Vaisala HUMICAP HMT140 User manual

M211488EN-G
User Guide
Vaisala Wi-Fi Data Logger
HMT140

PUBLISHED BY
Vaisala Oyj
Vanha Nurmijärventie 21, FI-01670 Vantaa, Finland
P.O. Box 26, FI-00421 Helsinki, Finland
+358 9 8949 1
Visit our Internet pages at www.vaisala.com.
© Vaisala Oyj 2019
No part of this document may be
reproduced, published or publicly
displayed in any form or by any
means, electronic or mechanical
(including photocopying), nor
may its contents be modified,
translated, adapted, sold or
disclosed to a third party without
prior written permission of the
copyright holder. Translated
documents and translated
portions of multilingual
documents are based on the
original English versions. In
ambiguous cases, the English
versions are applicable, not the
translations.
The contents of this document are
subject to change without prior
notice.
Local rules and regulations may
vary and they shall take
precedence over the information
contained in this document.
Vaisala makes no representations
on this document’s compliance
with the local rules and
regulations applicable at any
given time, and hereby disclaims
any and all responsibilities related
thereto.
This document does not create
any legally binding obligations for
Vaisala towards customers or end
users. All legally binding
obligations and agreements are
included exclusively in the
applicable supply contract or the
General Conditions of Sale and
General Conditions of Service of
Vaisala.
This product contains software
developed by Vaisala or third
parties. Use of the software is
governed by license terms and
conditions included in the
applicable supply contract or, in
the absence of separate license
terms and conditions, by the
General License Conditions of
Vaisala Group.
This product may contain open
source software (OSS)
components. In the event this
product contains OSS
components, then such OSS is
governed by the terms and
conditions of the applicable OSS
licenses, and you are bound by
the terms and conditions of such
licenses in connection with your
use and distribution of the OSS in
this product. Applicable OSS
licenses are included in the
product itself or provided to you
on any other applicable media,
depending on each individual
product and the product items
delivered to you.

Table of Contents
1. About This Document................................................................................... 7
1.1 Version Information.......................................................................................... 7
1.2 Related Manuals................................................................................................ 7
1.3 Documentation Conventions...........................................................................7
1.4 Trademarks........................................................................................................ 8
2. Product Overview........................................................................................... 9
2.1 Overview of HMT140 Features........................................................................ 9
2.1.1 Data Management..................................................................................... 9
2.1.2 Models.........................................................................................................9
2.1.3 Options and Accessories.........................................................................10
2.1.4 Fixed or Remote Probe........................................................................... 10
2.1.5 Interchangeable Probe............................................................................10
2.1.6 Optional Display........................................................................................ 11
2.2 HMT140 Components......................................................................................12
2.2.1 Installation Kit...........................................................................................12
2.2.2 HMT140 Utility and Configuration Cable............................................... 13
2.2.3 Connections and Wiring..........................................................................14
2.3 Safety................................................................................................................16
2.3.1 ESD Protection..........................................................................................17
2.4 Regulatory Compliances.................................................................................17
2.4.1 FCC Compliance Statement....................................................................18
2.4.2 ISED Compliance Statement...................................................................18
3. Installation........................................................................................................ 19
3.1 Mounting...........................................................................................................19
3.1.1 Opening Data Logger Cover...................................................................19
3.1.2 Wall Mounting......................................................................................... 20
3.1.3 Duct Installation.......................................................................................22
3.1.4 Probe Assembly with Duct Installation Kit...........................................22
3.2 Optional Mounting Accessories....................................................................24
3.2.1 Probe Mounting Flange..........................................................................24
3.2.2 Probe Mounting Clamp...........................................................................24
4. Set Up and Operation................................................................................. 25
4.1 Connect and Configure..................................................................................25
4.1.1 Installing HMT140 Utility Sofware......................................................... 25
4.1.2 Setting up HMT140 Hardware................................................................25
4.1.3 Configuring the HMT140 to Your Network...........................................26
4.1.4 Setup Consideration............................................................................... 28
4.2 Custom Settings............................................................................................. 28
4.2.1 Modify Transmission Settings................................................................ 28
4.2.2 Modify HMT140 Description....................................................................31
4.2.3 Check Transmission Status......................................................................31
Table of Contents
1

