A&D AD-8923-CC User manual

AD-8923-CC
Remote Controller
(CC-Link)
INSTRUCTION MANUAL
1WMPD4002124

© 2010 A&D Company, Limited. All rights reserved.
No part of this publication may be reproduced, transmitted, transcribed, or translated
into any language in any form by any means without the written permission of A&D
Company, Limited.
The contents of this manual and the specifications of the instrument covered by this
manual are subject to change for improvement without notice.
.

1
CONTENTS
1. INTRODUCTION ...........................................................................................................2
1.1. Features ..................................................................................................................2
2. DESCRIPTION OF EACH PART ..................................................................................3
2.1. Display ....................................................................................................................4
2.2. Keys ........................................................................................................................4
2.3. Connectors..............................................................................................................4
3. CONNECTION ..............................................................................................................5
3.1. Setting the AD-4212C and the AD-8923-CC ...........................................................5
3.2. Connecting the cables.............................................................................................5
3.3. Turning the power on...............................................................................................6
3.4. Operation ................................................................................................................9
3.5. Calibrating the AD-4212C........................................................................................9
4. FUNCTION SETTING ................................................................................................. 11
4.1. Display and keys ...................................................................................................12
4.2. Function table........................................................................................................12
4.3. Initializing the AD-8923-CC ...................................................................................13
5. RS-232C CONNECTOR..............................................................................................14
5.1. RS-232C serial interface specifications.................................................................14
6. CC-Link CONNECTOR...............................................................................................15
6.1. CC-Link interface specifications ............................................................................15
6.2. Timing chart...........................................................................................................20
6.3. Fixing the decimal point position ...........................................................................22
7. TROUBLESHOOTING ................................................................................................23
8. SPECIFICATIONS.......................................................................................................24
9. EXTERNAL DIMENSIONS..........................................................................................24

2
1. INTRODUCTION
This manual describes how the AD-8923-CC remote controller works and how
to get the most out of it in terms of performance.
Read this manual thoroughly before using the AD-8923-CC and keep it at
hand for future reference.
1.1. Features
Connecting the AD-8923-CC remote controller and the AD-4212C series
production weighing unit will enable transmission of the weight data to a
PLC using CC-Link.
zDisplays the weight data transmitted from the AD-4212C.
zCan change the weighing speed of the AD-4212C, calibrate the
AD-4212C using an external calibration weight and share the power
supply with the AD-4212C.
Note
When the AD-4212C is used as the weighing instrument,
connecting power to either the AD-8923-CC or the AD-4212C
will supply power to both devices. Refer to “3.3. Turning the
power on.”
zUsing the CC-Link interface, one CC-Link master station can control up to
42 units consisting of an AD-8923-CC and AD-4212C, to receive data or
perform re-zeroing.
Note
CC-Link is a high-speed field network able to simultaneously
handle both control and information data. With a high
communication speed of 10 Mbps, CC-Link can achieve the
maximum transmission distance of 100 meters and connect
to 64 stations.
When a CC-Link network is configured using the AD-8923-CC,
the maximum number of stations (or units) will be 42.

3
2. DESCRIPTION OF EACH PART
See the note
below.
Note
When the AD-8923-CC is to be
built into a weighing system, be
sure to ground the FG terminal.
Please confirm that the AC
adapter type is correct for your
local voltage and receptacle type.
CC-Link plug 1 pc.
C
onnector operation lever 1 pc.
AC adapter 1 pc.
Accessories
Display
A
C adapte
r
input jack
Rear View
Keys
DC in
p
ut terminal
CC-Link connecto
r
RS-232C connecto
r
(D-Sub 9-pin)
FG terminal
A
D adapte
r
ID label

