ADTEMP 424 User manual

PLEASE NOTE:
THIS MEDICAL INSTRUMENT MUST BE
USED ACCORDING TO INSTRUCTIONS
TO ENSURE ACCURATE READINGS.
Questions?
Call ADC toll free at 1-800-232-2670
Adtemp
TM
Digital Ear
Thermometer
424 Digital Ear Thermometer
Instruction Manual
IB p/n 93-424-00 rev 1 Printed in China
Dist. by: ADC®
Hauppauge, NY 11788
Inspected in the USA
Made in China
Printed in China
tel: 631-273-9600
toll free: 1-800-232-2670
fax: 631-273-9659
www.adctoday.com
email: info@adctoday.com
0044
Onbo Electronic (Shenzhen) Co., Ltd.
No. 497, Ta Laneg Nan Road, Ta Laneg Street
Baoan District, Shenzhen, China
Microlife AG, Espenstrasse
139, 9443 Widnau, Switzerland

19
AdtempTM 424 Digital Ear Thermometer
1. Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3
2. Important Saf ty Instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-4
3. Product D scription . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4
4. How to M asur Body T mp ratur . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5
5. How to R ad th Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5
6. How to Apply a N w Prob Cov r . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6
7. Dir ctions for Us . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-8
8. Changing from Fahr nh it to C lsius . . . . . . . . . . . . . . . . . . . . . . . . . . . .8
9. Error M ssag s . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9
10. Cl aning and Storag . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10
11. R placing th Batt ry . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10
12. Warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10
13. T chnical Sp cifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11
14.Contact Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .19
14. Contact Information / Contact Información
To register your pro uct an obtain further etaile user
information about our pro ucts an services visit us at:
www.adctoday.com
an follow the links.
For questions, comments, or suggestions
call us toll free at:
1-800-232-2670
Para registrar su pro ucto y obtener información aún más
etalla a e usuario sobre nuestros pro uctos y los servicios nos visitan en:
www.adctoday.com
y sigue los lazos.
Para preguntas, los comentarios, o las sugerencias llaman nosotros
tocamos libertamos en:
1-800-232-2670
Am rican Diagnostic Corporation
55 Commerce Drive, Hauppauge, New York 11788
Telephone: 631-273-9600 • Fax: 631-273-9659
Email: info@a cto ay.com

1. INTRODUCTION
Congratulations on your purchase of the ADC®A tempTM 424 Tympanic Digital Thermometer. ADC®profes-
sional iagnostic pro ucts are the instruments of choice in hospitals an physician's offices throughout the
worl , where accuracy an epen ability are critical.
Now, you too can enjoy the benefits of ADC®engineering an quality in the home. This feature rich instru-
ment was esigne to simplify use an eliver consistent, epen able results.
Your ADC®A tempTM 424 Digital Tympanic Thermometer is an automatic measuring evice that
offers the A vantages of Multiple Use (Wi e Range Measurement). The A tempTM 424 offers a wi e
range measurement feature ranging from 89.6°F ~ 108°F (32°C ~ 42.2°C); the pro uct can be use
as an ear thermometer to measure bo y temperature.
Quick M asur m nt
The innovative infrare technology allows measurement of ear temperature in only 1 secon .
Accurat and r liabl
Due to the unique probe assembly construction, the a vance infrare sensor, an the
complete calibration process this unit can offer a very accurate an reliable ear
temperature measurement.
G ntl and Easy to Us
• Special ergonomic esign enables simple an easy use of the thermometer.
• This thermometer can be use without interference to aily lifestyle. A measurement can
be taken even while a chil is sleeping.
• This thermometer is convenient to use for chil ren.
• This thermometer is less threatening to a chil than a rectal thermometer an more
pleasant to use than an oral thermometer.
Auto-Display M mory
The pro uct isplays the last rea ing automatically for 3 secon s when the unit is switche ON.
Saf and Hygi nic
• No risk of broken glass or mercury ingestion
• Completely safe for use on chil ren
• Disposable probe covers make A tempTM 424 completely hygienic
2. Important Saf ty Instructions
• Never use the thermometer for purposes other than bo y or surface temperature measurement.
Please follow the safety precautions when using on chil ren.
