1. General Information
Your EndoLight® 3
Accessories 3
Intended Use 4
Contraindications 4
2. Safety
Signal words and warning symbols 4
General 5
Safety instructions 5
Type label 7
Guidelines, regulations and standards 8
3. Your EndoLight®
Material 9
Diodes 9
Display 10
Battery 10
4. Setup
Charge your EndoLight® 10
Put your EndoLight® on 11
Additional adapter 11
Function Overview 12
-Turn on 12
-Select mode 12
-Select intensity and start session 13
-Turn off 13
Change the wristband 13
5. Recommendations for use 14
6. Storage and cleaning 14
7. Product durability 14
8. Firmware update procedure 15
9. Troubleshooting 16
10. Warranty and guarantee 16
11. Quality system - Serial number 17
12. Manufacturer and service contact details 17
13. Product specifications 18

3
1. General information
Your EndoLight®
From now on, the EndoLight® is your daily companion, enhancing your well-being on all levels.
You can wear it anytime and anywhere: At home, in the office, at the gym, in a café, while traveling or
on the go.
Attention: Take a moment to carefully read all the important information in these instructions,
especially the "Safety" chapter.
Accessories
EndoLight® with Additional wristband Adapter for smaller wrists
pre-installed wristband for smaller wrists
Power adapter plug Universal charger and data cable
USB-C to USB Type A
Europe: 220VAC to 5V 1A
USA: 110VAC to 5V 1A
Travel and storage case Cleaning cloth Quick Guide

4
Intended Use
The EndoLight® is a portable device that is based on the principles of Low-Level Laser Therapy
(LLLT). It is worn around the left wrist for a duration of 30 min/day.
The EndoLight® has 7 laser diodes and one LED diode that are grouped in two light units. They are
positioned so that the light hits the main arteries in the wrist (Arteria Ulnaris and Arteria Radialis) for
an effect on the entire body.
The diodes irradiate the circulating blood with different wavelengths of light and intensities, depending
on the user settings.
The device is powered by a built-in lithium-ion battery.
The EndoLight® is intended for private use only and not for industrial/commercial use. According to
the Product Safety Act, the EndoLight® is not a medical product but a wellness product. Its intended
use is for the improvement of general well-being and was not designed to diagnose, cure, alleviate, or
treat any disease, symptoms, or other medical problems.
Please note that in the event of improper use, liability is void.
Contraindications
Use in infants and children under 16 years of age, pregnant women and in case of pre-existing
conditions only after consultation with a doctor.
2. Safety
Signal words and warning symbols
A WARNING message contains special safety emphasis and must be
observed at all times. Failure to observe a WARNING message could result in
serious personal injury.
Failure to observe a CAUTION associated with use could result in minor injury.
Please observe the following instructions carefully to avoid property damage.
ADVICE: Please observe the information of this user guide. Do not use the
EndoLight® without reading this guide.
Electrical voltage warning
EndoLight® not for general waste.
Battery not for general waste.

5
General
Some basic safety precautions must be taken into consideration when using electrical equipment.
Failure to take precautions outlined in this section could result in fire, electric shock, other injuries,
damage to the EndoLight® or other property. Please read the following safety instructions carefully
before use.
This information relates to the EndoLight®, the charging and data cable, the power adapter, and all
accessories.
Always keep these operating instructions on hand and store them in a safe place so that you can pass
them on to the new owner in the event of a sale.
Safe use of the EndoLight® is the primary responsibility of the user. If the EndoLight® appears to be
operating incorrectly, please refer to the troubleshooting guide. Contact us if the problem persists.
Safety instructions
This hazard warning indicates a potentially hazardous situation which, if not avoided, may result in
minor or moderate injury.
Skin reactions
Wearing the EndoLight® can cause skin irritation or allergic reactions. To avoid this, you should
always keep the device clean and dry. Please do not wear it too tight around your wrist.
If the irritation persists, take the EndoLight® off your wrist, and contact your doctor if the symptoms
persist for more than 3 days.
If you experience pain, numbness, or a burning or tingling sensation while or after wearing the device,
remove it immediately.
Due to the high-power output, it is normal that the EndoLight® generates heat (approx. 40 degrees
Celsius or 104 degrees Fahrenheit). Skin with a higher proportion of melanin absorbs a higher amount
of laser light and can therefore heat up more likely. Observe your skin reaction particularly carefully if
you have dark skin (skin type V or VI).
If the EndoLight® becomes uncomfortably hot while it is being used or charged, or if it produces
unusual noises or smoke, immediately remove it from your wrist or unplug it from the socket.
Physical discomfort
Consult your doctor and temporarily stop using the device if you experience physical symptoms such
as seizures, fainting, eye aches or headaches that you suspect may be caused using the EndoLight®.
Product damage
If your EndoLight® has been damaged, do not use it anymore.
To avoid injury, do not try to open or disassemble the EndoLight® yourself.

