XS Instruments 70 Vio Series User manual

1
Rev. 1 24/01/2020
EN
70 Vio
INSTRUCTIONS MANUAL
pH - Cond - PC

2
Rev. 1 24/01/2020
EN

3
Rev. 1 24/01/2020
EN
Index
Introduction .............................................................................................................................................6
Safety information ...................................................................................................................................7
•Reporting terms: .................................................................................................................................7
•Additional documents for safety.........................................................................................................8
•Use according to destination...............................................................................................................8
•Basic requirements for a safe use .......................................................................................................8
•Unauthorized use ................................................................................................................................8
•Device maintenance............................................................................................................................8
•Responsibility of the owner of the instrument ...................................................................................9
Instrumental features ..............................................................................................................................9
•Parameters ..........................................................................................................................................9
•Datasheet .........................................................................................................................................10
Instrument description ..........................................................................................................................11
•Display ...............................................................................................................................................11
•Keyboard ...........................................................................................................................................11
•LED.....................................................................................................................................................12
Installation .............................................................................................................................................12
•Supplied components........................................................................................................................12
•Start-up..............................................................................................................................................12
•Connection of the power supply.......................................................................................................12
•Power on, date and time update, power off.....................................................................................13
•Replacement of batteries..................................................................................................................13
•Instrument transportation ................................................................................................................14
•Key functions.....................................................................................................................................14
•Inputs / Outputs connections ...........................................................................................................15
READ THE MANUAL BEFORE PROCEEDING TO CONNECT THE PROBES OR PERIPHERALS ........................15
•Symbols and icons on the display .....................................................................................................15
Operation of the device.........................................................................................................................16
Setup menu ...........................................................................................................................................17
•Setup Menu Structure .....................................................................................................................18
Temperature measurement ATC – MTC ...............................................................................................19
pH Parameter ........................................................................................................................................19
•pH parameter Setup .........................................................................................................................19
•Composition of the Setup menu for pH parameter ..........................................................................20
•Automatic pH calibration .................................................................................................................21
•Calibration with manual values.........................................................................................................23
•Performing pH measurement ...........................................................................................................23

4
Rev. 1 24/01/2020
EN
•Sensors with DHS technology............................................................................................................24
•Errors during calibration....................................................................................................................25
mV Parameter ...................................................................................................................................25
ORP Parameter (Oxide-Reduction potential) ..................................................................................25
•ORP parameter Setup .......................................................................................................................25
•Composition of the setup Menu for the ORP parameter..................................................................26
•ORP automatic calibration ................................................................................................................27
Conductivity Parameter ....................................................................................................................27
•…how to get Conductivity? ...............................................................................................................27
•Setup for Conductivity parameter.....................................................................................................28
•Automatic COND calibration ............................................................................................................31
•Manual COND calibration .................................................................................................................32
•Errors during calibration ..................................................................................................................33
•Performing Conductivity measurement ...........................................................................................33
Other measurements carried out with the Conductivity cell............................................................34
•TDS Parameter .................................................................................................................................34
•Salinity ..............................................................................................................................................35
Data Logger Function ........................................................................................................................35
•Setup for Data Logger parameter .....................................................................................................35
•Composition of setup menu for the Data Logger menu ...................................................................36
•Example of automatic Data Logger mode .........................................................................................36
•Example of manual Data Logger mode .............................................................................................37
•Recall Memory .................................................................................................................................37
•Clear the saved data..........................................................................................................................37
Instrument Setup menu ..................................................................................................................37
•Composition of the setup menu for Setting menu ...........................................................................38
DataLink+ Software (per Windows 7/8/XP/10) .................................................................................39
•Functions ...........................................................................................................................................40
Warranty ...........................................................................................................................................41
•Warranty period and limitations .......................................................................................................41
Disposal of electrical devices ............................................................................................................41
XS Instruments
Via della Meccanica n.25 41012 Carpi (MO) ITALY
Tel. +39 059.653274 Fax +39 059653282
www.xsinstruments.com

