YOKOGAWA GC8000 User manual

Technical
Information
Contents
Introduction...................................................................................................2
Safety Precautions..................................................................................................3
1. How to install ................................................................................................6
1.1 Data required for AI analysis .............................................................................. 6
1.2 Download software/initial free trial license....................................................... 6
1.3 Performing an Installation .................................................................................. 6
2. Initial Installation ..........................................................................................7
2.1 Activation of initial free trial license .................................................................. 7
2.2 Conrmationofoperation................................................................................... 8
2.2.1 Conrmationofhealthdiagnosisoperation .......................................... 8
2.2.2 Operationcheckoftheresolutionpredictionfunction ........................ 10
2.3 Checkitemsbeforemovingtothepaidversion............................................. 11
3. Paid License............................................................................................... 12
3.1 Paid License Usage Notes ................................................................................ 12
3.2 TypesofPaidLicenses...................................................................................... 12
3.3 Purchase Flow of Paid Licenses...................................................................... 13
Revision Information ........................................................................................... 16
ForGC8000
GasChromatograph
AIMaintenanceSupport
InstallationandLicenseGuide
TI 11B06H01-01EN
TI 11B06H01-01EN
©CopyrightMar.2023(YK)
1stEditionMar.2023(YK)

For GC8000 Gas Chromatograph AI Maintenance Support Installation and License Guide
2
TI 11B06H01-01EN Mar. 17, 2023-00
Introduction
Thismanualprovidestechnicalinformationfortheinitialsoftwareinstallationandrenewal
methodsforGasChromatographAIMaintenanceSupport(hereinafterreferredtoasGCAI):a
supportsoftwareforgaschromatographGC8000(hereinafterreferredtoasGC).
Afreetriallicenseofthissoftwareallowsyoutocheckcompatibilitywiththeindividual
applicationsyouneed.Afterconrmingthatitisapplicable,referto“GasChromatographAI
MaintenanceSupportHandbook”(TI11B06H01-02EN)andperformadetailedsetupformore
accuratedetection.
Therelevantdocumentsareasfollows:
Contents Document number Note
GasChromatographAIMaintenanceSupport GS11B06H01-01EN Generalspecications
Onlinemanual
GasChromatographAIMaintenanceSupport
InstallationandLicenseGuide
TI 11B06H01-01EN Licenseguide
Onlinemanual(thismanual)
GasChromatographAIMaintenanceSupport
Handbook
TI 11B06H01-02EN Onlinemanual
GasChromatographAIMaintenanceSupport
OperationManual
IM 11B06H01-01EN Onlinemanual
GC8000ProcessGasChromatograph IM 11B08A01-01E Onlinemanual
ASETAnalyzingServerEngineeringTerminal
Software
IM 11B06C01-01E Onlinemanual
PCASPCAnalyzerServerSoftware IM 11B06B01-01E Onlinemanual
nPrerequisites for this document
• Abilitytoreadandinterpretchromatogramstodetermineiftheyarenormalornot.
• KnowledgeofPCASoperationsandtheconstructionofstoreddata.
• KnowledgeofWindowssoftwareinstallationanduse.
nNotes
• Pleasereadthedocumentsabovementioned(GS,TI,IM)carefullyandunderstandthe
contentsbeforeoperatingthisproduct.
• Themanualisintendedtodescribethefunctionsoftheproductanddoesnotguaranteeits
suitabilityforyourparticularpurpose.
• Nopartoftheuser’smanualsmaybetransferredorreproducedwithoutpriorwritten
consentfromYOKOGAWA.
• Thecontentsofthismanualaresubjecttochangewithoutnotice.
• Ifyouhaveanyquestions,oryoundmistakesoromissionsintheuser’smanuals,please
contactoursalesrepresentativeoryourlocaldistributor.
lProduct Disclaimer
• Wemakenowarrantieswithrespecttotheproductsexceptassetforthinthewarranty
terms.
• Wearenotresponsibleforanydamagesincurredbyyouoranythirdpartyasaresultofthe
useofthisproduct,orforanydamagesorindirectdamagesofanykindincurredbyyouor
anythirdpartyduetodefectsinthisproductthatwecannotforeseeorpredict.
No.IM11B06H01-01EN 1stEdition:Mar.2023(YK)
AllRightsReservedCopyright©2023,YokogawaElectricCorporation

