A.R.C. Laser CLASSIC 514 User manual

User Manual
A.R.C. Laser GmbH
0123
Bessemerstr. 14
D-90411 Nürnberg
Tel.: +49 911 21779-0
Fax: +49 911 21779-99
www.arclaser.de

User Manual CLASSIC 514
August 2023
File: User Manual_CLASSIC 514_rev5.docx 2
Please contact for service:
A.R.C. Laser GmbH Headquarters:
A.R.C. Laser GmbH
Bessemerstr. 14
D-90411 Nürnberg
Tel.: +49 911 21779-0
Fax: +49 911 21779-99
service@arclaser.de
Copyright © A.R.C. Laser GmbH. All rights reserved.
This document contains confidential and copyrighted information and may only be used
with the devices described herein. Reproduction is prohibited without written permission
from A.R.C. Laser GmbH.
Local distributor

User Manual CLASSIC 514
August 2023
File: User Manual_CLASSIC 514_rev5.docx 3
1Introduction......................................................................................................................................5
1.1 Marking and symbols................................................................................................................5
1.2 Intended use .............................................................................................................................6
1.3 General contraindications .........................................................................................................6
1.4 Theory and technical set-up......................................................................................................7
2Transport and storage.....................................................................................................................8
2.1 Shipping and unpacking the device ..........................................................................................8
2.2 Return shipment........................................................................................................................9
3Set-up and installation ..................................................................................................................10
3.1 Installation site ........................................................................................................................10
3.2 Room requirement ..................................................................................................................10
3.2.1 Marking access points ........................................................................................................10
3.2.2 Window shielding................................................................................................................10
3.2.3 Reflecting surfaces .............................................................................................................11
3.3 Electrical connection...............................................................................................................11
4Safety information .........................................................................................................................12
4.1 General....................................................................................................................................12
4.2 Eye safety................................................................................................................................13
4.3 Electrical protection.................................................................................................................13
4.4 Explosion and fire hazard........................................................................................................14
4.5 Protection against undesired radiation....................................................................................14
4.6 NOHD safety distance.............................................................................................................14
4.7 CE-Regulations.......................................................................................................................14
4.8 RoHS3-Regulations ................................................................................................................14
4.9 Protective housing...................................................................................................................15
4.10 Connectors and switches........................................................................................................15
4.11 External interlock switch..........................................................................................................16
4.12 Safety-Shutter / Aiming beam.................................................................................................16
4.13 Manual reset ...........................................................................................................................16
4.14 Reset by power failure ............................................................................................................16
4.15 Lables and markings...............................................................................................................17
4.16 Operating conditions ...............................................................................................................18
4.17 Electromagnetic compatibility..................................................................................................18
5Advice for users.............................................................................................................................20
5.1 Technical instruction ...............................................................................................................20
5.2 Laser-Safety-Training..............................................................................................................20
5.3 Medical instruction ..................................................................................................................20
5.4 Device parts and accessories.................................................................................................20
5.4.1 Instrument table..................................................................................................................21
5.4.2 Foot Switch (optional).........................................................................................................21
5.4.3 Slit lamp adaption of the laser ............................................................................................22
5.4.4 Combinations......................................................................................................................23
6Operation........................................................................................................................................24
6.1 Preparation..............................................................................................................................26
6.2 Starting the laser.....................................................................................................................26
6.3 Programs and settings ............................................................................................................27
6.3.1 Main menu..........................................................................................................................27
6.3.2 Energy settings...................................................................................................................28
6.3.3 Pulse length settings...........................................................................................................28
6.3.4 Frequency/ pulse pause settings........................................................................................28
6.3.5 APL Mode (All Pulse Laser) (optionally).............................................................................29
6.3.6 Pulse counter......................................................................................................................29
6.4 Other settings..........................................................................................................................30
6.4.1 Password............................................................................................................................31
6.4.2 Info menu............................................................................................................................31
6.5 Laser STANDBY / READY......................................................................................................31
6.6 Release laser radiation ...........................................................................................................32
6.7 Treatment................................................................................................................................32

