OPERATOR’S MANUAL
Also available in other languages
OPERATOR’S MANUAL
25356_cover.qxd 12/3/2004 2:19 PM Page 1
To Purchase, Visit Avobus.com or call 1-800-674-3655

46
EnteraLite®Infinity™Operator’s Manual © Copyright 2004 ZEVEX Incorporated
TABLE OF CONTENTS
i
Page
PRODUCT OVERVIEW ........................................................................... iii
LIST OF WARNINGS AND CAUTIONS ......................................................... iv
1. PUMP COMPONENTS .......................................................................... 1
EnteraLite®Infinity™Enteral Feeding Pump ........................................ 1
Keypad ....................................................................................... 3
Display ....................................................................................... 4
Disposable Set ............................................................................. 5
Symbols ...................................................................................... 6
2. DIRECTIONS FOR USE ........................................................................ 7
Priming and Loading the Disposable Set ........................................... 7
Operating the Pump
A Single Feeding Example ...................................................... 10
An Interval Feeding Example .................................................. 13
The Battery
Running on Battery Power ..................................................... 17
Charging the Battery ............................................................. 18
User Preference Settings .............................................................. 19
3. ALARMS, MESSAGES, AND INDICATIONS .............................................. 21
4. CLEANING ..................................................................................... 28
5. ACCESSORIES ................................................................................. 29
6. SPECIFICATIONS ............................................................................. 31
7. IMPORTANT PEDIATRIC CONSIDERATIONS ............................................ 34
8. ADDITIONAL TECHNICAL INFORMATION .............................................. 36
9. WARRANTY .................................................................................... 37
LIST OF FIGURES ................................................................................ 38
WHO TO CALL .................................................................................... 41
INDEX .............................................................................................. 43
NOTES
Notes
25356_pages.qxd 2/2/2005 2:49 PM Page 1

ii
EnteraLite®Infinity™Operator’s Manual © Copyright 2004 ZEVEX Incorporated
45
Index
P
Pediatric Considerations 34
Pole Clamp 2, 7, 29
Prime Feature 3, 8, 26
Prime Key 3
Priming the Set 7
Program Memory 12, 16, 32
Pump Components
Air Sensor 1
Charger/Power Connector 2
Display 1, 4
Door 1
Downstream Pressure Sensor 1
Keypad 1, 3
Pump Instruction Label 2
Pump Wheel 1
Receptacle for Cassette 1
Receptacle for Pole Clamp 2
Speaker 2
Upstream Pressure Sensor 1
R
RATE/DOSE Key 3
Rinse 28
RUN/PAUSE Key 3
S
Self Test 9
Serial Number 2, 9
Settings Lock 19
Single Feeding Example 10
Size (Pump Specifications) 31
Storage Environment 33, 36
Symbols 6
U
User Preference Settings
Alarm Volume 19
Settings Lock 19
Dose Done Alarm 19
Backlight 19
V
VOL/TOTAL Key 3
W
Water 28
Warranty 37
25356_pages.qxd 2/2/2005 2:49 PM Page 2

44
EnteraLite®Infinity™Operator’s Manual © Copyright 2004 ZEVEX Incorporated
PRODUCT OVERVIEW
The EnteraLite®Infinity™is a rotary peristaltic enteral feeding pump designed
to deliver programmed doses of enteral nutrition solutions at selectable rates.
It is easy to use with a simple user interface and keypad. ZEVEX technology
allows an accuracy rate of +/- 5%, and the ability to safely operate in any ori-
entation. The EnteraLite®Infinity™system includes a complete line of DEHP-free
pump delivery sets with automatic free-flow protection.
The EnteraLite®Infinity™is designed for both hospital and alternate care use.
It is durable, water resistant, and easily carried in a ZEVEX carry pack. The
EnteraLite®Infinity™‘s size, weight, accuracy, and portability promote and sup-
port physical activity.
iii
Index
F
Feed Int Key 3
Feed Interval 3, 13, 25, 26, 32
Feeding Completion 12, 16
Feeding Example
Single Feeding Example 10
Interval Feeding Example 13
Flow Rate 10, 14, 31, 34
Foam 7, 24
Free-flow 33
Fuel Gauge 4, 18
I
Increase (+) Key 3
Interval Feeding Example 13
Indications
Battery Doesn’t Hold Its Charge 27
Battery Doesn’t Charge 27
CHEK 26
Charger Installed but No Plug
Symbol Visible 27
TO STOP LET GO 9, 26
Instructions for Use
Loading the Set 7
Priming the Set 7
Removing Air 8
K
Keypad 3
+ (Increase) Key 3
- (Decrease) Key 3
CLEAR Key 3
FEED INT Key 3, 13
ON/OFF Key 3
PRIME Key 3, 9, 26
RATE/DOSE Key 3
RUN/PAUSE Key 3
VOL/TOTAL Key 3, 11, 15
L
Less (-) Key 3
Lock Settings 19
M
Maintenance 36
Messages
DOSE DONE 12, 25
NEXT DOSE 16, 25
Mini Backpack 30
More (+) Key 3
N
Next Feeding 13, 16
O
Occlusion (Blockage) 23, 32, 34
ON/OFF Key 3
Operational Conditions 33, 36
25356_pages.qxd 2/2/2005 2:49 PM Page 3
To Purchase, Visit Avobus.com or call 1-800-674-3655

