schupp Trendino 139235 User manual

EN
INSTRUCTIONS FOR USE
Trendino
Therapy couch
139235
[UDI-TRENDINO]
[Current user manual]

220822SCHUPP GmbH & Co. KG
72280 Dornstetten-Aach | Glattal Straße 78 | T: +49 (0) 7443/243-0 | F: +49 (0) 7443/243-255 | info@schupp-gmbh.de | www.schupp.shop
2
Trendino
Imprint
Imprint
Manufacturer
Production
DE-MF-000010127
SCHUPP GmbH & Co KG
Glattal Street 78
D-72280 Dornstetten-Aach
Internet: www.schupp.shop
SCHUPP GmbH & Co KG
Neunecker Street 18
D-72290 Lossburg-Wittendorf
Copyright
The document including all illustrations is protected by copyright. Any
use outside the limits of copyright law without the consent of the
manufacturer is prohibited and punishable by law.
This applies in particular to duplications, translations, as well as
storage and processing in electronic systems.
Infringements give rise to a claim for damages and may have penal
consequences.
All rights reserved.
© 2022 SCHUPP GmbH & Co KG
Warranty / Liability
These instructions for use are obligatory before
carefully and completely before putting the product into
operation!
For correct use of the medical device, observe all instructions and
safety notes!
The instructions for use must always be accessible to all users!
You can call up the current status of the instruction manual online:
We reserve the right to make technical changes within the scope of
further development of the therapy couch without prior notice. All
information is non-binding. Misprints excepted.
We accept no liability for damage and malfunctions caused by
operating errors and non-compliance with these operating instructions.

220822SCHUPP GmbH & Co. KG
72280 Dornstetten-Aach | Glattal Straße 78 | T: +49 (0) 7443/243-0 | F: +49 (0) 7443/243-255 | info@schupp-gmbh.de | www.schupp.shop
3
Trendino
Index
TABLE OF CONTENTS
Direct link enabled [press the CTRL key and click on the chapter name].
1.0 General........................................................................................... 4
2.0 Safety regulations......................................................................... 4
2.1 Safety signs - Symbols................................................................................4
2.2 Applied standards..........................................................................................5
2.3 Intended use.....................................................................................................5
3.0 Safety instructions........................................................................ 6
3.1 General ...............................................................................................................6
3.2 Safety instructions for delivery / unpacking......................................7
3.3 Safety instructions for commissioning................................................7
4.0 Warranty / Guarantee.................................................................... 8
5.0 Disclaimer...................................................................................... 8
6.0 Function and operation................................................................ 9
6.1 Set-up and installation.................................................................................9
6.2 Height adjustment with electric motor .................................................9
6.3 Adjusting the head section cushion (1 & 3 parts).........................11
6.4 Adjusting the armrest upholstery (3 parts) ......................................11
6.5 Anti-crash monitoring................................................................................12
6.6 Paper roll holder...........................................................................................13
6.7 Equipment, optional assemblies...........................................................14
6.8 Accessories....................................................................................................15
7.0 Spare and wear parts.................................................................. 15
8.0 Service life, maintenance and care........................................... 15
8.1 Maintenance...................................................................................................15
8.2 Personal precautions.................................................................................16
8.3 Care and cleaning........................................................................................17
8.4 Surface disinfection....................................................................................17
9.0 Help with malfunctions............................................................... 18
10.0 Nameplate.................................................................................... 19
11.0 Disposal....................................................................................... 20
12.0 Technical data............................................................................. 20
12.1 Switching diagrams - Enable coding ..................................................22
13.0 Declaration of conformity .......................................................... 22
14.0 Contact.........................................Fehler! Textmarke nicht definiert.

