VALES & HILLS BIOMEDICAL TECH iCV200S User manual

ECG Acquisition
ECG Acquisition
Systems
User Manua
l
iCV200S
Systems
l

No
:
iCV200S User Manual
User Manual
ECG Acquisition Systems
Model: iCV200S
:
VH-JS-08(63F)
Date: 2019.09.0
Software Version: V3.4
Cardio View
®
Family
Date: 2019.09.0
6

Manufacturer information
Company name:
VALES & HILLS BIOMEDICA
Address: Bldg. 46-
1 No.2 North Street Jingyuan, Beijing Economic
Development Area, 100176 Beijing, China
Tel: 0086-010-
51665548
Fax: 0086-010-
67856343
E-mail:
sales@vhmedical.com
Website:
http://www.vh
CARDIO SYSTEMES 6 rue Descartes 78370 PLAISIR
iCV200S User Manual
Manufacturer information
:
VALES & HILLS BIOMEDICA
L
TECH. LTD.
1 No.2 North Street Jingyuan, Beijing Economic
Development Area, 100176 Beijing, China
51665548
67856343
sales@vhmedical.com
http://www.vh
ECG.com
CARDIO SYSTEMES 6 rue Descartes 78370 PLAISIR
Cardio View
®
Family
TECH. LTD.
1 No.2 North Street Jingyuan, Beijing Economic
-technological
CARDIO SYSTEMES 6 rue Descartes 78370 PLAISIR
France

Table of Contents
Preface
................................
1. iCV200S
................................
1.1 About iCV200S
................................
1.2 List of packing
................................
1.3 Accessories
................................
2. Installation
................................
2.1 Hardware Connection
2.2
Download vhECG Pro from Apple App Store
3. Electrodes Hookup
................................
3.1 Hookup
................................
3.2 Electrode connection and position
4. Getting Start
................................
4.1 Recorder Setting
................................
4.2 Communication between Recorder and iPad
4.3 Communication between Recorder and Patient
5. Basic operation of vhECG Pro
5.1 Settings for vhECG Pro
5.2 Recording
................................
5.3 Preview
................................
5.4 Operation after reviewing
5.5 Confirm vhECG
................................
5.6 HIS search and patients information management
5.7 Cloud
................................
6. Cleaning and Maintenance
7. Guidance and Manufacture’s Declarations
Annex I: iCV200S ECG Recorder Technical Specifications
Annex II: Definitions and Rules for ECG Measurements
iCV200S User Manual
................................
................................................................
................................
................................
................................
................................
................................
................................
................................
................................
................................
................................
................................
................................
................................
................................
................................
................................
2.1 Hardware Connection
................................................................
................................
Download vhECG Pro from Apple App Store
................................
................................
................................
................................
................................
................................
................................
................................
3.2 Electrode connection and position
................................
................................
................................
................................
................................
................................
................................
................................
4.2 Communication between Recorder and iPad
................................
................................
4.3 Communication between Recorder and Patient
................................
................................
5. Basic operation of vhECG Pro
................................................................
................................
5.1 Settings for vhECG Pro
................................................................
................................
................................
................................
................................
................................
................................
................................
5.4 Operation after reviewing
................................................................
................................
................................
................................
................................
5.6 HIS search and patients information management
................................
................................
................................
................................
6. Cleaning and Maintenance
................................................................
................................
7. Guidance and Manufacture’s Declarations
................................
................................
Annex I: iCV200S ECG Recorder Technical Specifications
................................
Annex II: Definitions and Rules for ECG Measurements
................................
Cardio View
®
Family
................................
......................... 1
................................
.................................................. 6
................................
................................... 6
................................
.................................... 6
................................
......................................... 7
................................
................................................ 8
................................
.......................... 8
................................
....................... 9
................................
................................... 10
................................
............................................. 10
................................
...................................... 11
................................
............................................ 12
................................
................................ 12
................................
........................ 12
................................
.................... 13
................................
.................. 14
................................
..................... 15
................................
......................................... 18
................................
............................................ 24
................................
.................. 30
................................
................................ 40
................................
................................................ 43
................................
............................................... 46
................................
........................ 52
................................
................................. 54
................................
........................................... 56
................................
................................................ 57

