ZOLL Propaq MD User manual

Propaq
®
MD Service Manual
9650-0801-01 Rev. BApril 2013

The issue date for the Propaq MD Service Manual (REF 9650-0801-01 Rev. B) is April, 2013.
If more than 3 years have elapsed since the issue date, contact ZOLL Medical Corporation to determine if
additional product information updates are available.
Copyright © 2013 ZOLL Medical Corporation. All rights reserved. Rectilinear Biphasic, SurePower, and
ZOLL are trademarks or registered trademarks of ZOLL Medical Corporation in the United States and/or other
countries.
Propaq is a registered trademark of Welch Allyn Inc.
Nellcor is a trademark of Covidien.
Oridion, Microstream, and FilterLine are registered trademarks of Oridion Systems, Ltd.
0123
ZOLL Medical Corporation
269 Mill Road
Chelmsford, MA USA
01824-4105
ZOLL International Holding B.V.
Newtonweg 18
6662 PV ELST
The Netherlands

9650-0801-01 Rev.B ZOLL Propaq MD Service Manual i
Table of Contents
Preface
Overview................................................................................................................................ 1
Safety Considerations............................................................................................................ 2
Additional Reference Material................................................................................................ 2
Conventions........................................................................................................................... 3
Service Policy Warranty.........................................................................................................3
Technical Service................................................................................................................... 3
Technical Service for International Customers ......................................................................4
Chapter 1 Maintenance Tests
Propaq MD Overview.............................................................................................................5
Physical Inspection of the Unit ............................................................................................. 8
Operational Tests ................................................................................................................. 9
Leads Test.......................................................................................................................... 15
Heart Rate Display Test ...................................................................................................... 16
Heart Rate Alarm Test ........................................................................................................ 17
Power Supply Test ..............................................................................................................19
Leakage Current Test ......................................................................................................... 21
Shock Test .......................................................................................................................... 22
Synchronized Cardioversion Test ....................................................................................... 26
Pacer Test .......................................................................................................................... 27
Printer Test ......................................................................................................................... 28
Treatment Report Test ........................................................................................................29
SpO2 Test ........................................................................................................................... 30
EtCO2 Calibration (required for annual calibration) ............................................................ 31
EtCO2 Calibration Check .................................................................................................... 32
EtCO2 Test ......................................................................................................................... 33
IBP Test .............................................................................................................................. 34
NIBP Calibration Check ...................................................................................................... 36
NIBP Leak Test ................................................................................................................... 39
Deflation Test ...................................................................................................................... 40
Inflation Test ....................................................................................................................... 41
NIBP Test ........................................................................................................................... 42
Temperature Test ...............................................................................................................43
Chapter 2 Troubleshooting
Overview.............................................................................................................................. 45
Propaq MD Error Messages ................................................................................................ 45

ii 9650-0801-01 Rev.B
Chapter 3 Disassembly Procedures
Overview.............................................................................................................................. 55
Required Equipment............................................................................................................ 55
Safety Precautions............................................................................................................... 56
Modules ............................................................................................................................... 56
Removing the Handle .......................................................................................................... 57
Removing the Printer Module.............................................................................................. 58
Disassembling the Main Unit ............................................................................................... 60
Removing the Front Enclosure Module ............................................................................... 63
Removing the Display Module............................................................................................. 65
Removing the EtCO2 and NIBP Modules............................................................................ 67
Removing the Monitor Board and SpO2 Modules ............................................................... 71
Removing the Defibrillator Board and Capacitor Bank Modules.......................................... 73
Chapter 4 Life Cycle Counters
Overview.............................................................................................................................. 77
Chapter 5 Replacement Parts
Overview.............................................................................................................................. 79
Replacement Parts ..............................................................................................................80
Chapter 6 Functional Description
Overview.............................................................................................................................. 83
Appendix A
Overview.............................................................................................................................. 89
Propaq MD Electrical Hardware Interconnect Diagram.......................................................90
Propaq MD Signal and Power Interconnect Schematic....................................................... 91
Appendix B
Overview.............................................................................................................................. 93

