IntelliPack Series 801T Transmitter/Alarm User's Manual mV/TC/RTD Input
___________________________________________________________________________________________
- 3 -
Front panel push buttons are available to reset a latched alarm
and to facilitate field configuration without connecting to a host
computer.
The module uses a high resolution, low noise, Sigma-Delta,
Analog to Digital Converter (-ADC) to accurately convert the
input signal into a digitized value. An optically isolated Digital-to-
Analog Converter (DAC) provides the corresponding process
current or voltage output. A separate alarm circuit controls the
relay contacts.
Flexible transmitter functionality and an optional alarm
combined in a single device, plus convenient reprogrammability,
makes this instrument extremely powerful and useful over a
broad range of applications. This transmitter may be configured
for several input transfer functions (with millivolt input), such as,
proportional, 24-segment linearizer, and square root. Input
averaging may also be combined with any function. The output
may produce a normal (ascending), or reverse (descending)
response. The optional relay of 801T-1500 models may be
configured as a limit alarm with deadband, with latching or non-
latching contacts, and for failsafe or non-failsafe modes. A
programmed relay time delay may be implemented to filter
transients and minimize nuisance alarms.
All IntelliPack modules are designed to withstand harsh
industrial environments. They feature RFI, EMI, ESD, EFT, and
surge protection, plus low temperature drift, wide ambient
temperature operation, and isolation between input, power,
output, and relay contacts. They also have low radiated
emissions per CE requirements. As a wide-range DC-powered
device, the unit may be powered from DC power networks
incorporating battery backup. Additionally, the input power
terminal is diode-coupled, providing reverse polarity protection.
This allows the unit to be connected to redundant power supplies,
or several units to safely share a single DC supply.
Units are DIN-rail mounted and removable terminal blocks
facilitate ease of installation and replacement, without having to
remove wiring. Transmitter power, output, and relay wiring are
inserted at one side of the unit, while input wiring is inserted at
the other side. Connectors are an industry standard screw clamp
type and accept a wide range of wire sizes.
The safe, compact, rugged, reconfigurable, and reliable
design of this transmitter makes it an ideal choice for control
room and field applications. Custom IntelliPack module
configurations are also possible (please consult the factory).
Key IntelliPack 801T Features
•Agency Approvals - cULusa Listed.
•Easy WindowsConfiguration - Fully reconfigurable via
our user-friendly Windows 95or NTIntelliPack
Configuration Program.
•Nonvolatile Reprogrammable Memory - This module has
an advanced technology microcontroller with integrated,
non-volatile, downloadable flash memory and EEPROM.
This allows the functionality of this device to be reliably
reprogrammed thousands of times.
•Convenient Field Reprogrammability - This unit allows
transmitter zero and span calibration, plus alarm setpoint
and deadband adjustments, to be made via module push-
buttons and LED’s, thus facilitating in-field changes without
having to connect a host computer.
Key IntelliPack 801T Features…continued
•Fully Isolated - Input, power, outputs, and relay contacts
are all isolated from each other for safety and increased
noise immunity.
•Wide Input Transfer Function Variability - Proportional,
24 segment linearizer, or square-root transfer function.
Averaging may be applied to any input function. Other
functions are possible—please consult the factory.
•Universal Analog Outputs - Supports process current
output ranges of 0-20mA, 4-20mA, and 0-1mA, and 0-5V or
0-10V outputs. Current outputs drive up to 550, typical.
Voltage outputs include short-circuit protection.
•Normal Or Reverse Acting Output Direction - The output
of this transmitter may be software configured for a normal
(ascending), or reverse (descending) response.
•Flexible DC Millivolt, Thermocouple, or RTD Inputs -
Accepts either DC millivolt, thermocouple, or RTD input
signals. Linearization, lead break detection, and TC
reference junction compensation are included.
•True 4-Wire Kelvin Measurement - Improves performance
of 4-wire RTD measurements.
•Self-Diagnostics - Built-in routines operate upon power-up
for reliable service, easy maintenance, and troubleshooting.
•Wide-Range DC-Powered - This device receives power
over a wide DC supply range and the power terminal is
diode-coupled. This makes this transmitter useful for
systems with redundant supplies, and/or battery back-up.
•Wide Ambient Operation - The unit is designed for reliable
operation over a wide ambient temperature range.
•Hardened For Harsh Environments - The unit will operate
reliably in harsh industrial environments and include
protection from RFI, EMI, ESD, EFT, and surges, plus low
radiated emissions per CE requirements.
•Convenient Mounting, Removal, & Replacement - The
DIN-rail mount and plug-in type terminal blocks make
module removal and replacement easy.
•High-Resolution Precise A/D Conversion - Transmitters
include a high-resolution, low noise, Sigma-Delta Analog to
Digital Converter (-ADC) for high accuracy and reliability.
•High-Resolution Precise D/A Conversion - Transmitter
output is driven with a high-resolution, low noise, Sigma-
Delta, Digital-to-Analog Converter (-DAC) for high
accuracy and reliability.
•LED Indicators - A green LED indicates power. A yellow
status LED will turn on if input signal is out of calibrated
range. A yellow alarm LED indicates when a relay is in
alarm. These LED’s also have other functions in field
program mode. A zero/full-scale LED is used to calibrate
transmitter zero and full-scale values.
•Automatic Self-Calibration - Self-calibration is built-in to
correct for temperature drift of the input circuit.
Alarm Relay Models (801T-1500) Only
•Optional Alarm Functionality - May be programmed for
limit alarms with deadband, latching/non-latching contacts,
and failsafe/non-failsafe operation.
•High-Power SPDT Relay Contacts - Includes a Single-
Pole-Double-Throw (SPDT) electromechanical alarm relay
for switching voltages to 230VAC at currents up to 5A.
•Failsafe or Non-Failsafe Relay Operation - May be
configured for failsafe or non-failsafe relay operation.
•Configurable Setpoint With Deadband - Includes
programmable deadband to help eliminate relay “chatter”
and prolong contact life.