UEi C164-N User manual

( I
C 1
TEST INSTRUMENTS ™
C162/C163/C164
ENGLISH
RoHS
Compliant
REACH
Compliant
CO2
TECHNOLOGY
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176 - TestEquipmentDepot.com

2
FUNCTIONS
•Measures: CO, CO2, NO (C164 only), Flue Temperature, Inlet temperature
•Calculates:
O2, Differential Temperature, CO/CO2Ratio, Efficiency (Net, Gross, High C), Excess Air,
CO/CO2, Losses, Differential
Pressure (C163 and C164 only)
FEATURES
TABLE OF CONTENTS
Functions.............................................2
Features..............................................2
General Specifications ................................3
Important Safety Warnings ............................3
Symbols..............................................4
Analyzer Overview ..................................4-5
Overview ...........................................6
Pre Test Checklist ...................................6
Setting Inlet Temperature ...........................6
Analyzer Connections ...............................6
Emptying & Cleaning the In-Line Water Trap ........7
Changing The Particle Filter.........................7
Quick Start ..........................................7
Fresh Air Purge......................................7
Over-Range Pump ...................................7
Measuring Flue Gases...............................7
Display Parameters................................8-9
Status Screen .......................................9
Status Bar ..........................................10
Status Bar Icons ...................................10
Status Bar Icon Layout ..............................10
Status Bar Menu Options ............................10
Status Bar Options Table ............................10
Menu ............................................10-11
Using the Menu ...................................10
Menu Items.......................................11
Stored Logs Memory.............................11-13
Menu Options .....................................12
Viewing Stored Logs................................12
Log View Menu Options .............................12
Navigating Stored Logs .............................12
Log Navigation Menu Options ........................13
Pressure/Temperature Testing ...................13-14
Pressure Testing (if Pressure Fitted). . . . . . . . . . . . . . . . . . . .13
Pressure Measurement Good Practice..................13
Large Bore Tubing Issues ............................13
Heat Exchanger Test..................................14
Viewing/Printing .....................................14
Printouts ...........................................15
Specifications......................................16
Certification........................................16
Where to Test ...................................17-18
What Results are Generally Acceptable ............18
What Results Are Generally Acceptable ................18
Typical Excess Air Level .............................18
Powering Off .......................................19
Post Test ...........................................19
General Maintenance ..............................19
Cold Weather Precautions..........................19
Replacing the Batteries ............................20
Annual Service & Recertification...................20
Recertification Services............................21
Returning Your Analyzer............................21
Where to Send Your Analyzer. . . . . . . . . . . . . . . . . . . . . . .21
Other Important Factors Relating to Combustion ....22
Combustion Measurement Terms ...................22
Net Temperature...................................22
Draft ............................................22
Efficiency.........................................22
Combustion Efficiency Calculations ....................22
CO Air Free .......................................22
Disposal............................................24
Cleaning ...........................................24
Storage.............................................24
Warranty ...........................................24
• EOS Technology
• Over-Range Protection Pump
•
NOx Filtered CO Sensor
•
Large 6 line Backlit Display
•
Water Trap Indicator
•
High Altitude Compensation
• Wireless BT Module
• Low Flow Detection

3
GENERAL SPECIFICATIONS
IMPORTANT SAFETY WARNINGS
Read entire Safety Notes section regarding potential hazard and proper instructions before using this analyzer. In this manual the
word “WARNING” is used to indicate conditions or actions that may pose physical hazards to the user. The word “CAUTION” is used to
indicate conditions or actions that may damage this instrument. This analyzer must only be used in well-ventilated locations by trained
and competent persons after due consideration of all potential hazards.
WARNING
To ensure safe operation and service of the tester, follow these instructions. Failure to observe these warnings can result in severe
injury or death.
WARNING
• Do not use this analyzer during electrical storms or in wet weather.
• To avoid false readings, charge batteries if a low battery indicator appears. (If fitted with rechargeable batteries)
• Always adhere to national and local safety codes. Use proper personal protective equipment (PPE).
WARNING
This analyzer extracts combustion gases that may be toxic in relatively low concentrations. These gases are exhausted from the back
of the analyzer.
WARNING
This analyzer is designed for trade professionals who are familiar with the hazards of their trade. Observe all recommended safety
procedures that include proper lockout utilization and use of personal protective equipment that includes safety glasses, gloves and
flame resistant clothing.
Users of portable gas detectors are recommended to conduct a “bump” check before relying on the unit to verify an atmosphere is free
from hazard. A ‘bump” test is a means of verifying that an instrument is working within acceptable limits by briefly exposing to a known
gas mixture formulated to change the output of all the sensors present.
NOTE: This is different from a calibration where the instrument is also exposed to a known gas mixture but is allowed to settle to a
steady figure and the reading adjusted to the stated gas concentration of the gas of the test gas.
•Operating Temperature: 0˚ to 113˚F (-18˚ to 45˚C)
•Storage Temperature: 0˚ to 113˚F (-18˚ to 45˚C)
•Operating Humidity: 15% to 90% R.H.
•Back light: Yes
•Dimensions: 8.54 x 4.18 x 1.86 inch
• Item Weight: 1.5 lb
•Calibration: Recommended Annually
•Certification: CE Conformity, RoHS, REACH Compliant,
AHRI 1260 standard
•Battery Type: NiMH (AA) 3
•Accuracy: ± (% of reading + # of least significant digits)

