Tsuruga 4021 User manual

-1-
Instruction Manual
Digital Large Display Meter, Model 4021
I-01950
1. Preface
●Please take care that this instruction manual is certainly delivered to the person in charge of operating this instrument.
●Unpack the product and confirm that the following items are included.
(1) 4021 itself (2) Instruction manual (3)CJS (for thermocouple only. It is installed in the terminals)
For safety and proper use of this product, please observe the following caution and also read the instruction manuals to follow before
the initial operation.
CAUTION
●The 4021 becomes active as soon as it is connected to a power source.
It has no On or Off switch. But warm-up time is necessary at least 15 minutes.
●Do not install under the following conditions.
-Where it is exposed to direct sunlight, dust, corrosive gases, rain, etc.
-Where ambient temperature or humidity is high.
-Where it is exposed to excessive noise or static electricity.
-Where there is constant vibration or shock
2. Standard Specifications
2.1 Model Name
4021– □– □– □– □– □
( 1 ) ( 2 ) ( 3 ) ( 4 ) ( 5 )
(1) Measuring Input
[ Thermocouple input ]
Model Sensor Measuring range Display range Accuracy <Note> Overload
4021-02 K -200 〜999
oC -270 〜999
oC ±(0.3% of rdg +1oC) DC±10 V
4021-04 J -200 〜999
oC -210 〜999
oC ±(0.3% of rdg +1oC) DC±10 V
4021-05 T -200 〜400
oC -270 〜420
oC ±(0.3% of rdg +1oC) DC±10 V
< NOTE > Accuracy: Defined at 23oC±5oC, 45 to 75%RH.
Temperature coefficient: ±300ppm/oC within the 0 to 50oC temperature range.
Compensation of standard contact: ±1oC within the 0 to 50oC temperature range.
Calibration: In conjunction with JIS C 1602-1995, by each standard thermoelectric power mV input.
[ RTD input ]
Model Sensor Measuring range Display range Accuracy <Note> Overload
4021-11 Pt100Ω-200 〜850
oC -200 〜870
oC ±(0.4% of rdg +1oC) DC±10 V
4021-12 Pt100Ω-99.9 〜99.9
oC -99.9 〜99.9
oC ±(0.2% of rdg +0.3oC) DC±10 V
< NOTE > Accuracy: Defined at 23oC±5oC, 45 to 75%RH.
Temperature coefficient: ±200ppm/oC within the 0 to 50oC temperature range.
Calibration: In conjunction with JIS C 1604-1997, by the resistive value of standard resistive elements.
(2) Power Supply
No. Power
3 AC100 〜120V
5 AC200 〜240V
9 DC 24V
(3) Mounting
No. Type
51 Wall-mount
52 Hanging-mount
53 Sticking-mount
61 Wall-mount
62 Hanging-mount
63 Sticking-mount
Integral Pt sensor
(for RTD input only)
(4) Color
No Type
(Null) Red LED
G Green LED
(5) Option
No. Function
(Null) Peak Hold
A01 Bottom Hold
A02 Change of the burnout direction <NOTE>
<NOTE>: Thermocouple input only.

