Acteon Satelec Piezotome 2 User manual

User’s manual • Manuel d’utilisation • Manual de uso
Benutzerhandbuch • Manuale d’uso • Handleiding voor de gebruiker
Manual de utilização • Användningshandbok • Εγχειρίδιο χρήσης

2

English 4
Français 26
Español 48
Deutsch 70
Italiano 92
Nederlands 114
Português 136
Svenska 158
Ελληνικά 180
3

1 INTRODUCTION 5
2 WARNINGS 5-6
3 DESCRIPTION 7
Physical description 7
Technical description 7-9
4 INSTALLATION / FIRST USE 10
Unpacking the device 10
Recommendations 10
Installation 10
First use 10
5 IRRIGATION SOLUTION 10
6 IRRIGATION LINES 11
7 ADJUSTMENTS / SETTINGS / MODES
INTERFACE 11
Start up 11
Settings 11
Irrigation function 11
Light function 12
PIEZOTOME function 12
NEWTRON function 12-13
TOOLBOX 13-14
8 SAFETY 14
9 USING THE DEVICE 14-15
10 SHUTTING DOWN THE DEVICE 16
11 ROUTINE MAINTENANCE /
STERILIZATION 16
Irrigation line maintenance 16
Device maintenance 17
Scaler cord maintenance 17
Ultrasonic handpiece maintenance 17
Tip maintenance 17-18
12 MONITORING MAINTENANCE 19
Monitoring 19
Preventive and corrective
maintenance 19
Fuse replacement 19
Operating faults 19-20
13 ELECTROMAGNETIC COMPATIBILITY 21
Electromagnetic emissions 21
Electromagnetic immunity 21
Electromagnetic immunity / Radiofrequency
mobile equipment 22
Recommended separation
distances 23
Cable lengths 24
14 DISPOSAL AND RECYCLING 24
15 MANUFACTURER'S LIABILITY 24
16 REGULATIONS 24
17 SYMBOLS 202-203
18 CUSTOMER RELATIONS 204
Manufacturer identification 204
Addresses of subsidiaries 204-205
The list of accessories referred to in this manual
is non exhaustive. Consult your local dealer for
further details.
4
ENGLISH
CONTENTS

5
I - INTRODUCTION
Congratulations! You have just taken possession
of your PIEZOTOME™2.
Designed by SATELEC®, the PIEZOTOME 2 is a
multifunction device intended for:
• surgical procedures (osteotomy, osteoplasty,
periodontal and implant surgery) with its
PIEZOTOME 2 or PIEZOTOME 2 LED handpiece.
• mechanical ultrasonic treatment (prophylaxis,
periodontics, endodontics) with its NEWTRON®or
NEWTRON LED handpiece (according to option).
In order to get the most out of this high-
technology device and ensure it has a long
working life, please read this manual carefully
before using or maintaining it.
Points preceded by the symbol should be given
particular attention.
II - WARNINGS
CAUTION:
United States Federal Law restricts the use of
this device solely to qualified, trained and
competent dental health practitioners or under
their supervision.
The tips designed for PIEZOTOME 2 and IMPLANT
CENTER 2 are not compatible with PIEZOTOME and
IMPLANT CENTER, and vice versa.
To reduce the risk of accidents, the precautions
stipulated below must be taken:
Device users:
- Use of the PIEZOTOME 2 is restricted solely to
qualified, trained and competent dental health
practitioners in the normal context of their
work.
- If you have received this device by error, please
contact the supplier so that it can be removed.
Interactions:
- PIEZOTOME 2 must not be used if the patient
and/or the operator has a cardiac stimulator
(pacemaker) or any other active implant (e.g. a
cochlear implant).
- The device complies with applicable
electromagnetic compatibility standards. The
user should nevertheless ensure that any
potential electromagnetic interference does
not cause an additional risk (presence of
radiofrequency emitters, electronic devices,
etc.).
- The device is not designed to withstand shocks
delivered by an electric defibrillator.
- Do not attempt to connect to the PIEZOTOME 2
connectors any other accessories than those
supplied by SATELEC.
Electrical connection:
- Your PIEZOTOME 2 must be connected to the
electric power supply by a certified dental
installation technician.
- Warning: To avoid any risk of electric shock, this
device must be connected to an electrical
power supply with a protective ground.
- The electric supply to which the device is
connected must comply with the standards in
force in your country.
- If loss of electrical power during use is likely to
generate an unacceptable risk, the device must
be connected to a suitable power source (e.g.
UPS).