6
Handling
Handle your EndoLight® with care. Do not drop it or expose it to an open flame.
Do not use accessories or detachable parts not described in this user guide.
Do not interconnect the EndoLight® with other equipment not described in this user guide or modify it
in any way.
This WARNING message contains valuable safety information, and it must be taken seriously at all
times. Failure to observe this WARNING message could result in serious personal injury.
Never look directly into the diodes –visible and invisible laser radiation may
be harmful to the eye.
The use of controls, adjustments, or performance of procedures other than those specified in this user
guide may result in exposure to hazardous laser radiation.
Suitable for mains supply voltages 110-240 V/AC.
Authorized and non-authorized users
The EndoLight® may be used by children under 16 years and persons with limited physical, sensory or
motor skills and inexperienced/untrained individuals, provided they are supervised or instructed by a
responsible adult regarding the safe use of the device and are aware of the possible risks. Use in
children under 16 years only after consultation with a doctor.
Cleaning and care must not be carried out by children without supervision.
The EndoLight® must not be used as a toy. If the device is used by children or in their vicinity, carefully
supervise the situation. Children cannot always correctly recognize possible dangers.
Keep the EndoLight® away from babies, infants and young children. Packaging materials or
components that come loose can pose a choking hazard.
The charging cable poses a strangulation risk. Store appropriately when not in use.
Battery
The EndoLight® is powered by a rechargeable lithium-ion battery. Please do not try to change the
battery yourself. Doing so may result in a hazard. Only trained and authorized service personnel may
change the battery. Contact our service team to have it replaced.
Charge the battery of your EndoLight® according to the instructions (see chapter 4). Use only the
supplied data/charging cable and the associated power supply unit.
The EndoLight® is disabled for use during charging. It automatically turns itself off as soon as it is
connected to a charger. When connected to a charger, it is not possible to turn the EndoLight® on.
If the EndoLight® becomes damp or wet, the battery must not be charged.
The battery contains lithium. In the case of fire use an ABC or BC type fire extinguisher.

7
ADVICE: This notice indicates possible damage to property, which may occur if the safety measures
are not followed.
ATTENTION! Please observe the following instructions to avoid damage to property:
Do not try to open or disassemble the EndoLight® yourself to repair it. Disregarding this can lead to
void guarantee and overheating. Always contact customer service representatives in case your
EndoLight® requires repair or maintenance.
If an electrostatic discharge should occur, the EndoLight® turns off automatically. You can simply turn
it on again by pressing the On/Off button.
The EndoLight® offers no special protection from water (IP Class IP0X). So please be careful that it
does not get damp or wet. For example, do not use it outside in the rain, when exercising on / in the
water, when showering, bathing, in the sauna or steam room. Do not use the EndoLight® if moisture
has penetrated.
Do not expose the EndoLight® to extreme temperatures or direct sunlight. It is designed for ambient
temperatures and storage temperatures between 0°C and 30°C (32°F and 86°F). Operating or storing
outside of these temperatures can damage the EndoLight® and reduce battery life. Do not expose the
EndoLight® to direct sunlight or sudden changes in air temperature or humidity (permissible relative
humidity: 8 % –90 %).
Maximum operating altitude is 3,000 m above sea level, without pressure compensation.
Avoid contact with dust or sand.
Never use force when pressing the buttons or connecting a cable or plug as this may cause damage
that is not under guarantee. If the plug does not connect easily, it is likely not compatible with the
EndoLight® or the charging port is blocked. Make sure that you are only using the accessories
provided and that the charging port is free from debris.
To prevent the cable from breaking, roll it up after use instead of folding the cable.
Disposal
When the EndoLight® has reached the end of life, do not dispose of in general waste, take it to
an e-waste recycler.
Dispose of the EndoLight® and the battery separately from household waste and in
accordance with the applicable environmental laws and regulations. Improper disposal can
have significant negative effects on the environment.
Batteries are not accepted via household recycling collections however they can be recycled at your
local recycling center or nearest battery recycling collection point found in most supermarkets.
Type label

