ViscoTec PreeFlow eco-CONTROL EC200 DUO Manual

Operating and
maintenance
instructions
eco-CONTROL
EC200 DUO
Dosing system
Translation of the German original operating manual

Item number 20202 print 21.2.12

Dosing system eco-CONTROL EC200 DUO
Operating and maintenance instructions
Copyright © Version 1.4 3 / 43
1Introduction 5
2Features 6
3Scope of supply 6
3.1 Mixer set - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 6
4Safety 7
4.1 Informal safety measures - - - - - - - - - - - - - - - - - - - - - - - 7
4.2 Use of symbols - - - - - - - - - - - - - - - - - - - - - - - - - - - - 7
4.3 Correct use, warranty - - - - - - - - - - - - - - - - - - - - - - - - - 7
4.4 Qualifications of the operators and maintenance personnel - - - - - - 8
4.5 Organisational measures - - - - - - - - - - - - - - - - - - - - - - - 8
4.6 Preparation for commissioning - visual inspection - - - - - - - - - - - 8
4.7 Preventing material damage to the motors of the 2K-dispenser - - - - 8
4.8 Sources of danger due to Dosing system - - - - - - - - - - - - - - - 9
5Operation 10
5.1 Displays and controls - - - - - - - - - - - - - - - - - - - - - - - - - 10
5.2 Connections - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 11
5.3 Symbols on the screen - - - - - - - - - - - - - - - - - - - - - - - - 12
5.4 Functional and operational schematic - - - - - - - - - - - - - - - - - 13
5.5 Starting up for the first time - - - - - - - - - - - - - - - - - - - - - - 14
5.5.1 Save the calibration values of the dosing pressure sensors 14
5.6 Starting up - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 15
5.6.1 Important settings 15
5.7 Shutting down - - - - - - - - - - - - - - - - - - - - - - - - - - - - 15
5.8 The dosing programs and their areas of application - - - - - - - - - - 15
5.8.1 Dosing programs 16
5.8.2 Operating modes 16
5.9 Calibrating - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 17
5.9.1 Automatic calibration 17
5.9.2 Calibration manual 18
Contents

Dosing system eco-CONTROL EC200 DUO
Operating and maintenance instructions
4 / 43 Version 1.4 Copyright ©
5.10 Settings - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 19
5.10.1 Mixing ratio 19
5.10.2 Mixer size 19
5.10.3 Setting the pot life, flush the mixer 19
5.10.4 Dosing quantity 20
5.10.5 Dosing time 20
5.10.6 Flow rate 20
5.10.9 Dosing pressure monitor 21
5.10.7 Sucking back 21
5.10.8 Dosing time 21
5.10.10 Tolerance value primary pressure monitoring 22
5.10.11 Password protection 22
5.10.12 Analogue Input 25
5.10.14 System and error messages 26
5.10.13 Time and date 26
5.11 Creating dosing programs - - - - - - - - - - - - - - - - - - - - - - - 27
5.12 Selecting the dosing program - - - - - - - - - - - - - - - - - - - - - 27
5.13 Copying the dosing program - - - - - - - - - - - - - - - - - - - - - 28
5.14 Changing a dosing program - - - - - - - - - - - - - - - - - - - - - - 29
5.15 Dosing, selected directly without dosing program - - - - - - - - - - - 30
5.16 Dosing, using a dosing program - - - - - - - - - - - - - - - - - - - - 31
5.17 Monitoring / displaying dosing pressure - - - - - - - - - - - - - - - - 31
5.18 Manual flush - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 31
5.19 Bleeding the 2K-dispenser after filling / refilling it and after cleaning - - 32
5.20 Error messages - - - - - - - - - - - - - - - - - - - - - - - - - - - - 32
5.21 Clearing error messages - - - - - - - - - - - - - - - - - - - - - - - 32
5.22 Error message dosing pressure monitor error - - - - - - - - - - - - - 32
5.23 Service - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 33
5.23.1 Operating information 33
5.23.2 Formatting the memory card 33
6Accessories / spare parts 33
7Troubleshooting / Maintenance 34
8Technical data 35
8.1 Dosing control - - - - - - - - - - - - - - - - - - - - - - - - - - - - 35
8.2 Dispenser - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 35
8.3 Interface description - - - - - - - - - - - - - - - - - - - - - - - - - 36
8.3.1 System plugs 36
8.3.2 Logical links of the outputs 38
9Disposal 42
10 EC-Declaration of Conformity 43

