6.0 Checking Ifthereisaninstalledoptional6.5”DisplayandKeypad–ObservetheMainMenumessagesandstatus
information.
6.1 IfthereisnoinstalledUserInterface,thenconnectalaptopPCviaEthernettotheSBCmountedonthe
backplane. Initiate the supplied UltraVNC software from the laptop to initiate a VNC session with the
‘blind’ analyzer and observe the analyzer Main Menu via the laptop.
6.2 AT this time there may be one or more alarm message due to low transmission, out of range parameters or other –
nal system conguration is still required!
6.3 IftheanalyzerdisplaysaWarning“ValidationRequired”thenthisindicatesthatthereisnotargetgas
absorption peak found at start-up. Either, shut down the analyzer, introduce some measured gas into the
owcellandre-startorperformavalidation.Thiswillensurethattheanalyzeriscorrectlytunedtothe
measurement gas absorption peak. If this Warning cannot be cleared by either method, please contact
YokogawaLaserAnalysisDivisionoryourlocalagentforfurtherassistance.
7.0 Congure
BASIC
By way of the appropriate user interface, the correct process parameters and other parameters can now
be entered.
7.1 EntertheBasicMenuandgotoCongure.
7.2 Gas Pressure Enterinthecorrectprocessgaspressure(ifActive,seeAdvancedCongure).
7.3 Gas Temperature Enterinthecorrectprocessgastemperature(ifActive,seeAdvancedCongure).
7.4 Ifanyotherparametersarerequiredtobeset(suchasanalogI/Oranges,alarmslevels,AutoValidation
sequences) then the Advanced Menu needs to be accessed.
Advanced Menu access is Password protected (default 1234, can be changed by user if necessary) and should only be
used by skilled and trained persons - Contact Yokogawa Laser Analysis Division or Local Agent if any doubts!
GototheDatasectionunderBasicandconguretheappropriate‘RecordResultData’settings.Thiswill
ensuretheanalyzerstoresimportantinformationduringoperationthatmaybeusedtoverifyoperation/
status/diagnosticsandothertroubleshooting.
8.0 Congure
ADVANCED
Usingthecorrectpassword(Default1234),enterintotheAdvancedMenu,thentheCongure.
8.1 Select the desired measurement units for path length, pressure and temperature.
8.2 Select Fixed or Active.
If Fixed, enter in the correct process gas pressure.
If Active, enter in the 4-20mA input signal range proportional to the pressure transmitter output range.
8.3 Select Fixed or Active.
If Fixed, enter in the correct process gas temperature.
If Active, enter in the 4-20mA input signal range proportional to the temperature range.
8.4 CongurethesystemI/ObyenteringintotheSystemI/OsubmenuinCongure.
8.5 IftheoptionalAnalogI/Oboardisinstalled,thenselectAnalogOutputandsettheappropriate4mAand
20mA values for Ch 1 Concentration and Ch 2 Transmission.
8.6 Selectwhatmode(Block,TrackorHold)the4-20mAoutputsaretobewhentheanalyzerisinWarning,
Fault, Export Data and Calibration Modes.
8.7 CongureDigitalI/O–WarningsandFaults.Manyofthesewillbefactorypre-setsoifunsureaboutany
settings then leave as Factory Default.
SelectandsetlevelforAlarmLimittoeithertheMeasuredGasorTransmission.
8.8 Go to the Data screen and set the appropriate parameters for ‘Record Result Data’ and ‘Spectrum
Capture’. These will ensure the analyzer stores important information during operation that may be used
toverifyoperation/status/diagnosticsandothertroubleshooting.