8
Guidelines, regulations and standards
This device is classified as a wellness device and meets UL/CE standards.
The EndoLight® is compliant according to:
•IEC EN 60335 1:2012 + A11:2014 + A13:2017 + A1:2019 + A14:2019 + A2:2019 + A15:2021 –
EN 62233:2008 Safety of household and similar electrical appliances
It complies with all European regulations on electromagnetic compatibility (EMC):
•Emission EN 61000-6-3:2007 + A1:2011
•Immunity EN 61000-6-1:2007
•EN 55032:2015 + A11:2020 FCC CFR 47 part 15 subpart B ICES-003:2016 updated April 2019
The EndoLight® is RoHS 3 compliant and contains no toxic chemicals.
Graphic 1: CE Declaration of Conformity

9
3. Your EndoLight®
Material
The structure of the EndoLight® is made from aluminum.
The wristbands are made of a hard-wearing silicone rubber that is particularly comfortable to wear
thanks to its high-quality material properties.
Other materials used are quartz glass and medical grade polycarbonate plastic.
All materials are BPA-free.
If the original wristband is too large for you, you can replace it with the smaller one, which is included
with the EndoLight®. Please refer to chapter 4 on further information on how to change the wristband.
The buckle that fastens the wristband is made of stainless steel.
Diodes
The device is a Class 1C laser device. Special proximity sensors ensure that the diodes are only
activated when they are in contact with skin. This reduces the risk of eye damage to the user or a by-
stander through direct exposure to LED or laser light.
Graphic 2: Safety label on proximity sensor
The core feature of the EndoLight® are the two uniquely designed diode units that deliver their power
in an optimally dosed manner. Special lenses collimate and guide the light directly through the skin,
into the arteries. This lens is made of PMMA and has features that, in addition to the contact sensor,
ensure user safety.
Graphic 3: Diode set-up

10
Warning: Please never look directly into the light and do not shine the light into anyone else's
eyes, as this can lead to serious eye damage.
Display
The EndoLight® features an IPS LCD panel.
Battery
The EndoLight® is powered by a rechargeable lithium-ion battery.
4. Setup
Charge your EndoLight®
Fully charge your EndoLight® before first use:
•Connect the USB Type A end of the charging cable to the power adapter and plug the adapter
into a socket. Alternatively, plug the Type A end into the USB port on a computer.
•The other end of the charging cable has a USB Type C connector. Plug it into the connection
socket provided on the EndoLight® (see graphic 4):
Graphic 4: USB Port
When the device is powered off and a charger is connected, the screen shows the charge level.
Graphic 5: Battery display
Charge time is approx. 2 hours from 0 % - 100 %. You can view the current charge status by pressing
one of the buttons.

11
Graphic 6: Battery charging status
The charging process is complete when a fully charged battery symbol appears on the display.
A fully charged EndoLight® has a battery life of 2-3 days when worn 30 minutes/day in “Recover”
mode at full intensity.
The EndoLight® will go into low power mode (low screen brightness) when battery level is less than
15 %. It will shut down when battery level is less than 2 %.
Put your EndoLight® on
Place the device around your left wrist and fasten the buckle.
The EndoLight® should sit comfortably on your wrist, about one finger's breadth
below the palm. The diodes rest on the arteries inside the wrist. If the display is upside
down, the EndoLight® has been put on the wrong way. The display is located on the
extension of the thumb.
If the wristband is too wide, see the instructions for changing the wristband
(chapter 4).
Graphic 7: EndoLight® at wrist
Make sure the diodes are resting on your skin. The EndoLight® is equipped with a contact sensor and
is only activated when contact with skin is made.
The EndoLight® must not be too tight so as not to hinder blood circulation (tight, but not constricting).
Additional adapter
In case you have small wrists, and you feel like light is leaking, please apply the adapter as shown on
graphics 8 and 9:
Graphic 8: Adapter covers bottom diode row Graphic 9: Adapter covers upper diode row
The EndoLight® is fully charged
More than 90 %
65 % - 90 %
40 % - 65 %
15 % - 40 %
Less than 15 %