Model 4021
-2-
2.2 General Specifications
Display : Red or Green LED (character height 56mm) with zero-suppress function.
Over-range indication : Flashes at minimum or maximum value of the display range when exceeding.
Burnout : Blinking with maximum value of display range.
Hold Function : Measured data is held (Not isolated from input).
Peak Hold : Maximum value of the measured data is held (Not isolated from input).
Resolution : Thermocouple input; 1oC
RTD input; 1oC (-11) or 0.1oC(-12)
Allowable external resistance : Thermocouple input; 500Ωor less
RTD input; 5Ωor less / lead wire (resistance value of 3 wires shall be same)
Display Cycle : Approx. 200ms
Input Type : Single ended
A/D Conversion : ∆-Σconversion system.
Noise Rejection : Normal mode (NMR) - 50dB or more.
Common mode (CMR)-110dB or more.
Noise through power supply line
: 1000V
Insulation Resistance : Input terminals – Case : DC500V 100MΩor more.
Power terminals – Case : DC500V 100MΩor more.
Power terminals - Input terminals : DC500V 100MΩor more.
Withstanding Voltage : Input terminals – Case : AC1500V each for 1 min.
Power terminals – Case : AC1500V each for 1 min.
Power terminals - Input terminals : AC1500V each for 1min.
(in case of DC power supply, AC 500V each for 1 min.)
Power Supply : AC100 to 120V 50/60Hz, AC200 to 240V 50/60Hz, DC24V
Allowable power : AC85 to 132V 50/60Hz, AC170 to 250V 50/60HZ, DC20 to 30V
Power Consumption : Approx. 6VA at AC100V, Approx.10VA at AC200V, Approx.100mA at DC24V
Operating Temperature : 0 to 50oC
Storage Temperature : -20 to 70oC
Weight : Approx. 1.6kg
Unit :
oC
Protection : IP44 (Wall-mount and hanging-mount)
IP65 (Sticking-mount with water-proof works)
2.3 Optional specifications
●Bottom Hold function
Display the minimum value of the measured data (Not isolated from input).
In this case, Peak Hold function is disabled.
●Change of the burnout direction
It is Available to flash the minimum value when using the thermocouple input.

Model 4021
-3-
2.4 Mounting
●Wall-mount (model code-51)
NOTE: Cut the case bottom or side to pull out lead wires.
●Hanging-mount (model code-52)
NOTE: Fixing points should be locating 10mm away from the wall.
●Sticking-mount (model code-53)
NOTE: Use coaching bond or appropriate sealant to keep IP65 protection.
CAUTION
・Hanger hole of the mounting panel should be
upright position as shown in the drawing.
Wall
Mount to the top roof
Wall
Make the wall in waterproof
by the sealant
Hanger
hole

Model 4021
-4-
3. Setting of Each Function
●Name of each part
●Display
Lit-up display
Blinking display
Turned-off display
●Function of each switch
Functions switch F : Change between the measuring
mode and the setting mode.
Mode switch M : Change parameter item at the setting
mode.
Shift switch >: Move to the next action or shift the
value.
Up switch ∧: Scroll or increase the value or figure
●Setting mode
- Changes to the settings are saved to EEPROM when returning to the measuring mode. Display temporally turns off the light
- Once flashing the new setting value in the setting mode, it is saved into the memory, and peak value (or bottom value) goes clear.
- If you do not touch any switch for 5 minutes at the setting mode, it automatically returns to the measuring mode without saving
changes.
3.1 Setting method
3.1.1 Average operation
Average operates the interval average or the moving average.
Example) Changes the non-average operation to the moving average of 8-data.
- Since the interval average operates 6-data, display cycle is approx. 1.2s.
- In case of moving average, display cycle is approx. 200ms.
[ Default ]
Function Setting value
Average operation Enable Interval average
[ Contents of average operation ]
Display Contents
Without average operation
Interval average
Moving average, 2-time
Moving average, 4-time
Moving average, 8-time
Moving average, 16-time
Moving average, 32-time
℃
P
DP0DP1DP2
MF
※
(BH LED)
PH LEDPolarity LED
Unit
Terminals
Setting switch※
>∧
P
P
P
Measuring mode
Display the existing value
F switch
Setting mode
Push the >switch to move to the setting mode.
(Blinking display)
Push the Λ switch to set the average calculation
value.
Exit this parameter by pressing M switch and display
F switch Approx. 3 sec.
M switch
℃
P
℃
P