Dosing system eco-CONTROL EC200 DUO
Operating and maintenance instructions
Copyright © Version 1.4 5 / 43
1Introduction
Dear customer,
we congratulate you on buying your Dosing system. We are certain that it will fully meet
your requirements. We wish you trouble-free and successful operation.
The dosing system eco-CONTROL EC200-K is made up of the dosing control and the
2K dispensers.
The Dosing system has been designed and tested for the most precise work with our
dispensers. The Dosing system has a wide variety of setting options for the dosing
quantity and time. All the values that are related to production can be saved and changed
at any time. Operation is done via an intuitive operator guidance system using a graphic
user interface. It is possible to operate two dispensers at the same time. It is possible to
set it up for other dispensers at any time without noteworthy setting-up times.
The dispensers have been developed and tested for high-precision work with products
ranging from low to high viscosity with extremely high repeat precision.
They are rotating displacement systems consisting of a rotor and stator and can be
dismantled in no time. A number of voids are produced as a result of the various
geometries of the conveying elements. Turning the rotor in the stator creates conveyance
which is either proportional to the angle of rotation or rpm-dependent.
Since the direction of flow is reversible, the medium can be sucked back to allow a clean
break of the thread. Self-sealing depends on the viscosity.
We would be glad of your help as part of our efforts to maintain our standard of quality at
the highest possible level. Please tell us about any possible ways we can improve our
product.

Dosing system eco-CONTROL EC200 DUO
Operating and maintenance instructions
6 / 43 Version 1.4 Copyright ©
2Features
3Scope of supply
Dosing system eco-CONTROL EC200 DUO (delivery as a set with 2K-dispenser)
- 2K dispenser
- Calibration adapter for 2K-
dispenser
- Mixer set
- Mains adapter with cable
- SD memory card
- Operating and maintenance
instructions
- 2 pressure sensors with
installation wrench
- Sticker set for marking
components
3.1 Mixer set
The mixer set contains three different mixers in the sizes: Small, medium and large.
(Fig. in original size)
Functions in brief with your transactions in the document:
Feature Section Page
3 dosing programs: Quantity Start/stop Time program 5.8.1 16
24 different dosing programs (operation with memory card) 5.11 27
Sucking back of the medium to prevent dripping 5.10.7 21
Adjustable pot life, automatic mixer filling 5.10.3 19
Adjustable flushflush time, prevents the medium from hardening in the
mixer
5.10.3 19
Flow quantity per minute can be set 5.10.6 20
Calibration of the controller / 2K- dispenser for exact working 5.9 17
Connection for foot switch and / or external signal 5.2 11
Pressure monitoring at the medium supply with primary pressure 5.10.10 22
Dosing pressure monitor 5.10.9 21
Level monitoring for supply tank / cartridge / tank 8.3 36
Overcurrent monitoring of the connected dispenser to protect against
damage due to sticking medium, for example.
5.10.14 26
Visualisation of the operating states via coloured displays 5.4 13
Small
Medium
Large

Dosing system eco-CONTROL EC200 DUO
Operating and maintenance instructions
Copyright © Version 1.4 7 / 43
4Safety
4.1 Informal safety measures
- The operating and maintenance manual must be kept, together with the operating and
maintenance manual of the dispenser(s), at the site where the dosing system is used
at all times.
- General and local regulations on health, safety and environmental protection must also
be provided and complied with.
4.2 Use of symbols
The following symbols are used in these instructions:
4.3 Correct use, warranty
The Dosing system is designed for controlling our dispensers in non explosion-protected
environments.
Any
- modifications and additions,
- use of non-genuine spare parts,
- repairs by persons or organisations not authorised by the manufacturer
- use without original sensors for monitoring dosing pressure
that are done without the explicit and written approval of the manufacturer, can lead to
the warranty being rendered null and void.
The manufacturer shall have no liability whatsoever for damage resulting from failure to
follow the Operating and maintenance instructions.
The chemical resistance of the parts which come into contact with the product (see the
commissioning and maintenance manual of the dispenser) must be ensured prior to
commissioning.
Text in
italics
Names of keys/buttons, connectors, chapters, screen displays, proper
names and input boxes
• Listing of the work flow sequence
-List
1. Numbered listing of a work flow sequence
1 Legend number in an illustration
Warning note. Failure to observe these notes may result in injury and
damage to the Dosing system.
Reference to technical information about operation and / or about
preventing damage.

