Viasys Healthcare LTV 1200 User manual

LTV®1200 Ventilator
Operator’s Manual
P/N 18247-001, Rev. B May 2006
www.ardusmedical.com

Operator’s Manual LTV®1200 Ventilator Page i
p/n 18247-001, Rev.B
Document Revision History
Document Revision Level Revision Date
Rev A, ECO 3891 April 2006
Rev B, ECO 4016 May 2006
Contact Information
Pulmonetic Systems, Inc.
17400 Medina Rd., Suite 100
Minneapolis, Minnesota 55447-1341
Phone: (763) 398-8300
Customer Care Center: (800) 754-1914
Fax: (763) 398-8400
Website: http://www.pulmonetic.com
LTV®1200 is a trademark belonging to Pulmonetic Systems, Inc.
Copyright © 2006 Pulmonetic Systems, Inc., Minneapolis, Minnesota.

Page ii LTV®1200 Ventilator Operator’s Manual
p/n 18247-001, Rev. B
Warranty
Pulmonetic Systems warrants that the LTV®1200 ventilator is free from defects in material and
workmanship for a period of one (1) year from the date of shipment, or 8,800 hours as measured on
the usage meter, whichever comes first, with the following limitations:
1) The internal battery is warranted for ninety (90) days from date of shipment.
Pulmonetic Systems will, at its option, either repair, replace, or issue credit for products that prove to
be defective during the warranty period.
For warranty service or repair, the product must be returned to Pulmonetic Systems or a service
facility designated by Pulmonetic Systems, shipping prepaid by the Buyer.
LIMITATION OF WARRANTY
Ordinary maintenance, as specified in the LTV®1200 Ventilator Operator’s and Service Manuals, is
not covered under the forgoing warranty.
The forgoing warranty does not apply to defects or damage to the unit resulting from:
•Improper use or misuse
•Improper or inadequate maintenance
•Unauthorized modifications or repairs
•Use of the unit with unauthorized accessories, e.g. external battery or AC adapter
•Use or storage outside the specified environment
NO IMPLIED WARRANTIES
This warranty is exclusive. There are no other warranties expressed or implied.
LIMITATION OF LIABILITY
Pulmonetic Systems shall not be liable for loss of profits, loss of use, consequential damages, or any
other claim based on breach of warranty. Pulmonetic Systems, Inc.’s liability for damages of any kind
shall be limited to the purchase price of the defective unit.

Operator’s Manual LTV®1200 Ventilator Page iii
p/n 18247-001, Rev.B
Notices
The LTV®1200 ventilator complies with limitations as specified in IEC 601-1-2 for Medical Products. It
does however, use and radiate radio frequency energy.
The function of this machine may be adversely affected by the operation of other nearby equipment,
such as high frequency surgical diathermy equipment, short-wave therapy equipment, defibrillators or
MRI equipment.
European Regulatory Requirements per 93/42/EEC Medical Device Directives
Pulmonetic Systems, Inc.’s European Representative for vigilance reporting within the European
Community is:
MediMark® Europe Sarl.
11, rue Emile Zola. BP 2332
F-38033 Grenoble Cedex 2. France
Tel: +33 (0)4 76 86 43 22
Fax: +33 (0)4 76 17 19 82
E-mail: info@medimark-europe.com
Any product malfunctioning issues that fall under Medical Device Directives Essential Requirements
should be directed to MediMark.
FM 75092
CE 02342

Page iv LTV®1200 Ventilator Operator’s Manual
p/n 18247-001, Rev. B
Notice To Operators
Unsafe Operation - Operating the LTV®1200 ventilator without a complete and thorough
understanding of its attributes is unsafe and may cause harm to the patient. It is important that this
manual be read and understood in its entirety before operating the ventilator.
Warnings and Cautions Section - Read the section on Warnings and Cautions carefully before
operating the LTV®1200 ventilator.
Use and Maintenance - Any questions regarding installing, operating, or maintaining the LTV®1200
ventilator, should be directed to a certified Pulmonetic Systems’ service technician or Pulmonetic
Systems, Inc.
Avis important
Fonctionnement dangereux - L'opération d'un ventilateur de la LTV® 1200 sans une excellente
compréhension de ses attributs est dangereuse et risque de blesser le patient. Il est très important de
lire et de comprendre entièrement ce manuel avant de faire fonctionner le ventilateur.
Section Avertissements et Attention - Lire attentivement la section Avertissements et Attention
avant de procéder à l'opération des ventilateurs de la LTV® 1200.
Utilisation et entretien - En cas de questions concernant l'installation, l'opération ou l'entretien des
ventilateurs de la série LTV®, veuillez vous adresser à un technicien de service certifié de Pulmonetic
Systems ou directement à Pulmonetic Systems, Inc.