User Manual CLASSIC 514
August 2023
File: User Manual_CLASSIC 514_rev5.docx 4
6.8 Laser-Stop...............................................................................................................................33
6.9 Switch Off................................................................................................................................33
6.10 Symbols...................................................................................................................................34
7Accesory L-ORD ............................................................................................................................35
7.1 Transport and Storage ............................................................................................................35
7.2 CE Regulations .......................................................................................................................35
7.3 Accessories.............................................................................................................................35
7.4 Stickers and Labels.................................................................................................................35
7.5 Operating conditions ...............................................................................................................36
7.6 Installation...............................................................................................................................37
7.7 Operation.................................................................................................................................39
7.8 The Treatment.........................................................................................................................40
7.9 Maintenance and Cleaning .....................................................................................................40
8Technical data................................................................................................................................41
8.1 CLASSIC 514..........................................................................................................................41
8.1.1 General...............................................................................................................................41
8.1.2 Laser Data ..........................................................................................................................41
8.1.3 Electrical Connection Data .................................................................................................41
8.1.4 Classification.......................................................................................................................41
8.2 L-ORD.....................................................................................................................................41
8.2.1 General...............................................................................................................................41
8.2.1 Laser-Data to be used with the L-ORD ..............................................................................41
8.2.2 Classification.......................................................................................................................41
9Maintenance ...................................................................................................................................42
9.1 Introduction..............................................................................................................................42
9.2 Safety Check (STK) ................................................................................................................42
9.3 Care by the user......................................................................................................................42
9.4 Cleaning of Accesories ...........................................................................................................44
9.4.1 Foot switch..........................................................................................................................44
9.4.2 Laser safety goggles...........................................................................................................44
9.4.3 Slit Lamp.............................................................................................................................44
9.4.3.1 Slit Lamp mirror ..............................................................................................................44
9.4.3.2 Dust Cover......................................................................................................................44
9.4.3.3 Testing the Slit Lamp Focus...........................................................................................44
9.5 Error Detection........................................................................................................................45
9.6 Disposal...................................................................................................................................46
10 Customer service...........................................................................................................................47
10.1 Warranty Information...............................................................................................................47
10.2 Warranty, Shipment, Packing .................................................................................................47
10.3 Sales and Service Informations ..............................................................................................47
11 Guidelines and manufacturers declaration.................................................................................48
11.1 Electromagnetic Emissions.....................................................................................................48
11.2 Electromagnetic Immunity (1) .................................................................................................49
11.3 Electromagnetic Immunity (2) .................................................................................................51
11.4 Recommended separation distances between portable and mobile RF telecommunications
equipment and the laser.......................................................................................................................52

User Manual CLASSIC 514
August 2023
File: User Manual_CLASSIC 514_rev5.docx 5
1 Introduction
We thank you for having chosen our CLASSIC 514 and wish you every success in using
this versatile laser system.
The CLASSIC 514 laser generates a laser beam of high intensity, which can cause
injuries if handled improperly. The user manual should therefore be read carefully before
starting up the device. If you have any further questions regarding the safety or the use
of the device or regarding the laser and laser radiation, please contact A.R.C. Laser
GmbH.
1.1 Marking and symbols
The symbol "Attention" is attached to all surfaces that mean danger to
the user. Before carrying out any further work on such marked parts, the
operating instructions should be consulted or the service of A.R.C. Laser
GmbH should be contacted.
The symbol "Attention laser radiation" is intended to warn the user
against directing laser beams to unwanted areas through improper
handling or improper operation of the device.

User Manual CLASSIC 514
August 2023
File: User Manual_CLASSIC 514_rev5.docx 6
1.2 Intended use
The CLASSIC 514 is intended to be used to treat diseases of the eye. In particular
diseases of the retina, which can be treated with coagulation and diseases in the
anterior segment, which can also be treated by coagulation, or partial evaporation of
tissue are indicated. These include most retinal diseases, glaucoma using iridotomy and
trabeculoplasty.
For this purpose, the laser is used together with a slit lamp. Here, the laser radiation is
directed via an imaging optics and mirrors on the target tissue.
Medical
Applications
Indications
Side and adverse effects
Contraindications
Ophthalmology
Retinal diseases
requiring
coagulation or
subthreshold
tissue alteration
which does not
cause
coagulation
•Infection
•inflammation, iritis (from
iridotomy or
trabeculoplasty)
•posterior synechiae
(from iridotomy)
•anterior synechiae (from
trabeculoplasty)
•bleeding
•not persistent IOP
increase after treatment
•macular edema
(thickening), choroidal
effusion, exudative
retinal detachment,
chorioretinal effusion,
choroidal neovascular
membranes, subretinal
fibrosis, visual field loss,
vitreous haemorrhage,
choroidal detachment,
keratopathy, cataract
•intransparent or
partly intransparent
or scattering media
•inflammation
(intraocular or at/
around the eye)
Glaucoma
treatable by
tissue
vaporization or
coagulation
1.3 General contraindications
Other intraocular or other medical use is not intended.
A T T E N T I O N
The device may only be operated by specially trained personnel who are experts in the
medical effects and possible dangers of the device. You must have the necessary skills
to use the laser in accordance with this instruction manual.
When not in use, the device should be protected against unauthorized use.