5
Rev. 1 24/01/2020
EN

6
Rev. 1 24/01/2020
EN
Introduction
XS Instruments, globally recognized as a leading brand in the field of electrochemical measurements, has
developed this new line of portable instruments completely produced in Italy, finding the perfect balance
between performance, attractive design and ease of use.
The robustness and integrity of the case, the integrated brightness sensor and the practical carrying case
make this instrument ideal for measurements directly in the field.
Thanks to the triple power supply and the ability to manually change the contrast and brightness of the
display, this instrument is also suitable for use in the laboratory.
The innovative high definition colour LCD display shows all the necessary information, such as the
measurement, the temperature, the buffers used for the last calibration (also custom), the condition of
stability.
Everyone can use these tools thanks to the instructions that appear directly on the display. The calibration
is guided step by step and the instrument configuration menu is easy to consult. In addition, a LED
indicates the status of the system to the user.
Up to 3 pH calibration points can be carried out between 10 automatically recognized values and 5 points
for Conductivity; in addition, buffers chosen by the operator can be used.
It is also possible to perform mV calibration for Redox sensors.
For an accurate measurement of the conductivity value, it is possible to work with 3 different cell
constants and to modify the compensation coefficient and the reference temperature.
It is possible to consult the calibration data anytime and the representation makes the calibration process
more efficient, through the icons of the buffers used.
Automatic or manual Data Logger function with values that can be stored in different GLP formats on the
internal memory (1000 data) or on the PC.
The ideal solution for an accurate and precise measurement is to use an XS Sensor electrochemical
electrode with an XS Instruments device and perform the calibrations with XS Solution certified
calibration solutions.

7
Rev. 1 24/01/2020
EN
Safety information
•Definition of warning words and symbols
This manual contains extremely important safety information, in order to avoid personal injury, damage to
the instrument, malfunctions or incorrect results due to failure to comply with them. Read entirely and
carefully this manual and be sure to familiarize with the tool before starting to work with it.
This manual must be kept near to the instrument, so that the operator can consult it easily, if necessary.
Safety provisions are indicated with warning terms or symbols.
•Reporting terms:
ATTENTION for a medium-risk hazardous situation, which could lead to serious injury or death, if
not avoided.
ATTENTION for a dangerous situation with reduced risk which can cause material damage, data loss
or minor or medium-sized accidents, if not avoided.
WARNING for important information about the product
NOTE for useful information about the product
Warning symbols:
Attention
This symbol indicates a potential risk and warns you to proceed with
caution.
Attention
The instrument must be used following the indications of the reference
manual. Read the instructions carefully.
Advice
This symbol draws attention to possible damage to the instrument or
instrumental parts.
Note
This symbol highlights further information and tips.
Attention
This symbol draws attention to a possible danger from electric current.

8
Rev. 1 24/01/2020
EN
•Additional documents for safety
The following documents can provide the operator with additional information to work with the measuring
system safely:
•operating manual for electrochemical sensors;
•safety data sheets for buffer solutions and other maintenance solutions (e.g. storage);
•specific notes on product safety.
•Use according to destination
This instrument is designed exclusively for electrochemical measurements both in the laboratory and
directly in the field.
Pay attention to the technical specifications shown in the INSTRUMENT FEATURES / TECHNICAL DATA table;
any other use is to be considered unauthorized.
This instrument has been manufactured and tested in compliance with EN 61010-1 safety standards relating
to electronic instruments and has left the factory in perfect technical and safety conditions (see test report
in each package).
The regular functionality of the device and the operator safety are guaranteed only if all the normal
laboratory safety standards are respected and if all the specific safety measures described in this manual
are observed.
•Basic requirements for a safe use
The regular functionality of the device and the operator safety are guaranteed only if all the following
indications are respected:
•the instrument can be used in accordance with the specifications mentioned above only;
•use the supplied power supply only. If you need to replace the power supply, contact your local
distributor;
•the instrument must operate exclusively in the environmental conditions indicated in this manual;
no part of the instrument can be opened by the user.
Do this only if explicitly authorized by the manufacturer.
•Unauthorized use
The instrument must not run, if:
•it is visibly damaged (for example due to transportation);
•it has been stored for a long period of time in adverse conditions (exposure to direct light, heat sources
or places saturated by gas or vapours) or in environments with conditions different from those
mentioned in this manual.
•Device maintenance
If used correctly and in a suitable environment, the instrument does not require maintenance procedures.
It is recommended to occasionally clean the instrument case with a damp cloth and a mild detergent. This
operation must be performed with the instrument off, disconnected from the power supply and by
authorized personnel only.