Operator’s Manual LTV®1200 Ventilator Page v
p/n 18247-001, Rev.B
Contents
Warranty ............................................................................................................................................. ii
Notices................................................................................................................................................iii
Chapter 1 - Introduction .......................................................................................1-1
Operator’s Safety Information..........................................................................................................1-2
Warnings..........................................................................................................................................1-3
Cautions.........................................................................................................................................1-11
Symbols .........................................................................................................................................1-17
Chapter 2 - Ventilator Overview........................................................................... 2-1
Intended Use ...................................................................................................................................2-1
Power/Supplies Required ................................................................................................................2-2
Information/Assistance ....................................................................................................................2-3
Chapter 3 - Breath Types ..................................................................................... 3-1
Breath Types ...................................................................................................................................3-1
Volume Control Breaths ...............................................................................................................3-2
Pressure Control Breaths.............................................................................................................3-3
Pressure Support Breaths............................................................................................................3-5
Spontaneous Breaths...................................................................................................................3-6
Chapter 4 - Ventilation Modes..............................................................................4-1
Control Mode ...................................................................................................................................4-1
Assist/Control Mode.........................................................................................................................4-1
SIMV Mode ......................................................................................................................................4-2
CPAP Mode .....................................................................................................................................4-3
NPPV Mode .....................................................................................................................................4-4
Apnea Backup .................................................................................................................................4-5
Volume Pressure Ventilation ...........................................................................................................4-6
Bias Flow .........................................................................................................................................4-6
Chapter 5 - Using the Controls and Indicators................................................... 5-1
Ventilator Controls ...........................................................................................................................5-1
Setting a Control..............................................................................................................................5-2
Variable Controls .............................................................................................................................5-2
Buttons.............................................................................................................................................5-3
Set Value Knob................................................................................................................................5-3
Extended Features ..........................................................................................................................5-3
Bright, Dim and Blank Control Displays...........................................................................................5-4
Flashing Controls.............................................................................................................................5-5
Dashes.............................................................................................................................................5-5
Control Limiting................................................................................................................................5-5
Control Locking................................................................................................................................5-6
Control Retention.............................................................................................................................5-6

Page vi LTV®1200 Ventilator Operator’s Manual
p/n 18247-001, Rev. B
Chapter 6 - Controls..............................................................................................6-1
Assist/Control - SIMV/CPAP - NPPV .............................................................................................. 6-1
Breath Rate ..................................................................................................................................... 6-2
Control Lock .................................................................................................................................... 6-3
High Pressure Limit......................................................................................................................... 6-4
Inspiratory / Expiratory Hold............................................................................................................ 6-5
Inspiratory Hold............................................................................................................................ 6-6
Expiratory Hold ............................................................................................................................ 6-8
Inspiratory Time............................................................................................................................. 6-10
Low Minute Volume....................................................................................................................... 6-11
Low Pressure ................................................................................................................................ 6-12
Low Pressure O2Source............................................................................................................... 6-13
Manual Breath............................................................................................................................... 6-17
O2% (O2Flush)............................................................................................................................. 6-18
On/Standby ................................................................................................................................... 6-20
PEEP Control ................................................................................................................................ 6-21
PEEP Control in NPPV Mode.................................................................................................... 6-21
Pressure Control ........................................................................................................................... 6-22
Pressure Support .......................................................................................................................... 6-23
Pressure Support in NPPV Mode .............................................................................................. 6-23
Select ............................................................................................................................................ 6-24
Sensitivity ...................................................................................................................................... 6-25
Set Value Knob ............................................................................................................................. 6-26
Silence Reset ................................................................................................................................ 6-27
Tidal Volume ................................................................................................................................. 6-28
Volume Pressure Mode................................................................................................................. 6-29
Chapter 7 - Displays and Indicators ....................................................................7-1
Airway Pressure .............................................................................................................................. 7-1
Display Window............................................................................................................................... 7-1
Indicators......................................................................................................................................... 7-1
Battery Level ................................................................................................................................... 7-2
Charge Status ................................................................................................................................. 7-4
External Power................................................................................................................................ 7-5
NPPV............................................................................................................................................... 7-6
Patient Effort.................................................................................................................................... 7-6
Vent Inop......................................................................................................................................... 7-6
Chapter 8 - Monitored Data ..................................................................................8-1
Automatic or Manual Data Display Scrolling................................................................................... 8-2
PIP xxx cmH2O............................................................................................................................ 8-3
MAP xx cmH2O ........................................................................................................................... 8-3
PEEP xx cmH2O.......................................................................................................................... 8-3
f xxx bpm...................................................................................................................................... 8-3
Vte xxx ml .................................................................................................................................... 8-3
VE xx.x L...................................................................................................................................... 8-3

Operator’s Manual LTV®1200 Ventilator Page vii
p/n 18247-001, Rev.B
I:E xx:xx........................................................................................................................................8-3
I:Ecalc xx:xx .................................................................................................................................8-4
Vcalc xxx Lpm ..............................................................................................................................8-4
SBT xxx min .................................................................................................................................8-4
xxx f/Vt xx f ...................................................................................................................................8-4
Chapter 9 - Ventilator Alarms............................................................................... 9-1
Alarms..............................................................................................................................................9-2
APNEA, APNEA xx bpm ..............................................................................................................9-2
BAT EMPTY .................................................................................................................................9-3
BAT LOW .....................................................................................................................................9-5
DEFAULTS...................................................................................................................................9-7
DEFAULTS SET...........................................................................................................................9-9
DISC/SENSE..............................................................................................................................9-10
HIGH f.........................................................................................................................................9-11
HIGH O2PRES...........................................................................................................................9-12
HIGH PEEP................................................................................................................................9-13
HIGH PRES................................................................................................................................9-14
HW FAULT .................................................................................................................................9-16
INOP...........................................................................................................................................9-17
LOW MIN VOL ...........................................................................................................................9-18
LOW O2 PRES............................................................................................................................9-19
LOW PEEP.................................................................................................................................9-20
LOW PRES ................................................................................................................................9-21
NO CAL DATA, NO CAL Monitor Display..................................................................................9-22
POWER LOST ...........................................................................................................................9-23
POWER LOW.............................................................................................................................9-24
REMOVE PTNT .........................................................................................................................9-25
RESET........................................................................................................................................9-26
SBT < f .......................................................................................................................................9-27
SBT > f .......................................................................................................................................9-28
SBT < f/Vt ...................................................................................................................................9-29
SBT > f/Vt ...................................................................................................................................9-30
SBT OFF ....................................................................................................................................9-31
XDCR FAULT.............................................................................................................................9-32
Alarm Status Messages.................................................................................................................9-33
f PEEP OFF................................................................................................................................9-33
HI f/Vt OFF .................................................................................................................................9-33
HI PEEP OFF .............................................................................................................................9-33
H&L PEEP OFF..........................................................................................................................9-34
HIGH f OFF ................................................................................................................................9-34
LMV LPPS OFF..........................................................................................................................9-34
LMV OFF....................................................................................................................................9-35
LO PEEP OFF............................................................................................................................9-35
LOCKED.....................................................................................................................................9-35
LO f/Vt OFF ................................................................................................................................9-36