4.3 HMT140 Function Modes............................................................................... 33
4.3.1 Operating Modes.....................................................................................33
4.3.2 Transmission of Data.............................................................................. 34
4.3.3 Transmit Attempts.................................................................................. 34
4.3.4 Alarm-triggered Transmission...............................................................34
4.4 Additional Settings.........................................................................................35
4.4.1 Change Input/Output Scale...................................................................35
4.4.2 Save and Recall Settings........................................................................ 36
4.4.3 Delete Saved Settings.............................................................................36
4.4.4 Restart the HMT140................................................................................ 36
4.4.5 Get General Information.........................................................................36
4.5 Battery Usage................................................................................................. 36
4.5.1 Viewing Battery Life Remaining............................................................36
4.5.2 Units and Temperature........................................................................... 37
4.5.3 Connection Performance....................................................................... 38
4.6 Advanced Options..........................................................................................39
4.6.1 Passwords................................................................................................40
4.6.2 Set Up a New Channel............................................................................. 41
4.6.3 Find Additional Networks...................................................................... 42
4.6.4 Test Network Connectivity.....................................................................43
4.7 Additional Program Parameters...................................................................43
5. Maintenance....................................................................................................45
5.1 Improving Sensor Response Time................................................................45
5.2 Removing and Fastening Probes..................................................................45
5.2.1 Fixed Probe Model.................................................................................. 45
5.2.2 Remote Probe Model..............................................................................46
5.3 Calibration and Adjustment.......................................................................... 47
5.4 Battery Replacement.....................................................................................48
6. Troubleshooting............................................................................................49
6.1 HMT140 Operation Mode.............................................................................. 49
6.2 Signal Strength...............................................................................................49
6.3 Indicator Lights...............................................................................................49
6.4 Access Points..................................................................................................49
6.5 Transmit Cycle................................................................................................ 50
6.6 Frequently Asked Questions........................................................................ 50
7. Technical Data................................................................................................54
7.1 HMT140 Dimensions (Fixed Probe)..............................................................54
7.2 HMT140 Dimensions (Remote Probe)..........................................................55
7.3 HMT140 Product Specifications....................................................................56
Index...................................................................................................................63
Warranty............................................................................................................65
Technical Support........................................................................................... 65
HMT140 User Guide M211488EN-G
2

List of Figures
Figure 1 HMT140 Features.............................................................................................. 12
Figure 2 HMT140 Components Diagram.....................................................................14
Figure 3 RTD Wiring Diagram........................................................................................ 15
Figure 4 RTD and Boolean Wiring Diagram............................................................... 15
Figure 5 Boolean Wiring Diagram.................................................................................15
Figure 6 Voltage Wiring Diagram................................................................................. 16
Figure 7 Current Wiring Diagram..................................................................................16
Figure 8 Opening Data Logger Cover..........................................................................19
Figure 9 Mounting Diagram............................................................................................21
Figure 10 Probe Installation with Duct Installation Kit.............................................22
Figure 11 Dimensions of HMP110 Probe with Duct Installation Kit....................... 23
Figure 12 Optional Probe Mounting Flange................................................................ 24
Figure 13 Optional Probe Mounting Clamp.................................................................24
Figure 14 Fixed Probe Model...........................................................................................46
Figure 15 Remote Probe Model...................................................................................... 47
Figure 16 HMT140 Fixed Probe Dimensions in mm (in)...........................................54
Figure 17 HMT140 Remote Probe Dimensions in mm (in)...................................... 55
HMT140 User Guide M211488EN-G
4

List of Tables
Table 1 Document Versions (English)............................................................................7
Table 2 Related Manuals....................................................................................................7
Table 3 Connect Time Components ............................................................................38
Table 4 HUMICAPâHumidity and Temperature Probe HMP110..........................56
Table 5 Temperature Probes..........................................................................................56
Table 6 Analog Inputs...................................................................................................... 57
Table 7 Measurement Variants...................................................................................... 57
Table 8 Operating Environment....................................................................................58
Table 9 General Specifications...................................................................................... 58
Table 10 Mechanical Specifications............................................................................... 58
Table 11 Wireless................................................................................................................ 59
Table 12 Standards............................................................................................................. 59
Table 13 Spare Parts and Accessories.......................................................................... 60
List of Tables
5