User Manual CLASSIC 514
August 2023
File: User Manual_CLASSIC 514_rev5.docx 7
1.4 Theory and technical set-up
This laser device has a laser diode as the beam source which emits a wavelength of
514 nm.
The electrons of the atoms may have different energy states, which can generate
electromagnetic phosound radiation changing from one state to the next by means of
absorption and consequently by radiation. In their normal state, the electrons of an atom
are in the so-called basic or normal state. This case is called cast inversion.
By applying a low voltage and a high power a diode laser is put into the state of the cast
inversion to create and emit light. Forward electrical bias across the laser diode causes
the emission by means of two species of charge carrier –holes and electrons –to be
"injected" from opposite sides of the p-n junction into the depletion region. When an
electron and a hole are present in the same region, they may recombine or "annihilate"
producing an ongoing emission, emitting further photons. The avalanche effect of the
stimulated emission arises.
To improve this effect the laser is equipped with a resonator –in our case the photon-
emitting semiconductor itself. The plan parallel properties of the crystal with its high
index of refraction take over the function of a resonator mirror. The end face of the
semiconductor is covered with a highly reflecting mirror to force the photons to be
reflected several times from each end face before they exit.
Through special manufacturing processes of the semiconductor material, A.R.C. Laser is
able to provide this laser system in its specific wavelength property. The systems
respective maximum output power varies in height and is set by the user via the display -
just like the pulse length and pulse pause.
In general laser diodes are extremely sensitive towards voltage fluctuations. An internal
microchip controller cushions and controls the high voltage to run the diode.
Via beam deflection mirrors and coupling ceramic elements the laser light is coupled into
a silica glass fiber. The transmission of the various wavelengths is independent from the
fiber. Transmissions ranging from 80 to 90 % of the laser power are possible. The laser
source is cooled directly to maintain the mandatory operating. The cooling circuit is
passive and controlled by fans.

User Manual CLASSIC 514
August 2023
File: User Manual_CLASSIC 514_rev5.docx 8
2 Transport and storage
We at A.R.C. Laser GmbH will make sure that the device is packed and transported with
the greatest possible care.
Before unpacking the laser, please check the packaging for damage and report any
damage immediately to the shipping agent and A.R.C. Laser GmbH. Only remove the
packaging in the presence of a representative of the carrier. Make a list of the damaged
parts and have this list signed by the courier.
The device must be transported at temperatures between -10° C and +60° C. The air
pressure during transport must be between 1080 hPa and 750 hPa. During storage, a
temperature range of +5° C to +40° C must be maintained. The environment/air must be
dry and clean. The relative humidity during transport and storage must be between 5 %
and 85 %.
A T T E N T I O N
If the laser is transported or stored at a temperature below +5° C, it can be damaged
when starting. Unpack the laser and leave it for at least four hours at normal room
temperature so that the system reaches room temperature.
2.1 Shipping and unpacking the device
Unpacking and installation of the device must be carried out by an authorized technician
or authorized representative of A.R.C. Laser GmbH. After the installation and the correct
connection of the device, the technician will put the device into operation and explain the
function of the device. All functions and safety procedures are discussed and
demonstrated here.

User Manual CLASSIC 514
August 2023
File: User Manual_CLASSIC 514_rev5.docx 9
2.2 Return shipment
The device will be shipped in a specially developed transport packaging. Please keep
this. If you ship the device back to A.R.C. Laser GmbH, be sure to use the intended
transport packaging. Normally, the CLASSIC is delivered together with the slit lamp. For
transportation the device is packed as follows:
The transport packaging is designed so
that the device fits snugly into the foam
parts.
All accessories are stowed in an extra padding inside the
transport packaging.
If all foam parts are in the right place and the box is
closed, the device is ready for shipping.

