ABB AX400 Series User manual

AX400 Series User Guide
Models AX460 and AX466
Single and Dual Input pH/Redox
(ORP) Analyzers
7.00
Monitoring pH
7.00
pH
pH
7.00
Monitoring pH
7.00
pH
pH

ABB
The Company
We are an established world force in the design and manufacture of instrumentation for industrial
process control, flow measurement, gas and liquid analysis and environmental applications.
As a part of ABB, a world leader in process automation technology, we offer customers
application expertise, service and support worldwide.
We are committed to teamwork, high quality manufacturing, advanced technology and unrivalled
service and support.
The quality, accuracy and performance of the Company’s products result from over 100 years
experience, combined with a continuous program of innovative design and development to
incorporate the latest technology.
The NAMAS Calibration Laboratory No. 0255 is just one of the ten flow calibration plants
operated by the Company, and is indicative of our dedication to quality
and accuracy.
Use of Instructions
Warning.
An instruction that draws attention to the risk of injury or
death.
Caution.
An instruction that draws attention to the risk of damage to
the product, process or surroundings.
✶
Note.
Clarification of an instruction or additional information.
Information.
Further reference for more detailed information or technical
details.
Although Warning hazards are related to personal injury, and Caution hazards are associated with equipment or property damage, it
must be understood that operation of damaged equipment could, under certain operational conditions, result in degraded process
system performance leading to personal injury or death. Therefore, comply fully with all Warning and Caution notices.
Information in this manual is intended only to assist our customers in the efficient operation of our equipment. Use of this manual for
any other purpose is specifically prohibited and its contents are not to be reproduced in full or part without prior approval of the
Marketing Communications Department.
Health and Safety
To ensure that our products are safe and without risk to health, the following points must be noted:
1. The relevant sections of these instructions must be read carefully before proceeding.
2. Warning labels on containers and packages must be observed.
3. Installation, operation, maintenance and servicing must only be carried out by suitably trained personnel and in accordance with the
information given.
4. Normal safety precautions must be taken to avoid the possibility of an accident occurring when operating in conditions of high pressure and/
or temperature.
5. Chemicals must be stored away from heat, protected from temperature extremes and powders kept dry. Normal safe handling procedures
must be used.
6. When disposing of chemicals ensure that no two chemicals are mixed.
Safety advice concerning the use of the equipment described in this manual or any relevant hazard data sheets (where applicable) may be
obtained from the Company address on the back cover, together with servicing and spares information.
BS EN ISO 9001:1994
Cert. No. Q05907
R
E
G
I
S
T
E
R
E
D
EN 29001 (ISO 9001)
Lenno, Italy – Cert. No. 9/90A
0255
Stonehouse, U.K.

