manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. YOKOGAWA
  6. •
  7. Measuring Instrument
  8. •
  9. YOKOGAWA EXA100 Series Installation guide

YOKOGAWA EXA100 Series Installation guide

[䉭] key
[䉮] key
This Quick Manual briefly explains the operation,
calibration and routine maintenance of the EXA Series
SC100 Converter and the dedicated conductivity sensor,
SC10XB. Please read the appropriate user's manuals
thoroughly before using this system.
References: User's Manuals
SC100 Panel Mount Conductivity Converter:
IM 12D11A01-01E
SC10XB Conductivity Sensor:
IM 12D11C01-01E
1. Key Operations
• To cancel the operation: Simultaneously press [䉭] and
[䉮] keys.
• To move from measurement display to setup display :
Hold down [ENT] key for at least 3 sec.
• To move from measurement display to calibration display :
Hold down [䉮] key for at least 3 sec.
• To move from setup/calibration display to measurement
display :
Hold down [䉭] key for at least 3 sec.
Main functions of keys
[ENT] key : Confirms the status or setting.
[䉭] key: Moves up a menu item or increases the setting.
[䉮] key: Moves down a menu item or decreases the setting.
2. Start-up
For start-up instructions, refer to the SC100 Panel Mount
Conductivity Converter's User's Manual, IM 12D11A01-01E.
3. Calibration
(Using Standard Calibration Solution)
Refer to Chapter 6 of the SC100 Panel Mount Converter's
User's Manual, IM 12D11A01-01E.
SC100 Panel Mount Conductivity Measurement System
[Quick Manual]
Status Display
• Status of
contact output
(specified by
suffix code)
[ENT] key
Main Display
• Conductivity reading
during measurement
• Setting and data during
calibration/setup
Sub-Display
• Temperature during
measurement
• Parameter during setup
LEDLED NumberAlphabetAlphabet
ZM
YL
XK
9WJ
8VI
7UH
6TG
5SF
4RE
3QD
2PC
1OB
0NA
Display and Operation Keys
Alphanumeric Display Table
IM 12D11A01-11E 1st Edition: Mar. 2003
EXA100 Series
* Conductivity meter does not require calibration using standard solution as routine maintenance if the
correct cell constant has been defined upon start-up.
* For details on the calibration using standard solution, refer to Section 6.5 of the SC100 Panel Mount
Converter's User's Manual, IM 12D11A01-01E.
LED
µS/cm
mS/cm
CANCEL
MEASCALSET
S4
S3
S2
S1
ENT
SC100
EXA
* Calibration should be performed if the cell constant has changed due to prolonged operation.
Press [ENT]
key once.
Water
Calibration
Display Top
Replace the sensor in the process.
Calibration result
is reflected to the
measured value.
Main display stops
flashing and sub-display
is flashing.
Set the conductivity
value of the standard
solution using [䉭] and
[䉮] keys.
Press [䉭] key
for 3 sec.
Measurement
Display
1st digit in Main
display is flashing.
Wait until reading
has stabilized.
Press [䉮] key
for 3 sec.
Measurement
Display
Press [ENT] key twice.
Main and sub displays
flash upon each press.
Rinse the conductivity
sensor well with water
or appropriate solution.
Standard solution
Press [ENT]
key once.
µS/cm
mS/cmmS/cm
EXA SC100
S1
S2
S3
S4
mS/cm
µS/cm
EXA SC100
ENT
S1
S2
S3
S4
SET CAL MEAS
CANCEL
µS/cm
mS/cm
CANCEL
MEASCALSET
S4
S3
S2
S1
ENT
SC100EXA
µS/cm
mS/cm
S4
S3
S2
S1
SC100EXA
Wipe thoroughly after rinse,
and then soak the sensor in
the standard solution.
µS/cm
mS/cmmS/cm
EXA SC100
S1
S2
S3
S4
Decimal point
is flashing.
