Wide CL24s User manual

Professional Dispaly Solution
Before connecting, operating or adjusting this product,
please read this instruction booklet carefully and completely.
在连接、操作或调整本产品之前,请仔细并完整地阅读本使
用说明书。

Specications and features are subject to change without notice. Images shown are
for illustrative purpose only. All names of the products are trademarks or registered
trademarks of their respective companies. IQ Sensor® is registered trademark of WIDE
Corporation. Created in Korea. 2014. 09. Ver 2.0 规格和功能如有变更,如不另行通知。显示的图像仅用于图示说
明。所有产品的名称都是其各自公司的商标或注册商标。IQ Sensor® 是 WIDE Corporation 的注册商标。韩国制
造。2014. 09. 2.0 版本
WIDE KOREA (Gomae-Dong) 12 Wongomae-Ro, Giheung-Gu, Yongin-Si, Gyeonggi-Do 446-901, Korea Tel: +82-31-
218-1600 Fax: +82-31-274-7400 [email protected] WIDE USA 8 Hammond St., Suite #114 Irvine, CA 92618, USA
Amsterdam, The Netherlands Tel: +31-20-311-9797 Fax: +31-20-311-9790 [email protected] WIDE JAPAN Shinjuku
Suzuki Bldg A 1-6-8 Shinjuku, Shinjuku-Ku, Tokyo160-0022, Japan Tel : +81-3-6457-8371 Fax : +81-3-6457-8372 info-
[email protected] WIDE ASIA 1350/88-89, Thairong tower, Pattanakarn road, Suanluang, Bangkok 10250, Thailand
Tel:+66- 2-717-1843 Fax: +66-2-717-1844 [email protected] Foreseeson Custom Displays, Inc. 2210 E. Win-
ston Road Anaheim, CA 92806 USA Tel: +1-714-300-0540 Fax: +1-714-300 0546 Foreseeson GmbH Industriestrasse
38a 63150 Heusenstamm, Germany Tel: +49-6104-64398-0 Fax: +49-6104-64398-11 [email protected] Foreseeson
UK Ltd. Unit 2 Kingsmill Business Park Chapel Mill Road Kingston upon Thames Surrey KT1 3GZ Tel: +44-208-546-1047
Fax: +44-208-546-3931 [email protected]

Professional Dispaly Solution
Before connecting, operating or adjusting this product,
please read this instruction booklet carefully and completely.
在连接、操作或调整本产品之前,请仔细并完整地阅读本使
用说明书。

1
Table of Contents
Overview ---------------------------------------- 2
Safety Symbol ---------------------------------------- 3
Safety Introductions ---------------------------------------- 5
Cautions ---------------------------------------- 7
FCC Information ---------------------------------------- 10
Parts ---------------------------------------- 15
Mechanical Dimension ---------------------------------------- 17
Controls ---------------------------------------- 18
Power Management ---------------------------------------- 19
On Screen Display (OSD) Menus ---------------------------------------- 19
On Screen Display(OSD) System Overview ---------------------------------------- 24
Standard Signal Table ---------------------------------------- 29
Signal Connector Pin Assignments ---------------------------------------- 30
Specication ---------------------------------------- 32
Cleaning Instructions ---------------------------------------- 33
Service ---------------------------------------- 33

CL24s USER’S MANUAL
Overview
CL24s is the perfect display for creative professionals. With its 24-inch wide LCD
panel, 10-bit gamma correction, wide gamut, and full HD support, CL24 is the
ideal tool for both still and moving images. Furthermore CL24 features IPS (In
Plane Switching) panel to take full advantage of LCD technology. It is impossible to
miss the improved color accuracy, rich detail, and extended viewing angles when
combining LCD with IPS.
CL24s gives you the multiple selection of color settings among. Users can also adjust
settings to suit their needs of their work places.