Using the device:
- Do not use the device if it appears to be
damaged or faulty.
- Turn the device OFF before unplugging the
power cord.
- To unplug the power cord, grip the cord plug
and hold the wall socket.
- Never use any other irrigation solution
containers than those intended for suspension
from the supplied brackets.
- The device must only be used with bottles or
bags of physiological saline or sterile water.
- The capacity of the irrigation solution
containers used must not exceed one litre.
- When the device is not to be used for a long
period of time, unplug the device from the
electric supply.
- Do not exert excessive force on the screen.
- Do not move the device during use.
Environment:
- Do not cover the device or obstruct the
ventilation vents.
- Do not immerse the device in liquid, and do not
use it outdoors.
- Do not tilt the device at an angle greater than
5°.
- Do not place the device near a heat source.
- Make sure that the cords are not in a traffic
path.
- The device must be stored in its original
packaging in an appropriate and safe place.
- The device is not designed for use in the
presence of anaesthetic gases or any other
flammable gas.
- Do not expose the device to water vapour, or
splashes.
- Any condensation inside an electrical device
may be dangerous.
- If the device has to be moved from a cool place
to a hot place, it must be allowed to warm up
to the new room temperature before being
used.
- The device is not designed to work near ionizing
radiation.
- Do not insert metal objects into the device (risk
of electric shock, short-circuit or emission of
hazardous substances).
Maintenance:
- Before and after each use, your device must be
disinfected with products recommended by
SATELEC.
- Before each procedure, it is essential to make
sure that the accessories to be used have been
cleaned, disinfected and sterilized.
Accessories:
- The device has been designed and developed
with its accessories in order to ensure maximum
safety and performance.
- Use of accessories from other manufacturers is
a potential hazard for you, your patients or your
device.
Repair:
- Warning: Do not repair or modify the device
without prior authorization from SATELEC.
- Warning: If the device is modified or repaired,
specific checks and tests must be performed to
ensure that the device can still be used safely.
- In the case of a fault, contact the supplier of
your device. Do not use unauthorized repairers,
who might make your device dangerous for you
and your patients.
If in any doubt, contact an approved dealer or our
customer support department:
- www.acteongroup.com
6

III - DESCRIPTION
3. 1 PHYSICAL DESCRIPTION
PIEZOTOME 2 has the following components:
- a control unit (Fig. 1),
- a PIEZOTOME 2 LED handpiece with its cord (Fig. 1-
2) (according to option),
- a NEWTRON LED handpiece (Fig. 1-3) (according to
option),
- a NEWTRON scaler cord (Fig. 1-4) (according to
option),
- a grounded power cord (Fig. 1-5),
- an LCD (Liquid Crystal Display) touch screen
(Fig. 1-6),
- two cord connectors for the two ultrasonic
functions (Fig. 1-7 and 8),
- two peristaltic pump housings (Fig. 1-9),
- two brackets for irrigation solution (Fig. 1-10),
- a multifunction footswitch (Fig. 8).
Located at the back of the unit are:
- 1 power cord socket with ground pin (Fig. 2-1),
- 1 footswitch connector (Fig. 2-2),
- 1 fan (Fig. 2-3),
- 2 bracket holders (Fig. 2-4),
- 1 power switch (Fig. 2-5).
3. 2 TECHNICAL DESCRIPTION
a) LCD touch screen
The LCD touch screen (Fig. 3) is used to define
the settings of your PIEZOTOME 2.
The PIEZOTOME 2 settings are adjusted by applying
reasonable pressure to the screen keys.
Identification of key areas common to all modes
(Fig. 3):
- 4: Reduce the irrigation flow rate
- 5: Increase the irrigation flow rate
- 6: Reduce value
- 7: Increase value
Identification of information display areas
common to all modes (Fig. 3)
- 2: Irrigation flow rate value
- 9: Operating fault indicator
Identification of key areas common to all modes
that also display information (Fig. 3)
- 1: Flush/prime
- 3: Irrigation ON/OFF
- 8: Save, store data
- 10: Select footswitch mode
- 11: Handpiece light ON/OFF
- 12: Select mode (according to connected
handpiece type)
Key areas and displays specific to the selected
mode
PIEZOTOME mode:
Identification of information display areas (Fig. 4)
- 13: Selected program power level
Identification of key areas that also display
information (Fig. 4)
- 14: Select program: D1, D2, D3, D4
NEWTRON mode:
Identification of information display areas (Fig. 5)
- 15: Selected program power level
Identification of key areas that also display
information (Fig. 5)
- 16: Select program: Soft, Medium, High, Boost
Startup page:
Identification of key areas (Fig.6)
- 17: Enter TOOLBOX mode
7