3
TI 11B06H01-01EN
For GC8000 Gas Chromatograph AI Maintenance Support Installation and License Guide
Mar. 17, 2023-00
Safety Precautions
n Precautionsfortheproduct’sprotection,safety,andmodication
• Inordertoprotectthesystemcontrolledbytheproductandtheproductitselfandensure
safeoperation,observethesafetyprecautionsdescribedinthisuser’smanual.Weassume
noliabilityforsafetyifusersfailtoobservetheseinstructionswhenoperatingtheproduct.
• Useoftheproductinamannernotspeciedinthismanualmayimpairitsprotective
functions.
• Wheninstallingprotectionorsafetycircuitsforthisproductorthesystemitcontrols,install
themseparatelyoutsideofthisproduct.
• Whenreplacingpartsorconsumablesofthisproduct,besuretouseourspeciedproducts.
• Modicationofthisproductisstrictlyprohibited.
n NotesonHandlingUser’sManuals
• Pleasehandovertheuser’smanualstoyourenduserssothattheycankeeptheuser’s
manualsonhandforconvenientreference.
• Pleasereadtheinformationthoroughlybeforeusingtheproduct.
• Thepurposeoftheseuser’smanualsisnottowarrantthattheproductiswellsuitedtoany
particularpurposebutrathertodescribethefunctionaldetailsoftheproduct.
• Nopartoftheuser’smanualsmaybetransferredorreproducedwithoutpriorwritten
consentfromYOKOGAWA.
• YOKOGAWAreservestherighttomakeimprovementsintheuser’smanualsandproductat
anytime,withoutnoticeorobligation.
• Ifyouhaveanyquestions,oryoundmistakesoromissionsintheuser’smanuals,please
contactoursalesrepresentativeoryourlocaldistributor.
nWarning and Disclaimer
YOKOGAWAmakenoguaranteeofanykindwithrespecttothisproduct,exceptasstatedinthe
warrantystatement.
Weshallnotbeliableforanydamagesorforanyindirectdamagesincurredbyyouoranythird
partyasaresultoftheuseofthisproductorforanyunforeseeabledefectinthisproduct.
nNotes on Software
• YOKOGAWAmakesnowarranties,eitherexpressedorimplied,withrespecttothe
software’smerchantabilityorsuitabilityforanyparticularpurpose,exceptasspeciedinthe
termsofwarranty.
• Thisproductmaybeusedonamachineonly.Ifyouneedtousetheproductonanother
machine,youmustpurchaseanotherproduct.
• Itisstrictlyprohibitedtoreproducetheproductexceptforthepurposeofbackup.
• Itisstrictlyprohibitedtoperformanyreverse-engineeringoperation,suchasreverse
compilationorreverseassemblingontheproduct.
• Nopartoftheproductmaybetransferred,convertedorsubletforusebyanythirdparty,
withoutpriorwrittenconsentfromYOKOGAWA.