9650-0801-01 Rev B ZOLL Propaq MD Service Manual 1
Preface
Overview
ZOLL®Medical Corporation’s Propaq®MD Service Manual is intended for the trained
biomedical technician whose responsibility is to identify malfunctions and/or make repairs at
the subassembly level. The Propaq MD Service Manual has six main sections and two
appendixes.
Preface—Contains safety warnings and an overview of the manual’s contents. Be sure to
review this section thoroughly before attempting to use or service the Propaq MD unit.
Chapter 1—Maintenance Tests describes the Annual Calibration Tests for NIBP and EtCO2
modules, Functional Tests that need to be run when a module has been repaired, and
Operational Tests that need to be performed at regular intervals to augment the automatic
testing the Propaq MD does to ensure readiness.
Chapter 2—Troubleshooting provides a listing of error messages to help the service
technician detect faults and repair them.
Chapter 3—Disassembly Procedures describes step-by-step procedures for removing
subassemblies from the Propaq MD unit.
Chapter 4—Life Cycle Counters describes step-by-step procedures for setting life cycle
counters after replacing modules in the Propaq MD unit.
Chapter 5—Replacement Parts List displays a complete list of ZOLL part numbers for field
replaceable parts available for the Propaq MD unit, allowing the service person to identify and
order replacement parts from ZOLL.

PREFACE
29650-0801-01 Rev B
Chapter 6—Functional Description provides technical descriptions for the Propaq MD major
subassembly modules.
Appendix A—Propaq MD Electrical Hardware Interconnect Diagram and Propaq MD Signal
and Power Interconnect Schematic.
Appendix B—Contains simulators and settings that may be used to assess the performance of
the NIBP module.
Safety Considerations
The following section describes general warnings and safety considerations for operators and
patients. Service technicians should review the safety considerations prior to servicing any
equipment and read the manual carefully before attempting to disassemble the unit. Only
qualified personnel should service the Propaq MD unit.
Federal (U.S.A.) law restricts this unit for use by or on the order of a physician.
Safety and effectiveness data submitted by ZOLL Medical Corporation to the Food and Drug
Administration (FDA) under section 510(K) of the Medical Device Act to obtain approval to
market is based upon the use of ZOLL accessories such as disposable electrodes, patient cables
and batteries. The use of external pacing/defibrillation electrodes and adapter units from
sources other than ZOLL is not recommended. ZOLL makes no representations or warranties
regarding the performance or effectiveness of its products when used in conjunction with
pacing/defibrillation electrodes and adapter units from other sources. If unit failure is
attributable to pacing/defibrillation electrodes or adapter units not manufactured by ZOLL, this
may void ZOLL's warranty.
Only qualified personnel should disassemble the Propaq MD unit.
WARNING! This unit can generate up to 2775 volts with sufficient current to cause lethal shocks.
All persons near the equipment must be warned to “STAND CLEAR” prior to discharging the
defibrillator.
Do not discharge the unit’s internal defibrillator energy more than three times in one minute or
damage to the unit may result.
Do not discharge a battery pack except in a ZOLL SurePowerTM Charging Station.
Do not use the Propaq MD in the presence of flammable agents (such as gasoline), oxygen-rich
atmospheres, or flammable anesthetics. Using the unit near the site of a gasoline spill may
cause an explosion.
Do not use the unit near or within puddles of water.
Additional Reference Material
In addition to this guide, there is a Propaq MD Operator’s Guide which is a comprehensive
reference work that describes all the user tasks needed to operate the Propaq MD.