LIST OF WARNINGS AND CAUTIONS
iv
EnteraLite®Infinity™Operator’s Manual © Copyright 2004 ZEVEX Incorporated
43
WARNINGS
Use ONLY feeding solutions prescribed by the responsible physician, regis-
tered dietitian, registered nurse, or other licensed practitioner.
Use ONLY EnteraLite®Infinity™Disposable Sets to ensure proper fluid deliv-
ery. Others will not deliver the correct dose, may allow dangerous free-flow
conditions, and may generate hazardous fluid pressures which may activate
occlusion alarms at unpredictable pressures.
Low flow rates, combined with high dose settings may exceed the life of the
disposable set. The disposable set must be replaced every 24 hours to main-
tain delivery accuracy, allow proper air and occlusion sensing, and prevent
growth of bacteria. DO NOT PROGRAM RATE AND DOSE COMBINATIONS WHICH
EXCEED A 24 HOUR FEED REGIMEN.
The battery capacity is an approximation. If you are unsure that enough
capacity remains for your intended use, recharge it.
To avoid electrical shock, never clean pump with charger plugged into an
outlet or when pump is on.
Do not use EnteraLite®Infinity™Enteral Feeding Pump for delivery of non-
enteral solutions. Serious injury may result.
Proper operation of pump requires door is closed and latched. Make sure
door is closed and latched when motor is running.
INDEX
Index
+ Key 3
- Key 3
A
AC Adapter/Charger 18, 29
Accuracy 31,34
Accessories
AC Adapter/Charger 18, 29
Backpack 30
Pole Clamp 2, 7, 29
Air Travel 36
Alarms 21
ERRA-ERRZ 22
LOAD SET 22
LOW BATT 22
NO FLOW IN 23
NO FLOW OUT 23
NO FOOD 24
PUSH RUN TO FEED 25
Alarm Volume 19
B
Battery
AC Adapter/Charger 18, 29
Battery Symbol 17
Charge Level Indicator 17, 31
Charge Time 17, 31
Charging the Battery 18, 31
Fuel Gauge 4, 18
LOW BATT Alarm 22
Power Indicator 4, 17, 18
Blockage (Occlusion) 23, 32, 34
C
Carry Packs 30
Classification Information 33
Cleaning 28
Clear Key 3
Compact Wall Charger 18, 29
D
Decrease (-) Key 3
Display
Battery Symbol 4, 18
Display Light 4
Fuel Gauge 4, 18
Run Symbol 4
Wall Plug Symbol 4, 17
Directions for Use
Loading the Set 7
Priming the Set 7
Removing Air 8
Disposable Set
500ml or 1200ml Bag 5
Adapter Protective Cover 5, 7
Cassette 5
Spike 5
Spike Protected Cover 5, 7
Stepped Enteral Adapter 5
Disposable Set Displacement 36
Dose 10, 14, 31
25356_pages.qxd 2/2/2005 2:49 PM Page 4

42
EnteraLite®Infinity™Operator’s Manual © Copyright 2004 ZEVEX Incorporated
v
Total Volume may be incorrect due to an interruption of battery power if the
overcurrent or temperature protection circuits are triggered. In such a case
one of the following conditions occurs:
During Single Feeding: Pump enters pause mode, but does not dis-
play “DOSE DONE” or turns off (but not due to a low battery).
During Interval Feeding: Pump enters pause mode or turns off (but
not due to low battery).
CAUTIONS
Federal law (U.S.A.) restricts this device to sale by or on the order of a
physician, registered dietitian, registered nurse, or other licensed practi-
tioner.
Dispose of EnteraLite®Infinity™Disposable Sets properly, as required by
local law.
If any leaks are detected in the disposable set, stop pump operation and
disconnect set from patient and replace with a new disposable set.
Care should be used when manually priming delivery set to ensure cassette
is not damaged by excess force.
25356_pages.qxd 2/2/2005 2:49 PM Page 5

© Copyright 2004 ZEVEX Incorporated
41
WHO TO CALL
Additional Information:
YOUR HEALTHCARE PROVIDER:
_______________________________________________________________
_______________________________________________________________
_______________________________________________________________
_______________________________________________________________
_______________________________________________________________
YOUR PHYSICIAN:
_______________________________________________________________
_______________________________________________________________
_______________________________________________________________
_______________________________________________________________
_______________________________________________________________
ZEVEX Incorporated
4314 ZEVEX Park Lane
Salt Lake City, Utah 84123 USA
Customer Service and Technical Support: (800) 970-2337
www.zevex.com
vi
EnteraLite®Infinity™Operator’s Manual
25356_pages.qxd 2/2/2005 2:49 PM Page 6
To Purchase, Visit Avobus.com or call 1-800-674-3655