For GC8000 Gas Chromatograph AI Maintenance Support Installation and License Guide
4
TI 11B06H01-01EN Mar. 17, 2023-00
nDocumentation Conventions
l SymbolMarks
Throughoutthisuser’smanual,youwillndseveraldierenttypesofsymbolsareusedtoidentify
dierentsectionsoftext.Thissectiondescribestheseicons.
SEE
ALSO
Indicatesreferences.
CAUTION
Indicatesthatoperatingthehardwareorsoftwareinthismannermaydamageitorleadtosystem
failure
TIP
Givesinformationthatcomplementsthepresenttopic.
l KeyboardInscriptions
Charactersenclosedbybrackets[],greater-thansign>,BoldItalicfont
Whencharactersareenclosedbybracketsinthedescriptionofakeyorbuttonoperation,it
indicatesakeyonthekeyboard,abuttonnameinawindow,oraniteminalistboxdisplayedina
window.TheBoldItalicfontindicatesthelename.
(Inscriptionexample)
Select[LicenseActivation]>[First-trial.lic]
(Meaning)
Thisindicatesthatontheconrmationscreenfortherst-timeuse,onthelicenseactivation
screen,click[LicenseActivation],thenselectthe[First-Trial.lic]le.
lDrawing Conventions
Somedrawingsmaybepartiallyemphasized,simplied,oromitted,fortheconvenienceof
description.
Somescreenimagesdepictedintheuser’smanualmayhavedierentdisplaypositionsor
charactertypes(e.g.,theupper/lowercase).Alsonotethatsomeoftheimagescontainedinthis
user’smanualaredisplayexamples.

5
TI 11B06H01-01EN
For GC8000 Gas Chromatograph AI Maintenance Support Installation and License Guide
Mar. 17, 2023-00
n CopyrightandTrademarkNotices
lAll Rights Reserved
Thecopyrightsoftheprogramsandon-linemanualarereserved.
Theon-linemanualisprotectedbythePDFsecurityfrommodication,however,itcanbeoutput
viaaprinter.Printingouttheon-linemanualisonlyallowedforthepurposeofusingtheproduct.
Whenusingtheprintedinformationoftheon-linemanual,checkiftheversionisthemostrecent
one.
Nopartoftheon-linemanualmaybetransferred,sold,distributed(includingdeliveryviaa
commercialPCnetworkorthelike),orregisteredorrecordedonvideotapes.
lTrademark Acknowledgments
• MicrosoftandWindowsareeitherregisteredtrademarksortrademarksofMicrosoft
CorporationintheUnitedStatesand/orothercountries.
• Adobe,AdobeAcrobat,andAcrobatReaderaretrademarksorregisteredtrademarksof
AdobeSystemsIncorporatedintheUnitedStatesand/orothercountries.
• Allothercompanyandproductnamesmentionedinthisuser’smanualaretrademarksor
registeredtrademarksoftheirrespectivecompanies.
• WedonotuseTMor®marktoindicatethosetrademarksorregisteredtrademarksinthis
user’smanual.

6
TI 11B06H01-01EN
For GC8000 Gas Chromatograph AI Maintenance Support Installation and License Guide
Mar. 17, 2023-00
1. How to install
1.1 Data required for AI analysis
Toinstallthesoftware,astablechromatogramofGCmeasurementresultsandatleastoneday
ofdataontheanalysisresultsarerequired.
Formoreaccuratesettings,werecommendatleast1monthofdata.Congurationdatacanbe
addedorchangedlater.
SEE
ALSO
RefertoGasChromatographAIMaintenanceSupportHandbook(TI11B06H01-02EN)formoreinformation.
1.2 Download software/initial free trial license
DownloadtheGCAIsoftwarefromtheYOKOGAWAwebsite.ConsultaYOKOGAWAsales
representativeinchargefortheURL.
Afterregisteringasamember,youcandownloaditfromtheproductpage.
CAUTION
• Aseparatelicensepurchaseisrequiredforcontinueduseaftertheendofthefreeperiod.
• IfyoucannotdownloaddirectlytoPCAS,pleaseusetheappropriaterecordingmediato
migratetoaPCwithPCASinstalled.
1.3 Performing an Installation
Downloadthesoftwarefromourdedicatedwebsite.Afterextractingtheziple(*zip)ofthe
software,executetheinstaller(GCAI_Install.exe)onthePConwhichPCASisinstalled.
Followtheinstructionsonthestartupscreentoproceedwiththeinstallation.
• Agreetothelicenseagreementandproceedtothenextstep.
• ThesoftwareisinstalledbydefaultinC:\GCAI.Tochangetheinstallationlocation,click
“Change”andspecifythefolder.
Afterinstallationiscomplete,astoragefolderwhereashortcuticonisstoredwillappearonthe
desktop.
Figure 1 GCAI shortcut icon
TIP OnceGCAIisinstalled,itwillbestoredinastartupprogram.GCAIwillautomaticallystartwhen
thePCstartsup.

