Unicare Health Kinetic Spectra Instruction Manual

Kinetic Spectra
Instruction & Safety Manual
Please read this manual before operation, for
maintenance instructions and safe usage
The enabling power
of applied knowledge

Kinetic Spectra, Instruction & Safety Manual
T08 9351 7500 E[email protected] Wwww.unicarehealth.com.au
Index
Defi nition ................................................................................................................................ 2
Description.............................................................................................................................3
Safety.... 4
Electrical connection ............................................................................................................5
Use of the hand control
Select your language ...............................................................................................6
Locking the hand control setting ..............................................................................7
START/STOP/REVERSE function ...........................................................................7
Quick Start ...............................................................................................................8
Possible values for each parameter.........................................................................8
How to adjust the basic parameters of the movement .............................................9
How to modify programs PROGRAM MODE...........................................................10-12
Using Programs .......................................................................................................13
Reading the values of a program .............................................................................14
How to use the WARM UP key ................................................................................15
How to use the MODULATION key..........................................................................16
How to use the BYPASS key....................................................................................17
Session Time............................................................................................................18
Patient Pad Kit .......................................................................................................................19
Setting up the Patient............................................................................................................20
Options................................................................................................................................... 21
Product information
Maintenance.............................................................................................................22
Troubleshooting........................................................................................................ 22
Cleaning ...................................................................................................................23
Technical specifi cations............................................................................................23
Symbols used...........................................................................................................24
Warranty...................................................................................................................24
Maintenance sheet...................................................................................................25
Safety Recommendations
• The physician defi nes the protocol and ensures that it is correctly implemented (adjustments,
session time and frequency of use).
• We recommend that you lock the hand control when you give it to the patient.
• For optimum safety, always give the hand control to the patient before starting the system.
• Explosion hazard: KINETEC Spectra is not designed for use in the presence of fl ammable
anaesthetics.
• Check that the electrical socket is in good condition and is suitable for the splint power supply cord.
The latter complies with current standards and has a grounding socket. The plug may be connected
to any standard socket. The socket must however have a grounded pin. To connect the power
supply, only use the original cable supplied with the machine. Check that the cables remain free
around the device so that they do not get damaged.

2
T08 9351 7500 E[email protected] Wwww.unicarehealth.com.au
Defi nition
The KINETEC Spectra is a Knee PASSIVE mobilization device enabling the extension and fl exion
movement from -10° to 120°.
Indications
• Knee replacement surgery.
• Fractures (patellar, tibia plateau, femoral,...).
• Arthrolysis
• Hip surgery, including hip replacement, hip pinning, osteotomy,...).
• Ligament repairs.
• Arthroscopic surgery (menisectomies, patellectomies,...)
• Burns, joints sepsis,...
Clinical Benefi ts
• Breaks the cycle of trauma, infl ammation and the loss of range of motion.
• Prevents joint stiffness.
• Speeds the reco very of post-operative range of motion.
• Maintains the quality of the joint surface.
• Reduces pain and edema.
• Promotes joint cartilage healing.
• Reduces hospitalization time
• Reduces the need for pain medication.
• Provides immediate post-operative continuous passive motion.
• Digital ROM readout on the patient hand control for positive reinforcement.
• Maintains desired positions for stretching and muscular rest.
Contraindications
Bone Cancer, Warped joint surfaces, Spastic paralyses, Unstable fractures, Uncontrolled infection.
The machine are not adapted for patients height more 2 m(6’7”) or under 1,40m (4’7”).

3
Kinetic Spectra, Instruction & Safety Manual
T08 9351 7500 E[email protected] Wwww.unicarehealth.com.au
Description
KINETEC Spectra consists of the following components:
1 Lower limb support.
2 high support.
3 Foot support and hand control location for transport.
4 Hand control.
5 Thigh support setting lock.
6 Lower limb support setting lock .
7 Foot support positioning setting lock.
8 Transport handle.
9 ON/OFF switch and fuses.
10 Liquid-crystal display (2 lines of 16 characters).
11 SESSION TIME display key.
12 BYPASS mode key.
13 MODULATION key.
14 WARM UP key.
15 EXTENSION setting key.
16 FLEXION setting key.
17 Increase / decrease keys.
18 START key.
19 STOP key.
20 FORCE key.
21 SPEED key.
22 PAUSE key.
23 TIMER key.
24 PROGRAM access key.

