manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. ADC
  6. •
  7. Measuring Instrument
  8. •
  9. ADC Prosphyg 760 User manual

ADC Prosphyg 760 User manual

IB p/n 9360N-00 rev 13 (08/31/2021) Printed in the U.S.A.
Low-Level Disinfection
Prepare Enzol enzymatic detergent according to the manufacturer’s instructions. Spray
detergent solution liberally onto cuff and use a sterile brush to agitate the detergent so-
lution over entire cuff surface for five minutes. Rinse continuously with distilled water
for five minutes. To disinfect, first follow the cleaning steps above, then spray cuff with
10% bleach solution until saturated, agitate with a sterile brush over entire cuff surface
for five minutes. Rinse continuously with distilled water for five minutes. Wipe off excess
water with sterile cloth and allow cuff to air dry.
CAUTION: Do not iron cuff.
CAUTION: Do not heat or steam sterilize cuff.
Manometer Quality Control
A serial number and lot number are
automatically assigned to every aneroid
during manufacturing, ensuring every item
is "controlled."
The serial number can be located on the
faceplate of each aneroid (Figure 4).
The lot number is located on the box
end label (Figure 5).
Standards
ANSI/AAMI/ISO 81060-1 • EN/ISO 81060-1
Disposal
When your sphygmomanometer or any of its parts have reached their end of life,
please be sure to dispose of them in accordance with all regional and national
environmental regulations. Devices that have become contaminated should be
disposed of in accordance with all local ordinances and regulations.
Warranty
ADC®warrants its products against defects in materials and workmanship under
normal use and service as follows:
•Warranty service extends to the original retail purchaser only and commences
with the date of delivery.
•See end panel label for package contents and to determine warranty duration.
Warranty duration is as follows:
Series Manometer Inflation System
760, 768, 769 20 Years 3 Years
780, 785, 790, 770, 775 10 Years 1 Year
What Is Covered: Replacement of parts and labor.
What Is Not Covered: Transportation charges. Damages caused by abuse, misuse,
accident, or negligence. Incidental, special, or consequential damages. Some states
do not allow the exclusion or limitation of incidental, special, or consequential
damages, so this limitation may not apply to you.
Implied Warranty: Any implied warranty shall be limited in duration to the terms
of this warranty and in no case beyond the original selling price (except where
prohibited by law). This warranty gives you specific legal rights and you may have
other rights which vary from state to state.
To Obtain Warranty Service: Send item(s) postage paid to: ADC, Attn: Repair
Dept., 55 Commerce Dr., Hauppauge, NY 11788. Please include your name and
address, phone no., proof of purchase, and a brief note explaining the problem.
For Australian Consumers
Our goods come with guarantees that cannot be excluded under the Australian
Consumer Law. You are entitled to a replacement or refund for a major failure and
compensation for any other reasonable foreseeable loss or damage. You are also
entitled to have the goods repaired or replaced if the goods fail to be of acceptable
quality and the failure does not amount to a major failure.
Serial Number
ProsphygTM Aneriods
Instructions for Use
Cuff Cleaning and Disinfecting
Use one or more of the following methods and allow to air dry:
• Wipe with mild detergent and water solution (1:9 solution). Rinse.
• Wipe with Enzol per manufacturer’s instructions. Rinse.
• Wipe with .5% bleach and water solution. Rinse.
• Wipe with 70% isopropyl alcohol.
• Launder with mild detergent in warm water, normal wash cycle.
Remove bladder first. Cuff is compatible with 5 wash cycles.
(Figure 5)
LATEX
(Figure 4)
FOR QUESTIONS, COMMENTS, OR SUGGESTIONS CALL TOLL FREE:
1-800-ADC-2670 or
OR VISIT www.adctoday.com/feedback
To register your product, visit us at
www.adctoday.com/support/warranty-registration
Product Registration
ADC
55 Commerce Drive
Hauppauge, NY 11788
U.S.A.
SC Cattus SRL
Aleea Carpati nr.53 Sc. B Ap. 64
Târgu-Mures, România, EU
Inspected, assembled
and packaged in the U.S.A.
tel: 631-273-9600
1-800-232-2670
fax: 631-273-9659
www.adctoday.com
email: info@adctoday.com