7
TI 11B06H01-01EN
For GC8000 Gas Chromatograph AI Maintenance Support Installation and License Guide
Mar. 17, 2023-00
2. Initial Installation
SomeapplicationsmaynotbecompatiblewiththissoftwareduetoGCmeasurement
specications.Pleaseusetheinitialfreetriallicenseinaccordancewiththismanual,and
purchaseapaidlicenseafterconrmingcompatibility.
Inaddition,asthenumberofmonitoredunitsincreases,dependingontheprocessingpowerof
thePCsused,thesystemmayinrarecasesexceedthespecications.Pleasecheckbackduring
thefreetriallicenseperiod.
CAUTION
• Thefreetriallicenseisintendedtoensurecompatibilitywiththeapplication.Wealsodonot
extendthefreeperiodforanyreason.
• Thesoftwarenotiesanomaliesonlyifathresholdexceedanceeventoccursduringthe
period.
2.1 Activation of initial free trial license
Theinitialfreetriallicenseleis“First-trial.lic”,beingenclosedwiththedownloadedcompressed
le.
Thefollowingscreenfortherst-timeusewillbedisplayedwhenthesystemisstartedfortherst
time.
Select“LicenseActivation”andactivatethefreetriallicenseaccordingtotheinstructionsonthe
screen.
Figure 2 Initialstartupscreen
lHow to Start Free trial licensing
OntheInitialstartupscreen,select[LicenseActivation]>[First-trial.lic]>clickcheck-boxtostart
fortrialID-Tag>[ActivatetheselectedID]
CAUTION
IfPCASfolder’ssetupisnotC:\PCAS,selectonthestartupfortherst-timeuse[License
Activation]or[createrequestle]andfollowthepop-upinstructionstospecifyaPCASfolderand
changethesetting.

8
TI 11B06H01-01EN
For GC8000 Gas Chromatograph AI Maintenance Support Installation and License Guide
Mar. 17, 2023-00
SEE
ALSO
Fordetailsonhowtoauthorizelicensesotherthanthefreetriallicense,seeChapter1ofIM11B06H01-01EN.
Tocontinuetousethesoftwareafterthefreetriallicenseexpires,youwillneedtoapplyforapaid
license,forwhichyouwillneedtocreatearequestle.
SEE
ALSO
See“3.3 PurchaseFlowofPaidLicenses”fordetailsonhowtocreatearequestle.
2.2 Conrmationofoperation
2.2.1 Conrmationofhealthdiagnosisoperation
Afterthelicenseactivation,itautomaticallyloadsthechromatographdatafromapreviousday
storedinPCASandstartstraining.Thetrainingmaytakeseveralminutes.
WhenthechromatogramoftherstGCanalysisresultisoutputafterthetrainingiscompleted,
theclassicationstarts.Followtheprocedurebelowtocheckifthetraininghasbeencompleted
successfullyandtheclassicationhasbeenmade.
Click“AllGCs”tabonthemainscreentoseethelistofGCstowhichlicenseshavebeen
assigned.
IftheiconprecedingtheIDnumberisanyoneofthefollowings:
training/classicationhasbeensuccessfullyperformedindefaultmode.
TheiconoftheIDnumberindicates astatusofUntraineduntiltheinitialclassicationhas
beenmade.
Figure 3 All GCs tab