User Manual CLASSIC 514
August 2023
File: User Manual_CLASSIC 514_rev5.docx 10
3 Set-up and installation
3.1 Installation site
Before the device is delivered, it must be ensured that the laser can be set up in a
suitable location.
The CLASSIC 514 should be operated in an easily accessible place. The laser should
not be operated near a heater, since the device only works when the ambient
temperature is not higher than 35° C. Higher temperatures can cause the device to
switch off due to overheating. A place of installation in direct sunlight can cause an early
shutdown and must be avoided in any case.
If the temperature is too low (below +18° C), the device cannot be started to prevent
condensation on the internal optics; this could result in permanent damage to the laser.
•The device should be set up so that the laser beam is not aimed directly at a door,
window or reflective material.
•The device should be operated in a dust-free room. There should be no carpets on
the floor or the walls.
•When the device is not in use, it should be covered to prevent contamination. For
this purpose, a suitable dust cover is included.
The wiring must be installed so that there are no tripping hazards or other hazards
All control elements must also be free and easily accessible. The air humidity is
monitored internally and must be below 75 %.
3.2 Room requirement
The legislation imposes the following requirements to any room in which a class 4 laser
according to IEC 60825-1 is operated.
3.2.1 Marking access points
All entrances must be clearly marked so that unintentional or unauthorized entry, which
can lead to a hazard, is avoided.
•Laser warning signs (triangle with laser symbol) and the wavelength labeling must
be attached to each access door.
•A warning lamp must be attached above each access door. This must always light
up when the laser is in operation.
Unintentional entry into the room without safety glasses is avoided.
•The laser safety goggles must be easily accessible at the entrance.
3.2.2 Window shielding
It must be ensured that no laser radiation can escape the room. In particular, the
windows must be secured with suitable materials. If you have any questions, your
contact person at A.R.C. Laser GmbH is available to assist you.

User Manual CLASSIC 514
August 2023
File: User Manual_CLASSIC 514_rev5.docx 11
3.2.3 Reflecting surfaces
To avoid danger from reflected direct and scattered radiation, there must be no reflective
surfaces in the room.
These can be:
•Mirrors
•Pictures behind glass
•Chrome surfaces
•Windows
These surfaces must be removed or suspended or matted. In the area of laser use, only
matted, non-reflective and non-flammable instruments and materials should be used.
3.3 Electrical connection
The laser is operated with a DC voltage of 24 V. The power is supplied by an external
power supply. The power supply unit can be connected to an AC voltage ranging from
from 100 V –240 V (50/ 60 Hz).
The power connection is made via the table column and the supplied connector.
Make sure that the plug is accessible at any time so that the laser can be disconnected
from the mains after use.

User Manual CLASSIC 514
August 2023
File: User Manual_CLASSIC 514_rev5.docx 12
4 Safety information
4.1 General
The CLASSIC 514 laser is a precision instrument for medical applications. The system
has been carefully developed and tested through intensive testing before shipping. In
order to offer you and the operating personnel all possible protection, we recommend
that you read this section of the operating instructions carefully.
The CLASSIC 514 laser falls under laser class 4 according to EN 60601-2-22, EN
60825-1 respectively.
Class 4 in the standard describes high-energy lasers and therefore special measures
have to be considered befor use in order to ensure safe and trouble-free working with
the device. It is particularly important to protect the eyes and skin of the operator, the
patient and third parties. Laser safety glasses must be used for eye protection.
The following explanations are not exhaustive. All users of laser devices should enclose
applicable legal regulations and provisions with the device and inform the staff
accordingly.
If the device is operated outside Germany, provisions of the American National Standard
Office ANSI Z136.3-2018 "American National Standard for the Safe Use of Lasers in
Health Care Facilities" and ANSI Z136.1-2014 "American National Standard for the Use
of Lasers” must be respected.
This manual is limited to the operation, maintenance and control of the device. The
manual is not a guide for the treatment of diseases that can be treated with the laser.
With regard to the devices supplied, such as slit lamps or instrument tables, the safety,
operating and maintenance instructions in the corresponding manuals must be
observed. With regard to the instrument table, reference is only made here to the danger
from improper use as a seat or storage area. When operating the height adjustment,
make sure that no one can be harmed.
A.R.C. Laser GmbH cannot be held responsible for damage or damage resulting from
improper use.
The warranty for the device expires if the laser has been opened (even partially),
modified or repaired by unqualified personnel.