Version: A/1
Preface
Copyright
Copyright© 2014,
Vales and Hills (V&H)
reproduce or duplicate this manual or any part thereof, in any form, without prior written permission
from V&H, V&H assumes no responsibility for any injury to anyone, or for any illegal or improper use
of the product, that
may result from failure to use this product in accordance with the instructions,
cautions, warnings, or statement of intended use published in this manual. Unauthorized copying of this
publication may not only infringe copyright but also reduce the abilit
up-to-
date information to users and operators alike.
The information contained in this manual is subject to change without notice. All changes will be in
compliance with regulations governing manufacture of medical equipmen
General Description and Intended Use
iCV200S
is a portable ECG system
iCV200S 12-lead
patient cable
vhECG Pro App.
The system is designed and manufactured by V&H for patient ECG recording with
automatic measurements
and interpretations.
This product is not intended for use as sustain or support life devices.
This product is not intended for introcardiac use.
This prod
uct is not intended for use in operating room or ICU
This product is not intended for
ECG Acquisition Systems will be used in the professional healthcare facility environment and the
product
is intended to provide reference for medical diagnosis,
diagnosis clinicians.
Precautions
Radio Frequency (RF) interference between the
existing RF transmitting or receiving equipment at the installation site, including electrosur
equipment, in close proximity to the electrocardiograph should be evaluated before the equipment is
operated as they may seriously degrade performance.
The iCV200S
is susceptible to interference from RF energy sources (lowered RF immunity) which
exceed the IEC 60601-1-
2 limits, such as power line bursts, other medical equipment, cellular products,
information technology equipment and radio/television transmission.
To reduce EMC interference the cardiograph shall be separated from the emitting source
possible. If assistance is needed, call V&H service.
Artifact on the ECG caused by electromagnetic interference should be evaluated by a physician or
physician authorized personnel to determine if it will negatively impact diagnosis or treatm
iCV200S User Manual
Vales and Hills (V&H). All rights reserved. N
reproduce or duplicate this manual or any part thereof, in any form, without prior written permission
from V&H, V&H assumes no responsibility for any injury to anyone, or for any illegal or improper use
may result from failure to use this product in accordance with the instructions,
cautions, warnings, or statement of intended use published in this manual. Unauthorized copying of this
publication may not only infringe copyright but also reduce the ability of V&H to provide accurate and
date information to users and operators alike.
The information contained in this manual is subject to change without notice. All changes will be in
compliance with regulations governing manufacture of medical equipmen
General Description and Intended Use
is a portable ECG system
with CardioView family
. It includes a data acquisition recorder and
patient cable
. The ECG waveform can be
displayed
The system is designed and manufactured by V&H for patient ECG recording with
and interpretations.
This product is not intended for use as sustain or support life devices.
This product is not intended for introcardiac use.
uct is not intended for use in operating room or ICU
.
This product is not intended for
infant.
ECG Acquisition Systems will be used in the professional healthcare facility environment and the
is intended to provide reference for medical diagnosis, not intended for a replacement of
Radio Frequency (RF) interference between the
iCV200S
recorder or electrocardiograph and any
existing RF transmitting or receiving equipment at the installation site, including electrosur
equipment, in close proximity to the electrocardiograph should be evaluated before the equipment is
operated as they may seriously degrade performance.
is susceptible to interference from RF energy sources (lowered RF immunity) which
2 limits, such as power line bursts, other medical equipment, cellular products,
information technology equipment and radio/television transmission.
To reduce EMC interference the cardiograph shall be separated from the emitting source
possible. If assistance is needed, call V&H service.
Artifact on the ECG caused by electromagnetic interference should be evaluated by a physician or
physician authorized personnel to determine if it will negatively impact diagnosis or treatm
Cardio View
®
Family
Page
1 of 58
. All rights reserved. N
o one is permitted to translate,
reproduce or duplicate this manual or any part thereof, in any form, without prior written permission
from V&H, V&H assumes no responsibility for any injury to anyone, or for any illegal or improper use
may result from failure to use this product in accordance with the instructions,
cautions, warnings, or statement of intended use published in this manual. Unauthorized copying of this
y of V&H to provide accurate and
The information contained in this manual is subject to change without notice. All changes will be in
compliance with regulations governing manufacture of medical equipmen
t.
. It includes a data acquisition recorder and
displayed
on an iOS equipment installed a
The system is designed and manufactured by V&H for patient ECG recording with
This product is not intended for use as sustain or support life devices.
ECG Acquisition Systems will be used in the professional healthcare facility environment and the
not intended for a replacement of
recorder or electrocardiograph and any
existing RF transmitting or receiving equipment at the installation site, including electrosur
gical
equipment, in close proximity to the electrocardiograph should be evaluated before the equipment is
is susceptible to interference from RF energy sources (lowered RF immunity) which
2 limits, such as power line bursts, other medical equipment, cellular products,
To reduce EMC interference the cardiograph shall be separated from the emitting source
s as much as
Artifact on the ECG caused by electromagnetic interference should be evaluated by a physician or
physician authorized personnel to determine if it will negatively impact diagnosis or treatm
ent.