9
TI 11B06H01-01EN
For GC8000 Gas Chromatograph AI Maintenance Support Installation and License Guide
Mar. 17, 2023-00
When a training failed
IDnumberindicates :astatusofUntrained,evenafteranewchromatogramisgenerated,
checkthefollowingitems.
1. Thehealthdiagnosesettingscreenisnotautomaticallyupdated.Youneedtodisplaythe
Listtabagainafterdisplayinganotherscreen.Thedisplayisupdated.
2. HealthdiagnoseisperformedbyusingPCASchromatogram.Therefore,ifthePCAS
foldernameisdierentfromtheoneinthesetting,thechromatogramisnotread.
CheckthePCASfoldernameasfollows.
Select[Conguration]>[SetPCASFolder]onthemenubar.
CheckthefoldernametherematchesthedirectoryoftheactualPCASfolderthatGCAI
refersto.
3. CheckifthefolderwiththecorrespondingIDinthePCASfoldercontainschromatograms
forthepreviousday’sdate.Ifthechromatogramofthepreviousdayisnotstored,the
trainingcannotbeperformedinthedefaultmode.
4. Aftertrainingiscompleted, remainsdisplayeduntilarstresultcomesout.
CheckwhetheranewchromatogramisgeneratedinthefolderofthecorrespondingIDin
PCASfolder.
5. Iftherearenoproblemsaftercheckingtheaboveitemsandtheautomaticallyregion
settingisconducted,checktheeachregionofthechromatogramleshowsstablevalue.
Ifthevalueisnotstable,itispossiblethatthedefaultregionisnotsetproperly.From
eachID’sheathdiagnosesettingsscreen,gototheregionsettingscreenandcheckto
seeifthereisatleastoneregionthatisenabled.
Ifallaredisabled,youneedtosetuparegionandstarttraining.
6. Otherwise,ifyoudonotidentifythesourceoftheproblem,reviewthelogsinthe
dashboard,andsystemmessagesaswell.
SEE
ALSO
Formoreinformationontheregionsetting,refertothefollowingdocuments:
• “GasChromatographAIMaintenanceSupportHandbook”(TI11B06H01-02EN)Chapter2
• “GasChromatographAIMaintenanceSupportOperatingManual”(IM11B06H01-01EN)Chapter3
TIP Ifyoudonotusethehealthdiagnosefunction,disableallregions.
TIP <HowtogeneratePre-anomalyorAnomalyarticially(reference)>
BygeneratinghistoricalchromatogramsinPCASchromatogramfolderfor
(Installation folder) \ PCAS \ data \ (applicable ID) \ (applicable sensor) \ FullCro \ (any date)
youcansimulatethesituationwhereachromatogramisgenerated.Tochecktheobjective,copythedata
withdierencesorpreviouschromatogramsofdatesfarenoughfromthetrainingdates(=theprevious
day)againstthetrainingdatainthedefaultmode,andduplicatetheminthesamefolder.
DuplicatingchromatogramshavenoeectonPCAS

10
TI 11B06H01-01EN
For GC8000 Gas Chromatograph AI Maintenance Support Installation and License Guide
Mar. 17, 2023-00
2.2.2 Operationcheckoftheresolutionpredictionfunction
Uponactivationofthelicense,GCAIautomaticallyreadsthedataoftheanalysisresultsstoredin
PCASandstartstraining.Trainingmaytakeseveralminutes.
1. Aftertrainingiscompleted,select[AllGCs]>[ID-Tag]>[ResolutionPred.]onthemain
screen.
Ifthe“DayBefore”and“30daysPrediction”aredisplayed,theoperationisnormal.
Ifyoudonotwanttousetheresolutionpredictionfunction,changethesettingfrom
“Monitoring”to“Disabled”.
Thethresholdissettothedefaultvalue(1.5);refertoChapter3ofTI11B06H01-02ENto
changetotheappropriatethreshold.
2. Iftheyarenotdisplayed,checkthefollowingitems.
a. Numbersinthescreenarenotautomaticallyupdated.Afterdisplayinganotherscreen,
displaythelisttabagain.Thedisplayisupdated.Conrmagainthattheresolution-predictive
functionisworkingproperly.
b. Select[Conguration]>[SetPCASFolder]onthemenubar.Makesurethatitmatches
theactualPCASfolderdirectory.
c. Checkwhethertheanalysisresultle(*.anx)isstoredinthefolderofthecorresponding
IDofthePCASfolder.Theanalysisresultisshownasbelowaccordingtothe
correspondingle.
“DayBefore” Displayedwhenlesfromthepreviousdayarestored
“30daysPrediction” Displayedwhenleshavebeenstoredfor14
consecutivedaysuptothepreviousday
Figure 4 Resolutionpredictionscreen