• Always use the thermometer with a new un amage Probe Cover for each measurement to prevent
infection. R adings tak n without prob cov rs will not b accurat . Only ADC®brand
prob cov rs can nsur that you g t an accurat m asur m nt from th 424
Digital Ear Th rmom t r.
• N v r imm rs th th rmom t r into wat r or oth r liquids. (not wat rproof) For
cl aning and disinf cting pl as follow th instructions in s ction 10, "Cl aning
and Storag ”.
18
us ag nt s limpiador s abrasivos, d s ngrasant s o
b nc no para limpiarlo ni sum rja l t rmóm tro n agua
o n cualqui r otro líquido limpiador. Tenga cui a o e no
romper la superficie e la pantalla. Saque la pila el termómetro si no
lo va a utilizar urante un largo perio o e tiempo para evitar años
causa os por una pila que haya per i o líqui o.
Nota: Cera e oi o ó sucio en el lente el sensor resultará en una
me i a no precisa. Limpie el lente según se in ica.
11. Sustitución d la Bat ría
El termómetro instantáneo e oi o funciona con una
pila e litio, tipo CR2032, que viene inclui a en ca a
uni a . Cuan o en el isplay aparezca el símbolo e la
batería , uste ebe cambiarla. Utilice un estornilla or
a ecua o para quitar el tornillo que sujeta la tapa protectora
e la pila, quite la pila agota a y ponga una nueva.
No a juste emasia o el turnillo cuan o instale la
tapa protectora.
12. Garantía
Este instrumento e me ición e alta cali a tiene una garantía e 2 años contra efectos e fabricación a
partir e la fecha e compra. Las reclamaciones e garantía eben efectuarse urante el perio o e vigencia
e la garantía.
Este pro ucto ha si o fabrica o con el máximo cui a o e acuer o con los están ares internacionales e
cali a . Si a pesar e ello, uste tiene motivos e queja, por favor, acu a a la farmacia o parafarmacia on e
a quirió el pro ucto y allí le informarán sobre cómo contactar con el istribui or ADC®e su país.
Los años causa os por un uso incorrecto no serán motivo e reclamación e la garantía. La pila y el
envase están exclui os e la garantía. Reclamaciones por estos motivos, inclui as las reclamaciones por
años que an exclui as.
13. Caract rísticas Técnicas
Mod lo: Termómetro Infrarojo Digital 421
Intérvalo d M dición: 32.0 F a 212.0 F (0 C~100 C)
Pr cisión: Laboratorio: : ±0.4 F: 89.6~108.0 F (±0.2 C: 32.0~42.2 C)
±2 F: 32.0~89.5 F, 108.1~212.0 F (±1 C: 0~31.9 C, 42.3~100.0 C)
Pantalla: De Cristal Líquido LCD de unidad de medición 0.1 C (0.1 F)
Alarma: a. Al poner en marcha el termómetro se oye 1 “bip” corto.
b. Una vez finalizada la medición se oye 1 “biiip” largo.
c. Error del sistema o mal funcionamiento: se oyen 3 “Bips” cortos.
M moria: Muestra automáticamente el valor de la última medición.
Luz Nocturna: a. El display se ilumina durante 4 segundos al poner en marcha el termómetro.
b. El display se ilumina de nuevo durante 5 segundos cuando la medición
ha terminado.
T mp ratura op rativa: 10 C a 40 C (50.0 F a 104 F)
T mp. d Cons rvación y Transport : -25 C a +55 C (-13 F a 131 F)
Apagado automático: Aprox. 1 minuto después de haber realizado la medición.
Bat ría: CR2032 BATTERY (X1) - 1000 mediciones como mínimo.
Tamaño: 141mm (largo) x 26 mm (ancho) x 20 mm (alto)
P so: 45g (con pila), 40g (sin pila)
Normativa d r f r ncia: Manufacturado para cumplir con las normativas PrEN12470-5 y ASTM E-1965
Empuje contre los resortes
para re-insertar la bateria.
Resortes
Empuje
para
Pry Out
3

•Keep the instrument an the probe covers away from irect exposure to the sun an keep it in a
ust-free, ry area at a temperature between 10° - 40° (50°F - 104°F).