TOOLBOX:
Identification of information display areas (Fig. 7)
- 19: Audio volume symbol
- 20: Screen brightness symbol
- 21: Light off time-out symbol
- 25: Relative audio volume value
- 26: Relative brightness level value
- 27: Selected time-out value
- 28: Software version
Identification of key areas (Fig. 7)
- 18: Reset factory configuration
- 22: Reduce value
- 23: Increase value
- 24: Validate the configuration
b) Control unit back panel
The power receptacle (Fig. 2-1) with its
grounding pin is used to connect the PIEZOTOME 2
to the power supply using a plug-in power cord.
The footswitch connector (Fig. 2-2) is used to
connect the PIEZOTOME 2 to the multifunction
control footswitch.
The fan (Fig. 2-3) protected by a metal grille,
keeps the PIEZOTOME 2 at its optimum performance
level.
The bracket holders (Fig. 2-4) are used to install
the brackets.
The power switch (Fig. 2-5) is used to switch the
device ON or OFF.
c) Control unit sides
The pump housings (Fig. 1-9) are designed to
accommodate SATELEC irrigation line cassettes.
The irrigation cassettes are installed by lifting
the pump covers and inserting the cassettes
horizontally into the cassette housings.
d) Front panel
The two connectors are designed to accept the
connector of the NEWTRON / NEWTRON LED cord or
the connector of the PIEZOTOME 2 / PIEZOTOME 2 LED
handpiece cord.
e) Control footswitch
The access to a large number of functions on the
control footswitch allows the practitioner to work
in a perfectly sterile environment, avoiding the
risk of cross-contamination.
Once the various settings have been adjusted,
the user has no further need to touch the keypad.
Depending on the mode, the control footswitch
buttons have different functions.
PIEZOTOME mode:
Footswitch button definition (Fig. 8)
- 1: Ultrasonics control (ON/OFF or progressive)
- 2: Flush/prime
- 3: Change program (D1 to D4)
- 4: Irrigation ON/OFF
- 5: Select active handpiece
NEWTRON mode:
Footswitch button definition (Fig. 8)
- 1: Ultrasonics control (ON/OFF or progressive)
- 2: Flush/prime
- 3: Change program (Soft to Boost)
- 4: Irrigation ON/OFF
- 5: Select active handpiece
f) Technical characteristics
Manufacturer: SATELEC
Device name: PIEZOTOME 2
Electrical power supply:
- Voltage: 100 VAC to 230 VAC
- Frequency: 50 Hz/60 Hz
- Rated power: 150 VA at 230 VAC
8

9
PIEZOTOME function
Operation:
Intermittent service: 10 min. ON / 5 min. OFF
Output characteristics:
No-load voltage: 250 volts (nominal value without
handpiece)
Min. ultrasonic frequency: 28 kHz
Irrigation flow rate: 10 to 120 ml/min. (nominal
value)
Adjustment in 10 ml/min. steps (nominal value)
Flush flow rate: 120 ml/min. (nominal value)
NEWTRON function
Operation:
Intermittent service: 10 min. ON / 5 min. OFF
Output characteristics:
No-load voltage: 150 volts (nominal value without
handpiece)
Min. ultrasonic frequency: 28 kHz
Irrigation flow rate: 10 to 40 ml/min. (nominal
value)
Adjustment in 1 ml/min. steps (nominal value)
Flush flow rate: 120 ml/min. (nominal value)
Protection
Electrical equipment class: Class 1
Electrical safety class:
Type BF in PIEZOTOME mode
Type BF in NEWTRON mode
Safety devices
Safety shut off if internal malfunction
2 fuses (power receptacle): 5 mm x 20 mm / 2 AT
for 100 VAC to 230 VAC
1 internal fuse not accessible to the user
reference F1: 5 mm x 20 mm – 10 AT / 250 VAC
Overall dimensions
Control unit:
Width: 472.9 mm
Height: 149.5 mm
Height with bracket: 471.1 mm
Depth: 339.9 mm
Weight: 5 kg without accessories
Footswitch:
Width: 311 mm
Depth: 209 mm
Height: 181 mm
Weight: approximately 3.5 kg
LCD screen:
Height: 86 mm
Width: 115 mm
Scaler handpiece cord: 2 000 mm (2 900 mm
optional)
PIEZOTOME 2 handpiece cord: 2 000 mm (2 900 mm
optional)
Temperatures
Operation: +10°C to +40°C
Storage: -20°C to +70°C
Humidity
Operation: 30% to 75%
Storage: 10% to 100% including condensation
Atmospheric pressure
Between 500 hPa and 1060 hPa