Model 4021
-5-
4. Wiring
CAUTION
●Do not miswiring. Otherwise, the meter is broken.
●To avoid an electrical shock, turn the power off when wiring.
●Do not wire with moistened hands. Locate away from the wet place.
●Do not touch terminals when turning the power on.
4.1 Terminal Arrangement
[ Thermocouple input ]
NC + - CJS+ CJS- COM HOLD PH NC P2(+) P1(-)
Terminal 1 2 3 4 5 6 7 8 9 10 11
Function
−
Input CJS Common Hold Peak Hold
−
Power supply
[ RTD input ]
A B B NC NC COM HOLD PH NC P2(+) P1(-)
Terminal 1 2 3 4 5 6 7 8 9 10 11
Function Input
−
Common Hold Peak Hold
−
Power supply
Terminal screws: M3
Fastening torque: 0.46~0.62N m
Crimp terminal: As shown on the right.
●Input terminals
[ Thermocouple input ]
Pay attention to the polarity when wiring.
CJS is not compatible.
Make sure that the serial number
of the CJS and the meter must match.
[ RTD input ]
Connect 3-wire of the RTD.
●Hold (HOLD)
Display can be held by connecting the Hold terminal and the Common terminal.
Active “L” IIL≦-1mA, “L”=0 to 0.8V, “H”=3.5 to 5V
●Peak Hold (PH)
Maximum value is displayed by connecting the Peak Hold terminal and the Common terminal. “PH” and LED are lit during the
peak hold function. This function is released by disconnecting the PH terminal.
Active “L” IIL≦-1mA, “L”=0 to 0.8V, “H”=3.5 to 5V
●Bottom Hold (PH)
If your meter code is –A01, PH terminal is for the bottom hold. Minimum value is displayed by connecting the Peak Hold terminal
and the Common terminal. “BH” and LED are lit during the bottom hold function. This function is released by disconnecting the
PH terminal.
Active “L” IIL≦-1mA, “L”=0 to 0.8V, “H”=3.5 to 5V
●Common (COM)
“COM” terminal is used for Hold and Peak (or Bottom) Hold terminals.
NOTE: Those terminals, COM, HOLD and PH, are not isolated to the input.
Use a photo-coupler or switch to insulate.
It is essential when using the input floating.
●Power Supply (P1(-), P2(+))
The power supply voltage is specified on the terminal plate at delivery from our factory.
- AC power supply(3): Use the meter within the range AC85 to 132V.
- AC power supply(5): Use the meter within the range AC170 to 250V.
- DC power supply(9): Use the meter within the range DC20 to 30V.
Connect +24V positive of DC power to P2 (+), and 0V negative side to P1 (-).
CAUTION
●Do not use the product with the voltage out of the rated range as it may cause breakage of the products.
φ3.2min.
6max.
5.3min. 4.4max.
42
+
1
NC
53
−
CJS
+
−
Gray
Purple
Thermocouple
CJS−CJS+
32
B
1
A
4 5
RTD
BNCNC
+5V
1mA or less
HOLD
photo‑coupler
COMswitch
IN Lo
4021

Model 4021
-6-
5. Maintenance
Store the meter within the rated storage temperature (from -20 to +70oC). When the front panel or the case is cleaned, use soft cloth
dipped with cleaner liquid. Do not use organic solvent like benzene or paint thinner as they may deform or discolor the case.
6. Calibration
In order to maintain long term accuracy, periodical calibration at an interval of about one year is recommended. Make a calibration in
the ambient condition of 23oC±5oC, 75%RH or less.
6.1 Calibration of Thermocouple input
Prepare the standard voltage generator, cold contact circuit (ice-bath), and standard thermocouple for calibration.
To enter the calibration mode, turn the power on with pushing M switch.
Connect the voltage generator and the CJS thermocouple as shown in the Figure-1.
Turn the power on with pushing the M switch.
Hold down the M switch for over five seconds.
After the lump test (all LED light up), display shows the sensor number, and then proceeds to the
Calibration mode with “
” message.
Press the M switch to enter ZERO and MAX calibration.
Display shows the measuring value. Resolution is 0.1oC for 100oC or lower, and 1oC for more than
100oC.
-ZERO calibration
Set the voltage generator to 0.00mV as shown in the Figure-1.
Press the >switch to momentarily memorizes the ZERO value. NOTE1
- MAX calibration
Set the voltage generator to the maximum voltage.
Press the Λ switch to momentarily memorize
the MAX value. NOTE1
Press the M switch to enter GAIN calibration
of the thermocouple in the ice-bath.
Display shows the measuring value. NOTE2
Set the voltage generator to 500.0mV as shown in the Figure-1.
Press the Λ switch to momentarily memorize the calibration value.
Press the M switch to enter ZERO calibration of the thermocouple in the ice-bath.
Connect the voltage generator, the CJS thermocouple, and the ice-bath as shown in the Figure-2.
Display shows the measuring value. NOTE2
Set the voltage generator to 0.00mV.
Press the >switch to momentarily memorizes the ZERO value.
Press the M switch to show “
”
Press the Λ switch to be complete your calibration. Display flashes to confirm your calibration.
Do not turn the power off to confirm your calibration.
An error message “
” will be displayed if you calibrate out of range.
Power OFF
Power OFF
No. Sensor MAX
K 950℃39.314mV
J 950℃54.956mV
T 400℃20.872mV
NOTE1: An error message “
” will be displayed
if you calibrate out of range.
NOTE2: Resolution is 0.1mV for 100mV or lower,
and 1mV for more than 100mV.
4
2
+
1
NC
53
−
CJS
CU+ CU−
generator
Standard voltage
Gray
Purple
3
2
+
1
NC
4
CJS
5
−
+−Gray
Purple
generator
Standard voltage
ice water 0℃
Standard
Thermocouple
CU−
CU+
Fig.‑1 Fig.‑2