4
T08 9351 7500 E[email protected] Wwww.unicarehealth.com.au
Safety
The physician defi nes the protocol and ensures that it is correctly implemented
(adjustments, session duration and frequency of use).
The patient must know the start/stop/reverse function on the control handle. Hand control must be
accessible to patient at
all times. (See page 7).
KINETEC Spectra complies with Directive 93/42/CEE.
EXPLOSION HAZARD:
KINETEC Spectra is not designed for use in the presence of fl ammable anesthetics.
In case of electromagnetic interference with other devices move the device.
KINETEC Spectra is in compliance with standards in force (IEC 601.1.2), electromagnetic compatibility
standard for medical
devices.

5
Kinetic Spectra, Instruction & Safety Manual
T08 9351 7500 E[email protected] Wwww.unicarehealth.com.au
Electrical Connection
KINETEC Spectra is a type B class I device.
Before connecting the device to the power supply, check
that the mains voltage
matches that shown on the identifi cation plate (100-240 V~
50-60Hz).
Connect the hand control (4)
Connect the power supply cable (25).
IMPORTANT
Check that the electrical socket is in good condition
and is suitable for the splint power supply cord.
The latter complies with current standards and
has a grounding socket.
The plug may be connected to any standard socket.
The socket must however have a grounded pin.
To connect the power supply, only use the original
cable supplied with the machine.
Check that the cables remain free around the device
so that they do not get damaged.
Starting the Unit
Switch on (9).
While the unit begins an auto diagnostic, the display shows the following:
Your KINETEC Spectra is ready to be used.
KINETIC Please Kinetic
Spectra V01.0 Wait 30 STOP 15 90

6
T08 9351 7500 E[email protected] Wwww.unicarehealth.com.au
Use of the Hand Control
Select your language
Begining Keys to press Display Remarks
Switch the Unit ON KINETIC
Spectra V01.0 Check if the hand
Please wait control is not blocked
Kinetic (See page 7)
30 STOP 15 90
Press the 2 keys in Language: The display indicates
the same time English the language selected
To change the Language: The English language
language English is selected
English
French
German
Italian
Spanish
To validate the new OK To exit and confi rm
language Switch on/off the new language,
switch OFF and Switch
IN the unit

7
Kinetic Spectra, Instruction & Safety Manual
T08 9351 7500 E[email protected] Wwww.unicarehealth.com.au
Use of the Hand Control
Locking the hand control setting
The hand control allows the patient to control the machine as appropriate.
Simultaneously press the and keys to lock the hand control,
The display reads “LOCK”, you can not change the parameters, if you try the display reads “LOCK”.
To unlock the hand control, simultaneously press the and keys, The display reads
“UNLOCK”.
We recommend that you lock the hand control when you give it to the patient.
Comment: The hand control is unlocked when you switch the unit ON/OFF.
START/STOP/REVERSE function
As with all KINETEC systems, KINETEC Spectra is equipped with a START/STOP/REVERSE
function.
When the unit is running, the display reads RUN
Press the key of the hand control. The movement stops. The display reads STOP.
Press the key of the hand control. The movement starts in the opposite direction and the display
reads RUN.
Caution:
For optimum safety, always give the hand control to the patient before starting the system.

8
T08 9351 7500 E[email protected] Wwww.unicarehealth.com.au
Use of the Hand Control
Quick Start
Begining Keys to press Display Remarks
Switch the Unit ON KINETIC
Spectra V01.0 Check if the hand
Please wait control is not blocked
Kinetic (See page 7)
30 STOP 15 90
or If the warm up mode
WARMUP is selected, switch OFF
30 STOP 15 90 this mode by pressing
on the key
Start the session with WARMUP The value change
the original 30 RUN 45 90 at the speed of the
Parameters of the movement
movement (default
setting)
Possible values for each parameter:
Possible values Default setting
Treatement Mode Normal
Extension limit -10 to 115° 30°
Flexion limit -5° to 120° 90°
Speed 1 to 5 (from 30° to 160° per minute) 2
Force 1 to 6 6
Extension pause 0 to 900 seconds (15minutes) 0
Flexion pause 0 to 900 seconds (15minutes) 0
Timer No timer(00H00) to 24H00 0

9
Kinetic Spectra, Instruction & Safety Manual
T08 9351 7500 E[email protected] Wwww.unicarehealth.com.au
Use of the Hand Control
How to adjust the basic parameters of the movement
Begining Keys to press Display Remarks
To stop the unit KINETIC Check if the hand
30 STOP 15 90 control is not blocked
(See page 7)
KINETIC
To display the extension 30 Ext 15 90
or fl exion of the KINETIC The value blinks
movement 40 Flex 15 90
To change the limit KINETIC The new value blinks
if necessary 40 Flex 15 110
To validate the new KINETIC While the value
value, press another SPEED 2 blinks press
key or wait more than FORCE the
4 seconds No timer to change if
necessary
Or to display pause in KINETIC Successive presses
extension or fl exion limit PAUSE HIGH 0 on this key selects
PAUSE LOW 0 the pause at the
extension or fl exion
limit
To change the pause KINETIC The new pause
if necessary PAUSE low 15 value blinks
To validate the new value KINETIC The unit is ready to
press another key or 40 Flex 15 110 start with the new
wait more than 3 seconds paremeters
The display shows the
selected mode
Or
Or
Or Key