11
TI 11B06H01-01EN
For GC8000 Gas Chromatograph AI Maintenance Support Installation and License Guide
Mar. 17, 2023-00
2.3 Checkitemsbeforemovingtothepaid
version
PurchasingapaidlicenseallowsyoutouseGCAIevenafterthefreeperiodends.Priorto
purchasing,makesurethatallofthefollowingitemsapply:
□ ConfirmationofnormaloperationofthemonitoredGCasdescribedin“2.2 Confirmationof
operation”.
□ Datahasbeenstoredformorethanaday.
□ Thedesiredmonitoringregionsetupoftheanomalydetectionfunctionissettoenable.
□ Theresultsoftheclassicationofthedesiredmonitoringmeasuringstream/detectorofthe
anomalydetectionfunctioncanbeconrmed.
□ Thepredictionresultsoftheresolutionpredictionfunctioncanbechecked.
□ Theerrorlog“Thenumberofoperationrequestsexceedstheprocessinglimit”thatoccurs
duetoinsucientPCspecicationsisnotgenerated.(Thismayoccurduetotheincreased
loadatthetimeoftraining,butyoucanusethesoftwareifnoerrorlogsaregenerated
duringnormaloperation.)Or,thesoftwarestopsorothermalfunctionsoccur.
SEE
ALSO
If“Thenumberofoperationrequestsexceedstheprocessinglimit”errorlogoccurs,seethechapter4ofIM
11B06H01-01EN
Whentheaboveconditionsaremet,youmaycontinuetousetheserviceafterthefreeperiodby
purchasingapaidlicense.Tracklong-termtrendstohelpmaintainyourGC.
Ifyoucannotconrmtheaboveitemsandyourfreeperiodhasexpired,pleasetryduringthepaid
period.

12
TI 11B06H01-01EN
For GC8000 Gas Chromatograph AI Maintenance Support Installation and License Guide
Mar. 17, 2023-00
3. Paid License
Thissectiondescribestheusageconsiderations,types,andpurchaseowforpaidlicenses.
3.1 Paid License Usage Notes
● AlicenseforthissoftwarecanbepurchasedforeachmonitoredGC.
● ThisserviceisnotavailableforGCswithanIDdierentfromthatofthecontractatthetime
ofpurchase.
● Renewthepaidlicensebeforethelicenseexpirationdatetocontinuetouse.
● Thelicenseperiodiscontinuousfromthestartofusetotheendofuse,andcannotbe
stoppedorcanceledinthemiddleofthelicenseperiodduetoanyreasonsuchasstoppage
ofGCmainunit,PCAS,GCAI.
● Onthequotation,licensesformorethan12monthsaresplitinto12months(suxcodefor
12months)+1to11months(suxcodefor1to11months),butaredeliveredasasingle
license.
(Example)
Ifyouwanttopurchasealicensefor22months:
22months(singlelicense)=12months+10months(twosuxcodesonthequotation)
Nocongurationchangesarerequiredtousethelicense.
3.2 TypesofPaidLicenses
Thepaidlicensesinclude“NewLicense”,“Additionallicense”,whichaddGCtobemonitored,
and“RenewalLicense”,whichextendsanexistinglicenseperiod.
Theexpirationdateofapaidlicensedependsonthetypeoflicense/situationasfollows;
Table 1 Typesofpaidlicenses
Type*1 Beforepurchasethelicense Eective
Newlicence(Additional)*2 Nopaidlicenseisapplied Fromthestartmonthtotheendof
thepurchasedlicenseperiod
Renewallicense Whenoneormorepaidlicenses
areappliedandtheirusageperiod
needstobeextended
Validuntiltheextensionperiod
expireswhichwasaddedtoan
existinglicenseperiod.
Additionallicense Whenoneormorepaidlicenses
arealreadyappliedandwhenyou
needtoaddanewlicense
Validuntilthesamemonthasthe
endoftheexistinglicense
*1: AdditionallicenseandRenewallicensescanbepurchasedasasinglelicenseinasingleorder/single
requestle.
*2: Thesystemautomaticallyselects“New”or“Add”whenyoupurchaseanewlicenseoranadditionallicense
accordingtoyourorder.

