Absolute Process Instruments APD 4152 User manual

Universal Thermistor to DC Isolated Transmitter, Field Configurable APD 4152
Input: Thermistor, or Custom Sensor
Output: 0-1 V to 0-10 V, ±5 V, ±10 V, 0-2 mA to 4-20 mA, Reverse Acting Optional
OOne Model Covers All Common Thermistor Sensors
OEasy Setup—No Computer or Software Needed
OZero and Span Output Calibration Buttons
OFull 1200 V Isolation
OInput LoopTracker®LED, Output Test Function
OBuilt-In Loop Power Supply for Sink/Source Output
QConvert/Isolate Thermistors for PLC Input, Control
and/or Validation
QInterface Thermistors with Panel Meters, PLCs,
Recorders, Data Acq., DCS, & SCADA Systems
Applications
Description
The APD 4152 accepts a thermistor temperature input and
provides an optically isolated and linearized DC voltage or
current output.
The sensor type, temperature range and output range are field
configurable. The input type is set with switches and its range
is configured using front buttons, a multimeter and an input
simulator. This provides a versatile solution that works with all
commonly available thermistor sensors.
Microprocessor-based linearization uses 41 to 55 segments or
up to a 14th order polynomial depending on the sensor type.
The input is sampled, digitally converted to a linearized tem-
perature signal and then passed through an optocoupler to the
output stage.
Full 3-way isolation (input, output, power) make this module
useful for ground loop elimination, common mode signal rejec-
tion, and noise pickup reduction.
The low noise 18 bit analog output is isolated and can be set up
for common voltage and milliamp output types.
Output Sink/Source Versatility
Standard on the APD 4152 is a 20 VDC loop excitation supply
for the milliamp output. The output can be selectively wired for
sinking or sourcing allowing use with a powered or unpowered
milliamp device.
LoopTracker
An API exclusive feature includes a green LoopTracker LED that
varies in intensity with changes in the process input signal.
It provides a quick visual picture of your process input at all
times and can greatly aid in saving time during initial startup
and troubleshooting.
Output Test
An API exclusive feature includes an output test switch to
provide a fixed output (independent of the input) when pressed.
The output test greatly aids in saving time during initial startup
and/or troubleshooting.
The test output level is adjustable from 0 to 100% of the
output span.
Input Types, Field Selectable
Thermistors: 44004/44033 2.252 kΩat 25°C
44005/44030 3.000 kΩat 25°C
44007/44034 5.000 kΩat 25°C
44006/44031 10.00 kΩat 25°C
44008/44032 30.00 kΩat 25°C
YSI 400 2.252 kΩat 25°C
Spectrum 1003k 1 kΩ
Custom: Provide sensor specifications, temperature
curve data, and temperature range
LoopTracker
Variable brightness green LED indicates input level and status
Status LED
Yellow LED for setup and operational status
DC Output Ranges, Field Selectable
Voltage: 0-1 V, 0-2 V, 0-4 V, 0-5 V, 1-5 V,
0-8 V, 0-10 V, 2-10 V,
±5 VDC, ±10 VDC
Current: 0-2 mA, 0-4 mA, 0-8 mA,
0-10 mA, 2-10 mA, 0-16 mA, 0-20 mA,
4-20 mA
20 V compliance, 1000 Ωat 20 mA
Reverse Acting Output, Factory Set
Roption: Reverse acting output
increasing input = decreasing output signal
Reverse acting models cannot be converted to direct acting
Output Calibration
Zero and span set by using up/down buttons, ±10% range
Output Loop Power Supply
20 VDC nom., regulated, 25 mADC, <10 mVRMS max. ripple
May be selectively wired for sinking or sourcing mA output
Output Test
Front push button switch enables/disables test level output
Adjustable 0-100% of span via up/down buttons
Output Resolution
18 bit
Output Ripple and Noise
Less than ±0.2% of span
Accuracy
±0.1°C accuracy and 0.001°C resolution
Ambient Temperature Range and Stability
–10°C to +60°C operating ambient
Better than ±0.02% of span per °C stability
Response Time
300 milliseconds nominal
Isolation
Full 3-way isolation: input, output, power, 1200 VRMS min.
600 VACp or 600 VDC common mode protection
75 dB minimum common mode rejection
Simultaneous 50 Hz and 60 Hz rejection
Housing and Connectors
IP 40, requires vertical installation on a 35 mm DIN rail inside
a panel or enclosure
For use in Pollution Degree 2 Environment
Four 4-terminal removable connectors, 14 AWG max. wire size
Power
85-265 VAC, 50/60 Hz or 60-300 VDC, 3 W maximum
D versions: 9-30 VDC or 10-32 VAC 50/60 Hz, 3 W maximum
Removable Plugs
Sink or Source
mA Output
Setup and Status
LED
Input LoopTracker
LED
Zero and Span for
Output
Universal
Thermistor Input
Adjustable Output
Test Function
How to Order