Page viii LTV®1200 Ventilator Operator’s Manual
p/n 18247-001, Rev. B
LPPS OFF.................................................................................................................................. 9-36
SBT f OFF.................................................................................................................................. 9-36
SBT f/Vt OFF ............................................................................................................................. 9-37
SBT HI f OFF ............................................................................................................................. 9-37
SBT LO f OFF............................................................................................................................ 9-37
SBT TIME .................................................................................................................................. 9-38
WARMUP xx.............................................................................................................................. 9-38
Chapter 10 - Extended Features.........................................................................10-1
Navigating the Extended Features Menus.................................................................................... 10-2
Alarm Operations .......................................................................................................................... 10-3
Alarm Volume ............................................................................................................................ 10-3
Apnea Interval............................................................................................................................ 10-4
High Pressure Alarm Delay .......................................................................................................10-4
Low Peak Pressure Alarm .........................................................................................................10-4
High f.......................................................................................................................................... 10-5
High PEEP................................................................................................................................. 10-5
Low PEEP.................................................................................................................................. 10-5
Patient Assist ............................................................................................................................. 10-6
Exit ............................................................................................................................................. 10-6
Vent Operations ............................................................................................................................ 10-7
Variable Rise Time..................................................................................................................... 10-8
Variable Flow Termination ......................................................................................................... 10-9
Variable Time Termination.......................................................................................................10-10
Pressure Control Flow Termination ......................................................................................... 10-11
Leak Compensation................................................................................................................. 10-12
O2Flush ................................................................................................................................... 10-13
Control Unlock ......................................................................................................................... 10-14
Language Selection ................................................................................................................. 10-14
Software Version...................................................................................................................... 10-15
Usage Meter ............................................................................................................................ 10-15
Communications Setting..........................................................................................................10-15
Set Date ................................................................................................................................... 10-16
Set Time................................................................................................................................... 10-17
Date Format............................................................................................................................. 10-17
PIP LED ................................................................................................................................... 10-18
Model Number / Serial Number ............................................................................................... 10-18
Valve Home Position ............................................................................................................... 10-19
Set Defaults ............................................................................................................................. 10-19
O2Cylinder Duration................................................................................................................ 10-20
Presets ........................................................................................................................................ 10-22
Turning Patient Query on or off ............................................................................................... 10-22
Selecting Patient Size.............................................................................................................. 10-22
LTV®1200 Presets Table ........................................................................................................ 10-23
SBT (Spontaneous Breathing Trial) Operations ......................................................................... 10-24
SBT Alarms ................................................................................................................................. 10-26

Operator’s Manual LTV®1200 Ventilator Page ix
p/n 18247-001, Rev.B
Transducer Autozero ...................................................................................................................10-27
Airway Pressure Transducer Autozero ....................................................................................10-27
Bi-directional Flow Transducer Differential Autozero...............................................................10-28
Exhalation Flow Transducer Differential Autozero - Narrow ....................................................10-29
Exhalation Flow Transducer Differential Autozero - Wide .......................................................10-30
Real Time Transducers ...............................................................................................................10-31
Chapter 11 - Ventilator Checkout Tests ............................................................ 11-1
Alarm Test .....................................................................................................................................11-3
Display Test ...................................................................................................................................11-4
Control Test ...................................................................................................................................11-6
Leak Test .......................................................................................................................................11-8
Vent Inop Alarm Test...................................................................................................................11-10
Exit...............................................................................................................................................11-12
Chapter 12 - Operating Procedure..................................................................... 12-1
To Turn the Ventilator On ..............................................................................................................12-1
Before Connecting the Ventilator to a Patient ...............................................................................12-2
Procedure for Control Mode Set Up ..............................................................................................12-4
Procedure for Assist/Control Mode Set Up....................................................................................12-5
Procedure for SIMV Mode Set Up .................................................................................................12-6
Procedure for CPAP Mode Set Up ................................................................................................12-7
Procedure for NPPV Mode Set Up ................................................................................................12-8
To Turn the Ventilator Off ..............................................................................................................12-9
LTV® Ventilator Settings Checklist..............................................................................................12-10
Chapter 13 - Cleaning, Disinfecting and Sterilizing ......................................... 13-1
Cleaning the Ventilator ..................................................................................................................13-1
Cleaning or replacing the Fan Filter ..............................................................................................13-2
Cleaning or replacing the Inlet Filter..............................................................................................13-3
Cleaning or Replacing the O2Inlet Filter .......................................................................................13-4
Cleaning the Exhalation Valve and Reusable Patient Circuit........................................................13-6
Chapter 14 - Power and Battery Operation....................................................... 14-1
Using the AC Adapter....................................................................................................................14-2
Using an External Battery..............................................................................................................14-3
Using the Automobile Cigarette Lighter Adapter ...........................................................................14-6
Replacing the Automobile Adapter Fuse..................................................................................14-10
The Universal Power Supply (UPS) ............................................................................................14-10
The Transport Battery System (TBS) ..........................................................................................14-10
Caring for the Internal Battery .....................................................................................................14-11
Battery Replacement................................................................................................................14-11
Battery Disposal .......................................................................................................................14-11
Chapter 15 - Troubleshooting............................................................................ 15-1
Displays and Buttons.....................................................................................................................15-2