Conventions
9650-0801-01 Rev B ZOLL Propaq MD Service Manual 3
Conventions
WARNING! Warning statements describe conditions or actions that can result in personal injury or death.
Caution Caution statements describe conditions or actions that can result in damage to the unit.
Note: Notes contain additional information on using the defibrillator.
Service Policy Warranty
In North America: Consult your purchasing agreement for terms and conditions associated with
your warranty. Outside of North America, consult ZOLL authorized representative.
In order to maintain this warranty, the instructions and procedures contained in this manual
must be strictly followed. For additional information, please call the ZOLL Technical Service
Department 1-800-348-9011 in North America.
Technical Service
If the ZOLL Propaq MD unit requires service, contact the ZOLL Technical Service
Department:
Telephone: 1-978-421-9655; 1-800-348-9011
Fax 1-978-421-0010
Have the following information available for the Technical Service representative:
•Unit serial number.
•Description of the problem.
•Department where equipment is used.
•Sample chart recorder strips or electronic log files documenting the problem, if applicable.
•Purchase Order to allow tracking of loan equipment.
•Purchase Order for a unit with an expired warranty.
If the unit needs to be sent to ZOLL Medical Corporation, obtain a service request number
(SR#) from the Technical Service representative. Return the unit in its original container to:
ZOLL Medical Corporation
269 Mill Road
Chelmsford, Massachusetts 01824-4105
Attn: Technical Service Department (SR#)
Telephone: 1-800-348-9011; 1-978-421-9655 FAX: 978-421-0010

PREFACE
49650-0801-01 Rev B
Technical Service for International Customers
International customers should return the unit in its original container to the nearest authorized
ZOLL Medical Corporation Service Center. To locate an authorized service center, contact the
International Sales Department at ZOLL Medical at the above address.

9650-0801-01 Rev. B Propag MD Service Manual 5
Chapter 1
Maintenance Tests
Propaq MD Overview
This chapter includes step-by-step procedures for various maintenance tests. At a minimum, a
trained biomedical technician must perform a CO2calibration and an NIBP calibration check
once per year. A CO2calibration check may be performed at any time to verify the CO2
calibration. It is necessary to perform maintenance tests after a module is repaired to ensure that
the functions of the Propaq MD unit work properly. Use the checklist at the back of this
document (ZOLL Propaq MD Maintenance Tests Checklist) to record your results of the
maintenance tests.
Because the Propaq MD units must be maintained ready for immediate use, it is important for
users to conduct the Operator’s Shift Checklist procedure at the beginning of every shift. This
procedure can be completed in a few minutes and requires no additional test equipment. (See
the ZOLL Propaq MD Operator’s Guide for the Operator’s Shift Checklist.)

CHAPTER 1MAINTENANCE TESTS
69650-0801-01 Rev. B
This chapter describes the following maintenance tests:
•Physical Inspection of the Unit
•Operational Tests
•Leads Test
•Heart Rate Display Test
•Heart Rate Alarm Test
•Leakage Current Test
•Shock Test
•Advisory Test
•Synchronized Cardioversion Test
•Pacer Test
•Printer Test
•Treatment Report Test
•SpO2 Monitor Test
•EtCO2Calibration
•EtCO2Calibration Check
•EtCO2 Monitor Test
•IBP Monitor Test
•NIBP Calibration Test
•NIBP Leak Test
•Deflation Test
•Inflation Test
•NIBP Monitor Test
•Temperature Test
Before You Begin the Maintenance Tests
•Assemble the tools listed in following section, “Equipment You Need to Perform the
Maintenance Tests.”
•Keep an extra fully charged Propaq MD battery available.
•Schedule an hour to conduct the entire maintenance test.
•Perform the tests in the order presented.
•Perform all the steps of each test procedure.
•Complete all the steps of the procedure before evaluating the test results.