HMT140 User Guide M211488EN-G
6

1. About This Document
1.1 Version Information
This document provides information for installing, operating, and maintaining the Vaisala Wi-Fi
Data Logger HMT140. It also details the operation of the Vaisala HMT140 Utility software.
Table 1 Document Versions (English)
Document Code Date Description
M211488EN-G April 2019 New document template. Added chapters Regional Versions
(page 10), FCC Compliance Statement (page 18), and ISED
Compliance Statement (page 18). Updated power supply
specifications and chapter HMT140 Product Specifications
(page 56).
1.2 Related Manuals
Table 2 Related Manuals
Document Code Name
M211484EN Vaisala Wi-Fi Data Logger HMT140 Quick Guide
M211975EN Vaisala viewLinc Enterprise Server Version 5.0 User Guide
1.3 Documentation Conventions
Warning alerts you to a serious hazard. If you do not read and
follow instructions carefully at this point, there is a risk of injury or even death.
WARNING!
Caution warns you of a potential hazard. If you do not read and
follow instructions carefully at this point, the product could be damaged or
important data could be lost.
CAUTION!
Note highlights important information on using the product.
Chapter 1 – About This Document
7

Tip gives information for using the product more eciently.
Lists tools needed to perform the task.
Indicates that you need to take some notes during the task.
1.4 Trademarks
Vaisalaâand HUMICAPâare registered trademarks of Vaisala Oyj.
All other product or company names that may be mentioned in this publication are trade
names, trademarks, or registered trademarks of their respective owners.
HMT140 User Guide M211488EN-G
8

2. Product Overview
2.1 Overview of HMT140 Features
The Vaisala Wi-Fi Data Logger HMT140 measures relative humidity and temperature using the
connected probe and analog signals – RTD, Voltage, Current Loops and Boolean Contacts. The
HMT140 comes with or without an optional LCD display.
The HMT140 supports several dierent signal measurements (channels), which are monitored
in Vaisala viewLinc Enterprise Server software:
• Relative Humidity (RH)
• Temperature (T)
• Switch Contact
• Voltage and Current
Channel types are set at the factory during the time of order. Other wireless parameters can be
changed at a later date using an HMT140 Configuration Cable and Vaisala HMT140 Utility
software.
More Information
‣Options and Accessories (page 10)
2.1.1 Data Management
An HMT140 can store several thousands of records before needing to overwrite data. In the
event of a communication problem, the HMT140 will continue to log and store data until it has
collected 3060 records per channel and then it will begin overwriting data.
For example, if an HMT140 is configured with a 5-minute sample rate, it will continue to log
data for approximately 10 days before overwriting data.
2.1.2 Models
The following HMT140 models and options are available:
• HMT141: Vaisala HUMICAPâHumidity and Temperature Probe (RH/T)
• HMT143: Dual RTD Temperature Inputs w/5 m Probes (-90 to +90 °C (-130 to +194 °F))
• HMT144: Voltage Inputs (0 ... 5 VDC)
• HMT145: Voltage inputs (0 ... 10 VDC)
• HMT146: Current Inputs (0 ... 20 mA)
• HMT147: Boolean Inputs
• HMT148: RTD Temperature Input w/5 m Probe and Boolean Input
• HMT14C: RTD Inputs (customer-supplied Probe)
• HMT14D: RTD and Current Input
• HMT14E: RTD and Voltage Input (0 ... 5 VDC)
• HMT14F: RTD and Voltage Input (0 ... 10VDC)
Chapter 2 – Product Overview
9