220822SCHUPP GmbH & Co. KG
72280 Dornstetten-Aach | Glattal Straße 78 | T: +49 (0) 7443/243-0 | F: +49 (0) 7443/243-255 | info@schupp-gmbh.de | www.schupp.shop
4
Trendino
SECURITY
1.0 General
With the requirements and specifications of physiotherapists, this therapy couch was developed in close cooperation for
the professional therapy area. Special attention was paid to the safety of the couch, the components of the integrated
safety, the comfort of the couch as well as durability and service life.
The design and manufacture of the therapy couch comply with the current safety principles of the BfArM and take into
account the generally accepted state of the art. The couch does not impair the safety or condition of patients or the
safety and health of users or third parties.
"Made in Germany" as well as the use of the highest quality materials guarantee a long and reliable functionality of the
therapy couch when used as intended.
All therapy tables from SCHUPP GmbH & Co.KG are developed in the Black Forest / Baden Württemberg and
manufactured with regional suppliers. The assembly of the therapy tables takes place in our ISO certified production
facility and thus guarantees a high and always consistent processing quality.
2.0 Safety regulations
2.1 Safety signs - symbols
Safety-relevant sections, instructions or notes are highlighted by the following symbols. Please familiarize yourself with
the definitions before reading the operating instructions.
Indicates an immediate danger to users, patients and third parties which can result in death or
serious injury.
DANGER
Indicates a potentially hazardous situation that could result in a low or moderate injury
to the user, patient, third parties or damage to the medical device.
or on other objects.
Warning
Warning of danger of trapping and crushing when adjusting the therapy couch
Warning
Warning of electrical voltage
Warning
Indicates a possible danger to property, which may result in property damage.
Caution
It is essential to observe the instructions for use
Additional notes, application tips, useful information

220822SCHUPP GmbH & Co. KG
72280 Dornstetten-Aach | Glattal Straße 78 | T: +49 (0) 7443/243-0 | F: +49 (0) 7443/243-255 | info@schupp-gmbh.de | www.schupp.shop
5
Trendino
SECURITY
2.2 Applied standards
The therapy couch has been developed and manufactured according to the latest state of the art and in compliance
with national/international guidelines. It complies with the following guidelines:
EU MDR (Medical Device
Regulation) 2017/745
This product is a class I medical device according to the EU Medical Device
Regulation.
DIN EN ISO 10993-1:2017-04
Biological evaluation of medical devices
DIN EN 60601-1:2013-12
Medical electrical equipment: General requirements for safety including
essential performance characteristics
DIN EN 62366-1:2015 + A1:2020
Medical devices - Application of fitness for use to medical devices
Machinery Directive 2006/42/EC
Medical hydraulic equipment - General requirements for safety
2.3 Intended use
The therapy couch of the company Schupp GmbH & Co. KG is a medical device of class 1
(EU MDR 2017/745) and is intended exclusively for professional use by medically trained personnel and instructed
users for the proper positioning and storage of patients during therapeutic healing and medical treatments or for the
short-term examination of patients in the rooms provided for this purpose.
The Schupp company does not give any guarantee regarding the suitability of the therapy couch for a specific
therapeutic or diagnostic purpose. The therapist determines the appropriate form of application.
The therapy couch is not intended or equipped for hydrotherapy or for domestic or radiological applications. Also, the
electrically operated couches must not be used in areas with flammable gases, vapours or increased humidity. In case
of non-observance, this can lead to considerable damage to the couch as well as endanger the patient, user or third
parties.

220822SCHUPP GmbH & Co. KG
72280 Dornstetten-Aach | Glattal Straße 78 | T: +49 (0) 7443/243-0 | F: +49 (0) 7443/243-255 | info@schupp-gmbh.de | www.schupp.shop
6
Trendino
SECURITY
3.0 Safety instructions
3.1 General
For your own safety as well as the safety of the patient and third parties, familiarize yourself with the limits and dangers
associated with the use of the "electrically height-adjustable therapy couch" before using it for the first time.
Observe all safety and warning instructions! Also observe the safety and warning instructions when using devices
together !
These operating instructions contain important information on safe handling, operation, care and
maintenance.
The instructions are an integral part of the therapy couch and must always be kept within reach for
all users.
In the event of a change of ownership or location, these instructions must be handed over with the
couch. Failure to comply with these instructions will result in the loss of warranty and guarantee
claims!
Observe the additional regional and country-specific rules and regulations for occupational safety
and accident prevention that apply to the place of use.
The operator may only use the therapy couch within its intended purpose for therapy and treatment
The operator may only allow qualified personnel or persons instructed and authorised in work safety
and accident prevention regulations to work on the therapy couch.
Caution
The operator must ensure that all users have read and understood the instructions, warnings and
safety notes in these instructions and are sufficiently informed about their operation. This must be
checked at regular intervals.
The operator must ensure that the therapy couch cannot be operated by unauthorized persons
when unattended.
The user must ensure that children are always supervised and do not stay in the area of the therapy
couch. This can lead to serious accidents.
The operator may only allow trained persons to work on the therapy couch for cleaning and care.
When leaving the therapy couch, it must be secured in such a way that unauthorized adjustment is
not possible. This can lead to serious accidents.
The therapy couch / pads are not sterile. Open wounds or skin irritations must not come into contact
with the padding material.
DANGER
When leaving the therapy couch, it must be secured in such a way that unauthorized adjustment is
not possible. For absolute safety, we recommend pulling the mains plug out of the socket when the
therapy couch is unattended.
When using additional devices (e.g. red light or heat emitters), the user must ensure that the
therapy couch cannot be adjusted unintentionally or accidentally. Safety distances must be
observed exactly.
The operator has a duty of care towards his employees. We therefore recommend safe footwear to
prevent the risk of injury.

