SOUND LEVEL METER
Sound Level Meter Model UT351 and UT352 (hereafter
referred to as “the Meter”) is a stable, safe and reliable
sound level meter. The Meter is suitable to use in noise
control, quality control, health care and all different kind of
environmental noise testing. For example: factory, road,
family, musical instrument and all kind of places which
need noise testing.
Unpacking Inspection
Open the package case and take out the Meter. Check
the following items carefully to see any missing or
damaged part:
Item Description Qty
1 English Operating Manual 1 piece
2 Windscreen 1 piece
3 1.5V Batteries (AA) 4 pieces
In the event you find any missing or damage, please
contact your dealer immediately.
Rules For Safe Operation
Soft cloth and mild detergent should be used to clean
the surface of the Meter when servicing. No abrasive
and solvent should be used to prevent the surface of the
Meter from corrosion, damage and accident.
Constantly check the battery as it may leak when it has
been using for some time, replace the battery as soon
as leaking appears. A leaking battery will damage the
Meter.
International Electrical Symbols
Conforms to Standards of European Union
The Meter Structure (see figure 1)
figure 1
A. Meter Front
1. Housing
2. Windscreen
3. Microphone
4. LCD Display
5. Functional Buttons
6. Signal output and power terminals
figure 2
Side Panel (see figure 3)
figure 3
DC Output Terminal
AC Output Terminal
CAL potentiometer
External DC6V
No. Meaning
1 Data Store is full
2 Date and Time display
3 Data Store
4 Decibel
5 C-Weighting
6 A-Weighting
7 Sound value display
8 Range display
9 Over range
10 Slow response
11 Fast response
12 Low battery display
13 Data Hold is on
14 Auto power off enabled
15 Under range
16 Analogue bar graph display
17 Symbol of Sound Pressure Level
18 Auto ranging enabled
19 Date display
20 Time display
21 Maximum value display
22 Minimum value display
23 Data Store enabled
Measurement Operation and Functional
Buttons
Below table indicated for information about the functional
button operations
Button Operation Performed
HOLD
Turn the Meter on and off. Press once
to turn the Meter on. Press and hold for
around 1 second to turn the Meter off.
HOLD feature: During sound measure
ment, press once to freeze the current
reading in the display. Press the button
again to resume normal operation.
A/C During sound measurement, press
A/C button to select “A” or “C” frequency
weighting. With “A” weighting selected,
the frequency response of the Meter is
similar to the response of the human ear.
“A” weighting is commonly used for envi
ronmental or hearing conservation
programs ‘C” Weighting is a much flatter
response and is suitable for the sound
level analysis of machines, engines, etc.
Most noise measurements are performed
using “A” Weighting and SLOW
Response.
For the Model UT352, press A/C button
also could recall and clear data:
RECALL:
Press and hold this button, the LCD
displays the last data stored and the
index number.
Press button to recall additional
stored reading.
Press button to exit the RECALL
mode
CLR:
Press and hold A/C button when turning
on the Meter until the LCD displays CLR
and RECORD. All the data stored will
be cleared.
OPERATING MANUAL
Before using the Meter inspect the case and
accessories. Do not use the Meter if it is damaged, LCD
cannot display, the case (or part of the case) is removed
or you consider the Meter does not work properly. Look
for cracks or missing plastic.
When using the Meter, must follow the instruction
manual.
The internal circuit of the Meter shall not be altered at
will to avoid damage of the Meter and any accident.
Replace the battery as soon as the battery indicator
appears.
Turn the Meter power off when it is not in use and take
out the battery when not using for a long time.
Do not use or store the Meter in an environment of
high temperature, humidity, explosive, inflammable and
strong magnetic field. The performance of the Meter
may deteriorate after dampened.
B.Display Symbols (see figure 2)
1. DC Output Terminal: DC analogue signal output.
Output impedance is around 100Ω (10mV/dB)
2. AC Output Terminal: AC analogue signal output.
Output impedance is around 600Ω (0.707V/ each
range scale)
3. CAL potentiometer: Calibration
4. External DC6V:Using power adaptor DC6V, output
plug ( 3.5) to plug in the terminal. It can use 4pcs
of 1.5V batteries or power adaptor to power up the
Meter

9 108 11
1312
Button Operation Performed
LEVEL Press to selection of Auto ranging, Manual
ranging.
The Meter is default to auto ranging.
Press LEVEL button to switch to manual
ranging. Press button to toggle from
low to high range or from high to low
range. Press and hold LEVEL button to
exit the manual ranging mode.
FAST/
SLOW
Press to select a FAST (125ms) or a
LOW (1 second) response time. Select
FAST to capture noise peaks and noises
that occur very quickly. Select the LOW
response to monitor a sound source that
has a consistent noise level or to aver
age quickly changing levels. Select LOW
response for most application.
Press and hold FAST/SLOW button to
enable display backlight. Press and hold
FAST/SLOW button again to disable
display backlight.
For UT352, press FAST/SLOW button to
store data:
Press HOLD button to freeze data, the
LCD displays HOLD and M symbol and
the data being stored.
Press FAST/SLOW to store the data,
the symbol RECORD and the number
of index blink for 0.5 seconds.
The Meter will automatically exit HOLD
mode.
The Meter can store up to 63 data.
MAX/
MIN
Press MAX/MIN to display the maxi
mum or minimum reading. The display
will update only when the measured value
exceeds the value presently in the display.