9
Rev. 1 24/01/2020
EN
The housing is in ABS / PC (acrylonitrile butadiene styrene / polycarbonate). This material is sensitive to
some organic solvents, for example toluene, xylene and methyl ethyl ketone (MEK).
If liquids get into the housing, they could damage the instrument.
In case of prolonged non-use of the device, cover the BNC connectors with the special cap.
Do not open the instrument housing: it does not contain parts that can be maintained, repaired or replaced
by the user. In case of problems with the instrument, contact your local distributor.
It is recommended to use original spare parts only. Contact your local distributor for information. The use
of non-original spare parts can lead to malfunction or permanent damage to the instrument. Moreover, the
use of spare parts not guaranteed by the supplier can be dangerous for the user himself.
For the maintenance of the electrochemical sensors, refer to the documentation present in their packaging
or contact the supplier.
•Responsibility of the owner of the instrument
The person who owns and uses the tool or authorizes its use by other people is the owner of the tool and
is responsible for the safety of all users of the tool and third parties.
The owner of the instrument must inform users of the use of the same safely in their workplace and on the
management of potential risks, also providing the required protective devices.
When using chemicals or solvents, follow the manufacturer's safety data sheets.
Instrumental features
•Parameters
pH 70 Vio: pH, mV, ORP, Temp
COND 70 Vio: Cond, TDS, Sal, Temp
PC 70 Vio: pH, mV, ORP, Cond, TDS, Sal, Temp

10
Rev. 1 24/01/2020
EN
•Datasheet
Series 70 Vio
pH
pH 70 Vio -PC 70 Vio
Measuring range
-2…16
Resolution / Accuracy
0.1, 0.01 / +0.02
Recognized calibration points and
buffers
AUTO: 1…3 / USA, NIST
CUS: 2 user values
Buffers indication
Yes
Calibration report
Yes
Automatic DHS recognition
Yes
Stability filter
Low – Med - High
mV
pH 70 Vio -PC 70 Vio
Range / Resolution
Range: -1000 … +1900 / Resolution: 0.1 / 1
ORP
pH 70 Vio -PC 70 Vio
Calibrations points
1 point / 475 mV
Conductivity
COND 70 Vio -PC 70 Vio
Range / Resolution
0,00 – 20,00 – 200,0 – 2000 μS /
2,00 – 20,00 – 200,0 mS
Automatic scale
Recognized calibration points and
buffers
1…5 / 84, 147, 1413 μS, 12.88, 111.8 mS,
1 user value
Reference temperature
15…30 °C
Temperature coefficient
0,00…10,00 %/°C
TDS
COND 70 Vio -PC 70 Vio
Measuring range / TDS Factor
0,1mg/l…200 gr/l/ 0.40…1.00
Salinity
COND 70 Vio -PC 70 Vio
Measuring range
0,1ppm …100ppt
Temperature
pH 70 Vio - COND 70 Vio -PC 70 Vio
Measuring range
-10…110 °C
Resolution / Accuracy
0,1 / ± 0,5°C
Temperature compensation ATC
(NTC30KΩ) and MTC
0…100 °C
System
GLP with calibration Timer
Yes
Internal memory
1000 Dates
Display
High definition colours LCD
Brightness and contrast management
Manual and automatic with integrated sensor
IP protection
IP 57
Power supply
AA 1,5 V – 3 batteries / 5 V power supply with USB cable
Sound level during standard operation
< 80 dB
Environmental operating conditions
0 … +60 °C
Maximum permissible humidity
< 95 % non-condensing
Maximum altitude of use
2000 m
System dimensions
185 x 85 x 45 mm
System weight
400 g
Reference regulations
EMC 2014/30/UE
RoHS 2011/65/EU
EN 61326-1
EN 61010-1

11
Rev. 1 24/01/2020
EN
Instrument description
•Display
•Keyboard
Down directional key to scroll through
the menus and parameters and to
manage the recall of data
MODE key to scroll through the
various measurement parameters
Enter / Menu key to confirm the chosen values
and enter the setup from the measure mode
Measurement
parameter
Indication of
instrumental
mode
Representation of
buffers used for
calibration
Error symbol
Actual value / Measure unit
Temperature and type of
compensation
ATC - automatic NTC 30KΩ
MTC - manual
Information icons
on connectivity and Data
Logger
Progressive number of data saved in
Data Logger mode
Stability
indicator
Text string
Date and time
Battery charge level
Text string
On-Off button to
switch the device on
and off
Meas / Cal key to return to measure
mode and to start calibration
Up directional key for scrolling
through menus and parameters
and managing data saving