220822SCHUPP GmbH & Co. KG
72280 Dornstetten-Aach | Glattal Straße 78 | T: +49 (0) 7443/243-0 | F: +49 (0) 7443/243-255 | info@schupp-gmbh.de | www.schupp.shop
7
Trendino
SECURITY
3.2 Safety instructions for delivery / unpacking
The therapy couches are 100% visually inspected and checked for electrical and mechanical
function before shipment. The couch has left the factory in perfect and operational condition.
Remove the transport securing strap under the head section and in the frame!
Attention when opening the packaging! Do not use pointed or sharp objects, the padding could be
directly damaged.
Check immediately after receipt of the shipment for completeness of the delivery as well as for
packaging damage. In the event of damage, this must be claimed immediately from the carrier and
must be noted immediately on the shipping documents.
If damage to the medical device is detected, it must be reported in writing to Schupp GmbH & Co.
KG within 2 working days.
After this period, warranty claims arising from transport damage can no longer be asserted.
3.3 Safety instructions for commissioning
Treatments may only be performed on a therapy couch that is in perfect safety condition; there is no
danger from electrical energy.
The therapy couch may only be actively connected to the mains during treatment.
The voltage and frequency of the house power supply must correspond to the specifications on the
type plate. Otherwise, operation is prohibited.
The couch was designed exclusively for use in normal conditions indoors.
The therapy couch should be placed on a firm and level surface and have adequate free space for
the user.
Warning
Observe the acclimatisation time before commissioning.
The power cable must be positioned so that it does not get jammed between the moving parts of
the couch - and - so that no one can trip over it. Strong tensile loads as well as running over with
the couch must be avoided.
Do not fix any cables or hoses to the rack frame, as they can be torn off by the couch movement
The mains plug must be disconnected for all activities inside the couch.
When using additional devices (e.g. red light or heat emitters), safety distances must be strictly
observed. Uncontrolled adjustment may result in burns to the skin or upholstery.
When moving or lifting the wheel frame of the couch, no patient may be on the couch (overloading
and danger of tipping over).
Treatments are only allowed with the wheel frame retracted.
Caution
Personally keep a sufficient distance when adjusting and driving the couch.
Before or during each position adjustment of the couch, make sure that no persons or objects
(children, animals, chairs, stools....) are in the moving area of the couch.
Also do not store any objects for use (e.g. covers, towels, belts, cushions....) in the area of the
frame!
Never load the arm, foot and/or head section with the total weight of the patient (danger of tipping
over) - a maximum load of 30kg is permitted.
Warning
Neither the therapist nor the patient may use the arm or head section as a sitting, kneeling, or
abusing the ascender.

220822SCHUPP GmbH & Co. KG
72280 Dornstetten-Aach | Glattal Straße 78 | T: +49 (0) 7443/243-0 | F: +49 (0) 7443/243-255 | info@schupp-gmbh.de | www.schupp.shop
8
Trendino
SECURITY
There is a risk of entrapment when adjusting the height!
Observe during each position adjustment,
that no part of the patient's own or body parts
or third parties between the moving parts or
between the upholstered part and the frame.
The danger points are marked on the couch with
marked with appropriate warning signs.
Warning
Never exceed the maximum static load capacity of:
150kg (139235)!
Regular inspection of the technical condition by the operator is required.
Regular monitoring or maintenance of wear parts such as operating and adjustment elements, gas
springs, hydraulic cylinders, motors and pads.
Faults must be rectified immediately, damaged components must be replaced without delay and
the couch must not be used until it has been repaired.
To ensure safety, only use original accessories from Schupp. Incorrect accessories can lead to
material damage to the couch or to injuries!
Caution !
Repair and maintenance only by the manufacturer or personnel authorized by the manufacturer.
The couch may not be modified without the written permission of the manufacturer. Conversions
and technical modifications are not permitted.
When used outside the intended purpose or when using additional devices, it is the responsibility
of the operator to comply with the limit values for the safety of medical electrical systems
according to DIN EN 60601-1 and following.
The customary national laws / regulations must be observed when operating this therapy couch.
Observe the necessary safety checks according to MPBetreibV (Medical Devices Operator
Ordinance), BetrSichV (Operational Safety Ordinance) and DGUV (German Social Accident
Insurance).
4.0 Warranty / Warranty
Schupp GmbH & Co KG guarantees you a therapy couch free of material and processing defects. The couch was
thoroughly tested before shipment and 100% approved in the final production test.
The warranty period is 60 months for the couch frame, 12 months for upholstered parts, 24 months for all other
components of the therapy couch, with the exception of wear parts, and begins at the time of delivery. Please be sure to
keep the invoice as proof of purchase.
Parts or components of the couch which prove to be defective within these periods can be sent to the service department
for claim settlement after prior consultation. Product liability ends 10 years after the product has been placed on the
market. The manufacturer is no longer obliged to provide replacement parts or accessories.
The General Terms and Conditions for Deliveries and Services as well as the warranty and guarantee provisions of
Schupp GmbH & Co. KG [current version at: www.schupp.shop] shall apply.
The therapy couch may only be repaired or modified by the manufacturer or by a person or company authorized by the
manufacturer! Otherwise, the warranty claim will expire!
5.0 Disclaimer
We expressly draw your attention to the fact that no liability is accepted for personal injury and/or damage to property
caused by failure to observe the instructions for use, by incorrect or negligent operation, maintenance, servicing and
repair or by improper use of the therapy couch or its accessories, or by exceeding the performance limits of the couch.
This also applies to the effects of foreign bodies, modifications, additions and conversions to the table which impair or
could impair safety.