Page x LTV®1200 Ventilator Operator’s Manual
p/n 18247-001, Rev. B
Ventilator Performance ................................................................................................................. 15-6
Power and Battery Operation...................................................................................................... 15-15
Alarms ......................................................................................................................................... 15-17
Checkout Test Failures ............................................................................................................... 15-23
Test Lung Operations.................................................................................................................. 15-25
Appendix A - Ventilator Specifications ..............................................................A-1
Appendix B - Set Up / Maintenance ....................................................................B-1
Recommended Maintenance Schedule ..........................................................................................B-1
Service Assistance..........................................................................................................................B-2
Appendix C - Installation and Checkout.............................................................C-1
Installation and Setup......................................................................................................................C-1
Unpacking the Ventilator – Instructions .......................................................................................C-1
Protective Boots...........................................................................................................................C-2
Protective Boot Removal .............................................................................................................C-3
Protective Boot Installation ..........................................................................................................C-6
Patient Breathing Circuit – Connection Instructions ..................................................................C-10
Oxygen Lines – Connection Instructions ...................................................................................C-12
Patient Assist Call System – Connection Instructions...............................................................C-14
Communications Port ................................................................................................................C-15
LTM™ Graphics Monitor............................................................................................................C-15
Using the Remote Alarm Cable .................................................................................................C-16
Checking the Ventilator for Proper Operation ...............................................................................C-18
Ventilator Proper Operation Worksheet.....................................................................................C-20
Appendix D - Principles of Operation.................................................................D-1
Overview .........................................................................................................................................D-1
Appendix E - Event Trace .................................................................................... E-1
Event Codes....................................................................................................................................E-3
Event Codes by Code #...............................................................................................................E-3
Event Codes by Event Name.......................................................................................................E-6
Appendix F - Glossary ......................................................................................... F-1
Appendix G - Index...............................................................................................G-1

Operator’s Manual LTV®1200 Ventilator Page 1-1
p/n 18247-001, Rev.B
Chapter 1 -
I
NTRODUCTION
This Operator’s Manual contains detailed information and instructions which when adhered to, ensure
the safe and effective set up, use and simple maintenance of the LTV® 1200 ventilator.
It is designed for use by Respiratory Therapists or other qualified and trained personnel under the
direction of a physician and in accordance with applicable state laws and regulations. It contains the
following:
•Ventilator Overview
•Installation and Checkout
•Using the Controls and Indicators
•Monitored Data
•Ventilator Alarms
•Extended Features
•Ventilator Checkout tests
•Operating Procedure
•Troubleshooting
•Cleaning, Disinfecting and Sterilizing
•Set Up / Maintenance
•Power and Battery Operation
Service tests, calibration, and major maintenance operations are described in the LTV®Series
Ventilator’s Service Manual (P/N 10665).
NOTE
Pressure Support and Pressure Control breaths on the LTV®1200 are compensated for PEEP.
Delivered pressure is controlled by the Pressure Support or Pressure Control setting and is affected
by the PEEP setting. For example, a Pressure Support setting of 20cmH2O and a PEEP setting of
10cmH2O results in a Peak Inspiratory Pressure (PIP) of 30cmH2O.
REMARQUE
La pression de support et le contrôle de la pression des respirations sur le LTV®1200 sont
compensés pour la pression expiratoire positive (PEP). La pression administrée est contrôlée par le
réglage de la pression de support ou par le réglage du contrôle de la pression et elle dépend du
réglage PEP. Par exemple, un réglage de la pression de support de 20 cmH2O et un réglage PEP de
10 cmH2O donnent une pression inspiratoire maximale (PImax) de 30 cmH2O.

Page 1-2 LTV®1200 Ventilator Operator’s Manual
p/n 18247-001, Rev. B
Operator’s Safety Information
All Operators are to read and understand the following information about Warning, Caution and
Note statements before operating the LTV®1200 ventilator.
WARNING
“WARNING” statements alert the reader to potentially hazardous situations which, if not avoided,
could result in death or serious injury.
AVERTISSEMENT
Les énoncés « AVERTISSEMENT » informent le lecteur de situations dangereuses qui, si elles ne
sont pas évitées, peuvent entraîner la mort ou des blessures graves.
CAUTION
“CAUTION” statements alert the reader to potentially hazardous situations which, if not avoided,
could result in equipment damage.
ATTENTION
Les énoncés « ATTENTION » informent le lecteur de situations dangereuses qui, si elles ne sont pas
évitées, peuvent causer des dommages à l’équipement.
NOTE
“NOTE” statements contain additional information to assist in the proper operation of the LTV®1200
ventilator.
REMARQUE
Les énoncés « REMARQUE » contiennent des informations supplémentaires pour aider à l'opération
adéquate des ventilateurs de la LTV® 1200.