Model Description Power
APD 4152 Thermistor input
to DC output
isolated
transmitter
85-265 VAC, 50/60 Hz or
60-300 VDC
APD 4152 D 9-30 VDC or 10-32 VAC
Options and Accessory
Universal
Power
Note: An appropriate simulator and a multimeter are required
for setup. We can set up the I/O ranges at no extra charge.
Models are field rangeable. For free setup specify the following.
Thermistor: Sensor model/type, resistance
Custom: Complete sensor data over temperature range
Temperature: Range in °F or °C
Output: Range and type (mV, V, mA)
Options—add to end of model number
NC5 5 point NIST traceable calibration certificate
NC11 11 point NIST traceable calibration certificate
UConformal coating for moisture resistance
RReverse acting output
Accessory—order as separate line item
API BP4 Spare removable 4 terminal plug, black
Dimensions
Height includes connectors
0.89" W x 4.62" H x 4.81" D (22.5 x 117 x 122 mm)
Quick Link: api-usa.com/apd4152
1 2 3 4
5 6 7 8
9 10 11 12
13 14 15 16
See Wiring
Diagrams on
Next Page
© 11-19
api-usa.com
1220 American Way Libertyville, IL 60048
Phone: 800-942-0315 Fax: 800-949-7502
AABSOLUTEBSOLUTE PPROCESSROCESS IINSTRUMENTSNSTRUMENTS
Minute
Setup!
1
Free Factory
I/O Setup!
HHHHHH
HHHHH
HHHHHH
HHHHH
HHHHHH
HHHHH
HHHHHH
HHHHH
HHHHHH
Made in USA
File E145968
85-265 VAC, 60-300 VDC
model only
Applications Link
api-usa.com/apps
Pb Lead
Free
Adjusting Output After Installation
It may be necessary to fine-tune the output signal after instal-
lation to account for offset, tare, lead length, or operating
temperature.
1. Press and release the Set button. This will turn on the yel-
low Status LED.
2. Use the Up and Down buttons to adjust the output to
the desired level. The Status LED will turn off during the
adjustment.
3. Once the desired output level has been met, press and
release the Set button to save the adjustment. The
“Status” LED will flash indicating that the change has been
made.
The unit has an auto Zero/Span detection for knowing which to
adjust. If the output signal is greater than 50% of the Span, the
unit will adjust the output signal Span.
If the output signal is less than 50% of the Span, the unit will
adjust the output signal Zero.
Changing I/O Setup
To reset the unit back to factory default without changing any
input switch settings press and hold the Set button while the
module is being powered up.
If using a new input switch setting, the unit will automatically
start in setup mode to allow you to calibrate and store your new
configuration.
Output Test Function
When the Test button is pressed it will drive the output with a
known good signal that can be used as a diagnostic aid during
initial start-up or troubleshooting. When pressed again, the out-
put will return to normal. The button allows hands-free operation
of the Test Mode.
The Test level can be adjusted by using the Up and Down
buttons.
The level can be set by pressing the Set button, or it can default
back to the setup value by not pressing the Set button.
Operation
The APD 4152 accepts a thermistor input and provides a linear-
ized and optically isolated DC voltage or current output.
The green LoopTracker®input LED provides a visual indication
that a signal is being sensed by the input circuitry of the mod-
ule. It also indicates the input signal strength by changing in
intensity as the process changes from minimum to maximum.
If the LED fails to illuminate, or fails to change in intensity as the
process changes, check the module power or signal input wiring.
Note that it may be difficult to see the LEDs under bright lighting
conditions.
The yellow status LED provides a visual indication of operational
modes.
Normal operation: Off
Push-to-Test mode: Steadily on
User setup mode: Blinking
Blinking Yellow LED Operational Error Codes
If an error occurs during operation, the yellow Status LED blinks
an error code. Check sensor, wiring, or consult factory.
1 1 Analog-digital converter out-of-
range
1 2 Sensor under range
1 3 Sensor over range
1 6 Hard ADC out-of-range
1 7 Sensor hard fault, open circuit,
hard ADC fault
Precautions
WARNING! All wiring must be performed by a qualified electrician
or instrumentation engineer. See diagram for terminal designa-
tions and wiring examples. Consult factory for assistance.
WARNING! Avoid shock hazards! Turn signal input, output, and
power off before connecting or disconnecting wiring, or remov-
ing or installing module.
Précautions
ATTENTION! Tout le câblage doit être effectué par un élec-
tricien ou ingénieur en instrumentation qualifié. Voir le dia-