• If the instrument is not going to be use for a prolonge perio the battery shoul be remove .
• Ensure that chil ren o not use the instrument unsupervise ; some parts are small enough to be
swallowe .
• Do not use the thermometer if there are signs of amage on the probe or on the
instrument itself. If amage , o not attempt to repair the instrument!
• Earwax in the ear canal may cause a lower temperature rea ing. Make sure subject's ear canal is
clean to ensure an accurate rea ing. Use caution when cleaning ears.
• The ADC®424 Digital Ear Thermometer consists of high quality precision parts. Protect it from
severe impact an shock. Keep the unit in its storage case when not in use.
WARNING:
• Please keep the probe covers out of the reach of chil ren. They may present a choking hazar .
• Use of this thermometer is not inten e as a substitute for consultation with your physician.
• This thermometer is not waterproof! NEVER immerse into liqui s!
• Cleaning the ear canal using improper techniques can harm the inner ear. Follow octors
instructions for proper cleaning.
•
This pro uct may contain a chemical known to the state of California to cause cancer, birth efects, or other
repro uctive harm.
3. Product D scription
(1) Probe Cover
(2) Probe
(3) Start button/
mo e change
(4) Display
(5) ON/OFF button
(6) Battery Cover
(7) Storage case/
probe cover applicator
4 17
2) Niños menores e 3 años con el sistema inmunológico comprometi o y para los cuales resulta crítica la
presencia o ausencia e fiebre.
3) Cuan o el usuario utiliza el termómetro por primera vez y hasta que no se familiarice con su uso y
obtenga resulta os fiables.
8. Cambio d Fahr nh it a C lsius y vic v rsa
El termómetro instantanéo e oi o pue e mostrar los resulta os e la me ición en gra os Fahrenheit o Celsius. Para
cambiar la pantalla entre ºC y ºF, simplemente apague el termómetro, presione y mantenga presiona o urante 5
segun os el botón e inicio (START). Pasa o este tiempo aparecerá intermitente en el isplay la escala e me ición
(ºC o ºF) . Presionan o e nuevo el botón e inicio po remos pasar e una escala a otra. Una vez escogi a la escala,
espere 5 segun os más hasta que el termómetro se pone punto a para iniciar automáticamente la me ición.
9. M nsaj s d Error
Display / Probl ma Significado Display Posibl causa y r m dio
10. Limpi za y Cons rvación
Use un paño o algo ón impregna o e alcohol (Isopropílico 70%) para limpiar la carcasa exterior el ter-
mómetro y el extremo e me ición. Asegúrese e que no entre líqui o en el interior el termómetro. Nunca
(1)
(2)
(3)
(4)
(5)
(6)
(7) Probe Cover
Applicator Hole
Listo para m dir
Temperatura me i a
emasia o alta
Aparece “H” cuan o la temperatura me i a
es superior a 42.2 °C o 108.0 °F.
Aviso! Fi br alta r qui r
at nción médica inm diata.
Temperatura me i a
emasia o baja
Aparece “L” cuan o la temperatura me i a
es inferior a 32 °C o 89.6 °F.
Temperatura ambiente
emasia o alta
Aparece „H“ junto „“ cuan o
la temperatura ambiente es
superior a 40 °C o 104 °F.
Temperatura ambiente
emasia o baja
Aparece „L“ junto „“ cuan o la
temperatura ambiente es inferior a 10.0°C
o 50.0 °F.
Pantalla e Error e
Funcionamiento
Error e funcionamiento el
Sistema.
Display en blanco Por favor, compruebe si la pila está coloca-
a correctamente. Compruebe también la
polari a e las pilas (<+> y <->).
In icación e fallo e las
pilas
Si sólo aparece permanentemente el
símbolo e la batería en el isplay, la
pila ebe ser inme iatemente substitui a.

16 5
4. How to M asur Body T mp ratur
The ADC®424 Digital Ear Thermometer measures the heat ra iate from the ear rum an the sur-
roun ing tissue. This energy is collecte through the lens an converte to a temperature value. The
measure rea ing obtaine irectly from the ear rum (Tympanic Membrane) can ensure the most
accurate bo y temperature. Measurements taken from the surroun ing tissue of the ear canal generate
lower rea ings an may result in mis iagnosis of a fever.