4
2.1. Display
zDisplays the weight data received. When the unit is “g” (gram), the unit
indicator turns on.
zWhen the weight value is stable (the header of the weight data received is
“ST”), the STABLE indicator turns on.
zIf the AD-8923-CC does not receive the weight data for two seconds or
more, - - - - - is displayed (Bar display).
zDisplays the AD-4212C weighing speed that is currently set, by turning on
the corresponding indicator.
2.2. Keys
zOperates the AD-4212C. For details, refer to “3.4. Operation”.
zTo enter the function setting of the AD-8923-CC, press the CAL key while
holding down the ON:OFF key. For details, refer to “4. FUNCTION
SETTING”.)
2.3. Connectors
zRS-232C connector ...... D-Sub 9-pin (male)
Used to connect to the AD-4212C weighing unit. For details, refer to “5.
RS-232C CONNECTOR”.
zCC-Link connector ........ 5-pin (male)
Used to connect to other AD-8923-CCs, PLCs or other CC-Link devices.
For details, refer to “6. CC-Link CONNECTOR”.
zDC input terminal (24 DCV) / AC adapter input jack
Either power supply can be used. For details, refer to “3.3. Turing the
power on”.
Weighing
speed
indicators
Turns on while the AD-4212C
is in the standby mode.
STABLE indicato
r
Unit indicato
r
AD-8923
REMOTE CONTROLLER
ON:OFF
CAL
SAMPLE PRINT
MODE
RE-ZERO
STABLE FAST MID.
RESPONSE
SLOW

5
3. CONNECTION
3.1. Setting the AD-4212C and the AD-8923-CC
Set the following items so that the AD-4212C and the AD-8923-CC have the
same value for each item.
Item AD-4212C AD-8923-CC
Baud rate 600, 1200, 2400*, 4800, 9600, 19200 bps
Data bits, parity 7 bits EVEN*
Stop bits 1 bit*
Terminator <CR><LF>*
Data format A&D standard format −
Communication control No RTS/CTS control −
Data output mode Stream mode −
* Factory setting for the AD-8923-CC. The factory setting for the AD-4212C
is the same unless otherwise specified.
3.2. Connecting the cables
Connect the cables using the connectors located on the rear of the AD-8923-CC.
Connection example to the AD-4212C and a PLC
Note
Be sure to ground the AD-4212C and the AD-8923-CC.
PLC
(CC-Link master station)
A
D-8923-CC
RS-232C connector (D-Sub 9-pin)
Connect to the power supply
Refer to “3.3. Turning the power on”.
RS-232C cable
provided with the D-4212C
CC-Link connector (5-pin)
G
round
CC-Link network
C
C-Link network
To other remote stations
such as the AD-8923-CC.
AD-4212C
Ground

6
Connection example to the CC-Link network (weighing
instruments No.1 through No.4)
No.1 No.2 No.3 No.4
zThe value of the terminating resistor varies depending on the CC-Link
cable used.
zUse the same resistor at each end of the network.
Cable FANC-SB FANC-110SBH FANC-SBH
Terminating resistor 110Ω1/2 W 130Ω1/2 W
Terminating resistors are not provided.
3.3. Turning the power on
As a power supply, an external 24-VDC power supply (24 VDC±10% /
700mA) or a 12-VDC AC adapter can be used.
Note
When the AD-4212C is used, connecting power to either the
AD-8923-CC or the AD-4212C will supply power to both devices.
So, instead of the AD-8923-CC, using the AC adapter on the
AD-4212C will supply power to the AD-8923-CC.
If the power is connected to both, no problems occur because the
power to be used is selected automatically.
Connect a terminating resistor
only to the stations at each end
of the network.
AD-8923-CC
AD-4212C
AD-8923-CC
AD-4212C
AD-8923-CC
AD-4212C
CC-Link
master station
AD-8923-CC
AD-4212C
(Ground)
Weighing
instrument

7
When the external 24-VDC power supply is used
Connect an external 24-VDC power supply to the DC input terminal located
on the rear of the AD-8923-CC.
Precautions on using the external power supply
CAUTION
zUse a power supply within the rated voltage range (24 VDC±10%).
Never use a power supply with a voltage exceeding the rated range.
•It may cause damage or heat buildup.
•The AD-8923-CC may not function properly.
zGround the FG terminal of the switching power supply used.
•To avoid electrical shock and increase the system safety.
•To increase the resistance against noises.
zDo not share the power line with other devices.
•Strong noises introduced from other devices may cause damage to the
AD-8923-CC.
•Inrush current from other devices may cause the AD-8923-CC not to start
up properly.
•Circuit configuration of the AD-8923-CC may affect other devices to
prevent them from functioning properly.
zSelect a switching power supply with a capacity of approximately 700mA
for each AD-8923-CC. Note that the AD-8923-CC may not start up with a
capacity less than 700mA.
•If the power supply capacity is not sufficient, the AD-8923-CC may not
function properly.
zBe sure to add a noise filter on the front end of the switching power supply
and ground the FG terminal.
•This will increase the resistance against noises.
zBe sure to ground the FG terminal of the AD-8923-CC and the AD-4212C.
•This will increase the resistance against noises.