IV - INSTALLATION /
FIRST USE
4. 1 UNPACKING THE DEVICE
Upon receipt of the device, check for any damage
caused in transit.
Contact your supplier if necessary.
4. 2 RECOMMENDATIONS
Have your PIEZOTOME 2 connected to the power
electricity supply by a certified dental
installation technician.
The electrical connection of the PIEZOTOME 2 must
comply with the applicable standards in your
country.
Warning:
To avoid any risk of electric shock, this device
must be connected to an electrical power
supply with a protective ground.
4. 3 INSTALLATION
Important:
Do not place the PIEZOTOME 2 close to or on top
of another device.
Do not place the power cord and the footswitch
cord in a wire cover or in a cable gland.
- Place the control unit in position on a fixed
surface, horizontal or with a slope of no more
than 5°.
- Check that the power switch (Fig. 2-5) is in
position O (OFF).
- Connect the power cord to the device.
- Connect the power cord to a power socket with
a ground pin.
- Connect the control footswitch to the
footswitch connector (Fig. 2-2).
- Position the footswitch so that it is easily
accessible for your feet.
- Insert the brackets into the bracket holders
(Fig. 2-4).
- Connect the NEWTRON scaler cord or the
PIEZOTOME 2 handpiece cord to the connector
(Fig. 1-7 and/or 1-8) (according to option).
- Connect the NEWTRON handpiece to the scaler
cord (delivered according to option).
- Check that the device is close enough to the
work area to be used without pulling on the
cords. If it is not, move the device closer.
- Place the NEWTRON or PIEZOTOME 2 handpiece on
the handpiece support.
- Suspend the bottles or bags of physiological
saline or sterile water from the brackets.
- Adjust the position of your device to suit your
angle of view.
4. 4 FIRST USE
Before using the PIEZOTOME 2 for the first time, all
the equipment must be given routine
maintenance and/or sterilized according to the
procedures defined in chapter 11.
V - IRRIGATION SOLUTION
PIEZOTOME 2 is not designed to administer drugs. It
must be used only with bottles or bags of
physiological saline or sterile water with an
individual capacity not exceeding one litre.
VI - IRRIGATION LINES
PIEZOTOME 2 must be used exclusively with SATELEC
irrigation lines.
10

VII - ADJUSTMENTS /
SETTINGS / MODES /
INTERFACE
7. 1 STARTUP
Important:
After stopping the device, wait four seconds
before restarting it.
By default, when the device is switched on, the
right-hand handpiece is selected.
If no handpiece is connected, the device selects
the right side in PIEZOTOME mode.
7. 2 SETTINGS
The settings are saved when each program is
configured and are retrieved whenever that
program is selected.
7. 3 IRRIGATION FUNCTION
a) Flush / prime
The flush / prime function is available on the LCD
screen and on the footswitch.
Press button (Fig. 3-1) or the footswitch
button (Fig. 8-2) to switch on the flush / prime
function.
The flush / prime function stays on as long as the
button is pressed.
The flush can be switched on while the device is
being used.
b) Flow rate adjustment
The irrigation flow rate is adjusted using buttons
and (Fig. 3-4 and 5).
Each time the corresponding button is pressed,
the flow rate is adjusted. Press and hold buttons
and (Fig. 3-4 and 5) to scroll the flow
rate values.
The adjusted value is displayed on the LCD screen
(Fig. 3-2) and stored in the current program.
The flow rate can be adjusted while the device is
being used.
c) Irrigation ON/OFF
The irrigation is switched ON or OFF by pressing
button (Fig. 3-3) or the footswitch button
(Fig. 8-4).
When irrigation is selected, the symbol is
highlighted.
When irrigation is not selected, the symbol
is darkened.
7. 4 LIGHT FUNCTION
The light function is displayed on the LCD screen.
Pressing button (Fig. 3-11) switches the light
function available for the NEWTRON and PIEZOTOME
functions ON or OFF.
The light function has a default light-off time-out
of nine seconds.
Switching off the light function becomes
effective when the time-out has elapsed.
If necessary, you can lengthen the time-out (refer
to 7.7).
11