Model 4021
-7-
6.2 Calibration of RTD input
Prepare the standard variable resistor.
To enter the calibration mode, turn the power on with pushing M switch.
Connect the standard variable resistor as shown in the Figure-3.
Turn the power on with pushing the M switch.
Hold down the M switch for over five seconds.
After the lump test (all LED lit up), display shows the sensor number, and then proceeds to the
Calibration mode with “
” message.
Press the M switch to enter ZERO and MAX calibration.
Display shows the measuring value. Resolution is 0.1oC for 100oC or lower, and 1oC for more than
100oC.
“
” in case of Pt100Ω(-200 to +850oC)
-ZERO calibration
Set the variable resistor to 100.0Ω(0.0oC).
Press the >switch to momentarily
memorizes the ZERO value. NOTE1
- MAX calibration
Set the variable resistor to 375.70Ω(800oC).
Press the Λ switch to momentarily memorize the MAX value. NOTE1
Press the M switch to show “
”.
Press the Λ switch to be complete your calibration. Display flashes to confirm your calibration.
Do not turn the power off to confirm your calibration.
An error message “
” will be displayed if you calibrate out of range.
Power OFF
Power OFF
No. Sensor (range) MAX
Pt100Ω(-200〜850℃) 800℃375.70Ω
Pt100Ω(-99.9〜99.9℃) 95.0℃136.61Ω
NOTE1: An error message “
” will be displayed if you
calibrate out of range.
2
B
1
A
3
B
BB
A
Standaed variable resistor
Fig.‑3

Model 4021
-8-
7. Dimensions
7.1 Basic type
●Wall-mount (-51)
●Hanging-mount (-52)
●Sticking-mount (-53)
Unit: mm
P
2‑φ4.5
℃
P
2‑φ4.5
P
℃
℃
R2.25
R4
4‑φ4.5
R2.25
R4
Provided at both side
and bottom
Provided at both side
and bottom
3
52 10max
135
270
40
245
33.5
18
245
52
55
135
270
33.5
18
4‑R5
25
52
245
70
55
135
270
8
81211
8
8
175
8
55
Split‑able
Split‑able

Model 4021
-9-
7.2 Integral Pt sensor
●Wall-mount (-61)
●Hanging-mount (-62)
●Sticking-mount (-63)
Unit: mm
℃
P
2‑φ4.5
℃
P
135 40
270
245
33.5
18
55
135
245
270
33.5
18
2‑φ4.5
℃
P
4‑R5
25
55
135
245
70
270
R2.25
R4
52
8
Split‑able
Provided at both side
side and bottom
R2.25
R4
52
8
4‑φ4.5
3
52 10max
81211
55
8
175
Split‑able
Provided at both side
side and bottom
8
Contact Information
Name : Tsuruga Electric Corporation
Address : 1-3-23 Minami-Sumiyoshi, Sumiyoshi-ku, Osaka-shi
558-0041 Japan