8
Cable connection
CAUTION
Before inserting the power line, make sure that the power to the
AD-8923-CC is turned off.
(1) Inserting the power line
Press down the release button on
the DC input terminal using a
screwdriver and insert the power
line.
The recommended stripping
length for the power line is 10
mm.
Applicable wire range
•Single wire: φ1.0 mm (AWG 26) to φ1.2 mm (AWG 16)
•Twisted wire: 0.3 mm2(AWG 22) to 0.75 mm2(AWG 20)
Individual wire diameter φ0.18 mm or greater
(2) Securing or removing the power line
To secure the power line, return the release button to the initial position
using the screwdriver. The power line will be locked.
To remove the power line, press the release button again using the
screwdriver, unlocking the power line.
When the AC adapter is used
Insert the AC adapter plug into the
AC adapter input jack located on
the rear of the AD-8923-CC and
insert the AC adapter into an
electrical outlet.
DC in
p
ut terminal
AD-8923-CC rear
Release button Screwdrive
r
Power line
A
C ada
p
ter in
p
ut
j
ac
k
DC in
p
ut terminal
A
C adapter plug
AD-8923-CC rear

9
3.4. Operation
zDisplays the data transmitted by the weighing instrument connected.
zThe AD-8923-CC key functions when connected to a weighing instrument
are listed below: (e.g. when the AD-4212C is connected)
Keys of the AD-8923-CC
Model ON:OFF CAL SAMPLE PRINT MODE RE-ZERO
AD-4212C
Switches
between the
weighing
mode and
the standby
mode.
Enters the
calibration
mode
using a
weight.
Switches
the
minimum
display.
Used for the
function
setting mode
and
calibration
mode.
Switches
the
weighing
speed.
Sets the
display to
zero.
3.5. Calibrating the AD-4212C
The following is the calibration procedure when the AD-4212C is connected.
(A calibration weight is used.)
Caution
zDo not allow vibration, drafts or temperature change to affect the
AD-4212C during calibration.
Caution on using an external calibration weight
zThe accuracy of the weight can influence the accuracy of weighing.
Select an appropriate weight as listed below.
A calibration weight of 200 g (conforming to OIML, Class E2 or
equivalent) is provided with the AD-4212C as a standard accessory.
Weighing instrument Usable calibration weight
AD-4212C-300 50g, 100g,
200 g
,
300g
AD-4212C-600 50g, 100g,
200 g
,
300g, 400 g, 500 g, 600 g
AD-4212C-3000 50g, 100g,
200 g
, 300g, 400 g, 500 g, 1000 g, 2000g, 3000g
AD-4212C-6000
200 g
, 500 g, 1000 g, 2000g, 3000g, 4000 g, 5000 g, 6000 g
The calibration weight in bold type: factory setting
Display
zThis indicator means “the AD-4212C is measuring
calibration data”. Do not allow vibration, drafts or other
external disturbances to affect AD-4212C while this
indicator is displayed.

10
Calibration procedure
Calibrates the AD-4212C using the calibration weight.
Operation
1. Warm up the AD-4212C for 30 minutes or more with
nothing on the pan.
2. Press the CAL key. Cal 0 is displayed.
zIf you want to cancel calibration, press the CAL
key. The display will return to the weighing mode.
zIf you want to change the calibration mass value,
press the SAMPLE key.
Press the RE-ZERO key to select the mass
value, and press the PRINT key to store it. Cal
0 is displayed.
3. Confirm that there is nothing on the pan and press
the PRINT key. The AD-4212C measures the zero
point. Do not allow vibration or drafts to affect the
AD-4212C.
The calibration weight value is displayed.
4. Place a calibration weight, of the weight value
displayed, on the pan and press the PRINT
key. The AD-4212C measures the calibration
weight. Do not allow vibration or drafts to affect
the AD-4212C.
5. end is displayed. Remove the weight from the
pan.
6. The display will automatically return to the weighing
mode.
7. Place the calibration weight on the pan and confirm
that calibration was performed correctly. If not,
check the ambient conditions such as drafts or
vibration, and repeat steps 2 through 7.
Press
Calibration
weight
Press
Press
STABLE FAST MID.
RESPONSE
SLOW
CAL
STABLE FAST MID.
RESPONSE
SLOW
PRINT
STABLE FAST MID.
RESPONSE
SLOW
STABLE FAST MID.
RESPONSE
SLOW
PRINT
STABLE FAST MID.
RESPONSE
SLOW
STABLE FAST MID.
RESPONSE
SLOW
STABLE FAST MID.
RESPONSE
SLOW