Dosing system eco-CONTROL EC200 DUO
Operating and maintenance instructions
8 / 43 Version 1.4 Copyright ©
4.4 Qualifications of the operators and maintenance personnel
The operating organisation is responsible for ensuring that the operators and
maintenance personnel are suitably qualified. The Operating and maintenance
instructions must have been read and understood. The relevant technical rules and safety
regulations must be complied with.
4.5 Organisational measures
The necessary personal protective equipment must be
provided by the operating organisation. All safety devices
that are fitted must be checked regularly. Safety glasses
and overalls must be worn during operation and cleaning
to provide protection against any splashes of chemicals.
All of the safety information contained in the respective
Operating and maintenance instructions for the dispenser(s) that are connected to the
Dosing system must be complied with.
4.6 Preparation for commissioning - visual inspection
The Dosing system must be visually examined each day before the start of work and
before all shift changes. If there is any doubt as to the system's readiness for operation,
it must be shut down immediately and inspected by a specialist before operation
resumes.
4.7 Preventing material damage to the motors of the 2K-dispenser
The 2K-dispenser lead (connector 16, Chapter 5.2, Connections, on page 11)
may only be connected and disconnected when the power supply is isolated.
The electronics in the drive motor could be damaged if this precaution is not
taken.
The 2K-dispenser should only be operated with the original pressure sensors
connected, which have been adapted to the controller.

Dosing system eco-CONTROL EC200 DUO
Operating and maintenance instructions
Copyright © Version 1.4 9 / 43
4.8 Sources of danger due to Dosing system
- Very high dosing pressures can be produced, depending on the viscosity and the speed
of rotation, and this could result in unintended spurting out of the medium. Check the
flow quantity in relation to the dosing needle used (needle cross-section).
- When it is started up for the first time and after being refilled, air bubbles that are
included in the medium could cause an uncontrollable spurting out of the outlet nozzle.
Only start production operation once the Dosing system has been completely
bled.
- Wear suitable protective clothing if chemical, corrosive or dangerous products are
being used. Note and comply with the safety stipulations and the information from
the manufacturer. Ensure sufficient bleeding or extraction of air. Take special safety
precautions if working with dangerous media, for example, provide eye flushing
facilities if working with corrosive chemicals.
Failure to follow the instructions given below can result in damage to the unit and
possible severe injury to persons in the vicinity.

Dosing system eco-CONTROL EC200 DUO
Operating and maintenance instructions
10 / 43 Version 1.4 Copyright ©
5Operation
5.1 Displays and controls
Before commissioning the Dosing system, the safety information in
Chapter 4, Safety, starting on page 7, must have been read and understood.
Item Function, description
1+2 Graphic display
3Mains switch
4NAVI wheel Select and set variable values and input dialog
confirmations.
5Key (without Fig., below) For service, software update.
6KeySTART Starts the dosing / function
7Key STOP Stops the dosing / function; emergency stop
8KeyPRG / Esc Selection of programming, quitting an input dialog.
9Key INFO Brings further information onto the display, used to
quit page 2 in the dosing programs.
10 Return Confirmation of a selection / value input.
Shortcut keys for dosing programs
11 Time program on the right, next to the key
is a
status LED
12 Start / stop program
13 Quantity program