40
EnteraLite®Infinity™Operator’s Manual © Copyright 2004 ZEVEX Incorporated
1
1. PUMP COMPONENTS
1. Pump Components
EnteraLite®Infinity™Enteral Feeding Pump
1. Door
2. Pump Wheel
3. Display
4. Upstream Pressure Sensor
5. Air Sensor
6. Downstream Pressure Sensor
7. Receptacle for Cassette
8. Keypad
Figure 1-1 Pump: Front
2
3
1
4
5
6
7
8
Figure Description Page
2-29 Backlight Settings: LITE OFF and LITE ON ............................. 19
3-1 ERRA and ERRZ Alarm ........................................................ 22
3-2 LOAD SET Alarm ................................................................ 22
3-3 LOW BATT Alarm ............................................................... 22
3-4 NO FLOW IN Alarm ............................................................ 23
3-5 NO FLOW OUT Alarm .......................................................... 23
3-6 NO FOOD Alarm ................................................................. 24
3-7 PUSH RUN TO FEED Alarm ................................................... 25
3-8 DOSE DONE Message .......................................................... 25
3-9 NEXT DOSE Message ........................................................... 25
3-10 CHECK Indication .............................................................. 26
3-11 TO STOP LET GO Indication ................................................. 26
3-12 Battery Symbol and the E and F of the Fuel Gauge .................. 27
4-1 Rinsing Pump Under Stream of Water .................................... 28
5-1 AC Adapter/Charger ........................................................... 29
5-2 Pole Clamp ...................................................................... 29
5-3 Pole Clamp with Pump Attached to IV Pole ............................ 29
5-4 Backpack ........................................................................ 30
5-5 Backpack (Cutaway View of Internal Compartments) ............... 30
5-6 Backpack (Cutaway View of Internal Straps) .......................... 30
25356_pages.qxd 2/2/2005 2:49 PM Page 7

EnteraLite®Infinity™Enteral Feeding Pump
1. Pump Instruction Label
2. Speaker
3. Receptacle for Pole Clamp
4. Serial Number Label
5. Charger/Power Connector
2
EnteraLite®Infinity™Operator’s Manual © Copyright 2004 ZEVEX Incorporated
39
1. Pump Components
Figure 1-2 Pump: Rear
3
1
25
Figure Description Page
2-6 Tubing Segment Being Pinched ............................................ 8
2-7 Seat Cassette ..................................................................... 9
2-8 Close Pump Door ................................................................ 9
2-9 Priming Indication ............................................................. 9
2-10 Program Rate to 120 ml/hr ................................................ 10
2-11 Program Dose to 500 ml/hr ............................................... 10
2-12 Clear Feed Interval ........................................................... 11
2-13 Pump is Running ............................................................. 11
2-14 Amount Currently Delivered ............................................... 11
2-15 Total Amount Delivered .................................................... 11
2-16 DOSE DONE ..................................................................... 12
2-17 Program Rate to 50 ml/hr ................................................. 14
2-18 Program Dose to 100 ml/hr ............................................... 14
2-19 Program Feed Interval to 6.00 hr ........................................ 14
2-20 Pump is Running ............................................................. 15
2-21 Amount Currently Delivered ............................................... 15
2-22 Total Amount Delivered .................................................... 15
2-23 Time Until Next Feeding .................................................... 16
2-24 Battery Power Indicator ..................................................... 17
2-25 Fuel Gauge ...................................................................... 18
2-26 Alarm Volume Settings: BEEP LOW and BEEP HIGH .................. 19
2-27 Settings Lock: UNLK and LOCK ............................................ 19
2-28 Dose Done Settings: BEEP WHEN DONE and MUTE WHEN DONE .. 19
4
25356_pages.qxd 2/2/2005 2:49 PM Page 8

38
EnteraLite®Infinity™Operator’s Manual © Copyright 2004 ZEVEX Incorporated
3
1. Pump Components
Keypad
ON/OFF Press key for 1.5 seconds to
either turn EnteraLite®Infinity™on or
off.
PRIME Press and hold key to rapidly fill
disposable set with fluid. Release key to
stop.
RUN/PAUSE Press key to either start
pump or place it in pause mode.
FEED INT Press key to display feed interval setting.
VOL/TOTAL Press key once to display volume delivered for current
feeding. Press key again to display total cumulative volume delivered
in all feed cycles since total volume was last cleared.
RATE/DOSE Press key to transition between rate and dose settings.
CLEAR Press key to reset displayed function to its minimum value.
+Press key to increase displayed feed function. Press and hold key to
increase value rapidly.
-Press key to decrease displayed feed function. Press and hold key to
decrease value rapidly.
Note: The following keys only function when pump is in pause mode:
PRIME,CLEAR,+,and-.
Figure 1-3 Keypad Layout
LIST OF FIGURES
Figure Description Page
1-1 Pump: Front ...................................................................... 1
1-2 Pump: Rear ....................................................................... 2
1-3 Keypad Layout ................................................................... 3
1-4 Display Segments ............................................................... 4
1-5 Alphanumeric Characters and Units ....................................... 4
1-6 Setting Indicators .............................................................. 4
1-7 Run Indicators ................................................................... 4
1-8 Power Indicators ................................................................ 4
1-9 EnteraLite®Infinity™500ml or 1200ml Bag Set ........................ 5
1-10 EnteraLite®Infinity™Spike Set .............................................. 5
1-11 Symbol: Consult Accompanying Documents ............................. 6
1-12 Symbol: Feeding Bag Connection .......................................... 6
1-13 Symbol: Patient Connection ................................................. 6
1-14 Symbol: Type BF Shock Protection ......................................... 6
1-15 Symbol: IEC 529 ................................................................. 6
1-16 Symbol: Single Patient Use ................................................... 6
1-17 Symbol: DEHP-free ............................................................. 6
2-1 Spike Container .................................................................. 7
2-2 Fill Bag ............................................................................ 7
2-3 Remove Cover from Barbed Adapter /
Place Cover on Pole Clamp ................................................... 7
2-4 Squeeze Bag ...................................................................... 8
2-5 Pinch Tubing ..................................................................... 8
25356_pages.qxd 2/2/2005 2:49 PM Page 9
To Purchase, Visit Avobus.com or call 1-800-674-3655