12
Rev. 1 24/01/2020
EN
•LED
All the instruments are equipped with a two-colours LED (red and green) which provides the user with
important information on the status of the system:
Function
LED
Description
Power on
Fixed
Power off
Fixed
Standby
Flashing every 20 s
Stable measure
Flashing every 3 s
Errors during calibration
Flashing every 1 s
Errors during measurement
Flashing every 3 s
Time of saving the data
On / Off in rapid succession
Recall Memory mode
Alternate green and red, pause 5 s
Selection confirmation
Switched on for 1 s
Timed screens
Fixed
DHS deactivation
Fixed
Installation
•Supplied components
The instrument is always supplied inside the specific carrying case; the version without sensor is
always supplied with:
instrument with batteries, 5V adapter with USB cable, 1m S7/BNC connection cable, NT55 temperature
probe, buffer solutions in single-dose bottle and / or sachet, paper tissues, screwdriver, beaker, electrode
holder support -only for multi-parameter- multilingual user manual and test report.
There are versions with the sensor(s) already included. Contact your local distributor to be updated on the
correct composition of the sales kit.
•Start-up
•The device leaves the factory ready to be used by the user.
•Batteries are included.
•Connection of the power supply
•in addition to batteries, the instrument can be powered through electricity grid;
•check that the electrical standards of the line on which the instrumentation is to be installed
comply with the voltage and operating frequency of the instrument;
•use the original power supply only;
•connect the power supply to the USB cable and the other end of the cable (Micro USB) to the
Micro USB port located on the front of the instrument;
•Connect the power supply to an electric socket easy to reach.

13
Rev. 1 24/01/2020
EN
The electricity supply can be originated from the power grid and from the USB port of a PC too.
If the instrument is powered by PC, the icon indicating the battery status disappears from the display .
Opening the DataLink+ software, on the display it is shown this icon .
•Power on, date and time update, power off
Turn on the system by pressing the button . The display initially activates all segments and then
appears:
•model and firmware of the instrument;
•settings relating to the most important parameters and possible information about the DHS sensor;
On first use, and after each battery replacement, the instrument during the start-up phase will request the
updating of the date and time.
•Using the directional keys, update the year and confirm with the key . Carry out the same
operation with the month and day, and subsequently with hours and minutes.
•The instrument will enter measure mode in the last parameter used.
To switch off the instrument, press the key in measure mode.
•Replacement of batteries
The instrument works with 3 AA 1.5V batteries.
To proceed with the replacement:
1. Turn off the device.
Turn the instrument over with the display facing down and place it on a stable surface.
It is advisable to put a cloth to avoid any scratching on display.
2. Using the screwdriver supplied, completely unscrew the screw close to the battery symbol.
3. Remove the battery stopper cap with the help of the lanyard.
4. Remove the 3 exhausted batteries (one in the left compartment and two in the right compartment)
and insert the new ones. Pay attention to the correct polarity. Follow the diagram above the battery
symbol in the rear compartment of the instrument.
5. Reinsert the battery holder and tighten the screw.
ATTENTION
Danger of death or serious injury from electric shock.
Contact with live components can lead to injury or death.
•Use the adapter supplied only.
•Do not put the power supply in contact with liquids nor in a condensing environment. Avoid
thermal shock.
•All electrical cables and connections must be kept away from moisture or liquids.
•Check that the cables and plugs are not damaged, otherwise replace them.
•During use, do not cover the power supply and/or do not place it inside containers.

14
Rev. 1 24/01/2020
EN
•Instrument transportation
The instrument is always supplied with the appropriate carrying case. Use the original case only, to transport
the instrument. If you need to buy it again, contact your local distributor.
The interior of the case is shaped to be able to house the instrument and the sensors still connected.
•Key functions
Button
Pression
Function
Short
Press to turn the device on or off.
Short
In measure mode, press to scroll through the different parameters:
•pH 70 Vio: pH mV ORP
•COND 70 Vio: Cond TDS Sal
•PC 70 Vio: pH mV ORP Cond TDS Sal
Short
•In calibration, setup and memory recall mode, press to return to measure
mode.
•In measure mode, press to start the calibration.
Short
In measure mode, press to enter the setup.
In the setup menus, press to select the desired program and / or value.
During calibration, press to confirm the value.
Short
In the setup and subset menus press to scroll
In the setup submenus, press to change the value
In memory recall mode, press to scroll through the saved values.
In MTC and customer calibration mode, press to change the value.
: In measure mode, press to save the data (manual Data Logger) or start
and end the recording (Automatic Data Logger).
: In measure mode, press to recall the saved data.
Long-press
(3s)
In measure mode, keep one of the two keys pressed to change the temperature
in MTC mode (manual compensation, without probe). When the value starts to
flash, the user can change the temperature value by entering the correct one and
confirming with .
IMPORTANT:
•When the Sleep mode is active (by default after two minutes of inactivity of the instrument) press
any key to reactivate the brightness of the display.
•Only at this point do the keys regain their function.