220822SCHUPP GmbH & Co. KG
72280 Dornstetten-Aach | Glattal Straße 78 | T: +49 (0) 7443/243-0 | F: +49 (0) 7443/243-255 | info@schupp-gmbh.de | www.schupp.shop
9
Trendino
FUNCTION
6.0 Function and operation
6.1 Set up and installation
Set up the therapy couch at the desired treatment site and remove the protective foil from the couch. Do not use a
cutter, knife or similar pointed tools for this purpose. The upholstery could be damaged.
The packaging is made of environmentally friendly and recyclable materials and can be disposed of properly at a
collection point. Also note when setting up:
Carry the couch only by the lower metal frame, never use the head section or arm sections as carrying handles!
Remove the transport securing tape under the head section and in the rack frame before startup!
Due to the special design of the couch, you do not need an extra wheeled frame to move it. Just lift the couch by
the metal frame below the headrest cushion and move it to the desired installation location.
Observe couch width - Do not push the couch against a door frame! Upholstery can burst open!
The stability in case of small unevennesses in the floor can
be improved by unscrewing the level adjustment on the feet
of the couch frame. For adjustment, you need an open-end
wrench with a width across flats of 16 to loosen the lock nut.
In case of an energetically operated therapy couch, it should be positioned near a socket!
Lay the mains cable in such a way that there is no danger, use cable bridges if necessary!
Check whether the operating voltage of the therapy couch (label) corresponds to the mains voltage!
The therapy couch is immediately ready for operation after connection to the mains voltage.
6.2 Adjusting the height with electric motor
The adjustment of the height by means of an electric motor is carried out by means of a hand or foot switch.

220822SCHUPP GmbH & Co. KG
72280 Dornstetten-Aach | Glattal Straße 78 | T: +49 (0) 7443/243-0 | F: +49 (0) 7443/243-255 | info@schupp-gmbh.de | www.schupp.shop
10
Trendino
FUNCTION
HAND & FOOT SWITCH
The electric height adjustment is actuated by double-
pressure actuation of the manual switch (release coding) or
the foot switch to prevent unintentional or unauthorized
adjustment.
To adjust the height, the same direction key on the hand
switch or foot switch must be pressed twice in quick
succession. The drive is enabled after the second actuation.
Now the couch surface can be raised or lowered. With
Lying surface moves upwards
Lying surface moves down
the drive moves in the desired direction as long as the
switch remains pressed. Releasing the key immediately
interrupts the adjustment process.
Manual switch
Foot switch
Once the height adjustment has been completed, within
3 seconds a correction is made before the couch switches
off automatically.
If the interruption is greater than 3 seconds, you must
execute the double-pressure release again. For switching
diagram, see Technical data.
The times are hard-coded in the drive and cannot be
changed.
►The therapy couch must not be operated by unauthorized persons when unattended!
►Children must always be supervised and are not allowed to stay in the area of the therapy couch!
►Cleaning and care personnel must always be instructed in occupational safety.
DANGER!
►Moving parts - Keep your distance when adjusting the position!
►Always detach the hand control from the frame to adjust the couch!
Warning.
►When adjusting, make sure that the cable of the hand switch, foot switch cannot be crushed or otherwise
damaged!
►Make sure that no one gets trapped between the arm section, head section and floor when the therapy
couch is at a low height.
►Accidental, single or permanent actuation by a weight of an object or a person does not cause the rack to
lower.
The screw jack motor is designed for short-time operation. In case of strong and continuous overload, the motor
is switched off automatically by the integrated thermal fuse to prevent further damage. After approx. 20-30
minutes, the motor is ready for operation again!
When adjusting the lying surface with very heavy patients, the motor operates more slowly than with light
patients. This condition is due to the design and does not represent faulty operation of the adjustment drive.
►Observe the maximum dynamic load capacity of the couch of:
150kg (139235)!