1. Pump Components
4
EnteraLite®Infinity™Operator’s Manual © Copyright 2004 ZEVEX Incorporated
37
Display
The display includes large alphanumeric
characters, as well as smaller symbols and
words. All segments and symbols on the dis-
play are shown in Figure 1-4.
Settings and messages are displayed in
large alphanumeric characters. (Figure 1-5)
When the large alphanumeric characters dis-
play numbers, one of the smaller words
below the characters indicates which func-
tion is being displayed. (Figure 1-6)
The run symbol indicates pump is running.
(Figure 1-7)
The wall plug symbol indicates the charger
is plugged in, and connected to a live power
outlet. The battery symbol indicates pump
is running on battery power only. The fuel
gauge symbol indicates how much battery
charge remains. (Figure 1-8)
NOTE: The display light will automatically
turn off 10 seconds after the last key is
pressed. The light will also remain on for 10
seconds after the charger is connected
between pump and live power outlet. If you
would prefer the light to remain on when
connected to an outlet, you can change the
pump light setting to ON. See the User
Preference Settings section (Page 19) for
instructions to change this setting.
Figure 1-4 Display Segments
Figure 1-5 Alphanumeric Characters
and Units
Figure 1-6 Setting Indicators
Figure 1-7 Run Indicators
Figure 1-8 Power Indicators
9. WARRANTY
9. Warranty
Solely for the benefit of the original buyer, ZEVEX INCORPORATED ("ZEVEX"),
warrants all new EnteraLite®Infinity™products of its manufacture to be free
from defects in material and workmanship, and will replace or repair, F.O.B., at
its factory in Salt Lake City, Utah, or other location designated by ZEVEX, any
EnteraLite®Infinity™products returned to it within twenty-four (24) months of
original purchase by the buyer. Such repair or replacement shall be free of
charge.
ZEVEX warrants to the original buyer, all repaired or replaced products to be
free from defects in material and workmanship and will replace or repair such
products F.O.B., at its factory in Salt Lake City, Utah, or other location desig-
nated by ZEVEX. Such repair or replacement shall carry a warranty of ninety
(90) days from the date of repair or replacement or the balance of the new
product warranty as described above, whichever is greater.
This Warranty applies to all EnteraLite®Infinity™products manufactured by
ZEVEX and is the ONLY WARRANTY GIVEN FOR THE SALE OF PRODUCTS OR SER-
VICES. NO WARRANTIES IMPLIED IN LAW, INCLUDING, BUT NOT LIMITED TO THE
IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR PARTICULAR
PURPOSE, SHALL APPLY. ZEVEX WILL BE LIABLE, IN ANY EVENT, ONLY FOR THE
PURCHASE PRICE OF THE DEFECTIVE PRODUCT, BUT NOT FOR ANY CONSEQUEN-
TIAL DAMAGES.
This Warranty may not be modified, amended or otherwise changed, except by
a written document properly executed by a corporate officer of ZEVEX.
ZEVEX, Inc. voids the warranty if the EnteraLite®Infinity™pump is opened or
tampered with in any way without prior authorization from ZEVEX, Inc.
25356_pages.qxd 2/2/2005 2:49 PM Page 10

36
EnteraLite®Infinity™Operator’s Manual © Copyright 2004 ZEVEX Incorporated
1. Pump Components
5
Disposable Set
The only disposable sets approved for use with EnteraLite®Infinity™pump are:
EnteraLite®Infinity™500 ml Bag Set, Order Number INF0500
EnteraLite®Infinity™1200 ml Bag Set, Order Number INF1200
EnteraLite®Infinity™Spike Set, Order Number INF0010
Figure 1-10 EnteraLite®Infinity™Spike Set
Figure 1-9 EnteraLite®Infinity™500ml or 1200ml Bag Set
5
7
6
8
9
23 4
1. “ “ Symbol
2. Cassette
3. Downstream Tubing
4. Upstream Tubing
5. 500ml or 1200ml Bag
6. Barbed Enteral Adapter
7. Adapter Protective Cover
8. Spike Protective Cover
9. Spike
1
8. ADDITIONAL TECHNICAL INFORMATION
8. Additional Technical Information
Operational Conditions
EnteraLite®Infinity™is designed in accordance with EN 60601-1-2, EN
60601-1-4, and RTCA DO-160D standards for electromagnetic emissions and
immunity. Normal operation should not interfere with nor be affected by
other electronic equipment. EnteraLite®Infinity™can safely be operated on
commercial aircraft.
EnteraLite® Infinity™ Disposable Set Displacement
EnteraLite®Infinity™Disposable Set displaces approximately 15ml of fluid
when primed. Approximately 3.5 ml of air is drawn into the tubing before
the enteral pump detects that bag is empty. Therefore, approximately 11.5
ml of fluid remains in the disposable set when the pump stops feeding.
(These values vary by approximately 0.5 ml, depending on whether battery
is low or fully charged).
Routine Maintenance
There are no user serviceable parts or routine calibration or adjustment pro-
cedures required for EnteraLite®Infinity™. All service requests should be
referred to ZEVEX technical service.
25356_pages.qxd 2/2/2005 2:49 PM Page 11