7. 5 PIEZOTOME FUNCTION
Important:
The PIEZOTOME function settings can be
customized by the user.
The PIEZOTOME function settings can be adjusted
when no PIEZOTOME 2 handpiece is connected to
the device.
The program giving the highest power is D1.
Select the PIEZOTOME function by pressing button
(Fig. 3-12) or by pressing the footswitch
button (Fig. 8-5).
On the LCD screen, the PIEZOTOME tab must be
highlighted.
Select the program to be used by pressing button
, , or (Fig. 4-14) or
using the footswitch button (Fig. 8-3); the
selected program is highlighted on the LCD
screen.
Adjust the programs and the power level
according to the tip data sheet
recommendations.
If necessary, modify the irrigation function (flow
rate value, etc.) (refer to 7.3).
If necessary, modify the program power level (1 to
5) using buttons (Fig. 3-6) and (Fig. 3-7).
The power level of the program concerned is shown
as a number and a bar-graph (Fig. 4-13).
Modify the footswitch type if necessary by
pressing button (Fig. 3-10) to toggle between
ON/OFF and progressive .
Important:
In progressive mode, for programs D1 to D3,
the delivered power ranges between power
level 1 of program D3 and the configuration
defined by the user (program and power level).
In progressive mode, the power delivered by
program D4 ranges only between levels 1 and 5,
according to the value defined by the user.
If necessary, select handpiece light function
ON/OFF by pressing button (Fig. 3-11).
Confirm the new settings by pressing button
(Fig. 3-8).
The saving of the various settings is confirmed by
an audible signal and a visual signal (button
flashing).
Repeat this sequence if necessary in order to
adjust all the programs.
Caution: The tips designed for PIEZOTOME 2 and
IMPLANT CENTER 2 are not compatible with
PIEZOTOME and IMPLANT CENTER, and vice versa.
7. 6 NEWTRON FUNCTION
Important:
A scaler cord and its NEWTRON handpiece must
be connected to the device in order to be able
to configure the various settings.
The NEWTRON function settings can be
customized by the user.
Program Main functions
Very powerful D1
Osteotomy, osteoplastyPowerful D2
Medium D3
Low D4 Soft tissue detachment
12

Select the NEWTRON function by pressing button
(Fig. 3-12) or by pressing the footswitch
button (Fig. 8-5).
On the LCD screen, the NEWTRON tab must be
highlighted.
Select the program for which settings are to be
adjusted by pressing buttons , ,
or (Fig. 5-16) or by pressing the
footswitch button (Fig. 8-3).
Adjust the power level according to the TIPBOOK
recommendations.
If necessary, modify the program power level (1 to
10) using buttons (Fig. 3-6) and (Fig. 3-7).
The power level of the program concerned is shown
as a number and a bar-graph (Fig. 5-15).
Modify the footswitch type if necessary by
pressing button (Fig. 3-10) to toggle between
ON/OFF and progressive.
If necessary, select handpiece light function
ON/OFF by pressing button (Fig. 3-11).
Confirm the new settings by pressing button
(Fig. 3-8). The saving of the various settings is
confirmed by an audible signal and a visual signal
(flashing button).
Repeat this sequence if necessary in order to
adjust all the programs.
7. 7 TOOLBOX
Important:
PIEZOTOME 2 has a TOOLBOX function used to make
various adjustments, including brightness, light
off time-out, audio volume and factory
configuration reset.
To access the TOOLBOX function, switch OFF the
device, wait four seconds, switch the device ON
again, then press button (Fig. 6-17) on the
welcome screen (Fig. 6).
a) Audio volume adjustment
Adjust the audio volume value (Fig. 7-19)
using buttons and (Fig. 7-22 and 23). The
audio volume is adjustable from 0% to 100%.
The relative value of the audio volume is
displayed (Fig. 7-25).
b) Screen brightness adjustment
Adjust the brightness value (Fig. 7-20) using
buttons and (Fig. 7-22 and 23). The
brightness is adjustable from 30% to 100%.
The brightness level setting is displayed (Fig. 7-26).
c) Light off time-out adjustment
Adjust the time-out length value (Fig. 7-21)
by pressing buttons and (Fig. 7-22 and 23).
The length of the adjustable time-out (from nine
seconds to 18 seconds) is displayed (Fig. 7-27).
13

d) Storing the settings
To store the modified settings, press button
(Fig. 7-24). The screen shown in Fig. 6 reappears.
e) Factory configuration reset
Press button (Fig. 7-18) to restore the
factory configuration settings.
The factory configurations for the PIEZOTOME and
NEWTRON functions are effective when button
(Fig. 7-24) is pressed. The screen shown in Fig. 6
reappears.
f) Software version
The version of the software is displayed at the
bottom of the screen (Fig. 7-28).
PIEZOTOME 2 is configured in the factory with the
following settings:
PIEZOTOME mode:
NEWTRON mode:
VIII - SAFETY
PIEZOTOME 2 has a system that detects
malfunctions of the device.
If an internal operation error occurs, symbol 1
(Chap XVII - Symbol) is displayed in the error zone
(Fig. 3-9) and the device beeps four times.
It is recommended that the device be switched
OFF (Fig. 2-5) and switched back ON again.
IX - USING THE DEVICE
Important:
- Do not disconnect the handpiece cords when
the device is switched ON and the footswitch
is pressed.
- Do not disconnect the NEWTRON handpiece
when the device is switched ON and the
footswitch is pressed.
- Do not screw or unscrew tips when the
handpieces are on.
- All accessories must be cleaned, disinfected
and sterilized before use.
- For your safety and that of your patient,
PIEZOTOME 2 must not be used with accessories
other than those supplied or recommended by
SATELEC.
Program Power Irrigation Main functions
Very
powerful D1 360
ml/min.
Osteotomy,
osteoplasty
Powerful D2 360
ml/min.
Medium D3 360
ml/min.
Low D4 360
ml/min.
Soft tissue
detachment
Program Power Irrigation Main functions
Green Soft P=5 15
ml/min. Periodontics
Yellow Medium P=5 15
ml/min. Endodontics
Blue High P=5 15
ml/min. Scaling
Orange Boost P=5 15
ml/min. Loosening
14

