manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. YOKOGAWA
  6. •
  7. Measuring Instrument
  8. •
  9. YOKOGAWA EXA100 Series Installation guide

YOKOGAWA EXA100 Series Installation guide

[䉭] key
[䉮] key
This Quick Manual briefly explains the operation,
calibration and routine maintenance of the EXA Series
SC100 Converter and the dedicated conductivity sensor,
SC10XB. Please read the appropriate user's manuals
thoroughly before using this system.
References: User's Manuals
SC100 Panel Mount Conductivity Converter:
IM 12D11A01-01E
SC10XB Conductivity Sensor:
IM 12D11C01-01E
1. Key Operations
• To cancel the operation: Simultaneously press [䉭] and
[䉮] keys.
• To move from measurement display to setup display :
Hold down [ENT] key for at least 3 sec.
• To move from measurement display to calibration display :
Hold down [䉮] key for at least 3 sec.
• To move from setup/calibration display to measurement
display :
Hold down [䉭] key for at least 3 sec.
Main functions of keys
[ENT] key : Confirms the status or setting.
[䉭] key: Moves up a menu item or increases the setting.
[䉮] key: Moves down a menu item or decreases the setting.
2. Start-up
For start-up instructions, refer to the SC100 Panel Mount
Conductivity Converter's User's Manual, IM 12D11A01-01E.
3. Calibration
(Using Standard Calibration Solution)
Refer to Chapter 6 of the SC100 Panel Mount Converter's
User's Manual, IM 12D11A01-01E.
SC100 Panel Mount Conductivity Measurement System
[Quick Manual]
Status Display
• Status of
contact output
(specified by
suffix code)
[ENT] key
Main Display
• Conductivity reading
during measurement
• Setting and data during
calibration/setup
Sub-Display
• Temperature during
measurement
• Parameter during setup
LEDLED NumberAlphabetAlphabet
ZM
YL
XK
9WJ
8VI
7UH
6TG
5SF
4RE
3QD
2PC
1OB
0NA
Display and Operation Keys
Alphanumeric Display Table
IM 12D11A01-11E 1st Edition: Mar. 2003
EXA100 Series
* Conductivity meter does not require calibration using standard solution as routine maintenance if the
correct cell constant has been defined upon start-up.
* For details on the calibration using standard solution, refer to Section 6.5 of the SC100 Panel Mount
Converter's User's Manual, IM 12D11A01-01E.
LED
µS/cm
mS/cm
CANCEL
MEASCALSET
S4
S3
S2
S1
ENT
SC100
EXA
* Calibration should be performed if the cell constant has changed due to prolonged operation.
Press [ENT]
key once.
Water
Calibration
Display Top
Replace the sensor in the process.
Calibration result
is reflected to the
measured value.
Main display stops
flashing and sub-display
is flashing.
Set the conductivity
value of the standard
solution using [䉭] and
[䉮] keys.
Press [䉭] key
for 3 sec.
Measurement
Display
1st digit in Main
display is flashing.
Wait until reading
has stabilized.
Press [䉮] key
for 3 sec.
Measurement
Display
Press [ENT] key twice.
Main and sub displays
flash upon each press.
Rinse the conductivity
sensor well with water
or appropriate solution.
Standard solution
Press [ENT]
key once.
µS/cm
mS/cmmS/cm
EXA SC100
S1
S2
S3
S4
mS/cm
µS/cm
EXA SC100
ENT
S1
S2
S3
S4
SET CAL MEAS
CANCEL
µS/cm
mS/cm
CANCEL
MEASCALSET
S4
S3
S2
S1
ENT
SC100EXA
µS/cm
mS/cm
S4
S3
S2
S1
SC100EXA
Wipe thoroughly after rinse,
and then soak the sensor in
the standard solution.
µS/cm
mS/cmmS/cm
EXA SC100
S1
S2
S3
S4
Decimal point
is flashing.