User Manual CLASSIC 514
August 2023
File: User Manual_CLASSIC 514_rev5.docx 13
4.2 Eye safety
As a safety measure against direct or indirect laser radiation, it is necessary that
everyone in the room wear laser safety glasses. The treating doctor is protected from
radiation by the laser protection integrated (eye safety filter) into the slit lamp. Eye
protection appropriate to the patient must be guaranteed for the patient.
When using the laser, only those laser safety glasses that are designed for the
wavelength of 514 nm and on which both the CE mark and the appropriate protection
level are noted.
The following laser safety glasses from A.R.C. Laser GmbH are suitable for the
CLASSIC 514:
•AS01003 for 514 nm: Protection level D LB6+ I LB 6 + RM LB8 / Optical density
OD8+
There are different types of laser safety glasses that also allow spectacle wearers to
wear a tight goggle on top of their own which are protective to all sides. In particular, the
scattered radiation that does not directly come from the front of the eye may present
risks resulting from internal reflections on the glasses. Therefore, we encourage you to
wear laser safety glasses which also guarantee a full protection to the side parts.
The eye safety of the operator is ensured by the
eye safety filter which is firmly installed into the slit
lamp. The eye safety filter is already included and
may not be removed at any time. If the CLASSIC
514 is used with the laser adaptor L-ORD (chapter
7), the eye safety filter is included. Therefore, no
further safety measures are necessary for the
operator.
A T T E N T I O N
Never look directly at the laser beam or the light reflected by the laser beam. This will
cause serious damage to your eye.
4.3 Electrical protection
Never remove the housing parts of the laser. Work on the device may only be carried out
by authorized A.R.C. Laser GmbH service personnel. The mains may only be replaced
by authorized service personnel.
The room in which the laser is operated should be kept dry. In the event that cleaning is
necessary, make sure that the floor is dry before starting up the device.
A T T E N T I O N
Never use the laser if you notice any visible defect on the device.
Never use the device if you notice any visible defect on the power plug, wires or supply
lines are exposed.

User Manual CLASSIC 514
August 2023
File: User Manual_CLASSIC 514_rev5.docx 14
4.4 Explosion and fire hazard
Never work with the laser near flammable anesthetics, flammable solutions or other
flammable materials. Flammable plastic parts or paper parts in particular should also be
removed from the working area of the laser. There is a fire or explosion hazard if the
laser is used in the presence of flammable materials, solutions, or gases or an oxygen-
enriched environment.
4.5 Protection against undesired radiation
If a footswitch is used, it should never be outside the area of the attending doctor. It is
forbidden for anyone other than the attending doctor to trigger the footswitch.
Especially in operating theaters where there are multiple footswitches, it is important to
ensure that the laser footswitch is close to the laser device.
Also, the hand trigger of the slit lamp may only be used by the attending doctor.
4.6 NOHD safety distance
The NOHD (Nominal Ocular Hazard Distance) is the distance at which the irradiance is
equal to the exposure limit value of the cornea of the eye. The NOHD identifies the
danger area within which there is a risk of damage to the health of the eye if you look
directly into the laser beam without protection.
The NOHD is calculated according to EN 60825-1 and taking into account the permitted
power fluctuations (+/- 20%) according to EN 60601-2-22 using the following equation:
𝑁𝑂𝐻𝐷 =√4𝑃
𝑀𝑍𝐵 ∗ 𝜋 –𝐷𝑖𝑎𝑚𝑒𝑡𝑒𝑟 𝐿𝑎𝑠𝑒𝑟 𝐵𝑒𝑎𝑚
𝐵𝑒𝑎𝑚𝑠 𝑑𝑖𝑣𝑒𝑟𝑔𝑒𝑛𝑐𝑒
For the CLASSIC 514 with slit lamp the following applies:
Wavelength λ: 514 nm
NOHD: 12.2 m
Beam divergence α(full angle): 1.1°
Permitted Maximum Radiation: 1.9*10-³ Jm-2
4.7 CE-Regulations
The laser system CLASSIC 514 has been approved by the notified body in accordance
with the European Directive 93/42/EEC for medical devices. Accordingly, the device
bears the CE mark CE 0123.
The device has been checked for electrical and mechanical safety. All parts we use
comply with the CE regulations or have been tested for approval or suitability by the
notified body.
Additional devices that you attach to the device require approval from an official test
center. Changes to the device or interventions on your part will void the approval and
warranty. The test approval number is included with the device.
4.8 RoHS3-Regulations