Propaq MD Overview
9650-0801-01 Rev. B Propag MD Service Manual 7
Equipment You Need to Perform the Maintenance Tests
The equipment listed below is used in the maintenance procedures in this chapter. You can
substitute an equivalent device; however, not all simulators and analyzers will produce the
same results. Be sure to follow the manufacturer’s recommendations for conducting the
maintenance tests.
•Fluke Impulse 7000DP Analyzer, with appropriate interface adapter (Fluke Biomedical
BP Pump/Dynatech Nevada Inc. part number 3010-0378)
•ECG simulator
•Fluke Index 2 SpO2simulator
•IBP/Temp simulator
•NIBP simulator
•5% CO2calibration gas cylinder with CO2sample line (kit available from Scott Medical
Products, PN 4653ORF-BD)
•CO2sampling line airway adapter
•Stop watch
•Battery Current Test Fixture 9100-3055-TF
•Agilent E3633A Power Supply
•Agilent 34401 Digital Multi-meter
Accessories You Need for the Propaq MD Options Maintenance Tests
•3-lead, 5-lead and 12-lead ECG cables. (12-lead cable needed if 12-lead option is
installed.)
•Dual lumen NIBP hose
•IBP cable
•SpO2 cable and sensor
•EtCO2 cable and sensor
•Paddles
•Printer paper
•Battery
•Auxiliary power source
•USB thumb drive

CHAPTER 1MAINTENANCE TESTS
89650-0801-01 Rev. B
1.0Physical Inspection of the Unit
Observe this... Pass/Fail
Housing
Is the unit clean and undamaged? oo
1.1 Does the unit show signs of excessive wear? oo
1.2 Does the handle work properly? oo
1.3 Does the recorder drawer open and close properly? oo
1.4 Are input connectors clean and undamaged? oo
1.5 Are there any cracks in the housing? oo
1.6 Does the front panel have any damage or cracks? oo
1.7 Are there any loose housing parts? oo
1.8 Paddles (if applicable)
Do the adult and pedi plates have major scratches or show signs of damage? o o o
1.9 Do the adult shoes slide on and off easily to expose the covered pedi plates? o o o
1.10 Are the paddles clean (e.g., free of gel) and undamaged? (if applicable) o o o
1.11 Cables
Are all cables free of cracks, cuts, exposed or broken wires? oo
1.12 Are all bend/strain reliefs undamaged and free of excessive cable wear? oo
1.13 Battery
Is the Propaq MD battery fully charged? oo
1.14 Is the battery seated in the battery well correctly? oo
1.15 Is battery housing free of cracks and damage? oo
1.16 Are connectors free of damage and excessive wear? oo

Propaq MD Overview
9650-0801-01 Rev. B Propag MD Service Manual 9
2.0Operational Tests
The Propaq MD unit comes equipped with routine operational tests that can be performed
periodically. With the exception of the Physical Inspection of the Unit test, these tests contain
instructions within them that provide guidance during the tests.
The operational tests on the unit can be accessed by doing the following:
1. Turn on the Propaq MD unit.
Note: If the unit is in AED mode, press the Manual quick access key to go into manual
mode. Use the navigation keys to select the four digits in the Manual Mode pass code.
Press SAVE when you are finished. If the unit has not been configured to enter a pass
code, the message Exit to Manual Mode is displayed. Use the navigation keys to select
Yes.
2. Press the More ( ) then the Setup ( ) quick access keys.
3. Use the navigation keys to select Supervisor>Service>Device Tests. The operational tests
are displayed. Use the navigation keys to select one of the tests.
Note: You need a supervisor passcode to enter the Supervisor menu.

CHAPTER 1MAINTENANCE TESTS
10 9650-0801-01 Rev. B
Keypad Test
This test checks all the front panel buttons on the unit to make sure they are working correctly.
Follow the instructions and use the navigation keys to move around the display and make
selections.
LED Test
This test checks all the Light Emitting Diodes in the unit to make sure they are working
correctly.
Follow the instructions and use the navigation keys to move around the display and make
selections.
Note: The speed of the flash varies by color.