OF
·c
® □
®
T A
•
4 Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176 - TestEquipmentDepot.com




8
DISPLAY PARAMETERS
The large display is backlit with 6 lines, the last line is the Status Bar line.
Menu Screen
Time (settable parameter; HH:MM:SS, displayed in 24 hour clock format)
Date (settable parameter; DD:MM:YY)
Header (settable parameter; 16 character, 2 line; 10 characters can be added to Line 1 of the header, 6 characters to Line 2)
IR Print (KMIRP2 or IRP-2)
˚C/˚F (settable parameter)
NOx Ref (settable parameter; select 0.0% to 20.0%)
Gas Unit (settable parameter; select between ppm(n), ppm, mg/kWh(n), mg/kWh, mgm3(n), mgm3)
Language (settable parameter; select between English, Espanol, Francais)
EFF (Efficiency - settable parameter; select between Nett or Gross
O2 Ref (settable parameter; select 0.0% to 20.0%)
Logs (if logs are saved, options displayed; COMB., AUX, PRS/TMP, MEM)
Code (for Authorized Service Personnel only)
AUX Screen
(Default selections can be changed by user to desired selections; select between O2, CO, NO, Display Fuel Type, CAL, XAIR, LOSS,
EfGc, ATM, Ti, ∆T, T2, T1, CO2, COa, or COn, P1, P2, (if fitted for each line))
Default Selections Are:
O2 (displayed in %)
COn (displayed in p)
CO2 (displayed in %)
NOn (displayed in p)
NOxn (displayed in p)
Flue 1 Screen
Default Selections Are:
O2 (displayed in %)
COn (displayed in p)
CO2 (displayed in %)
NOn (displayed in p)
NOxn (displayed in p)

con
09:42 ► I
9
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176 - TestEquipmentDepot.com







16
SPECIFICATIONS
Parameter Range Resolution Accuracy
Temperature Measurement
Flue Temperature
Inlet Temperature (Internal Sensor)
Inlet Temperature (External Sensor)
32˚ to 1112˚F (0˚ to 600˚)
32˚ to 122˚F (0° to 50°C)
32˚ to 1112˚F (0˚ to 600˚)
0.1˚F (0.1˚C)
0.1˚F (0.1˚C)
0.1˚F (0.1˚C)
±0.5˚F (0.5˚C)
±1˚F (1˚C)
±0.5˚F (0.5˚C)
Flue Gas Measurement
Carbon Monoxide
Nitric Oxide (if fitted)
0 - 2000ppm
0 to 600ppm
1ppm
1ppm
±3ppm or ±5% rdg
(whichever is greater)
±5ppm or 5%
(whichever is greater)
Calculations
Oxygen
CO/CO2Ratio
Efficiency (Net or Gross)
Efficiency High (C)
Excess Air
0 - 21%
0 - 0.9999
0 - 99.9%
0 - 119.9%
0 - 119.9%
0.1%
0.0001
0.1%
0.1%
0.1%
±0.3% volume
±5% rdg
±1% rdg
±1% rdg
±0.2% rdg
Pre-programmed Fuels Pellets, Light Oil, LPG, Butane, Propane, Natural Gas, Bio Oil, Heavy Oil
Battery Life >8 hours (continuous with pump on)
Certification The C162/C163/C164 are TUV-tested and certified to EN 50379, Parts 1-3 in accordance to 1st
German Federal Emission Control Ordinance (Bim5chV); Meets requirements for AHRI 1260
Operating Conditions
Temperatures
Humidity
32˚ to 113˚F (0˚ to 45˚C)
15 to 90% RH, (non-condensing)
Power Supply Rechargeable batteries, USB Charging
Physical Characteristics
Weight
Dimensions
1.5 lb. (680g)
H: 8.54”x W: 4.18”x D: 1.86” (H: 217 mm x W:106 mm x D: 47mm)
These analyzers are in conformity with the relevant Union harmonization legislation listed below:
Directive Title
201430EU Electromagnetic Compatibility (EMC)
201165EU Restriction of the use of certain hazardous substances in electrical and electronic equipment (RoHS)
CERTIFICATION
The C162/C163/C164 are TUV-tested and certified to EN 50379, Parts 1, 2 & 3 in accordance to 1st German Federal Emission Control
Ordinance (BlmSchV)
EMC
EN507270:2015
Safety
EN61010-1:2010
RoHS
IEC62321-2:2013, IEC62321-1:2013; IEC62321-3-1;2013, IEC63321-5:2013, IEC623321-4:2013, IEC62321-7-2:2017, IEC62321-7-1:2015,
IEC62321-6:2015
AHRI
1260
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176 - TestEquipmentDepot.com



-----------------�
°
T2 ---F
09: 42 I
19
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176 - TestEquipmentDepot.com

This manual suits for next models
4
Table of contents
Other UEi Measuring Instrument manuals
Popular Measuring Instrument manuals by other brands

Thermo Scientific
Thermo Scientific Orion Star A111 reference guide

UE Systems
UE Systems ULTRAPROBE 10000 General Setup and Knowledge

MTS Systems
MTS Systems Level Plus STI Installation & operation manual

TRIMOS
TRIMOS OPTIMA manual

ComSonics
ComSonics Qualifier X-Ray user guide

Tektronix