Move the decimal point
using [䉭] and [䉮] keys.
Press [ENT] key to confirm.
µS/cm
mS/cm
EXA SC100
S1
S2
S3
S4
µS/cm
mS/cm
EXA SC100
S1
S2
S3
S4
µS/cm
mS/cm
EXA SC100
S1
S2
S3
S4
Data updated
µS/cm
mS/cm
S4
S3
S2
S1
SC100EXA
Press [ENT]
key once.
Press [ENT]
key once.
IM 12D11A01-11E 1st Edition: Apr. 2003
4. Measurement
Sample Temperature Conditions
SC10XB Conductivity Sensor : 70 8C or lower
Note: Specification requirements vary depending on sample composition.
• The SC10XB can be used in conjunction with a dedicated adapter.
5. Maintenance
• Cleaning the Electrode <Periodical cleaning, sample dependent>
• Cleaning the Electrode
6. Error Messages, Error checking flowchart
• An error message during calibration
For details on error messages, refer to Chapter 8 of the SC100 Panel Mount Converter's User's Manual, IM 12D11A01-01E
Clean the electrode.
Check connections.
NO
NO
NO
NO
NO
The electrode is wet.
The electrode is dirty.
YES
YES
YES
YES
YES
The electrode has
deteriorated. Replace
the sensor with a new
one.
Water
Electrode
Cotton wab or Soft Material
Check connections.
Clean the electrode.
NO
NO
NO
The temperature
sensor is disconnected
or short-circuited.
The electrode is dirty.
Cable connections
are not appropriate.
YES
YES
YES
The electrode has
deteriorated.
Replace the sensor
with a new one.
Zero Calibration (Air Calibration: Eliminating an error resulting from wiring capacity)
Refer to Section 6.4 of the SC100 Panel Mount Converter's User's Manual, IM 12D11A01-01E.
Replace the
sensor.
• An error message during measurement
NO
YES
Dry it thoroughly.
Water
Calibration
Display Top
Replace the sensor
in the process.
Calibration result
is reflected to the
measured value.
Main display stops
flashing and
sub-display is flashing.
Press [ENT] key once.
Press [n] key
for 3 sec.
Measurement
Display
Wait until air
calibration has
been complete
(60 sec).
"Air" in sub-display
is flashing.
Press [,] key
for 3 sec.
Measurement
Display
Press [ENT]
key once.
Rinse the conductivity
sensor well with water
or appropriate solution.
Press [n] key.
µS/cm
mS/cm
EXA SC100
S1
S2 S3
S4
mS/cm
µS/cm
EXA SC100
ENT
S1
S2 S3
S4
SET CAL MEAS
CANCEL
µS/cm
mS/cm
CANCEL
MEASCALSET
S4
S3
S2
S1
ENT
SC100EXA
µS/cm
mS/cm
S4
S3
S2
S1
SC100EXA
Press [ENT]
key once.
In the air
µS/cm
mS/cm
EXA SC100
S1
S2 S3
S4
µS/cm
mS/cm
S4
S3
S2
S1
SC100EXA
µS/cm
mS/cm
EXA SC100
S1
S2 S3
S4
Data updated
µS/cm
mS/cm
EXA SC100
S1
S2 S3
S4
Main display is
flashing and "WAI" in
sub-display is flashing.
µS/cm
mS/cm
EXA SC100
S1
S2 S3
S4
"Str" in sub-display
is flashing.
"CAL" in sub-display
is flashing.
Press [ENT]
key once.
Error message C-2, C-3, or C-4 appears
during calibration.
Error message E-1, E-2, E-3, or
E-5 appears during measurement.
Use standard solution with
correct concentration.
Cable connections
are not appropriate.
Dry the electrode in zero (air)
calibration.
When error message
C-1 appears, bring the
standard solution within
the operating temperature.
* Correct temperature
coefficient should be set when error message C-4 appears.
If an E-4 error message
appears, the converter has malfunctioned.
Contact your sales representative or the nearest service office.
Converter failure has
been detected.
Replace the
converter.
Converter failure
has been detected.
Replace the
converter.
The concentration
of standard solution is
incorrect.