10
T08 9351 7500 E[email protected] Wwww.unicarehealth.com.au
Use of the Hand Control
How to modify programs PROGRAM MODE
Begining Keys to press Display Remarks
To switch off the unit Check if the hand
control is not blocked
(See page 7)
Press the two keys at KINETIC Welcome text during
the same time and Spectra V01.0 3 seconds
switch the unit ON
Then Program 2 The program number
EMPTY blinks
To change the program Program 10 The new program
if necessary 25 kinetic 110 number blinks
To choose the Program 10 The display indicates
treatment mode 25 warmup 110 the selected treatment
mode, the program
number blinks again
Or Program 10
25 kinetic 110
To display the extension Program 10 The value blinks
or fl exion limit of the 30 warmup 100
movement Program 10
30 warmup 110
Or
Or

11
Kinetic Spectra, Instruction & Safety Manual
T08 9351 7500 E[email protected] Wwww.unicarehealth.com.au
Use of the Hand Control
How to modify programs PROGRAM MODE (continued)
Begining Keys to press Display Remarks
To change the limit PROGRAM 10 The new value
of movement 30 warmup 120 blinks
To validate the new PROGRAM 10 While the value blinks
value press another SPEED: 2 press the
key or wait more FORCE
No timer to change if
Pause low 0 necessary
To record the program 10 PROGRAM 10
Save: + clear:-
Then PROGRAM 10 The program 10 has
Save been recorded and
PROGRAM 11 the display indicates
EMPTY the next program
so you can change
another program
Or PROGRAM 10 The program 10 has
to cancel the program clearing been cancelled and
PROGRAM 11 the display indicates
EMPTY the next program
so you can change
another program
To exit program mode, KINETIC To use the modifi ed
switch OFF and switch Spectra C01.0 program see page 10
ON the unit
Or
Or

12
T08 9351 7500 E[email protected] Wwww.unicarehealth.com.au
Use of the Hand Control
Comments:
When a program has been deleted, the display shows
Program table:
Program Treatment Flexion Extension Speed Force Flexion Extension Timer
Number Mode Limit Limit Pause Pause
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16

13
Kinetic Spectra, Instruction & Safety Manual
T08 9351 7500 E[email protected] Wwww.unicarehealth.com.au
Use of the Hand Control
Using Programs
The KINETEC Spectra allows you to store up to 16 programs, including the type of treatment, ROM,
speed, load, pauses and timer.
The original parameter values of the program are empty.
These values can be modifi ed and recorded at any time (see ‘How to enter a program’ page 10).
Begining Keys to press Display Remarks
To stop the unit PROGRAM 10 Check of the hand
40 STOP 15 100 control is not blocked
(See page 7)
To access the program PROGRAM 1 The program number
mode 70 WARMUP 10 blinks
To change the program PROGRAM 3 The new program
if necessary 25 WARMUP 90 number blinks
To exit and validate KINETIC The current
the selected program Please Wait parameters have
been recorded in
program 3
To exit without KINETIC Back to the starting
validation of selected 40 STOP 15 110 parameters
program
Start the unit warning The value change at
25 run 20 90 the speed of the
movement
Comments:
• The values shown in the ‘Display’ column are examples. They actually depend on the stored
programs.
• The current movement parameters can be changed while using that program but no data will be
stored in the original program.
See the programming mode (p 10) to modify programs.
Or

14
T08 9351 7500 E[email protected] Wwww.unicarehealth.com.au
Use of the Hand Control
Reading the values of a program: example SPEED
Begining Keys to press Display Remarks
To stop the unit PROGRAM 10 Check of the hand
40 STOP 15 100 control is not blocked
(See page 7)
To access the program PROGRAM 1 The program number
mode 70 WARMUP 10 blinks
To change the program PROGRAM 3 The new program
if necessary 25 WARMUP 90 number blinks
To read the speed PROGRAM 3 Displays the speed
value SPEED 5 value
After 15 seconds or PROGRAM 3
after pressing on 70 WARMUP 110
another key
To exit and validate KINETIC The current
the selected program 40 STOP 15 110 parameters have
been recorded in
program 3
Start the unit KINETIC The value change at
40 RUN 15 110 the speed of the
movement
Or