Press the MAX/MIN button and the MAX
icon will display on the reading. The reading
displayed is the highest reading encountered
since the MAX mode was entered.
Press and hold MAX/MIN button to
disable or enable auto power off feature.
The Meter will automatically off after
approximately 15 minutes of inactivity.
The icon display indicates that the
auto power off feature is active.
Button Operation Performed
Calibration (see figure 4)
1. Turn the Meter on.
2. Put the Meter in the “A” weighting mode, FAST
response mode, range set to 60~110dB, lock to MAX.
3. Place the microphone onto the calibrator’s 1/2 inches
sound source hole.
4. Turn the calibrator on, using 94dB@1kHz standard
sound source.
5. Adjust the Meter’s CAL potentiometer located on the
side panel until the LCD displays 94.0dB
figure 4
Specifications
A.General Specifications
Display: 3 1/2 digits, 1999 maximum
Overloading:
Under range displays UNDER
Over range displays OVER
Battery Deficiency:
Change batteries as soon as is displayed.
Sampling Rate:
Fast Speed: 125 microseconds
Slow Speed: 1 second
Microphone: 1/2” electret condenser
Drop Test: 1 meter pass
Battery: 4 x 1.5V batteries (AA)
Battery Life: Typical 20 hours continuous
Dimension: 273 x 69 x 39mm
Weight: around 386 g (including batteries)
B.Environmental Requirements
For indoor use only.
Altitude: 2000m
Temperature and humidity:
Operating:
0 ~30 (≤80%R.H)
30 ~40 (≤75%R.H)
40 ~50 (≤45%R.H)
Storage:
-20 ~ +60 (≤80%R.H)
Safety/ Compliances:
EN61326:1997+A1:1998+A2:2001+A3:2003,
EN61672-1: 2002 Class 2 and IEC60641:1979 Type 2,
ANSI S1.4:1983 Type 2
Certification:
Accuracy Specifications
Accuracy:±(a% reading + b digits).guarantee for 1 year.
Operating temperature:23 ±5 .
Relative humidity:≤80%.
Temperature coefficient: 0.1 x (specified accuracy) / 1
A. Decibel dB Maintenance
This section provides basic maintenance information
including battery replacement instruction.
Warning
Do not attempt to repair or service your Meter unless
you are qualified to do so and have the relevant
calibration, performance test, and service information.
To avoid electrical shock or damage to the Meter, do
not get water inside the case.
In order not to affect the Meter accuracy or damge to
the Meter, do not open the Meter housing
A.General Service
Periodically wipe the case with damp cloth and mild de
tergent. Do not use chemical solvent.
To clean the terminals with cotton-tipped swab with de
tergent, as dirt or moisture in the terminals can affect
readings.
Press the Meter power off when it is not in use and take
out the battery when not using for a long time.
Do not store the Meter in place of humidity, high tem
perature, explosive, inflammable and strong magnetic
field
B.Replacing the Battery (see figure 5)
figure 5
Warning
To avoid false readings replace the battery as
soon as the battery indicator appears.
To replace battery:
1.Press the Meter power off
2.Remove the screw from the battery compartment, and
then take out the battery door from the battery compart
ment.
3.Remove the battery from the battery compartment.
4.Replace the battery with 4pcs new 1.5V AA batteries
5.Rejoin the battery door and the battery compartment,
and install the screw
Copyright 2008 Uni-Trend Group Limited.
All rights reserved.
Manufacturer:
Uni-Trend Technology (Dongguan) Limited
Dong Fang Da Dao
Bei Shan Dong Fang Industrial Development District
Hu Men Town, Dongguan City
Guang Dong Province
China
Postal Code: 523 925
Headquarters:
Uni-Trend Group Limited
Rm901, 9/F, Nanyang Plaza
57 Hung To Road
Kwun Tong
Kowloon, Hong Kong
Tel: (852) 2950 9168
Fax: (852) 2950 9303
Email: [email protected]
http://www.uni-trend.com
14
MAX/
MIN
Function Range Resolution Accuracy Remarks
Sampling Time: 125 microsecond
Sampling Time: 1 second
1 dB per scale,
Sampling time: 200 times per second
Over range display: OVER
Under range display: UNDER
Has input terminal
Has input terminal
Turn on and off the Meter and data
holding
Selecting auto and manual ranging
RECALL/CLR features are for model
UT352 only. Press to recall and clear
data
STORE feature is for model UT352
only. Press to store data
30~80dB
50~100dB
60~110dB
80~130dB
FAST
SLOW
30~130dB
Output impedance around
100Ω, 10mV/dB
Output impedance around
600Ω, 0.707V/ each scale
A-Weighting and
C-Weighting selection
Selection of fast or slow
sampling rate and turn on
and off of display backlight
Selection of maximum and
minimum value.
Selection of auto power on
and off
A-Weighting and
C-Weighting
Sampling Rate
Analogue Bar Graph
Overloading
DC analogue signal
output
AC analogue signal
output
Power (HOLD)
LEVEL (AUTO)
A/C (RECALL/CLR)
FAST/SLOW
(STORE/BL)
MAX/MIN (Auto
ranging)
0.1dB ±1.5dB
1dB
Frequency Response
31.5~8kHz
Press MAX/MIN button again. The MIN
icon will appear on the display. The read
ing displayed is the lowest reading
encountered since the MIN mode was
entered.
Press the MAX/MIN button again to exit
the MAX/MIN display mode