This manual suits for next models

1

Other YOKOGAWA Measuring Instrument manuals

YOKOGAWA PH71 User manual

YOKOGAWA

YOKOGAWA PH71 User manual

YOKOGAWA 735201 User manual

YOKOGAWA

YOKOGAWA 735201 User manual

YOKOGAWA 702907 User manual

YOKOGAWA

YOKOGAWA 702907 User manual

YOKOGAWA EXA TB TB400G User manual

YOKOGAWA

YOKOGAWA EXA TB TB400G User manual

YOKOGAWA ZR802S User manual

YOKOGAWA

YOKOGAWA ZR802S User manual

YOKOGAWA 701957 User manual

YOKOGAWA

YOKOGAWA 701957 User manual

YOKOGAWA CL155 User manual

YOKOGAWA

YOKOGAWA CL155 User manual

YOKOGAWA ZR802G Installation manual

YOKOGAWA

YOKOGAWA ZR802G Installation manual

YOKOGAWA 701944 User manual

YOKOGAWA

YOKOGAWA 701944 User manual

YOKOGAWA EXAxt ZR22G Manual

YOKOGAWA

YOKOGAWA EXAxt ZR22G Manual

YOKOGAWA RAGN Manual

YOKOGAWA

YOKOGAWA RAGN Manual

YOKOGAWA WE440 User manual

YOKOGAWA

YOKOGAWA WE440 User manual

YOKOGAWA 2038 User manual

YOKOGAWA

YOKOGAWA 2038 User manual

YOKOGAWA Total Insight series User manual

YOKOGAWA

YOKOGAWA Total Insight series User manual

YOKOGAWA AQ7277B User manual

YOKOGAWA

YOKOGAWA AQ7277B User manual

YOKOGAWA ZX8D User manual

YOKOGAWA

YOKOGAWA ZX8D User manual

YOKOGAWA AQ2200 Series User manual

YOKOGAWA

YOKOGAWA AQ2200 Series User manual

YOKOGAWA ADMAG TI Series User manual

YOKOGAWA

YOKOGAWA ADMAG TI Series User manual

YOKOGAWA AQ6373B User manual

YOKOGAWA

YOKOGAWA AQ6373B User manual

YOKOGAWA WT500 User manual

YOKOGAWA

YOKOGAWA WT500 User manual

YOKOGAWA ST401G User manual

YOKOGAWA

YOKOGAWA ST401G User manual

YOKOGAWA AQ6370D User manual

YOKOGAWA

YOKOGAWA AQ6370D User manual

YOKOGAWA 99025 User manual

YOKOGAWA

YOKOGAWA 99025 User manual

YOKOGAWA WT200 User manual

YOKOGAWA

YOKOGAWA WT200 User manual

Popular Measuring Instrument manuals by other brands

CLA-VAL e-FlowMeter user manual

CLA-VAL

CLA-VAL e-FlowMeter user manual

Raycus RFL-C Series user guide

Raycus

Raycus RFL-C Series user guide

Teledyne Lecroy PP011-1 Operator's manual

Teledyne Lecroy

Teledyne Lecroy PP011-1 Operator's manual

Good Will Instrument GSP-827 Operation manual

Good Will Instrument

Good Will Instrument GSP-827 Operation manual

General DSM402SD user manual

General

General DSM402SD user manual

Hunter HC manual

Hunter

Hunter HC manual

Intec I-Starter 2.8 user manual

Intec

Intec I-Starter 2.8 user manual

Siemens SITRANS LU AO quick start guide

Siemens

Siemens SITRANS LU AO quick start guide

Campbell CS616 Features

Campbell

Campbell CS616 Features

Orflo Moxi Flow user guide

Orflo

Orflo Moxi Flow user guide

MMF VC120 instruction manual

MMF

MMF VC120 instruction manual

SK telecom Air Cube SN-100T manual

SK telecom

SK telecom Air Cube SN-100T manual

Fluke RaySafe 452 user manual

Fluke

Fluke RaySafe 452 user manual

Oval EX DELTA II instruction manual

Oval

Oval EX DELTA II instruction manual

ASM NMG2 Installation and operation manual

ASM

ASM NMG2 Installation and operation manual

Endress+Hauser Proline Promass 40E technical information

Endress+Hauser

Endress+Hauser Proline Promass 40E technical information

Simplay Labs SL-863 user guide

Simplay Labs

Simplay Labs SL-863 user guide

Honeywell H Series installation instructions

Honeywell

Honeywell H Series installation instructions

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.