- Check the integrity of the device and its
accessories before and after each use in order
to detect any problem.
- If necessary, do not use the device and
replace any faulty component.
The PIEZOTOME 2 is prepared for use as follows:
- Make sure that the device is connected
correctly and properly insulated.
- Switch the power switch (Fig. 2-5) to position I
(ON).
- Suspend the bottles or bags of physiological
saline or sterile water from the brackets.
- Raise the irrigation pump covers.
- Insert the irrigation line cassettes horizontally
into their housings.
- Close the covers.
- Insert the irrigation line perforating pins into
the bottles or pouches of physiological saline
and open the cap on the pins.
- Make any necessary setting adjustments
(irrigation flow rate, power mode, program,
etc.) (refer to chapter 7).
Note: This sequence must be completed under
conditions of controlled asepsis for the safety of
your patients. An operating assistant may be
needed.
Note: The previously-used display settings are
saved automatically when switching between
modes.
PIEZOTOME function
- Connect the PIEZOTOME 2 handpiece cord to one
of the connectors of the device (Fig. 1-7 and 8).
- Screw the selected tip onto the handpiece using
a tip wrench (according to the
recommendations of the clinical booklets).
- Fasten the irrigation line to the cord using line
clips.
- Connect the end of the irrigation line to the
PIEZOTOME 2 handpiece.
- Prime the irrigation circuit by pressing the
flush/prime button on the screen (Fig. 3-1) or
the multifunction footswitch (Fig. 8-2).
- When the irrigation solution reaches the end of
the tip, release the footswitch.
- Use the PIEZOTOME 2 in accordance with good
dental practice.
NEWTRON function
- Check that there are no traces of moisture on
the NEWTRON handpiece or the scaler cord
connections. Remove any moisture by wiping
and by drying with the multifunction syringe fed
with medical-quality filtered air.
- Connect the scaler cord to one of the
connectors of the device (Fig. 1-7 and 8).
- Align the electrical contacts of the NEWTRON
handpiece and the scaler cord, then plug the
handpiece into the cord.
- Screw the selected tip onto the handpiece using
a tip wrench (according to the TIPBOOK
recommendations).
- Fasten the irrigation line to the cord using line clips.
- Connect the end of the irrigation line to the
scaler cord.
- Prime the irrigation circuit by pressing the
flush/prime button on the screen (Fig. 3-1) or
the multifunction footswitch (Fig. 8-2).
- When the irrigation solution reaches the end of
the tip, release the footswitch.
- Use the PIEZOTOME 2 in accordance with good
dental practice.
15

X - SHUTTING DOWN THE
DEVICE
Upon completion of the dental procedure:
- Remove the bottles or bags of physiological
saline or sterile water from the brackets.
- Remove the irrigation line perforating pins from
the bottles or bags.
- Immerse the irrigation line perforating pins in a
vessel containing distilled water.
- Rinse each irrigation line alternately, along with
the handpiece, by running the flush function
until the vessel and the irrigation lines are
completely empty.
- Remove the irrigation line clips.
- Disconnect the single-use irrigation lines from
the handpieces and dispose of them in a safety
container for soiled medical items.
- Disconnect the PIEZOTOME 2 handpiece or the
NEWTRON handpiece and the scaler cord.
- Switch the device OFF (O).
XI - ROUTINE MAINTENANCE/
STERILIZATION
Important:
The device must be switched OFF during
cleaning and disinfection procedures.
The only irrigation lines that can be sterilized
are those identified as sterilizable.
During sterilization, items made of different
metals must not come into contact. Any contact
would establish galvanic couples and cause
localized damage.
To avoid this happening, place the items in
individual sterilizable bags or in a sterilization
box.
To maintain the sterility or asepsis of the
accessories (cords, handpieces, etc.), store
them in sealed bags or containers suitable for
medical practice.
The maintenance and/or sterilization
instructions below must be applied before each
use of the device.
Avoid using cleaning and disinfection agents
containing flammable substances.
If this is not possible, make sure that all the
agent has evaporated and that there is no
combustible matter on the device or its
accessories before it is switched on.
Sterilized items must be allowed to dry and
cool to room temperature before they are
reused.
Before sterilization, check that your autoclave
is clean and that the water used is of adequate
quality.
After each sterilization cycle, remove the
items from the autoclave immediately in order
to minimize the risk of metal corrosion.
11. 1 IRRIGATION LINE MAINTENANCE
Sterile irrigation line
The irrigation lines supplied by SATELEC are
disposable and must always be discarded after
use. Re-use of an irrigation line may result in
contamination of your patients and expose you to
de facto liability. Do not attempt to resterilize
the single-use irrigation lines. Do not attempt to
modify the irrigation lines.
Sterilizable irrigation line
Refer to the user's manual delivered with the
equipment.
16