1
CONTENTS
Section Page
1 INTRODUCTION .......................................................... 2
1.1 System Description ............................................ 2
1.2 AX400 Series Analyzer Options .......................... 2
2 OPERATION ................................................................. 3
2.1 Powering Up the Analyzer .................................. 3
2.2 Displays and Controls ........................................ 3
2.2.1 Key Functions ....................................... 3
2.3 Operating Page .................................................. 6
2.3.1 Single Input pH ..................................... 6
2.3.2 Dual Input pH ....................................... 7
2.3.3 Wash Function ...................................... 8
2.3.4 Single Input Redox (ORP) ..................... 9
2.3.5 Dual Input Redox (ORP) ...................... 10
2.3.6 Dual Input pH and Redox (ORP) ......... 11
3 OPERATOR VIEWS .................................................... 12
3.1 View Set Points ................................................ 12
3.2 View Outputs .................................................... 13
3.3 View Hardware ................................................. 13
3.4 View Software .................................................. 14
3.5 View Logbook .................................................. 14
3.6 View Clock ....................................................... 17
4 SETUP ........................................................................ 18
4.1 Sensor Calibration ............................................ 18
4.1.1 Set Buffer Type (pH Only) .................... 18
4.1.2 Set Up User Defined Buffers
(pH Only) ............................................. 20
4.1.3 Adjust Offset (Redox/ORP Only) .........21
4.1.4 Automatic and Manual, Single-
and Two-Point Calibration (pH Only) ... 22
4.1.5 Grab Calibration (pH Only) .................. 24
5 PROGRAMMING ........................................................ 25
5.1 Security Code .................................................. 25
5.2 Configure Display ............................................. 26
5.3 Configure Sensors ............................................ 27
5.4 Configure Alarms .............................................. 30
5.4.1 Wash Cycle Configuration
(Applicable Only to Alarm 3) ................ 32
5.5 Configure Outputs ............................................ 34
5.6 Configure Clock................................................ 36
5.7 Configure Security ............................................ 37
5.8 Configure Logbook........................................... 37
5.9 Test Outputs and Maintenance ........................ 38
Section Page
6 INSTALLATION ........................................................... 39
6.1 Siting Requirements ......................................... 39
6.2 Mounting .......................................................... 40
6.2.1 Wall-/Pipe-mount Analyzers ................ 40
6.2.2 Panel-mount Analyzers ....................... 41
6.3 Connections, General ....................................... 42
6.3.1 Relay Contact Protection and
Interference Suppression .................... 43
6.3.2 Cable Entry Knockouts,
Wall-/Pipe-mount Analyzer ................. 44
6.4 Wall-/Pipe-mount Analyzer Connections ..........45
6.4.1 Access to Terminals ............................ 45
6.4.2 Connections ....................................... 46
6.5 Panel-mount Analyzer Connections ................. 47
6.5.1 Access to Terminals ............................ 47
6.5.2 Connections ....................................... 48
7 CALIBRATION ............................................................ 49
7.1 Equipment Required ......................................... 49
7.2 Preparation ....................................................... 49
7.3 Factory Settings ............................................... 50
8 SIMPLE FAULT FINDING ........................................... 55
8.1 Error Messages ................................................ 55
8.2 Calibration Fail Message or
no Response to pH/Redox Changes................ 55
8.3 Checking the Temperature Input ...................... 55
APPENDIX A ....................................................................... 56
A1 Buffer Solutions ................................................ 56
SPECIFICATION .................................................................. 59

2
1.1 System Description
The AX400 Series pH/Redox (ORP) analyzers and associated
electrode systems have been designed for continuous
monitoring and control of pH and Redox (ORP). The electrode
system can be standardized to the analyzer using the built-in
calibration facility and a single point buffering facility provides
easy re-calibration after initial standardization.
The analyzer is available in wall-/pipe-mount or panel-mount
versions with either one or two programmable, pH or Redox
(ORP) input channels, each with its own associated temperature
input channel. When making temperature compensated
measurements, the sample temperature is sensed by a
resistance thermometer (Pt100, Pt1000 or Balco 3K) mounted
in the electrode system.
The analyzer can be configured for and connected to either a
standard pH input (single, high impedence input >1013Ω) or
differential pH input (dual, high impedence inputs, both >1013Ω).
Differential pH input is designed for use with pH electrode
systems that incorporate a solution earth (ground) rod. The
measuring electrode and reference electrode signals are
measured separately using two, high impedence amplifiers and
compared with the solution earth (ground) potential. The
difference between the results is the value used for the pH
measurement.
All models incorporate a wash facility for system cleaning; the
alarm 3 relay can be configured to control the wash system
either automatically or manually. The relay can be programmed
to deliver either a continuous or pulsed signal to control an
external power supply to a solenoid or pump and the frequency,
duration and recovery time for the wash cycle are also
programmable. During a wash cycle, the analog output value is
held in its pre-cycle condition.
Analyzer operation and programming are performed using five
tactile membrane keys on the front panel. Programmed
functions are protected from unauthorized alteration by a four-
digit security code.
1 INTRODUCTION
1.2 AX400 Series Analyzer Options
Table 1.1 shows the range of configurations that are possible for
the AX400 Series analyzers. The analyzer automatically detects
the type of input board fitted for each input and displays only the
frames applicable to that input board type. If no input board is
fitted for Sensor B input, Sensor B frames are not displayed.
rebmuNledoMrezylanArezylanAfonoitpircseDArosneSBrosneS
014XA000,01ot0(ytivitcudnoCtupnIelgniS µ)mc/SytivitcudnoCelbacilppAtoN
114XA000,01ot0(ytivitcudnoCtupnIlauD µ)mc/SytivitcudnoCytivitcudnoC
614XA)PRO(xodeR/HpdnaytivitcudnoCtupnIlauDytivitcudnoC)PRO(xodeR/Hp
054XA)PSU(ytivitcudnoCtupnIelgniSytivitcudnoCelbacilppAtoN
554XA)PSU(ytivitcudnoCtupnIlauDytivitcudnoCytivitcudnoC
064XA)PRO(xodeR/HptupnIelgniS)PRO(xodeR/HpelbacilppAtoN
664XA)PRO(xodeR/HptupnIlauD)PRO(xodeR/Hp)PRO(xodeR/Hp
Table 1.1 AX400 Series Analyzer Options