2.1.2.1 Regional Versions
There are 2 regional versions of each HMT140 model, with dierent applicable standards and
regulations. The versions are distinguished by the letters A(for EU countries) and N(for non-
EU countries) on the HMT140 Order Form.
For more information, see Table 12 (page 59).
2.1.3 Options and Accessories
• Humidity and temperature probe HMP110
• Humidity and temperature replacement probe HMP110R (Requires a separate order form.
For more information, contact Vaisala.)
• Humidity sensor HUMICAP180R
• HMP110 Probe mounting flange 226061
• HMP110 Probe mounting clamps, 10 pcs 226067
• HMP110 Probe cable 3 m (9.8 ft) HMT120Z300
• HMP110 Probe cable 5 m (16.4 ft) HMT120Z500
• HMP110 Probe cable 10 m (32.8 ft) HMT120Z1000
• Duct installation kit 215619
• Probe mounting flange 226061
• Boolean Switch Kit with Door Contact 236319
• Thermal Dampener Blocks 236310SP
• Recommended battery: Xeno XL-060F
• Form Factor: AA
• Chemistry: Lithium-Thionyl chloride (Li-SOCl2)
• Nominal Voltage: 3.6V
• Nominal Capacity: 2400mAh
• Pulse Current Capability: 120mA
• Operating Temperature Range: -40 °C... +60 °C (-40 °F... +140 °F)
• Voltage @ 2mA & -20 °C (-4 °F): 2.6V
• Voltage @ 2mA & +60 °C (+140 °F): 2.6V
• HMP110 sensor protection:
• Plastic grid filter DRW010522SP
• Plastic grid with membrane filter DRW010525
• Stainless steel sintered filter HM46670SP
2.1.4 Fixed or Remote Probe
The HMT140 is available either with a fixed probe directly attached to the housing or a remote
probe with dierent cable lengths (3/5/10 m (9.8/16.4/32.8 ft)). Extension cables can be easily
cascaded in order to obtain longer reach, to a maximum length of 10 m (32.8 ft).
2.1.5 Interchangeable Probe
The HMP110 HUMICAPârelative humidity probe used in the HMT140 is fully interchangeable.
You can easily remove the probe and replace it with a new one without having to adjust the
HMT140. You have the following options when purchasing a new probe from Vaisala:
• Order a new probe and keep your current one.
HMT140 User Guide M211488EN-G
10

• Order a new probe and return the old one to Vaisala (replacement probe).
Only probes that have a compatible digital output (VDIGI mode) can be used with
the HMT140. Compatible probes have the letter "V" as the first letter in their order
code. The order code is written on the probe.
2.1.6 Optional Display
The HMT140 is also available with an optional graphical 128x64 pixel resolution LCD display.
The display shows the measurement results of selected parameters in selected units (defined
at the time of ordering). The parameters are displayed simultaneously as two separate rows on
the display.
Chapter 2 – Product Overview
11

2.2 HMT140 Components
The following figure illustrates the main features of the HMT140. On the left is a remote probe
model without display, and on the right is a fixed probe model with the optional display. The
numbers and arrows indicate the main components of the HMT140 models.
Figure 1 HMT140 Features
1HMT140 enclosure
2Cable bushing: cable gland
3Probe cable
4HMP110 probe
5Plastic grid filter
2.2.1 Installation Kit
The HMT140 is typically installed mounted on the wall with up to four screws (not included).
For duct mounting, Duct Installation Kits are available as accessories.
To help you position the screws correctly, use the Drilling Template, available inside the back
cover of the HMT140 Quick Guide.
HMT140 User Guide M211488EN-G
12

More Information
‣Options and Accessories (page 10)
‣Wall Mounting (page 20)
2.2.2 HMT140 Utility and Configuration Cable
To connect an HMT140 to your network, install the HMT140 Utility software and connect the
HMT140 to your PC with the HMT140 Configuration Cable (software and cable shipped with
viewLinc).
More Information
‣Configuring the HMT140 to Your Network (page 26)
Chapter 2 – Product Overview
13

2.2.3 Connections and Wiring
Figure 2 HMT140 Components Diagram
HMT140 User Guide M211488EN-G
14

Figure 3 RTD Wiring Diagram
Figure 4 RTD and Boolean Wiring Diagram
Figure 5 Boolean Wiring Diagram
Chapter 2 – Product Overview
15

Figure 6 Voltage Wiring Diagram
Figure 7 Current Wiring Diagram
2.3 Safety
This product has been tested for safety. Note the following precautions:
Remove batteries before shipping.WARNING!
Do not modify the unit or use it in ways not described in the
documentation. Improper modification may lead to safety hazards, equipment
damage, failure to perform according to specification, or decreased equipment
lifetime.
CAUTION!
HMT140 User Guide M211488EN-G
16