Dosing system eco-CONTROL EC200 DUO
Operating and maintenance instructions
Copyright © Version 1.4 11 / 43
5.2 Connections
Item Function, description
14 + 15 System plugs *
16 Motor A / B
17 USB Data transfer with PC, interface
18 Foot switch Foot switch for hands-free operation
19 System plugs *
20 RS 232 Data transfer with PC, interface
21 +24 V/DC 100VA * Mains plug, connection for power supply
22 Memory Card Card reader for MMC/SD memory card
24 Primary pressure monitoring Pressure sensor connection for medium
components A and B
25 Dosing pressure monitor Pressure sensor connection for medium
components A and B
* See section 8.3, page 36 for pin-outs
(The primary pressure does not apply for self-levelling liquids)
A Product supply A C2 Compressed air
B Product supply B C3 Dosing pressure monitor
C1 Primary pressure monitoring D Power supply dispenser
15 14
19
16 18
20
21
22
17
24 25
ViscoTecPumpen und Dosiertechnik GmbH
Amperstraße 4 D-84513 Töging
AB
C1
C2
C
C3
CB A
D

Dosing system eco-CONTROL EC200 DUO
Operating and maintenance instructions
12 / 43 Version 1.4 Copyright ©
5.3 Symbols on the screen
Designation / function Description
Dosing program
Quantity
See 5.8.1, Dosing programs,page 16
Quantity program
Dosing program
Start / stop
See 5.8.1, Dosing programs,page 16
Start / stop program
Dosing program
Time
See 5.8.1, Dosing programs,page 16
Time program
flow rate Dosing quantity per minute1.
1 In ml or g
Sucking back
Preventing dripping of the medium
The amount of medium that is "sucked back"
into the dispenser in connection with the
dosing process.
Calibrating Matching the dispenser by determining the
effective flow rate.
Saving Used to save the last values that had been
amended. Only applies if a SD memory card
has been inserted.
Start / stop
Shows the current controlling of the dispenser
(high/low). The switch is set to I during dosing.
Relative quantity,
capacity Dynamic display of the flow rate, for example
Fault
Plain text messages in the status line give
information on the relevant error message.
Details can be called up with the Info key.
O
O
0% 100%
Fault

Dosing system eco-CONTROL EC200 DUO
Operating and maintenance instructions
Copyright © Version 1.4 13 / 43
5.4 Functional and operational schematic
The interaction of the screen displays, the input dialog and the function keys:
The control system has an intuitive input dialog. The
selection or the change of a value is offered on the screen to
suit the relevant program section. The corresponding area
(A) on the screen is then shown as inverse or framed.
The status bar (B) provides plain text messages in
accordance with the input dialogs.
NAVI wheel (See 5.1, Displays and controls, page 10)
Enables rapid selection, input and confirmation of different functions.
The rotation of the NAVI wheel switches to the next or previous selection, depending on
the direction of rotation, or changes a value.
A press of the NAVI wheel confirms a selection or a change in value that had just been
made. (Alternatively, the Return key can also be used for confirmation)
The illumination of the NAVI-wheel signalises the following operating states: Blue =
Ready for operation, Green = Dosing in progress, Red = Fault.
Changing a setting value
1. Select the value to be changed by rotating the NAVI wheel, the value is shown in
inverse.
2. Press the NAVI wheel, numbered items can be selected individually.
3. Select a numbered item by rotating the NAVI wheel, the numbered item flashes.
4. Press the NAVI wheel, the numbered item is shown in inverse.
5. Change the value by rotating the NAVI wheel and press the NAVI wheel for
confirmation, the numbered item flashes again. Another numbered item can be
selected.
6. If further numbered items are to be changed, start again from 3., and if not, continue.
7. Press the PRG key to confirm the change, the value that has just been set is shown
in inverse, if other values can be changed in the input dialog, these can now be
reached by rotating the NAVI wheel.
Esc
Used to quit an input dialog. In some cases it also assumes other functions. If that is the
case, this is shown in plain text in the status line.
Press Start
Esc
Calibration auto
Motor A
ml
Desired qty
Actual qty
ml ml
00.500ml
00.000ml
A
B
00.500