gramme pour désignations des bornes et des exemples de
câblage. Consulter l'usine pour assistance.
ATTENTION! Éviter les risques de choc! Fermez le signal
d'entrée, le signal de sortie et l'alimentation électrique avant
de connecter ou de déconnecter le câblage, ou de retirer ou
d'installer le module.
Electrical Connections
See wiring diagrams. Observe polarity.
*Do not make any connections to unused terminals or use
them as wiring junctions for external devices. This may cause
permanent damage to the module!
Range Selection
Select ranges before installation. Use the tables on the next
page to select the I/O ranges. The module side label lists output
ranges. Ranges can also be found at api-usa.com/4152
Check the model/serial number label for module power,
options, or custom range information. A custom range uses
switch settings described in the Custom Range Table.
Models with Rreverse acting output use the same switch
settings, except the output range is reversed (4-20 mA is
20-4 mA).
1. Set switches A, B, and C from the table to set input type and
range.
2. Set switches D and E from the table to set the output range
and set switch E: V for voltage or I for current output.
For output ranges that fall between the listed ranges use the
next highest setting. The output can be trimmed using the
zero and span buttons.
Electrical Connections
See wiring diagrams at right. A multimeter and a resistance
simulator are required for setup. Observe polarity. If the output
does not function, check wiring and polarity.
The power supplies are fuse protected and the unit may be
returned to API for fuse replacement.
Input
The thermistor sensor input is connected as shown in the wir-
ing diagrams at right. If a custom input was specified, see the
model/serial number label for sensor type, temperature range
and options.
Output
For milliamp ranges, determine if your device provides power
to the current loop or if the loop must be powered by the APD
module. Typical voltage may be 9-24 VDC at your device's
terminals if it provides power to the loop.
Module Power
Check model/serial number label for module operating voltage
to make sure it matches available power.
When using DC power, either polarity is acceptable, but for
consistency with similar API products, positive (+) can be wired
to terminal 13 and negative (–) can be wired to terminal 16.
Range Calibration
Input and output ranges, if specified on your order, are factory
pre-configured (at 24°C ±1°C).
Note: Perform the following calibration procedure any time
input or output settings are changed.
1. Connect a multimeter to the output terminals 2 - 3, or 3 - 4
depending on output type. See wiring diagram at right.
2. Connect a resistance simulator to the input of the module.
3. Connect power to the unit (terminals 13, 14, and 16) and
apply power to the module.
4. Wait until the yellow Status LED starts blinking once per
second.
Low End Input Calibration
5. Use the simulator to apply the low end of the input signal.
6. Push the Set button to store the low end input value.
7. The Status LED will turn on to indicate the reading was
saved.
8. Use the Up and Down buttons to adjust the output to the
desired low output reading. For example: 4 mA for a 4-20
mA output or –10 V for a ±10V output.
9. Press and release the Set button to store the low output.
High End Input Calibration
10. Wait until the Status LED blinks once per second.
11. Use the simulator to apply the high end of the input signal.
12. Push the Set button to store the high end input value.
13. The Status LED will turn on to indicate the reading was
saved.
14. Use the Up and Down buttons to adjust the output to the
desired high output reading (i.e. 20 mA for a 4-20 mA output).
15. Press and release the Set button to store the high output.
Blinking Yellow LED Setup Error Codes
If an error occurred or invalid selection was made, the yellow
Status LED blinks an error code. Check switches A, B, C, and
input wiring.
2 1 Invalid sensor selected
2 4 Invalid thermistor selected
2 7 Invalid input setting
(Zero greater than Span)
Output Test Level Adjustment
16. Wait until the Status LED turns on and stays on.
17. Using the Up and Down buttons adjust the test output for
the desired level (i.e. 12 mA for a 4-20 mA output).
18. Press and release the Set button to store the test output.
19. Wait until the Status LED starts blinks once per second.
20. To change any value, turn off the power and repeat steps 1 to
19.
Saving Setup
21. Press and release the Set button to store the settings and
lock them into memory. The Status LED will be on during this