15
Rev. 1 24/01/2020
EN
•Inputs / Outputs connections
Use original accessories guaranteed by the manufacturer only.
If necessary, contact your local distributor.
The BNC connectors are protected by a plastic cap.
Remove the cap before connecting the probes.
PC 70 Vio top panel
READ THE MANUAL BEFORE PROCEEDING TO CONNECT THE PROBES OR PERIPHERALS
•Symbols and icons on the display
Symbol
Description
Symbol
Description
M+
Number of data stored in Data Logger
mode on instrumental memory
Error in measurement or calibration
Instrument connected to the DataLink+
software
FIXED: Automatic Data Logger set
INTERMITTENT:
Automatic Data
Logger in operation
FIXED: Calibration deadline set for that
parameter
INTERMITTENT: Calibration deadline
active for that parameter
Press the directional keys to change
the parameter or value on the
display
DHS digital sensor active
Battery charge indication
Measurement stability indicator
The bars scroll if the measurement
is not stable
RCA for pH temperature
probe
BNC for pH/ORP electrodes
BNC for conductivity cell
RCA for Conductivity
temperature probe
Micro USB for connection
to PC and power supply

16
Rev. 1 24/01/2020
EN
Operation of the device
•After the switching on, the instrument enters measure mode in the last parameter used.
•To scroll through the different parameter screens, press the key ; the current measurement
parameter is shown in the display on the top left (e.g.: ).
Sequence of parameters in measure mode:
pH 70 Vio
COND 70 Vio
PC 70 Vio
pH Conductivity pH
mV TDS mV
ORP Salinity ORP
Conductivity
TDS
Salinity
Note: Pressing the button after the last parameter, the instrument automatically restarts from the
first one.
In the measurement screens for the pH, ORP and Conductivity parameters, press the key to start the
calibration of the active parameter. (next paragraphs).
On the left side of the display, through a string of different colours, it is always indicated how the
instrument is located.
Note: in order to confirm to the user the switching from one mode to another, the string flashes.

17
Rev. 1 24/01/2020
EN
String
Meaning
The instrument is in measure mode.
The instrument is in calibration (automatic or manual in relation to
the user's choice).
The user is in the setup mode.
The configuration menus can concern the characteristics of the
parameters or the general setting of the instrument.
The instrument is in the Recall Memory mode.
The data stored are being displayed by running the manual or
automatic Data Logger.
Setup menu
•In measure mode, press the key to enter SETUP mode, select the parameter you want to edit
by using the directional keys and confirming with .
pH 70 Vio
COND 70 Vio
PC 70 Vio
PH SETTINGS COND SETTINGS PH SETTINGS
ORP SETTINGS TDS SETTINGS ORP SETTINGS
LOG SETTINGS LOG SETTINGS COND SETTINGS
SETTINGS SETTINGS TDS SETTINGS
LOG SETTINGS

18
Rev. 1 24/01/2020
EN
SETTINGS
•Within the selected menu, move between the different programs using the directional buttons and
press the button to access the submenu you want to edit.
•Using the keys and choose the desired option or change the numerical value and confirm
with .
•The value or parameter that is being edited is recognizable as it flashes on the display.
•The icon indicates that the value or parameter to choose is editable using the directional keys.
•Press the key to return to the previous menu.
•Setup Menu Structure
P1.0
PH SETTINGS
P1.1
Buffer Selection
P1.2
Resolution
P1.3
Set Stability
P1.6
View pH Cal
P1.7
Due pH Cal
P1.8
Reset pH Setting
P1.9
Temp Cal pH
P2.0
ORP SETTINGS
P2.6
View ORP Cal
P2.7
Due ORP Cal
P2.8
Reset ORP Setting
P2.9
Temp Cal ORP
P3.0
COND SETTINGS
P3.1
Cell Constant
P3.2
Buffer Selection
P3.3
Reference Temp
P3.4
Temp. Compensation Factor
P3.6
View Cond Cal
P3.7
Due Cond Cal
P3.8
Reset Cond Setting
P3.9
Temp Cond pH
P4.0
TDS SETTING
P4.1
TDS Factor
P8.0
LOG SETTINGS
P8.1
Data Logger Type