3
Fig. 2.1 Location of Controls and Displays
Fig. 2.2 Membrane Key Functions
0.00
Monitoring pH
0.00
Alarm
LEDs
Display
Lines
Lower
Display Line
Membrane Keys
pH
Units
Menu Key
Sidescroll Key
Downscroll Key
Up Key
Down Key
pH
B – Advancing to Next Page
C – Moving Between Parameters
D – Adjusting and Storing a Parameter Value
E – Selecting and Storing a Parameter Choice
A – Moving Between Menus
For majority
of parameters
Parameter 1
Parameter 2
Parameter 3
Parameter 4
Page 1
Parameter 1
Parameter 2
Parameter 3
Page 2
Advance to
next page
or
Parameter 1
Parameter 2
Parameter 3
Page X
Parameter 4
Advance to
next parameter
New value is
stored automatically
Parameter Value
Adjust
Parameter X
Y
Z
Select
New value is
automatically stored
Menu 1
Menu 2
Advance to
next menu
2.1 Powering Up the Analyzer
Caution. Ensure all connections are made
correctly, especially to the earth stud – see Section 6.3.
1) Ensure the input sensor(s) is/are connected correctly.
2) Switch on the power supply to the analyzer. A start-up
screen is displayed while internal checks are performed,
then the Operating Page (see Section 2.3) is displayed as the
pH or Redox (ORP) monitoring operation starts.
2.2 Displays and Controls – Fig. 2.1
The display comprises two rows of 41/2digit, 7-segment digital
displays, which show the actual values of the measured
parameters and alarm set points, and a 6-character dot matrix
display showing the associated units. The lower display line is a
16-character dot matrix display showing the programming
information.
2.2.1 Key Functions
2 OPERATION

4
…2 OPERATION
Use the Menu
Key to scroll
through
the Menus
Available only if the
analog option board is fitted
Key
VIEW SETPOINTS VIEW OUTPUTS VIEW HARDWARE VIEW SOFTWARE
A1: Setpoint Analog Output 1 Sensor A Mo ule AX400/2000 Issue
A2: Setpoint Analog Output 2 Sensor B Mo ule
A3: Setpoint Analog Output 3 Option Boar
A4: Setpoint Analog Output 4
A5: Setpoint
Security Co e
SENSOR CAL. Cal. User Co e Sensor Cal. A
A: Buffer Metho
A: Immerse Buf.1
A: Cal. Buffer 1
A: Immerse Buf.2
A: Cal. Buffer 2
A: Calibration
CONFIG.SENSORS
B: Probe Type
B: Diff. Input
B: Electro e
B: Temp. Comp
B: Temp. Sensor
B: Preset Temp.
B: Sample Comp.
CONFIG. DISPLAY Set Language Set Temp. Units Set Backlight
LED Backlight
Temp. UnitsEnglish
B: Sample Coeff.
Use the Sidescroll Key to scroll through the Pages within each Menu
To
CONFIG. ALARMS
(see Fig. 2.3B)
Use the Downscroll
Key to scroll through
the Parameters
within each Page
OPERATING PAGE
Section 2.3, Page 6 Section 3.1, Page 12 Section 3.2, Page 13 Section 3.3, Page 13 Section 3.4, Page 14 Section 3.5, Page 14
Section 4.1, Page 18
Section 5.1, Page 25
Section 5.2, Page 26
Section 5.3, Page 27
Section 3.6, Page 17
Set Auto Buffers
Buffer Type
Set Buffer 1
Set Buffer 2
A: Slope & Check
Sensor Cal. B
B: Buffer Metho
B: Immerse Buf.1
B: Cal. Buffer 1
B: Immerse Buf.2
B: Cal. Buffer 2
B: Calibration
B: Slope & Check
B: Set Min Slope
Config. Sensor B
Config. Sensor A
A: Probe Type
A: Diff. Input
A: Temp. Comp
A: Temp. Sensor
A: Preset Temp.
A: Sample Comp.
A: Sample Coeff.
A: Set Min Slope
A: Electro e
A: Enter Point 1
A: Enter Point 2
B: Enter Point 1
A: Enter Point 5
A: Enter Point 4
A: Enter Point 3
B: Enter Point 2
B: Enter Point 5
B: Enter Point 4
B: Enter Point 3
#### 100% ####
#### 100% ####
#### 100% ####
#### 100% ####
Alarms
Errors
Power
Cals
VIEW LOGBOOK
Dual input analyzer only
VIEW CLOCK
Date 01:01:03
Time 12:00
Note. Sensor calibration frames shown
above are for 2-point calibration only.
For other calibration options, refer to Section 4.1.
A: Enable Cals B: Enable Cals
**
*
Applicable only if Probe Type is set to Redox or ORP
Fig. 2.3A Overall Programming Chart