1. Pump Components
6
EnteraLite®Infinity™Operator’s Manual
Symbols
© Copyright 2004 ZEVEX Incorporated
35
Brief instructions for operation of pump as
well as a brief explanation of each alarm
message which pump may communicate are
printed on the label attached to pump.
These are not intended to be used in place
of the Operator's Manual. They are simply a
quick reference guide. Please read the
Operator's Manual before operating the
EnteraLite®Infinity™!
The bag symbol is printed on pump door
over port where upstream tubing enters
pump.
The patient symbol is printed on pump door
over port where downstream tubing exits
pump.
EN 60601-1 Type BF degree of protection
against electrical shock. No electrical con-
nection to patient. Drop from any angle
from height of 3 feet shall not damage
pump operation.
IEC 529 degree of protection against water
entering the enclosure. Water jets from any
direction shall have no effect.
This symbol is printed on pump delivery
sets. It indicates only one patient should
use each disposable set.
This symbol is printed on pump delivery
sets. It indicates pump delivery sets are
made with materials that do not contain the
plasticizer DEHP.
Figure 1-12
Feeding Bag Connection
Figure 1-13
Patient Connection
Figure 1-14
Type BF Shock Protection
Figure 1-15
IEC 529
IPX5
Figure 1-11
Consult Accompanying
Documents
Consult Accompanying
Documents!
Figure 1-16
Single Patient Use
Figure 1-17
DEHP-free
7. Important Pediatric Considerations
Occlusion alarm pressure of EnteraLite®Infinity™is 5 psi (34kPa) for
upstream occlusion and 12 psi (83 kPa) for a downstream occlusion.
If these specifications meet the required feed regimen, EnteraLite®Infinity™
can be used to deliver enteral formulas to pediatric patients. ALWAYS VERIFY
RATE, DOSE AND FEED INTERVAL BEFORE PROCEEDING TO FEED.
25356_pages.qxd 2/2/2005 2:49 PM Page 12
To Purchase, Visit Avobus.com or call 1-800-674-3655

34
EnteraLite®Infinity™Operator’s Manual © Copyright 2004 ZEVEX Incorporated
2. Directions for Use
7
2. DIRECTIONS FOR USE
Recommendation for First Use:
Since battery may not be fully charged when pump is first received, it is rec-
ommended that battery be charged for 6 hours prior to operating on battery
power (see Page 18 for additional information).
Priming and Loading the
Disposable Set
Step 1:
If you are using the spike set with a vented
bottle or pre-filled bag:
Remove protective cover from spike, and insert
spike into spike adapter of formula container
(Figure 2-1).
If you are using the 500ml or 1200ml dispos-
able set:
Hold bag upright and pour in feeding solution
(Figure 2-2). Close cap securely.
NOTE: Blenderized or aggressively mixed solu-
tions may have foam. If using this type of
solution, allow it to sit for 10 to 15 minutes
before pouring into bag. This will reduce the
chance of an alarm due to air in the tubing.
Step 2:
Remove protective cover from barbed adapter. If
using pole clamp, protective cover may be held
in the groove on back of clamp (Figure 2-3).
Figure 2-1 Spike Container
Figure 2-2 Fill Bag
Figure 2-3 Remove Cover from
Barbed Adapter / Place Cover
on Pole Clamp
7. IMPORTANT PEDIATRIC CONSIDERATIONS
7. Important Pediatric Considerations
EnteraLite®Infinity™can be used on pediatric patients if the specifications of
pump meet delivery requirements of the patient. The specifications of primary
importance are:
The flow rate range of EnteraLite®Infinity™is 0.1 ml/hr to 10 ml/hr in 0.1
ml/hr increments and 10 ml/hr to 600 ml/hr in 1 ml/hr increments.
The EnteraLite®Infinity™pump delivers the dose at the specified rate with-
in +/- 5% volumetric flow rate accuracy under the following conditions:
NOTE: For test methodology refer to ANSI/AAMI ID26-1992 American National
Standard for Infusion Devices. Reported accuracy is based upon a generic vol-
ume accuracy calculation:
Accuracy % = (Volume Delivered - Volume Programmed) x 100
Volume Programmed
utilizing ZEVEX EnteraLite® Infinity™ disposable sets
fluid head height at +6.0 inches ± 0.3 inches (+15.24 cm ± 0.76 cm)
with respect to center of rotor.
over the range of environmental operating conditions (Specified per
IEC 60601-1)
Operating Temperature Range: 41 °F to 104 °F (5 °C to 40 °C)
Operating Humidity Range: 30 to 75% noncondensing
25356_pages.qxd 2/2/2005 2:49 PM Page 13