Dosing system eco-CONTROL EC200 DUO
Operating and maintenance instructions
14 / 43 Version 1.4 Copyright ©
5.5 Starting up for the first time
• Operate the 2K-dispenser in accordance with the operating manual.
• Ensure that the mains switch has been turned off.
• Funktionsbereitschaft des 2K-Dispensers gemäß der mitgelieferten Inbetriebnahme-
und Wartungsanleitung herstellen.
• Connect the plugs of the 2K-dispenser(s).
• Connect the plugs of both 2K-dispenser pressure sensors.
• Connect the power supply to the control system.
• Connect the mains plug of the mains adapter to the electrical mains.
• Turns on the mains switch.
5.5.1 Save the calibration values of the dosing pressure sensors
Notice: The Dosing system is only ready for operation when the 2K-dispenser is
connected.
Tip: Use the stickers provided (letters A and B), to mark the following components.
- Both motors and dispensers and their plugs
- Pressure sensors and their plugs
- Supply lines for the media used
Caution: The 2K-dispenser lead (connector 16, Chapter 5.2, Connections on page 11)
may only be connected and disconnected when the power supply is isolated. The
electronics in the drive motor could be damaged if this precaution is not taken.
• Press PRG, system selection menu,
display 2 appears.
• Select Administration, display 25 appears.
Display 2
• Select Calib. outp. pressure A , display 51 appears.
Display 25
• Enter the calibration data supplied with the sensor
• Carry out the same setting with pressure sensor B.
Display 51
Calibration auto
Calibration manuell
Service
Flush manually A
Flush manually B
Administration
Programs
10 47h
Fr 18.03.11
Insert value
Esc
Calib. Outp. Press. A
Cal. factor
OK
output signal 094.7 mV

Dosing system eco-CONTROL EC200 DUO
Operating and maintenance instructions
Copyright © Version 1.4 15 / 43
5.6 Starting up
• Turns on the mains switch.
• Position the new mixer where necessary. The dosing system set is ready for operation.
5.6.1 Important settings
Some important settings must be made to ensure problem-free operation of the 2K-
dosing station. The following set-up makes the process easier:
- Mixer used
- Mixing ratio
- Pot life
- or flush time
- Pressure tolerance of the input pressure monitor, where used
5.7 Shutting down
The unit is shut down in reverse order of setting up.
5.8 The dosing programs and their areas of application
General
In order to achieve teh moste accurate results with the eco-CONTROL EC200 DUO, all
the main factors for a reproducible dosing operation can be set.
The dosing system must be recalibrated each time the medium is changed.
See 5.9, Calibrating page 17.
All the parameters of a dosing program can be saved. When using a memory card, 24
program slots are available for permanent storage.
Without the memory card, the data is stored temporarily on program slot 00. It is kept until
the dosing system is switched off.
The following values
can be defined,
depending on the
dosing program used:
Dosing program
Flow quantity
Flow rate
Dosing time
Sucked back quantity
Suck back speed
Pause before Sucking back
Material density
Quantity program x x x x x x
Start / stop program x x x x
Time program x x x x x

Dosing system eco-CONTROL EC200 DUO
Operating and maintenance instructions
16 / 43 Version 1.4 Copyright ©
5.8.1 Dosing programs
Quantity program
Quantity program is used to give out a fixed and defined quantity of medium. Depending
on the 2K-dispenser and the medium used, it is possible to set very small amouts per
dosing. The dosing time is determined by the volume flow. The dosing operation can be
broken off at any time until the set time has expired (STOP key).
Start / stop program
The Start / stop program is used if the 2K-dispenser is set to convey the medium over an
individual period controlled by the operator (or external control). The output quantity can
be defined. The dosing quantity is determined by the volume flow.
Time program
The time program is used if the dosing is to be done for a set period. This is triggered by
the operator (or an external controller). The output quantity can be defined. The dosing
operation can be broken off at any time until the set time has expired (STOP key). The
maximum dosing time that can be set is 99.99 seconds.
5.8.2 Operating modes
Dosing manually
In the three dosing programs the dosing is initiated by pressing the START key. The
quantity program and the time program end the dosing by itself once the set values have
been reached or at the end of the set dosing time.
In the Start / stop program dosing can be ended by pressing the STOP key.
Dosing with the foot switch
The dosing system reacts to the foot switch being activated as follows:
In the quantity and time programs the dosing can be stopped by pressing the STOP key
before the specified values have been reached (EMERGENCY STOP). A set pause and
a return flow are carried out as well.
Dosing program Press the foot switch ... and release it again
Quantity program Quantity program ends
Start / stop program Dosing on Dosing off
Time program Time program ends