Move the decimal point
using [䉭] and [䉮] keys.
Press [ENT] key to confirm.
µS/cm
mS/cm
EXA SC100
S1
S2
S3
S4
µS/cm
mS/cm
EXA SC100
S1
S2
S3
S4
µS/cm
mS/cm
EXA SC100
S1
S2
S3
S4
Data updated
µS/cm
mS/cm
S4
S3
S2
S1
SC100EXA
Press [ENT]
key once.
Press [ENT]
key once.
IM 12D11A01-11E 1st Edition: Apr. 2003
4. Measurement
Sample Temperature Conditions
SC10XB Conductivity Sensor : 70 8C or lower
Note: Specification requirements vary depending on sample composition.
• The SC10XB can be used in conjunction with a dedicated adapter.
5. Maintenance
• Cleaning the Electrode <Periodical cleaning, sample dependent>
• Cleaning the Electrode
6. Error Messages, Error checking flowchart
• An error message during calibration
For details on error messages, refer to Chapter 8 of the SC100 Panel Mount Converter's User's Manual, IM 12D11A01-01E
Clean the electrode.
Check connections.
NO
NO
NO
NO
NO
The electrode is wet.
The electrode is dirty.
YES
YES
YES
YES
YES
The electrode has
deteriorated. Replace
the sensor with a new
one.
Water
Electrode
Cotton wab or Soft Material
Check connections.
Clean the electrode.
NO
NO
NO
The temperature
sensor is disconnected
or short-circuited.
The electrode is dirty.
Cable connections
are not appropriate.
YES
YES
YES
The electrode has
deteriorated.
Replace the sensor
with a new one.
Zero Calibration (Air Calibration: Eliminating an error resulting from wiring capacity)
Refer to Section 6.4 of the SC100 Panel Mount Converter's User's Manual, IM 12D11A01-01E.
Replace the
sensor.
• An error message during measurement
NO
YES
Dry it thoroughly.
Water
Calibration
Display Top
Replace the sensor
in the process.
Calibration result
is reflected to the
measured value.
Main display stops
flashing and
sub-display is flashing.
Press [ENT] key once.
Press [n] key
for 3 sec.
Measurement
Display
Wait until air
calibration has
been complete
(60 sec).
"Air" in sub-display
is flashing.
Press [,] key
for 3 sec.
Measurement
Display
Press [ENT]
key once.
Rinse the conductivity
sensor well with water
or appropriate solution.
Press [n] key.
µS/cm
mS/cm
EXA SC100
S1
S2 S3
S4
mS/cm
µS/cm
EXA SC100
ENT
S1
S2 S3
S4
SET CAL MEAS
CANCEL
µS/cm
mS/cm
CANCEL
MEASCALSET
S4
S3
S2
S1
ENT
SC100EXA
µS/cm
mS/cm
S4
S3
S2
S1
SC100EXA
Press [ENT]
key once.
In the air
µS/cm
mS/cm
EXA SC100
S1
S2 S3
S4
µS/cm
mS/cm
S4
S3
S2
S1
SC100EXA
µS/cm
mS/cm
EXA SC100
S1
S2 S3
S4
Data updated
µS/cm
mS/cm
EXA SC100
S1
S2 S3
S4
Main display is
flashing and "WAI" in
sub-display is flashing.
µS/cm
mS/cm
EXA SC100
S1
S2 S3
S4
"Str" in sub-display
is flashing.
"CAL" in sub-display
is flashing.
Press [ENT]
key once.
Error message C-2, C-3, or C-4 appears
during calibration.
Error message E-1, E-2, E-3, or
E-5 appears during measurement.
Use standard solution with
correct concentration.
Cable connections
are not appropriate.
Dry the electrode in zero (air)
calibration.
When error message
C-1 appears, bring the
standard solution within
the operating temperature.
* Correct temperature
coefficient should be set when error message C-4 appears.
If an E-4 error message
appears, the converter has malfunctioned.
Contact your sales representative or the nearest service office.
Converter failure has
been detected.
Replace the
converter.
Converter failure
has been detected.
Replace the
converter.
The concentration
of standard solution is
incorrect.