1434
9360N-00 rev 13.qxp_Layout 1 9/2/21 3:30 PM Page 1
Thank you for choosing an ADC®ProsphygTM Aneroid
sphygmomanometer.
Device Description and Intended Use
An ADC®aneroid sphygmomanometer is used by professional healthcare providers
and individuals trained in the auscultatory blood pressure technique to determine
systolic and diastolic blood pressure in humans.
Contraindications
Aneroid sphygmomanometers are contraindicated for neonate use. Do not use with
neonatal cuffs or neonate patients. Review the size chart (below) for proper limb range.
Symbol Definitions
The following symbols are associated with your ADC sphygmomanometer.
General Warnings
A warning statement in this manual identifies a condition or practice
which, if not corrected or discontinued immediately could lead to patient
injury, illness, or death.
WARNING:
Do not allow a blood pressure cuff to remain on patient for more than 10
minutes when inflated above 10 mmHg. This may cause patient distress, disturb blood
circulation, and contribute to the injury of peripheral nerves.
WARNING:
If luer lock connectors are used in the construction of tubing, there is a
possibility that they might be inadvertently connected to intra-vascular fluid systems,
allowing air to be pumped into a blood vessel. Immediately consult a physician if this occurs.
WARNING: Safety and effectiveness with neonate cuff sizes 1 through 5 is not established.
WARNING: If this equipment is modified, appropriate inspection and testing must be
conducted to ensure its continued safe use.
WARNING: Do not apply cuff to delicate or damaged skin. Check cuff site frequently
for irritation.
WARNING: Only use the cuff when the range markings indicated on the cuff show that
the proper cuff size is selected, otherwise erroneous readings may result.
WARNING: Allow space between patient and cuff. Two fingers should fit in this space
if the cuff is correctly positioned.
WARNING: Do not apply cuff to limbs used for IV infusion.
WARNING: Patient should remain still during measurement to avoid erroneous readings.
WARNING: When using with an infant or child cuff, extra care must be taken to
prevent over-inflation. With smaller cuffs (infant or child) the cuff can inflate to over
300mmHg with just two full compressions of the bulb. To prevent discomfort or injury
to the patient and damage to the instrument, bulb should only be partially squeezed,
so that each “stroke” inflates the cuff in 40mmHg to 60mmHg increments until inflated
to the desired level.
CAUTION:
To obtain the greatest accuracy from your blood pressure instrument, it is
recommended that the instrument be used within a temperature range of 50°F to
104°F (10°C to 40°C), with a relative humidity range of 15%-85% (non-condensing).
CAUTION: Extreme altitudes may affect blood pressure readings. Your device has
been designed for normal environmental conditions.
Operation of Pocket Aneroids
This booklet contains operating and
maintenance information for pocket aneroid
sphygmomanometers. Please read and retain.
Your pocket aneroid
sphygmomanometer consists of
an aneroid manometer (gauge),
complete inflation system,
(latex-free inflation bladder, squeeze bulb,
and the valve), a zippered carrying case,
and operating instructions.
Most models are preassembled and ready for use.
In units requiring assembly, the bulb and valve should connect
to the tube closest to the Index Line. The gauge connects to the remaining tube.
Apply a small amount of alcohol or soapy water to tubing if needed.
Measurement Procedure
Patient Position
The patient should sit or lie comfortably. The arm should be fully supported on a flat
surface at heart level. (If the arm’s position varies, or is not level with the heart, meas-
urement values obtained will not be consistent with the patient’s true blood pressure.)
When seated, the patient should have their back and arm supported, and their legs
should not be crossed. The patient should relax prior to measurement comfortably
for five minutes and should refrain from talking or moving during measurement.
Observer should view manometer in a direct line to avoid “parallax error.”
Apply the Cuff
Nylon cuffs are specially designed to
promote the precisely accurate
determination of blood pressure.