12
Function overview
Graphic 10: Buttons
Graphic 11: Display
Turn on
Briefly press the On/Off button to turn on.
Select mode
The EndoLight® has the following three operating modes that you can choose based on your wellness
goals:
Recharge: Activates the green diode and both red and infrared diodes
Relax: Activates the yellow, green and blue diodes
Turn on/off and confirm settings
Scroll through the modes
Select intensity
Minutes until session is finished. One session is 30 minutes.
Mode: Recharge, Relax or Recover (Icon)
Intensity: 25 % | 50 % | 75 % | 100 % (Ring closure)
Active diodes: The ring colors indicate which diodes are active
Battery status

13
Recover: Activates all diodes simultaneously
#
Graphic 12-14: Modes
Press the button to "scroll" through the modes. Confirm your selection with On/Off.
Select intensity and start session
Set intensity via the rocker switch. You can choose from the following intensity levels:
25 % 50 % 75 % 100 %
Graphic 15: Intensity
The button increases the intensity. The button decreases it. Press the On/Off button to confirm
your selection and start the session.
The selection applies to all diodes simultaneously. Individual diodes cannot be selected separately.
Turn off
After 30 minutes the EndoLight® beeps three times before turning off automatically. Alternatively, you
can end the session early by pressing and holding the On/Off button.
If the contact sensor is not near the skin, the diodes will not turn on, or stop if already
on. The timer will pause, and this screen will show up together with an acoustic signal.
Please correct the fit on your wrist to start or continue the session.
Graphic 16: Notice Contact sensor
Change the wristband
You do not need any tools to replace the wristband.
Detach the wristband
Both parts of the wristband have a small spring lock that secures the band to the device. To mount the
smaller wristband, you only need to detach the part with the holes.
Slide the silver spring pin of the band down to release it. You can use your finger or a pointed object.

14
Mount the wristband
The new wristband is clicked back into place at the bars. First insert the lower end of the spring pin
into the hole provided. Slide the spring pin down and position it over the upper hole. Release the spring
pin so that it falls into the hole and fixes the band.
Important: MAKE SURE THAT THE PIN HAS FALLEN SECURELY IN POSITION TO AVOID ACCIDENTAL
DETACHMENT AND DROPPING OF THE DEVICE.
5. Recommendations for use
Wear your EndoLight® whenever you want it to support you in everyday life. We recommend one daily
session of 30 minutes. Maximum sessions per day: 2.
At the beginning, start with low intensity (25 % - 50 %) and observe your reaction. You can then
gradually increase the power from session to session depending on how you feel.
6. Storage and cleaning
Store your EndoLight® in the provided storage case when not in use to prevent dust and debris getting
inside the device.
Perspiration can develop during the session. We recommend wiping the surfaces after use with the
provided cleaning cloth. Alternatively, you can use an abrasion-resistant, lint-free cloth for cleaning.
You can use water or cleaning alcohol (such as isopropanol 70 %) to moisten the cloth. First spray this
onto the cloth and not directly onto the device. Make sure that the cloth is only damp and not wet.
Never spray cleaner directly on the device. Damage due to the ingress of liquids or solvents will void
the manufacturer’s guarantee. Use of other cleaning products may damage your EndoLight®.
The EndoLight® should only be cleaned when it is turned off and not while it is charging.
CAUTION: Do not use cleaning products that contain ammonium chloride. These chemicals
may cause cracking of the material.
7. Product durability
The EndoLight® is designed to be durable under typical use.
The device is equipped with parts that are subject to natural wear and tear (e.g., the rechargeable
battery). The service life of these parts depends either on their frequency of use or is limited in time.
These parts can possibly be replaced by our service team.

15
8. Firmware update procedure
If a firmware update becomes necessary, please follow the steps listed below.
Boot Version: 1.00.00.04
Main Version: 1.00.01.04
Firmware Erase:
1. Power off the device.
2. Press and hold down both arrow buttons and On/Off button.
3. The device will power up and start a long beep after 10 seconds.
4. Release all buttons.
Firmware Download:
1. Copy the update utility (WeberUsbFw.exe) and firmware file (WW100007.bin) to the same folder.
2. Press On/Off button.
3. The device will start in BOOT mode.
4. Connect USB cable to PC.
5. The PC will show a new device detected message, only at first time, and install drivers
automatically.
6. When done, run the update utility (WeberUsbFw.exe).
7. Click "Start". The utility and the device screen both will show the download progress.
8. When done, the device will show a success message.
9. Disconnect USB cable.
10. The device will power off.
11. Press On/Off button to turn on the device in normal mode.
Note: the device will again power on in BOOT mode if the firmware file is not compatible with the
device's serial number, or download process interrupted.
Retrieving Information:
In BOOT mode:
Press and hold down On/Off button at any time to show BOOT serial number and firmware revision.
In normal mode:
Power up the device while pressing and holding down the power button for 10 seconds. Serial number
and firmware revision will appear at the bottom of the screen.