19
Rev. 1 24/01/2020
EN
P8.2
Clear Data
P9.0
SETTINGS
P9.1
Temperature U.M.
P9.2
Time and data Set
P9.3
Backlight mode
P9.4
Brightness
P9.5
Sleep Mode
P9.6
Parameters Setup
P9.8
Reset
P9.9
Auto Power-Off
Temperature measurement ATC – MTC
•ATC: The direct measurement of the sample temperature for all parameters is carried out through the
NTC 30KΩprobe, which can be either integrated into the sensor (electrode and / or cell) or external.
•MTC: If no temperature probe is connected, the value must be changed manually: keep pressed
or until the value starts to flash; then adjust it by continuing to use the directional keys; press
to confirm.
pH Parameter
pH 70 Vio; PC 70 Vio
On this series of devices, it is possible to use pH sensors with integrated temperature probe or to connect
two different sensors. Connect the pH electrode to the BNC type connector marked in green. Connect the
temperature probe to the RCA / CINCH Temp connector always marked with a green background.
The instrument is also able to recognize the DHS sensor, an innovative electrode that stores calibration data
and that can be used immediately after on any enabled instrument.
•pH parameter Setup
•In measure mode press to access the SETUP menu.
•Press the button to access the pH SETTINGS P1.0 menu.
•Move with the keys and to select the program you want to access.
The table below shows the setup menu structure for the pH parameter, and for each program the options
that the user can choose and the default value:

20
Rev. 1 24/01/2020
EN
•Composition of the Setup menu for pH parameter
Program
Description
Options
Factory Default Settings
P1.1
CAL BUFFER SELECT
USA – NIST – Custom
USA
P1.2
SELECT RESOLUTION
0.1 – 0.01
0.01
P1.3
STABILITY FILTER
LOW – MEDIUM - HIGH
MED
P1.6
CALIBRATION DATA
-
-
P1.7
SET DUE CAL
NO – HOURS - DAYS
NO
P1.8
RESET SETTINGS
YES – NO
NO
P1.9
TEMPERATURE CAL
YES – NO
-
P1.1 pH buffer selection
•Access this setup to select the buffer family for performing the electrode calibration.
Calibration lines from 1 to 3 points can be made.
During calibration, press to exit and save the points calibrated up to that moment
(see calibration paragraph).
The instrument automatically recognizes 2 families of buffers (USA and NIST); in addition, the user has
the option of performing a manual calibration of up to 2 points with customizable values:
USA Buffers: 1,68 - 4,01 - 7,00** - 10,01 - 12,45 (Factory setting)
NIST Buffers: 1,68 - 4,00 - 6,86** - 9,18 - 12,46
**Neutral point always requested as first
In measure mode at the bottom left of the display, a series of
beakers indicates the buffers with which the last automatic and
manual calibration was carried out.
P1.2 Resolution
Access this menu to choose the resolution needed, when reading the
pH parameter:
•0.1
•0.01 -default-
P1.3 Stability criterion in pH measurement
To consider the reading of a value truthful, we recommend waiting for the measurement stability, indicated
by the icon . When the measurement is not stable, four red flashing bands appear on the display.
Access this menu to change the measurement stability criterion:
“LOW”: choose this option to bring up the stability icon even in conditions of poor stability.
Readings included within 1.2 mV.
“MEDIUM” (default value): readings included within 0.6 mV.
“HIGH”: choose this option to display the stability icon only in conditions of high measurement stability,
readings included within 0.3 mV.
P1.6 pH calibration data
Beaker
pH value of the buffer
Acid
< 6.5
Neutral
6.5 ~ 7.5
Basic
> 6.5
This manual suits for next models
3
Table of contents
Other XS Instruments Measuring Instrument manuals

XS Instruments
XS Instruments 7 Vio User manual

XS Instruments
XS Instruments OXY 7 Vio User manual

XS Instruments
XS Instruments pH 70 Plus User manual

XS Instruments
XS Instruments COND 7 Plus User manual

XS Instruments
XS Instruments Temp70 PT100 vio User manual

XS Instruments
XS Instruments 60 VioLab Series User manual
Popular Measuring Instrument manuals by other brands

GHM
GHM Delta OHM LPPYRA03 operating manual

Eastron
Eastron SDM230 Series manual

Tradinco
Tradinco TRAQC-3 LL user manual

Merlin
Merlin EVO SM user manual

Ecolab
Ecolab Nalco Water STORM Settling Rate Monitor Installation, operation & maintenance manual

GARO
GARO GNM3D-RS485 Installation and programming manual