11
4. FUNCTION SETTING
Function setting specifies the AD-8923-CC performance. The parameters are
stored in non-volatile memory, and are maintained even if the power line or AC
adapter is removed.
The function setting menu consists of two layers. The first layer is the “Class”
and the second layer is the “Item”. Each item stores a parameter.
Press the SAMPLE key to select an item and press the RE-ZERO key to
change the parameter. Then, press the PRINT key to store the new
parameter.
Example
This example sets “Baud rate” to “9600 bps”.
Note
The AD-8923-CC may not function properly, depending on the
settings and operating environment. Check the settings and
change them as necessary.
START
While holding
down
Item
Class
Item
"Baud rate"
Parameter
"9600 bps"
Press twice
press
END

12
4.1. Display and keys
The STABLE indicator turns on to indicate that the parameter displayed is in effect.
Selects a class or item.
Changes the parameter.
When a class is displayed, moves to an item in the class.
When an item is displayed, stores the new parameter and displays the
next class.
When an item is displayed, cancels the new parameter and displays the
next class.
When a class is displayed, exits the function setting mode and returns to
the weighing mode.
4.2. Function table
Class Item and Parameter Description
- Not fixed
Displays the decimal point position of the
weight data received.
0
dpp
Decimal point
position
5
Fixed
Fixes the decimal point at the set digit.
Even if the minimum display is switched
using the SAMPLE key, the decimal point
position does not change. For details,
refer to “6.3. Fixed decimal point position.”
0 Disabled
Disables the SAMPLE key function.
fnc
Environment
Display
5apl
Sample key
function 1 Enabled
Enables the SAMPLE key function.
0 600 bps
1 1200 bps
2 2400 bps
3 4800 bps
4 9600 bps
5if
Serial
interface
bp5
Baud rate
5 19200 bps
Set the same value as that of the
weighing instrument to be connected.
5tn
Station
number
1
64
Station
number
0 156 Kbps
1
625 Kbps
2 2.5 Mbps
3 5 Mbps
CCl
CC-Link
interface
C-bp
CC-Link
baud rate
4 10 Mbps
Set the same value as that of the
CC-Link master station to be
connected
Factory setting

13
4.3. Initializing the AD-8923-CC
Initialization restores the function settings of the AD-8923-CC to factory
settings.
Operation
1 Turn the power on. - - - - - or weighing
mode display appears.
2 While holding down the ON:OFF key, press
the PRINT key. Clr is displayed.
3 Press the PRINT key.
To cancel this operation, press the CAL key
4 Press the RE-ZERO key to select “go”.
5 Press the PRINT key to perform
initialization.
After initialization, - - - - - or weighing
mode display appears.
.
ON:OFF
PRINT
PRINT
RE-ZERO
PRINT
Press
While holding
down
Press
Press
Press

14
9876
(Va)
(Vs)
SG
TXD
RXD
(CTS)
(RTS)
(DSR)
54321
AD-8923-CC
interior
Inch screw
(#4-40UNC)
To the weighing instrument
5. RS-232C CONNECTOR
The RS-232C cable provided with the AD-4212C can be connected directly.
5.1. RS-232C serial interface specifications
RS-232C
Transmission system
: EIA RS-232C
Transmission form : Asynchronous, bi-directional, half duplex
Data format : Baud rate : 600, 1200, 2400, 4800, 9600, 19200 bps
Data bits : 7 or 8 bits
Parity : EVEN, ODD (Data bits 7 bits)
NONE (Data bits 8 bits)
Stop bits : 1 bit or 2 bits
Code : ASCII
Terminator: <CR> or <CR><LF>
Circuit
Connection to the weighing instrument
D-Sub 9-pin
Pin
No.
Signal
name
Direction Description
1 (Vs) −Used internally
2 RXD Input Receive data
3 TXD Output Transmit data
4 −−N.C.
5 SG −Signal ground
6 (DSR) −Used internally
7 (RTS) −Used internally
8 (CTS)
−Used internally
9 (Va) −Used internally
Note
When the user prepares a cable, do not connect to the pins that are
used internally.
(The AD-8923-CC is a DTE device.
Connect to a DCE device such as the
AD-4212C, using a straight through cable.)