Propaq MD Overview
9650-0801-01 Rev. B Propag MD Service Manual 11
LCD Test
This test checks the colors of the Liquid Crystal Display on the unit.
Follow the instructions and use the navigation keys to move around the display and make
selections.
RFU Indicator Test
This test checks the Ready For Use Indicator on the unit to make sure it is working correctly.
Follow the instructions shown on the display.

CHAPTER 1MAINTENANCE TESTS
12 9650-0801-01 Rev. B
Audio Test
This test checks the audio tones on the unit to make sure they are working correctly.
Follow the instructions and use the navigation keys to move around the display and make
selections.
Printer Test
This test checks the printer to make sure it is working correctly.
Follow the instructions shown on the display.

Propaq MD Overview
9650-0801-01 Rev. B Propag MD Service Manual 13
Battery Test
This test checks the battery to make sure it is working correctly.
Follow the instructions on the display.
Fan Test
This test checks the fan in the unit to make sure it is working correctly.
Follow the instructions and use the navigation keys to move around the display and make
selections.

CHAPTER 1MAINTENANCE TESTS
14 9650-0801-01 Rev. B
USB Test (Optional)
Note: Before starting this test, contact the ZOLL Technical Service Department to obtain a
test file, or create a file (see instructions below). Once you have obtained a copy, save
it on a USB thumb drive.
This test allows you to perform the USB test to make sure it is working correctly.
Follow the instructions and use the navigation keys to move around the display and make
selections.
When these tests are completed, use the navigation keys to exit out of the Service menu.
Creating a Test File
1. Create a Text file using Microsoft Notebook.
2. Open new file in Notebook.
3. Click in the first line of the file and type “Thunderbolt Testfile".
4. Save file as Tbolt_usb.txt.

Propaq MD Overview
9650-0801-01 Rev. B Propag MD Service Manual 15
3.0Leads Test
Equipment 3-lead, 5-lead, and 12-lead cables, Fluke Impulse 7000DP Analyzer
Note: Test each cable separately.
Test Setup Connect the lead wires appropriate for each test to the Fluke Impulse 7000DP Analyzer or
equivalent, then connect the ECG cable to the Propaq MD unit.
Do this... Observe this... Pass/Fail
3.1 Turn on the Propaq MD unit.
Note: Make sure that the unit is not
in PADS.
No ECG LEAD FAULT message
displayed. oo
3.2 Disconnect one lead from the
simulator. The ECG LEAD FAULT message displays
within 3 seconds (if configured). oo
3.3 Reconnect the lead. Repeat step 2
with the remaining leads. Wait for ECG LEAD FAULT message to
clear from the display (if configured). oo
3.4 Repeat steps 2 and 3 for 5 lead and 12
lead cables. Note: If heart rate alarm sounds, on the
Propaq MD unit:
Press the More quick access key ( ) key
to access the second set of quick access
keys.
Press the Alarm quick access key ().
Press the Alarm Cancel quick access key
().
oo

CHAPTER 1MAINTENANCE TESTS
16 9650-0801-01 Rev. B
4.0Heart Rate Display Test
Equipment Fluke Impulse 7000DP Analyzer
ECG Cable (3-lead, 5-lead or 12-lead)
Test Setup Press Lead quick access key until II displays.
Connect the ECG leads to the ECG studs on Analyzer.
Do this... Observe this... Pass/Fail
4.1 Set the ECG Simulator to 120BPM. The Heart Rate displays as 120 +/- 4 bpm. oo
Other manuals for Propaq MD
5
Table of contents
Other ZOLL Health Care Product manuals
Popular Health Care Product manuals by other brands

Algostim
Algostim Algovita 4300 user manual

ResMed
ResMed AIRCurve 10 Vauto user guide

Medela
Medela Supplemental Nursing System Instructions for use

BodyMedia
BodyMedia SenseWear operating manual

Seating Matters
Seating Matters Sorrento Operation manual & product specification

Unicare Health
Unicare Health SALSA 200 Instruction & safety manual