2. Directions for Use
8
EnteraLite®Infinity™Operator’s Manual © Copyright 2004 ZEVEX Incorporated
33
NOTE: If set is to be used with a carry pack, all
air must be removed from bag and tubing.
Continue to step 3 for instructions on
removing air. If set is to be hung above pump,
i.e. on an IV pole, you may skip to step 4.
Step 3:
Turn bag upside down and gently squeeze. Tilt bag
as needed to keep air at tubing port (Figure 2-4).
Step 4:
Gently pinch teal colored tubing below “ “
symbol. Hold this position until air is removed
from tubing. Gently squeeze bag at same time
to assist fluid flow. If fluid does not flow, pinch
may be too tight (Figure 2-5).
NOTE: Inside the teal colored tubing, below the
“ “ symbol is the in-line occluder. By pinching
tubing gently, the tubing moves away from the
in-line occluder allowing fluid to flow (Figure 2-
6). It is important to only pinch the tubing
below the “ “ symbol to avoid damaging the
in-line occluder.
NOTE: Air may also be removed from tubing
using the pump prime feature. See step 7 for
instructions on using the pump prime feature.
Step 5:
Loop silicone tubing around pump wheel
stretching lightly. Seat cassette into pump
(Figure 2-7). Close pump door (Figure 2-8).
Figure 2-4 Squeeze Bag
Figure 2-5 Pinch Tubing
Tubing
Port
Figure 2-6 Tubing Segment
Being Pinched
6. Specifications
DISPOSABLE SET FEATURES
Automatic Free-flow Protection
No Drip Chamber
DEHP-free
ACCESSORIES
AC Adapter/Charger
Multi-position Pole Clamp (rotatable 360°)
Multi-purpose EnteraLite®Infinity™Carry Packs
WARRANTY
2 years
OPERATING ENVIRONMENT
Temperature: 41 °F to 104 °F (5 °C to 40 °C)
Humidity: 30% to 75% noncondensing
STORAGE ENVIRONMENT
Temperature: -4 °F to 149 °F (-20 °C to 65 °C)
Humidity: 10% to 95% noncondensing
CLASSIFICATION INFORMATION
EnteraLite®Infinity™Enteral Feeding Pump with internal battery –
Type BF Applied Part
EnteraLite®Infinity™AC Adapter / Battery Charger – Class II
25356_pages.qxd 2/2/2005 2:50 PM Page 14

32
EnteraLite®Infinity™Operator’s Manual © Copyright 2004 ZEVEX Incorporated
2. Directions for Use
9
Step 6:
Press and hold ON/OFF key for 1.5 seconds
to turn pump on. While pump runs through
a self test, display will light and an audi-
ble alarm will sound as pump displays the nine
digit serial number three digits at a time for
one second each. The next number displayed
for one second, a number beginning with an
"R", is the software revision.
Next, alarm will stop and all segments of dis-
play will be shown for 2 seconds. Verify all
display segments and symbols are active.
The self test is completed and pump will then
display last programmed rate and will be in
pause mode.
If any air is still in the tubing, continue to
Step 7 to use pump prime feature.
Step 7:
Press and hold PRIME key. Alarm will sound
once and pump will begin pumping at the
maximum rate of 600 ml/hr. Display will
read "TO STOP LET GO" (Figure 2-9). Once all air
is removed from tubing, release key. Pump will
stop, display will revert to last programmed
rate, and pump will be in pause mode.
For pump operation instructions:
For a Single Feeding Example, go to Page 10.
For an Interval Feeding Example, go to Page 13.
Figure 2-9
Priming Indication
Figure 2-8 Close Pump Door
Figure 2-7 Seat Cassette
6. Specifications
VOLUME DISPLAY
Current Dose Status
Accumulated
INTERVAL FEED
Interval Feed Indication
Displays Time to Next Feed
PROGRAM MEMORY RETAINED
When Pump is on or off
PEDIATRIC USE Yes
PRIME Yes
ALARMS
Chek Programming error
Load Set Set not loaded properly
Low Batt Battery Low
No Flow In Upstream occlusion
-5 psi (-34 kPa)
Tolerance = ± 3psi (21 kPa)
psi (pounds per square inch)
kPa (kilopascals)
No Flow Out Downstream occlusion,
12 psi (83 kPa)
Tolerance = ± 3psi (21 kPa)
psi (pounds per square inch)
kPa (kilopascals)
No Food Bag is empty
ERRA - ERRZ Self-test error
25356_pages.qxd 2/2/2005 2:50 PM Page 15
To Purchase, Visit Avobus.com or call 1-800-674-3655