Dosing system eco-CONTROL EC200 DUO
Operating and maintenance instructions
Copyright © Version 1.4 17 / 43
Dosing by means of an external controller
As in the case of dosing with the foot switch, but the triggering and termination of the
dosing are done by supplying an electrical signal. See 8.3, Interface description, page 36.
Notice: If a memory card is being used, press the foot switch briefly to load the saved and
most recently run program and to display it on the screen. Press the foot switch once
again to start the dosing. It behaves in exactly the same way with an external signal.
5.9 Calibrating
Preparation
• Mount the calibration adapter (see 2K-dispenser operating manual).
• Connect to the controller the 2K-dispenser that has been made ready for operation
and bled.
Caution: The 2K-dispenser lead (connector 16, Chapter 5.2, Connectionson page
11) may only be connected and disconnected when the power supply is isolated. The
electronics in the drive motor could be damaged if this precaution is not taken.
• Keep at hand a suitable vessel (calibration vessel) to catch and measure the quantity
required for calibration.
5.9.1 Automatic calibration
Notice: During the calibration operation the Dosing system cannot be operated by
either the foot switch or an external control signal.
• Press PRG, system selection menu,
display 2 appears.
• Select Calibration auto
• Select motor Aor B, menu Calibration auto, display 15
appears.
Display 2
Select whether the calibration is to be done with quantity
unit ml(µl)or g(mg)*and confirm this, the set quantity can
be set as desired.
Display 15
• Set and confirm the set quantity.
The value preset in the Dosing system is to be regarded as the ideal value.
• Position the calibration vessel under the calibration adapter of the 2K-dispenser.
Calibration auto
Calibration manuell
Service
Flush manually A
Flush manually B
Administration
Programs
10 47h
Fr 18.03.11
Press Start
Esc
Calibration auto
Motor A
ml
Desired qty
Actual qty
ml ml
00.500ml
00.000ml

Dosing system eco-CONTROL EC200 DUO
Operating and maintenance instructions
18 / 43 Version 1.4 Copyright ©
• Press the START key, the calibration starts, the 2K-dispenser is switched on and
conveys the quantity specified by the system. This process can be repeated as often
as desired. e.g., for comparative measurements.
• The ACTUAL quantity that goes into the calibration vessel is determined in the
previously set measuring unit. See also 5.10.4, Dosing quantity.
• Input and confirm the ACTUAL quantity that has been determined.
• Select the OK button and confirm, the next dialog to save the calibration value
appears (display 17).
• Accept or discard the suggested program slot, confirm this,
and the system has been calibrated.
Display 17
5.9.2 Calibration manual
The Calibration manual function provides a simplified option to recalibrate the system
without having to run through the entire process in the Calibration auto menu.
This can be helpful, for example, after changing the stator, in the event of batch-
related product variations, etc., whereby the set dosing value is not to be changed
(QA documents, etc.).
Caution: Setting the flow rate in the Calibration manual menu also changes the
dosing results of all the dosing programs by a linear factor.
• Press the PRG key, system selection menu, display 2
appears.
• Select Calibration manual.
• Select motor Aor B, menu Calibration manual, display 40
appears.
Display 2
• Confirm the flow rate and set the desired value with the
NAVI wheel
• Press the NAVI-wheel key, system selection menu, display
2 appears.
Display 40
Save with OK
Esc
Save calibration value
at prog. no.: 07
OK
Calibration auto
Calibration auto
Calibration manuell
Service
Flush manually A
Flush manually B
Administration
Programs
10 47h
Fr 18.03.11
Accept with OK
Esc
Calibration manual
OK
Degree of dosing qty
Adjustment range 90 ..110%
100.0 %