Version: A/1
Like all electronic devices, this electrocardiograph is susceptible to electrostatic discharge (ESD). ESD
typically occurs when electrostatic energy is transferred to the patient, the electrodes, or the
electrocardiograph. ESD may result in ECG artif
display or on the printed report. When ESD occurs, the electrocardiograph’s interpretation may be
inconsistent with the physician’s interpretation.
V&H assumes no liability for failures resulting from
series and any RF or ESD generating equipment when these levels exceed those established by
applicable standard.
Warnings
The electrocardiograph has not been designed for use with high frequency (HF) surgical
and does
not protect against hazards to the p
Ensure the location of the electrode and associated cables provides maximum separation away from
all sources of HF energy. The best way to ensure patient saf
ele
ctrodes and cables from the patient when exposed to HF energy.
Other medical equipment, including but no
pacemakers, and other simulators, may be used simultaneously with the electrocardiograph.
However, suc
h devices may disturb the electrocardiograph signal.
Portable RF communications equipment including peripherals such as antenna cables and external
antennas should be used no closer than 30 cm (12 inches) to any part of the device including cables
specifie
d by the manufacturer. Otherwise degradation of the performance of this equipment could
result.
This equipment has been tested and found to comply with the limits for a Class B digital device,
pursuant to part 15 of the FCC Rules. These limits are designed
against harmful interference in a residential installation. This equipment generates uses and can
radiate radio frequency energy and, if not installed and used in accordance with the instructions,
may cause harmful interfe
interference will not occur in a particular installation. If this equipment does cause harmful
interference to radio or television reception, which can be determined by turning the equipmen
and on, the user is encouraged to try to correct the interference by one or more of the following
measures:
-
Reorient or relocate the receiving antenna.
-
Increase the separation between the equipment and receiver.
-
Connect the equipment int
connected.
-
Consult the dealer or an experienced radio/TV technician for help
This device complies with Part 15 of the FCC Rules.
approved by the pa
rty responsible for compliance
equipment.
Operation is subject to the following two conditions:
(1) T
his device may not cause ha
(2) T
his device must acc
undesired operation.
iCV200S User Manual
Like all electronic devices, this electrocardiograph is susceptible to electrostatic discharge (ESD). ESD
typically occurs when electrostatic energy is transferred to the patient, the electrodes, or the
electrocardiograph. ESD may result in ECG artifact that may appear as narrow spikes on the cardiograph
display or on the printed report. When ESD occurs, the electrocardiograph’s interpretation may be
inconsistent with the physician’s interpretation.
V&H assumes no liability for failures resulting from
RF or ESD interference between V&H CardioView
series and any RF or ESD generating equipment when these levels exceed those established by
The electrocardiograph has not been designed for use with high frequency (HF) surgical
not protect against hazards to the p
atient.
Ensure the location of the electrode and associated cables provides maximum separation away from
all sources of HF energy. The best way to ensure patient saf
ctrodes and cables from the patient when exposed to HF energy.
Other medical equipment, including but no
t
limited to defibrillators, ultrasound machines,
pacemakers, and other simulators, may be used simultaneously with the electrocardiograph.
h devices may disturb the electrocardiograph signal.
Portable RF communications equipment including peripherals such as antenna cables and external
antennas should be used no closer than 30 cm (12 inches) to any part of the device including cables
d by the manufacturer. Otherwise degradation of the performance of this equipment could
This equipment has been tested and found to comply with the limits for a Class B digital device,
pursuant to part 15 of the FCC Rules. These limits are designed
against harmful interference in a residential installation. This equipment generates uses and can
radiate radio frequency energy and, if not installed and used in accordance with the instructions,
may cause harmful interference to radio communications. However, there is no guarantee that
interference will not occur in a particular installation. If this equipment does cause harmful
interference to radio or television reception, which can be determined by turning the equipmen
and on, the user is encouraged to try to correct the interference by one or more of the following
Reorient or relocate the receiving antenna.
Increase the separation between the equipment and receiver.
Connect the equipment into an outlet on a circuit different from that to which the receiver is
Consult the dealer or an experienced radio/TV technician for help
This device complies with Part 15 of the FCC Rules. Changes or modifications not expressly
rty responsible for compliance could void the user's authority to operate the
Operation is subject to the following two conditions:
his device may not cause ha
rmful interference, and
his device must acc
ept any interference received,
including interference that may cause
Cardio View
®
Family
Page
2 of 58
Like all electronic devices, this electrocardiograph is susceptible to electrostatic discharge (ESD). ESD
typically occurs when electrostatic energy is transferred to the patient, the electrodes, or the
act that may appear as narrow spikes on the cardiograph
display or on the printed report. When ESD occurs, the electrocardiograph’s interpretation may be
RF or ESD interference between V&H CardioView
series and any RF or ESD generating equipment when these levels exceed those established by
The electrocardiograph has not been designed for use with high frequency (HF) surgical
equipment
Ensure the location of the electrode and associated cables provides maximum separation away from
all sources of HF energy. The best way to ensure patient saf
ety is to completely remove all
ctrodes and cables from the patient when exposed to HF energy.
limited to defibrillators, ultrasound machines,
pacemakers, and other simulators, may be used simultaneously with the electrocardiograph.
Portable RF communications equipment including peripherals such as antenna cables and external
antennas should be used no closer than 30 cm (12 inches) to any part of the device including cables
d by the manufacturer. Otherwise degradation of the performance of this equipment could
This equipment has been tested and found to comply with the limits for a Class B digital device,
pursuant to part 15 of the FCC Rules. These limits are designed
to provide reasonable protection
against harmful interference in a residential installation. This equipment generates uses and can
radiate radio frequency energy and, if not installed and used in accordance with the instructions,
rence to radio communications. However, there is no guarantee that
interference will not occur in a particular installation. If this equipment does cause harmful
interference to radio or television reception, which can be determined by turning the equipmen
t off
and on, the user is encouraged to try to correct the interference by one or more of the following
o an outlet on a circuit different from that to which the receiver is
Consult the dealer or an experienced radio/TV technician for help
.
Changes or modifications not expressly
could void the user's authority to operate the
including interference that may cause