3
Safety Symbol
The following symbols appear on the product, its labeling, or the product packing.
Each symbol carries a special denition, as dened below:
Dangerous : High Voltage Consult accompanying documents
Direct Current Power adapter
Indicates protective earth ground lndicates equipotential earth ground
DC Power control switch Indicates top-bottom direction
Do not get wet Fragile
Consult the operating instructions. 6Maximum Stacking
Indicates the manufacturing date Indicates the manufacturer
Representative in the USA Authorized representative in the
European community
Serial Number Humidity limitation
Temperature limitation Atmospheric pressure limitation

CL24s USER’S MANUAL
Indicates proof of conformity to applicable European Economic Community Council
directives and to harmonized standards published in the ofcial journal of the European
Communities.
Medical Equipment is in accordance with UL 60601-1 and CAN/CSA C22.2 No.601.1 in
regards to electric shock, re hazards, and mechanical hazard.
Tested to comply with FCC Class B standard.
Indicates the display is approved according to the CCC regulations.
a RoHS labels.
This symbol indicates that the waste of electronic equipment must not be disposed
as unsorted municipal waste and must be collected separately. Please contact the
manufacturer or other authorized disposal company to decommission your equipment.

5
Safety Introductions
On Safety
1. Before connecting the AC power cord to the DC adapter outlet make sure the
voltage designation of the DC adapter corresponds to the local electrical supply.
2. Never insert anything metallic into the cabinet openings of the medical LCD
monitor. Doing so may create the danger of electric shock.
3. To reduce the risk of electric shock, do not remove cover. No user-serviceable
parts inside. Only a qualied technician should open the case of the medical LCD
monitor.
4. Never use your medical LCD monitor if the power cord has been damaged. Do
not allow anything to rest on the power cord, and keep the cord away from areas
where people can trip over it.
5. Be sure to hold the plug, not the cord,when disconnecting the medical LCD
monitor power cord from an electric socket.
6. Unplug your medical LCD monitor power cord when it is going to be left unused
for an extended period of time.
7. Unplug your medical LCD monitor power cord from the AC outlet before any
service.
8. If your medical LCD monitor does not operate normally, in particular, if there are
any unusual sounds or smells coming from it, unplug it immediately and contact
an authorized dealer or service center.
9. Please contact the manufacturer if the set should be installed in an inaccessible
area.
Warning
yDo not touch input or output connectors and the patient simultaneously.
y This medical LCD monitor is intended for connection to input/output signals and
other connectors that comply with relevant IEC standard (e.g.,IEC60950 for IT
equipment and IEC60601 series for medical electrical equipment). In addition,
all such combination-system shall comply with the standard IEC 60601-1-1
or clause 16 of the 3Ed. of IEC60601-1, respectively, safety requirements for

CL24s USER’S MANUAL
medical electrical systems. Any person who has formed a combination-system is
responsible for the system to comply with the requirements of IEC 60601-1-1 or
clause 16 of the 3Ed. of IEC60601-1, respectively. If in doubt, contact qualied
technician or your local representative.
yTo avoid risk of electric shock, this equipment must only be connected to a supply
mains with protective earth.
yPower supply (AC/DC Adapter) is specied as a part of LCD Color Display.
yNot to position equipment to make it difcult to disconnect power cord plug
fromappliance inlet.
yDo not modify this equipment without authorization of the manufacturer.
On installation
1. Openings in the medical LCD monitor cabinet are provided for ventilation. To
prevent overheating, these openings should not be blocked or covered. If you put
the medical LCD monitor in a bookcase or some other enclosed space, be sure to
provide adequate ventilation.
2. Put your medical LCD monitor in a location with low humidity and a minimum of
dust.
3. Do not expose the medical LCD monitor to rain or use it near water (in kitchens,
near swimming pools, etc.). If the medical LCD monitor accidentally gets wet,
unplug it and contact an authorized dealer immediately. You can clean the medical
LCD monitor with a damp cloth if necessary, but be sure to unplug the medical
LCD monitor rst.
4. Place your medical LCD monitor near an easily accessible AC outlet.
5. High temperature can cause problems. Don’t use your medical LCD monitor in
direct sunlight and keep it away from heaters, stoves, replaces,and sources of
heat.
6. Don’t place your medical LCD Monitor on an unstable stand, Medical LCD monitor
may malfunction or fall.
7. This medical LCD monitor should not topple over when tilted at a 5° angle, in any
position, during NORMAL USE, excluding transport.