To avoid an inaccurat m asur m nt:
• Use ONLY ADC®probe covers, as this unit is calibrate to work with probe covers.
• Straighten the ear canal by gently pulling the mi le of the ear back an up. This ensures mea-
surement of heat from the ear rum. (Please note: Pull outer ear out for newborns - age 1)
• Place the probe firmly into the ear canal. Press the Start button an keep the probe in the ear until
the thermometer beeps to in icate completion of the measurement.
Th 424 Digital Ear Th rmom t r has b n clinically t st d and prov n to b saf and
accurat wh n us d in accordanc with op rating instructions.
5. How to R ad th Display
Display Display M aning D scription
7. Instruccion s d Uso
Important : Ant s d r alizar cada m dición, coloqu un capuchón nu vo intacto n l
xtr mo d m dición.
1. Presione el botón ON/OFF. La pantalla se activa y muestra to os los segmentos
urante 2 segun os.
2. El valor e la última me ición es mostra o en el pantalla automáticamente urante
2 segun os junto con el símbolo “M”.
3. Cuan o el símbolo °C o °F está parpa ean o, suena un bip y el termómetro está
listo realizar la me ición.
4. Hale e la oreja hacia arriba y para atrás para obtener una visión clara el tímpano.
• En los niños menores e un año:
- tire bien para atrás e la oreja.
• En los niños mayores e un año y a ultos:
- hale e la oreja hacia arriba y para atrás.
(Precaución: Muchos pe iatras sugieren usar el termómetro e oi o
en niños mayores e 6 meses.)
5. Mientras hale e la oreja, inserte correctamente el extremo e me ición en el canal au itivo
y presione el botón e inicio (START). Cuan o oiga "bip" pue e soltarlo. Esta es la señal
que avisa e que la me ición ha termina o.
6. Retire el termómetro el con ucto au itivo. La pantalla LCD muestra la temperatura me i a.
7. Después e ca a me ición se ebe esechar el capuchón. Para hacer esta operación, siga, por favor las
instrucciones el aparta o 6: Cómo cambiar el Capuchón Protector.
8. Para asegurar una total exactitu el resulta o e la me ición, espere, por favor, un mínimo e 30
segun os tras haber efectua o e 3 a 5 me iciones segui as.
NOTA:
• Es mejor que los niños pequeños estén recosta o e la o, para que la oreja que e miran o hacia arriba. Para
niños mayores o a ultos, es mejor ponerse etrás e ellos y ligeramente hacia a un la o.
• Tome siempre la temperatura en el mismo oí o, ya que los resulta os pue en variar si se toman en el oí o erecho
o en el izquier o.
• Por favor, espere unos minutos antes e tomar la temperatura si el paciente estaba ormi o.
• Está recomen a o realizar 3 me iciones en el mismo oí o y tomar como resulta o e la me ición el valor más
alto. Esto es importante con:
1) Recién naci os e menos e 100 ias e vi a.
Seen when you turn on the unit.
Shown for 2 secon s.
Self-check ensures
proper functioning.
Press the ON/OFF button to turn
on the unit. All icons will be
shown for 2 secon s.
Last rea ing store in
memory.
The last rea ing will be shown on
the isplay automatically for
3 secon s.
Rea y to measure. The unit beeps once an the
°F or °C icon is flashing.
Measurement complete
°F or °C stops flashing.
Unit beeps once.
When the °F or °C starts to flash
you can take another temperature.
Low battery in ication.
Unit will not take accurate
measurements.
The battery icon will flash to
remin the user to replace the
battery.
NOTA:
• Para evitar contaminaciones cruza as, por favor utilice un capuchón nuevo para ca a me ición.
• Por favor, compruebe que el capuchón haya que a o perfectamente coloca o y ajusta o en el extremo e
me ición el termómetro antes e utilizar el termómetro (por favor, fíjese en las fotografías que aparecen
ebajo). Si el capuchón está roto, eséchelo y coloque uno e nuevo.
(X) No está bien coloca o ()Correcto

6 15
6. How to Apply a N w Prob Cov r 5. Control d l Display y Símbolos
Pantalla LCD Significado d la Pantalla D scripción
6. Cómo cambiar l Capuchón Prot ctor
(1) Place a probe cover onto the
storage case hole with paper si e
upwar s. Align the cover
to fit the hole exactly.