- 1 -
Instruction Manual
Digital Large Display Meter, Model 4022 I-01951
1. Preface
●Please take care that this instruction manual is certainly delivered to the person in charge of operating this instrument.
●Unpack the product and confirm that the following items are included.
(1) 4022 itself (2) Instruction manual
For safety and proper use of this product, please observe the following caution and also read the instruction manuals to follow before
the initial operation.
CAUTION
●The 4022 becomes active as soon as it is connected to a power source.
It has no On or Off switch. But warm-up time is necessary at least 15 minutes.
● Do not install under the following conditions.
-Where it is exposed to direct sunlight, dust, corrosive gases, rain, etc.
-Where ambient temperature or humidity is high.
-Where it is exposed to excessive noise or static electricity.
-Where there is constant vibration or shock.
2. Standard Specifications
2.1 Model Name
4022– □– □– □– □– □
( 1 ) ( 2 ) ( 3 ) ( 4 ) ( 5 )
(1) Measuring Input
Model
Measuring range
Input resistance
Accuracy <Note>
Overload
4022-03
± 999 mV
1MΩ
± (0.1% of rdg +1 digit)
DC ±250 V
4022-04
± 9.99 V
1MΩ
± (0.1% of rdg +1 digit)
DC ±250 V
4022-09
DC 1
~
5 V
1MΩ
± (0.1% of rdg +2 digit)
DC ±250 V
4022-19
DC 4
~
20 mA
12.5Ω
± (0.1% of rdg +2 digit)
DC ±150 mA
< NOTE > Accuracy: Defined at 23oC±5oC, 45 to 75%RH.
Temperature coefficient: ± 100ppm / oC for the 4022-03, ± 150ppm / oC for the 4022-04, -09, and -19
Temperature range: 0 to 50oC
(2) Power Supply
No.
Power
3
AC100
~
120V
5
AC200
~
240V
9
DC 24V
(3) Mounting
No.
Type
51
Wall-mount
52
Hanging-mount
53
Sticking-mount
(4) Color
No
Type
(Null)
Red LED
G
Green LED
(5) Option
No.
Function
(Null)
Peak Hold
A01
Bottom Hold

Model 4022
- 2 -
2.2 General Specifications
Display : 0 to 999, Red or Green LED (character height 56mm) with zero-suppress function.
Scaling function : Full scale display -999 to +999, programmable.
Offset display -999 to +999, programmable.
Hold function: : Measured data is held (Not isolated from input).
Peak hold: : Maximum value of the measured data is held (Not isolated from input).
Decimal point : Programmable by the terminal plate (Not isolated from input).
Over-range indication : It blinks at130% display. When exceeded 999, blinking with 000.
Resolution : 1/1000
Display cycle : Approx. 67ms.
Zero set function : Reset the input signal to default.
Offset fixing function : Fix a display reading of input less than offset value to the offset value.
Average function : Average operates the interval average or the moving average.
Input type : Single ended.
A/D conversion : ∆-Σ conversion system.
Noise rejection : Normal mode (NMR) 50dB or more.
Common mode (CMR) 110dB or more.
Noise through power supply line
: 1000V
Insulation resistance : Input terminals – Case : DC500V 100MΩor more.
Power terminals – Case : DC500V 100MΩor more.
Power terminals – Input terminals : DC500V 100MΩor more.
Withstanding voltage : Input terminals – Case : AC1500V each for 1 min.
Power terminals – Case : AC1500V each for 1 min.
Power terminals – Input terminals : AC1500V each for 1min.
(in case of DC power supply, AC 500V each for 1 min.)
Power supply : AC100 to 120V 50/60Hz, AC200 to 240V 50/60Hz, DC24V
Allowable power : AC85 to 132V 50/60Hz, AC170 to 250V 50/60HZ, DC20 to 30V
Power consumption : Approx. 6VA at AC100V, Approx.10VA at AC200V, Approx.100mA at DC24V
Operating temperature : 0 to 50oC
Storage temperature : -20 to 70oC
Weight : Approx. 1.6kg
Unit : Specified character
Protection : IP44 (Wall-mount and hanging-mount)
IP65 (Sticking-mount with water-proof works)
2.3 Optional specifications
●Bottom Hold function
Display the minimum value of the measured data (Not isolated from input).
In this case, Peak Hold function is disabled.