Operator’s Manual LTV®1200 Ventilator Page 1-3
p/n 18247-001, Rev.B
Warnings
WARNING
Untrained Personnel – Only properly trained personnel should operate the ventilator. The LTV®
1200 ventilator is a restricted medical device designed for use by Respiratory Therapists or other
properly trained and qualified personnel under the direction of a physician and in accordance with
applicable state laws and regulations.
Leak Testing the Patient Breathing Circuit – The patient circuit must be leak tested in the VENT
CHECK mode before connection to the patient. In addition, the Ventilator Checkout mode should be
used to check for correct operation of the ventilator alarm, displays and controls. Harm to the patient
or ineffective ventilation may result from failure to leak test the patient breathing circuit before
connection to a patient. When using a heated humidifier, include it in the circuit when performing leak
testing.
Adjustable and Critical Alarms – For safety purposes, all adjustable alarms and all critical alarms
must be checked to insure proper operation.
Alarms Function Verification - All alarms must be verified as functioning properly on a daily basis. If
any alarm malfunctions, immediately contact a certified Pulmonetic Systems’ service technician or
Pulmonetic Systems, Inc.
Patient Monitoring - Patients who are dependent on a ventilator should be constantly monitored by
qualified personnel. Such personnel should be prepared to address equipment malfunctions and
circumstances where equipment becomes inoperative. An alternative method of ventilation should be
available for all patients dependent on the ventilator, and qualified personnel should be fully familiar
with emergency ventilation procedures.
Alternative Ventilation - It is recommended that an alternative means of ventilating the patient be
available at all times and that all ventilator operators be fully familiar with emergency ventilation
procedures.
Fire or Explosion - Operation of the LTV®1200 ventilator in the presence of flammable gases could
cause a fire or explosion. Under no circumstances is the ventilator to be operated when explosive
gases are present. The presence of nitrous oxide or flammable anesthetics presents a danger to the
patient and operator.
Patient Breathing Circuit Disconnection - Inadvertent disconnection of the patient from the patient
breathing circuit can be dangerous.
Critical Alarms - Failure to set the critical alarms such as the Low Minute Volume alarm and the Low
Pressure alarm may cause non-detection (no alarm) for a disconnection of the lower sense line or the
exhalation valve drive line.
Exhalation Valve Diaphragm – Patient ventilation may be ineffective or dangerous if the exhalation
valve diaphragm is damaged or worn out. The exhalation valve diaphragm must be inspected on a
daily basis and replaced whenever necessary.
Sustained HIGH PRES Alarm - During a sustained High Pressure alarm condition (HIGH PRES), the
ventilator’s turbine is stopped and gas is not delivered to the patient. Disconnect the patient from the
ventilator and ventilate the patient using an alternative method. See Chapter 15 - Troubleshooting,
Alarms for additional information concerning the HIGH PRES alarm.
BAT EMPTY Alarm - A BAT EMPTY alarm indicates the internal battery is almost depleted. Connect
the ventilator to an external power source immediately.

Page 1-4 LTV®1200 Ventilator Operator’s Manual
p/n 18247-001, Rev. B
WARNING
Battery run time - When the battery reaches the BAT LOW level, the ventilator will only run for
approximately 10 minutes before generating a battery empty alarm (BAT EMPTY). The approximate
time shown is based on tests using the nominal settings, a new battery and a full 8 hour charge cycle
as specified in Appendix A - Ventilator Specifications. Actual run time may be more or less depending
on ventilator settings, patient demand, and battery age or condition. It is highly recommended that an
alternate power source is connected PRIOR to the ventilator reaching the BAT EMPTY alarm
condition to ensure continuous, uninterrupted patient ventilation
INOP Alarm - If an INOP alarm occurs during operation, ventilate the patient using an alternative
method, disconnect the ventilator, and immediately contact a certified Pulmonetic Systems’ service
technician or Pulmonetic Systems, Inc.
NO CAL Condition - Operation of the LTV®1200 ventilator under a NO CAL condition may result in
inaccurate pressure and volume measurements. Should this condition occur, disconnect the patient
from the ventilator, provide an alternative method of ventilation and immediately contact a certified
Pulmonetic Systems’ service technician or Pulmonetic Systems, Inc.
XDCR FAULT Alarm - Continued operation of the LTV®1200 ventilator with an activated XDCR
FAULT alarm may result in inaccurate flow and volume measurements. Should this condition occur,
disconnect the patient from the ventilator, provide an alternative method of ventilation and
immediately contact a certified Pulmonetic Systems’ service technician or Pulmonetic Systems, Inc.
Personal Injury and Electric Shock - Operation of the LTV®1200 ventilator if any of its panels have
been removed may result in electrical shock to the patient or operator. All servicing must be
performed by a certified Pulmonetic Systems’ service technician.
Audible Alarms - Failure to immediately identify and correct audible alarm situations may result in
serious patient injury.
Equipment Malfunction or Failure - The LTV® 1200 ventilator has alarms to notify operators of
certain conditions and to cease operating upon detecting possible danger. In the event of equipment
failure, all ventilator operators should have an alternative method of ventilation available and be fully
familiar with emergency ventilation procedures.
Improperly Functioning Ventilator - Operation of a ventilator that does not appear to be working
properly may be hazardous. If the ventilator is damaged, fails Ventilator Checkout tests or
malfunctions in any way, discontinue its use and immediately contact a certified Pulmonetic Systems’
service technician or Pulmonetic Systems, Inc.
Ventilator Checkout Tests – Be aware that gas is not delivered to the patient during these tests.
Disconnect the patient from the ventilator and ventilate the patient using an alternative method before
running the Ventilator Checkout tests.
Ventilator Checkout and Maintenance Modes - The LTV® 1200 ventilator does not deliver gas
during the Ventilator Checkout mode (VENT CHECK) or Ventilator Maintenance mode (VENT
MTNCE) and should not be used to ventilate a patient during these tests.
Inspired Oxygen (FIO2) Concentration – If the patient has a variable respiratory rate, his/her minute
ventilation will fluctuate. If exact concentrations of inspired oxygen (FIO2) are required to be delivered
to the patient, it is recommended that an accurate oxygen analyzer with alarms be used.
O2Cylinder Duration Information - The accuracy of the displayed useable amount of oxygen
remaining in an external O2cylinder (O2 DUR hh:mm) is dependant on the precision of the pressure
gauge used on the O2cylinder and the accuracy of the information provided by the operator in the O2
CYL DUR menu settings. The calculated/displayed useable amount of oxygen information is to be
used for reference purposes only.