5
2 OPERATION…
CONFIG.ALARMS Config. Alarm 1
A1: Type
A1: Assign
A1: Failsafe
A1: Action
A1: Setpoint
A1: Hysteresis
A1: Delay
Config. Alarm 2
A2: Type
A2: Assign
A2: Failsafe
A2: Action
A2: Setpoint
A2: Hysteresis
A2: Delay
A3: Assign
A3: Failsafe
A3: Action
A3: Setpoint
A3: Hysteresis
A3: Delay
CONFIG.OUTPUTS Config. Output 1
AO1: Assign
AO1: Range
AO1: Span Value
AO1: Zero Value
CONFIG.SECURITY Alter Sec.Co e
Alter Cal.Co e
TEST/MAINTENANCE Test Outputs
Test Output 1
Test Output 2
Maintenance
Hol Outputs
To
FACTORY SETTINGS
(see Section 7.3, Page 50)
Config. Alarm 3
A3: Type
AO1: Default
AO1: Default Val
Available only if the
analog option board is fitted
Key
Config. Output 2
AO2: Assign
AO2: Range
AO2: Span Value
AO2: Zero Value
AO2: Default
AO2: Default Val
Config. Output 3
AO3: Assign
AO3: Range
AO3: Span Value
AO3: Zero Value
AO3: Default
AO3: Default Val
Test Output 3
Test Output 4
Use the Sidescroll Key to scroll through the Pages within each Menu
Section 5.4, Page 30
Section 5.5, Page 34
Section 5.7, Page 37
Section 5.9, Page 38
Use the Menu
Key to scroll
through
the Menus
Use the Downscroll
Key to scroll through
the Parameters
within each Page
CONFIG.CLOCK Set Clock?
Format /mm/yy
Date 01:01:02
Time 12:00
Press To AbortPress To Set
Section 5.6, Page 36
CONFIG.LOGBOOK
Section 5.8, Page 37
Logbook
Config. Alarm 4
A4: Type
A4: Assign
A4: Failsafe
A4: Action
A4: Setpoint
A4: Hysteresis
A4: Delay
Config. Alarm 5
A5: Type
A5: Assign
A5: Failsafe
A5: Action
A5: Setpoint
A5: Hysteresis
A5: Delay
Config. Output 4
AO4: Assign
AO4: Range
AO4: Span Value
AO4: Zero Value
AO4: Default
AO4: Default Val
Wash Frequency
Wash Duration
Recovery Perio
Wash Mo e
*
*
Applicable only
to Alarm 3
Fig. 2.3B Overall Programming Chart

6
Measured Values
pH.
Temperature.
Measured Millivolts
Millivolts.
% Slope and pH Check Value
% slope value.
A value between the programmed minimum % slope value (see Set Min Slope in the
CONFIG. SENSORS page – Section 5.3) and 105% is displayed. If the value is outside these
limits, check the electrode system.
pH check value (zero point).
Displayed as an additional indication of pH electrode system condition; 7pH is the
optimum value for glass electrodes and 0pH for Antimony electrodes.
Alarm 3 set to Wash (A3: Type, Section 5.4) – see Section 2.3.3.
Alarm 3 not set to Wash.
2.3 Operating Page
2.3.1 Single Input pH
…2 OPERATION
Monitoring pH
7.00
pH
24.4
Deg.C
Wash Function
Monitoring pH
Millivolts
404
mV
A: Slope & Check
100.0
%
7.00
pH