Version: A/1
The conductor parts of the electrode of ECG Acquisition Box
neutral electrode) should not contact with ot
There wi
ll be potential risks when using the product along with installation of cardiac rhythm or
other electrical stimulator at the same time. Pay full attention to changes of the ECG to ensure
safety and to ensure that the leakage current is within the allowable
Do not use the product with Pacemaker.
During defibrillation,
Unpolarized electrode
Warranty and Service
V&H provides a one-
year
products under warranty must be performed or approved by V&H. Unauthorized repairs will void the
warranty. In addition, whether or not covered under warranty, any product repair shall ex
performed by V&H certified service per
Limited warranty statement
The warranty will
not cover the damages caused by:
1) Use
or maintenance contrary to labeled instructions
2) A
lteration or repair by anyone not authorized by V&H
3) Accidents.
Product return
If the product has to be returned for service, V&H recommends these parking instructions:
Wherever possible, use the original shipping carton and packing materials.
Including a packing list.
Including a problem description report
It is recommended that all returned goods be insured. Claim for loss or damage to the product must be
initialed by the sender.
Unit disposal Instructions
Disposal of battery and the unit after the end of device life in municipally approved recycling a
life of device is 5 years.
iCV200S User Manual
The conductor parts of the electrode of ECG Acquisition Box
neutral electrode) should not contact with ot
her conductor parts,
including
ll be potential risks when using the product along with installation of cardiac rhythm or
other electrical stimulator at the same time. Pay full attention to changes of the ECG to ensure
safety and to ensure that the leakage current is within the allowable
Do not use the product with Pacemaker.
do not touch patients, ele
ctrode and patient cable.
Unpolarized electrode
(silver/silver chloridoe) can be used for def
i
Warranty and Service
year
warranty and life-time service to all Car
dioView series products. Repairs on
products under warranty must be performed or approved by V&H. Unauthorized repairs will void the
warranty. In addition, whether or not covered under warranty, any product repair shall ex
performed by V&H certified service per
.
Limited warranty statement
not cover the damages caused by:
or maintenance contrary to labeled instructions
;
lteration or repair by anyone not authorized by V&H
;
If the product has to be returned for service, V&H recommends these parking instructions:
Wherever possible, use the original shipping carton and packing materials.
Including a packing list.
Including a problem description report
.
It is recommended that all returned goods be insured. Claim for loss or damage to the product must be
Unit disposal Instructions
Disposal of battery and the unit after the end of device life in municipally approved recycling a
Cardio View
®
Family
Page
3 of 58
and its connector (including the
including
the earth.
ll be potential risks when using the product along with installation of cardiac rhythm or
other electrical stimulator at the same time. Pay full attention to changes of the ECG to ensure
safety and to ensure that the leakage current is within the allowable
range.
ctrode and patient cable.
i
brillation.
dioView series products. Repairs on
products under warranty must be performed or approved by V&H. Unauthorized repairs will void the
warranty. In addition, whether or not covered under warranty, any product repair shall ex
clusively be
If the product has to be returned for service, V&H recommends these parking instructions:
Wherever possible, use the original shipping carton and packing materials.
It is recommended that all returned goods be insured. Claim for loss or damage to the product must be
Disposal of battery and the unit after the end of device life in municipally approved recycling a
rea. The

Version: A/1
Classification
Degree of protection against electrical
shock
The power supply of ECG Acquisition box
Degree of protection against harmful
ingress of water
Degree of safety of application in the
presence of a flammable anesthetic mixture
with air or with oxygen or nitrous oxide
Method(s) of sterilization or disinfection
recommended by the manufacturer
Mode of operation
The defibrillation-
proof recovery time
iCV200S User Manual
Degree of protection against electrical
Type CF, defibrillation
The power supply of ECG Acquisition box
2×AAA Battery
Degree of protection against harmful
Enclosed
equipment without protection against
ingress of water
Degree of safety of application in the
presence of a flammable anesthetic mixture
with air or with oxygen or nitrous oxide
Not suitable for use in the presence of a flammable
anesthetic mixture with a
nitrous oxide.
Method(s) of sterilization or disinfection
recommended by the manufacturer
Sterilization: not applicable
disinfection: see user manual
Continuous operation
proof recovery time
5s
Cardio View
®
Family
Page
4 of 58
Type CF, defibrillation
-proof
equipment without protection against
Not suitable for use in the presence of a flammable
anesthetic mixture with a
ir or with oxygen or
Sterilization: not applicable
disinfection: see user manual
Continuous operation

Version: A/1
List of Symbols
Symbols
SN
IPX0
Warning:
Warning statements describe conditions or actions that can result in personal injury or lost of life.
Caution:
Warning statements describe conditions or actions that can result in damage to the equipment of
software.
Note:
Notes contain additional
iCV200S User Manual
Meaning
Caution, consult accompanying
documents.
Serial Number(xxxxxxxxxxxxxxx)
Manufacturer
Date of manufacture
Authorized representative in the European community
The CE marking with the Registration Number of the Notified
Body.
This denotes the compliance of
Directive 93/42/EEC
Waste Electrical and Electronic Equipment (WEEE) It is the
responsibility of the end user to dispose of this equipment at a
designated collection point for recycling.
DEFIBRILLATION-PROOF T
YPE CF APPLIED PART
Bluetooth
Function switch button
Keep dry
temperature limits
Indicates the range of humidity to which the medical device can be
safely exposed.
Against ingress of water with harmful effects: non
Warning statements describe conditions or actions that can result in personal injury or lost of life.
Warning statements describe conditions or actions that can result in damage to the equipment of
Notes contain additional
information on iCV200S usage.
Cardio View
®
Family
Page
5 of 58
Meaning
documents.
Authorized representative in the European community
The CE marking with the Registration Number of the Notified
This denotes the compliance of
European Medical Device
Waste Electrical and Electronic Equipment (WEEE) It is the
responsibility of the end user to dispose of this equipment at a
designated collection point for recycling.
YPE CF APPLIED PART
Indicates the range of humidity to which the medical device can be
Against ingress of water with harmful effects: non
-protected
Warning statements describe conditions or actions that can result in personal injury or lost of life.
Warning statements describe conditions or actions that can result in damage to the equipment of

Version: A/1
1. iCV200S
iCV200S
Rest ECG for i
Thank you for your purch
working with it.
If you have any comments or suggestions, we would app
Please contact support
@vh
1.1 About iCV200S
iCV200S
is a portable ECG system. It includes a data acquisition recorder
product can connect with iOS equipment such as
system is designed and manufactured by V&H
recording and analyzing patients resting ECG. This system is applicable to the heart disease analysis for
the medical treatment institution.
Fig1.1
iCV200S
1.2 List of packing
The ECG Acquisition Systems consists of the following items:
Part No.
CV001008
CV002019
CV004005
iCV200S User Manual
Rest ECG for i
OS
Thank you for your purch
ase of iCV200S Resting ECG System
and we hope that you
If you have any comments or suggestions, we would app
@vh
medical.com.
is a portable ECG system. It includes a data acquisition recorder
product can connect with iOS equipment such as
iPad, iPad-mini
and iPhone
system is designed and manufactured by V&H
, The ECG A
cquisition System is capable of sampling,
recording and analyzing patients resting ECG. This system is applicable to the heart disease analysis for
the medical treatment institution.
iCV200S
ECG Acquisition Systems connect w
ith iOS equipment
The ECG Acquisition Systems consists of the following items:
Description
ECG Acquisition Box
ECG cable (Type: ECG-
FD10X4(ZX)
User manual
Cardio View
®
Family
Page
6 of 58
and we hope that you
will enjoy
If you have any comments or suggestions, we would app
reciate hearing from you.
is a portable ECG system. It includes a data acquisition recorder
and patient cable. The
and iPhone
through Bluetooth. The
cquisition System is capable of sampling,
recording and analyzing patients resting ECG. This system is applicable to the heart disease analysis for
ith iOS equipment
QTY
1
FD10X4(ZX)
)
1
1