If the equipment is used in a manner not specified by Vaisala, the
protection provided by the equipment may be impaired.
WARNING!
This device requires a separation distance of at least 20 cm. This
distance must be maintained between the user and the device when the device
is operating.
CAUTION!
Cet appareil nécessite une distance de séparation d'au moins
20 cm. Cette distance doit être maintenue entre l'utilisateur et l'appareil lorsque
l'appareil est en fonctionnement.
CAUTION!
2.3.1 ESD Protection
Electrostatic Discharge (ESD) can damage electronic circuits. Vaisala products are adequately
protected against ESD for their intended use. However, it is possible to damage the product by
delivering electrostatic discharges when touching, removing, or inserting any objects in the
equipment housing.
To avoid delivering high static voltages to the product, handle ESD-sensitive components on a
properly grounded and protected ESD workbench or by grounding yourself to the equipment
chassis with a wrist strap and a resistive connection cord.
2.4 Regulatory Compliances
This product complies with the following standards:
• EN 61326-1B, EN 301 489-1 1)
• EN 300 328 V2.1.1
This product complies with the following safety standard:
• EN 61010-1
The HMT140 contains a radio module with the following module approvals:
• FCC ID: U3O-G2M5477
• IC ID: 8169A-G2M5477
• NCC ID: CCAF11LP0240T6
•
R
2
0
1
-
1
2
5
7
6
5
当該機器には電波法に基づく、技術基準適合証明等を受けた特定無線
設備を装着している。
1) EN 61326-1B and EN 301 489-1 apply only to the EU specific versions of HMT140 loggers with CE
marking.
Chapter 2 – Product Overview
17

2.4.1 FCC Compliance Statement
This equipment has been tested and found to comply with the limits for a Class B digital
device, pursuant to Part 15 of the FCC rules. These limits are designed to provide reasonable
protection against harmful interference in a residential installation. This equipment generates,
uses and can radiate radio frequency energy and, if not installed and used in accordance with
the instructions, may cause harmful interference to radio communications. However, there is
no guarantee that the interference will not occur in a particular installation. If this equipment
does cause harmful interference to radio or television reception, which can be determined by
turning the equipment o and on, the user is encouraged to try to correct the interference by
one or more of the following measures:
• Reorient or relocate the receiving antenna.
• Increase the separation between the equipment and receiver.
• Connect the equipment into an outlet on a circuit dierent from that of the receiver.
• Consult the dealer or an experienced radio/TV technician for help.
This device complies with part 15 of the FCC Rules. Operation is
subject to the following two conditions: (1) This device may not cause harmful
interference, and (2) this device must accept any interference received,
including interference that may cause undesired operation.
WARNING!
Changes or modifications to this equipment not expressly
approved by the party responsible for compliance could void the user’s
authority to operate the equipment.
WARNING!
2.4.2 ISED Compliance Statement
This device complies with Industry Canada licence-exempt RSS standard(s). Operation is
subject to the following two conditions:
1. This device may not cause interference; and
2. This device must accept any interference, including interference that may cause undesired
operation of the device.
This device has a whip antenna with a gain of 2 dBi.
Le présent appareil est conforme aux CNR d’lndustrie Canada applicables aux appareils radio
exempts de licence. L’exploitation est autorisée aux deux conditions suivantes:
1. l’appareil ne doit pas produire de brouillage, et
2. l’utilisateur de l’appareil doit accepter tout brouillage radioélectrique subi, même si le
brouillage est susceptible d’en compromettre le fonctionnement.
Le présent appareil dispose d'une antenne fouet avec un gain de 2 dBi.
HMT140 User Guide M211488EN-G
18
Other manuals for HUMICAP HMT140
4
Table of contents
Other Vaisala Data Logger manuals

Vaisala
Vaisala VaiNet RFL100 User manual

Vaisala
Vaisala VaiNet RFL100 User manual

Vaisala
Vaisala HUMICAP HMT140 User manual

Vaisala
Vaisala HUMICAP HMT140 User manual

Vaisala
Vaisala CWL100 User manual

Vaisala
Vaisala MRG112 Use and care manual

Vaisala
Vaisala VaiNet RFL100 User manual

Vaisala
Vaisala VaiNet RFL100 User manual