Model 4022
- 3 -
2.4 Mounting
●Wall-mount (model code-51)
NOTE: Cut the case bottom or side to pull out lead wires.
●Hanging-mount (model code-52)
NOTE: Fixing points should be locating 10mm away from the wall.
●Sticking-mount (model code-53)
NOTE: Use coaching bond or appropriate sealant to keep IP65 protection.
CAUTION
・
Hanger hole of the mounting panel should be
upright position as shown in the drawing.
Wal l
Mount to the ceiling.

Model 4022
- 4 -
3. Setting of Each Function
●Name of each part
●Display
Lit-up display
Blinking display
Turned-off display
●Function of each switch
Functions switch F : Change between the measuring
mode and the setting mode.
Mode switch M : Change parameter item at the setting
mode.
Shift switch >: Move to the next action or shift the
value.
Up switch ∧: Scroll or increase the value or figure.
●Programming configuration
- Push the >switch to move to the setting mode.
- Changes to the settings are saved to EEPROM when returning to the measuring mode and adjustment mode. Display temporally
turns off the light.
- In case of moving average, display cycle is fixed to approx. 67ms.
- Once flashing the new setting value in the setting mode, it is saved into the memory, and peak value (or bottom value) goes clear.
- If you do not touch any switch for 5 minutes at the setting mode, it automatically returns to the measuring mode without saving
changes.
offset full scale offset fixing moving average zero-set
Setting function
M
M
M
M
M
M
F
Measuring mode
Setting mode
ZERO MAX
Adjustment
function
M
M
M
UP, DOWN setting
F
F
F
6seconds
3seconds
F
℃
P
DP0DP1DP2
M
F
※
(BH LED)
PH LED
Polarity LED
Unit
Terminals
Setting switch※
>∧

Model 4022
- 5 -
3.1 Setting method
3.1.1 Display scaling - Offset setting
Offset display can be set to an arbitrary value.
Example: With the rated input 1to 5V, the display at the input 1V is adjusted from 0 to 100.
- Offset display means the
display at 0mV input for the
voltage measuring range, at 1V
input for the 1-5VDC range,
and at 4mA input for 4-20mA
range.
Λswitch:
0→1→2→・・・・→9→0→1
In case of 102-digit:
0→1→2→・・・・→9→-9→-8→
・・・・→-1→-0→0→1
3.1.2 Display scaling - Full scale setting
Full scale display can be set to an arbitrary value.
Example: With the rated input 1 to 5V, the display at the input 5V is adjusted from 999 to 599.
- Full scale display means the
display at maximum value input.
Λswitch:
0→1→2→・・・・→9→0→1
In case of 102-digit:
0→1→2→・・・・→9→-9→-8→
・・・・→-1→-0→0→1
3.1.3 Offset fixing
Display can be fixed to the offset value when the input value is lower than the offset value.
Example: Change the offset fixing function from disable to enable.
- If your meter chooses
the4-20mA range and sets the
display scaling to 100-500,
display shows 100 even though
the input signal is less than
4mA.
:Enable the offset fixing.
:Disable the offset fixing.
Measuring mode
Display shows offset value. (DP0 and DP1 lit up)
DP0 and DP1 Blinking.
Setting mode
Push the >switch to move to the setting mode.
Push the >switch to scroll digits to select the desired digit
to modify (blinking the selected digit).
Push the Λswitch to set the new setting value.
It change display scaling - full scale setting by pressing
M
switch.
Push
F
switch for 3seconds.
Push M switch.
Display shows full scale value. (DP1 and DP2 lit up)
DP1 and DP2 Blinking.
Setting mode
Push the >switch to move to the setting mode.
Push the >switch to scroll digits to select the desired digit
to modify (blinking the selected digit).
Push the Λswitch to set the new setting value.
It change offset fixing by pressing
M
switch.
Display shows setting condition
Setting mode
Push the >switch to move to the setting mode. (blinking )
Push the Λswitch to set “on”or “off”.
It change average operation by pressing
M
switch.
P