process.
22. Once the Status LED turns off, setup and configuration
is complete. Turn off power to the unit and remove the
simulator and multimeter.
Instructions APD 4152
2
api-usa.com
1220 American Way Libertyville, IL 60048
Phone: 800-942-0315 Fax: 800-949-7502
AABSOLUTEBSOLUTE PPROCESSROCESS IINSTRUMENTSNSTRUMENTS
API maintains a constant effort to upgrade and improve its products.
Specifications are subject to change without notice. See api-usa.com for
latest product information. Consult factory for your specific requirements.
WARNING: This product can expose you to chemicals includ-
ing nickel, which are known to the State of California to cause
cancer or birth defects or other reproductive harm. For more
information go to www.P65Warnings.ca.gov
Mounting to a DIN Rail
Install module vertically on a 35 mm DIN rail in a protective
enclosure away from heat sources. Do not block air flow. Allow
1" (25 mm) above and below housing vents for air circulation.
1. Tilt front of module downward and clip
the lower mount with spring clips to the
bottom edge of DIN rail.
2. Push front of module upward until upper
mount snaps into place.
Removal
Avoid shock hazards! Turn signal input,
output, and power off before removing
module.
1. Push up on bottom back of module.
2. Tilt the front of module downward to
release upper mount from top edge of
DIN rail.
3. The module can now be removed from
the DIN rail.
– +
Voltage
Device
21 3 4
Voltage Output
Switch E set to “V”
– + 20V
mA
Device
Ri
21 3 4
Current Sourcing Output
Switch E set to “I”
Thermistor Input 109 11 12
65 7 8
Do Not Connect to Unused
Terminals 1, 5, 6, 7, 8
Spring
Clip
Upper
Mount
Lower
Mount
3
To maintain full isolation and avoid
malfunctions, do not connect power
supplies in common with input, out-
put or unit power.
Yellow status LED
Setup: blinks once per second
Off: normal operation
2 digit code: error code
Green LoopTracker LED brightness
varies with input level
13 Power AC or DC +
14 Earth Ground
16 Power AC or DC –
1413 15 16
Cu 60/75°C
conductors
14 AWG
max.
Thermistor Output ±10 V 0-10 V ±5 V 2-10 V 0-8 V 0-5 V 1-5 V 0-4 V 0-2 V 0-1 V 0-20 mA 4-20 mA 0-16 mA 0-10 mA 2-10 mA 0-8 mA 0-4 mA 0-2 mA
Switches ABCDE ABCDE ABCDE ABCDE ABCDE ABCDE ABCDE ABCDE ABCDE ABCDE ABCDE ABCDE ABCDE ABCDE ABCDE ABCDE ABCDE ABCDE
Type Ohms
44004/44033 2.252 k 6185V 6183V 6184V 6187V 6182V 6189V 6186V 6181V 6188V 6180V 6183I 6187I 6182I 6189I 6186I 6181I 6188I 6180I
44005/44030 3 k 6285V 6283V 6284V 6287V 6282V 6289V 6286V 6281V 6288V 6280V 6283I 6287I 6282I 6289I 6286I 6281I 6288I 6280I
44007/44034 5 k 6385V 6383V 6384V 6387V 6382V 6389V 6386V 6381V 6388V 6380V 6383I 6387I 6382I 6389I 6386I 6381I 6388I 6380I
44006/44031 10 k 6485V 6483V 6484V 6487V 6482V 6489V 6486V 6481V 6488V 6480V 6483I 6487I 6482I 6489I 6486I 6481I 6488I 6480I
44008/44032 30 k 6585V 6583V 6584V 6587V 6582V 6589V 6586V 6581V 6588V 6580V 6583I 6587I 6582I 6589I 6586I 6581I 6588I 6580I
YSI 400 2.252 k 6685V 6683V 6684V 6687V 6682V 6689V 6686V 6681V 6688V 6680V 6683I 6687I 6682I 6689I 6686I 6681I 6688I 6680I
Spectrum 1003 1 k 6785V 6783V 6784V 6787V 6782V 6789V 6786V 6781V 6788V 6780V 6783I 6787I 6782I 6789I 6786I 6781I 6788I 6780I
Range Table: Thermistor APD 4152
Custom Input Output ±10 V 0-10 V ±5 V 2-10 V 0-8 V 0-5 V 1-5 V 0-4 V 0-2 V 0-1 V 0-20 mA 4-20 mA 0-16 mA 0-10 mA 2-10 mA 0-8 mA 0-4 mA 0-2 mA
Switches ABCDE ABCDE ABCDE ABCDE ABCDE ABCDE ABCDE ABCDE ABCDE ABCDE ABCDE ABCDE ABCDE ABCDE ABCDE ABCDE ABCDE ABCDE
Type
See module side label FF85V FF83V FF84V FF87V FF82V FF89V FF86V FF81V FF88V FF80V FF83I FF87I FF82I FF89I FF86I FF81I FF88I FF80I
Range Table: Custom Input Range APD 4152
For models with "R" option, output ranges are reversed
For models with "R" option, output ranges are reversed
Current Sinking Output
switch E set to “I”
External loop supply
2134
– +
+–
Loop
Power
Source
+
–
4-20 mA
Device
Ri
api-usa.com
1220 American Way Libertyville, IL 60048
Phone: 800-942-0315 Fax: 800-949-7502
AABSOLUTEBSOLUTE PPROCESSROCESS IINSTRUMENTSNSTRUMENTS
12 3 4
5 6 7 8
9 10 11 12
13 14 15 16
Input
Not used
Output
Power
Wire terminal torque
0.5 to 0.6 Nm or
4.4 to 5.3 in-lbs
Module powers mA output loop
Module mA output is unpowered
* Do not make
connections to unused
terminals!
* Do not make connections
to unused terminals!
api-usa.com
1220 American Way Libertyville, IL 60048
Phone: 800-942-0315 Fax: 800-949-7502
AABSOLUTEBSOLUTE PPROCESSROCESS IINSTRUMENTSNSTRUMENTS
Notes APD 4152
Date installed Model Serial number Location Input range Output range A B C D E
Range Setup Record