7
Monitoring pH
7.00
pH
7.00
pH
Wash Function
Monitoring pH
Temperature
25.6
Deg.C
Millivolts
404
mV
A: Slope & Check
100.0
%
7.00
pH
24.4
Deg.C
-256
mV
B: Slope & Check
%
pH
100.0
7.00
Measured pH
Sensor A.
Sensor B.
Measured Temperature
Sensor A.
Sensor B.
Measured Millivolts
Sensor A.
Sensor B.
Sensor A % Slope and pH Check Value
% slope value.
A value between the programmed minimum % slope value (see Set Min Slope in the
CONFIG. SENSORS page – Section 5.3) and 105% is displayed. If the value is outside these
limits, check the electrode system.
pH check value (zero point).
Displayed as an additional indication of pH electrode system condition; 7pH being the
optimum value for glass electrodes and 0pH for Antimony electrodes.
Sensor B % Slope and pH Check Value
% slope value.
See Sensor A above.
pH check value (zero point).
Alarm 3 set to Wash (A3: Type, Section 5.4) – see Section 2.3.3.
Alarm 3 not set to Wash.
…2.3 Operating Page
2.3.2 Dual Input pH
2 OPERATION…

8
…2.3 Operating Page
2.3.3 Wash Function
Note. Applicable only if Alarm 3 (A3: Type) is set to Wash in the CONFIG. ALARMS page – see Section 5.4.
Wash Function
Off – Wash function off. Lower display line of Operating Page shows
WASH INHIBITED.
On – Wash function controlled automatically. Lower display line of Operating Page
shows WASH IN PROGRESS.
Manual – Enables wash function to be initiated manually – see below.
Caution. Set Wash Function to Off before removing the sensor from the process.
See Section 3.1.
Probe Type set to pH (for either sensor if dual input analyzer) – see Section 5.3.
See Section 4.1.
Probe Type set to Redox or ORP (for both sensors in any combination if dual input analyzer)
– see Section 5.3.
Sensor calibration enabled (Section 5.3) – see Section 4.1.
Alter Sec. Code not set to zero (Section 5.7) – see Section 5.1.
Alter Sec. Code set to zero (Section 5.7) – see Section 5.2.
Wash Function set to Manual – see below.
Wash Function not set to Manual.
Press to Wash (Manual Wash only)
Press to Wash and Press to Abort are shown alternately on the lower display line.
Press the key to initiate the wash cycle. The display returns to the top of
the Operating Page and the lower display line shows WASH IN PROGRESS until
the wash cycle is completed. The Wash Function selection resets to the one
that was set before Manual was selected.
Press the key to abort the wash cycle. The display returns to the top of
the Operating Page.
VIEW SETPOINTS
SENSOR CAL.
Wash Function
-----
Manual
On
Off
-----
Press To Wash
Press To Abort
SENSOR CAL.
Security Co e
CONFIG. DISPLAY
Press To Wash
MONITORING PH
WASH IN PROGRESS
MONITORING PH
…2 OPERATION

9
…2.3 Operating Page
2.3.4 Single Input Redox (ORP)
Measured Values
Millivolts.
Temperature.
Note. If Probe Type is set to ORP in the CONFIG. SENSORS page (see Section 5.3),
the lower display line shows Monitoring ORP.
Sensor A Offset
Displays the offset value for the sensor set in A: Adjust Offset – see Section 4.1.3.
Alarm 3 set to Wash (A3: Type, Section 5.4) – see Section 2.3.3.
Alarm 3 not set to Wash.
Monitoring Re ox
404
mV
25.6
Deg.C
A: mV Offset
0
mV
Wash Function
Dual Re ox
2 OPERATION…

10
Measured Millivolts
Sensor A.
Sensor B.
Note. If Probe Type for both Sensor A and Sensor B is set to ORP in the
CONFIG. SENSORS page (see Section 5.3), the lower display line shows Dual ORP.
Sample Temperature
Sensor A.
Sensor B.
Note. The measured temperature is displayed only if Temp. Sensor is not set to None
in the CONFIG. SENSORS page – see Section 5.3.
Sensor A Offset
Displays the offset value for Sensor A set in A: Adjust Offset – see Section 4.1.3.
Sensor B Offset
Displays the offset value for Sensor B set in B: Adjust Offset – see Section 4.1.3.
Alarm 3 set to Wash (A3: Type, Section 5.4) – see Section 2.3.3.
Alarm 3 not set to Wash.
…2.3 Operating Page
2.3.5 Dual Input Redox (ORP)
…2 OPERATION
Dual Re ox
404
mV
-256
Temperature
25.6
Deg.C
24.4
Deg.C
mV
A: mV Offset
0
mV
B: mV Offset
0
mV
Wash Function
Dual Re ox