7
8. In the position specied for transport, medical LCD monitor shall not overbalance
when tilted at a 10 degree angle.
9. When carrying this product, please use both handles (if included) on the left and
right side of the product, and carry using two people. If you want the product to
be installed in another place, please call A/S center.
10. Do not use other cables or accessories that are not provided.
11. Do not lay this monitor on the other equipment.
Environmental Conditions for Operation and Storage
yTemperature range within 0°C to 40°C (operation), -20°C to 60°C (storage)
yRelative humidity range 30% to 80%
yAtmospheric pressure range within 500 to 1060hPa.
Intended Use
This Medical LCD Monitor is an accessory intended for use with Medical Equipment to
display alphabetical, numerical and graphical data.
Cautions
CAUTION
RISK OF ELECTRIC SHOCK
DO NOT OPEN
This symbol alerts the user that important literature concerning the operation
of this unit has been included. Therefore, it should be read carefully in order
to avoid potential problems.
This symbol warns users that un-insulated voltage within the unit may have
sufcient magnitude to cause electrical shock. Therefore, it is dangerous to
make contact with any part inside the unit. To reduce the risk of electrical shock, DO
NOT remove cover (or back). There are no user-serviceable parts inside. Refer
servicing to qualied service personnel. To prevent re or shock hazards, do not

CL24s USER’S MANUAL
expose this unit to rain or moisture. Do not use this unit’s polarized plug with an
extension cord receptacle or other outlets unless the prongs can be fully inserted.
This display is designed to meet the medical safety requirements for a patient vicinity
device. This device may not be used in connection with life support equipment.
Underwriters Laboratories (UL) Classication:
UL safety Compliance:
This medical LCD monitor is U.L. Classied WITH RESPECT TO ELECTRIC SHOCK,
FIRE AND MECHANICAL HAZARDS ONLY IN ACCORDANCE WITH UL 60601-1/CAN/
CSA C22.2 NO. 601.1
EEC Safety Compliance:
This medical LCD monitor unit meets the requirements of EN-60601-1 so as to
conform to the Medical Device Directive 93/42/EEC (general safety information). Use
120V rating 5-15P type plug only in the U.S.
This medical LCD monitor complies to the above standards only when used with the
supplied medical grade power supply.
24.1” Monitor (FS-L2403D) - ATM090-P240(ADAPTER)
Caution: Make sure the power cord is the correct type that is required in your area. This
medical LCD monitor has a universal power supply that allows operation in either 100-120V AC or
200-240V AC voltage areas (no user adjustment is required).
Use the proper power cord with correct attachment plug type. If the power source is 120 V AC,
use a power cord which is a Hospital Grade Power Cord with NEMA 5-15 style plug, labeled for 125
volts AC with UL and C-UL approvals. If the power source is a 240 V AC supply, use the tandem (T
blade) type attachment plug with ground conductor power cord that meets the respectiveEuropean
country’s safety regulations.
The hospital-grade plug for medical products intended for use in Denmark has DEMKO approval
and is rated 13 amps at 250Vac. Plug is recommended for use in medical applications and
specications are being added to the standard SB 107-2-D1. Plug mates with maker’s Danish
hospital-grade socket. Hospital sockets have slightly different shaped openings allowing only the
hospital plug, not the standard Danish plug, to be inserted, to protect the ac circuit in specic
medical settings. A ground post, located on the back of the display, may be used for the purpose
of grounding the display’s chassis. Any such ground must be installed in accordance with applicable
electrical codes. The groundpost is shown on the mechanical drawing found in this user’s guide.

9
Recycling
Follow local governing ordinances and recycling plans regarding the recycling
or disposal of this equipment.
Cleaning Instructions
Follow your hospital protocol for the handling of blood and body uids. Clean
the display with a diluted mixture of mild detergent and water. Use a soft
towel or swab. Use of certain detergents may cause degradation to the labels and
plastic components of the product. Consult cleanser manufacturer to see if agent is
compatible. Do not allow liquid to enter the display.
Servicing
Do not attempt to service the medical LCD monitor yourself, as opening or removing
covers may expose you to dangerous voltages or other hazards, and will void the
warranty. Refer all servicing to qualied service personnel. Unplug the medical LCD
monitor from its power source and refer servicing to qualied personnel under the
following conditions:
yIf the power cord or plug is damaged or frayed.
yIf liquid has been spilled into the medical LCD monitor.
yIf objects have fallen into the medical LCD monitor.
yIf the medical LCD monitor has been exposed to rain or moisture.
yIf the medical LCD monitor has been subjected to excessive shock by being
dropped.
yIf the cabinet has been damaged.
yIf the medical LCD monitor seems to be overheated.
yIf the medical LCD monitor emits smoke or abnormal odor.
yIf the medical LCD monitor fails to operate in accordance with the operating
instructions.

