Westlab FXP Series User manual

FXP8M
OPERATING MANUAL
WESTLAB
TABLE OF
CONTENT
1. Applications 1
2. Specifications 1
3. Construction 1
4. Approvals 1
5. Guidelines 2
6. Do/ Don’t 2
8. General Cleaning 3
9. Installation 3
10. Operation 3
11. Servicing 4
12. Performance Test 4
13. Warranty Statement 5
14. Warning 5
*Please read this instructions
carefully before use.

1
WESTLAB
FXP8M
OPERATING
MANUAL
1. APPLICATIONS
The FXP range is a single chamber device used
to clean instruments and other hardware with
ultrasonic energy in a mixed water/chemical solution.
3. CONSTRUCTION
2. SPECIFICATIONS
•
•
•
•
•
•
Certified by the Australian Made Campaign
as Made and Owned in Australia.
Power is supplied by one single phase 10A plug.
Operating temperature should not exceed 60°C
to maintain reliability and efficiency.
Minimum operating depth should not be less than 80mm.
Care should always be taken to avoid excessive
spillage of solution when draining fluid.
Resin coated transducers and circuit board
for moisture protection.
4. APPROVALS
•
•
•
•
AS/NZS 3760:2010
AS/NZS 3100:2017
Certificate of Conformity No:E990002-(C-Tick)
Certificate for inclusion of medical device – Class 1 (T.G.A.)
FXP8MModel
Timer Mechanical (30min)
Heang Power Nil
Tank Material Pressed 304 Stainless Steel
Tank Capacity 1.7L
Transducer Frequency
40kHz
Ultrasonic Power 50W
Overall Dimensions 225 x 165 x 240mm
Internal Tank 150 x 140 x 100mm
Gross Weight 2.5kg
Basket Standard: 120 x 100 x 65mm with 9mm Mesh Aperture
Opon: 5mm Mesh Aperture
Lid Standard: Clear Plasc Included
Opon: PVC or Stainless Steel
Valve Standard: No

2
5. GUIDELINES
Always allow 5 minutes for fresh liquids to degas.
Degassing is the action of removing air bubbles from the
solution, which is achieved from turning the ultrasonics on.
It is not recommended to operate unit for longer
than 30 minutes.
Fluid temperature will increase under operating conditions.
The operator should always be aware that it only takes
a small amount of moisture contamination on the power
connections to cause electrical breakdown.
•
•
•
•
•
6. DO:
Use a suitable chemical applicable to your application.
Check the suitability of the chemical for use in ultrasonics,
as some chemicals can emit a dangerous toxic gas.
Use beakers and positioning covers or auxiliary pans
for additional solutions.
Rinse thoroughly after each cleaning application.
Keep tank bottom free of silt and dirt accumulation.
Always drain the basket fully of fluid before leaving
the cleaning chamber.
✓
✓
✓
✓
✓
7. DON’T:
Run machine without solution in the tank.
Do not allow liquid level to drop below 80mm from
tank bottom.
Use heaters without a thermostat control.
Setting for the thermostat on heated models should
not exceed 60°C.
Place items directly on the tank bottom, always suspend
in basket or rack.
Drop items into stainless steel bath.
Do not pour hot water directly into the tank.
Use chemicals with very high or low pH level.
Use wax and compound removers on plastic or acrylics.
Use a flammable or combustible solution.
Subject the jacket and controls to splashing liquid
and fluid overflows.
✕
✕
✕
✕
✕
✕
✕
✕
✕
✕
✕
WESTLAB
FXP8M
OPERATING
MANUAL

3
8. GENERAL CLEANING:
The turnover of solution must be determined by the user
to satisfy acceptable cleaning results of the finished article.
If the contamination is heavy and difficult to remove,
a pre-rinse in a heated solution could be enough to soften
and loosen unwanted debris for ultrasonic cleaning.
At the other end of the scale, after the items have been
removed it is essential that a final rinse be performed
to remove any residue that remains on the cleaned parts.
•
•
•
9. INSTALLATION:
When locating a position for the cleaner, consideration
should be given to the following:
FOR M:
Turn the Mechanical timer switch to the desired time
setting and ultrasonic cavitation will occur in the liquid.
FOR D:
Generous circulation of air all around the cleaner.
Surrounding air free of dirt, contamination
and high temperatures.
Away from possible liquid splatter.
Easy access for manual adjustment.
•
•
•
•
10. OPERATION:
Insert the single phase 10A plug into a 240-volt, AC outlet.
Pour cleaning fluid of your choice into the tank.
Switch the supply mains ON.
The unit is now ready to operate.
•
•
•
•
To set interval time, press min or sec for desired setting.
Press start/stop for countdown start.
When countdown finishes, 30 seconds of beeps are given,
and the relay is off.
To terminate the countdown process, press the start/stop key.
To re-start the same cycle of countdown,
press start/stop key once.
To change the current count down setting, press min
or sec to set the desired value.
•
•
•
•
•
•
WESTLAB
FXP8M
OPERATING
MANUAL