15
6. CC-Link CONNECTOR
The AD-8923-CC CC-Link is a remote device station of CC-Link ver.1.10.
When a CC-Link is used, the AD-8923-CC can be controlled by the PLC
remote I/O or remote registers. So, the program can be simple. And
connection to a PLC is simple, thus, a weighing system can be built easily.
The setting values of CC-Link are changed in the function setting CCl.
6.1. CC-Link interface specifications
Station number 1 to 64
Baud rate 156 kbps, 625 kbps, 2.5 Mbps, 5 Mbps, 10 Mbps
Communications connector
The connector used can be attached or removed while the power is ON.
The function of each signal line is as follows.
DA Signal DA
DB Signal DB
DG Signal ground
SLD Shield
FG Frame ground
Status LEDs
LED ON OFF Blinking
RUN Normal •Resetting
•No signal −
SD Transmitting −−
RD Receiving −−
ERR
•Setting error
•CRC error
•Station trouble
Normal When the setting
values are changed
CC-Link connector and LEDs

16
Memory map
Remote register (Number of occupied stations: 1)
Blank “Name” column: internally reserved (not used).
AD-8923-CC --> Master station Master station --> AD-8923-CC
Remote
register
Buffer
memory Name Remote
register
Buffer
memory Name
RWr0000 2E0 RWw0000 2E0
RWr0001 2E1 Weight value RWw0001 2E1
RWr0002 2E2 Error code RWw0002 2E2
RWr0003 2E3 Error sub-code RWw0003 2E3

17
Remote I/O (Number of occupied stations: 1)
Blank “Name” column: internally reserved (not used).
AD-8923-CC --> Master station Master station --> AD-8923-CC
Remote
input
Buffer
memory Name Remote
output
Buffer
memory Name
RX0000 RY0000 Re-zero
RX0001 RY0001
RX0002 RY0002 Tare (Re-zero)
RX0003 RY0003
RX0004 RY0004
RX0005 RY0005
RX0006 CPU operation RY0006
RX0007 Stable / Unstable RY0007
RX0008 Decimal point 20
RY0008
RX0009 Decimal point 21RY0009
RX000A Decimal point 22RY000A
RX000B RY000B
RX000C RY000C
RX000D RY000D
RX000E
RY000E
RX000F
0E0
RY000F
160
RX0010 RY0010
RX0011 RY0011
RX0012 RY0012
RX0013 RY0013
RX0014 RY0014
RX0015 RY0015
RX0016 RY0016
RX0017
RY0017
RX0018 Request flag of
initialization
RY0018 Reply flag of
initialization
RX0019 Reply flag of initial
data setting
RY0019 Request flag of
initial
data setting
RX001A Error status flag RY001A Request flag of
error reset
RX001B
Remote READY flag
RY001B
RX001C
RY001C
RX001D
RY001D
RX001E
RY001E
RX001F
0E1
RY001F
161

18
Numeric values of the remote register
All the values are hexadecimal. Negative values are expressed by the two’s
complement.
Decimal Hexadecimal (32 bits)
-10 FFFFFFF6
-1 FFFFFFFF
0 00000000
1 00000001
10 0000000A
Weight value examples
1.000 will be 1000, thus expressed as 0x000003E8. (RWr0001: 0x0000,
RWr0000: 0x03E8)
-1.000 will be -1000, thus expressed as 0xFFFFFC18. (RWr0001: 0xFFFF,
RWr0000: 0xFC18)
Prohibited writing in the internally reserved areas
Writing is prohibited in the internally reserved areas.
Writing in the remote output (RY) and the remote register (RWw) of the
internally reserved areas may cause the AD-8923-CC to malfunction.
Values of the remote input (RX) and the remote register (RWr) of the
internally reserved areas are not fixed.
Error code
Error code Error status flag (Instrument error)
0 No error
1 -
2 EEPROM error (Writing error)
3 -
4 Calibration error
5 Weighing instrument error
6 RS-232C error
7 Over (+)
8 Over (-)
Stable / Unstable
RX0007
0 Unstable
1 Stable