Operating the Pump
A Single Feeding Example
If you want to deliver 500 ml of enteral feeding solution at a rate of 120 ml/hr
and then stop the pump:
Step 1:
Prepare the disposable set with enteral feeding solution as described in the
Priming and Loading the Disposable Set instructions (page 7) .
Step 2:
Hang feeding bag or container so that the bottom of bag is at or above the
level of the pump door.
OR - If an EnteraLite®Infinity™carry pack is to be used, load pump and
feeding bag into the proper compartments, securing pump, bag and tubing
with the pack's straps (Page 30).
Step 3:
Turn pump on by pressing the ON/OFF key. After running
the self test, display will show last programmed rate.
Step 4:
Press the +or -key to change the rate to 120 ml/hr.
Hold down either key to change rapidly (Figure 2-10).
Step 5:
Press the RATE/DOSE key to display dose. Press the +or
-key until a dose of 500 ml is displayed (Figure 2-11).
2. Directions for Use
10
EnteraLite®Infinity™Operator’s Manual © Copyright 2004 ZEVEX Incorporated
31
Key Sequence for “A Single Feeding Example”
120 ml/h 500 ml
Figure 2-10 Program
Rate to 120 ml/hr
Figure 2-11 Program
Dose to 500 ml
6. SPECIFICATIONS
6. Specifications
SIZE
Pump Dimensions: 1.95” H x 5.65” W x 4.05” D
Pump Weight: 14.4 oz (411.0 g)
BATTERY
Type: Lithium Ion
Life: 24 hours @ 125 ml/hr
Charge Time: approximately 6 hours
Charge Level Indicator Yes
Compact Wall Charger Yes
DISPLAY
Backlit LCD
OPERATING ORIENTATION
Any
FLOW RATE
Range: 0.1 - 600 ml/hr
Increment: 0.1 ml/hr from 0.1 ml/hr to 9.9 ml/hr
1 ml/hr from 10 ml/hr to 600 ml/hr
Accuracy: ± 5%
DOSE
Range: 0.1 ml to 3000 ml or infinite dose
Increment: 0.1 ml from 0.1 - 9.9 ml
1 ml from 10 - 999 ml
10 ml from 1000 - 3000 ml
25356_pages.qxd 2/2/2005 2:50 PM Page 16

30
EnteraLite®Infinity™Operator’s Manual © Copyright 2004 ZEVEX Incorporated
2. Directions for Use
NOTE: For a single feeding the Feed interval setting
should be programmed to NONE (Figure 2-12). When
you have finished setting rate and dose check the dis-
play to see if it reads FEED INTERVAL. If it does, press
the FEED INT key, then press the CLEAR key. Display
will read NONE.
Step 6
Connect barbed adapter to the patient's enteral feeding tube.
Step 7
Press the RUN/PAUSE key. Pump will begin running.
Display will show programmed rate and the arcs
around the run symbol will rotate (Figure 2-13).
While pump is running, the following may be viewed
by pressing the appropriate key: Rate, Dose, and Feed
Interval.
To view the amount delivered in the current feeding,
press the VOL/TOTAL key (Figure 2-14). This counter
will reset itself when the previously programmed dose
has been completed and a new feeding is started, or
if a feeding is interrupted and Rate, Dose or Feed
Interval has been changed.
To view the amount delivered over the course of sev-
eral feedings press the VOL/TOTAL key a second time.
Display will read TOTL then the amount. This counter
never resets itself. It can only be reset by the user
(Figure 2-15).
While pump is running, the settings cannot be changed and the Prime fea-
ture is disabled.
If you want to stop pump at any time, press the RUN/PAUSE key. Display
will show rate, and pump will be in pause mode. Or, turn pump off by press-
ing the ON/OFF key.
11
Figure 2-12
Clear Feed Interval
Figure 2-13
Pump is Running
Figure 2-14 Amount
Currently Delivered
Figure 2-15 Total
Amount Delivered
5. Accessories
EnteraLite®Infinity™Mini-Backpack
Order Numbers PCK1001 (red), PCK1002 (teal)
Pack holds EnteraLite®Infinity™with a 500 ml bag in the front section or can
accommodate a 1200 ml bag in the rear section. Pump is secured with a Velcro®
strap. A Velcro®strap secures the neck of the 500 ml or 1200 ml bag. The tub-
ing is loaded into pump and any excess length of tubing can be secured with
a Velcro®tab. The downstream tubing then feeds through a port at bottom of
pack (either side). Also included in this pack is a pocket that may be used to
hold an ice pack.
Figure 5-4 Backpack
Ice Pack
Pocket
500ml Bag
Compartment
Pump
Compartment
1200ml Bag
Compartment
Figure 5-5 Backpack
(Cutaway View of Internal
Compartments)
Figure 5-6 Backpack
(Cutaway View of
Internal Straps)
Feeding
Bag Strap
Tubing
Strap
Tubing
Strap
25356_pages.qxd 2/2/2005 2:50 PM Page 17