Model 4022
- 6 -
3.1.4 Average operation
Average operates the interval average or the moving average.
Example: Changes the non-average operation to the moving average of 8-data.
- Since the interval average operates 6-data, display cycle is approx. 400ms.
- In case of moving average, display cycle is approx. 67ms.
3.1.5 Zero set
Reset the input signal electrically to zero.
Example: Change the zero set function from disable to enable.
- If you enable the zero set and
cause short circuit between the
zero set terminal and the COM
terminal, display shows the
offset value (as same as the
setting value at chapter 3.1.1)
Example:
If you execute the zero set at
100 when scaling 0-500,
display would show -100 to
0 to +400.
:Enable the zero set.
:Disable the zero set.
[ Default ]
Function Setting value
Offset
0
Full scale
999
Offset fixing
Disable
Average operation
Enable interval average
Zero set
Disable
[ Contents of average operation ]
Display
Contents
Without average operation
Interval average
Moving average, 2-time
Moving average, 4-time
Moving average, 8-time
Moving average, 16-time
Moving average, 32-time
Display shows setting condition
Setting mode
Push the >switch to move to the setting mode. (blinking )
Push the Λswitch to set the contents of average operation.
It change zero set by pressing
M
switch.
P
Display shows that zero set function is now disable .
Setting mode
Push the >switch to move to the setting mode. (blinking )
Push the Λswitch to set “on”or “off”.
Exit this parameter by pressing
M
switch.
Measuring mode
Push F switch
.
P

Model 4022
- 7 -
3.2 Adjustment function
3.2.1 Zero adjustment
Fine calibration of the offset displayis available to meet the real input value.
Example: your meter is 1-5V input, and when inputting 1V, displayed value is 5, not 0.
3.2.2 MAX adjustment
Fine calibration of the maximum display is available to meet the real input value.
Supply the input as close as the rated maximum value.
Display shows the input value. (DP1 and DP2 lit up)
Exit this parameter by pressing
M
switch.
Measuring mode
Push F switch. Blinking for 1seconds.
Push the >Λswitch to adjusted to the full
scale value.
Displayed straight line lists.
Display
Input
>switch to scroll the setting value
to decrement.
Λswitch to scroll the setting value
to increments.
It may take a while if the scaling width
is narrow. Keep the switches pushing
for some time.
Measuring mode
Display shows offset value. (DP0 and DP1 lit up)
Setting mode
Push the >Λswitch to adjusted to the offset
value.
It change MAX adjustment by pressing
M
switch.
Push
F
switch for 6seconds.
Push M switch.
Displayed straight line translates.
Display
Input
>switch to scroll the setting value
to decrement.
Λswitch to scroll the setting value
to increments.
It may take a while if the scaling width
is narrow. Keep the switches pushing
for some time.