Version: A/1
Note:
If you have received
the above items, you can
App Store. Please see “
2.2 Downl
1.3 Accessories
Item No.
Description
CV002019
iCV200S patient cable
(Type: ECG
CV008002
Battery
Notes:
1)
If any lead wire of lead part is failure, the user is unable to replace the lead wire, please contact with
V&H.
2)
Please use the original standard patient cable which needs to be used to provide protection against the
effect of the discharge
of a cardiac defibrillator and against high
3)
If any lead wire of lead part is failure, the user can replace the standard 6511ECG cable, the ECG
cable safety requirements tested for compliance with 93/42/EEC, should have the CE certif
& Hills recommends Qingdao Bright medical, the model is M02010
attachment may reduce the safety of the products.
4) iCV200S
is powered by two AA
comm
unication between the recorder and i
before using.
If the power is low,
battery model is AAA
LR03
5)
For the sake of environmental protection, please recycle disposal of the battery.
6)
For products and accessories that are no longer used, it will be the cause of environmental pollution,
they should be pr
operly treated.
7) Not including S
tandard Patient
8)
If you find anything above is not included in the packing, please contact us by the following address:
Technical Support Dept. of Vales
Address: Bldg. 46-
1 No.2 North Street Jingyuan, Beijing Economic
CARDIO SYSTEMES, 6 rue Desc
iCV200S User Manual
the above items, you can
download
iCV200S software vhECG Pro from Apple
2.2 Download vhECG Pro from Apple App Store
Description
Intended Use
iCV200S patient cable
(Type: ECG
-FD10X4(ZX))
Cables allow connection to the electrode leads,
which adhere to
the patient for detection of ECG
signals.
Battery
Power for ECG acquisition box
If any lead wire of lead part is failure, the user is unable to replace the lead wire, please contact with
Please use the original standard patient cable which needs to be used to provide protection against the
of a cardiac defibrillator and against high
-
frequency burns
If any lead wire of lead part is failure, the user can replace the standard 6511ECG cable, the ECG
cable safety requirements tested for compliance with 93/42/EEC, should have the CE certif
& Hills recommends Qingdao Bright medical, the model is M0201044
attachment may reduce the safety of the products.
is powered by two AA
A LR03 batteries.
Insufficient of power may affect the
unication between the recorder and i
OS equipment.
Check the batteries with adequacy of power
If the power is low,
the u
ser can replace the new battery.
LR03
. The product can be used continuou
sly for at least
For the sake of environmental protection, please recycle disposal of the battery.
For products and accessories that are no longer used, it will be the cause of environmental pollution,
operly treated.
tandard Patient
Cable and Battery when we sale.
If you find anything above is not included in the packing, please contact us by the following address:
Technical Support Dept. of Vales
&
Hills Biomedical Tech. Ltd.
Tel: 0086-010-51665548
E-mail: support@vhmedical.com
Website: http://www.vhecg.com
Fax: 0086-010-67856343
1 No.2 North Street Jingyuan, Beijing Economic
-
technological Development Area,
100176 Beijing, China
CARDIO SYSTEMES, 6 rue Descartes 78370 PLAISIR, France
Cardio View
®
Family
Page
7 of 58
iCV200S software vhECG Pro from Apple
oad vhECG Pro from Apple App Store
”.
Cables allow connection to the electrode leads,
the patient for detection of ECG
Power for ECG acquisition box
If any lead wire of lead part is failure, the user is unable to replace the lead wire, please contact with
Please use the original standard patient cable which needs to be used to provide protection against the
frequency burns
.
If any lead wire of lead part is failure, the user can replace the standard 6511ECG cable, the ECG
cable safety requirements tested for compliance with 93/42/EEC, should have the CE certif
icate. Vales
44
. If use the manufacturer approval
Insufficient of power may affect the
Check the batteries with adequacy of power
ser can replace the new battery.
Vales & Hills recommends the
sly for at least
8 hours under normal use.
For the sake of environmental protection, please recycle disposal of the battery.
For products and accessories that are no longer used, it will be the cause of environmental pollution,
If you find anything above is not included in the packing, please contact us by the following address:
Hills Biomedical Tech. Ltd.
technological Development Area,
artes 78370 PLAISIR, France

Version: A/1
2. Installation
A complete iCV200S
system consists of
patient cable. T
his system can connect
section will guide you
iCV200S
2.1
Hardware Connection
iCV200S
ECG Acquisition Recorder
Notes:
1) Power Indicator: Blue
light for sufficient power
2)
Lead abnormal indicator:
Button
Power Indicator
Lead abnormal indicator
iCV200S Pa
tient Cable Connector
iCV200S User Manual
system consists of
iCV200S
data acquisition recorder and
his system can connect
iPad, iPad-mini or iPhone
with Ap
iCV200S
recorder composition and vhECG Pro
Hardware Connection
ECG Acquisition Recorder
iCV200S Recorder Frontside
iCV200S Recorder Backside
Fig 2.1 iCV200S Recorder
light for sufficient power
and g
lint for power shortage
Lead abnormal indicator:
Green light for sufficient power.
Lead abnormal indicator
15-Pin-Share-Socket
tient Cable Connector
Electrode connector
Cardio View
®
Family
Page
8 of 58
data acquisition recorder and
iCV200S 12-lead
with Ap
ple App – vhECG Pro. This
installation.
lint for power shortage
.
Battery Compartment Cover
Electrode connector