User Manual CLASSIC 514
August 2023
File: User Manual_CLASSIC 514_rev5.docx 15
Our company operates worldwide and regards the protection of the environment and
natural resources as an entrepreneurial obligation. Based on individual tests, A.R.C.
Laser GmbH confirms that, to the best of our knowledge, our products do not contain
any substances in concentrations whose placing on the market is prohibited according to
the applicable requirements of Directive 2015/863/EU (RoHS 3).
4.9 Protective housing
The CLASSIC 514 laser has a protective housing. It prevents radiation from the laser
from escaping and protects users from touching live parts. This housing must not be
removed.
4.10 Connectors and switches
At the rear panel of the laser following connector can be found:
On- /Off switch
Service port
Warning light
Eye safety filter
Interlock switch
Foot switch
Power supply

User Manual CLASSIC 514
August 2023
File: User Manual_CLASSIC 514_rev5.docx 16
It is prohibited to use the device’s socket for other usages than those detailed here.
Connector
Description
Usage
Color
Service Port
Service port, do not use
The service port may only be
used by trained A.R.C. Laser
service personal
Green
Warning light
Connection for the external
warning light of the door
Only use the A.R.C. Laser
GmbH connection for the
CLASSIC 514
Orange
Eye safety filter
Connector for the mechnical
eye safety filter
Not in use anymore, please
insert the dummy plug.
Black
Interlock
switchr
Interlock switch for the door
inter lock or fort he L-ORD
Der interlock switch can
either be used for the door
interlock or when conecting
the L-ORD with the CLASSIC
514
Red
Foot switch
Connector for the A.R.C.
Laser GmbH footswitch
Only use the A.R.C. Laser
GmbH foot switch for the
CLASSIC 514
Blue
Power supply
Connector for the medical
certified A.R.C. Laser power
supply
Only use the A.R.C. Laser
GmbH power supply for the
CLASSIC 514
Yellow
4.11 External interlock switch
A door interlock switch is required by the accident prevention regulations. The device is
equipped with an interlock connector as standard, which can be replaced by a door
interlock switch. The laser switches off when the door is opened. In addition, the laser
cannot be switched to the READY mode when the door is open. An error message
appears on the control panel. When the door is closed, the error message disappears
and the laser can be switched to READY again.
The door interlock plug should be inserted instead of the dummy plug. Only connect
plugs with 12V and 20mA.
Make sure the switch stucks firmly to prevent unexpected system interlock problems.
4.12 Safety-Shutter / Aiming beam
The CLASSIC 514 has an internal safety shutter. This shutter opens by pressing the
READY button and allows the user to release laser radiation.
Thereafter, the aiming beam is only visible when the laser is in the READY mode. This
aiming beam is a laser beam with very low power.
4.13 Manual reset
If an error occurs, the system switches into STANDBY mode. In that case, you should
switch the laser off and on again. Wait for 20 seconds when switching the device on
again otherwise the error F00f is shown and you have to restart again. The restart
should fix the error due to the automatic recalibration.
If the error still occurs, this can only be fixed by trained personnel. Please contact the
service of A.R.C. Laser GmbH in this case.
4.14 Reset by power failure
If the device is disconnected from the mains accidentally - e.g. due to a power failure - it

User Manual CLASSIC 514
August 2023
File: User Manual_CLASSIC 514_rev5.docx 17
usually restarts automatically. The system recalibrates. If there is still an error, it can only
be corrected by trained personnel. Please contact A.R.C. Laser GmbH service
department should this occur.
4.15 Lables and markings
The CLASSIC 514 is provided with various warnings in accordance with European and
worldwide guidelines to prevent the laser user from being exposed to laser radiation by
negligence. The following labels can be found on the CLASSIC 514:
Type plate
1) Manufacturer
2) Laser name
3) Wavelength and power
4) Power input
5) Serial number (SN)
6) Productions date
7) Unique Device Identifier (UDI = Date
+ SN + GTIN)
8) Global Trade Item Number (GTIN)
9) Do not dispose with household waste
10) Applied part type BF
11) Follow user manual
12) CE sign
Further Labels
The modification label shows the current state oft he
device.