11. 2 DEVICE MAINTENANCE
Important:
Never use an abrasive cleaning agent on the
device.
Do not use sprays or liquids to clean and
disinfect the PIEZOTOME 2 control unit.
The device, the brackets and the multifunction
footswitch are not sterilizable.
The PIEZOTOME 2 control unit, the control
footswitch and the bracket must always be
cleaned and disinfected after each procedure,
using alcohol, disinfectant agents or disinfectant
wipes for dental practices, such as SEPTOL™ wipes.
Regular monitoring of the PIEZOTOME 2 unit is
necessary in order to detect any problem.
It is important to keep the control unit
ventilation vents clean in order to avoid
abnormal heating.
11. 3 SCALER CORD MAINTENANCE
Important:
Never use an abrasive cleaning agent on the
cords.
The cords must not be immersed or disinfected
using agents containing acetone, chlorine or
bleach.
The cords must not be cleaned in an ultrasonic
bath.
a) Cleaning and disinfection
The cords must always be cleaned, disinfected
and sterilized after each procedure, using
alcohol, disinfectant agents or disinfectant wipes
for dental practices, such as SEPTOL™ wipes.
Disconnect the NEWTRON handpiece from its cord
before cleaning, disinfection and sterilization.
b) Sterilization
The cords can be sterilized in an autoclave under
the following conditions:
- Autoclave: Class B
- Sterilization temperature: 134°C at 2 Bar
- Time at sterilization temperature: 18 minutes
CAUTION: Not all autoclaves can reach 134°C.
Not all autoclaves draw a pre-vacuum. Please
refer to your autoclave manufacturer for
specific sterilization instructions.
11. 4 ULTRASONIC HANDPIECE MAINTENANCE
After each use, the handpiece irrigation circuit
must be rinsed with distilled or demineralized
water for 20 to 30 seconds.
Cleaning and disinfection:
The handpieces must always be cleaned,
disinfected and sterilized after each procedure,
using alcohol, disinfectant agents or disinfectant
wipes for dental practices, such as SEPTOL™ wipes.
For handpiece sterilization, refer to the manual
for each product.
11. 5 TIP MAINTENANCE
a) Lifetime
Tip shape and weight are the determining
characteristics for obtaining maximal efficiency
of the ultrasonic generator.
Careful monitoring of these two characteristics
by the user will ensure that optimal performance
of the device is maintained.
Consequently, you are strongly advised not to
alter the structure of the tips by filing or twisting
them or modifying them in any other way.
Similarly, through normal wear, tip aging leads to
alteration of its characteristics. Always replace a
17

18
tip that has been damaged by wear or by
accidental impact (dropped, bent, etc.).
Use the TIPCARD to check scaler tip wear. Replace
routinely-used tips at least once a year.
b) Pre-disinfection / chemical cleaning
To be carried out immediately after treatment,
wearing thick gloves. Immerse in an ultrasonic
bath (enzymatic, quaternary ammonium base
detergent / disinfectant solution), carefully
respecting the concentration and contact time
recommended by the manufacturer of the
solution.
Use a solution that bears the EC marking or that
is in compliance with any standard required by
the national regulations.
Rinse under running water for at least 30
seconds.
c) Mechanical / chemical cleaning
To be carried out immediately after pre-
disinfection/chemical cleaning, wearing thick
gloves. Scrub the products in a new bath of
enzymatic, quaternary ammonium base
detergent / disinfectant solution, for at least 30
seconds, and in any case until all remaining
traces of contamination have been removed.
Rinse again under running water for at least 30
seconds.
d) Drying
Dry with a clean, non-woven single-use cloth, so
as to remove any traces of liquid. Pack in single-
use sterilization pouches or bags in compliance
with the specifications defined in standard EN ISO
11607-1 or in any equivalent standard required by
the national regulations.
e) Sterilization
The tips must be sterilized individually in an
autoclave, according to the following
parameters:
- autoclave: type B, compliant with standard EN
13060
- sterilization temperature: 134°C
- sterilization stabilization time: 18 minutes
- pressure: 2 Bars min.
f) Storage
Store the sterilized products in a dry place
protected from dust. Before re-use, if the
integrity of the packaging is not certain,
repackage then re-sterilize in accordance with
the defined protocol.
If contamination is visible inside the bag, destroy
the product.
g) Disposal of the product
Dispose of the product in receptacles for waste
materials of healthcare activities involving
infectious risks.
CAUTION: Not all autoclaves can reach 134°C.
Not all autoclaves draw a pre-vacuum. Please
refer to your autoclave manufacturer for
specific sterilization instructions.