4
11. SERVICING
Unauthorized repairs will void warranty.
Only Unisonics Australia employees should service
this ultrasonic cleaner.
The circuit board presents high-level voltage.
Improper testing or repair procedures could result in injury
to the service person and damage to the product.
All repair requests are supplied as a no obligation free quote.
•
•
•
•
12. PERFORMANCE TEST:
Please find as follows the testing process for efficiency
levels in the ultrasonic cleaner.
TESTING PROCEDURE:
The aluminium foil test is an economical and consistent
test for units under 60 Litres to gauge performance.
FXP8 cleaners over 60 Litres should use a cavitation
meter for more accurate testing.
Always degas the solution prior to the performance test to
allow for the elimination of air found in fresh solution.
This is achieved by turning the ultrasonics on for 5 minutes.
Testing should always be carried out in a solution only
chamber (basket and cleaning items removed).
The foil must be suspended in the solution for a duration
of 30 seconds.
A consistent pattern of perforation should result
with hole sizes varying from one to four mm in diameter.
Tested fluid should be disposed of, as aluminium residue
will remain and deposit as a contaminant.
PLEASE NOTE: Inconsistency in foil thickness
and manual holding can create a degree of error
that will vary the results of the foil test.
1.
2.
3.
4.
5.
WESTLAB
FXP8M
OPERATING
MANUAL

5
13. WARRANTY STATEMENT:
The MANUFACTURER of this machine, WESTLAB
AUSTRALIA PTY. LTD., warrants it to be free from defects
in material and workmanship under normal use
and service and the purchaser shall be entitled
to the following consideration:
14. WARNING:
UNAUTHORIZED MODIFICATIONS AND/OR REPAIRS
WILL VOID WARRANTY.
Only a full authorized service person should attempt to service
this product. The circuits present high-level voltage. Improper
testing or repair procedures could result in injury to the service
person and damage to the product.
If any part or parts of the cleaner prove defective within
twelve months of the date of installation, in the original
purchasers’ premises, the defective part or parts will be
repaired or replaced free of charge. The cost of
transportation both ways of the part or parts or even
the unit itself, to the premises of the MANUFACTURER
is to be paid by the claimant.
The MANUFACTURER will not be liable for any loss,
damage or expense caused directly or indirectly by
any fault developing in this machine or its accessories.
This WARRANTY applies only to installations within the
Commonwealth of Australia and its Territories. Removal of the
original serial numbers voids this WARRANTY in its entirety.
The MANUFACTURES decision shall be final and binding
in respect to any claim or dispute arising out of a defect in
any machine parts.
The MANUFACTURER recognises no other WARRANTY,
expressed or implied by others.
At the discretion of the MANUFACTURER, this warranty
shall not apply to: -
(a) Parts of proprietary manufacture supplied
or incorporated in this machine, which are covered
by their manufactures separate guarantee.
(b) Parts made defective through accident, incorrect
voltage or current characteristics, alteration misuse,
neglect or wear and tear.
order and delivery details at the end
of the day.
ONLINE
www.westlab.com.au
Order online with our easy Express
Order form for fast, secure ordering.
PHONE
1800 358 101
Speak to one of our
friendly, efficient staff
from 8.00am to 5.00pm Monday to
Friday.
EMAIL
sales@westlab.com.au
24 hours, 7 days a week convenience.
FAX
1300 725 903
Order via fax and we will confirm your
ORDERING WITH
WESTLAB IS EASY...
WESTLAB
FXP8M
OPERATING
MANUAL
This manual suits for next models
1
Table of contents
Other Westlab Ultrasonic Jewelry Cleaner manuals