Version: A/1
iCV200S
Connecting further accessories
The
illustration on the Fig 2.2 shows
connected.
2.2 Download
vhECG Pro
iCV200S
Resting ECG System
vhECG Pro approved by
Apple
Free Download
You can download
vhECG Pro
following steps:
Step 1.
Login with your Apple ID (Settings → Store). If you do not have an Apple ID, you can creat
one with your e-
mail address
Step 2
. Entry the APP store of your iPad/iPad
Step 3
. Search “vhecg pro”;
Step 4
. Download the software of “vhecg pro”,
vhECG Pro “ ”.
Note:
If you have any questions about
Patients
outside
environment
iPad
/iPhone
iCV200S User Manual
Connecting further accessories
illustration on the Fig 2.2 shows
further accessories and the interface where they are
Fig 2.2 Operation schematic
vhECG Pro
from Apple App Store
Resting ECG System
can connect the software running on
iPad
Apple
“ ”.
vhECG Pro
to your iPad, iPad-mini or iPhone
from Apple App Store
Login with your Apple ID (Settings → Store). If you do not have an Apple ID, you can creat
mail address
;
. Entry the APP store of your iPad/iPad
-mini/iPhone;
. Search “vhecg pro”;
. Download the software of “vhecg pro”, then install it by operational guide
If you have any questions about
vhECG Pro download, please contact
s
Patients within
environment
outside
environment
/iPhone
Bluetooth
Patient
Patient cable (F-
TYPE APPLIED PART)
iCV200S ECG acquisition box
Cardio View
®
Family
Page
9 of 57
further accessories and the interface where they are
iPad
, iPad-mini or iPhone named
from Apple App Store
for free as
Login with your Apple ID (Settings → Store). If you do not have an Apple ID, you can creat
e
then install it by operational guide
and you will get
s
upport@vhmedical.com.
TYPE APPLIED PART)
iCV200S ECG acquisition box

Version: A/1
3. Electrodes Ho
okup
3.1 Hookup
iCV200S User Manual
okup
Fig 3.1 Electrode location
Cardio View
®
Family
Page
10 of 58

Version: A/1
3.2 Electrode connection and position
Precordial lead:
Coat the patient’s chest with con
correct position, then attach the lead and
Electrode clip:
Limb lead:
1) Clean the skin of the patient
2) Apply conductive paste or wipe with alcohol cotton balls
3) Attach the electrode clip, preferably on the inside of limbs
4) To prevent oxidation and poor connection, please clean the clips after use
Note:
1)
It is recommended that electrodes be placed by medical trained personnel.
2)
To obtain correct ECG record, please previously clean the skin.
iCV200S User Manual
3.2 Electrode connection and position
Coat the patient’s chest with conductive paste or wipe with alcohol cotton balls in
correct position, then attach the lead and
suck ball to the patie
nt in the order as below:
1) Clean the skin of the patient
.
2) Apply conductive paste or wipe with alcohol cotton balls
.
3) Attach the electrode clip, preferably on the inside of limbs
.
4) To prevent oxidation and poor connection, please clean the clips after use
It is recommended that electrodes be placed by medical trained personnel.
To obtain correct ECG record, please previously clean the skin.
Limb leas position:
RA: Right arm
LA: Left arm
RL: Right leg
LL: Left keg
Precordial lead position:
C1: The fourth intercostal space to the right of the sternum.
C2: The fourth intercostal space to
C3: Mid-
point between V2 and V4.
C4: The fifth intercostal space in the left clavicular midline.
C5: The fifth intercostal space in the axillary line.
C6: The fifth intercostal space in the axillary midline.
Cardio View
®
Family
Page
11 of 58
ductive paste or wipe with alcohol cotton balls in
nt in the order as below:
4) To prevent oxidation and poor connection, please clean the clips after use
.
It is recommended that electrodes be placed by medical trained personnel.
Limb leas position:
C1: The fourth intercostal space to the right of the sternum.
C2: The fourth intercostal space to
the left of the sternum.
point between V2 and V4.
C4: The fifth intercostal space in the left clavicular midline.
C5: The fifth intercostal space in the axillary line.
C6: The fifth intercostal space in the axillary midline.