11
…2.3 Operating Page
2.3.6 Dual Input pH and Redox (ORP)
Measured pH and Millivolts
Sensor A.
Sensor B.
Note. The Probe Type for Sensors A and B can be set to any combination of pH,
Redox or ORP in the CONFIG. SENSORS page – see Section 5.3. The display indications
change depending on Probe Type settings, e.g. if Sensor A is set to Redox and Sensor B
to pH, the lower display shows Re ox/pH.
Measured Temperature
Sensor A.
Sensor B.
Note. The measured temperature is displayed only if Temp. Sensor is not set to None
in the CONFIG. SENSORS page – see Section 5.3.
Measured Millivolts
Sensor A.
Sensor B.
Sensor B Offset
Displays the offset value for Sensor B set in B: Adjust Offset – see Section 4.1.3.
Sensor A % Slope and pH Check Value
% slope value.
A value between the programmed minimum % slope value (see Set Min Slope in the
CONFIG. SENSORS page – Section 5.3) and 105% is displayed. If the value is outside these
limits, check the electrode system.
pH check value (zero value).
Displayed as an additional indication of pH electrode system condition; 7pH is the
optimum value for glass electrodes and 0pH for Antimony electrodes.
Alarm 3 set to Wash (A3: Type, Section 5.4) – see Section 2.3.3.
Alarm 3 not set to Wash.
pH/Re ox
7.00
pH
-256
mV
Temperature
25.6
Deg.C
Millivolts
404
mV
A: Slope & Check
100.0
%
7.00
pH
24.4
Deg.C
-256
mV
B: mV Offset
0
mV
Wash Function
pH/Re ox
2 OPERATION…

12
3 OPERATOR VIEWS
View Set Points
This page shows alarm set points. The value of each of the set points is shown, together
with the name of the parameter it is assigned to.
Set point values and relay/LED actions are programmable – see Section 5.4.
Sensor A (pH), Alarm 1 Set Point
Sensor A (Temperature), Alarm 2 Set Point
Sensor B (pH), Alarm 3 Set Point – Dual input analyzers only
Sensor B (Temperature), Alarm 4 Set Point – Dual input analyzers only
Note. Alarm 4 is available only if the optional analog output board is fitted.
Alarm 5 Set Point
Note. Alarm 5 is available only if the optional analog output board is fitted.
Return to main menu.
Sensor calibration enabled (Section 5.3) – see Section 4.1.
Alter Sec. Code not set to zero (Section 5.7) – see Section 5.1.
Alter Sec. Code set to zero (Section 5.7) – see Section 5.2.
8.30
pH
A1: Setpoint
Sen.A
Temp.A
VIEW SETPOINTS
-----
35.0
Deg.C
A2: Setpoint
Sen.B
6.80
A3: Setpoint
Temp.B
55.0
Deg.C
A4: Setpoint
-----
Off
A5: Setpoint
VIEW OUTPUTS
SENSOR CAL.
Security Co e
CONFIG. DISPLAY
VIEW SETPOINTS
pH
3.1 View Set Points
Note. The parameter names and units of measurement displayed in the View Set Points page depend on the Probe Type
settings for Sensors A and B in the CONFIG. SENSORS page – see Section 5.3. Those shown below are given as examples only.

13
3 OPERATOR VIEWS…
3.2 View Outputs
Theoretical Analog Output
There are up to four analog outputs, each showing information for one sensor.
Note. Analog outputs 3 and 4 are available only if the optional analog output board
is fitted.
Live current output value being retransmitted.
Current output shown as a percentage of full scale for the output range set in CONFIG.
OUPUTS – see Section 5.6.
See Section 3.3.
Sensor calibration enabled (Section 5.3) – see Section 4.1.
Alter Sec. Code not set to zero (Section 5.7) – see Section 5.1.
Alter Sec. Code set to zero (Section 5.7) – see Section 5.2.
Analog Output 2 – repeat for output 2 (and outputs 3 and 4 if optional
analog output board fitted).
3.3 View Hardware
Sensor A Module
Shows the type of input board fitted to the analyzer for the Sensor A input.
Sensor B Module – Dual input analyzers only
Shows the type of input board fitted to the analyzer for the Sensor B input.
Option Board
Shows the type of option board fitted to the analyzer (if applicable).
See Section 3.4.
Sensor calibration enabled (Section 5.3) – see Section 4.1.
Alter Sec. Code not set to zero (Section 5.7) – see Section 5.1.
Alter Sec. Code set to zero (Section 5.7) – see Section 5.2.
50.0
%
Analog Output 1
12.00
mA
VIEW OUTPUTS
-----
VIEW HARDWARE
SENSOR CAL.
Security Co e
CONFIG. DISPLAY
Analog Output 2
pH
pH
-----
Sensor A Mo ule
VIEW HARDWARE
-----
Option Boar
-----
Sensor B Mo ule
-----
Off
Mo bus
Analog
VIEW SOFTWARE
SENSOR CAL.
Security Co e
CONFIG. DISPLAY
VIEW HARDWARE