16
9. Troubleshooting
Problem
Possible cause
Solution
The EndoLight® …
does not turn on
Empty battery
Make sure the battery is
charged and try again.
Device is charging (usage
during charging is disabled for
safety)
Wait until the device is fully
charged, disconnect from USB
cable and try again.
does not turn off
Software error
Wait for the battery to drain
until the EndoLight® turns
itself off. Fully charge it and
turn it back on/off.
gets very hot
(a slight warming is normal)
Battery is overheated
Unplug the EndoLight® from
the power supply if you are
charging it. Turn it off and only
turn it on again once it has
cooled down completely.
does not respond to button
presses
Software error
Turn off the EndoLight® and try
again to adjust your settings.
If troubleshooting does not solve the problem, please contact customer service.
10 . Warranty and guarantee
Warranty
We grant you a 24-month warranty. This covers manufacturing defects in your EndoLight® from the
date of purchase that were already present at the time of purchase.
Guarantee
What is covered by the guarantee?
The guarantee period for material and manufacturing defects under normal use is one year and begins
from the date of purchase. This applies to both the EndoLight® and any accessories.
No new guarantee period begins after repairs or replacements.
What is not covered?
We do not guarantee the repair or replacement of a product if the defect is due to:
•Accidental damage caused by improper care or misuse, as well as careless use or operation
which is not in conformity with this user guide.
•Use of the product beyond “normal use”, which corresponds to more than 1 hour a day.
•Use of parts that have not been assembled or installed in accordance with the manufacturer's
guidelines.
•Use of parts and accessories from other manufacturers.

17
•Repairs or changes that have not been made by the manufacturer or its authorized partners.
•Incorrect setup
•Normal wear and tear
Please contact our service team if you are unsure which cases are covered by your guarantee.
Replacement or repair of defective devices within the guarantee period
During the guarantee period defective devices or device parts can be sent to one of our service
addresses. Repair or replacement of the EndoLight® is at the discretion of the manufacturer.
Before returning a device, please contact us and have your purchase receipt or delivery note ready.
Therefore, please keep the invoice and delivery note carefully as proof of purchase. Without these
documents, you will be charged for the costs of all work involved.
We will then inform you about the further procedure.
Replacement or repair of defective devices outside of the guarantee period
Even after the guarantee period has expired, you can send the defective device for repair. Please
contact us first before sending in the device.
Repairs or replacements that occur after the guarantee period are subject to a charge.
11. Quality system - Serial number
During production, the device is assigned a serial number for traceability:
You can find it under the USB port door.
In case you experience a warranty or guarantee issue, please have the serial number of your
EndoLight® ready.
12. Manufacturer and service contact details
Manufacturer address:
Weber medical GmbH
Sohnreystr. 4
37697 Lauenförde
Germany
Website: www.webermedical.com

18
Service address for USA:
Weber Laser Systems
Attn: Repairs
1926 Bonus Drive
San Diego CA 92110
USA
General service address for all other countries:
W Medical Systems GmbH
Lönsstr. 10
37697 Lauenförde
Germany
Tel.: +49 5273 3894506
Email: info@wmedicalsystems.com
Website: www.wmedicalsystems.com
If you require any help with the setup, use or maintenance of your EndoLight®, or to report unexpected
operation or events, please contact the general service address: info@wmedicalsystems.com
13. Product specifications
Diode color
Diode type
Quantity
Wavelength
Infrared
Laser
2
808 nm
Red
Laser
2
660 nm
Yellow
LED
1
590 nm
Green
Laser
1
505 nm
Blue
Laser
2
450 nm
Output power per diode*
8,75/17,5/26,25/35 mW
Total output power*
70/140/210/280 mW in Recover mode
Input voltage
110-240VAC
Duration per session
30 minutes
Permissible environment temperature
0°C - 30°C (32°F - 86°F)
Permissible relative humidity
8 % - 90 %
Power source
Lithium-ion battery
Weight
0.13 kg
*Dependent on the selected intensity (25 % | 50 % | 75 % | 100 %)