This manual suits for next models

1

Other YOKOGAWA Measuring Instrument manuals

YOKOGAWA FU24 User manual

YOKOGAWA

YOKOGAWA FU24 User manual

YOKOGAWA AQ6376 User manual

YOKOGAWA

YOKOGAWA AQ6376 User manual

YOKOGAWA PH450G User manual

YOKOGAWA

YOKOGAWA PH450G User manual

YOKOGAWA DX Series Release note

YOKOGAWA

YOKOGAWA DX Series Release note

YOKOGAWA TA220 User manual

YOKOGAWA

YOKOGAWA TA220 User manual

YOKOGAWA FLEXA FLXA21-D-S Guide

YOKOGAWA

YOKOGAWA FLEXA FLXA21-D-S Guide

YOKOGAWA 800 plus User manual

YOKOGAWA

YOKOGAWA 800 plus User manual

YOKOGAWA vigilantplant OX102 User manual

YOKOGAWA

YOKOGAWA vigilantplant OX102 User manual

YOKOGAWA 701937 User manual

YOKOGAWA

YOKOGAWA 701937 User manual

YOKOGAWA 701931 User manual

YOKOGAWA

YOKOGAWA 701931 User manual

YOKOGAWA 96060 User manual

YOKOGAWA

YOKOGAWA 96060 User manual

YOKOGAWA AXF series User manual

YOKOGAWA

YOKOGAWA AXF series User manual

YOKOGAWA mR10000 436101 User manual

YOKOGAWA

YOKOGAWA mR10000 436101 User manual

YOKOGAWA wt1600 User manual

YOKOGAWA

YOKOGAWA wt1600 User manual

YOKOGAWA ZR802S User manual

YOKOGAWA

YOKOGAWA ZR802S User manual

YOKOGAWA Powercert PR201 User manual

YOKOGAWA

YOKOGAWA Powercert PR201 User manual

YOKOGAWA WT3001E User manual

YOKOGAWA

YOKOGAWA WT3001E User manual

YOKOGAWA 700929 User manual

YOKOGAWA

YOKOGAWA 700929 User manual

YOKOGAWA Power Series Plus User manual

YOKOGAWA

YOKOGAWA Power Series Plus User manual

YOKOGAWA FLXA202 Manual

YOKOGAWA

YOKOGAWA FLXA202 Manual

YOKOGAWA 701928 User manual

YOKOGAWA

YOKOGAWA 701928 User manual

YOKOGAWA WT5000 Specification sheet

YOKOGAWA

YOKOGAWA WT5000 Specification sheet

YOKOGAWA SR10000 series User manual

YOKOGAWA

YOKOGAWA SR10000 series User manual

YOKOGAWA TA520 User manual

YOKOGAWA

YOKOGAWA TA520 User manual

Popular Measuring Instrument manuals by other brands

Omron K3HB-R manual

Omron

Omron K3HB-R manual

TELEWAVE 44L1 Operation manual

TELEWAVE

TELEWAVE 44L1 Operation manual

Forward STM-6000 user manual

Forward

Forward STM-6000 user manual

Dräger X-AM 5000 Technical manual

Dräger

Dräger X-AM 5000 Technical manual

Bosch PTD 1 Original instructions

Bosch

Bosch PTD 1 Original instructions

SellEton Scales OP-903 user manual

SellEton Scales

SellEton Scales OP-903 user manual

Unipower Combicontrol PIC432B manual

Unipower

Unipower Combicontrol PIC432B manual

Applent Instruments AT512 user manual

Applent Instruments

Applent Instruments AT512 user manual

Olympus EPOCH 650 user manual

Olympus

Olympus EPOCH 650 user manual

Bosch DLR165 Operating/safety instructions

Bosch

Bosch DLR165 Operating/safety instructions

MRC LX-1108 Operation manual

MRC

MRC LX-1108 Operation manual

Hilti PR 3-HVSG Original operating instructions

Hilti

Hilti PR 3-HVSG Original operating instructions

Thermo Scientific Orion 8030cX instruction manual

Thermo Scientific

Thermo Scientific Orion 8030cX instruction manual

HT HTA106 user manual

HT

HT HTA106 user manual

Burkert ELEMENT neutrino 8202 operating instructions

Burkert

Burkert ELEMENT neutrino 8202 operating instructions

az-instrument 8902 Operation manual

az-instrument

az-instrument 8902 Operation manual

Toshiba LF434 manual

Toshiba

Toshiba LF434 manual

493K K-METRON User instructions

493K

493K K-METRON User instructions

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.