14
…3 OPERATOR VIEW
3.4 View Software
Issue
Shows the version number of the operating software.
Optional analog output board fitted and Logbook set to On in the CONFIG.
LOGBOOK page (Section 5.8) – see Section 3.5.
Operating Page (optional analog output board not fitted) – see Section 2.3.
Sensor calibration enabled (Section 5.3) – see Section 4.1.
Alter Sec. Code not set to zero (Section 5.7) – see Section 5.1.
Alter Sec. Code set to zero (Section 5.7) – see Section 5.2.
The logbook stores data entries for alarm events, sensor errors, power failures and pH
calibration information.
View Logbook
Use the and keys to access the Alarms logbook.
Alarms
The Alarms logbook contains up to 10 entries (entry 1 is the most recent), each comprising
an alarm number, alarm state (On or Off) and the date and time of the occurrence.
Optional analog input board fitted – see Section 3.6.
Sensor calibration enabled (Section 5.3) – see Section 4.1.
Alter Sec. Code not set to zero (Section 5.7) – see Section 5.1.
Alter Sec. Code set to zero (Section 5.7) – see Section 5.2.
Advance to entries 2 to 10.
Note. If no more entries are stored, the display shows No More Entries.
1.00
AX400/2000 Issue
VIEW SOFTWARE
-----
VIEW LOGBOOK
SENSOR CAL.
Security Co e
CONFIG. DISPLAY
VIEW SOFTWARE Monitoring pH
-----
11:09:02 12:34
VIEW LOGBOOK
-----
VIEW CLOCK
SENSOR CAL.
Security Co e
CONFIG. DISPLAY
A1
-----
VIEW LOGBOOK
1
2
Cals
Power
Errors
Alarms
A1
On
3.5 View Logbook
Note. The VIEW LOGBOOK function is available only if the optional analog output board is fitted and Logbook is set to On in
the CONFIG. LOGBOOK page – see Section 5.8.

15
3 OPERATOR VIEWS…
…3.5 Logbook
View Logbook
Use the and keys to access the Errors logbook.
Errors
The Errors logbook contains up to 5 entries (entry 1 is the most recent), each comprising
the sensor letter, error number and the date and time of the occurrence.
Optional analog input board fitted – see Section 3.6.
Sensor calibration enabled (Section 5.3) – see Section 4.1.
Alter Sec. Code not set to zero (Section 5.7) – see Section 5.1.
Alter Sec. Code set to zero (Section 5.7) – see Section 5.2.
Advance to entries 2 to 5.
Note. If no more entries are stored, the display shows No More Entries.
View Logbook
Use the and keys to access the Power logbook.
Power
The Power logbook contains up to 2 entries (entry 1 is the most recent), each comprising
the power state (On or Off) and the date and time of the occurrence.
Optional analog input board fitted – see Section 3.6.
Sensor calibration enabled (Section 5.3) – see Section 4.1.
Alter Sec. Code not set to zero (Section 5.7) – see Section 5.1.
Alter Sec. Code set to zero (Section 5.7) – see Section 5.2.
Advance to entry 2.
Note. If no more entries are stored, the display shows No More Entries.
-----
-----
11:09:02 12:34
VIEW CLOCK
SENSOR CAL.
Security Co e
CONFIG. DISPLAY
Sen.A
VIEW LOGBOOK
Errors
1
Sen.A
Pt100
2
Alarms
Cals
Power
Errors
-----
11:09:02 12:34
VIEW CLOCK
SENSOR CAL.
Security Co e
CONFIG. DISPLAY
-----
VIEW LOGBOOK
1
Off
2
Errors
Alarms
Cals
Power