Index and range markings ensure use
of the correct cuff size. The artery mark
indicates proper cuff positioning.
Place the cuff over the bare upper arm with
the artery mark positioned directly over the
brachial artery. The bottom edge of the cuff should be positioned approximately one
inch (2-3cm) above the antecubital fold. Wrap the end of the cuff not containing
the bladder around the arm snugly and smoothly and engage adhesive strips
(Figure 1).
NOTE: If the unit is
equipped with a calibrated
nylon cuff, featuring Index
and Range markings, a cor-
rect fit may be verified by
checking that the Index Line
falls between the two Range
Lines. (Figure 2).
Inflate the Cuff
Close the valve by turning thumbscrew clockwise.
Palpate the radial artery while inflating the cuff. Be sure to inflate cuff quickly by
squeezing bulb rapidly.
Inflate cuff 20-30 mmHg above the point at which the radial pulse disappears.
NOTE: Cuff pressure range is 0 mmHg to 300 mmHg.
Position the Stethoscope
Position the chestpiece in the antecubital space below the cuff, distal to the brachium.
Do not place chestpiece underneath the cuff, as this impedes accurate measurement.
Use the bell side of a combination stethoscope for clearest detection of the low pitched
Korotkoff (pulse) sounds.
Deflate the Cuff
Open the valve to deflate the cuff gradually at a rate of 2-3 mmHg per second.
Measurement
Record the onset of Korotkoff sounds as the systolic pressure, and the disappearance
of these sounds as diastolic pressure. (Some healthcare professionals recommend
recording diastolic 1 and diastolic 2. Diastolic one occurs at phase 4.)
NOTE: It is recommended that K4 be used in children aged 3 to 12, and K5 should be
used for pregnant patients unless sounds are audible with the cuff deflated, in which
case K4 should be used. K5 should be used for all other adult patients.
After measurement is completed, open valve fully to release any remaining air in the
cuff. Remove cuff.
Care and Maintenance
STORAGE
Pocket Gauge: After measurement, fully exhaust cuff then wrap cuff around gauge
and bulb and store in zippered carrying case.
NOTE: This product will maintain the safety and performance characteristics
specified at temperatures ranging from 50°F to 104°F (10°C to 40°C) at a relative
humidity level of 15% to 85%.
This device can be safely stored at temperatures ranging from -4°F to 158°F
(-20°C to 70°C) with a relative humidity of 85% or below.
Manometer: Your pocket aneroid gauge requires minimal care and maintenance.
The manometer may be cleaned with a soft cloth but should not be dismantled under
any circumstances.
Should the indicator needle of the manometer rest
outside the oval calibration mark (Figure 3), then
the manometer must be re-calibrated to within ±3
mmHg when compared to a reference device that
has been certified to national or international
measurement standards. A manometer whose
indicator needle is resting outside of this mark
is NOT acceptable for use.
In the event that the gauge is ever in need of calibration, simply return for service.
Damaged or broken parts will be replaced as needed at a minimal charge. Refer to the
warranty for specific details of warranty coverage.
The manufacturer recommends a calibration check every 2 years.
Cuff Size Limb Range
Inches CM
Infant 7I 3.5 to 5.5 9 to 14
Child 9C 5.1 to 7.6 13 to 19.5
Sm. Adult 10SA 7.4 to 10.6 19 to 27
Adult 11A 9 to 15.7 23 to 40
Lg. Adult 12X 13.3 to 19.6 34 to 50
Thigh 13T 15.7 to 25.9 40 to 66
Adcuff Series Size Chart
Range Lines
Index Lines
NOTE: Store gauge with
valve in full exhaust position.
(Figure 3)
(Figure 2)
Cuff
Index Line
Hang
Tab
Gauge
Artery Mark
Bulb & Valve
Bladder &
Tubing
Left Arm
Tube
(Figure 1)
Symbol Definition
Important Warning/Caution
Not made with natural rubber latex
Circumference Size
Meets essentials requirements of
European Medical Device Directive
93/42/EEC
Consult instructions for use
LATEX
Symbol Definition
Authorized European Representative’s
Information
Manufacturer’s Information
Temperature Limit
Humidity Limitation
Medical Device
Cuff Size Limb Range
Inches CM
Sm. Adult 10SA 6.3 to 9.8 16 to 25
Adult 11A 9 to 13.3 23 to 34
Lg. Adult 12X 11.8 to 15.7 30 to 40
780-790 Series Size Chart
9360N-00 rev 13.qxp_Layout 1 9/2/21 3:30 PM Page 2