13
TI 11B06H01-01EN
For GC8000 Gas Chromatograph AI Maintenance Support Installation and License Guide
Mar. 17, 2023-00
3.3 Purchase Flow of Paid Licenses
Theowchartforpurchasingapaidlicenseisshownbelow.
Renewal license
2. Send Request File
1. Request File Output
3. Receipt of the quotation
6. License activation
5. Receipt of the license File
4. Ordering/Send application form
0. Check operation with Trial-License
Figure 5 Flowfromlicensepurchasetoactivation
Theprocedureforpurchasingapaidlicense
1: RequestFileOutput
For new/additional licenses only:
Select[License]>[PrepareforAddingaLicense]onthemenubarofthemainscreen.
Followtheon-screeninstructionstoselectGCyouwanttomonitor,andthencreatea
licensele.
Forrenewallicensesonly,andforrenewalandadditionallicenses:
Select[License]>[PrepareforRenewingaLicense/Licenseinfo...]onthemenubarofthe
mainscreen.
Followtheon-screeninstructionstoselectGCyouwanttomonitor,andthencreatea
licensele.
Figure 6 [License] on the menu bar is selected
RequestFileOutputConditions
Table 2 LicenseTypesandRequestFileOutputConditions
Type RequestleOutputconditions Requestleoutputavailableperiod
New(Add*) Anewpaidlicenseispurchasedorthelatest
paidlicensehasexpiredformorethan1
month.
Norestrictions.
Renewal Avalidpaidlicenseexists,orthemostrecent
paidlicenseexpiredwithinonemonth
Norestrictions.
Additional Ifavalidpaidlicenseexistsandthepaid
licenseforthesubjectIDhasexpiredformore
thanonemonth.
Availableupto2monthspriortotheexpiration
ofexistingpaidlicensesIfwithinthefreetrial
licenseexpirationdate,thefreetriallicense
expirationdatemustbeatleasttwomonths
shorterthanthepaidlicenseexpirationdate.
Renewal+
Additional
IDmustmeettheaboverenewalandaddition
requirements.
Untiltheexpirationdateoftheexistinglicense
*: Thesystemautomaticallyselects“New”or“Add”whenyoupurchaseanewlicenseoranadditionallicense
accordingtoyourorder.
2: Sendrequestle
Sendtherequestlecreatedin1toalocalYOKOGAWAsalesrepresentative.