Model 4022
- 8 -
4. Wiring
CAUTION
● Do not miswiring. Otherwise, the meteris broken.
● To avoid an electrical shock, turn the power off when wiring.
● Do not wire with moistened hands. Locate away from the wet place.
●Do not touch terminals when turning the power on.
4.1 Terminal Arrangement
Terminal
INHi
NC
INLo
DP1
DP2
COM
HOLD
PH
ZS
P2(+)
P1(-)
1
2
3
4
5
6
7
8
9
10
11
Function Input
- Input
101digit
102digit
Common
Hold Peak
hold Zero
set Power supply
Decimal point
Terminal screws: M3
Fastening torque: 0.46~0.62N・m
Crimp terminal: As shown on the right.
●Input terminals (INHi, INLo)
Pay attention to the polarity when wiring. Connect input of higher electric potential to Hi. Input and power line shall lay separately.
Otherwise, display may be unstable.
●Hold (HOLD)
Display can be held by connecting the Hold terminal and the Common terminal.
Active “L” IIL≦-1mA, “L”=0 to 0.8V, “H”=3.5 to 5V
●Decimal point (DP1, DP2)
Decimal point is programmable.
Connect and short-circuit the desired decimal point terminal,
DP1 or DP2, and the common terminal.
Active “L”, IIL ≦-1mA, “L”=0 to 0.8V, “H”=3.5 to 5V
If you duplicate both DP1 and DP2, decimal point does not light up.
●Peak Hold (PH)
Maximum value is displayed by connecting the PH terminal and the Common terminal. “PH” LED are lit during the peak hold
function. This function is released by disconnecting the PH terminal.
Active “L” IIL≦-1mA, “L”=0 to 0.8V, “H”=3.5 to 5V
●Bottom Hold (PH)
If your meter code is –A01, PH terminal is for the bottom hold. Minimum value is displayed by connecting the PH terminal and the
Common terminal. “PH” and LED are lit during the bottom hold function. This function is released by disconnecting the PH
terminal.
Active “L” IIL≦-1mA, “L”=0 to 0.8V, “H”=3.5 to 5V
●Zero set (ZS)
Zero set enables by choosing “on” at the rear switch. Zero set value is memorized into EEPROM for approx. 10 years.
Active “L” IIL≦-1mA, “L”=0 to 0.8V, “H”=3.5 to 5V
- Method:
1. Shift the rear switch of the zero set to on side.
2. Input the zero set value, and cause short circuit between the zero set terminal and the COM terminal. Display shows 0 (when you
set the offset value to 0).
3. If you disconnect the zero set terminal and the COM terminal, zero set value is memorized and zero set function starts.
Display value = Input value – Zero set value.
- Release
1. Shift the rear switch of the zero set to off side. But zero set value is still memorized.
φ3.2min.
6max.
5.3min. 4.4max.
DP1DP2COM

Model 4022
- 9 -
●Common (COM)
“COM” terminal is used for Hold and Peak (or Bottom) Hold, Zero set, and Decimal point terminals.
NOTE: Those terminals, COM, DP1, DP2, HOLD, ZS, PH, are not isolated to the input.
Use a photo-coupler or switch to insulate.
It is essential when using the input floating.
●Power Supply (P1(-), P2(+))
The power supply voltage is specified on the terminal plate at delivery from our factory.
- AC power supply(3): Use the meter within the range AC85 to 132V.
- AC power supply(5): Use the meter within the range AC170 to 250V.
- DC power supply(9): Use the meter within the range DC20 to 30V.
Connect +24V positive of DC power to P2 (+), and 0V negative side to P1 (-).
CAUTION
●Do not use the product with the voltage out of the rated range as it may cause breakage of the products.
5. Maintenance
Store the meter within the rated storage temperature (from -20 to +70oC). When the front panel or the case is cleaned, use soft cloth
dipped with cleaner liquid. Do not use organic solvent like benzene or paint thinner as they may deform or discolor the case.
6. Calibration
In order to maintain long term accuracy, periodical calibration at an interval of about one year is recommended.
Refer to the chapter 3.2 for calibration.
Make a calibration in the ambient condition of 23oC±5oC, 75%RH or less.

Model 4022
- 10 -
7. Dimensions
●Wall-mount (-51)
● Hanging-mount (-52)
● Sticking-mount (-53)
Unit: mm
P
2-φ4.5
℃
P
2-φ4.5
P
R2.25
R4
4-φ4.5
R2.25
R4
Provided at both side
and bottom
Provided at both side
and bottom
3
52 10max
135
270
40
245
33.5
18
245
52
55
135
270
33.5
18
4-R5
25
52
245
70
55
135
270
8
81211
8
8
175
8
55
Split-able
Split-able
ton
V