Dosing system eco-CONTROL EC200 DUO
Operating and maintenance instructions
Copyright © Version 1.4 19 / 43
5.10 Settings
5.10.1 Mixing ratio
5.10.2 Mixer size
5.10.3 Setting the pot life, flush the mixer
If a dosing program is active, the mixing ratio is shown in
the right-hand display and can be changed. The value for
dispenser A can be set
(1:1 to 5:1).
See also 5.14, Changing a dosing program
Calculate the mixing ratio always by volume, not by weight.
Display 26.3
If a dosing program is active, the mixer size is shown in the
right-hand display and can be changed.
The three sizes of the supplied mixer set are available for
selection: Small, medium and large.
See also 5.14, Changing a dosing program
The mixer size is taken into account in the flow quantity in
the Rinse mixer functions.
Display 26.3
The pot life is shown in display 26.3. It can be set in every
program (Chapter 5.14). It then starts running backwards on
every dosing. The remaining time is shown.
It is compulsory to enter the pot life.
Display 26.3
Once the pot life has expired, dosing takes place (empty shot) in accordance with the
set mixer size, filling it with "fresh" medium.
Empty shot = single dosing
Continuous = ongoing in accordance with the pot life.
The Rinse function can be activated after a pot life has been entered.
The value must be smaller than the pot life. It starts running backwards after the
mixer has been filled (while the pot life is running). The remaining time is shown.
Once the flush time has expired, the mixer is rinsed once with the component from
the selected dispenser (A or B). This ensures that the mixer is filled with just one
medium and protected against sticking. The quantity corresponds to the set mixer
size.
Mixture 1.0:1.0
Mixer small
Pot time 00 : 23 : 45
Blank shot
Flush
Flushing Time
Flush A
Flush B
00:23:00
endless
G
Program no. 07
Mixture 1.0:1.0
Mixer small
Pot time 00 : 23 : 45
Blank shot
Flush
Flushing Time
Flush A
Flush B
00:23:00
endless
G
Program no. 07
Mixture 1.0:1.0
Mixer small
Pot time 00 : 23 : 45
Blank shot
Flush
Flushing Time
Flush A
Flush B
00:23:00
endless
G
Program no. 07

Dosing system eco-CONTROL EC200 DUO
Operating and maintenance instructions
20 / 43 Version 1.4 Copyright ©
5.10.4 Dosing quantity
General on determining the dosing amount.
Notice: If a value that differs from 100% is set in the Calibration manual menu
(section 5.9.2), this affects the results of the dosing.
The smallest dosing quantities can be determined most easily via the weight. We
recommend setting the material density first.
Dose your required quantity with the Start / stop program and weigh it. Now input in the
quantity program the quantity in g.
Setting the dosing amount as a value
1. Press the Quantity program shortcut key, display 3
appears.
2. Ensure that the required quantity unit* has been selected
in the Quantity display area.
3. Set and confirm the dosing amount.
Display 3
5.10.5 Dosing time
1. Press the Time program shortcut key, display 10
appears.
2. Set, confirm and save the dosing time.
The set value remains active until the next change or until
the Dosing systems is switched off. Alternatively, a value of
its own for the flow rate can be assigned to each dosing
program (See 5.14, Changing a dosing program, page 29).
Display 10
5.10.6 Flow rate
The flow rate in quantity/min can be set in each dosing program. The
maximum flow rate that can be set is
6 ml/min, per dispenser (A and B). If a mix ratio of 1:1 is set, 12 ml/min is
conveyed.
• Selecting, changing and saving the displayed value.
Display 34
The bar chart shows in what range (%) the flow rate is.
The set value remains active until the next change or until the Dosing systems is
switched off. Alternatively, a value of its own for the flow rate can be assigned to each
dosing program (See 5.14, Changing a dosing program, page 29).
Time program
Esc
Time
0% 100%
ml/min
01.00 03.000 00.000
mlsec
SuckbackFlow rate
Flow rate
0% 100%
ml/min
03.000
Table of contents
Popular Laboratory Equipment manuals by other brands

Velp Scientifica
Velp Scientifica AREX DIGITAL PRO F20500410 instruction manual

MELAG
MELAG Vacuklav 24-B/L user manual

IKA
IKA RC 5 basic operating instructions

Gilson
Gilson Power Carrousel user guide

Markes International
Markes International Kori-xr installation manual

Cleaver Scientific
Cleaver Scientific CSL-CELLAS instruction manual