Operator’s Manual LTV®1200 Ventilator Page 1-5
p/n 18247-001, Rev.B
WARNING
Ventilation Variables and O2Consumption - Variations in the patient’s minute ventilation, I:E ratio
and/or ventilator setting changes or equipment status (i.e. circuit leaks) affect the consumption rate of
oxygen. When warranted by a patients condition, it is recommended that a back-up cylinder or
alternative source of oxygen be available at all times.
Before Using Automobile Cigarette Lighter or Power Outlets - Before using Automobile Cigarette
Lighter or Power Outlets as a power source for the LTV® 1200, assure that the ventilator’s internal
battery is in good condition and fully charged. Poor cigarette lighter or power outlet connections,
electrical system defects (battery, charging system, etc.), or use of vehicle accessories (air
conditioner, high current lights, high power audio equipment, etc.) could result in less than the
required voltage being delivered to the ventilator, generate a Power Lost alarm and switch the
ventilator’s power source to the internal battery.
Unauthorized Parts or Accessories – Serious harm to the patient may result from the use of
unauthorized parts or accessories. Only items expressly approved by Pulmonetic Systems may be
used in conjunction with the LTV® 1200 ventilators.
Unapproved Adapters – Only Pulmonetic Systems Accessories should be used to connect the
ventilator to Patient Assist Call Systems. These accessories incorporate safety features to reduce the
risk of shock. Do not attempt to modify these accessories in any way.
Patient Assist Call Connector – Do not apply more than 25V rms or 32VDC to the Patient Assist
Call connector.
Ventilator Service and Repair - All servicing or repair of the LTV® 1200 ventilator must be
performed only by a service technician certified by Pulmonetic Systems
Patient Circuits – Pulmonetic Systems Patient Circuits, Exhalation Valve Assemblies and Water
Traps are shipped clean, not sterile.
Ultra Violet Light Sensitivity – The material used in the tubing of the Reusable Patient Circuits is
not UV stable. Avoid exposure of the tubing to UV light.
Mounting Screws - Refer to the information contained in Pulmonetic Systems Replacement Screws
Kit, P/N 11149, to determine the appropriate accessories mounting screws or accessories
replacement screws location, type and length to use when removing or exchanging external
accessories on an LTV® ventilator.
Mounting Screw Use – Internal damage to the ventilator may result if the wrong length mounting
screws are used when installing or removing external accessories.
Specific Boot Replacement Screw Location - One leg of the upper protective boot has an
additional screw hole (furthest from the end of the leg); On earlier version ventilators (screw was
located in the upper hole in the leg of the boot) the use of a 3/16” mounting screw is required. On
current version ventilators (screw was located in the lower hole in the leg of the boot) the use of a 1/4”
mounting screw is required.
Specific Boot Installation Screw Location - One leg of the upper protective boot has an additional
screw hole (furthest from the end of the leg); On earlier version ventilators, the screw hole will align
with the upper hole in the boot and requires the use of the 1/4” mounting screw. On current version
ventilators, the screw hole will align with the lower hole in the boot and requires the use of the 3/8”
mounting screw.
Patient Circuit Accessories - The use of accessories such as Speaking Valves, Heat-Moisture
Exchangers and Filters create additional patient circuit resistance and in the event of a disconnection,
may impede the generation of a Low Pressure alarm. Ensure that the Low Pressure alarm settings
accommodate these types of accessories when used in combination with patient circuits.

Page 1-6 LTV®1200 Ventilator Operator’s Manual
p/n 18247-001, Rev. B
WARNING
Low Minute Volume Control Settings - The Low Min. Vol. control should be set to its highest
clinically appropriate value. If there is a clinical need to set the Low Minute Volume alarm to lower
values or off (“- - -“), perform a clinical assessment to determine if an alternative monitor (i.e. a Pulse
Oxymeter with an audible alarm, or a Cardio Respiratory Monitor) should be used.