220822SCHUPP GmbH & Co. KG
72280 Dornstetten-Aach | Glattal Straße 78 | T: +49 (0) 7443/243-0 | F: +49 (0) 7443/243-255 | info@schupp-gmbh.de | www.schupp.shop
11
Trendino
FUNCTION
6.3 Adjustment of the headboard cushion (1 & 3 parts)
You can raise or lower the headrest upholstery continuously
by pressing the operating lever under the headrest
upholstery in the direction of the upholstery surface and
simultaneously moving the headrest up and down.
After releasing the actuating lever, the gas pressure spring,
which is blocked in the direction of tension and compression,
holds the headrest cushion securely in its newly adjusted
position.
The nose slot insert can be inserted or removed depending
on the application.
►Moving parts - Do not press the operating lever of the head section upwards in an uncontrolled
manner!
►Neither the therapist nor the patient may sit, kneel or stand on the head section of the couch!
►The head section may be loaded with a maximum of 30 kg!
Warning.
6.4 Adjustment of the arm part-cushion (3 parts)
With the 3-piece headrest (optional), both armrests can be
adjusted completely independently of each other.
The lowering of the armrest cushions is infinitely variable
over a range of up to 18 cm.
By means of star handle the arm part can be fixed in the
individual position.
To adjust an armrest, hold the armrest cushion with one
hand and loosen the star handle (1) underneath with the
other hand.
Now move the arm section cushion along the rail to the
desired position and tighten the star grip (1) again.
Repeat this procedure for the opposite armrest.
Make sure that all star knobs are tightened again after
each adjustment.
Both armrests are detachably connected to the headboard.
To remove an armrest, hold the armrest cushion with one
hand and loosen the star grip (2) on the inside with the
other hand. Now the armrest can be pulled out to the side.
Repeat this process for the opposite side if necessary.
► Watch your hands when adjusting the arm section cushions!
Warning.
(2) Withdrawal
(1) Adjustment
(2)
Withdrawal

220822SCHUPP GmbH & Co. KG
72280 Dornstetten-Aach | Glattal Straße 78 | T: +49 (0) 7443/243-0 | F: +49 (0) 7443/243-255 | info@schupp-gmbh.de | www.schupp.shop
12
1.0
Trendino
FUNCTION
6.5 Anti-crash monitoring
The therapy couch is equipped with anti-crash monitoring
of the cushion positions.
(utility model registered: DBGM 20 2006 014 884.1)
The large height adjustment range of the couch down to
40cm allows adjustment of the:
1-part head section up to -90° resp.
3-piece headboard up to -40°.
actually not to.
Such an angled head section would inevitably hit the floor
when the couch is lowered.
This would result in material damage to the table or the
floor. A risk of injury for users and patients could also not
be ruled out (risk of crushing feet).
The anti-crash monitoring system detects dangerous
cushion positions during the lowering movement and
automatically switches off the motor drive when such a
position is reached.
The shutdown always refers to the lowering movement of
the lying surface into the hazardous area.
Protected areas of the anti-crash monitoring
If the 1-part head section is in the adjustment range of -30° ... -90°, the height of the lying surface can only be lowered
up to 58cm.
If the 3-part head section is in the adjustment range of 0° ... -40°, the height of the lying surface can only be lowered
up to 58cm.