(2) Take the unit, vertically
penetrate the probe into the
center part of the probe cover.
(3) Completely push the probe
into the probe cover hol er hole.
Push firmly.
(4) After feeling a slight click, take
out the probe. The cover shoul
be securely attache .
Pl as Not : Unit is calibrate to work with probe covers. Only use of ADC®probe covers ensures
accuracy an hygienic use.
NOTE
• In or er to avoi sprea ing germs, please use a new probe cover for each measurement.
• Check to ensure probe cover is fitte on firmly before use (please see the iagrams below).
If the probe cover is broken, iscar an apply a new one imme iately.
(X) Incorrect ()Correct
Auto-prueba asegura el
buen funcionamiento.
Presione el botón ON/OFF
para poner la uni a en
marcha. Aparecerán to os los
segmentos urante 2 segun os.
Ultima lectura guar a a
en memoria.
El último valor me i o aparecerá
en el isplay automáticamente
urante 3 segun os.
Listo para me ir. El aparato está prepara o para
Iniciar la me ición, el símbolo
°C o el °F aparecerá estellan o.
Me ición acaba a cuan o
°C ó °F no siguen estellan o.
El valor me i o aparecerá en la
pantalla LCD. Cuan o °C ó °F
empiezan a estellar, el aparato
está listo para la próxima me ición.
In icación e batería baja.
Uni a no ten rá me i as
precisas.
Cuan o se pone en marcha el
Termómetro, el símbolo e la
batería aparece estellan o para
recor ar al usuario que ebe
cambiar la pila.
(1) Coloque un capuchón en el orifico
el soporte protector el termómetro,
con la lámina e papel miran o hacia
arriba. Alinee el capuchón con el
agujero.
(2) Coja el termómetro e intro uzca
verticalmente el extremo e
me ición en el centro el
capuchón.
(3) Presione el termómetro sobre el
orificio el soporte el termómetro.
Empuje firmemente.
(4) Tras notar un suave „click“ levante
el extremo el termómetro con el
capuchón perfectamente ajusta o.

(1)
(2)
(3)
(4)
(5)
(6)
(7) Tiente
Aplica or e
Cobertura Hoyo
14 7
7. Dir ctions for Us
Important: Prior to v ry m asur m nt, fit a n w undamag d Prob Cov r on th
m asuring prob . Failur to do so may r sult in incorr ct t mp ratur r adings.
1. Press the ON/OFF button. The isplay is activate to show all
symbols for 2 secon s. (Fig. 1)
2. The last measurement rea ing will be shown on the isplay
automatically for 3 secon s with the “M“ symbol visible. (Fig. 2)
3. A beep soun is hear . The °C or °F symbol will flash. The
thermometer is rea y to measure. (Fig. 3)
4. Straighten the ear canal by pulling the outer
ear up an back to give a clear view of
the ear rum.
• Chil ren ages 1 year to a ult:
- Pull the ear up an back. (Fig. 4)
(Caution: Many pe iatricians suggest use of ear
thermometers for chil ren ol er than 6 months.)
• For chil ren un er 1 year:
- Pull the ear straight back (Fig. 5)
5. While pulling the outer ear, insert the probe snugly into
the ear canal an press the "START” button. Release it, when
you hear a beep soun . A beep in icates measurement
is complete. (Fig. 6)
6. Remove the thermometer from the ear. The isplay shows the
measure temperature.
7. Replace the probe cover after each measurement. To o this, follow the
instructions in section 6, (How to apply a new probe cover).
8. In or er to assure accurate rea ings, please wait at least 30 secon s after
3-5 continuous measurements.
• La presencia e cera en el con ucto au itivo pue e provocar lecturas inferiores a las normales.
Asegúrese e que el con ucto au itivo está totalmente limpio para asegurar la precisión e la me ición.
• Los componentes el Termómetro Instantáneo e Oí o ADC®son e alta cali a y precisión. Evite
que su termómetro se caiga. Protéjalo frente a fuertes golpes o impactos. Mantenga la uni a en sa
cubierta e almacenaje cuar o no se use.