Version: A/1
4. Getting Start
The communication between iPad and
iCV200S
Resting ECG System for iPad is vhECG
The running language o
f vhECG
English, that means if your iPad sets the language not belong to above selections, vhECG
English. The software
vhECG
participations in vhECG
development
4.1 Recorder Setting
Referring to 2.1 to get the <terms> explaining.
T
he communication between
Battery: Set two AAA
batteries
<Button> Control:
Turn on:
Click <button> to turn on the recorder.
Turn off:
Keep press (two seconds) <button> to turn off the recorder.
<LED (Green)> State:
Off: The recorder is off.
On: The recorder is on.
Slow Flash (on
ce two seconds
4
.2 Communication between Recorder and iPad
Step 1: Turn on the recorder.
Step 2: Click “
” on iPad to enter settings.
Step 3: Click “
Step 4: Open the vhECG software “
Connectivity” and select Bluetooth method.
iCV200S User Manual
The communication between iPad and
iCV200S
Recorder is through Bluetooth.
Resting ECG System for iPad is vhECG
Pro
. It currently runs with the following languages:
f vhECG
Pro
depends on your iPad language setting. The default language is
English, that means if your iPad sets the language not belong to above selections, vhECG
vhECG
Pro
will support more languages in progress. V&H
development
.
Referring to 2.1 to get the <terms> explaining.
he communication between
iCV200S
recorder and iPad is through Bluetooth.
batteries
into the recorder.
Click <button> to turn on the recorder.
Keep press (two seconds) <button> to turn off the recorder.
ce two seconds
): Power insufficient --
change the batteries
.2 Communication between Recorder and iPad
Step 1: Turn on the recorder.
” on iPad to enter settings.
” on iPad. To enable Bluetooth function
of
Step 4: Open the vhECG software “
”. Click “ ”
to enter vhECG settings, then click “
Connectivity” and select Bluetooth method.
Such as:
Cardio View
®
Family
Page
12 of 58
Recorder is through Bluetooth.
. It currently runs with the following languages:
depends on your iPad language setting. The default language is
English, that means if your iPad sets the language not belong to above selections, vhECG
Pro will run in
will support more languages in progress. V&H
appreciates your
recorder and iPad is through Bluetooth.
change the batteries
.
of
the iPad device.
to enter vhECG settings, then click “
ECG Device

Version: A/1
Step 5:
Come back to the home page,
When the connection
Click the round icon
Note:
The
last five digits of the device will show on the bottom of your iPad, you need to select the same
device number on the iPad with your recorder.
4.3
Communication between Recorde
Step 1: Connect the electrode on the electrode connector.
Step 2: Clean the skin.
Step 3: According to se
ction
Note:
If the connection of the lead wire and Acquisition bo
there’s any damage between the connection of the lead wire and Acquisition box, G
corresponding to the lead wire is
iCV200S User Manual
Fig 4.1 Select Bluetooth
Come back to the home page,
click “ ” on the top-
right corner of your
When the connection
successes, you will see “
Device Connected
Click the round icon
“ ” with numbers to start work.
Fig 4.2
Device connection diagram
last five digits of the device will show on the bottom of your iPad, you need to select the same
device number on the iPad with your recorder.
Communication between Recorde
r and Patient
Step 1: Connect the electrode on the electrode connector.
ction
“3.2 Electrode connection and position”.
If the connection of the lead wire and Acquisition bo
x is normal, g
reen
there’s any damage between the connection of the lead wire and Acquisition box, G
corresponding to the lead wire is
off
. Please check Patient cable and change Patient cable.
Cardio View
®
Family
Page
13 of 58
right corner of your
iPad.
Device Connected
” at the bottom of your iPad.
Device connection diagram
last five digits of the device will show on the bottom of your iPad, you need to select the same
reen
light for sufficient power. If
there’s any damage between the connection of the lead wire and Acquisition box, G
reen light for
. Please check Patient cable and change Patient cable.

Version: A/1
5.
Basic operation of vhECG Pro
vhECG Pro is an easy
-
experience. Most of the usage is the same as iOS. If you are familiar with Apple applications, there is no
p
roblem to operate vhECG Pro. This chapter will describe the basic operation of vhECG Pro, please
read it carefully.
Tap vhECG Pro icon “
System as below:
iCV200S User Manual
Basic operation of vhECG Pro
-
to-use application providing
professional
experience. Most of the usage is the same as iOS. If you are familiar with Apple applications, there is no
roblem to operate vhECG Pro. This chapter will describe the basic operation of vhECG Pro, please
”
on the desktop to run. Enter main interface
Fig 5.1 Main interface
Cardio View
®
Family
Page
14 of 58
professional
ECG tools and excellent user
experience. Most of the usage is the same as iOS. If you are familiar with Apple applications, there is no
roblem to operate vhECG Pro. This chapter will describe the basic operation of vhECG Pro, please
on the desktop to run. Enter main interface
of vhECG Resting ECG
Bluetooth
communication
Settings

Version: A/1
vhECG Pro mainly consist of
5.1 Settings for vhECG Pro
Users can customize vhECG Pro. Tap
Setting Options
Operation options
1) GENERAL Setting
Hospital Name
Input your hospital or
organization name, etc.
Doctor Name
Input the vhECG user’s name
2) LIVE MONITOR Setting
Use Demo Data
ON / OFF
Backward Record
ON / OFF
Pediatric
Mode
OFF / V1 V3 V5 /
V5R V3R V1 V3 V5 V7
Highpass
OFF / 0.05Hz / 0.08Hz / 0.1Hz
Lowpass
#1
OFF / 30Hz / 40Hz / 45Hz /
70Hz / 75Hz / 100Hz / 150Hz
AC Notch
#2
OFF / 50Hz/ 60Hz
EMG Filter
#3
ON / OFF
Settings
Hospital basic
information, filter,
recording time, hospital
info, report format and
other related parameters
iCV200S User Manual
vhECG Pro mainly consist of
four parts: Settings, Recording, Review and Cloud.
Fig 5.2 vhECG Pro composition
5.1 Settings for vhECG Pro
Users can customize vhECG Pro. Tap
“ ”
to enter settings. It consists of
Operation options
Operation results
Input your hospital or
organization name, etc.
Group name
Input the vhECG user’s name
Be used to sign fo
2) LIVE MONITOR Setting
ON / OFF
ON: Show the demo wave;
OFF: Not show the demo wave.
ON / OFF
ON: Record last setting
OFF: Not record last setting
OFF / V1 V3 V5 /
V5R V3R V1 V3 V5 V7
OFF: Not record new baby’s ECG data;
V1 V3 V5/ V5R V3R V1 V3 V5 V7:
Record Pediatric ECG data
OFF / 0.05Hz / 0.08Hz / 0.1Hz
F
ilter out noise with higher frequency, that is,
more noise can be filtered in this mode.
OFF / 30Hz / 40Hz / 45Hz /
70Hz / 75Hz / 100Hz / 150Hz
F
ilter out noise with lower frequency, that is,
less noise can be filtered in this mode.
OFF / 50Hz/ 60Hz
Clear noise of alternati
ON / OFF
ON: Filter out interference of EMG noise;
OFF: Not filter out interference of EMG
Recording
Review
vhECG Pro
New-built case,
ECG collection,
preview
Edit, delete, upload,
print and send email for
ECG
Cardio View
®
Family
Page
15 of 58
four parts: Settings, Recording, Review and Cloud.
to enter settings. It consists of
four parts.
Operation results
Group name
Be used to sign fo
r diagnosis
ON: Show the demo wave;
OFF: Not show the demo wave.
ON: Record last setting
-time ECG data;
OFF: Not record last setting
-time ECG data.
OFF: Not record new baby’s ECG data;
V1 V3 V5/ V5R V3R V1 V3 V5 V7:
Record Pediatric ECG data
ilter out noise with higher frequency, that is,
more noise can be filtered in this mode.
ilter out noise with lower frequency, that is,
less noise can be filtered in this mode.
Clear noise of alternati
on current.
ON: Filter out interference of EMG noise;
OFF: Not filter out interference of EMG
Cloud
Edit, delete, upload,
print and send email for
View, diagnostic and
review report uploaded