This manual suits for next models

6

Popular Accessories manuals by other brands

Banner SLU4 quick start guide

Banner

Banner SLU4 quick start guide

Bosch rexroth PSM instruction manual

Bosch

Bosch rexroth PSM instruction manual

Spice SPP030 user manual

Spice

Spice SPP030 user manual

Guzzanti GZ-24B instruction manual

Guzzanti

Guzzanti GZ-24B instruction manual

JUMO digiLine O-DO S10 operating manual

JUMO

JUMO digiLine O-DO S10 operating manual

Tri-Tronics M8 installation manual

Tri-Tronics

Tri-Tronics M8 installation manual

SPL Crimson 1250 manual&nbsp;

SPL

SPL Crimson 1250 manual&nbsp;

cimuka CT Series user manual

cimuka

cimuka CT Series user manual

STIEBEL ELTRON SRC C digital Operation and installation manual

STIEBEL ELTRON

STIEBEL ELTRON SRC C digital Operation and installation manual

Elsys EMS Door operating manual

Elsys

Elsys EMS Door operating manual

Asus ABTU001 user manual

Asus

Asus ABTU001 user manual

Vega SMART-SS0102 user manual

Vega

Vega SMART-SS0102 user manual

Rohde & Schwarz NRT-Z14 user manual

Rohde & Schwarz

Rohde & Schwarz NRT-Z14 user manual

LEGRAND Wattstopper CI-355 installation instructions

LEGRAND

LEGRAND Wattstopper CI-355 installation instructions

Optex BGS-V2000 Series instruction manual

Optex

Optex BGS-V2000 Series instruction manual

Philips LP2PB201CS user guide

Philips

Philips LP2PB201CS user guide

Lippert 2021130655 Installation and owner's manual

Lippert

Lippert 2021130655 Installation and owner's manual

Sandstrom S696PPW14 instruction manual

Sandstrom

Sandstrom S696PPW14 instruction manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.