Operator’s Manual LTV®1200 Ventilator Page 1-7
p/n 18247-001, Rev.B
AVERTISSEMENT
Personnel non qualifié - Seul le personnel qualifié doit opérer le ventilateur. Le ventilateur de la
1200 LTV®est un dispositif médical restreint conçu pour être utilisé par les inhalothérapeutes ou
autres personnes qualifiées, et par le personnel qualifié sous la supervision d'un médecin et en
conformité avec les lois et règlements applicables.
Contrôle de l’étanchéité du circuit respiratoire du patient – L'étanchéité du circuit respiratoire du
patient (vérification de ventilation) doit être vérifiée en mode VENT CHECK avant le raccordement au
patient. En outre, on doit utiliser le mode Ventilator Checkout (vérification du ventilateur) afin de
s’assurer du fonctionnement adéquat de l’alarme, des affichages et des commandes du ventilateur.
Le défaut de vérifier l’étanchéité du circuit respiratoire du patient avant le raccordement à un patient
peut être nocif pour le patient ou provoquer une ventilation inefficace. Lorsqu’un humidificateur
chauffant est employé, il convient de l’inclure dans le circuit en procédant à la vérification de
l’étanchéité.
Alarmes réglables et critiques – Pour assurer la sécurité et obtenir un fonctionnement adéquat,
toutes les alarmes réglables et critiques doivent être vérifiées.
Vérification du fonctionnement des alarmes - Toutes les alarmes sonores et visuelles doivent être
vérifiées quotidiennement. Si une des alarmes fonctionne de façon inadéquate, contactez votre
technicien de service certifié de Pulmonetic Systems ou Pulmonetic Systems Inc.
Surveillance du patient – Un personnel qualifié doit constamment surveiller les patients qui sont
reliés à un ventilateur. Le personnel doit être en mesure de s’occuper des défectuosités de
fonctionnement de l’équipement ainsi que des circonstances où ce dernier devient inopérant. Une
forme de ventilation alternative doit être disponible à tous les patients reliés au ventilateur et le
personnel qualifié devrait être pleinement familier avec les procédures de ventilation d’urgence.
Ventilation alternative - Il est recommandé qu'un moyen alternatif de ventilation soit disponible en
tout temps, et que tous les opérateurs de ventilateur soient pleinement familiers avec les procédures
de ventilation d'urgence.
Feu ou explosion - L'opération des ventilateurs de la 1200 LTV®en présence de gaz inflammables
peut causer un feu ou une explosion. Le ventilateur ne doit être opéré sous aucune circonstance en
présence de gaz. La présence d'oxyde nitreux ou d'anesthésiques inflammables représente un
danger pour le patient et l'opérateur.
Débranchement du circuit respiratoire du patient - Le débranchement accidentel du circuit
respiratoire du patient peut s'avérer dangereux.
Alarmes critiques – Le défaut de définir les alarmes critiques telles que l’alarme basse ventilation-
minute et l’alarme basse pression peut causer une non-détection (absence d’alarme) pour un
débranchement du tube de détection inférieur ou du tube d’entraînement de la soupape d’expiration.
Diaphragme de la soupape d'expiration - Une ventilation inefficace ou dangereuse pour le patient
peut résulter si le diaphragme de la soupape est endommagé ou usé. Le diaphragme de la soupape
d'expiration doit être vérifié quotidiennement, et remplacé au besoin.
Alarme ALARME PMAX continue — Dans des conditions d’alarme de haute pression prolongées
(ALARME PMAX), la turbine du ventilateur s'arrête et le gaz n’est plus transmis au patient.
Débranchez le patient du ventilateur et utilisez une autre méthode de ventilation. Pour plus de détails
sur l’état ALARME PMAX, reportez-vous au chapitre 15, Troubleshooting, Alarms.

Page 1-8 LTV®1200 Ventilator Operator’s Manual
p/n 18247-001, Rev. B
AVERTISSEMENT
Durée d’utilisation de la batterie – Lorsque la batterie atteint le niveau BAT INT BASS, le
ventilateur fonctionne pendant environ 10 minutes avant d’émettre une alarme de batterie faible (BAT
INT VIDE). Cette durée approximative est basée sur des tests avec des paramètres nominaux, une
nouvelle batterie et un cycle de chargement complet de 8 heures, tel que spécifié dans l’Annexe A –
Spécifications du ventilateur. La durée d’utilisation réelle pourrait être supérieure ou inférieure, selon
les paramètres du ventilateur, la demande du patient et l’âge ou l’état de la batterie. Il est fortement
recommandé qu’une source d’alimentation alternative soit connectée AVANT que le ventilateur
n’atteigne l’état d’alarme BAT INT VIDE afin d’assurer une ventilation continue et ininterrompue au
patient.
Alarme BAT EMPTY - Une alarme BAT EMPTY indique que la pile interne est pratiquement à plat.
Branchez immédiatement le ventilateur à une source d'alimentation externe.
Alarme INOP - Si une alarme INOP survient au cours de l'opération, ventilez le patient à l'aide de la
méthode alternative, retirez immédiatement le ventilateur du service, et contactez immédiatement
votre technicien de service certifié de Pulmonetic Systems ou Pulmonetic Systems Inc.
Condition NO CAL - L'opération continue du ventilateur de la série LTV® sous condition NO CAL
peut résulter en mesures de pression et de volume erronées. Si cette condition se présente, le
ventilateur doit être retiré du service, et vous devez immédiatement contacter votre technicien de
service certifié de Pulmonetic Systems ou Pulmonetic Systems Inc.
Alarme XDCR FAULT - L'opération continue du ventilateur de la série LTV® avec une alarme XDCR
FAULT activée peut résulter en mesures de débit et de volume erronées. Si cette condition se
présente, le ventilateur doit être retiré du service, et vous devez immédiatement contacter votre
technicien de service certifié de Pulmonetic Systems ou Pulmonetic Systems Inc.
Blessures personnelles et chocs électriques - L'opération d’un ventilateur de la série LTV® alors
que ses panneaux sont enlevés, peut causer un choc électrique au patient ou à l'opérateur. Tout
entretien doit être effectué par un technicien de service certifié de Pulmonetic Systems.
Alarmes sonores - L'échec à identifier et à corriger dans l'immédiat les situations d'alarmes sonores
peut causer des blessures au patient.
Mauvais fonctionnement ou panne de l'équipement - Des dispositifs électromécaniques peuvent
mal fonctionner ou subir une panne. Le ventilateur de la série LTV® a été conçu avec des alarmes,
pour détecter et aviser les opérateurs de certaines conditions, et pour cesser d'opérer en cas de
conditions d'opération dangereuses. En cas de panne de l'équipement, tous les opérateurs du
ventilateur devraient avoir une forme de ventilation alternative à leur disponibilité, et être pleinement
familiers avec les procédures de ventilation d'urgence.
Ventilateurs fonctionnant de façon inadéquate - L'opération d'un ventilateur dont le
fonctionnement semble inadéquat peut représenter un danger. Si le ventilateur est endommagé, s'il
échoue les tests de vérification du ventilateur ou s'il fonctionne de façon inadéquate, suspendez
l'utilisation de ce ventilateur et contactez immédiatement votre technicien de service certifié de
Pulmonetic Systems Inc.
Tests de vérification du ventilateur – Noter que le gaz n’est pas transmis au patient au cours de
ces tests. Débrancher le patient du ventilateur et ventiler le patient à l’aide d’une forme de ventilation
alternative avant de procéder aux tests de vérification du ventilateur.
Modes Vérification et Entretien du ventilateur - Le ventilateur de la série LTV® ne transmet pas le
mélange de gaz en mode Vérification du ventilateur (VENT CHECK) ou en mode Entretien du
ventilateur (VENT MTNCE), il ne devrait donc pas être utilisé pour ventiler un patient durant
l'exécution de ces tests.