14
TI 11B06H01-01EN
For GC8000 Gas Chromatograph AI Maintenance Support Installation and License Guide
Mar. 17, 2023-00
3: Receiptofthequotation
Dependingonthecontentsofyourrequestle,theYOKOGAWAsalesrepresentativein
chargewillsendyouaquotation.
4: Ordering
Checkthecontentsbeforeplacingyourorderwiththeapplicationformatthattime.
5: Receiptofthelicensele
Aftertheorderisconrmed,alicenselewhichismadeaccordingtothequotationissent
fromtheYOKOGAWArepresentative.
6: Licenseactivation
Select[License]>[LicenseActivation]onthemenubar,selectthelicenselethatwas
receivedin5.Verifythatthelicensehasthecorrectlicensedetails.Aftertheverication,
followtheon-screeninstructionstoactivatethelicense.
SEE
ALSO
Fordetailsonhowtoauthorizelicenses,refertoChapter1ofthe“GasChromatographAIMaintenance
SupportOperationManual”(IM11B06H01-01EN)
License activation validity
Table 3 LicenseTypesandRequestFileOutputConditions
Type Activation validity Licensevalidityperiod
New(Add) Untilthemonthinwhichthepurchasedlicense
expires.
Fromthestartmonthtotheendofthe
purchasedlicenseperiod
Renewal Untilthemonthinwhichthepurchasedlicense
expires.
Validuntiltheextensionperiodexpires
whichwasaddedtoanexistinglicense
period.
Add Activatebeforetheexpirationdateoftheexisting
license.Activationisnotpossibleafterthe
expirationdate.
Validuntilthesamemonthastheend
oftheexistinglicense
Renewal+
Additional
Activateduringthemonthpriortotheexpiration
dateoftheexistinglicenseuntiltheexpiration
date.
Activationisnotpossibleaftertheexpirationdate.
Validuntiltheextensionperiodexpires
whichwasaddedtoanexistinglicense
period.

15
TI 11B06H01-01EN
For GC8000 Gas Chromatograph AI Maintenance Support Installation and License Guide
Mar. 17, 2023-00
n Validityoffreetriallicenseswhenpurchasingpaidlicenses
Youcanrenewafreetriallicensetoapaidlicensefollowingafreetriallicenseperiod.(Case.1)
Inthiscase,thelicensefeewillbechargedafterthefreetriallicenseexpires.
Case1 WhenpurchasinganadditionallicenseforafreetriallicenseID6.
Year 20XX/ 20XY/
Month 456789• • • 12 1 2
ID:1
ID:3
ID:6
Beawareofthefollowingcase.(Case2)
IftheexpirationdateoftheexistinglicenseIDisearlierthantheexpirationdateofthefreetrial
license,thefreetriallicensewillexpireatthesametimeastheexistinglicenseexpirationdate.In
thesystem,thefreetriallicenseexpirationdatewillbeoverwritten.
Topreventthis,besuretorenewonlytheexistinglicenseIDrst.ThenorderthelicenseIDyou
areassignedforthefreetriallicense.Thatway,thefreeperiodwillnotbeoverwrittenandthe
maximumperiodwillbevalid.
Case2 WhenafreetriallicenseID:6isexpectedtoexpirebeyondtheexistingpaidlicenseexpirationdate.
Year 20XX/ 20XY/
Month 456789• • • 12 1 2
ID:1
ID:3
ID:6
Overwrite
Legend:
Existnglicense
Newlypurchasinglicense
Freetriallicense
Ifmultiplefreetriallicenseswithdierentexpirationdatesareconvertedtopaidlicensesat
once,theexpirationdateofalllicenseswillbe12monthsfromtheexpirationdateoftheearliest
expiringfreetriallicense.Therefore,tomaximizeyourfreetrialperiod,pleasepurchaseoneID
rstandthenorderadditionalremainingIDs.

16
TI 11B06H01-01EN Mar. 17, 2023-00
Revision Information
Title : ForGC8000GasChromatographAIMaintenanceSupportInstallationand
LicenseGuide
Manualnumber: TI11B06H01-01EN
Mar.2023/1stEdition
Newlyreleased
YokogawaElectricCorporation
2-9-32Nakacho,Musashino-shi,Tokyo180-8750,JAPAN
http://www.yokogawa.com/
Other manuals for GC8000
1
Table of contents
Other YOKOGAWA Laboratory Equipment manuals
Popular Laboratory Equipment manuals by other brands

SPS
SPS Polos SPIN150i infinite manual

Four E's Scientific
Four E's Scientific MI0102008 operating instructions

Sakura
Sakura Tissue-Tek Film Quick reference guide

Thermo Scientific
Thermo Scientific BF51842C Installation and operation manual

RS PRO
RS PRO RS-983 instruction manual

Bullard
Bullard T4X user manual