CL24s USER’S MANUAL
Accessories
Use only accessories specied by the manufacturer, or sold with the medical LCD
monitor.
Classication
yProtection against electric shock : Class I including AC/DC Adapter
yApplied Parts : No Applied Parts
yDegree of safety in the presence of ammable anesthetics mixture with air or with
oxygen or with nitrous oxide. Not suitable for use in the presence of a ammable
anesthetics mixture with oxygen or with nitrous oxide.
yMode of operation : Continuous.
FCC Information
This medical LCD monitor unit has been tested and found to comply with the limits
of a Class B digital device, pursuant to Part 15 of the FCC rules. These limits are
designed to provide reasonable protection against interference. This monitor can
radiate radio frequency energy and, if not installed and used in accordance with the
instructions, it may interfere with other radio communications equipment. There
is no guarantee that interference will not occur in a particular installation. If this
equipment is found to cause harmful interference to radio or television reception, the
user is encouraged to try to correct the interference by carrying out one or more of
the following measures:
1. Reorient or relocate the receiving antenna.
2. Increase the distance between the medical LCD monitor and the subject of
interference.
3. Plug the monitor into an outlet on a different electrical circuit than that to which
the subject of interference is connected.
4. Consult the dealer or an experienced radio/TV technician for help.

11
NOTICES TO USER
This device complies with part 15 of the FCC Rules. Operation is subject to the
following two conditions:
(1) this device may not cause harmful interference, and (2) this device must accept
any interference received, including interference that may cause undesired operation.
FCC WARNING
This medical LCD monitor generates or uses radio frequency energy. Changes or
modications to this medical LCD monitor may cause harmful interference unless
the modications are expressly approved in the instruction manual. The user could
lose authority to operate this equipment if an unauthorized change or modication is
made.
1. Guidance and manufacturer’s declaration - electromagnetic emission
The medical LCD monitor is intended for use in the electromagnetic environment specied below.
The customer or the user of the medical LCD monitor should assure that it is used in such an
environment.
Emission test Compliance Electromagnetic environment -guidance
RF Emissions CISPR 11 Group 1
The medical LCD monitor uses RF energy only for its
internal function. Therefore, its RF emissions are very
low and are not likely to cause any interference in
nearby electronic equipment.
RF Emissions CISPR 11 Class B The medical LCD monitor is suitable for use in all
establishments, including domestic establishments
and those directly connected to the public low-voltage
power supply network that supplies buildings used for
domestic purposes.
Harmonic emissions IEC
61000-3-2 D
Voltage uctuations IEC
61000-3-3 Compliance

CL24s USER’S MANUAL
2. Guidance and manufacturer’s declaration - electromagnetic immunity
This medical LCD monitor is intended for use in the electromagnetic environment specied below.
The customer or the user of the medical LCD monitor should assure that it is used in such an
environment.
Immunity test
IEC 60601
Test
level
Compliance
level
Electromagnetic environment-guidance
Electrostatic
discharge(ESD)
IEC 61000-4-2
6 kV contact
8 kV air
6 kV contact
8 kV air
Floors should be wood, concrete or
ceramic tile. If oors are covered with
synthetic material, the relative humidity
should be at least 30%
Electrical fast
transient/burst
IEC 61000-4-4
2 kV for
power
supply lines
1 kV for
input/
output lines
2 kV for
power
supply lines
1 kV for
input/
output lines
Main power quality should be that
of a typical commercial or hospital
environment.
Surge
IEC 61000-4-5
1 kV
differential
mode
2 kV
common
mode
1 kV
differential
mode
2 kV
common
mode
Main power quality should be that
of a typical commercial or hospital
environment.
Conducted RF
IEC 61000-4-6
3 Vrms
150 kHz to
80MHz
3 Vrms
150 kHz to
80MHz
Portable and mobile RF communications
equipment should be used no closer to
any part of the LCD monitor, including
cables, than the recommended
separation distance calculated from the
equation applicable to the frequency of
the transmitter.
Recommended separation distance :
d = [ 3.5 / V1 ] √P
Where P is the maximum output power
rating of the transmitter in watts (W)