2. Directions for Use
12
EnteraLite®Infinity™Operator’s Manual
When you would like to restart pump, press the ON/OFF and/or the
RUN/PAUSE key. Pump will save the memory of where it was in the feeding
before stopping. Press the RUN/PAUSE key and pump will restart at the
point where it was stopped.
NOTE: Any changes to pump settings during a feeding cycle will cause pump
to start a new feeding; it will not start where it left off.
Feeding Completion:
When the dose has been completely delivered, pump will stop running, and
display will read DOSE DONE. Pump will be in a pause mode until it is turned
off or feeding is restarted. To clear DOSE DONE, press and hold the ON/OFF
key for 1.5 seconds to turn pump off, or press the RUN/PAUSE key to put
pump in pause mode. Adjustments to settings can be made at this time, or
press the RUN/PAUSE key a second time to start another feeding using
same rate, dose, and feed interval settings.
If you would prefer the alarm to sound when the dose is
complete, you can change the pump dose complete
alarm setting to BEEP WHEN DONE. See the User
Preference Settings section (Page 19) for instructions to
change this setting. When set to BEEP WHEN DONE, the
pump will stop at the end of the feeding, beep inter-
mittently, and display will read DOSE DONE (Figure
2-16). To silence alarm when the dose is complete,
press and hold the ON/OFF key for 1.5 seconds to turn
pump off, or press the RUN/PAUSE key to put pump in
pause mode.
NOTE: To program an infinite dose, press and hold the +key until display
reads INF. When an infinite dose has been programmed, pump will not alarm
DOSE DONE. Pump will run continuously as long as feeding solution remains
in the bag. When empty, pump will alarm NO FOOD or NO FLOW IN.
© Copyright 2004 ZEVEX Incorporated
29
Figure 2-16 DOSE DONE
5. ACCESSORIES
5. Accessories
Additional accessories are available for use with EnteraLite®Infinity™:
EnteraLite®Infinity™AC Adapter/Charger
Order Number INFCH01
Plug AC Adapter/Charger into a wall outlet and plug
connector into pump to operate on AC power
and/or recharge battery.
EnteraLite®Infinity™Pole Clamp
Order Number Z-11981
Thread bolt into receptacle on the back of pump
and tighten gray wheel to mount pump on clamp.
If gray wheel is slightly loosened, pump can be
rotated to snap into different positions. Retighten
gray wheel when pump is in the desired position.
Tighten black wheel to mount clamp on a pole.
Figure 5-1 AC Adapter/Charger
Figure 5-2 Pole Clamp
Figure 5-3 Pole Clamp with
Pump Attached to IV Pole
25356_pages.qxd 2/2/2005 2:50 PM Page 18
To Purchase, Visit Avobus.com or call 1-800-674-3655

28
EnteraLite®Infinity™Operator’s Manual © Copyright 2004 ZEVEX Incorporated
2. Directions for Use
Next Feeding:
Turn pump on. Previous settings for rate, dose, and feed interval will be
saved in memory. If no changes are required, verify each setting is correct
and restart pump running.
NOTE: Rate, dose, and feed interval settings can be locked so that changes
cannot be made. See the User Preference Settings section (Page 19) for
instructions to change this feature.
Operating the Pump
An Interval Feeding Example
If you want to deliver 100 ml of enteral feeding solution at a rate of 50 ml/hr,
and repeat this feeding every 6 hours:
Step 1:
Prepare the disposable set with enteral feeding solution as described in the
Priming and Loading the Disposable Set instructions (Page 7).
Step 2:
Hang feeding bag or container so that the bottom of bag is at or above the
level of the pump door.
OR - If an EnteraLite®Infinity™pack is to be used, load the pump and feed-
ing bag into the proper compartments, securing pump, bag and tubing with
the pack's straps (see Page 30).
Step 3:
Turn pump on by pressing the ON/OFF key. After running the self test, dis-
play will show last programmed rate.
13
Key Sequence for “An Interval Feeding Example”
50 ml/h 100 ml 6.00 hr
4. CLEANING
4. Cleaning
WARNING: To avoid electrical shock, never clean pump or EnteraLite®Infinity™
AC Adapter/Charger with charger plugged into an outlet or pump turned on.
WARNING: Make sure the EnteraLite®Infinity™AC Adapter/Charger is com-
pletely dry before plugging into an electrical outlet.
To clean the EnteraLite®Infinity™Enteral Feeding Pump:
Pump may be cleaned with warm soapy water and
a nonabrasive sponge or soft cloth.
OR - Apply one of the following solutions for
approximately 10 minutes, then wipe pump clean
with a damp cloth or sponge:
5% bleach and water solution
Multipurpose disinfectant cleaner
You may also rinse the pump by holding it under
a stream of warm water. Then, dry with a clean cloth.
Use a cotton swab to gently clean pathways of cassette receptacle.
Compartments of the EnteraLite®Infinity™pack can be cleaned with a damp
sponge. The EnteraLite®Infinity™pack is machine-washable. Use warm water
and gentle cycle, whenever possible. Hang to dry.
EnteraLite®Infinity™AC Adapter/Charger normally does not require cleaning.
When desired, a dry or slightly damp cloth may be used to clean the outside
surface of AC Adapter/Charger while it is disconnected from the wall outlet.
Figure 4-1 Rinsing Pump
Under Stream of Water
25356_pages.qxd 2/2/2005 2:50 PM Page 19
Other manuals for EnteraLite Infinity
1
Table of contents
Other Zevex Medical Equipment manuals
Popular Medical Equipment manuals by other brands

3A HEALTH CARE
3A HEALTH CARE ASPEED 2 PROFESSIONAL instruction manual

Anjue
Anjue AJ-2402L operating instructions

OmniPod
OmniPod 5 quick start guide

IBIOM Instruments Ltd
IBIOM Instruments Ltd ERGO XR2 Operation manual

Autel
Autel MaxiTPMS TS501 manual

Dräger
Dräger VentStar Basic (P) 180 Instructions for use