
DOCUMENT: MAN-006-01
Effective: 28-Apr-2020
CONTROL NUMBER: CO-2020-382 Page 9 of 69
1.4 VentFree Contraindications
VentFree should NOT be used for the following situations or patients with:
●Do not use with demand type implanted pacemaker or defibrillator.
●Do not use electrical stimulation over pregnant uterus.
●Do not use on a patient with recent abdominal surgery with open abdominal wounds.
●Do not use on open or damaged skin.
●Do not use on patients under 18 years of age.
1.5 VentFree Warnings
Regional Use – ALWAYS use the AC/DC converter with the corresponding regional AC Inlet
plug .
Use Setting- This device should only be used in a hospital ICU setting.
Cleaning -Clean device with disinfectant wipe between uses (after use-before storage and
before use). Only use approved chemicals for cleaning/disinfection (refer to Section 6).
Patient Supervision - Clinicians should regularly monitor patient for adequate stimulation level
throughout the treatment.
Multiple Patient Use -If the device is used on more than one patient, the operator should follow
facility infection control procedures.
Prescription and Use – This device should only be prescribed and operated by a clinician that is
familiar with the precautionary measures and operational functions associated with the unit.
Electrode Placement- Electrodes should only be placed on the abdomen in compliance with the
directions for use (refer to Section 2)
Cutting Electrodes - Electrodes should not be cut other than as indicated by the dotted lines in
the back of the electrodes.
Reposition of Electrodes- Electrodes should not be repositioned during treatment sessions.
Throat Stimulation – Severe spasm of the laryngeal and pharyngeal muscles may occur when
the electrodes are placed across the throat or mouth. This may be strong enough to close off the
airway or cause breathing difficulty.
Transthoracic Stimulation – Do not apply electrical stimulation transthoracically (through the
chest area) in that the introduction of electrical current into the heart may cause cardiac
arrhythmias.
Other Electrode Placement -Stimulation should not be applied across or through the head,
directly on the eyes, covering the mouth, on the front of the neck(especially the carotid sinus), or
placed on the chest, upper back or crossing over the heart.
Electrical Shock – To prevent electrical shock, disconnect the unit from the power source and
remove the battery before attempting any maintenance procedures.
Accessories – Use only accessories that are specifically designed for this device. Do not use this
device with accessories manufactured by companies other than Liberate Medical. Liberate
Medical, LLC is not responsible for any consequence resulting from using products
manufactured by other companies. The use of accessories, transducers, or cables other than those
specified by the manufacturer, may result in increased emissions or decreased immunity of the
VentFree muscle stimulator.