13
3. Guidance and manufacturer’s declaration - electromagnetic immunity
This medical LCD monitor is intended for use in the electromagnetic environment specied below.
The customer or the user of monitor should assure that it is used in such an environment.
Immunity test IEC 60601 Test
level
Compliance
level
Electromagnetic environment-guidance
Power frequency
(50/60Hz) Magnetic
eld IEC 61000-4-8
3.0 A/m 3.0 A/m
Power frequency magnetic elds should
be at levels characteristic of a typical
location in a typical commercial or
hospital environment.
Voltage dips, short
Interruptions and
voltage variations
on power supply
input lines IEC
61000-4-l 1
<5%
U
T (>95% dip
in
U
T) for 0.5 cycle
40%
U
T (60% dip
in
U
T) for 5 cycle
70%
U
T (30% dip
in
U
T) for 25 cycle
<5%
U
T(<95% dip
in
U
T) for 5 sec.
<5%
U
T (>95% dip
in
U
T) for 0.5 cycle
40%
U
T (60% dip
in
U
T) for 5 cycle
70%
U
T (30% dip
in
U
T) for 25 cycle
<5%
U
T(<95% dip
in
U
T) for 5 sec.
Main power quality should be that of a
typical commercial or hospital
environment.
If the user of monitor requires
continued operation during power
mains interruptions, it is recommended
that monitor be powered from an
uninterruptible power supply or a
battery.
Note: UT is the A.C. mains voltage prior
to application of the test level.
Radiated RF
IEC 61000-4-3
3 V/m
80.0 MHz to
2.5 GHz
3 V/m
80.0 MHz to
2.5 GHz
Recommended separation distance
d = [ 3.5 / E1 ] √P 80MHz to
800MHz
d = [ 7 / E1 ] √P 80MHz to
2.5GHz
Where P is the maximum output
power rating of the transmitter
in watts (W) according to the
transmitter manufacturer and d is the
recommended separation distance in
meters(m). Field strengths from xed
RF transmitters, as determined by an
electromagnetic site survey, should be
less than the compliance level in each
frequency range.

CL24s USER’S MANUAL
4. Recommended separation distances between portable and mobile RF
communications equipment and this medical LCD monitor.
The medical LCD monitor is intended for use in an electromagnetic environment
in which radiated RF disturbances are controlled. The customer or the user of the
monitor can help prevent electromagnetic interference by maintaining a minimum
distance between portable and mobile RF communications equipment (transmitters)
and the medical LCD monitor as recommended below, according to the maximum
output power of the communications equipment.
Rated maximum
output power [W]
of transmitter
Separation distance according to frequency of transmitter[m]
150kHz to 80MHz
d = [ 3.5 / V1 ] √P
80MHz to 800MHz
d = [ 3.5 / E1 ] √P
800MHz to 2.5GHz
d = [ 7 / E1 ] √P
V1=3Vrms E1=3V / m E1=3V / m
0.01 0.116 0.116 0.2333
0.1 0.368 0.3687 0.7378
1 1.166 1.1660 0.2333
10 3.687 3.6872 0.7375
100 11.660 11.6600 23.333
For transmitters rated at a maximum output power not listed above, the recommended separation
distance(d) in meters (m) can be estimated using the equation applicable to the frequency of
the transmitter, where P is the maximum output power rating of the transmitter in watts (W)
according to the transmitter manufacturer.
Note
1. At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
2. These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reection from structures, objects and people.

15
Parts
24.1” CL24s (FS-L2403D) Monitor
Accessories
USB Driver
User Manual D-SUB Cable
(6ft/1.8m)
AC-DC Adapter
(ATM090-P240)
S-Video (6ft/1.8m)
DVI Cable
(6ft/1.8m)
BNC Cable
(6ft/1.8m)
AC Power cord
(6ft/1.8/m US,EU, JP)
(Hospital Grade)

CL24s USER’S MANUAL
Connectors
24.1” CL24s (FS-L2403D) Monitor Connector
Mechanical Product Drawing
24.1” CL24s (FS-L2403D) Dimension

17
Mechanical Dimension
Our ergonomic Standbase allows the monitor screen to be easily rotated between
portrait and landscape formats. You can also easily make height, tilt, and swivel
adjustment.
Height adjust
(0~108mm)
108mm
Rotate
Landscape / portrait
(clockwise)
Swivel
(-30° ~ +30°) 30° 30°
Tilt
(-3° ~ +30°)
30°
3°
To adjust the panel orientation:
yBefore you rotate the display panel, rst please make sure you pull in the bottom
of the monitor toward you at its maximum tilt position, If not, the corner of the
monitor may cause scratches on the surface of the Standbase during the rotation.
yTo change from landscape to portrait, turn clockwise.
yTo change from portrait to landscape, turn counterclockwise.