ADVERTENCIA
• Por favor, guar e los capuchones fuera el alcance e los niños. Estos pue en presentar riesgo e
sofocación.
• El uso e este termómetro no ebe ser utiliza o como sustitución e la consulta con su octor.
• Este termómetro no es resistente al agua. Por favor, NUNCA lo sumerja en líqui os.
• Limpiar el canal el oi o usan o técnicas ina ecua as pue e ser añino para el oi o interior.
Siga las recomen aciones e su mé ico para una limpieza correcta.
• Este pro ucto pue e contener una sustancia química que el esta o e California como causante e
cáncer, efectos e nacimiento y otros años repro uctivos.
3. D scripción d l Producto
(1) Capuchón Desechable
(2) Extremo e me ición
(3) Botón e Inicio (START)
(4) Display LCD
(5) Botón e puesta en marcha ON/OFF
(6) Cubierta protectora e las pilas
(7) Cubierta e almacenaje/aplica or el capuchón
4. Cómo l ADC®424 T rmóm tro Digital d Or ja Mid T mp ratura corporal
El termómetro instantaneo e oi o ADC®424 mi e la energía infrarroja irra ia a por el tímpano y los teji os
circun antes. Esta energía es recogi a por las lentes y converti a en un valor e temperatura. La me ición
obteni a irectamente el tímpano (Membrana timpánica) asegura la me ición e la temperatura el cuerpo
e la manera más exacta posible. Las me iciones realiza as en los teji os
circun antes el con ucto au itivo an lugar a valores más bajos que pue en con ucir al iagnóstico
erróneo e inexistencia e fiebre.
Cómo vitar una m dición incorr cta:
• Primero coloque el capuchón. La uni a está calibra a para lecturas precisas con los cupuchones.
• Tire e la zona me ia e la oreja hacia arriba y para atrás, para que el con ucto au itivo que e en línea recta.
• Intro uzca firmemente el extremo e me ición en el con ucto au itivo, presione el botón e inicio
(Start) y no retire el termómetro el con ucto au itivo hasta que el termómetro emita un bip in icativo e
que la me ición ha finaliza o.
El T rmóm tro Instantan o d oida ADC®424 ha sido clínicam nt validado y s ha
comprobado qu r sulta s guro y pr ciso si mpr y cuando s us d acu rdo con las
instruccion s d st manual.
(Fig. 1)
(Fig. 2)
(Fig. 3)
(Fig. 4) (Fig. 5)
(Fig. 6)

NOTE:
• For an infant, it is best to have the chil laying flat with hea si eways so the ear is facing
towar s you.
For an ol er chil or a ult, it is best to stan behin an slightly to the si e of the patient.
• Always take measurements in the same ear since temperature rea ings may be ifferent from the
right an left ear.
• Wait a few minutes before taking ear temperature after sleeping.
• It is recommen e that three temperatures in the same ear be taken an the highest one taken as
the rea ing. This is very important with:
1) Newborn infants in the first 100 ays.
2) Chil ren un er three years of age with a compromise immune system an for whom the presence
or absence of fever is critical.
3) When the user is learning how to use the IR thermometer for the first time until he/she has
familiarize himself/herself with the instrument an obtains consistent rea ings.
Caution:
If you ar not confid nt with r adings obtain d, confirm t mp ratur with a standard
digital th rmom t r.
8. Changing from Fahr nh it to C lsius and vic -v rsa
The ADC®Digital 424 Thermometer can isplay temperature measurements in either Fahrenheit or
Celsius. To switch the isplay between °F an °C, turn the unit off. Press an hol the Start button
for 5 secon s, the isplay will show ---°C or ---°F. Release the Start button. Push the Start button
again to toggle between °C or °F. Five secon s after completing the change to °C or °F, the unit will
prepare for a measurement, an beep when rea y.
813
R ady Mod or
1. INTRODUCTION
Las felicitaciones en su compra el ADC®A tempTM 424 Tympanic Termómetro Digital. ADC®pro uctos iagnós-
ticos profesionales son los instrumentos e elección en hospitales y las oficinas e mé ico a través el mun o,
on e certeza y serie a son críticas.