220822SCHUPP GmbH & Co. KG
72280 Dornstetten-Aach | Glattal Straße 78 | T: +49 (0) 7443/243-0 | F: +49 (0) 7443/243-255 | info@schupp-gmbh.de | www.schupp.shop
13
1.0
Trendino
FUNCTION
If the couch is to be lowered even further after a stop, the
head section must be adjusted upwards accordingly (see
Chapter 5.3).
As soon as the 1-part head section is no longer in the
area at risk of collision, the couch can be lowered to the
lower height of 40cm.
As soon as the 3-part head section is no longer in the
area at risk of collision, the couch can be lowered to the
lower height of 40cm.
After a shutdown, the couch can always be raised from
any position, even without headrest adjustment.
6.6 Paper roll holder
The therapy couch can be equipped with the "Paper roll holder" option. The paper roll holder is mounted under the
cushion or on the frame on the narrow side of the foot section. The holder can accommodate paper rolls with a width of
up to 59cm.
Place the paper roll on the peg of a holder.
With the other hand, pull the opposite bracket outward.
Place the paper roll on the pin of this holder from below.
Release the holder.
For optimal adjustment of the paper roll holder to the roll
width, the fastening screws can be loosened.

220822SCHUPP GmbH & Co. KG
72280 Dornstetten-Aach | Glattal Straße 78 | T: +49 (0) 7443/243-0 | F: +49 (0) 7443/243-255 | info@schupp-gmbh.de | www.schupp.shop
14
Trendino
EQUIPMENT
6.7 Equipment, optional assemblies
Couch type
TRENDINO
Item no:
139235
Medical device
●
Integrated safety [IS], drive enable coding
●
Foot switch
○
Manual switch
●
All-around switch bracket, railing
Height adjustment, hydraulic, operable from
both sides
Height adjustment, electric, 8000N
●
Height adjustment, electric, 10000N, extra
fast
Wheel frame, damped release, with
antistatic rollers, ball bearing guided
Special frame, color
Load capacity 150kg
●
Load capacity 200kg
Load capacity 250kg
Cushion
2
Cushion width 65 cm
●
Cushion width 80 cm
○
Imitation leather, 6cm, durable composite
structure
●
Imitation leather, 8cm, Comfort de Luxe
padding, extra soft
○
Head section incl. nose slot adjustable with
GDF, +30° / -90°.
●
Head section 3 pcs. incl. nose slot
adjustable with GDF, +30° / -40°, arm
sections adjustable up to 18cm
○
Nose slot insert
●
Headboard Pro Comfort, one-piece with 3
different inserts. Inserts
Lying surface 2 parts
●
Lying surface 4 parts
○
lying surface 3 parts, head part, foot part
adjustable with GDF, +85°.
Lying surface 5 parts, head part, foot part
adjustable with GDF, +85°.
lying surface 5 parts, head part, foot part
adjustable with GDF, +85°, roof kyphosis
with GDF
Lying surface widening due to swivel-swivel
armrest upholstery
Upholstery colors
37
Special upholstery, motif, embroidery
○
Paper roll holder
○
Lounger protective cover 600x400
○
● Standard equipment
○ optional

220822SCHUPP GmbH & Co. KG
72280 Dornstetten-Aach | Glattal Straße 78 | T: +49 (0) 7443/243-0 | F: +49 (0) 7443/243-255 | info@schupp-gmbh.de | www.schupp.shop
15
Trendino
MAINTENANCE & CARE
6.8 Accessories
Couch type
Item no:
TRENDINO
Item no:
139235
Manual switch L
443125
●
Foot switch + floor L
443123
●
Gas spring headboard
402099
●
Power cable PVC
Linak
423999
●
Protective cover
600x400
190999
●
● Accessories for couch type
7.0 Spare and wear parts
Repairs must not affect the safety of the couch. Therefore, they may only be carried out by the manufacturer or
persons authorized for this purpose. Only original spare parts from the manufacturer may be used for repairs.
If you notice any defects or wear on your therapy couch, please inform us using the contact details provided (see last
page). This also applies to type plates and labels.
When contacting us, always state the couch type, the article number and the serial number. This data can be found on
the type plate on the frame of the foot section end of the couch (see chapter 10).
Our service staff will be happy to assist you in defining the fault and in selecting the correct spare and wear parts such
as: Arm, foot, head or body pads, swivel castors, foot plugs, arm section guides as well as operating parts for the
electric height adjustment.
8.0 Service life, maintenance and care
The therapy couch you have purchased is designed for an extremely long service life. The limiting factor here is the
drive, which is designed for a lifetime of max. 10 years or 100000 cycles. After this time, depending on the condition of
the table, the drive can be replaced by our service department, provided regular inspection and maintenance.
8.1 Maintenance
In order to ensure the safety of the patient and user and to protect the therapy couch from damage, expert use and
regular maintenance in accordance with DIN 31051 are required. We recommend that you have the table serviced
every 1-2 years by professionally trained Schupp employees or by authorised specialist dealers. As part of the
maintenance, all prescribed safety checks are carried out in accordance with current safety regulations:
►Testing of medical technology according to DIN EN 62353 (VDE 0751-1), annually and in case of
change
►BetriebSichV (Ordinance on Industrial Safety and Health)
►MPBetreibV (Medical Devices Operator Ordinance)
►DGUV (German Social Accident Insurance); Regulation 3 (formerly BGV A3)
carried out.
►Electrical voltage - Keep your distance!
►Before carrying out any maintenance work, disconnect the couches with electric motor from the
power supply by removing the power plug of the motor connection cable from the power socket!
Warning.