15
Kinetic Spectra, Instruction & Safety Manual
T08 9351 7500 E[email protected] Wwww.unicarehealth.com.au
Use of the Hand Control
How to use the WARM UP key
Begining Keys to press Display Remarks
To stop the unit KINETIC Check of the hand
40 STOP 15 100 control is not blocked
(See page 7)
To select the warmup WARM UP The display shows the
mode 40 STOP 15 90 default settings
(see page 8)
To change the
movement value if
necessary
(See page 9)
Start the movement WARM UP The value change
15 RUN 15 110 at the speed of the
movement
Warm Up Rules:
The Kinetec Spectra starts at 70% of the full ROM, increasing 5% of the range
every other cycle until the pre-set ROM is reached.
Comments:
The calculation mode used enables reaching the pre-set ROM in approximately
seven full cycles.
Example:
For a Warm Up treatment with a pre-set ROM from 0° to 100°.
The fi rst cycle starts from 15° to 85° to 15° and increase 5% each cycle.
Cycle 1 (70%)
Cycle 2 (75%)
Cycle 3 (80%)
Cycle 4 (85%)
Cycle 5 (90%)
Cycle 6 (95%)
Cycle 7 (100%)
Pre-Set ROM
50°
100°
0°

16
T08 9351 7500 E[email protected] Wwww.unicarehealth.com.au
Use of the Hand Control
How to defi ne the patient tolerance • At the start of a session
This function, MODULATION Mode is a way to set within the tolerance of a patient at the beginning of
a session.
Set up the patient and proceed as below:
Begining Keys to press Display Remarks
To stop the unit KINETIC Check of the hand
Spectra V01.0 control is not blocked
(See page 7)
Please Wait
Kinetic If the Warm Up is
30 STOP 15 90 selected, switch OFF
WARM UP this mode by pressing
30 STOP 15 90 on the key
To select the MODUL.: use + or - The display includes
MODULATION Mode 30 STOP 15 90 the keys to be used
to deive the machine,
this message is
displayed 3 seconds
To select the pain level MODUL.: use + or - The unit is moving to
30manual 50 50 the selected way. It is
waiting for you to
select new limits
To set the pain level MODUL.: use + or - The new limit of the
when reached, 30manual 50 50 movement is recorded
immediately press
To start the session KINETIC The angle display
with the new 30 run 45 50 changes with current
movement limits movement
Or
Or

17
Kinetic Spectra, Instruction & Safety Manual
T08 9351 7500 E[email protected] Wwww.unicarehealth.com.au
Use of the Hand Control
How to defi ne the patient tolerance • During the session
Begining Keys to press Display Remarks
The unit is running KINETIC Check of the hand
30 RUN 15 90 control is not blocked
Or (See page 7)
WARMUP
30 RUN 15 90
To select the BYPASS BYPASS: use + or - The display indicates
mode 30 RUN 15 90 the keys used to run
the machine.
This message is
displayed 3 seconds
To select the NEW BYPASS: use + or - The unit is moving
pain level 30bypass 100 90 to the selected way.
It is waiting for you to
select new limits
To set the NEW pain bypass.: use + or - The new limit of the
level when reached, 30 manual 50 50 movement is recorded
immediately press
Continue the session WARMUP The angle display
with the new movement 30 RUN 45 50 changes with current
limits movement
Or
Or

18
T08 9351 7500 E[email protected] Wwww.unicarehealth.com.au
Use of the Hand Control
Session Time
This function shows the running time (in minutes) of the session (motor functioning).
• It directly accessible by the key, the display shows
This counter is reset each time the unit is switched ON.
TIME
30 STOP
02H25
15 90

19
Kinetic Spectra, Instruction & Safety Manual
T08 9351 7500 E[email protected] Wwww.unicarehealth.com.au
Use of the Kinetec Patient Pad Kit
How to defi ne the patient tolerance • During the session
FOR OPTIIMAL HYGIIENE,, A NEW SET OF PADS SHOULD BE USED FOR EACH PATIIENT..
Eacch ccovverr iiss prrovviided wiitth a llabell tto rreccorrd tthe pattiientt’’ss name
CLEANING:
- Sterilization of the pads ( if necessary): Sterilized at 134 °C during 18 minutes.
- Disinfecting of the pads: Washing at 30°C with use of a disinfecting solution during the rising cycle
Example of product that can be used: Solution “ Baclinge” at 0.125 % or “Souplanios” at 0,125%
from ANIOS Laboratory.
A complete list of distributors in your country is available on request.
The KINETEC Spectra is delivered with a complete set. Components:
- 4 straps
- 1 foot support
- 1 cover