Device Description and Intended Use
ADC®an roid sphygmomanom t rs ar us d by prof ssional h althcar provid rs and individuals
train d in auscultatory blood pr ssur t chniqu to d t rmin systolic and diastolic blood pr ssur
in humans and animals.
A warning statement in this manual identifies a condition or practice which, if not
corrected or discontinued immediately could lead to patient injury, illness, or death.
• CAUTION: F d ral law r stricts this d vic to sal by or on th ord r of a physician or lic ns d
h althcar practition r.
• WARNING:
If lu r lock conn ctors ar us d in th construction of tubing, th r is a possibility
that th y might b inadv rt ntly conn ct d to intra-vascular fluid syst ms, allowing air to b pump d
into a blood v ss l. Imm diat ly consult a physician if this occurs.
• WARNING:
Do not allow a blood pr ssur cuff to r main on pati nt for mor than 10 minut s
wh n inflat d abov 10 mmHg. This may caus pati nt distr ss, disturb blood circulation, and
contribut to th injury of p riph ral n rv s.
• WARNING: Saf ty and ff ctiv n ss with n onat cuff siz s 1 through 5 is not stablish d.
• WARNING:
Do not twist conn ctors wh n joining th blood pr ssur cuff to tubing as this
may caus th conn ctors to stick and mak changing th blood pr ssur cuff difficult.
• WARNING:
Wh n using with an infant or child cuff, xtra car must b tak n to pr v nt ov r-
inflation. With small r cuffs (infant or child) th cuff can inflat to ov r 300mmHg with just two full com-
pr ssions of th bulb. To pr v nt discomfort or injury to th pati nt and damag to th instrum nt,
bulb should only b partially squ z d, so that ach “strok ” inflat s th cuff in 40mmHg to
60mmHg incr m nts until inflat d to th d sir d l v l.
• WARNING:
This product may contain a ch mical known to th stat of California to caus canc r,
birth d f cts, or oth r r productiv harm.
Parts and Assem ly
This booklet contains operating and maintenance infor-
mation for the DIAGNOSTIX™922 series desktop mercu-
rial blood pressure instrument. Please read and retain.
Your DIAGNOSTIX™922 consists of a main unit with
4mm calibrated unbreakable plastic cartridge tube, in-
flation system (which includes the ADCUFF™nylon
cuff with Size Guide™ marking system, latex-free infla-
tion bladder, bulb, and the ADFLOW™ valve), luer con-
nector, and operating instructions.
Cartridge Tu e
Locking
Mechanism
Bul &
Valve
Main Unit
Reservoir
Cuff