This manual suits for next models

7

Other ADC Measuring Instrument manuals

ADC Diagnostix Palm Style Aneroid Maintenance manual

ADC

ADC Diagnostix Palm Style Aneroid Maintenance manual

ADC advantage connect 6024N User manual

ADC

ADC advantage connect 6024N User manual

ADC DIAGNOSTIX 922 series Maintenance manual

ADC

ADC DIAGNOSTIX 922 series Maintenance manual

ADC MGA3000 series User manual

ADC

ADC MGA3000 series User manual

ADC Diagnostix 750 Series User manual

ADC

ADC Diagnostix 750 Series User manual

ADC 774 Series User manual

ADC

ADC 774 Series User manual

ADC advantage connect 6024N User manual

ADC

ADC advantage connect 6024N User manual

ADC DIAGNOSTIX 932 series Maintenance manual

ADC

ADC DIAGNOSTIX 932 series Maintenance manual

ADC GNOSTIX 922 Maintenance manual

ADC

ADC GNOSTIX 922 Maintenance manual

Popular Measuring Instrument manuals by other brands

GREISINGER GMH 3831 operating manual

GREISINGER

GREISINGER GMH 3831 operating manual

Rayleigh Instruments RI-D384-C operating instructions

Rayleigh Instruments

Rayleigh Instruments RI-D384-C operating instructions

Gossen MetraWatt METRA HIT 22S operating instructions

Gossen MetraWatt

Gossen MetraWatt METRA HIT 22S operating instructions

Jinko JK2830 Operation manual

Jinko

Jinko JK2830 Operation manual

General DSM403SD user manual

General

General DSM403SD user manual

USA Measurements US-4011 User and technical manual

USA Measurements

USA Measurements US-4011 User and technical manual

evertz 3405 manual

evertz

evertz 3405 manual

Observint RATTLER2 quick start guide

Observint

Observint RATTLER2 quick start guide

Teac HD-P2 Service manual

Teac

Teac HD-P2 Service manual

ABB GLA133 Series user manual

ABB

ABB GLA133 Series user manual

Simex SRT-N118-XA user manual

Simex

Simex SRT-N118-XA user manual

Tegam 730A Operation manual

Tegam

Tegam 730A Operation manual

Hanna Instruments HI 9812-0 instruction manual

Hanna Instruments

Hanna Instruments HI 9812-0 instruction manual

Oven TPM138 Passport and operaton manual

Oven

Oven TPM138 Passport and operaton manual

Heyuan Intelligence DZ81-MS3UI7E3 user manual

Heyuan Intelligence

Heyuan Intelligence DZ81-MS3UI7E3 user manual

Sylvac S25 quick start

Sylvac

Sylvac S25 quick start

MITECH MT200 user manual

MITECH

MITECH MT200 user manual

Myron L PoolPro PS6FCE quick start guide

Myron L

Myron L PoolPro PS6FCE quick start guide

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.