User Manual CLASSIC 514
August 2023
File: User Manual_CLASSIC 514_rev5.docx 18
Laser warning labels
(attached on the side)
The device is marked with an additional label, which indicate to respect the user manual.
4.16 Operating conditions
The medical laser CLASSIC 514 is not suitable for use in connection with combustible
gas mixtures of all kinds.
The device is not approved for operation at altitudes above 2,000 m above sea level.
and only for an air pressure between 1080 hPa and 750 hPa.
The following environmental conditions must be met:
•Ambient temperature:
18° C to 35° C
•Relative humidity: <75 %
•
4.17 Electromagnetic compatibility
The CLASSIC 514 laser system meets the EMC requirements according to EN 60601-1-
2. Guidelines and the manufacturer's declaration are described in Chapter 11.
A T T E N T I O N
This device should not be operated immediately adjacent to or stacked with other
devices.
This device should not be operated immediately adjacent to portable or mobile wireless
communication devices or stacked with portable or mobile wireless communication
devices.
Laser stopp label and marking of the fiber exit
(attached on the device)

User Manual CLASSIC 514
August 2023
File: User Manual_CLASSIC 514_rev5.docx 19
For recommended separation distances from portable or mobile wireless
communication devices, please refer to Chap. 18.3 (ELECTROMAGNETIC
PROTECTION - II) and 18.4 (RECOMMENDED PROTECTION DISTANCES).
RFID systems shall not be used at separation distances less than 0.15 m from the
device. Otherwise, degradation of the performance of this device could result.

User Manual CLASSIC 514
August 2023
File: User Manual_CLASSIC 514_rev5.docx 20
5 Advice for users
5.1 Technical instruction
During the installation of the device, instruction is given by an A.R.C. Laser GmbH
employee or an authorized representative.
This first instruction essentially relates to the technical use of the device. In addition,
essential security-relevant points are dealt with. During the briefing, all persons working
in the vicinity of the laser should be present.
5.2 Laser-Safety-Training
The CLASSIC 514 laser is designed for medical users. It may only be used by personnel
who have been instructed in its operation. The A.R.C. Laser GmbH recommends, in
addition to the briefing, participation in seminars in which working with different laser
systems is dealt with. In addition, instructions are given on laser safety and the use of
lasers in general. It has also proven useful that people who do not work directly with the
laser attend courses on laser safety.
Training for the accompanying staff is additionally accompanied by an instruction from
an A.R.C. Laser GmbH employee or by an authorized representative when installing the
device. During instruction, the use of laser safety glasses and laser safety will be
specifically addressed.
A.R.C. Laser GmbH has a list of recommended courses as well as laser safety courses.
These can be obtained from us at any time.
5.3 Medical instruction
In the context of device instruction, only the general medical application is addressed.
The A.R.C. Laser GmbH only gives recommendations for applications.
If necessary, there is the possibility to take part in a training course with an experienced
doctor. Please contact your responsible administrator or A.R.C. Laser GmbH directly.
5.4 Device parts and accessories
The basic parts of the CLASSIC 514 include:
Part
Description
Article number
Power cord
EU power cord (type E/F)
UK power cord (type G)
US power cord (type B)
KB13003
KB13004
KB13005
Power supply
Power supply
PS01010
Interlock plug
Interlock plug for the door interlock
UG04022
Eye safety filter dummy
plug
Dummy plug for the eye safety filter
connection
BG01168
User manual
User manual
Transport box
Transport box for the CLASSIC 514
VP04002
Optionally, the CLASSIC 514 can be used with a foot switch instead of the hand trigger
Table of contents
Other A.R.C. Laser Measuring Instrument manuals
Popular Measuring Instrument manuals by other brands

Bio-Therapeutic
Bio-Therapeutic bt-analyze Software Quick Start Guide

Urrea
Urrea SURTEK 122230 User manual and warranty

Endress+Hauser
Endress+Hauser RID14 Brief operating instructions

ATAGO
ATAGO PAL-Easy ACID7 instruction manual

Wavetek
Wavetek Meterman CT-238 Operator's manual

Diehl
Diehl AQUARIUS S installation guide