Ahora, uste pue e isfrutar también e los beneficios e ADC®irigien o y la cali a en el hogar. Esta característica
instrumento que rico fue iseña o para simplificar el uso y entregar resulta os coherentes y fiables.
Su ADC®A tempTM 424 Termómetro Digital e Tympanic es un ispositivo automático e me ición ofrece las
Ventajas e Múltiples Uso (Me i a e gran varie a ). El A tempTM 424 ofertas una característica e me i a e gran
varie a que recorre e 32 °C a 42,2 °C (89,6 °F a 108,0 °F); el pro ucto pue e ser utiliza o como un termómetro e
oreja para me ir temperatura corporal.
M dición n 1 s gundo
La innova ora tecnología permite la me ición e la temperatura en el oí o en sólo 1 segun o.
Exactitud y Pr cisión
Por las características únicas el sistema e ensamblaje el capuchón, el avanza o sensor y el complejo
proceso e calibración al que ha si o someti o, este termómetro realiza me iciones e gran exactitu y
precisión.
Cómodo y Fácil Uso
• El iseño ergonómico especial hace que el termómetro sea muy fácil e usar.
• El termómetro instantáneo e Oí o pue e ser utiliza o a iario sin ningún tipo e interferencias. La
me ición pue e realizarse incluso mientras el niño uerme.
• El termómetro instantaneo e Oí o resulta fácil e usar en niños.
• El termómetro instantaneo e Oí o no asusta a los niños como lo hace un termómetro rectal y es más
agra able e usar que un termómetro oral.
Aparición n la pantalla d la t mp ratura m morizada
Al poner en marcha el termómetro aparece el valor e la última me ición automáticamente en la pantalla
urante 2 segun os.
Higiénico y S guro
• Sin riesgo e roturas el cristal o e ingestión el mercurio.
• Totalmente seguro para ser usa o en niños.
• Los capuchones esechables hacen que el uso el termómetro IR 1DA1 sea totalmente higiénico.
2. Instruccion s d S guridad Important s
• El termómetro nunca ebe ser usa o con fines iferentes e para los que ha si o iseña o.
• Use siempre el termómetro con un capuchón intacto para ca a me ición con el objeto e prevenir
infecciones. L cturas tomadas sin los capachon s no s rán pr cisas. Sólo los capu chones e
la marca ADC®pue en asegurar la máxima precisión en sus me iciones.
• Nunca sum rja l t rmóm tro instantán o d oído n agua o cualqui r otro líquido
(no s r sist nt al agua). Para su limpi za y d sinf cción, por favor, siga las
instruccion s d l apartado “Limpi za y Cons rvación”.
•Proteja su termómetro y los capuchones e la exposición irecta al sol y guár elos en un lugar seco, sin
polvo, y a una temperatura entre 10 y 40 °C (50 °F – 104 °F).
• No use el termómetro si parece que el extremo e me ición está aña o o si parece haber algún año en
el termómetro. Si realmente está aña o, no intente reparar el termómetro. Por favor, contacte con el
istribui or ADC®e su país.

12 9
9. Error M ssag s
Display Probl m Display M aning Possibl Caus & R m dy
* Caution: High fever requires imme iate me ical attention!
Measure Temperature
too high
Measure temperature higher
than 108.0 °F or 42.2 °C.
*Caution!
Measure Temperature
too low
Measure temperature lower
than 32 °C or 89.6 °F.
See section 7 for proper
technique.
Room temperature too
high
Displays “H” in conjunction
with the “ ” when room
temperature is higher than
40.0 °C or 104.0 °F.
Room temperature too low Display “L” in conjunction with
the “ ” when room
temperature is lower than
10.0 °C or 50.0 °F.
System malfunction Check battery. Possible amage
from ropping or getting wet.
Blank isplay Dea battery?
Check battery polarity.
( + an - )
Low battery in ication Battery shoul be replace
imme iately.