220822SCHUPP GmbH & Co. KG
72280 Dornstetten-Aach | Glattal Straße 78 | T: +49 (0) 7443/243-0 | F: +49 (0) 7443/243-255 | info@schupp-gmbh.de | www.schupp.shop
16
Trendino
MAINTENANCE & CARE
8.2 Personal precautions
To maintain functionality and comfort, it is recommended to check the therapy couch regularly by visual inspection (at
least 4 times per year). You can take some precautions yourself to prolong the life of the couch. These are:
Guides and joints
The therapy couch you have purchased is designed for an extremely long service life. The guides
and joints are therefore low-maintenance. If the height adjustment is not used for a longer period of
time or after very intensive use, the proper function should be checked regularly. If necessary, the
joints can be lightly re-oiled with a fine-mechanical oil. Also visually check the screw connections,
weld seams, locking pins and bolts for completeness and mechanical integrity.
Couch drive
The electric motor is maintenance-free and does not need to be oiled or greased. The piston rod of
the drive should be regularly cleaned of dust with a soft cloth.
When not in use for a longer period of time, disconnect the electric motor / therapy bed from the
mains supply.
Power cable
Check the power cord regularly. If it is damaged, it must be replaced immediately. Please contact
the Schupp service department for this purpose.
Wheels
To ensure smooth running of the wheel frame or wheels, the axles of the wheels can be lightly re-
oiled with a fine mechanical oil if necessary.
►Electrical voltage - Keep your distance!
►Before carrying out any maintenance work, disconnect the couch with electric motor from the power
supply by removing the mains plug of the motor connection cable from the mains socket!
►Make sure that no water or moisture enters the plugs or plug connections during cleaning.
Warning.
►Regular inspection/maintenance and compliance with technical tests do not relieve the user of his
responsibility for the safety of the patient and the proper functioning of his treatment systems!
Caution.
►No tools are to be used during the visual inspection. In case of any abnormalities, inform the
Service Department immediately!

220822SCHUPP GmbH & Co. KG
72280 Dornstetten-Aach | Glattal Straße 78 | T: +49 (0) 7443/243-0 | F: +49 (0) 7443/243-255 | info@schupp-gmbh.de | www.schupp.shop
17
Trendino
MAINTENANCE & CARE
8.3 Care and cleaning
The upholstery of the therapy couch must be treated very gently despite the highest leather quality. The surfaces
should be cleaned or disinfected after each treatment.
In case of light soiling, clean the upholstery immediately with a soft cloth, a household sponge or a soft household
brush in combination with lukewarm soapy water. The use of a commercially available microfiber cloth is particularly
recommended.
For heavy soiling, soapy water can be used in concentrated form. Repeat brushing if necessary. After cleaning, wipe
with clean water until there is no more cleaner on the artificial leather.
Moisture residues must be dried off with a soft cloth!
To avoid stains penetrating the material, they should be removed immediately. The use of abrasive cleaning agents
(scouring agents) or solvents (nitro, acetone, etc.) will damage and after a short time lead to brittleness and
discolouration of the artificial leather surface.
► The upholstery must not be cleaned with alcoholic or chemical solvents, e.g. thinner, benzine or
similar!
► Do not use chlorides, synthetic cleaning agents, wax polishes or aerosol sprays for cleaning.
Warning.
►Always disconnect the power plug before cleaning!
►Electrical voltage - never clean the therapy couch with water jets!
Warning.
Cleaning and disinfection of the couch, including the upholstery, may only be carried out with a
suitable and approved disinfectant cleaner
Discoloration - The manufacturer is not responsible for material discoloration of the synthetic
leather due to contact with denim or other textiles.
8.4 Surface disinfection
The cleaning and disinfection of the couch, including the upholstery, may only be performed with suitable disinfectant
cleaners (observe application concentration!).
The currently released disinfectants / disinfectant cleaners vary depending on the manufacturer and leather type. Our
sales and/or service staff can provide you with the latest release documents if required. You can also call up the
current status online:
[Plata, Pandoria Plus]
[AKV_Amalfi]