XII - MONITORING /
MAINTENANCE
Important:
In the case of a fault, you are advised to
contact the supplier of your device. Do not use
unauthorized repairers, who might make your
device dangerous for you and your patients.
12. 1 MONITORING
Regular monitoring of the device and its
accessories is necessary in order to detect any
insulation fault or damage. Replace them if
necessary.
It is important to keep the control unit
ventilation vents clean in order to avoid
abnormal heating.
12. 2 PREVENTIVE AND CORRECTIVE
MAINTENANCE
PIEZOTOME 2 does not need any preventive
maintenance other than monitoring of the
accessories and routine cleaning/disinfection.
12. 3 FUSE REPLACEMENT
PIEZOTOME 2 is protected by two fuses located in
the power receptacle (Fig. 2-1).
The replacement procedure is as follows:
- Switch the device OFF (position O).
- Unplug the power cord from the electricity
supply.
- Unplug the power cord from the receptacle
(Fig. 2-1). Insert the tip of a flat screwdriver
into the notch above the fuse drawer to release
it.
- Remove the old fuses.
- Replace the old fuses with fuses of the same
type and rating.
- Push the fuse drawer back into its housing until
it clicks to indicate the correct position.
- Plug the power cord into the receptacle
(Fig. 2-1). Plug the power cord into the
electrical supply.
Note:
The device also has an internal fuse not
accessible to the user.
Contact SATELEC after-sales service (see
Chapter 2 - Repairs).
All the information necessary for the repair of
faulty components is made available by SATELEC
to the technical personnel of the SATELEC-
approved dealer network who carry out the work.
12. 4 OPERATION FAULTS
Refer to the table below:
19

20
Fault Possible cause Remedy
Device does not operate
(LCD screen off)
Power cord not plugged in properly. Check the electric socket.
Return to SATELEC after-sales.
Power switch in position O (OFF). Place the power switch in position I (ON).
No power voltage. Call an electrician.
Power receptacle fuse(s) blown. Replace the power receptacle fuses.
Internal fuse blown Return to SATELEC after-sales.
Device does not operate
(LCD screen on) Transmission fault. Switch the device OFF then ON again.
Return to SATELEC after-sales.
No spray
Irrigation solution bag or bottle empty. Replace the bag or bottle of irrigation
solution.
Irrigation off. Press the irrigation ON/OFF button.
Irrigation line clogged. Change irrigation line.
Inadequate spray Incorrect irrigation flow rate adjustment. Adjust the irrigation flow rate.
Power fault
weak tip vibration
Tip worn or distorted. Replace the tip.
Incorrect power adjustment. Refer to the TIPBOOK or clinical booklet for
the necessary information.
Incorrect use: incorrect angle of attack or
inadequate pressure on the tooth.
Refer to the TIPBOOK or clinical booklet for
the necessary information.
Liquid or moisture present between
handpiece and cord. Dry the electrical contacts thoroughly.
No ultrasonic function
Incorrect tip tightening. Retighten the tip with the wrench.
Faulty connector contact. Clean the connector contacts.
Handpiece cord wire cut. Return to SATELEC after-sales to replace
the cord.
No light
Handpiece light ring absent. Install the light ring.
Faulty light ring. Replace the light ring.
Faulty light ring connector contacts. Clean the light ring connector contacts.
Light ring the wrong way around. Position the light ring according to the
foolproofing stud.
Faulty handpiece and/or cord connector
contacts.
Clean the handpiece and/or cord connector
contacts.
Other. Contact your installer-fitter.
Leak between the
NEWTRON handpiece and
the handpiece cord
Worn handpiece 1.15x1 seal Replace the seal (kit F12304).
Liquid leak in the
irrigation pump
Broken tube in the irrigation line
cassette. Replace the irrigation line.