
5
SC-50 Manual v3.1
c. Cap all Reaction Vials.
d. Keep out all Reaction Vials needed for MLF assay.
a. For Quick Check Mode, you will need one (1) reaction vial per wine sample.
b. For Quantify Mode you will need at least two (2) reaction vials, one (1) for the
Malic Acid Standard and one (1) per wine sample. (See Quantify Mode, pg 8, for
further information).
e. Place all remaining reaction vials in freezer until use.
f. Thoroughly rinse plastic measuring spoon with DI water.
Procedures
Biopressure Assay method:
There are two modes for using your SC-50 to assess MLF status by Biopressure assay. Both modes use
steps 1 through 18 below.
In the Quick-check Mode, you set the instrument such that if your sample has 0.1 g/L or more
of malic acid, you will see the STOP LED on the SC100/300 illuminate and the tone will sound,
alerting you that MLF is not complete. There is an optional confirmation step (see
Confirmation Procedure, page 8) that you can run after the test to verify the validity of a
result.
In the Quantify Mode, you set the instrument to higher sensitivity and run the assay with one or
more malic acid standards provided in the kit. This allows you to determine the actual malic
acid concentration in your wine samples. (See Quantify Mode Procedure on page 8).
We recommend running the Quantify Mode until you are familiar with the workings of the instrument.
Note: the following procedure is designed to be run at room temperature, between 66 and 73°F
(19 - 23°C). Results may differ if temperatures are outside this range. SeeAppendix for more
information.
1. Take out all Reaction vials needed for assay from freezer. One Reaction vial per sample is
necessary. If you will run the assay in Quantify Mode you will need at least one additional
reaction vial (See Quantify Mode Procedure on page 8). Allow Reaction vials to come to
room temperature before initiating assay. Note that you can run multiple wine samples at one
time.
2. Using the 10 mL serological pipette, measure a 10 mL wine sample and dispense into the 100
mL plastic beaker or other suitable container. (Serological pipette can be rinsed with DI water
for re-use; remove cotton plug if necessary)
3. Place wine sample in microwave and boil for 90 seconds at medium power. Actual time
depends on your microwave oven’s power; the object is to get the wine sample to a gentle boil
for 60 seconds without having the wine splash out of beaker. If you are unsure about the power
of your microwave, we suggest you experiment with a couple of wine samples beforehand to
determine what the best settings are for your microwave oven. You may also bring wine to a
gentle boil on a stove top for 60 seconds.