Operator’s Manual LTV®1200 Ventilator Page 1-9
p/n 18247-001, Rev.B
AVERTISSEMENT
Concentration d’oxygène inspiré (FIO2)– Si la fréquence respiratoire du patient est variable, sa
ventilation-minute va fluctuer. Lorsqu’une concentration exacte d’oxygène inspiré (FIO2) est
nécessaire pour une transmission au patient, il est recommandé d’utiliser un analyseur de niveau
d’oxygène2 précis, comportant des alarmes.
Informations sur la durée d'utilisation restante de la bouteille d’oxygène - La précision de
l’affichage de la quantité d’oxygène utilisable restante dans une bouteille d’oxygène externe (O2 DUR
hh:mm) dépend de la précision de la jauge de pression utilisée sur la bouteille et de l'exactitude des
informations fournies par l’opérateur dans les paramètres du menu DUREE CYL O2. Les
informations calculées et affichées sur la quantité d’oxygène utilisable ne doivent être utilisées qu’à
titre indicatif.
Variables de ventilation et consommation d’oxygène — Les variations dans la ventilation par
minute du patient et dans le rapport inspiration/expiration, la modification des paramètres ou l’état du
matériel (fuite dans le circuit, par exemple) modifient le taux de consommation de l’oxygène. Lorsque
la situation du patient le permet, il est recommandé qu'une bouteille d’oxygène de secours ou toute
autre source alternative d’oxygène soit disponible en permanence.
Avant toute utilisation d’une prise d’allume-cigare ou d’une prise de courant — Avant d’utiliser
un allume-cigare ou une prise de courant comme source d’alimentation du ventilateur LTV®, vérifiez
que la batterie interne du ventilateur est en bon état et entièrement chargée. L’utilisation d’un allume-
cigare ou d’une prise de courant fournissant un branchement de qualité médiocre, des défauts du
circuit électrique (batterie, système de charge, etc.), ou l’utilisation d’accessoires d’automobile
(climatisation, phares, chaîne stéréo et haut-parleurs à forte consommation, etc.) peuvent affecter le
voltage délivré au ventilateur et provoquer une sous-alimentation de celui-ci. Dans cette situation, le
ventilateur déclenche une alarme PAS ALIM SEC et utilise la batterie interne du ventilateur comme
source d’alimentation.
Pièces, accessoires et options non autorisées - Des dommages à l'équipement ou des blessures
au patient peuvent survenir suite à l'utilisation de pièces, accessoires et options non autorisées.
Seuls les éléments expressément approuvés par Pulmonetic Systems doivent être utilisés en
conjonction avec les ventilateurs de la série LTV®.
Accessoires non approuvés – L’utilisation d’accessoires qui ne sont pas expressément approuvés
par Pulmonetic Systems pourrait entraîner des conditions dangereuses. Seuls les accessoires de
Pulmonetic Systems devraient être utilisés pour brancher les ventilateurs aux systèmes d’aide aux
patients. Ces accessoires comportent des caractéristiques de sécurité pour réduire les risques de
choc. N’essayez pas de modifier ces accessoires d’aucune façon.
Connecteur d’appel d’aide aux patients – Ne mettez pas plus de 25 V efficace ou 32 V c.c. au
connecteur d’appel d’aide aux patients.
Entretien et réparation du ventilateur - Tout entretien ou réparation du ventilateur de la série LTV®
ne doit être effectué que par un technicien de service certifié de Pulmonetic Systems.
Circuits du patient – Les circuits du patient du Pulmonetic Systems, les valves expiratoires et les
collecteurs d’eau sont expédiés propres, mais pas stériles.
Sensibilité à la lumière ultraviolette – Les matériaux utilisés pour la tubulure des circuits du patient
ne sont pas stables sous rayons UV. Éviter d’exposer la tubulure à la lumière UV.
Vis de montage des accessoires – Voir les renseignements fournis dans la trousse de vis de
remplacement de Pulmonetic Systems, numéro de pièce 11149, pour déterminer l’emplacement, le
type et la longueur des vis de montage d’accessoires ou des vis de remplacement pour accessoires à
utiliser lors de la dépose ou de l’échange d’accessoires externes sur un ventilateur de la série LTV®.
Table of contents