220822SCHUPP GmbH & Co. KG
72280 Dornstetten-Aach | Glattal Straße 78 | T: +49 (0) 7443/243-0 | F: +49 (0) 7443/243-255 | info@schupp-gmbh.de | www.schupp.shop
18
Trendino
TROUBLESHOOTING
9.0 Help with malfunctions
Symptoms
Possible causes
Measures
Pads cannot be aligned
mechanically
If the couch has not been used
for a long time, the supporting
gas pressure springs may be
stiff.
Gas strut defective.
In this case, the lever of the gas pressure spring
must be pulled completely upwards. Use the
other hand to move the lying surface completely
up and down several times.
Exchange gas strut
No function of the couch drive
Motor is not connected to the
power supply
Fuse is defective
Cables are damaged
Check that the couch is properly connected to a
230V socket and that this socket is live.
Check fuse in house distribution box.
Check the power cable for damage.
Check the supply line of the hand or foot switch
for damage. Check whether the connecting
cables of the switching elements are correctly
plugged into the motor.
Engine sounds
Contact the Schupp service department (see last
page).
No or limited function Hand or
foot switch
Wear
Pollution
Switching elements defective
Exchange hand / foot switch
Open the controller box
By pressing both housing tabs simultaneously,
the housing cover can be swung upwards.
Disconnect the plug of the hand or foot switch
and reconnect the plug of the new switching
element. Pay attention to the plug coding.

220822SCHUPP GmbH & Co. KG
72280 Dornstetten-Aach | Glattal Straße 78 | T: +49 (0) 7443/243-0 | F: +49 (0) 7443/243-255 | info@schupp-gmbh.de | www.schupp.shop
19
Trendino
TYPE SHIELD
Symptoms
Possible causes
Measures
No function of the electric
drive
Activated motor protection due to
overheating
If the motor is operated for too long without
interruption (> 1 minute), the internal protective
device can switch off the drive when the couch is
under maximum load. Wait until the drive has
cooled down again.
Problem is not fixable
unknown
Contact the Schupp service department (see last
page).
Be sure to pay attention to the following features
of your therapy couch:
Type, part number
Serial number
This data can be found on the type plate on the
frame of the foot section end of the couch.
10.0 Nameplate
Applied symbols:
Medical device
Item number
Type no. - Date of manufacture -Serial
number
Single Registration Number
Unique Device Identification
Manufacturer, address
Attention ! Warnings present
Read operating instructions carefully,
observe warnings
Product for internal use only
CE Conformity marking (see chapter 13.0)
Product must not be disposed of with
household waste
Product compliant according to ROHS II
Directives 2011/65/EU

220822SCHUPP GmbH & Co. KG
72280 Dornstetten-Aach | Glattal Straße 78 | T: +49 (0) 7443/243-0 | F: +49 (0) 7443/243-255 | info@schupp-gmbh.de | www.schupp.shop
20
Trendino
TECHNICAL DATA
11.0 Disposal
Do not dispose of this device with normal household waste !
To ensure environmentally friendly disposal of the therapy couch, optional assemblies and spare parts at the end of their
service life or after decommissioning, please contact the Schupp service department. We will be happy to help you
dispose of your therapy couch.
If you dispose of the couch yourself the following information is important for you:
Couch frame , is made of lacquered steel
Upholstery parts, consist of a wooden board (MDF), composite foam, PU foam, imitation leather covering
Drive, consists of electric motor, hydraulic cylinder, gas springs
The resulting waste must be recycled or disposed of in a manner that is safe for people and the environment(*). Please
inform yourself in principle about the regulations applicable in your country or also about the local regulations for the
disposal of old devices and accessories.
(*) Watch out for physical hazards such as sharp edges during disassembly / disposal.
12.0 Technical data
Medical Devices
class
EU MDR
Class 1 medical device
Dimension couch
surface
Length
200 cm
Length
Headboard
upholstery
43 cm
Length
Body part cushion
153 cm
Wide
65 , 70 , 80cm
Reference
Faux leather
Base plate
Wood fiberboard P2 - 19 mm E1 [EN13986:2004, A1:2015]
Upholstery:
Standard
6 cm cushion thickness
Foam composite, padded edges
Upholstery:
Comfort de luxe
8 cm cushion thickness
Foam composite, padded edges
The 2 part lying
surface
includes
Body part cushion
1 part
Headboard
upholstery
1 part
Nose slit opening and cover
Manually adjustable with gas spring
Adjustment range
+30° to -90