16
…3.5 Logbook
View Logbook
Use the and keys to access the Cals logbook.
Calibration
The Cals logbook contains up to 5 entries (entry 1 is the most recent), each comprising 2
frames. Frame 1 contains the entry number, sensor letter and the calibration pass/fail
indication.
Frame 2 contains the % slope value, the pH check value and the date and time of the
occurrence.
Optional analog input board fitted – see Section 3.6.
Sensor calibration enabled (Section 5.3) – see Section 4.1.
Alter Sec. Code not set to zero (Section 5.7) – see Section 5.1.
Alter Sec. Code set to zero (Section 5.7) – see Section 5.2.
Advance to entries 2 to 5.
Note. If no more entries are stored, the display shows No More Entries.
VIEW CLOCK
-----
Calibration
-----
VIEW LOGBOOK
1
Sen.A
Passe
11:09:02 12:34
SENSOR CAL.
Security Co e
CONFIG. DISPLAY
Sen.A
100.0
%
pH
2
7.00
Power
Errors
Alarms
Cals
…3 OPERATOR VIEWS

17
Date
Shows the current date.
Time
Shows the current time.
Operating Page – see Section 2.3
Sensor calibration enabled (Section 5.3) – see Section 4.1.
Alter Sec. Code not set to zero (Section 5.7) – see Section 5.1.
Alter Sec. Code set to zero (Section 5.7) – see Section 5.2.
3.6 View Clock
Note. The VIEW CLOCK function is available only if the optional analog output board is fitted.
3 OPERATOR VIEWS
-----
Date 01:01:01
VIEW CLOCK
-----
-----
Time 12:00
Monitoring pH
SENSOR CAL.
Security Co e
CONFIG. DISPLAY
VIEW CLOCK

18
4 SETUP
4.1 Sensor Calibration
SENSOR CAL.
-----
Cal. User Co e
0000
Set Auto Buffers
Sensor Cal. A
4.1.1 Set Buffer Type (pH Only)
Sensor Calibration
Note. When Probe Type for either sensor (Sensor A only if single-input) is set to
Redox or ORP, this section is applicable only if Enable Cals. for that sensor is set to Ye s –
see Section 5.3.
Sensor Calibration Security Code
Enter the required code number (between 0000 and 19999), to gain access to the sensor
calibration procedure. If an incorrect value is entered, access to subsequent calibration
pages is prevented and the display reverts to the SENSOR CAL. page.
Note. Applicable only if Alter Cal. Code is not set to zero – see Section 5.7.
Probe Type set to pH (for either sensor if dual input analyzer) – continued
below.
Probe Type set to Redox or ORP (for both sensors if dual input analyzer) –
continued on page 21.
Set Auto Buffers
Continued on Page 21.
Alter Sec. Code not set to zero (Section 5.7) – see Section 5.1.
Alter Sec. Code set to zero (Section 5.7) – see Section 5.2.
Continued below.
Buffer Type
Select the relevant type of buffer solution (see Appendix A):
ABB – ABB supplied buffer solution.
NIST – NIST buffer solution.
DIN – DIN 19266 buffer solution.
MERCK – MERCK buffer solution
TECH – US Technical buffer solution
User – Buffer solution with a user defined pH value – see Section 4.1.2.
Buffer Type not set to User – continued on next page.
Buffer Type set to User – see Section 4.1.2.
Set Auto Buffers
-----
Sensor Cal. A
SECURITY CODE
CONFIG. DISPLAY
Buffer Type
-----
User
TECH.
MERCK
DIN
NIST
ABB
Set Buffer 1
Buffer Type
A: Enter Point 1
Other manuals for AX400 Series
2
This manual suits for next models
7
Table of contents
Other ABB Measuring Instrument manuals

ABB
ABB Aztec 600 User manual

ABB
ABB COPA-XM 3000 Series User manual

ABB
ABB EL3060 Series User manual

ABB
ABB Gas Sampling Probe 2 User manual

ABB
ABB Smart User manual

ABB
ABB MagMaster User manual

ABB
ABB MagMaster User manual

ABB
ABB EL3000 Series User manual

ABB
ABB XFC6200EX User guide

ABB
ABB 10DX3111G User manual
Popular Measuring Instrument manuals by other brands

DAKOTA ULTRASONICS
DAKOTA ULTRASONICS Z-300-0001 Operation manual

Bender
Bender ISOMETER IR427 + MK7 manual

Hanna Instruments
Hanna Instruments HI97700 instruction manual

Wildlife Acoustics
Wildlife Acoustics Song Meter Mini Bat 2 user guide

CTEK
CTEK BATTERY ANALYZER quick start guide

GOK
GOK ODR Assembly and operating manual