Version: A/1
Setting Options
Operation options
Connect to HIS
ON / OFF
Patient Queue
Mode
#4
ON / OFF
Enable Minnesota
Code
ON / OFF
Print Grid
ON / OFF
3) COMMUNICATION Setting
Ecg Device
Connectivity
Bluetooth
Device SN Last
Digit
Number
Device ID
Numbers
Ecg Cloud / FTP
Adress
#5
Input ECG cloud or FTP address
FTP File Format
XML / Report
4) REPORT Setting
Report title
Edit the title of report
Report format
PDF / Image
Report password
Edit password
Lead Layout
6×2 / 12×1
Paper orientation
A4 Portrait / A4 Landscape/ B5
Portrait / B5 Landscape
Chest leads voltage
halved
ON
Default mail
address
#7
E
dit E
Auto Delete Report
After Sent
ON
Notes:
#1: Lowpass:
Patented by V&H.
frequency but it will severely affect ECG high frequency component such us QRS complex.
iCV200S User Manual
Operation options
Operation results
noise.
ON / OFF
ON:
Connect to HIS;
OFF: Not
connect to HIS.
ON / OFF
ON: Detect from patient menu;
OFF: Not detect from patient menu.
ON / OFF
ON
: Show the diagnostic list by the code;
OFF: Not show the diagnostic list by the
code.
ON / OFF
ON: Grid;
OFF
: No grid.
3) COMMUNICATION Setting
Bluetooth
Connect with the Recorder and receive the
ECG from the the Recorder.
Number
Edit the SN Last Digit of the Recorder.
Numbers
Edit the ID of the Recorder.
Input ECG cloud or FTP address
Transfer the ECG to Cloud or FTP system
XML / Report
XML: transfer the ECG data to FTP in XML
format
Report: transfer the ECG data to FTP in
report format
Edit the title of report
Report title
PDF / Image
ECG Email format
Edit password
Patient report
6×2 / 12×1
Report lead layout
A4 Portrait / A4 Landscape/ B5
Portrait / B5 Landscape
Report paper orientation
ON
/ OFF
ON: voltage of chest leads (V1~V6) auto
halved when above 2.5mV on report
interface.
OFF:
no halved.
dit E
-mail address Report
default recipient
ON
/ OFF
ON: Auto Delete Reports After Sent;
OFF: Retain Reports After Sent.
Patented by V&H. Muscle noise can be removed by setting Lowpa
frequency but it will severely affect ECG high frequency component such us QRS complex.
Cardio View
®
Family
Page
16 of 58
Operation results
Connect to HIS;
connect to HIS.
ON: Detect from patient menu;
OFF: Not detect from patient menu.
: Show the diagnostic list by the code;
OFF: Not show the diagnostic list by the
: No grid.
Connect with the Recorder and receive the
ECG from the the Recorder.
Edit the SN Last Digit of the Recorder.
Edit the ID of the Recorder.
Transfer the ECG to Cloud or FTP system
XML: transfer the ECG data to FTP in XML
Report: transfer the ECG data to FTP in
report format
ECG Email format
Patient report
encryption
Report lead layout
Report paper orientation
ON: voltage of chest leads (V1~V6) auto
-
halved when above 2.5mV on report
no halved.
default recipient
ON: Auto Delete Reports After Sent;
OFF: Retain Reports After Sent.
Muscle noise can be removed by setting Lowpa
ss with low cutoff
frequency but it will severely affect ECG high frequency component such us QRS complex.
Other manuals for iCV200S
1
Table of contents
Other VALES & HILLS BIOMEDICAL TECH Medical Equipment manuals
Popular Medical Equipment manuals by other brands

Stealth Products
Stealth Products Diego owner's manual

CA.MI
CA.MI NEW VAPINAL instruction manual

CONCORD
CONCORD Emerald Operator's manual

schwa-medico
schwa-medico Easytrain Myo operating instructions

Drew Technologies
Drew Technologies IMclean manual

Smart Caregiver
Smart Caregiver TamperProof Caregiver Key TL-2016R3 Installation and use instructions