TEAR OR AT SELLER’S OPTION REPLACEMENT OF THE GOODS WITH NON-INFRINGING GOODS.
TO THE EXTENT PERMITTED BY LAW, THE EXCLUSIVE REMEDY OF THE USER OR BUYER, AND THE LIMIT OF SELLER'S
LIABILITY FOR ANY AND ALL LOSSES, INJURIES, OR DAMAGES CONCERNING THE GOODS (INCLUDING CLAIMS BASED ON
CONTRACT, NEGLIGENCE, TORT, STRICT LIABILITY OR OTHERWISE) SHALL BE THE RETURN OF GOODS TO SELLER AND
THE REFUND OF THE PURCHASE PRICE, OR, AT THE OPTION OF SELLER, THE REPAIR OR REPLACEMENT OF THE
GOODS. IN THE CASE OF SOFTWARE, SELLER WILL REPAIR OR REPLACE DEFECTIVE SOFTWARE OR IF UNABLE TO DO
SO, WILL REFUND THE PURCHASE PRICE OF THE SOFTWARE. IN NO EVENT SHALL SELLER BE LIABLE FOR LOST
PROFITS, BUSINESS INTERRUPTION, OR ANY SPECIAL, INDIRECT, CONSEQUENTIAL OR INCIDENTAL DAMAGES. SELLER
SHALL NOT BE RESPONSIBLE FOR INSTALLATION, DISMANTLING OR REINSTALLATION COSTS OR CHARGES. No Action,
regardless of form, may be brought against Seller more than 12 months after a cause of action has accrued. The goods returned
under warranty to Seller's factory shall be at Buyer's risk of loss, and will be returned, if at all, at Seller's risk of loss.
Buyer and all users are deemed to have accepted this LIMITATION OF WARRANTY AND LIABILITY, which contains the complete
and exclusive limited warranty of Seller. This LIMITATION OF WARRANTY AND LIABILITY may not be amended, modified or its
terms waived, except by writing signed by an Officer of Seller.
Service Policy
Knowing that inoperative or defective instruments are as detrimental to TSI as they are to our customers, our service policy is
designed to give prompt attention to any problems. If any malfunction is discovered, please contact your nearest sales office or
representative, or call TSI’s Customer Service department at 1-800-874-2811 (USA) or +001 (651) 490-2811 (International).

10 Warranty
SoundPro®Sound Level Meter Outdoor Measuring System
(This page intentionally left blank)

TSI Incorporated –Visit our website www.tsi.com for more information.
USA Tel: +1 800 874 2811
UK Tel: +44 149 4 459200
France Tel: +33 1 41 19 21 99
Germany Tel: +49 241 523030
India Tel: +91 80 67877200
China Tel: +86 10 8219 7688
Singapore Tel: +65 6595 6388
P/N 057-829 Rev. F ©2018 TSI Incorporated Printed in U.S.A.
Table of contents
Other TSI Quest Measuring Instrument manuals