1. La introducción . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .13
2. Las Instruccion s important s d la S guridad . . . . . . . . . . . . . . . .13-14
3. La D scripción d l producto . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14
4. Cómo M dir T mp ratura corporal . . . . . . . . . . . . . . . . . . . . . . . . . . .14
5. Cómo L r la Pr s ntación . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .15
6. Cómo Aplicar una Nu va Cob rtura d Ti nta . . . . . . . . . . . . . . . . .15-16
7. Las dir ccion s para l Uso . . . . . . . . . . . . . . . . . . . . . . . . . . . . .16-17
8. Cambiar d Fahr nh it a C ntígrado . . . . . . . . . . . . . . . . . . . . . . . . .17
9. Los M nsaj s d rror . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .17
10. Limpiar y Almac nami nto . . . . . . . . . . . . . . . . . . . . . . . . . . . . .17-18
11. R mplazar la Bat ría . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .18
12. La garantía . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .18
13. Las Esp cificacion s técnicas . . . . . . . . . . . . . . . . . . . . . . . . . . . . .18
14. El sitio w b . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .19
15. Contact Información . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .19

10 11
10. Cl aning and Storag
Use an alcohol swab or cotton swab moistene with alco-
hol (70% Isopropyl) to clean the thermometer casing an
the measuring probe. Ensure that no liqui enters the inte-
rior of the thermometer.
N v r us abrasiv cl aning ag nts, thinn rs or
gasolin for cl aning and n v r imm rs th
instrum nt in wat r or oth r cl aning liquids.
Take care not to scratch the surface of the lens or isplay.
Remove the battery from the instrument if it is not require
for exten e perio s of time in or er to avoi amage to
the thermometer resulting from a leaking battery.
Pl as Not : Ear wax or dirt on th l ns will r sult in an
inaccurat r ading. Cl an th l ns as indicat d abov .
11. R placing th Batt ry
The ADC®Digital 424 Ear Thermometer is
supplie with one lithium battery, type CR2032.
Replace with a new CR2032 battery when the
low battery symbol appears on the isplay.
Use a screw river to loosen the screw from
battery cover. Remove the battery cover an
replace battery. Do not overtighten screw when
replacing the cover.
12. Warranty
This pro uct was manufacture with the utmost care accor ing to international quality stan ar s. This
pro uct is warrantee for 2 years against manufacturing efects. Claims beyon this, inclu ing claims
for amages, are exclu e .
• The warranty covers the instrument an the batteries. The packaging is not inclu e .
• Opening or altering the instrument invali ates the warranty.
• The warranty oes not cover amage cause by improper han ling, ischarge batteries,
acci ents or non-compliance with the operating instructions. Contact ADC®Customer
Service Dept. for more information.
13. T chnical Sp cifications
Typ : Ear Thermometer 424
M asuring Rang : 32.0 F~212.0 F (0 C~100 C)
Accuracy: Laboratory: : ±0.4 F: 89.6~108.0 F
(±0.2 C: 32.0~42.2 C)
±2 F: 32.0~89.5 F, 108.1~212.0 F
(±1 C: 0~31.9 C, 42.3~100.0 C)
Display: Liquid Crystal Display with 0.1 C (0.1 F) resolution.
B ps Wh n: a. The unit is turned ON and ready for measurement:
(1 short “beep“ sound)
b. Measurement complete: 1 long beep sound.
c. System error, malfunction, or “Hi””Low” measurement.
3 short “beep“ sounds.
M mory: Automatically displays the last measured temperature.
Illuminat d Display: a. The display will be lighted for 4 seconds when the
(Back-Lite) unit is turned ON.
b. The display will be lighted again for 5 seconds when
the measurement has been completed.
Op rating T mp ratur : 10 C to 40 C (50.0 F to 104 F)
Storag /Transport T mp ratur : -25 C to +55 C (-13 F to 131 F)
Automatic Switch-off: Approx. 1 minute after last measurement has
been taken.
Batt ry: 1 each CR2032 BATTERY - at least 1000 measurements
Dim nsions: 5.5” x 1” x .79” (141mm L x 26 mm W x 20 mm H)
W ight: 1.59oz. (45g) (with battery), 1.41oz. (40g) (w/o battery)
D vic Standard: Device corresponds to the requirements of the standard
for infrared thermometers
ASTM E 1965, EN 12470-5, ICE 60601-1, IEC 60601-1-2
Electromagnetic Compatibility: Device fulfills the stipulations of the standard
IEC 60601-1-2
Push against contact prings
for re-insertion
Springs
Push to
Insert
Pry Out